This solicitation requests proposals to upgrade, renovate, and relocate Sterile Processing Services (SPS) at the Michael E. DeBakey Veterans Affairs Medical Center in Houston, Texas. The project involves renovating approximately 16,500 square feet of space for SPS, logistics, and laundry in the basement of Building 100. All construction work must comply with the latest JCAHO, NFPA, and life safety standards and be performed without disrupting medical center operations. Interested offerors should review the attached request for proposal, statement of work, specifications, and drawings for full details on the scope of work. Proposals are due by the date specified in the solicitation. The selected contractor will work with the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 16 on this project.
View the file
Other files
Show all 32
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
VISION® 1321/1327 CART AND
UTENSIL WASHER/DISINFECTOR
(Typical - details may vary.)
APPLICATION
The Vision 1321/1327 Cart and Utensil Washer/Disinfector is a high-capacity mechanical washer intended for use in the efficient cleaning, low-level disinfecting and drying of case carts, containers, utensils, beds and other miscellaneous reusable items used in the care of patients. In addition, soiled simple hard-surfaced rigid surgical instruments (e.g., forceps and clamps) are cleaned and intermediate level disinfected with a required option.
DESCRIPTION
The Vision 1321/1327 Cart and Utensil Washer/Disinfector is a mechanical unit equipped with a microprocessor control system. The unit is designed with eight factory-set adjustable cycles: CART STANDARD, CART LOW ECO, CONTAINERS, UTENSILS, BEDS, QUICK, ALUM SAFE and INSTRUMENTS.
Eleven additional cycles are available for customized programming to meet specific operating requirements.
Cycles are programmed with a drying phase, and with minimal wash and thermal rinse duration. Two exterior mounted fluorescent lights are included to illuminate the wash chamber.
The Vision 1321/1327 Cart and Utensil Washer/Disinfector is available in a double-door configuration. The unit can be built to seismic design.
Size (W x H x L)
1321 Version
• Maximum overall dimensions:
124 x 112.75 x 106" (3150 x 2864 x 2692 mm)
• Effective chamber load capacity:
40 x 80 x 82" (1016 x 2032 x 2083 mm)
1327 Version
• Overall dimensions:
124 x 112.75 x 130" (3150 x 2864 x 3302 mm)
• Effective chamber load capacity:
40 x 80 x 106" (1016 x 2032 x 2692 mm)
VOLTAGE
❑ 208 V, 3-Phase, 60 Hz ❑ 380-415 V, 3-Phase, 50 Hz ❑ 480 V, 3-Phase, 60 Hz
CONFIGURATION
❑ 1321 Double Power Door (Pass-Through) ❑ 1327 Double Power Door (Pass-Through)
ACCESSORIES
❑ Common Service Area ❑ Nonservice Side Pane ❑ Service Side Access P ❑ Air Compressor (avail
❑ 208 V ❏ 230 V ❏ ❑ Floor Ramps for Floor ❑ HEPA Air Filter ❑ Seismic Tie-Down Kit ❑ Bar Code Scanner ❑ Vision Cart Washer/D ❑ Vision Cart Washer/D
The Selections Checked Bel
STANDARDS
This washer/disinfector meets applicable requirements of the following standards:
• Governing Directive for the affixing of the CE Mark:
Medical Devices Directive 2007/47/EC.
• As certified by UL:
» Underwriters Laboratories (UL) Standard UL 61010-1, 2005.
» Canadian Standards Association (CSA) CAN/CSA- C22.2, No. 61010-1, Second Edition.
• International Standards:
» EN/IEC 61010-1, 2nd Edition; EN/IEC 61010-2-040, 2005; EN/IEC 61326-1, 2005.
» EN ISO 15883-1:2006, ANSI/AAMI ST 15583-1 and CSA Z15883-1:2009 - Washer-Disinfectors - General Requirements, Definitions and Tests
» EN ISO 15883-2:2006*, ANSI/AAMI ST 15583-2 (draft)* Requirements and Tests for Washer-Disinfectors Employing Thermal Disinfection for Surgical Instruments, Aesthesia Equipment, Hollowware, Utensils, Glassware, etc.
* When Instrument Washing Option is Present.
SD940 (02/01/12)
Item ________________________ Location(s)___________________
Junction Panels ls anels able only North America) 460 V Mounted Unit (two) isinfector Container Rack isinfector Utensil Rack
OPTIONS
❑ Pure Water Line ❑ Third Reservoir ❑ Instrument Cycle
REMOTE MONITORING
❑ ProConnect® Response Center (Remote
Monitoring, Priority Technical Support, Customer Care Center Access, Equipment Performance Reports) ow Apply To This Equipment ckeen Highlight ckeen Highlight ckeen Highlight
» EN ISO 15883-6:2011, Requirements and Tests for General Purpose Washer-Disinfectors Employing Thermal Disinfection for Non-Invasive Medical Devices, Washbowels, Utensils, Transit Containers, etc.
FEATURES
Sumpless Solution Delivery System. The washer/disinfector cabinet base is 6-3/4" (171 mm) deep.
Solutions are heated by an in-line stainless-steel heat exchanger. A thermal rinse tank includes a steam coil to maintain water temperature for thermal rinse phase.
Spraying System. The washer/disinfector includes two horizontally mounted spray headers, one on each side of the wash chamber to optimize load coverage and cycle time. Each spray header contains 17 jets directed to provide maximum full load coverage. Header height is adjusted to concentrate spray action to lower portion of chamber when processing carts 44" (1118 mm) high or lower. Control monitors spray system travel to stop movement if obstruction is detected or alert operator if full movement is not accomplished.
Horizontal Sliding Doors. Each chamber is provided with two automatic, powered, side-sliding doors made of two 1/4" (6 mm) tempered glass panes to minimize heat transfer and allow easy visual monitoring of the cleaning process. Units are equipped with sensors to detect obstructions and prevent unit operation if doors are not fully closed. Doors are dynamically pressed against a silicone gasket ensuring complete air and water seal of the wash chamber.
Interior Light. Two exterior mounted 18W fluorescent lamps are provided to illuminate the wash chamber. The life expectancy of the lamps is 10,000 hours.
Integral Self-Priming Automatic Chemical Dispenser.
Three peristaltic pumps are included in the standard washer/disinfector with up to two additional chemical injection pumps provided if Instrument Washing option and Three Reservoir option are installed. Standard pumps: one for neutral or alkaline detergent, one for rinse aid and one for the descaler agent. The pumps are located in a separate room, positioned near the detergent containers and connected to washer/disinfector with up to 100’ (30 m) of piping. Each tube is color coded to facilitate tracking through the facility.
Pumps allow the use of the Prolystica® Ultra Concentrate chemicals. Each ultra concentrated product is 10 times the concentration of a traditional product, therefore Only 1/10th of the chemical is injected to properly process the cycles.
The peristaltic pumps automatically add a selected quantity of detergent:
• Ultra Concentrates: 1/20 or 1/10 oz/gal (0.4 to 0.8 mL/L).
• Regular Chemicals: 1/8 to 2 oz/gal (1 to 16 mL/L)
• Rinse Aid: 1/50 oz/gal (0.16 mL/L).
A low-level sensor is included to indicate when the detergent level in the container is low or when insufficient chemical is available for the next cycle.
The control monitors the volume of chemicals injected and indicates if this parameter meets specified criteria during all specific phases (only for instrument cycle).
Wash Chamber. The chamber is constructed of 14 and 16 gauge, #304 stainless steel (No. 4 finish), argon-welded and polished. Base is made of #304 L stainless steel. Chamber flooring consists of two removable stainless-steel panels designed with gratings running lengthwise in the chamber allowing for quiet loading and unloading.
A removable debris screen is located in the bottom of the chamber. The debris screen prevents large debris from entering the piping system and pump. The screen is provided with a handle and is easily removed for cleaning under running water.
Other Components. All components of the wash/rinse system, including screens, spray headers, piping, and booster in-line heat exchanger are constructed of #304 stainless steel. High-pressure recirculating pump is made of #316 L stainless steel.
Ball valves are constructed of Teflon® and #316L stainless steel.1
Unit frame, mobile mechanical core and all fasteners are constructed of #304 stainless steel. Aluminum-sheathed rigid fiberglass insulation, 1" (25 mm) thick covering the top and sides of the chamber exterior, reduces heat loss and noise level to the work area.
Treatment Staging Tanks. The tanks are equipped with an automatic solution level control, automatic hot water fill and safety overflowing piping. Tanks are made of #304 L stainless steel and are fully insulated with 1" (25 mm) thick aluminum sheathed fiberglass to prevent heat loss and burn hazard. The bottom of the tanks are sloped toward the water outlet for optimum drainability. Each solution tank includes an internal baffle that deflects solutions to the tank walls to assure self-cleaning of the tank during the recirculation process. Capacity is 25 U.S. gal (94.62 L).
High Pressure Pump. Open impeller pump is powered by dual-speed motor permitting two ranges of flow rate/pressure.
Pump impeller, shaft and casing are of #316 L stainless-steel construction. The pump motor is a TEFC (totally enclosed fan cooled) class H motor, electro-polished, magnetic starter, overload protection and sealed bearings (requiring no lubrication).
Nonrecirculated, Vented Drying System. The vented drying system effectively dries the processed load at the completion of each cycle. Fresh, heated filtered air is blown at high velocity through all four corners of the wash chamber and to the load.
Dry air is then evacuated through the chamber vent opening.
Fresh air can be HEPA (High Efficiency Particulate Air) filtered as an accessory.
Automatic Floor Tilting System. System slopes the processed load at the start of the drying phase to properly drain flat surfaces of carts. The floor is automatically returned to its level position at completion of the cycle for smooth loading and unloading.
Integral Exhaust Fan. The fan assists the building ventilation system when evacuating vapor from the wash chamber. Fan impeller, casing and motor shaft are made of stainless steel.
1.Teflon® is a registered trademark of E.E. Dupont de Nemours and Company.
Heating System. Solutions are heated by an instantaneous, in-line, stainless-steel heat exchanger designed to reduce steam consumption and cycle time. A steam coil also exists in the thermal rinse tank to preheat and maintain water temperature to reduce thermal rinse phase readiness time.
Drain Discharge Cool Down. Unit is provided with a cold water connection for use with cold water selection and effluents cool down if sump or reservoir water temperature is higher than 140°F (60°C) while being discharged to building drain system.
Process Data Validation System. The washer/disinfector includes a Process Data Validation System (PDVS). This system documents and measures the following parameters:
• Sump temperature
• Volume of detergent injected during Instrument Cycle
Control alarms user if this independently recorded data falls outside passing criteria. The PDVS increases facility confidence level that the cycle has been successfully completed.
ProConnect Response Center. Minimize response time and minimize unscheduled downtime on your equipment. Secure, internet-based, 24/7 remote monitoring enables both Predictive Maintenance as well as instant alert to STERIS when there is an equipment alarm. Also included are priority technical support, online parts ordering, equipment performance dashboards, and scheduling service at eservice.steris.com.
CONTROL SYSTEM
The control system consists of an Allen-Bradley SoftLogix™ PC-based control system1. Operator interfaces are ergonomically located, easily accessible and viewable on both ends of the unit.Other control system features are as follows:
• Control system monitors and controls washer operations and functions. Cycle progresses automatically through the designated phases as programmed. Control system also stores all cycle data as protection against power disruption.
• Operator Interface is an 8 x 7" (203 x 179 mm) touch-sensitive color graphic display. Display permits operator to monitor current cycle status, including current chamber temperature and remaining phase time. Identical information displayed on both ends of the unit.
• Integral thermal printer with automatic paper take-up provides an easy-to-read record of all pertinent cycle data. Generated printout includes date, treatment type, cycle starting time and key cycle transition points. All cycle deviations are indicated by visual and audible means, recorded by the printer and need acknowledgement by operator.
• Security access code requires entry of a four-digit access code to change cycles, cycle values and to enter service mode.
• Service mode is accessible through main control panel for service and maintenance purposes.
• Preprogrammed parameters for each cycle. If operator selects an out-of-range setting when modifying the cycle values, the control system alerts operator with a reference message and halts further operation until the correct value is entered.
• USB Port is supplied so cycle data is available in PDF format and accessible through Service Mode.
1 SoftLogix™ is a trademark of Allen-Bradley, a Rockwell Automation Company.
• Ethernet port is available for remote monitoring and troubleshooting.
• Three Operator Modes are available. Supervisor, Service and Cycle. The first two modes are password protected.
Cycle mode is always available.
CYCLE DESCRIPTION
ADVISORY NOTE: This washer/disinfector is specifically designed to only process goods as outlined in this tech data.
If there is any doubt about the use of a specific material or product, contact the manufacturer of the product for recommended washing techniques.
STERIS does not intend, recommend, or represent in any way that this Vision 1321/1327 Cart and Utensil Washer/Disinfector be used for the terminal disinfection or sterilization of any regulated medical device. The Vision 1321/1327 Cart and Utensil Washer/Disinfectors are intended only to perform an initial step in the processing of soiled, reusable items used in the care of patients. If medical devices will be contacting blood or compromised tissues, such devices must be terminally processed in accordance with good hospital practices (GHP) before each use in human patients.
Items to be cleaned are placed on the appropriate accessory cart and/or rack and are positioned in front of the washer. When the washer is ready, the operator opens the doors and pushes the cart/rack into the chamber. Operator closes the door and initiates the cycle. Unit proceeds through treatment schedule and automatically opens the unloading door at cycle completion. A display message and audible alarm indicate that the load is ready for removal. Clean side operator removes clean/disinfected items.
Each cycle program is operator adjustable to meet specific processing needs. Cycle programming is protected by a security access code set by the supervisor. The standard control system features eight factory pre-programmed cycles
(CART STANDARD, CART LOW ECO, CONTAINERS,
UTENSILS, BEDS, QUICK, INSTRUMENTS and ALUM SAFE) and a special SELF DECONTAMINATION cycle for routine maintenance of the unit. Control is also equipped with open cycles that are programmable on-site at the user’s discretion.
Each cycle program is constructed of phases. The following phase descriptions are from the featured Cart and Utensil cycles developed and validated to provide a total cleaning solution by joining advanced washing technology with Prolystica Ultra Concentrate cleaning chemistries:
Wash phase - solution from Reservoir (Tank) 2 or 3 (if the third reservoir option is present) is recirculated through the spray system for the selected time period. Solution is heated to 135°F (57°C) in the second portion of the phase. After a programmable number of cycles, solutions are automatically drained and renewed.
Thermal Rinse phase - pre-heated, hot or pure water from Reservoir (Tank) 1 is sprayed over the load and recirculated for the selected time period to rinse and disinfect the load.
Solution is heated to 176°F (80°C). After a programmable number of cycles, solutions are automatically drained and renewed. Rinse Aid is automatically added during the reservoir (tank) filling.
HEPA-Filtered Air Drying - The drying phase consists of blowing heated nonrecirculated air on the load, and evacuating it through the vent connection. Selected time for this nonrecirculated phase must be between 02:00 and 15:00.
SAFETY FEATURES
Safety Door Switch. A microswitch prevents a cycle from starting if the doors are not fully closed, and also stops the unit if doors are opened during a cycle. Doors must be closed to continue operation. Also, if an obstruction is present when a door attempts to close, control senses obstruction and prevents door from closing.
Emergency Stop Pushbuttons. The washer/disinfector is equipped with two external Emergency Stop pushbuttons that automatically stop operation of the unit. Once pushed, door locking mechanism disengages allowing the door to be easily opened from inside or outside.
Door Interlock. The safety interlock mechanism prevents both doors from being opened simultaneously, preventing cross-contamination. The clean side/unload door cannot be opened until the cycle has been successfully completed.
Labeling. The washer/disinfector is labeled with warning and caution pictograms to warn the operators and service technicians of precautions to be taken.
Emergency Stop Cables. Located on each side of the interior wash chamber, instantly stop washer/disinfector operation if pulled. Once pulled, door locking mechanism disengages allowing the door to be easily opened from inside or outside.
ON/OFF Switch. Power switch is located on load side of the unit and is used to shut off control power and drain sump and reservoirs (tanks).
OPTIONAL FEATURES
Instrument Cycle Package allows the cleaning and intermediate level disinfection of soiled hard-surfaced rigid surgical instruments (e.g., forceps and clamps). Option comes with an Enzyme injection line, a coupling system for manifolded rack, a Vision instrument washing manifolded rack, detergent flow meters and an Instrument Cycle.
Third Reservoir Option allows an additional detergent to be introduced to wash cycle. Tank is constructed of #304 L stainless-steel, insulated with aluminum sheathed rigid fiberglass to prevent heat loss and burn hazard and with a capacity of 25 US gal (95 L).
Pure Water Line, constructed of 316 stainless-steel, delivers purified water to rinse phase.
ACCESSORIES
High Efficiency Particulate Air (HEPA) filter is located on the mechanical core of washer/disinfector. Air entering the chamber passes through the HEPA filter to prevent any contamination of the load.
Service Side Enclosure Panels and supports are provided to enclose the mechanical core.
Nonservice Side Enclosure Panels are provided to cover insulation on the nonservice side of unit cabinet.
Remote Air Compressor, complete with tank and pressure switch, is available in either 208 V or 460 V. Oilless air compressor operates at 59 dBa. Wiring at installation not provided by STERIS. Also required for opertation, a 120 V, 60 Hz outlet for auto drain valve.
Floor Ramps, 4' (1.2 m) long, are provided with guards on each side to help the operator guide the load into the washer.
Slope of the ramp not to exceed 6.5° to enable easy loading of carts in the wash chamber.
Seismic Tie-Down Kit includes hardware for properly securing the washer to the building floor. The washer is designed to comply with Seismic Zone 3 and 4 requirements.
Bar Code Scanner is provided to automatically identify and initiate the appropriate cycle by means of a bar code tag attached to load.
PREVENTIVE MAINTENANCE
A global network of skilled service specialists can provide periodic inspections and adjustments to help ensure low-cost peak performance. STERIS representatives can provide information regarding annual maintenance programs.
CHEMICAL ADDITIVES SPECIFICATIONS
Follow detergent label recommendations for the concentration of chemical to use.
To achieve maximum cleaning efficiency, STERIS recommends the following chemicals:
• Prolystica® Ultra Concentrate Enzymatic Presoak and Cleaner (Instrument Cycle).
• Prolystica® Ultra Concentrate Alkaline Detergent.
• Prolystica® Ultra Concentrate Neutral Detergent.
IMPORTANT: STERIS does not promote, recommend, or endorse the use of any other type of chemical additives in the processing of articles in the Vision 1321/1327 Cart and Utensil Washer/Disinfector, such as high alkaline detergents (pH >12), alcohol rinses, and liquid disinfectants, including hypochloric acid (bleach).
NOTES
1. Customers must ensure the washer/disinfector stands on a level non combustible floor.
2. STERIS recommends that shutoff valves and vacuum breakers (not provided by STERIS) be installed on service lines, and that fused disconnect switches (with lockout in OFF position; not provided by STERIS) be installed in electric supply lines near the equipment.
3. Pipe sizes shown indicate terminal outlets only. Building service lines provided (not provided by STERIS), must supply the specified pressures and flow rates.
4. For all ventilation ducting from the washer, STERIS recommends installation of a dedicated corrosion-proof, water-tight duct, rated to an operating temperature of 210°F (99°C) or more, to the exterior of the building, sloped toward the washer with condensate drain connection.
5. Refer to Equipment Drawing for specific installation details.
6. STERIS recommends a well-lighted service area (if applicable), along with the provision of a convenience outlet for maintenance.
7. Always follow local electrical codes and safety-related work practices for wiring.
UTILITY REQUIREMENTS
Important: Refer to Equipment Drawing (Vision 1321: 920-517- 287; Vision 1327: 920-513-916) for installation details and specifications.
Hot Water
1" NPT
Cold Water
1" NPT
Steam
1-1/2" NPT
Air
1/2" NPT
Ventilation 6" (152 mm) O.D.
Drain Recommended minimum 9" (229 mm) floor sink with minimum 4" (101 mm) drain outlet.
Pure Water
1" NPT
Condensate Return
1" NPT
Electricity » 200-208 V, 60 Hz, 3-Phase, 25.5-27.5 Amps
» 460-480 V, 60 Hz, 3-Phase, 13.5 Amps
» 380-415 V, 50 Hz, 3-Phase, 14.5 Amps
Telecommunications Requirements for ProConnect Response Center
• An active wired or wireless TCP/IP network, 10/100BaseT Ethernet connection at each piece of connected equipment, Internet access and an IP address on the facility network.
• 5 GB of available hard drive space to run the service agent. Can be installed on:
»Dedicated PC running Windows XP with 2.8GHz processor, 512MB of RAM
»Virtual Machine
»Server
• Local STERIS login at the PC with a username of STERIS and the password should be ProConnect (STERIS Customer Number).
Ethernet cable to connect each piece of STERIS equipment and the dedicated PC to the facility network.
The base language of this document is ENGLISH. Any translations must be made from the base language document.
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH
APPLICABLE LOCAL AND NATIONAL CODES AND
REGULATIONS.
ENGINEERING DATA
Shipping Weight (max):
Chamber
Mechanical Core
1920 lbs (871 kg)
1400 lbs (635 kg)
Operating Weight:
Cabinet, Water, Two Heavy Case Carts
Without Cabinet, With Water and Two Heavy Case Carts
7483 lbs (3394 kg)
6367 lbs (2888 kg)
Hot Water Consumption per cycle: 14 U.S. gal
(53 L)
Steam Consumption per cycle (with hot tap water heated at 140°F-60°C): 31 lbs
(14 kg)
Noise Level1:
1 Calculated as described in ISO-3746 standard
65 dBa
Heat Loss: 13,800 BTU/hr (4039 W)
NOTES
Recommended Air Compressor
1. Rotary scroll air compressor must be located in a clean, well lit, and ventilated area.
2. Never install the compressor where the ambient temperature is higher than 105°F (40°C), or where humidity is high. Clearance must allow for safe, effective inspection and maintenance. Minimum clearances required: above, 24" (610 mm); drive belt side, 12" (300 mm); and, other sides 20" (510 mm).
3. Never use any piping smaller than the compressor connection.
UTILITY REQUIREMENTS
Recommended Air Compressor
Electrical – Compressor Motor
208 or 460 Volt, 60 Hz, 3-Phase
Electrical – Auto Drain Valve
120 Volt, 60 Hz
ENGINEERING DATA - RECOMMENDED AIR COMPRESSOR WITH AUTOMATIC TANK DRAIN
CFM
Open Flow Weight Comp.
Stages
TANK MOTOR
Lubrication Size Capacity Max. Press Operating Speed
7.7 240 lbs (109 kg)
1 Oilless See Below 30 gal (114 L)
100 psig (689 kPa)
60 Hz (1725 rpm)
Recommended Air CompressorDimensions are typical -drawing is not to scale.
Auto drain 18"
(457 mm) 1/2" NPTF Discharge
Starter
Motor
Scroll Air End 40"
(1016 mm)
23" (584 mm)
34" (863 mm)
12" (304 mm)
16" (406 mm)Not F or In stallation
Refer to the Following Equipment Drawings for Installation Details
Equipment Drawing Number Equipment Drawing Title
920-517-287 Vision 1321 Cart and Utensil Washer/Disinfector
920-513-916 Vision 1327 Cart and Utensil Washer/Disinfector
Floor-Mounted UnitPit-Mounted Unit
FRONT VIEW
33 1/2 (851)
Access Ramp
TOP VIEW
FRONT VIEW
TOP VIEW
Loading Clearance
Height
Se rv ic e
Ar ea
Loading Clearance
Width
Service Door
98 1/2 (2502)
3 3/4 (95)
(1041)
80 (2032)
11 7/8 (301)
60 3/4 (1543)
(711)
119 3/4 (3042) 87 (2210)
40 (1016)12 3/8 (314)
84 1/2 (2146)
(813)
109 (2769)
120 (3048)
98 1/2 (2502)
Se rv ic e
Ar ea 80 (2032)
33 1/2 (851)
Service Door
3 3/4 (95)
112 3/4 (2864)
Loading Clearance
Height
80 (2032)
12 3/8 (314)
Loading Clearance
Width
40 (1016)
84 1/2 (2146) 120 (3048)
(813)
102 (2590)
Vision 1321 Cart and Utensil Washer/Disinfector
Not F or In st al la tio n
Dimensions are typical – drawing is not to scale.
Dimensions are in inches (mm)
For further information, contact:
STERIS Corporation 5960 Heisley Road Mentor, OH 44060-1834 • USA 440-354-2600 • 800-548-4873 www.steris.com
SD940 ©2012, STERIS Corporation. All rights reserved. (02/01/12)
This document is intended for the exclusive use of STERIS Customers, including architects or designers. Reproduction in whole or in part by any party other than a Customer is prohibited.
Floor-Mounted UnitPit-Mounted Unit
FRONT VIEW
33 1/2 (851)
Access Ramp
TOP VIEW
FRONT VIEW
TOP VIEW
Loading Clearance
Height Se rv ic e Ar ea
Loading Clearance
Width
Service Door
122 1/2 (3112)
3 3/4 (95)
(1041)
80 (2032)
11 7/8 (301)
60 3/4 (1543)
(711)
119 3/4 (3042) 87 (2210)
40 (1016)12 3/8 (314)
84 1/2 (2146)
(813)
109 (2769)
120 (3048)
122 1/2 (3112) Se rv ic e Ar ea 80 (2032)
33 1/2 (851)
Service Door
3 3/4 (95)
112 3/4 (2864)
Loading Clearance
Height
80 (2032)
12 3/8 (314)
Loading Clearance
Width
40 (1016)
84 1/2 (2146) 120 (3048)
(813)
102 (2768)
Vision 1327 Cart and Utensil Washer/Disinfector
Not F or In st al la tio n
Dimensions are in inches (mm)Dimensions are typical – drawing is not to scale.
| APPLICATION |
| DESCRIPTION |
| STANDARDS |
| FEATURES |
| Control System |
| CYCLE DESCRIPTION |
| SAFETY FEATURES |
| OPTIONAL FEATURES |
| ACCESSORIES |
| PREVENTIVE MAINTENANCE |
| CHEMICAL ADDITIVES SPECIFICATIONS |
| NOTES |
| UTILITY REQUIREMENTS |
| NOTES |
| UTILITY REQUIREMENTS |
| sd940-1: Off |
| sd940-2: Off |
| sd940-9: Off |
| sd940-9c: Off |
| sd940-9a: Off |
| sd940-3: Off |
| sd940-10: Off |
| sd940-11: Off |
| sd940-12: Off |
| sd940-13: Off |
| sd940-14: Off |
| sd940-15: Off |
| sd940-9b: Off |
| sd940-6: Off |
| sd940-7: Off |
| sd940-8: Off |
| sd940-4: Off |
| sd940-5Every exit is an entrance to somewhere else: |
| When you have obstacles and have a door close on you, another door opens: |
| Don't cry over the missed opportunity, look forward to the new opportunity and move on with your life: Off |
| sd940-16: Off |
| sd940-17: Off |
| sd940-18: Off |
| sd940-19: Off |
| SD940Text-1: |
| SD940Text-3: |
| SD940Text-2: |
File details come from the government source that posted it. Updated .