S0940_Steris_Amsco7052HPWasher.pdf
PDF 291 KB Posted
- Attached to
- Y1DA--Project No. 580-17-121 Upgrade/Renovate/Relocate Sterile Processing Services (SPS)
- Solicitation number
- 36C25621R0014
About this file
This document contains a technical specification for an AMSCO 7052HP Single-Chamber Washer/Disinfector to be provided under a federal solicitation. Key details include:
-
The washer/disinfector will be used to clean and disinfect reusable surgical instruments, utensils, rubber/plastic goods and related medical items. It features a touchscreen control system, vertical sliding doors, stainless steel construction and cycles that are customizable to meet specific operating requirements.
-
The unit must meet standards including CAN/CSA, UL and ISO regarding safety, performance and processing of reusable medical devices. It has dimensions of 42x80-3/4x32 inches and a load height of 31 inches. Safety features include door obstruction sensors and interlocks.
-
The related solicitation is for the upgrade, renovation and relocation of Sterile Processing Services at the Michael E. DeBakey VA Medical Center and involves approximately 16,500 square feet of space. All construction must adhere to JCAHO, NFPA and life safety standards while avoiding disruption to medical center operations.
-
The technical specification provides extensive details on the washer/disinfector features, cycles, control system, installation requirements, preventative maintenance options, notes and required utilities to inform evaluation of proposals to fulfill the stated need.
View the file
Other files
Show all 32
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
AMSCO® 7052HP SINGLE-CHAMBER
WASHER/DISINFECTOR
APPLICATION
The AMSCO 7052HP Single-Chamber Washer/ Disinfector with touch screen control is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans and urinals, trays, bowl, basin, kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
DESCRIPTION
The AMSCO 7052HP Single-Chamber Washer/ Disinfector is a cabinet-type mechanical washer that includes:
• Color touch screen control system and visual progress indicator
• Audible warning system with adjustable volume
• Space saving, power vertical sliding doors
• Eight factory-loaded (Instruments, Orthopedic Instruments, Utensils, Plastic Goods, Rigid M.I.S., Anesthesia / Respiratory Goods, Gentle and Decontamination) and up to 20 Custom cycles. Each cycle is effective, fast, energy efficient and may be customized (within allowed range of parameters) to meet specific operating requirements
• LED status indicator lighting
• Compact space saving footprint
• Conforms to seismic building code requirements The Selections Che
POWER*
❑ Steam-Heated Unit ❑ Electric-Heated Unit
VOLTAGE*
❑ 200 to 208 V, 60 Hz (Steam) ❑ 200 to 208 V, 50 Hz (Steam) ❑ 380 to 400 V, 60 Hz ❑ 380 to 415 V, 50 Hz ❑ 460 to 480 V, 60 Hz
LANGUAGE PACKAGE
❑ French (FD355) ❑ Spanish (FD356) ❑ German (FD357) ❑ Italian (FD358)
* Careful consideration must given to voltage selection prior to ordering. Later changes require substantial field modification.
† Refer to SD867 for Material Handling Accessories.
ACCESSOR
❑ Printer ❑ Seismic ❑ Conden ❑ Two-Le ❑ Two-Le
(FD79-1
❑ Three-L ❑ Four-Le ❑ Five-Le ❑ Multi-Fu ❑ Multi-Fu ❑ General ❑ Self Do ❑ Transfe ❑ Set of d ❑ Handhe ❑ Conduc ❑ Addition ❑ Addition ❑ Installa ❑ Installa (Typical - details may vary.)
• Patented spray arm system providing entire chamber coverage
• 180 degree reverse installable
STANDARDS
The AMSCO 7052HP Single-Chamber Washer/Disinfector complies with the following standards:
As certified by Intertek:
• CAN/CSA-C22.2 No. 61010-1, 3rd Edition (R2017)
• CAN/CSA-C22.2 No. 61010-2-040 2nd Edition Dated 07/01/2016
• UL 61010-1, 3rd Edition (2012-05-11) 10283376 (Rev A)
SD1058 (06/01/19)
Item ________________________
Location(s)___________________ cked Below Apply To This Equipment
IES†
(FD094)
Tie-Down Kit (FD354) sate Return To Drain (FD353) vel Manifold Rack (FD74-800) vel Manifold Rack with Equal Space 00) evel Manifold Rack (FD74-900) vel Manifold Rack (FD75-000) vel Manifold Rack (FD75-200) nction Rack for Large Items (FD75-500) nction Rack for Small Items (FD75-600)
Purpose Basket (FD77-100) cking Transfer Cart (FD199) r Cart Adapter (FD209) ry Contacts (FD362) ld Bar Code reader (FD363) tivity Probe (FD350) al Wall Flange (FD351) al Chemical Pump (FD349) tion Kit – Steam (FD088) tion Kit – Electric (FD037)
ACCESSORIES (CONT’D)†
❑ Offset Installation Air Management Kit (FD352) ❑ IQ-OQ-PQ Protocols (FD359) ❑ Machine Identifier (11000866) ❑ Drain Acu-CoolDown (FD361) ❑ ConnectAssure Technology for Data Export
(CAT3030)
❑ ProConnect® Technical Support Services
(Remote Monitoring, Priority Technical Support, Customer Care Center Access, Equipment Performance Reports). Available in U.S. and Canada only. (GP09-167)
OPTIONS
❑ Acu-Rinse Reservoir (FD000054) ❑ Acu-Wash Reservoir (FD000055) ❑ 40-Amp Min Circuit Protection (FD000056)
Governing Directive for the affixing of the CE mark:
• Medical Devices Directive 93/42/EEC as amended by 2007/
47/EC
• Machinery Directive 2006/42/EC
Standards Applied to Demonstrate Conformity to the Directive:
• IEC 61010-1 3rd Edition (2012-05-11), EN 61326-1, 2013, IEC 61326-1 Second Edition (July 2012) and IEC 61010-2- 040:2015
• EN ISO 15883-1/A1:2014, ANSI/AAMI ST15583-1:2009 and CAN/CSA Z15883-1-09 (R2104) Washer-Disinfectors – General Requirements, Definitions and Tests
• EN ISO 15883-2: 2009, ANSI/AAMI ST15583-2:2013 and CAN/CSA Z15883-2-09 (R2014) Requirements and Tests for Washer-Disinfectors Employing Thermal Disinfection for Surgical Instruments, Anesthesia Equipment, Holloware, Utensils, Glassware, etc.
SIZE (W X H X L)
Exterior dimensions:
• 42 x 80-3/4 x 32" (1067 x 2051 x 813 mm)
Interior chamber dimensions:
• 29-1/2 x 29-1/4 x 26-1/2" (749 x 743 x 673 mm)
Load height:
• 31" (787 mm) above finished floor
FEATURES
Vertical sliding power door:
• Door is constructed of double pane tempered glass to allow operator to view chamber interior with door closed.
• Color LED light illuminates top portion of each door to indicate machine status.
• Door is pneumatically activated using touch screens located on the control panel (one each side of unit) and is equipped with a built-in safety system.
• Each door is mounted on a compressed seal that reduces heat loss and increases heating capability.
• If a power failure occurs, door can be opened manually. A door interlock feature is provided to prevent cross-contamination.
• Unit can be configured in single- or double-door operation.
Stainless-steel pump is powered by a dual-speed motor. High pump speed provides equivalent capacity of a 14 hp (10.4 kW) motor, 165 U.S. gal/min at 196 ft (625 L/min at 60 m) head pressure.
Low pump speed provides equivalent capacity of a 3.5 HP (2.6 kW) motor, 90 U.S. gal/min at 46 ft (341 L/min at 14 m) head pressure. Pump impeller is mounted directly on motor shaft and does not require additional bearings. Pump motor is equipped with a drip-proof frame, magnetic starter, overload protection and sealed bearings (not requiring periodic lubrication).
Pump, spray system and all recirculating piping are of stainless-steel construction.
Wash chamber:
• Constructed of 16-gauge, 316L stainless steel (No. 4 finish), argon-welded and polished. Chamber is of sanitary-type design (horizontal fixed surfaces are sloped, without overlapping sheet metal) for complete drainability and reducing hard-to-clean locations.
• Single-walled, insulated construction of chamber exterior reduces heat loss and noise level to the work area.
• Rotary spray assemblies are positioned (one at top and one at bottom of chamber, each measuring 24-3/8" [620 mm] long) to reach all load surfaces. Spray assemblies are easily removable without tools to aid in cleaning and maintenance.
Sump:
• Constructed of 316L stainless steel with 7.2 gal (27 L) capacity with removable stainless-steel filter preventing debris from entering pump and piping system.
• Heating coil (steam or electric) at the bottom of the wash chamber (sump) raises and maintains water temperature up to 185°F (85°C) during the Wash phase and up to 194°F (90°C) during the Thermal Rinse phase.
Chemical peristaltic injection pumps:
• Pre-Programmed cycles and pumps are optimized to use Prolystica® Ultra Concentrate (PUC) HP chemicals. Each ultra concentrated product is 10 times more concentrated than equivalent traditional product; so,10 times less chemical is injected per cycle.
• Unit comes with one enzyme and one detergent injection peristaltic pump which automatically add a selected quantity of chemical from 1/40 to 2 oz/gal (0.2 to 16 mL/L).
• One pump is dedicated to automatically add a selected quantity of lubricant during Thermal Rinse phase.
• Pumps are positioned near chemical containers. A low-level sensor is included to indicate when detergent level is low, or when insufficient chemical is available for next cycle.
• 100’ (30 m) extension tubes are included to pump chemicals from canisters installed in a remote location to unit.
• Control monitors volume of chemicals injected and indicates if this parameter meets specified criteria during all specific phases.
Top Utility Connections (except drain connections) facilitate installation. All utilities (including vent, steam, electric, water and compressed air) are connected on washer top. Top water injection gives a secure air gap of 31-1/2" (800 mm). Backflow preventers are not required for hot and pure water.
Front Service Panels provide easy access to piping, valves, electrical components and wiring. Unit can be configured as either a Vented System that exhausts chamber vapors to building exhaust system through 3.0" (76 mm) OD vent connection located at top of washer or as a Non-Vented Drying System where chamber vapors are exhausted through cold water condenser to facility room.
Drying System uses uniquely designed four-sided inflow drying pattern to produce high-flow air curtain. Air curtain provides broad, efficient, drying within wash chamber. Recycled and non-recycled air is manifolded and circulated through piping and accessory providing energy efficient system while ensuring complete chamber air coverage. Fresh air is drawn through a HEPA filter. Drying system includes a 3 hp (2.2 kW) blower, to remove vapor from chamber prior to opening doors, and three electric heaters totaling 11.8 kW to heat and maintain chamber air temperature. When Acu-Drying is selected, drying time self-adjusts to meet drying time of loaded rack and items.
Pure Water Stainless-Steel Supply Valve is supplied on washer/disinfector so Pure Water Rinse Supply may be used with factory programmed cycles and optional Acu-Rinse Reservoir.
Drain Discharge Cool Down ensures water drained at the end of each phase, from the chamber sump to the building drain system, does not exceed 140°F (60°C). If the water temperature in sump is higher than 140°F (60°C), cold water is automatically added to reduce the temperature of the water discharged into the building drain system.
Sampling Valve is supplied in washer/disinfector recirculation piping for North American Customers that are not able to install sampling valves on facility supply lines near washer/disinfector.
Barrier Wall Kit Flanges for one wall (clean side) installation come standard. An optional kit to enclose second wall (dirty side) is also available (FD351).
Cycle Graph enables monitoring of injection volume, water circulating pressure and water temperature during a cycle.
Graphs are displayed at the end of each cycle and data can be exported using https interface or USB port.
CONTROL SYSTEM
The AMSCO 7052HP Single-Chamber Washer/Disinfector is equipped with a touch screen Customer interface. This 8.4" (213 mm) color touch screen is mounted at eye level beside chamber door (both sides of unit) and tilted for better visibility, allowing easy monitoring of all wash cycles. Control system monitors and controls all phases of each programmed cycle.
The control system features:
• Cycle selection is simplified through use of Icons representing devices and cycle confirmation (on-screen) to ensure correct cycle is selected.
• Service mode for preventive maintenance testing and to facilitate troubleshooting.
• Built-in service diagnostic program to permit system calibration and verification of component operations.
• Security lock-out feature that enables cycles and temperatures to be locked and unchangeable without proper access code.
• Cycle data is stored as a protection against power disruption. Data may be downloaded from controller using supplied USB ports.
• Permits operator to monitor current washer/disinfector status (including current chamber temperature, circulating pressure and time remaining in phase).
• Indicates any abnormal conditions with audible and visual alarm system.
CYCLE DESCRIPTION
ADVISORY NOTE: STERIS does not intend, recommend or represent in any way that this AMSCO 7052HP Single-Chamber Washer/Disinfector be used for the terminal disinfection or sterilization of any regulated medical device. AMSCO 7052HP Washer/Disinfector is intended only to perform an initial step in the processing of soiled, reusable medical devices. If medical devices contact blood or compromised tissues, such devices must be terminally processed in accordance with good hospital practices before each use in human patients.
Once treatment cycle is selected, washer/disinfector automatically processes load through standard phases (additional phases are included in certain treatment cycles depending on unit configuration) as described on Cycle Charts and in supplied Operator Manual.
AMSCO 7052HP Single-Chamber Washer/Disinfector enables Customers to use Dual Wash phase combining PUC HP enzyme and PUC HP detergent in the same filling which accelerates cycle time and reduces water consumption.
Custom cycles are available but must be validated by the Customer (not STERIS responsibility).
SAFETY FEATURES
Vertical chamber doors are equipped with an obstruction sensor to detect any door obstruction. If obstruction is present, door automatically opens.
The washer/disinfector is equipped with a safety lockout feature so a cycle cannot start unless the door is fully closed.
If the door is opened during a cycle, all utility services to chamber are shut off, and the cycle stops.
Door interlock feature is provided to prevent cross-contamination. Door interlock feature allows only one door to be opened at a time whenever power is on. When the cycle is in process, door interlock prevents either door from being opened. Access to the load is then restricted.
Safety stop pushbutton(s), one on the load and one on the unload side, automatically stop all unit operation when pressed.
Building electrical supply disconnect switch must be used to shut off power to the unit before servicing.
ACCESSORIES AND OPTIONS*
Printer (FD094), if provided, produces an easy-to-read printed record of whether load was properly processed at preset temperature, as well as a complete list of alarm and abort in-cycle messages. A paper out LED is included.
* See SD867 for information on material handling accessories, including the Rigid Minimal Invasive Surgery (MIS) Rack.
Acu-Rinse Reservoir (if equipped) pre-heats Pure Water for Rinse and Thermal Rinse phases significantly reducing cycle times. Constructed of 316L stainless steel.
Acu-Wash Reservoir recovers a portion of water used in Thermal Rinse treatment and stores it for use in Wash treatment of next cycle reducing overall cycle water consumption.
Steam Condensate Return To Drain (FD353) allows for connection of steam condensate return outlet to drain when steam condensate return line is not available in building. Cold water is also injected in drain piping when condensate return water temperature is too high. Condensate return cool down keeps temperature in drain piping below 140°F (60°C).
Seismic Tie-Down Kit (FD354) includes a seismic report for proper installing and securing of washer to building floor. Unit is designed to comply with seismic building code.
Additional Chemical Pump (FD349; up to two pumps) is available, giving flexibility to wash with neutral process, alkaline process, or to vary chemical used, depending on load type.
Drain Acu-CoolDown (FD361) precisely injects correct amount of potable water into facility drain system, minimizing volume of cold water needed.
Offset Installation Air Management Kit (FD352) controls airflow at top of door assemblies in a double wall installation.
ConnectAssure Data Export (CAT3030) exports XML cycle records from equipment to Customer tracking system.
ProConnect® Technical Support Services maximizes operational efficiencies with secure, Internet-based, real-time equipment monitoring. Data from Customer equipment is used by STERIS to provide pro-active Customer alert notifications, technical support and predictive maintenance. (ProConnect Technical Support Services is available in U.S. and Canada only.) Refer to Tech Data sheet SD983, PROCONNECT TECHNICAL SUPPORT SERVICES, for details.
INSTALLATION
The washer/disinfector is designed as a fully enclosed cabinet for freestanding or recessed installation. Clearance between top of the unit and the ceiling must be at least 14" (356 mm).
If the system is recessed through one or two barrier walls, stainless-steel barrier flanges are included to provide a finished wall appearance.
Once installed, system is designed to allow for easy access for maintenance purposes.
All configuration modifications are available upon request.
Contact STERIS to receive a quotation.
PREVENTIVE MAINTENANCE
Customers are encouraged to contact STERIS concerning our annual maintenance program. Under the terms of the program, preventive maintenance, adjustments and replacement of worn parts are provided on a scheduled basis to help ensure optimal equipment performance and help minimize untimely or costly schedule interruptions. STERIS maintains a worldwide staff of well-equipped, factory-trained technicians to provide these services, as well as on-site installation, training and expert repair services. Contact STERIS for details.
NOTES
1. Machine segments (unit is bolted onto a wooden skid) are shipped in one cardboard carton (W x H x L): 50 x 91 x 42" (1270 x 2311 x 1067 mm).
2. Customer must ensure the washer/disinfector stands on a noncombustible floor (floor should be level).
3. Customer must provide utility connections with shutoff disconnects within 2’ (0.6 m) of equipment perimeter and below ceiling deck or purchase Installation Kit from STERIS.
4. STERIS recommends vacuum breakers (not provided by STERIS) be installed on service lines, and disconnect switches (with lockout in OFF position; not provided by STERIS) be installed in electric supply lines near equipment.
5. For all ventilation ducting from the washer/disinfector, STERIS recommends installation of a dedicated, corrosion-proof, flexible watertight duct (3" [76 mm] OD) to the exterior of the building, sloped toward the washer/disinfector.
6. Minimum ceiling height for door removal is 94" (2388 mm).
7. STERIS recommends illumination of the service area along with providing a convenience outlet for maintenance.
8. Noise level of 60.6 dB; Operating weight of 1543 lb (700 kg).
9. Unit is shipped with an Operator Manual. Maintenance Manual is available for purchase.
10. ConnectAssure Data Export requires an ethernet drop and ethernet connection within 10 ft of the equipment. The Customer must supply the network cable.
UTILITY REQUIREMENTS
IMPORTANT: Refer to equipment drawing 10288794 for installation details and specifications.
For Further Information, contact:
STERIS Corporation 5960 Heisley Road Mentor, OH 44060-1834 • USA 440-354-2600 • 800-548-4873 www.steris.com
0283376 [SD1058] ©2019, STERIS Corporation. All rights reserved. Rev A (06/01/19)
This document is intended for the exclusive use of STERIS Customers, including architects or designers. Reproduction in whole or in part by any party other than a Customer is prohibited.
The base language of this document is ENGLISH. Any translations must be made from the base language document.
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH
APPLICABLE LOCAL AND NATIONAL CODES AND
REGULATIONS.
| APPLICATION |
| DESCRIPTION |
| STANDARDS |
| SIZE (W X H X L) |
| FEATURES |
| CONTROL SYSTEM |
| CYCLE DESCRIPTION |
| SAFETY FEATURES |
| ACCESSORIES AND OPTIONS* |
| INSTALLATION |
| PREVENTIVE MAINTENANCE |
| NOTES |
| UTILITY REQUIREMENTS |
| SteamHeated Unit: Off |
| ElectricHeated Unit: Off |
| 200 to 208 V 60 Hz Steam: Off |
| 200 to 208 V 50 Hz Steam: Off |
| 380 to 400 V 60 Hz: Off |
| 380 to 415 V 50 Hz: Off |
| 460 to 480 V 60 Hz: Off |
| French FD355: Off |
| Spanish FD356: Off |
| German FD357: Off |
| Italian FD358: Off |
| AcuRinse Reservoir FD000054: Off |
| AcuWash Reservoir FD000055: Off |
| 40Amp Min Circuit Protection FD000056: Off |
| Printer FD094: Off |
| Seismic TieDown Kit FD354: Off |
| Condensate Return To Drain FD353: Off |
| TwoLevel Manifold Rack FD74800: Off |
| TwoLevel Manifold Rack with Equal Space: Off |
| ThreeLevel Manifold Rack FD74900: Off |
| FourLevel Manifold Rack FD75000: Off |
| FiveLevel Manifold Rack FD75200: Off |
| MultiFunction Rack for Large Items FD75500: Off |
| MultiFunction Rack for Small Items FD75600: Off |
| General Purpose Basket FD77100: Off |
| Self Docking Transfer Cart FD199: Off |
| Transfer Cart Adapter FD209: Off |
| Set of dry Contacts FD362: Off |
| Handheld Bar Code reader FD363: Off |
| Conductivity Probe FD350: Off |
| Additional Wall Flange FD351: Off |
| Additional Chemical Pump FD349: Off |
| Installation Kit Steam FD088: Off |
| Installation Kit Electric FD037: Off |
| Offset Installation Air Management Kit FD352: Off |
| IQOQPQ Protocols FD359: Off |
| Machine Identifier 11000866: Off |
| Drain AcuCoolDown FD361: Off |
| ConnectAssure Technology for Data Export: Off |
| ProConnect Technical Support Services: Off |
| Item: |
| undefined: |
| Locations: |
File details come from the government source that posted it. Updated .