Equipment Drawing - Vision 1327 - STEAM PIT.pdf

PDF 556 KB Posted

Attached to
Y1DA--Project No. 580-17-121 Upgrade/Renovate/Relocate Sterile Processing Services (SPS)
Solicitation number
36C25621R0014
Issued by

About this file

This solicitation is for upgrading, renovating and relocating sterile processing services (SPS) at the Michael E. DeBakey Veterans Affairs Medical Center in Houston, Texas. The project will renovate approximately 16,500 square feet of SPS, logistics and laundry in the basement of Building 100. All construction work must adhere to the latest JCAHO, NFPA and life safety requirements and be performed without affecting ongoing medical center operations. The solicitation includes an RFP, statement of work, specifications and drawings providing additional details on the project's scope. The soliciting agency is the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 16. Responses are due based on the standard timeline outlined in the full solicitation document.

View the file

Other files

File Type Posted
087100 Door Hardware Revised.pdf PDF
36C25621R0014 0005.docx DOCX document
580-17-12_Renovate SPS_Project Drawings_1-15-21.pdf PDF
36C25621R0014 0004.docx DOCX document
36C25621R0014 0003.docx DOCX document
QH102.pdf PDF
36C25621R0014 0002.pdf PDF
Area B-F-compressed.pdf PDF
S3185_Steris_Vision1327-CartWash.pdf PDF
S9610_Steris_PrepTable.pdf PDF
S5505A_STERRAD-100NX.pdf PDF
S0235_Steris_Amsco400.pdf PDF
Area B-G-compressed.pdf PDF
Slab Infill Detail.pdf PDF
S0940_Steris_Amsco7052HPWasher.pdf PDF
105113 MetalLockers.docx DOCX document
Clarifications 12-15-2020.pdf PDF
S9815B_Steris_Amsco600.pdf PDF
S5505B_STERRAD-100S.pdf PDF
20201211_ADDENDUM_1.pdf PDF
Site Visit Attendees.docx DOCX document
DE-1R.pdf PDF
S5505C_STERRAD-NX.pdf PDF
36C25621R0014 0001.pdf PDF
580-17-121_Renovate SPS_Combined Specifications.pdf PDF
DOL Wate Determinations TX20200253 dated 9-11-20.pdf PDF
36C25621R0014.pdf PDF
580-17-121_Renovate SPS_Specifications.pdf PDF
Site Visit Instructions _ Renovate SPS_1.docx DOCX document
580-17-121 _ Renovate SPS _ Statement of Work.docx DOCX document
SF24 BID BOND.pdf PDF
580-17-121 _ Renovate SPS _ Drawings.pdf PDF
Show all 32

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

80[2032]

LOADING

CLEARANCE

HEIGHT

112-3/4[2864]

CLEARANCE

FOR RELIEF

VALVE

OPERATION

102[2591]

UNIT

HEIGHT

110-1/2[2807]

120[3048]

UNIT WIDTH

12-3/8[314] 40[1016]

LOADING

CLEARANCE

WIDTH

32[813]

13-1/8 [333]

84-1/2[2146]

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

TYPICAL INSTALLATION (1/2)

NOTES:

(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].

(B) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.

(C) ALLOW SUFFICIENT SPACE IN FRONT OF LOADING AREA TO LOAD / UNLOAD CARTS

ACCORDING TO CUSTOMER CARTS AND DIMENSION AND STERIS WASHING ACCESSORIES:

CONTAINER RACK 63 1/2 [1612] LG

UTENSIL RACK 65 1/2 [1663] LG

INSTRUMENT RACK 65 [1651] LG

(D) WALLS COULD BE BUILDING WALL OR SERVICE AND NON-SERVICE S/S PANELS ACCESSORIES

(E) FOR STANDARD THROUGH-WALL UNIT, CLEARANCE BETWEEN CLOSED DOOR OF ELECTRIC

CABINET AND FACILITY WALL, OR ANY OPPOSING FIXTURE, MUST MEET LOCAL CODES.

(F) SERVICE AREA ACCESS DOOR IS INSTALLED ON LOAD SIDE OF THE WASHER.

LOAD SIDE CAN BE EITHER ON LEFT OR RIGHT SIDE OF THE MECHANICAL CORE.

(G) IF RECOMMENDED FLOOR CLEARANCE FOR INSTALLATION IS NOT AVAILABLE

ADDITIONAL MANPOWER AND EQUIPMENT WILL BE REQUIRED.

FRONT VIEW

Page 1 of 17 920513916.E

Released

120[3048]

SERVICE CLEARANCE

ABOVE FINISHED

FLOOR

98 [2489] AT 104 [2642]

CLEARANCE MIN.

5-1/4[133]

2[51]

105[2667]

WALL OPENING

3-3/4[95]

122-1/2[3112]

UNIT LENGTH

124[3150]

WALL

OPENING

112[2845]

68[1727]

20[508]

80[2032]

33-1/2[851]

126[3200] 126[3200]

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

TOP VIEW

SERVICE

AREA

FLOOR CLEARANCE

FOR INSTALLATION

SEE PAGE 1 *(NOTE G) SEE PAGE 1 *(NOTE G)

FOR INSTALLATION

FLOOR CLEARANCE

SERVICE DOOR

SEE PAGE 1 *(NOTE F)

SEE PAGE 1 *(NOTE D)

SEE PAGE 1 *(NOTE D)

SIDE VIEW

SEE PAGE 1 *(NOTE E)

CEILING DECK (IF PRESENT)

TYPICAL INSTALLATION (2/2)

Page 2 of 17 920513916.E

85-1/2[2172]

MIN

102[2591]

UNIT HEIGHT

213[5410]

UNIT WIDTH

[76] 3[76]

78[1981]

217[5512]

WALL OPENING

[76]

78[1981]

3[76]

61[1549]

MIN

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SIDE BY SIDE

TYPICAL INSTALLATION

NOTES:

(A) IMPORTANT: FOR INVERTED WASHER, REFERENCE POINT "A" FOR UTILITIES POSITION IS REVERSED AS SHOWN.

(B) FOR TWO UNITS WITH SERVICE AREA BETWEEN, CLEARANCE BETWEEN CLOSED DOORS OF ELECTRIC CABINETS

AND NEAREST PART OF OTHER WASHER, OR ANY FIXTURE BETWEEN, MUST MEET LOCAL CODES.

MAIN WASHER INVERTED WASHER

SUMP

SUMP

COMMON

SERVICE

AREA

NON-SERVICE

SIDE WALL

NON-SERVICE

SIDE WALL

POINT A

SEE NOTE

*(NOTE A)

POINT A

SEE NOTE

*(NOTE A)

Page 3 of 17 920513916.E

73-1/2 [1867]

[3150]

112[2845] 3[76]

118[2997]

2X 12 [305]

[1981]

4X 16 [406]

2X 21-1/4 [540]

3[76]

*(NOTE E)

[559]

4-1/8 [105]

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

STANDARD CONSTRUCTION

(WITHOUT SEISMIC ANCHORAGE)

1/2

FLOOR SLOPE

TOWARD FLOOR

DRAIN DEGREE: 1

FLOOR DRAIN

LIMIT OF PIT

*(NOTE D)

SERVICE

AREA

NOTES:

(A) ALL MATERIALS (INCLUDING CONCRETE, LEVELING PADS, STAINLESS STEEL ANGLES, FLOOR SINK...) ARE

BY OTHERS THAN STERIS. SLOPE PIT FLOOR TO DRAIN. STAINLESS STEEL ANGLES ARE

RECOMMENDED BY STERIS BUT ARE NOT NECESSARY FOR WASHER INSTALLATION.

(B) REFER TO PAGE 13 FOR FLOOR DRAIN FLOW RATE CAPACITY REQUIREMENTS.

(C) STERIS RECOMMENDS A MAXIMUM 8 in [203 mm] ROUND OR SQUARE FLOOR SINK BY 6 in [152 mm] DEEP.

(D) THIS CART WASHER CAN FIT INTO AN EXISTING 76 in [1930 mm] PIT, BUT WITH RESTRICTED ACCESS TO

COMPONENTS LOCATED IN THE PIT. THE DRAIN LOCATION MAY NEED TO BE RELOCATED.

(E) FOR EXISTING 76 in [1930 mm] PIT WASHER, THE CLEARANCE WILL BE 2 in [51 mm] INSTEAD OF 3 in [76 mm].

SUMP

4X CORNER

LEVELING PAD

A

A

BB

Page 4 of 17 920513916.E

73-1/2[1867]

FLOOR DRAIN

4 [102] MIN

FLOOR SINK

8 [203] MAX.

2X 12 [305] 2X 21-1/4 [540]

4X 16 [406]

6[152] MIN

6-3/4 [171] MAX

6-3/4[171]

MIN

22[559]

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

STANDARD CONSTRUCTION

(WITHOUT SEISMIC ANCHORAGE)

2/2

RECOMMENDED

2 X 2 STAINLESS-STEEL

ANGLES AROUND

PERIMETER OF PIT

CORNER

LEVELING PAD

SEE PAGE 4 *(NOTE C)

FINISHED FLOOR

(RIGHT END)

WASHER DRAIN OUTLET

SEE PAGE 4 *(NOTE B)

B-BSECTION

RECOMMENDED

2 X 2 STAINLESS-STEEL

ANGLES AROUND

PERIMETER OF PIT

CORNER

LEVELING PAD

WASHER DRAIN OUTLET

SEE PAGE 4 *(NOTE B)

FINISHED FLOOR

(SERVICE AREA)

A-ASECTION

Page 5 of 17 920513916.E

22[559]

73-1/2 [1867]

[3150]

4-1/8[105]

112[2845]

[1981]

4X 8 [203]

68-3/4 [1746]

4X 10 [254]

118[2997]

3[76]

3[76]

*(NOTE E)

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

CONSTRUCTION WITH SEISMIC ANCHORAGE

(SEE REPORT #10183017 AGAINST CBC-2013)

1/2

SUMP

FLOOR DRAIN

FLOOR SLOPE

TOWARD FLOOR

DRAIN DEGREE: 1

LIMIT OF PIT

*(NOTE D)

SERVICE

AREA

NOTES:

(A) ALL MATERIALS (INCLUDING CONCRETE, STAINLESS STEEL ANGLES, FLOOR SINK...) ARE

BY OTHERS THAN STERIS. SLOPE PIT FLOOR TO DRAIN. STAINLESS STEEL ANGLES ARE

RECOMMENDED BY STERIS BUT ARE NOT NECESSARY FOR WASHER INSTALLATION.

(B) REFER TO PAGE 13 FOR FLOOR DRAIN FLOW RATE CAPACITY REQUIREMENTS.

(C) STERIS RECOMMENDS A MAXIMUM 8 in [203 mm] ROUND OR SQUARE FLOOR SINK BY 6 in [152 mm] DEEP.

(D) THIS CART WASHER CAN FIT INTO AN EXISTING 76 in [1930 mm] PIT, BUT WITH RESTRICTED ACCESS TO

COMPONENTS LOCATED IN THE PIT. THE DRAIN LOCATION MAY NEED TO BE RELOCATED.

(E) FOR EXISTING 76 in [1930 mm] PIT WASHER, THE CLEARANCE WILL BE 2 in [51 mm] INSTEAD OF 3 in [76 mm].

(F) WELDING OF THE SUMP ATTACHMENT TO FOUNDATION PLATE IS PERFORMED BY OTHERS.

(G) REFER TO SEISMIC REPORT FOR FOUNDATION PLATES AND CONCRETE CONSTRUCTION REQUIREMENTS.

STERIS OFFERS SEISMIC PRE-INSTALL ACCESSORY (FD33-5). WHICH INCLUDES THE FOUNDATION PLATES.

(H) ALL FOUR CORNER FOUNDATION PLATES UPPER SURFACE SHALL BE LEVELLED. THE PLATE STANDING ON HIGHEST PART OF THE CONCRETE FLOOR SHALL MEET THE SHOWN POSITION. STERIS PROVIDES SHIMS TO LEVEL THE OTHER CORNERS TO A MAXIMUM OF 1/2 in [12.7 mm] FROM NOMINAL POSITION.

4X CORNER

FOUNDATION PLATES

*(NOTE G)

A

A

BB

Page 6 of 17 920513916.E

73-1/2[1867]

22[559]

FLOOR DRAIN

4 [102] MIN

FLOOR SINK

8 [203] MAX.

4X 8 [203]

8-1/8[206]

REF

4X 10 [254]

68-3/4[1746]

10[254]

MIN

SEE PAGE 6 *(NOTE H)

7+1/4 -0 178+6

-0[ ]

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

A-ASECTION

4X CORNER

FOUNDATION PLATES

SEE PAGE 6

*(NOTE G)

FINISHED FLOOR

(RIGHT END)

WASHER DRAIN OUTLET

SEE PAGE 6

*(NOTE B)

B-BSECTION

RECOMMENDED

2 X 2 STAINLESS-STEEL

ANGLES AROUND

PERIMETER OF PIT

SEE PAGE 6 *(NOTE C)

FINISHED FLOOR

(SERVICE AREA)

WASHER DRAIN OUTLET

SEE PAGE 6 *(NOTE B)

RECOMMENDED

2 X 2 STAINLESS-STEEL

ANGLES AROUND

PERIMETER OF PIT

CONSTRUCTION WITH SEISMIC ANCHORAGE

(SEE REPORT #10183017 AGAINST CBC-2013)

2/2

Page 7 of 17 920513916.E

85-7/8 [2181]

80-3/8 [2042]

67-3/8 [1711]

2X 91 [2311]

91-1/4[2318]

68-5/8 [1743]

21-5/8[549]

63-1/2 [1613]

[2591]

93[2362]

73-1/2 [1867]

108-3/8 [2753]

107-1/2 [2731]

75-1/4 [1911]

87-3/8[2219]

79-7/8[2029]

[1753]

HW HOT WATER

CW COLD WATER

PW PURE WATER (OPTION)

A COMPRESSED AIR

D FLOOR DRAIN LOCATION *(NOTE F)

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

RELAY FUNCTION SEE PAGE 9 *(NOTE B)

CR10 HOT WATER REQUEST

CR11 PURE WATER REQUEST

CR12 COLD WATER REQUEST

UTILITY (PLUMBING)

CONNECTION POSITION

FRONT VIEW

HW

CWPW

A

D

CR S

NOTES:

(A) ALL CONNECTIONS SHOULD BE IN ACCORDANCE

WITH LOCAL CODES.

(B) EXTERNAL ISOLATION VALVES, UNIONS, VACUUM

BREAKERS ON UTILITY LINES TO WASHER ARE

NOT PROVIDED BY STERIS.

(C) ALL OVER PRESSURE DEVICES TO CONTROL

PRESSURE OF STEAM, WATER AND AIR ARE NOT

INCLUDED. IT IS RESPONSIBILITY OF CUSTOMER

TO INSTALL THOSE EQUIPMENT ON MAIN UTILITIES.

(D) DO NOT PLACE PIPING IN SERVICE ACCESS AREA.

(E) SIGNAL AVAILABLE FOR WATER REQUEST INSIDE

ELECTRICAL BOX.

(F) WASHER DRAIN OUTLET MAY NOT ALIGN PERFECTLY

WITH CENTER OF FLOOR SINK. EDGES OF FLOOR SINK

MUST BE AT LEAST 1 in [25 mm] FROM THE WASHER DRAIN.

(G) AIR LINE, ELECTRICAL CONDUIT, DUCT WORK, STEAM PIPING,

CONDENSATE RETURN PIPING AND WATER PIPING SHOULD

BE ROUTED ABOVE WASH CHAMBER TO AVOID INTERFERING

WITH SERVICE OF COMPONENTS ON WASHER ROOF.

SIDE VIEW

PW

A

HW

CW

D

CR

S

*(NOTE E)

TOP VIEW

HW

CW

PW

A

D

CR

S

SERVICE

AREA

*(NOTE D)

6-3/4 MIN

[171]

S STEAM *(NOTE C)

CR CONDENSATE RETURN

Page 8 of 17 920513916.E

23-7/8[606]

80-3/4[2051]

49-5/8[1260]

E ELECTRICAL

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

RELAY FUNCTION *(NOTE B)

CR5 CYCLE IN OPERATION

CR6 LOW VENTILATION REQUEST

CR7 HIGH VENTILATION REQUEST

CR8 ALARM

CR9 CYCLE COMPLETE

CR10 HOT WATER REQUEST

CR11 PURE WATER REQUEST

CR12 COLD WATER REQUEST

UTILITY (ELECTRICITY)

CONNECTION POSITION

NOTES:

(A) ALL CONNECTIONS SHOULD BE IN ACCORDANCE WITH

LOCAL CODES.

(B) AUXILIARY DRY CONTACTS FOR CUSTOMER UTILITY CONTROL

AVAILABLE INSIDE CONTROL BOX.

SEE ELECTRICAL SCHEMATIC: #10132900

(C) FOR STANDARD THROUGH-WALL UNIT, CLEARANCE BETWEEN

CLOSED DOOR OF ELECTRIC CABINET AND FACILITY WALL, OR

ANY OPPOSING FIXTURE, MUST MEET LOCAL CODES.

(D) STERIS RECOMMENDS A LIGHT IN SERVICE

AREA ALONG WITH A PROVISION OF AN

ELECTRICAL OUTLET FOR MAINTENANCE.

FRONT VIEW

E

SIDE VIEW

E

*(NOTE C)

TOP VIEW

E SERVICE

AREA

Page 9 of 17 920513916.E

108-1/2[2756]

42-1/2[1080]

50-3/8[1280]

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

RELAY FUNCTION *(NOTE F)

CR5 CYCLE IN OPERATION

CR6 LOW VENTILATION REQUEST (250 SCFM)

CR7 HIGH VENTILATION REQUEST (1000 SCFM)

UTILITY (VENTILATION)

CONNECTION POSITION

NOTES:

(A) INSTALLATION OF EXTERNAL DUCTING FROM WASHER EXHAUST FAN TO THE EXTERIOR OF THE BUILDING (NOT BY STERIS) MUST NOT EXCEED 1 in H2O IN PRESSURE LOSS TO ENSURE OPTIMAL DRYING PERFORMANCE OF THE WASHER AND FAN MOTOR EFFICIENCY.

IF PRESSURE LOSS OF EXTERNAL DUCTING EXCEEDS 1 in H2O, AN EXTERNAL FAN MUST BE INSTALLED TO COMPENSATE (NOT BY STERIS).

(B) BUILDING DUCT MUST BE SCALED UP TO 10 in [254 mm] O.D. MINIMUM AS CLOSE AS POSSIBLE TO WASHER DUCT CONNECTION (ADAPTOR FROM 6 in [152 mm] TO BUILDING DUCT O.D. NOT PROVIDED BY STERIS).

(C) FACILITY ROOM VENTILATION MUST COMPENSATE FOR WASHER EXHAUST FLOW RATE.

(D) STERIS DOES NOT RECOMMEND INSTALLATION OF SPRINKLER OVER THE UNIT, BUT IF REQUIRED BY LOCAL CODE IT MUST BE FUSED TO 260 F [126 C] OR ABOVE. ANY SPRINKLER HEADS IN FRONT OF THE LOAD/UNLOAD DOORS MUST ALSO BE FUSED TO 260 F [126 C] OR ABOVE.

(E) DO NOT PLACE DUCT WORK TO INTERFERE WITH SERVICE ACCES AREA.

(F) SIGNAL AVAILABLE FOR EXTERNAL VENTILATION CONTROL INSIDE ELECTRICAL BOX.

FRONT VIEW

EX

SIDE VIEW

EX

*(NOTE F)

TOP VIEW

EX

SERVICE

AREA

*(NOTE E)

EX VENTILATION

Page 10 of 17 920513916.E

132[3353]

MAX

17[432]

60[1524]

REF

8-3/8[213]

CONSUMPTION / CYCLE CYCLES STANDARD DECONTAM INSTRUMENT (OPTIONAL)

CHEMICAL

(CH)

ENZYME

(OPTIONAL) ALL N/A N/A 2.5 oz @ 1/10 oz/gal

[76 ml @ 0.8 ml/L]

DETERGENT

FIRST 2.5 oz @ 1/10 oz/gal

[76 ml @ 0.8 ml/L] N/A 2.5 oz @ 1/10 oz/gal [76 ml @ 0.8 ml/L]

CONSECUTIVE 0.7 oz @ 1/10 oz/gal [21.6 ml @ 0.8 ml/L] N/A N/A

RINSE AID

FIRST 0.5 oz @ 1/50 oz/gal

[15 ml @ 0.16 ml/L] N/A N/A

CONSECUTIVE 0.14 oz @ 1/50 oz/gal [4.3 ml @ 0.16 ml/L] N/A N/A

DESCALER ALL N/A 50 oz @ 2 oz/gal [1520 ml @ 16 ml/L] N/A

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

DOSING STATION

*NOTES:

(A) MAXIMUM CONTAINER HEIGHT

FROM FLOOR IS 24 in (610 mm).

AMBIENT TEMPERATURE SHALL

NOT EXCEED 40°C (104°F)

(B) UP TO 150 ft AVAILABLE ON

SPECIAL ORDER

FRONT VIEW

CEILING

CHEMICAL

CONTAINERS

*(NOTE A)

12 ft [3.6 m] OF

INJECTION LINE

(FROM CONTAINERS

TO DOSING STATION)

DOSING STATION

WEIGHT : 15 lbs [6.8 kg]

UP TO 5 TUBES

*(NOTE B)

100 ft (30 m) OF INJECTION LINE

AND SIGNAL CABLE *(NOTE B)

DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.

STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND

SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.

EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO WASHER/STATION

TO AVOID FORMING BUNDLES THAT COULD AFFECT THE CHEMICAL INJECTION

SYSTEM'S PERFORMANCE (DO NOT CUT SIGNAL CABLE).

Page 11 of 17 920513916.E

UTILITY REQUIREMENTS CHART (PLUMBING)

UTILITY PARAMETER REQUIREMENT NOTES

HOT WATER

(HW)

*(NOTE 6)

CONNECTION SIZE / TYPE 1 in / FEMALE NPT 1 in / MALE BSPT

1. INSTALLATION OF A WATER HAMMER ARRESTOR

(NOT PROVIDED BY STERIS) IS RECOMMENDED.

2. CUSTOMER MUST PROVIDE UTILITY

CONNECTIONS WITH SHUTOFFS DISCONNECTS

WITHIN 2 FEET OF THE PERIMETER OF THE

EQUIPMENT AND BELOW THE CEILING DECK.

3. OPTIMAL WASHER PERFORMANCES OBTAINED

WHEN UTILITIES ARE PROVIDED AT MAXIMUM

VALUE OF RANGES.

4. OPTIMAL CLEANING EFFICIENCY WHEN HOT

WATER IS AT A MAXIMUM OF 135 F [57.2 C].

5. IF CYCLE WASH PHASE SET POINT

TEMPERATURE IS SET HIGHER THAN 140 F

[60 C], CONSUMPTION FOR DRAIN COOLING

IS DOUBLED.

6. DEFAULT VALIDATED INSTRUMENT CYCLE

HOT WATER CONSUMPTION IS: 75 US gal [284 L].

THIS CONSUMPTION MAY BE REDUCED BY THE

USE OF PURE WATER FOR RINSE AND/OR

THERMAL RINSE PHASES.

7. IF DRAIN COOLING SYSTEM IS ENABLED,

COLD WATER MAXIMUM FLOW RATE IS

INCREASED TO: 60 US gpm [227 L/min]

8. DEFAULT VALIDATED INSTRUMENT CYCLE

COLD WATER CONSUMPTION IS: 50 US gal [189 L].

9. STANDARD VALIDATED CYCLES CAN BE SET

TO USE PW FOR THERMAL RINSE PHASE.

THEN, HW CONSUMPTION FOR THIS PHASE

WILL BE REPLACED BY PW.

10.VALIDATED INSTRUMENT CYCLES CAN BE

SET TO USE PW FOR RINSE AND/OR THERMAL

RINSE PHASES. THEN, HW CONSUMPTION FOR

THESE PHASES WILL BE REPLACED BY PW AT

A RATE OF 25 US gal [95 L] PER PHASE.

PRESSURE RANGE

15 - 50 psig DYNAMIC

[103-345 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING FLOW RATE

19 US GPM [72 L/min] AT 15 psig [103 kPa] TO

35 US GPM [132 L/min] AT 50 psig [345 kPa]

TOTAL CONSUMPTION

FOR STANDARD CYCLE

FIRST CYCLE: 50 US GAL [189 L]

(25 US GAL [95 L] PER PHASE)

CONSECUTIVE CYCLES: 14 US GAL [53 L]

(7 US GAL [27 L] PER PHASE)

TEMPERATURE RANGE 110°F [43°C] TO 150°F [65.5°C]

WATER QUALITY HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)

COLD WATER

(CW)

*(NOTE 7 - 8)

CONNECTION SIZE / TYPE 1 in / FEMALE NPT 1 in / MALE BSPT

PRESSURE RANGE

30 - 50 psig DYNAMIC

[206-345 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING FLOW RATE

30 US GPM [114 L/min] AT 30 psig [206 kPa] TO

38.5 US GPM [146 L/min] AT

50 psig [345 kPa]

TOTAL CONSUMPTION

FOR STANDARD CYCLE

FIRST CYCLE: 0 US GAL [0 L]

CONSECUTIVE CYCLES: 0 US GAL [0 L]

TOTAL CONSUMPTION

FOR DRAIN COOLING

IF RECYCLING: 7 US GAL [26 ]

NOT RECYCLING: 46 US GAL [175 L]

TEMPERATURE RANGE 40 - 70°F [4.4 - 21°C] MAXIMUM

WATER QUALITY HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)

PURE WATER

(PW)

*(NOTE 9 - 10)

CONNECTION SIZE / TYPE 3/4 in / FEMALE NPT 3/4 in / MALE BSPT

PRESSURE RANGE

5 - 50 psig DYNAMIC

[34 - 340 kPa] MAX. 90 psig STATIC [620 kPA]

OPERATING FLOW RATE

12 US GPM [45 L/min] AT 5 psig [34 kPa] TO

28 US GPM [106 L/min] AT 50 psig [345 kPa]

TOTAL CONSUMPTION

FOR STANDARD CYCLE

FIRST CYCLE: 0 US GAL [0 L]

CONSECUTIVE CYCLES: 0 US GAL [0 L]

TEMPERATURE RANGE 150°F [65.5°C] MAXIMUM

RESISTIVITY

(CONDUCTIVITY) 0.5-0.1 M -cm [2-10 µS/cm] pH 6.8 - 7.5

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

Page 12 of 17 920513916.E

UTILITY REQUIREMENTS CHART (PLUMBING)

COMPRESSED

AIR

(A)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT 1/2 in / MALE BSPT

1. AIR PURITY PER ISO-8573-1:2010 [7:4:4]

MAX. PARTICULATE DENSITY: 10 mg/m³

MAX. DEW POINT FOR WATER CONTENT:

3°C [37°F]

MAX. OIL CONCENTRATION FOR THE OIL

CONTENT: 5 mg/m³

2. STERIS RECOMMENDS A REFRIGERATED AIR

DRYER WHERE ENVIRONMENTAL DEW POINT

CONDITIONS ARE HIGHER THAN

RECOMMENDED (CONTACT STERIS SALES

REPRESENTATIVE).

3. STERIS RECOMMENDS A MAXIMUM 8 in [203 mm] ROUND OR SQUARE FLOOR SINK BY 6 in [152 mm] DEEP.

4. IF DRAIN COOLING SYSTEM IS ENABLED,

DRAIN MAXIMUM FLOW RATE IS INCREASED TO:

120 US GPM [454 L/min]

5. MAXIMUM FLOW RATE OCCURS AT RESERVOIR

COMPLETE DRAINING (EX: POWER OFF). IT WILL

BE MAINTAINED FOR A MAXIMUM OF:

50 SECONDS FOR UNITS WITH 2 RESERVOIRS;

75 SECONDS FOR UNITS WITH 3 RESERVOIRS.

PRESSURE RANGE MIN. 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]

OPERATING FLOW RATE 6 SCFM [170 L/min]

DRAIN

(D)

CONNECTION SIZE / TYPE

*(NOTE 3) 4 in [101 mm] OD

OPERATING FLOW RATE

*(NOTE 4 - 5) MAX. 60 US GPM [227 L/min]

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

STEAM INLET

(S)

CONNECTION SIZE / TYPE 1 1/2 in / FEMALE NPT 1 1/2 in / MALE BSPT

6. STERIS REQUIRES CLEAN AND AT LEAST 97%

SATURATED DRY STEAM.

7. PROPER OPERATION OF THE WASHER

REQUIRES A MINIMUM DYNAMIC PRESSURE

DIFFERENTIAL OF 21.5 psig [148 kPa] BETWEEN

STEAM INLET AND CONDENSATE RETURN.

8. PEAK OPERATING FLOW RATES OCCUR WHEN

RECIRCULATION HEATER AND RESERVOIR

HEATER ARE ACTIVATED AT THE SAME TIME.

IT WILL BE MAINTAINED FOR A MAXIMUM OF

90 SECONDS.

9. STERIS REQUIRES A LOW PRESSSURE

CONDENSATE RETURN SYSTEM.

PRESSURE RANGE

30 - 80 psig DYNAMIC

[206-552 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING

MASS FLOW RATE

PEAK 900 lb/h [409 kg/h] AT 30 psig [207 kPa]

PEAK 1400 lb/h [636 kg/h AT 80 psig [550 kPa]

TOTAL CONSUMPTION

FOR STANDARD CYCLES

FIRST CYCLE: 44 lb [20.0 kg] CONSECUTIVE CYCLES: 39 lb [17.7 kg]

CONDENSATE

RETURN

(CR)

CONNECTION SIZE / TYPE 1 in / FEMALE NPT 1 in / MALE BSPT

PRESSURE RANGE MAX. 8.5 psig [58.6 kPa] STATIC

OPERATING FLOW RATE PEAK : 2.9 US GPM [11 L/min]

Page 13 of 17 920513916.E

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

UTILITY REQUIREMENTS CHART (ELECTRIC)

UTILITY

PARAMETER AND REQUIREMENT NOTES

CONNECTION

SIZE/TYPE

NOMINAL

AMPERAGE

RECOMMENDED

CIRCUIT

ELECTRICITY

(EL)

200-208V 50 Hz, 3-Ph

1 - 1 1/4 in [35 - 44 mm]

CONDUIT

27.5A 40A

1. ALL CONNECTIONS SHOULD BE IN

ACCORDANCE WITH LOCAL CODES.

2. NEUTRAL WIRE NOT REQUIRED.

3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES

AND SAFETY-RELATED WORK PRATICES TO

DETERMINE THE VALUE OF PROTECTION.

4. WIRING ON THE EQUIPMENT TERMINATES AT

A JUNCTION BOX. WIRING BETWEEN JUNCTION

BOX AND BUILDING SERVICE LINES ARE NOT

SUPPLIED BY STERIS.

5. DISCONNECT SWITCHES WITH OFF POSITION

LOCKOUT (SUPPLIED BY OTHERS) MUST BE

INSTALLED IN ELECTRIC SUPPLY LINES NEAR THE

EQUIPMENT.

6. BREAKERS NOT PROVIDED BY STERIS

7. MAIN SUPPLY TERMINAL BLOCK WIRE

SIZE: 6 AWG MAX.

8. PROTECTIVE EARTH TERMINAL BLOCK WIRE

SIZE: 4 AWG MAX.

200-208V 60 Hz, 3-Ph

1 - 1 1/4 in [35 - 44 mm]

CONDUIT

25.5A 40A

380-400-415V 50 Hz, 3-Ph

1 - 1 1/4 in [35 - 44 mm]

CONDUIT

14.5A 20A

460-480V 60 Hz, 3-Ph

1 - 1 1/4 in [35 - 44 mm]

CONDUIT

13.5A 20A

380-400V 60 Hz, 3-Ph

1 - 1 1/4 in [35 - 44 mm]

CONDUIT

15A 20A

Page 14 of 17 920513916.E

UTILITY REQUIREMENTS CHART

VENTILATION

EXHAUST

(EX)

CONNECTION SIZE / TYPE

*(NOTE 1)

6 in [152 mm]

OUTSIDE DIAMETER

1. BUILDING DUCT MUST BE SCALED UP TO 10 in [254 mm] O.D.

MINIMUM AS CLOSE AS POSSIBLE TO WASHER DUCT

CONNECTION (ADAPTOR FROM 6 in [152 mm] TO BUILDING

DUCT O.D. NOT PROVIDED BY STERIS)

2. FOR ALL VENTILATION DUCTING FROM WASHER, STERIS

RECOMMENDS THE INSTALLATION OF A DEDICATED

CORROSION-PROOF AND WATERTIGHT DUCT RATED

TO AN OPERATING TEMPERATURE OF 210 F [99 C] OR

MORE TO THE EXTERIOR OF THE BUILDING, SLOPED

TOWARD THE WASHER AND FREE OF DEAD LEGS.

3. INTERNAL DRYING AND EXHAUST FANS PRODUCE 950 SCFM

AND BUILDING EXHAUST NEEDS TO EXHAUST AT THE RATE

OF 1000 SCFM.

PRESSURE DROP OF

BUILDING VENTILATION

DUCT WORK

< 1 in H2O @ 1000 scfm

OPERATING FLOW RATE

*(NOTE 3)

VAPOR EXHAUST AND DRYING:

1000 scfm [28.31 m

/min]

WASH AND RINSE:

250 scfm [7.07 m

/min]

TEMPERATURE RANGE MAX. 212°F [100°C]

PHYSICAL SPECIFICATIONS

PARAMETER SPECIFICATION NOTES

OPENING AREA FROM CLEAN SIDE

TO SERVICE AREA LESS THAN 30 in

[19354 mm

CHAMBER INTERNAL DIMENSIONS

W x H x L

59 ½ in X 88 in X 106 in [1511 mm X 2235 mm X 2692 mm]

CHAMBER INTERNAL VOLUME 321 ft3 [9.1 m3]

NOISE LEVEL: A-WEIGHTED EQUIVALENT SURFACE

SOUND PRESSURE LEVEL 65 dB (A) AS PER ISO 3746

PERMISSIBLE ENVIRONMENT

INDOOR USE ONLY:

- ALTITUDE OF OPERATION UP TO 6560 ft (2000 m);

- TEMPERATURE OF 41 TO 104 F (5 TO 40 C);

- MAXIMUM RELATIVE HUMIDITY IS 80% FOR

TEMPERATURES UP TO 88 F (31 C) DECREASING

LINEARLY TO 50% RELATIVE HUMIDITY AT 104 F (40 C);

- POLLUTION DEGREE 2: EQUIPMENT MUST BE

INSTALLED IN AN ENVIRONMENT WHERE NORMALLY

ONLY NON-CONDUCTIVE POLLUTION OCCURS BUT

WHERE OCCASIONAL, TEMPORARY CONDUCTIVITY

CAUSED BY CONDENSATION CAN BE EXPECTED.

AS DETERMINED

ACCORDING TO

INTERNATIONAL STANDARD

EN/IEC 61010-1, SECOND

EDITION

HEAT LOSS

AVERAGE STANDARD VALIDATED CYCLE WITH ROOM @ 75 F [24 C]

TOTAL HEAT LOSS

FREE-STANDING INSTALLATION (ONE WALL) THROUGH-WALLS INSTALLATION (TWO WALLS)

FRONT OF WALL BACK OF WALL AT EACH END BETWEEN WALLS

13800 Btu/h [4039 W]

4000 Btu/h [1171 W]

9800 Btu/h [2868 W]

4000 Btu/h [1171 W]

5800 Btu/h [1697 W]

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

Page 15 of 17 920513916.E

T I T L E

EQUIPMENT DRAWING OF

VISION® 1327 CART AND UTENSIL WASHER/DISINFECTOR

STEAM HEATED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Frederick Pothier 920513916

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

PHYSICAL REQUIREMENTS

ITEM REQUIREMENTS SPECIFICATIONS OR REFERENCE

TRANSPORT

VERIFY THAT HALLS, DOORS, ELEVATORS AND ALL OTHER ENTRANCES MEET WEIGHT AND DIMENSIONAL CLEARANCE

SPECIFICATIONS:

ACCESS TO INSTALLATION SITE ACCESSIBLE BY PALLET JACK AND/OR FORK LIFT.

SHIPPING WEIGHT AND

DIMENSIONS (W x H x L)

CRATE A: 2400 lb [1088 kg] 49 x 83 x 121 in [1245 x 2108 x 3074 mm]

CRATE B: 2815 lb [1277 kg] 49 x 83 x 121 in [1245 x 2108 x 3074 mm]

CRATE C: 1325 lb [601 kg] 37 x 71 x 91 in [940 x 1803 x 2311 mm]

CRATE D: 1510 lb [685 kg] 49 x 54 x 121 in [1245 x 1371 x 3074 mm]

LARGEST UNCRATED PIECE:

WEIGHT AND DIMENSIONS

(W x H x L)

SUMP (CRATE A):

WEIGHT: 436 lb [198 kg]

71 x 7 x 113 in [1803 x 178 x 2870 mm]

HEAVIEST UNCRATED PIECE:

WEIGHT AND DIMENSIONS

(W x H x L)

RESERVOIR MECHANICAL CORE (CRATE C):

MAXIMUM WEIGHT: 1000 lb [454 kg]

21 x 68 x 44 in [533 x 1727 x 1117 mm]

WALLS AND

CEILING

VERIFY THAT WALLS AND CEILING MEET DIMENSIONAL AND POSITION REQUIREMENTS:

WALL OPENING DIMENSIONS

AND CEILING HEIGHT SEE PAGE 1 AND 2

SERVICE AREA CLEARANCE SEE PAGE 1 AND 2

FIRE PREVENTION DEVICES

STERIS DOES NOT RECOMMEND INSTALLATION OF SPRINKLER OVER THE

UNIT BUT IF REQUIRED BY LOCAL CODE IT MUST BE FUSE TO 260 F (126 C)

OR ABOVE. ANY SPRINKLER HEADS IN FRONT OF THE LOAD/UNLOAD DOORS

MUST ALSO BE FUSE TO 260 F (126 C) OR ABOVE.

FLOOR

VERIFY THAT FLOOR AND PIT MEET DIMENSIONAL AND CONSTRUCTION REQUIREMENTS:

MAXIMUM OPERATING WEIGHT

(WATER AND TWO HEAVY CASE CARTS)

LOADED FREE-STANDING UNIT:

7266 lb [3296 kg]

LOADED THROUGH-WALLS UNIT:

6150 lb [2790 kg]

AVERAGE FLOOR LOAD

OVER WASHER FOOTPRINT 0.49 psi [3.38 kPa]

MAX. FLOOR LOAD

AT WASHER FOOT PADS

(1 ½ in [38 mm] DIAMETER PADS)

SUMP, NON-SERVICE SIDE (2): 490 psi [3378 kPa] SUMP, SERVICE SIDE (2): 1275 psi [8791 kPa]

MECHANICAL CORE (4): 280 psi [1930 kPa]

PIT DIMENSIONS

SEE PAGE 4 & 5 FOR STANDARD PIT MOUNTED UNIT

SEE PAGE 6 & 7 FOR PIT MOUNTED UNIT WITH SEISMIC

CORNER LEVELING PADS/

FOUNDATION PLATES POSITION

FLOOR DRAIN DIMENSIONS

AND POSITION

FLOOR CLEARANCE FOR INSTALLATION SEE PAGE 1-2

Page 16 of 17 920513916.E

Signature History Record Signature Role Vote Event Date

Frederick Pothier Author Approve 2016-12-23 09:35:00 EST

Maxime Metivier Approver Approve 2016-12-23 10:09:01 EST

Page 17 of 17 920513916.E

Sheet 1
Views
new_view_35
Zones
A1
Sheet 2
Views
new_view_37
new_view_57
Zones
A1
Sheet 3
Views
new_view_87
new_view_88
new_view_93
Zones
A1
Sheet 4
Views
new_view_80
Zones
A1
Sheet 5
Views
right_81
right_81
right_81
bottom_82
Zones
A1
Sheet 6
Views
new_view_55
Zones
A1
Sheet 7
Views
A
A
A
right_70
Zones
A1
Sheet 8
Views
new_view_38
new_view_39
new_view_40
Zones
A1
Sheet 9
Views
new_view_44
new_view_45
new_view_46
Zones
A1
Sheet 10
Views
new_view_41
new_view_42
new_view_43
Zones
A1
Sheet 11
Views
new_view_47
Zones
A1
Sheet 12
Zones
A1
Sheet 13
Zones
A1
Sheet 14
Zones
A1
Sheet 15
Zones
A1
Sheet 16
Zones
A1