Equipment Drawing - Vision 1327 - STEAM PIT.pdf
PDF 556 KB Posted
- Attached to
- Y1DA--Project No. 580-17-121 Upgrade/Renovate/Relocate Sterile Processing Services (SPS)
- Solicitation number
- 36C25621R0014
About this file
This solicitation is for upgrading, renovating and relocating sterile processing services (SPS) at the Michael E. DeBakey Veterans Affairs Medical Center in Houston, Texas. The project will renovate approximately 16,500 square feet of SPS, logistics and laundry in the basement of Building 100. All construction work must adhere to the latest JCAHO, NFPA and life safety requirements and be performed without affecting ongoing medical center operations. The solicitation includes an RFP, statement of work, specifications and drawings providing additional details on the project's scope. The soliciting agency is the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 16. Responses are due based on the standard timeline outlined in the full solicitation document.
View the file
Other files
Show all 32
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
80[2032]
LOADING
CLEARANCE
HEIGHT
112-3/4[2864]
CLEARANCE
FOR RELIEF
VALVE
OPERATION
102[2591]
UNIT
HEIGHT
110-1/2[2807]
120[3048]
UNIT WIDTH
12-3/8[314] 40[1016]
LOADING
CLEARANCE
WIDTH
32[813]
13-1/8 [333]
84-1/2[2146]
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
TYPICAL INSTALLATION (1/2)
NOTES:
(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].
(B) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.
(C) ALLOW SUFFICIENT SPACE IN FRONT OF LOADING AREA TO LOAD / UNLOAD CARTS
ACCORDING TO CUSTOMER CARTS AND DIMENSION AND STERIS WASHING ACCESSORIES:
CONTAINER RACK 63 1/2 [1612] LG
UTENSIL RACK 65 1/2 [1663] LG
INSTRUMENT RACK 65 [1651] LG
(D) WALLS COULD BE BUILDING WALL OR SERVICE AND NON-SERVICE S/S PANELS ACCESSORIES
(E) FOR STANDARD THROUGH-WALL UNIT, CLEARANCE BETWEEN CLOSED DOOR OF ELECTRIC
CABINET AND FACILITY WALL, OR ANY OPPOSING FIXTURE, MUST MEET LOCAL CODES.
(F) SERVICE AREA ACCESS DOOR IS INSTALLED ON LOAD SIDE OF THE WASHER.
LOAD SIDE CAN BE EITHER ON LEFT OR RIGHT SIDE OF THE MECHANICAL CORE.
(G) IF RECOMMENDED FLOOR CLEARANCE FOR INSTALLATION IS NOT AVAILABLE
ADDITIONAL MANPOWER AND EQUIPMENT WILL BE REQUIRED.
FRONT VIEW
Page 1 of 17 920513916.E
Released
120[3048]
SERVICE CLEARANCE
ABOVE FINISHED
FLOOR
98 [2489] AT 104 [2642]
CLEARANCE MIN.
5-1/4[133]
2[51]
105[2667]
WALL OPENING
3-3/4[95]
122-1/2[3112]
UNIT LENGTH
124[3150]
WALL
OPENING
112[2845]
68[1727]
20[508]
80[2032]
33-1/2[851]
126[3200] 126[3200]
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
TOP VIEW
SERVICE
AREA
FLOOR CLEARANCE
FOR INSTALLATION
SEE PAGE 1 *(NOTE G) SEE PAGE 1 *(NOTE G)
FOR INSTALLATION
FLOOR CLEARANCE
SERVICE DOOR
SEE PAGE 1 *(NOTE F)
SEE PAGE 1 *(NOTE D)
SEE PAGE 1 *(NOTE D)
SIDE VIEW
SEE PAGE 1 *(NOTE E)
CEILING DECK (IF PRESENT)
TYPICAL INSTALLATION (2/2)
Page 2 of 17 920513916.E
85-1/2[2172]
MIN
102[2591]
UNIT HEIGHT
213[5410]
UNIT WIDTH
[76] 3[76]
78[1981]
217[5512]
WALL OPENING
[76]
78[1981]
3[76]
61[1549]
MIN
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SIDE BY SIDE
TYPICAL INSTALLATION
NOTES:
(A) IMPORTANT: FOR INVERTED WASHER, REFERENCE POINT "A" FOR UTILITIES POSITION IS REVERSED AS SHOWN.
(B) FOR TWO UNITS WITH SERVICE AREA BETWEEN, CLEARANCE BETWEEN CLOSED DOORS OF ELECTRIC CABINETS
AND NEAREST PART OF OTHER WASHER, OR ANY FIXTURE BETWEEN, MUST MEET LOCAL CODES.
MAIN WASHER INVERTED WASHER
SUMP
SUMP
COMMON
SERVICE
AREA
NON-SERVICE
SIDE WALL
NON-SERVICE
SIDE WALL
POINT A
SEE NOTE
*(NOTE A)
POINT A
SEE NOTE
*(NOTE A)
Page 3 of 17 920513916.E
73-1/2 [1867]
[3150]
112[2845] 3[76]
118[2997]
2X 12 [305]
[1981]
4X 16 [406]
2X 21-1/4 [540]
3[76]
*(NOTE E)
[559]
4-1/8 [105]
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
STANDARD CONSTRUCTION
(WITHOUT SEISMIC ANCHORAGE)
1/2
FLOOR SLOPE
TOWARD FLOOR
DRAIN DEGREE: 1
FLOOR DRAIN
LIMIT OF PIT
*(NOTE D)
SERVICE
AREA
NOTES:
(A) ALL MATERIALS (INCLUDING CONCRETE, LEVELING PADS, STAINLESS STEEL ANGLES, FLOOR SINK...) ARE
BY OTHERS THAN STERIS. SLOPE PIT FLOOR TO DRAIN. STAINLESS STEEL ANGLES ARE
RECOMMENDED BY STERIS BUT ARE NOT NECESSARY FOR WASHER INSTALLATION.
(B) REFER TO PAGE 13 FOR FLOOR DRAIN FLOW RATE CAPACITY REQUIREMENTS.
(C) STERIS RECOMMENDS A MAXIMUM 8 in [203 mm] ROUND OR SQUARE FLOOR SINK BY 6 in [152 mm] DEEP.
(D) THIS CART WASHER CAN FIT INTO AN EXISTING 76 in [1930 mm] PIT, BUT WITH RESTRICTED ACCESS TO
COMPONENTS LOCATED IN THE PIT. THE DRAIN LOCATION MAY NEED TO BE RELOCATED.
(E) FOR EXISTING 76 in [1930 mm] PIT WASHER, THE CLEARANCE WILL BE 2 in [51 mm] INSTEAD OF 3 in [76 mm].
SUMP
4X CORNER
LEVELING PAD
A
A
BB
Page 4 of 17 920513916.E
73-1/2[1867]
FLOOR DRAIN
4 [102] MIN
FLOOR SINK
8 [203] MAX.
2X 12 [305] 2X 21-1/4 [540]
4X 16 [406]
6[152] MIN
6-3/4 [171] MAX
6-3/4[171]
MIN
22[559]
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
STANDARD CONSTRUCTION
(WITHOUT SEISMIC ANCHORAGE)
2/2
RECOMMENDED
2 X 2 STAINLESS-STEEL
ANGLES AROUND
PERIMETER OF PIT
CORNER
LEVELING PAD
SEE PAGE 4 *(NOTE C)
FINISHED FLOOR
(RIGHT END)
WASHER DRAIN OUTLET
SEE PAGE 4 *(NOTE B)
B-BSECTION
RECOMMENDED
2 X 2 STAINLESS-STEEL
ANGLES AROUND
PERIMETER OF PIT
CORNER
LEVELING PAD
WASHER DRAIN OUTLET
SEE PAGE 4 *(NOTE B)
FINISHED FLOOR
(SERVICE AREA)
A-ASECTION
Page 5 of 17 920513916.E
22[559]
73-1/2 [1867]
[3150]
4-1/8[105]
112[2845]
[1981]
4X 8 [203]
68-3/4 [1746]
4X 10 [254]
118[2997]
3[76]
3[76]
*(NOTE E)
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
CONSTRUCTION WITH SEISMIC ANCHORAGE
(SEE REPORT #10183017 AGAINST CBC-2013)
1/2
SUMP
FLOOR DRAIN
FLOOR SLOPE
TOWARD FLOOR
DRAIN DEGREE: 1
LIMIT OF PIT
*(NOTE D)
SERVICE
AREA
NOTES:
(A) ALL MATERIALS (INCLUDING CONCRETE, STAINLESS STEEL ANGLES, FLOOR SINK...) ARE
BY OTHERS THAN STERIS. SLOPE PIT FLOOR TO DRAIN. STAINLESS STEEL ANGLES ARE
RECOMMENDED BY STERIS BUT ARE NOT NECESSARY FOR WASHER INSTALLATION.
(B) REFER TO PAGE 13 FOR FLOOR DRAIN FLOW RATE CAPACITY REQUIREMENTS.
(C) STERIS RECOMMENDS A MAXIMUM 8 in [203 mm] ROUND OR SQUARE FLOOR SINK BY 6 in [152 mm] DEEP.
(D) THIS CART WASHER CAN FIT INTO AN EXISTING 76 in [1930 mm] PIT, BUT WITH RESTRICTED ACCESS TO
COMPONENTS LOCATED IN THE PIT. THE DRAIN LOCATION MAY NEED TO BE RELOCATED.
(E) FOR EXISTING 76 in [1930 mm] PIT WASHER, THE CLEARANCE WILL BE 2 in [51 mm] INSTEAD OF 3 in [76 mm].
(F) WELDING OF THE SUMP ATTACHMENT TO FOUNDATION PLATE IS PERFORMED BY OTHERS.
(G) REFER TO SEISMIC REPORT FOR FOUNDATION PLATES AND CONCRETE CONSTRUCTION REQUIREMENTS.
STERIS OFFERS SEISMIC PRE-INSTALL ACCESSORY (FD33-5). WHICH INCLUDES THE FOUNDATION PLATES.
(H) ALL FOUR CORNER FOUNDATION PLATES UPPER SURFACE SHALL BE LEVELLED. THE PLATE STANDING ON HIGHEST PART OF THE CONCRETE FLOOR SHALL MEET THE SHOWN POSITION. STERIS PROVIDES SHIMS TO LEVEL THE OTHER CORNERS TO A MAXIMUM OF 1/2 in [12.7 mm] FROM NOMINAL POSITION.
4X CORNER
FOUNDATION PLATES
*(NOTE G)
A
A
BB
Page 6 of 17 920513916.E
73-1/2[1867]
22[559]
FLOOR DRAIN
4 [102] MIN
FLOOR SINK
8 [203] MAX.
4X 8 [203]
8-1/8[206]
REF
4X 10 [254]
68-3/4[1746]
10[254]
MIN
SEE PAGE 6 *(NOTE H)
7+1/4 -0 178+6
-0[ ]
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
A-ASECTION
4X CORNER
FOUNDATION PLATES
SEE PAGE 6
*(NOTE G)
FINISHED FLOOR
(RIGHT END)
WASHER DRAIN OUTLET
SEE PAGE 6
*(NOTE B)
B-BSECTION
RECOMMENDED
2 X 2 STAINLESS-STEEL
ANGLES AROUND
PERIMETER OF PIT
SEE PAGE 6 *(NOTE C)
FINISHED FLOOR
(SERVICE AREA)
WASHER DRAIN OUTLET
SEE PAGE 6 *(NOTE B)
RECOMMENDED
2 X 2 STAINLESS-STEEL
ANGLES AROUND
PERIMETER OF PIT
CONSTRUCTION WITH SEISMIC ANCHORAGE
(SEE REPORT #10183017 AGAINST CBC-2013)
2/2
Page 7 of 17 920513916.E
85-7/8 [2181]
80-3/8 [2042]
67-3/8 [1711]
2X 91 [2311]
91-1/4[2318]
68-5/8 [1743]
21-5/8[549]
63-1/2 [1613]
[2591]
93[2362]
73-1/2 [1867]
108-3/8 [2753]
107-1/2 [2731]
75-1/4 [1911]
87-3/8[2219]
79-7/8[2029]
[1753]
HW HOT WATER
CW COLD WATER
PW PURE WATER (OPTION)
A COMPRESSED AIR
D FLOOR DRAIN LOCATION *(NOTE F)
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
RELAY FUNCTION SEE PAGE 9 *(NOTE B)
CR10 HOT WATER REQUEST
CR11 PURE WATER REQUEST
CR12 COLD WATER REQUEST
UTILITY (PLUMBING)
CONNECTION POSITION
FRONT VIEW
HW
CWPW
A
D
CR S
NOTES:
(A) ALL CONNECTIONS SHOULD BE IN ACCORDANCE
WITH LOCAL CODES.
(B) EXTERNAL ISOLATION VALVES, UNIONS, VACUUM
BREAKERS ON UTILITY LINES TO WASHER ARE
NOT PROVIDED BY STERIS.
(C) ALL OVER PRESSURE DEVICES TO CONTROL
PRESSURE OF STEAM, WATER AND AIR ARE NOT
INCLUDED. IT IS RESPONSIBILITY OF CUSTOMER
TO INSTALL THOSE EQUIPMENT ON MAIN UTILITIES.
(D) DO NOT PLACE PIPING IN SERVICE ACCESS AREA.
(E) SIGNAL AVAILABLE FOR WATER REQUEST INSIDE
ELECTRICAL BOX.
(F) WASHER DRAIN OUTLET MAY NOT ALIGN PERFECTLY
WITH CENTER OF FLOOR SINK. EDGES OF FLOOR SINK
MUST BE AT LEAST 1 in [25 mm] FROM THE WASHER DRAIN.
(G) AIR LINE, ELECTRICAL CONDUIT, DUCT WORK, STEAM PIPING,
CONDENSATE RETURN PIPING AND WATER PIPING SHOULD
BE ROUTED ABOVE WASH CHAMBER TO AVOID INTERFERING
WITH SERVICE OF COMPONENTS ON WASHER ROOF.
SIDE VIEW
PW
A
HW
CW
D
CR
S
*(NOTE E)
TOP VIEW
HW
CW
PW
A
D
CR
S
SERVICE
AREA
*(NOTE D)
6-3/4 MIN
[171]
S STEAM *(NOTE C)
CR CONDENSATE RETURN
Page 8 of 17 920513916.E
23-7/8[606]
80-3/4[2051]
49-5/8[1260]
E ELECTRICAL
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
RELAY FUNCTION *(NOTE B)
CR5 CYCLE IN OPERATION
CR6 LOW VENTILATION REQUEST
CR7 HIGH VENTILATION REQUEST
CR8 ALARM
CR9 CYCLE COMPLETE
CR10 HOT WATER REQUEST
CR11 PURE WATER REQUEST
CR12 COLD WATER REQUEST
UTILITY (ELECTRICITY)
CONNECTION POSITION
NOTES:
(A) ALL CONNECTIONS SHOULD BE IN ACCORDANCE WITH
LOCAL CODES.
(B) AUXILIARY DRY CONTACTS FOR CUSTOMER UTILITY CONTROL
AVAILABLE INSIDE CONTROL BOX.
SEE ELECTRICAL SCHEMATIC: #10132900
(C) FOR STANDARD THROUGH-WALL UNIT, CLEARANCE BETWEEN
CLOSED DOOR OF ELECTRIC CABINET AND FACILITY WALL, OR
ANY OPPOSING FIXTURE, MUST MEET LOCAL CODES.
(D) STERIS RECOMMENDS A LIGHT IN SERVICE
AREA ALONG WITH A PROVISION OF AN
ELECTRICAL OUTLET FOR MAINTENANCE.
FRONT VIEW
E
SIDE VIEW
E
*(NOTE C)
TOP VIEW
E SERVICE
AREA
Page 9 of 17 920513916.E
108-1/2[2756]
42-1/2[1080]
50-3/8[1280]
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
RELAY FUNCTION *(NOTE F)
CR5 CYCLE IN OPERATION
CR6 LOW VENTILATION REQUEST (250 SCFM)
CR7 HIGH VENTILATION REQUEST (1000 SCFM)
UTILITY (VENTILATION)
CONNECTION POSITION
NOTES:
(A) INSTALLATION OF EXTERNAL DUCTING FROM WASHER EXHAUST FAN TO THE EXTERIOR OF THE BUILDING (NOT BY STERIS) MUST NOT EXCEED 1 in H2O IN PRESSURE LOSS TO ENSURE OPTIMAL DRYING PERFORMANCE OF THE WASHER AND FAN MOTOR EFFICIENCY.
IF PRESSURE LOSS OF EXTERNAL DUCTING EXCEEDS 1 in H2O, AN EXTERNAL FAN MUST BE INSTALLED TO COMPENSATE (NOT BY STERIS).
(B) BUILDING DUCT MUST BE SCALED UP TO 10 in [254 mm] O.D. MINIMUM AS CLOSE AS POSSIBLE TO WASHER DUCT CONNECTION (ADAPTOR FROM 6 in [152 mm] TO BUILDING DUCT O.D. NOT PROVIDED BY STERIS).
(C) FACILITY ROOM VENTILATION MUST COMPENSATE FOR WASHER EXHAUST FLOW RATE.
(D) STERIS DOES NOT RECOMMEND INSTALLATION OF SPRINKLER OVER THE UNIT, BUT IF REQUIRED BY LOCAL CODE IT MUST BE FUSED TO 260 F [126 C] OR ABOVE. ANY SPRINKLER HEADS IN FRONT OF THE LOAD/UNLOAD DOORS MUST ALSO BE FUSED TO 260 F [126 C] OR ABOVE.
(E) DO NOT PLACE DUCT WORK TO INTERFERE WITH SERVICE ACCES AREA.
(F) SIGNAL AVAILABLE FOR EXTERNAL VENTILATION CONTROL INSIDE ELECTRICAL BOX.
FRONT VIEW
EX
SIDE VIEW
EX
*(NOTE F)
TOP VIEW
EX
SERVICE
AREA
*(NOTE E)
EX VENTILATION
Page 10 of 17 920513916.E
132[3353]
MAX
17[432]
60[1524]
REF
8-3/8[213]
CONSUMPTION / CYCLE CYCLES STANDARD DECONTAM INSTRUMENT (OPTIONAL)
CHEMICAL
(CH)
ENZYME
(OPTIONAL) ALL N/A N/A 2.5 oz @ 1/10 oz/gal
[76 ml @ 0.8 ml/L]
DETERGENT
FIRST 2.5 oz @ 1/10 oz/gal
[76 ml @ 0.8 ml/L] N/A 2.5 oz @ 1/10 oz/gal [76 ml @ 0.8 ml/L]
CONSECUTIVE 0.7 oz @ 1/10 oz/gal [21.6 ml @ 0.8 ml/L] N/A N/A
RINSE AID
FIRST 0.5 oz @ 1/50 oz/gal
[15 ml @ 0.16 ml/L] N/A N/A
CONSECUTIVE 0.14 oz @ 1/50 oz/gal [4.3 ml @ 0.16 ml/L] N/A N/A
DESCALER ALL N/A 50 oz @ 2 oz/gal [1520 ml @ 16 ml/L] N/A
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
DOSING STATION
*NOTES:
(A) MAXIMUM CONTAINER HEIGHT
FROM FLOOR IS 24 in (610 mm).
AMBIENT TEMPERATURE SHALL
NOT EXCEED 40°C (104°F)
(B) UP TO 150 ft AVAILABLE ON
SPECIAL ORDER
FRONT VIEW
CEILING
CHEMICAL
CONTAINERS
*(NOTE A)
12 ft [3.6 m] OF
INJECTION LINE
(FROM CONTAINERS
TO DOSING STATION)
DOSING STATION
WEIGHT : 15 lbs [6.8 kg]
UP TO 5 TUBES
*(NOTE B)
100 ft (30 m) OF INJECTION LINE
AND SIGNAL CABLE *(NOTE B)
DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.
STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND
SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.
EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO WASHER/STATION
TO AVOID FORMING BUNDLES THAT COULD AFFECT THE CHEMICAL INJECTION
SYSTEM'S PERFORMANCE (DO NOT CUT SIGNAL CABLE).
Page 11 of 17 920513916.E
UTILITY REQUIREMENTS CHART (PLUMBING)
UTILITY PARAMETER REQUIREMENT NOTES
HOT WATER
(HW)
*(NOTE 6)
CONNECTION SIZE / TYPE 1 in / FEMALE NPT 1 in / MALE BSPT
1. INSTALLATION OF A WATER HAMMER ARRESTOR
(NOT PROVIDED BY STERIS) IS RECOMMENDED.
2. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
3. OPTIMAL WASHER PERFORMANCES OBTAINED
WHEN UTILITIES ARE PROVIDED AT MAXIMUM
VALUE OF RANGES.
4. OPTIMAL CLEANING EFFICIENCY WHEN HOT
WATER IS AT A MAXIMUM OF 135 F [57.2 C].
5. IF CYCLE WASH PHASE SET POINT
TEMPERATURE IS SET HIGHER THAN 140 F
[60 C], CONSUMPTION FOR DRAIN COOLING
IS DOUBLED.
6. DEFAULT VALIDATED INSTRUMENT CYCLE
HOT WATER CONSUMPTION IS: 75 US gal [284 L].
THIS CONSUMPTION MAY BE REDUCED BY THE
USE OF PURE WATER FOR RINSE AND/OR
THERMAL RINSE PHASES.
7. IF DRAIN COOLING SYSTEM IS ENABLED,
COLD WATER MAXIMUM FLOW RATE IS
INCREASED TO: 60 US gpm [227 L/min]
8. DEFAULT VALIDATED INSTRUMENT CYCLE
COLD WATER CONSUMPTION IS: 50 US gal [189 L].
9. STANDARD VALIDATED CYCLES CAN BE SET
TO USE PW FOR THERMAL RINSE PHASE.
THEN, HW CONSUMPTION FOR THIS PHASE
WILL BE REPLACED BY PW.
10.VALIDATED INSTRUMENT CYCLES CAN BE
SET TO USE PW FOR RINSE AND/OR THERMAL
RINSE PHASES. THEN, HW CONSUMPTION FOR
THESE PHASES WILL BE REPLACED BY PW AT
A RATE OF 25 US gal [95 L] PER PHASE.
PRESSURE RANGE
15 - 50 psig DYNAMIC
[103-345 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE
19 US GPM [72 L/min] AT 15 psig [103 kPa] TO
35 US GPM [132 L/min] AT 50 psig [345 kPa]
TOTAL CONSUMPTION
FOR STANDARD CYCLE
FIRST CYCLE: 50 US GAL [189 L]
(25 US GAL [95 L] PER PHASE)
CONSECUTIVE CYCLES: 14 US GAL [53 L]
(7 US GAL [27 L] PER PHASE)
TEMPERATURE RANGE 110°F [43°C] TO 150°F [65.5°C]
WATER QUALITY HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
COLD WATER
(CW)
*(NOTE 7 - 8)
CONNECTION SIZE / TYPE 1 in / FEMALE NPT 1 in / MALE BSPT
PRESSURE RANGE
30 - 50 psig DYNAMIC
[206-345 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE
30 US GPM [114 L/min] AT 30 psig [206 kPa] TO
38.5 US GPM [146 L/min] AT
50 psig [345 kPa]
TOTAL CONSUMPTION
FOR STANDARD CYCLE
FIRST CYCLE: 0 US GAL [0 L]
CONSECUTIVE CYCLES: 0 US GAL [0 L]
TOTAL CONSUMPTION
FOR DRAIN COOLING
IF RECYCLING: 7 US GAL [26 ]
NOT RECYCLING: 46 US GAL [175 L]
TEMPERATURE RANGE 40 - 70°F [4.4 - 21°C] MAXIMUM
WATER QUALITY HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
PURE WATER
(PW)
*(NOTE 9 - 10)
CONNECTION SIZE / TYPE 3/4 in / FEMALE NPT 3/4 in / MALE BSPT
PRESSURE RANGE
5 - 50 psig DYNAMIC
[34 - 340 kPa] MAX. 90 psig STATIC [620 kPA]
OPERATING FLOW RATE
12 US GPM [45 L/min] AT 5 psig [34 kPa] TO
28 US GPM [106 L/min] AT 50 psig [345 kPa]
TOTAL CONSUMPTION
FOR STANDARD CYCLE
FIRST CYCLE: 0 US GAL [0 L]
CONSECUTIVE CYCLES: 0 US GAL [0 L]
TEMPERATURE RANGE 150°F [65.5°C] MAXIMUM
RESISTIVITY
(CONDUCTIVITY) 0.5-0.1 M -cm [2-10 µS/cm] pH 6.8 - 7.5
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 12 of 17 920513916.E
UTILITY REQUIREMENTS CHART (PLUMBING)
COMPRESSED
AIR
(A)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT 1/2 in / MALE BSPT
1. AIR PURITY PER ISO-8573-1:2010 [7:4:4]
MAX. PARTICULATE DENSITY: 10 mg/m³
MAX. DEW POINT FOR WATER CONTENT:
3°C [37°F]
MAX. OIL CONCENTRATION FOR THE OIL
CONTENT: 5 mg/m³
2. STERIS RECOMMENDS A REFRIGERATED AIR
DRYER WHERE ENVIRONMENTAL DEW POINT
CONDITIONS ARE HIGHER THAN
RECOMMENDED (CONTACT STERIS SALES
REPRESENTATIVE).
3. STERIS RECOMMENDS A MAXIMUM 8 in [203 mm] ROUND OR SQUARE FLOOR SINK BY 6 in [152 mm] DEEP.
4. IF DRAIN COOLING SYSTEM IS ENABLED,
DRAIN MAXIMUM FLOW RATE IS INCREASED TO:
120 US GPM [454 L/min]
5. MAXIMUM FLOW RATE OCCURS AT RESERVOIR
COMPLETE DRAINING (EX: POWER OFF). IT WILL
BE MAINTAINED FOR A MAXIMUM OF:
50 SECONDS FOR UNITS WITH 2 RESERVOIRS;
75 SECONDS FOR UNITS WITH 3 RESERVOIRS.
PRESSURE RANGE MIN. 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]
OPERATING FLOW RATE 6 SCFM [170 L/min]
DRAIN
(D)
CONNECTION SIZE / TYPE
*(NOTE 3) 4 in [101 mm] OD
OPERATING FLOW RATE
*(NOTE 4 - 5) MAX. 60 US GPM [227 L/min]
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
STEAM INLET
(S)
CONNECTION SIZE / TYPE 1 1/2 in / FEMALE NPT 1 1/2 in / MALE BSPT
6. STERIS REQUIRES CLEAN AND AT LEAST 97%
SATURATED DRY STEAM.
7. PROPER OPERATION OF THE WASHER
REQUIRES A MINIMUM DYNAMIC PRESSURE
DIFFERENTIAL OF 21.5 psig [148 kPa] BETWEEN
STEAM INLET AND CONDENSATE RETURN.
8. PEAK OPERATING FLOW RATES OCCUR WHEN
RECIRCULATION HEATER AND RESERVOIR
HEATER ARE ACTIVATED AT THE SAME TIME.
IT WILL BE MAINTAINED FOR A MAXIMUM OF
90 SECONDS.
9. STERIS REQUIRES A LOW PRESSSURE
CONDENSATE RETURN SYSTEM.
PRESSURE RANGE
30 - 80 psig DYNAMIC
[206-552 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING
MASS FLOW RATE
PEAK 900 lb/h [409 kg/h] AT 30 psig [207 kPa]
PEAK 1400 lb/h [636 kg/h AT 80 psig [550 kPa]
TOTAL CONSUMPTION
FOR STANDARD CYCLES
FIRST CYCLE: 44 lb [20.0 kg] CONSECUTIVE CYCLES: 39 lb [17.7 kg]
CONDENSATE
RETURN
(CR)
CONNECTION SIZE / TYPE 1 in / FEMALE NPT 1 in / MALE BSPT
PRESSURE RANGE MAX. 8.5 psig [58.6 kPa] STATIC
OPERATING FLOW RATE PEAK : 2.9 US GPM [11 L/min]
Page 13 of 17 920513916.E
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY REQUIREMENTS CHART (ELECTRIC)
UTILITY
PARAMETER AND REQUIREMENT NOTES
CONNECTION
SIZE/TYPE
NOMINAL
AMPERAGE
RECOMMENDED
CIRCUIT
ELECTRICITY
(EL)
200-208V 50 Hz, 3-Ph
1 - 1 1/4 in [35 - 44 mm]
CONDUIT
27.5A 40A
1. ALL CONNECTIONS SHOULD BE IN
ACCORDANCE WITH LOCAL CODES.
2. NEUTRAL WIRE NOT REQUIRED.
3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES
AND SAFETY-RELATED WORK PRATICES TO
DETERMINE THE VALUE OF PROTECTION.
4. WIRING ON THE EQUIPMENT TERMINATES AT
A JUNCTION BOX. WIRING BETWEEN JUNCTION
BOX AND BUILDING SERVICE LINES ARE NOT
SUPPLIED BY STERIS.
5. DISCONNECT SWITCHES WITH OFF POSITION
LOCKOUT (SUPPLIED BY OTHERS) MUST BE
INSTALLED IN ELECTRIC SUPPLY LINES NEAR THE
EQUIPMENT.
6. BREAKERS NOT PROVIDED BY STERIS
7. MAIN SUPPLY TERMINAL BLOCK WIRE
SIZE: 6 AWG MAX.
8. PROTECTIVE EARTH TERMINAL BLOCK WIRE
SIZE: 4 AWG MAX.
200-208V 60 Hz, 3-Ph
1 - 1 1/4 in [35 - 44 mm]
CONDUIT
25.5A 40A
380-400-415V 50 Hz, 3-Ph
1 - 1 1/4 in [35 - 44 mm]
CONDUIT
14.5A 20A
460-480V 60 Hz, 3-Ph
1 - 1 1/4 in [35 - 44 mm]
CONDUIT
13.5A 20A
380-400V 60 Hz, 3-Ph
1 - 1 1/4 in [35 - 44 mm]
CONDUIT
15A 20A
Page 14 of 17 920513916.E
UTILITY REQUIREMENTS CHART
VENTILATION
EXHAUST
(EX)
CONNECTION SIZE / TYPE
*(NOTE 1)
6 in [152 mm]
OUTSIDE DIAMETER
1. BUILDING DUCT MUST BE SCALED UP TO 10 in [254 mm] O.D.
MINIMUM AS CLOSE AS POSSIBLE TO WASHER DUCT
CONNECTION (ADAPTOR FROM 6 in [152 mm] TO BUILDING
DUCT O.D. NOT PROVIDED BY STERIS)
2. FOR ALL VENTILATION DUCTING FROM WASHER, STERIS
RECOMMENDS THE INSTALLATION OF A DEDICATED
CORROSION-PROOF AND WATERTIGHT DUCT RATED
TO AN OPERATING TEMPERATURE OF 210 F [99 C] OR
MORE TO THE EXTERIOR OF THE BUILDING, SLOPED
TOWARD THE WASHER AND FREE OF DEAD LEGS.
3. INTERNAL DRYING AND EXHAUST FANS PRODUCE 950 SCFM
AND BUILDING EXHAUST NEEDS TO EXHAUST AT THE RATE
OF 1000 SCFM.
PRESSURE DROP OF
BUILDING VENTILATION
DUCT WORK
< 1 in H2O @ 1000 scfm
OPERATING FLOW RATE
*(NOTE 3)
VAPOR EXHAUST AND DRYING:
1000 scfm [28.31 m
/min]
WASH AND RINSE:
250 scfm [7.07 m
/min]
TEMPERATURE RANGE MAX. 212°F [100°C]
PHYSICAL SPECIFICATIONS
PARAMETER SPECIFICATION NOTES
OPENING AREA FROM CLEAN SIDE
TO SERVICE AREA LESS THAN 30 in
[19354 mm
CHAMBER INTERNAL DIMENSIONS
W x H x L
59 ½ in X 88 in X 106 in [1511 mm X 2235 mm X 2692 mm]
CHAMBER INTERNAL VOLUME 321 ft3 [9.1 m3]
NOISE LEVEL: A-WEIGHTED EQUIVALENT SURFACE
SOUND PRESSURE LEVEL 65 dB (A) AS PER ISO 3746
PERMISSIBLE ENVIRONMENT
INDOOR USE ONLY:
- ALTITUDE OF OPERATION UP TO 6560 ft (2000 m);
- TEMPERATURE OF 41 TO 104 F (5 TO 40 C);
- MAXIMUM RELATIVE HUMIDITY IS 80% FOR
TEMPERATURES UP TO 88 F (31 C) DECREASING
LINEARLY TO 50% RELATIVE HUMIDITY AT 104 F (40 C);
- POLLUTION DEGREE 2: EQUIPMENT MUST BE
INSTALLED IN AN ENVIRONMENT WHERE NORMALLY
ONLY NON-CONDUCTIVE POLLUTION OCCURS BUT
WHERE OCCASIONAL, TEMPORARY CONDUCTIVITY
CAUSED BY CONDENSATION CAN BE EXPECTED.
AS DETERMINED
ACCORDING TO
INTERNATIONAL STANDARD
EN/IEC 61010-1, SECOND
EDITION
HEAT LOSS
AVERAGE STANDARD VALIDATED CYCLE WITH ROOM @ 75 F [24 C]
TOTAL HEAT LOSS
FREE-STANDING INSTALLATION (ONE WALL) THROUGH-WALLS INSTALLATION (TWO WALLS)
FRONT OF WALL BACK OF WALL AT EACH END BETWEEN WALLS
13800 Btu/h [4039 W]
4000 Btu/h [1171 W]
9800 Btu/h [2868 W]
4000 Btu/h [1171 W]
5800 Btu/h [1697 W]
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 15 of 17 920513916.E
T I T L E
EQUIPMENT DRAWING OF
VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR
STEAM HEATED, PIT MOUNTED UNIT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Frederick Pothier 920513916
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
PHYSICAL REQUIREMENTS
ITEM REQUIREMENTS SPECIFICATIONS OR REFERENCE
TRANSPORT
VERIFY THAT HALLS, DOORS, ELEVATORS AND ALL OTHER ENTRANCES MEET WEIGHT AND DIMENSIONAL CLEARANCE
SPECIFICATIONS:
ACCESS TO INSTALLATION SITE ACCESSIBLE BY PALLET JACK AND/OR FORK LIFT.
SHIPPING WEIGHT AND
DIMENSIONS (W x H x L)
CRATE A: 2400 lb [1088 kg] 49 x 83 x 121 in [1245 x 2108 x 3074 mm]
CRATE B: 2815 lb [1277 kg] 49 x 83 x 121 in [1245 x 2108 x 3074 mm]
CRATE C: 1325 lb [601 kg] 37 x 71 x 91 in [940 x 1803 x 2311 mm]
CRATE D: 1510 lb [685 kg] 49 x 54 x 121 in [1245 x 1371 x 3074 mm]
LARGEST UNCRATED PIECE:
WEIGHT AND DIMENSIONS
(W x H x L)
SUMP (CRATE A):
WEIGHT: 436 lb [198 kg]
71 x 7 x 113 in [1803 x 178 x 2870 mm]
HEAVIEST UNCRATED PIECE:
WEIGHT AND DIMENSIONS
(W x H x L)
RESERVOIR MECHANICAL CORE (CRATE C):
MAXIMUM WEIGHT: 1000 lb [454 kg]
21 x 68 x 44 in [533 x 1727 x 1117 mm]
WALLS AND
CEILING
VERIFY THAT WALLS AND CEILING MEET DIMENSIONAL AND POSITION REQUIREMENTS:
WALL OPENING DIMENSIONS
AND CEILING HEIGHT SEE PAGE 1 AND 2
SERVICE AREA CLEARANCE SEE PAGE 1 AND 2
FIRE PREVENTION DEVICES
STERIS DOES NOT RECOMMEND INSTALLATION OF SPRINKLER OVER THE
UNIT BUT IF REQUIRED BY LOCAL CODE IT MUST BE FUSE TO 260 F (126 C)
OR ABOVE. ANY SPRINKLER HEADS IN FRONT OF THE LOAD/UNLOAD DOORS
MUST ALSO BE FUSE TO 260 F (126 C) OR ABOVE.
FLOOR
VERIFY THAT FLOOR AND PIT MEET DIMENSIONAL AND CONSTRUCTION REQUIREMENTS:
MAXIMUM OPERATING WEIGHT
(WATER AND TWO HEAVY CASE CARTS)
LOADED FREE-STANDING UNIT:
7266 lb [3296 kg]
LOADED THROUGH-WALLS UNIT:
6150 lb [2790 kg]
AVERAGE FLOOR LOAD
OVER WASHER FOOTPRINT 0.49 psi [3.38 kPa]
MAX. FLOOR LOAD
AT WASHER FOOT PADS
(1 ½ in [38 mm] DIAMETER PADS)
SUMP, NON-SERVICE SIDE (2): 490 psi [3378 kPa] SUMP, SERVICE SIDE (2): 1275 psi [8791 kPa]
MECHANICAL CORE (4): 280 psi [1930 kPa]
PIT DIMENSIONS
SEE PAGE 4 & 5 FOR STANDARD PIT MOUNTED UNIT
SEE PAGE 6 & 7 FOR PIT MOUNTED UNIT WITH SEISMIC
CORNER LEVELING PADS/
FOUNDATION PLATES POSITION
FLOOR DRAIN DIMENSIONS
AND POSITION
FLOOR CLEARANCE FOR INSTALLATION SEE PAGE 1-2
Page 16 of 17 920513916.E
Signature History Record Signature Role Vote Event Date
Frederick Pothier Author Approve 2016-12-23 09:35:00 EST
Maxime Metivier Approver Approve 2016-12-23 10:09:01 EST
Page 17 of 17 920513916.E
| Sheet 1 |
| Views |
| new_view_35 |
| Zones |
| A1 |
| Sheet 2 |
| Views |
| new_view_37 |
| new_view_57 |
| Zones |
| A1 |
| Sheet 3 |
| Views |
| new_view_87 |
| new_view_88 |
| new_view_93 |
| Zones |
| A1 |
| Sheet 4 |
| Views |
| new_view_80 |
| Zones |
| A1 |
| Sheet 5 |
| Views |
| right_81 |
| right_81 |
| right_81 |
| bottom_82 |
| Zones |
| A1 |
| Sheet 6 |
| Views |
| new_view_55 |
| Zones |
| A1 |
| Sheet 7 |
| Views |
| A |
| A |
| A |
| right_70 |
| Zones |
| A1 |
| Sheet 8 |
| Views |
| new_view_38 |
| new_view_39 |
| new_view_40 |
| Zones |
| A1 |
| Sheet 9 |
| Views |
| new_view_44 |
| new_view_45 |
| new_view_46 |
| Zones |
| A1 |
| Sheet 10 |
| Views |
| new_view_41 |
| new_view_42 |
| new_view_43 |
| Zones |
| A1 |
| Sheet 11 |
| Views |
| new_view_47 |
| Zones |
| A1 |
| Sheet 12 |
| Zones |
| A1 |
| Sheet 13 |
| Zones |
| A1 |
| Sheet 14 |
| Zones |
| A1 |
| Sheet 15 |
| Zones |
| A1 |
| Sheet 16 |
| Zones |
| A1 |
File details come from the government source that posted it. Updated .