This document is a comprehensive Medical Gas Annual Inspection Report for the Birmingham VA Medical Center (VAMC) conducted on July 15th-18th, 2024. The inspection covered multiple medical gas systems across the facility's ground through 8th floors, including oxygen, medical air, vacuum, nitrous oxide, nitrogen, and other critical medical gas infrastructure. The detailed report reveals an extensive evaluation of source equipment, master alarms, zone valves, patient outlets, and overall system performance, with a total of 1,152 medical gas outlets tested and a 97.7% pass rate.
Key findings include several corrective actions required to meet NFPA 99 2024 code standards, such as updating zone valve and area alarm labels, addressing medical air gauge installations, and resolving ventilation issues in the nitrous oxide and nitrogen manifold rooms. Notable technical observations include specific outlet leakage issues on various floors, particularly in operating rooms, critical care units, and patient care areas. While most systems passed inspection, the report highlights 26 outlet failures across different gas types and recommends systematic upgrades to ensure full compliance with the latest medical gas system safety regulations.
View the file
Other files
Show all 29
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PREPARED BY: BRAYDEN SEGER
JULY 15TH-18TH, 2024
BIRMINGHAM VAMC
MEDICAL GAS
ANNUAL
INSPECTION
700 19TH STREET SOUTH
BIRMINGHAM, AL 35233
ITEMS TESTED/INSPECTED
OXYGEN MEDICAL AIR
MEDICAL
VACUUM
NITROUS
OXIDE
NITROGEN
CARBON
DIOXIDE
COMP AIR WAGD
SOURCE EQUIPMENT P P P P P
MASTER ALARMS P P P P P
ZONE VALVES P P P P P P P
ZONE VERIFICATION P P P P P P P
CROSS CONNECTION P P P P P P P
AREA ALARMS P P P P P
PATIENT OUTLETS P P P P P P P
PARTICLE ANALYSIS P P P P
PURITY ANALYSIS P P P P
MGI MEDICAL GAS VERIFIER: BRAYDEN SEGER
SIGNATURE:
DATE: 7/15/2024
MEDICAL GAS ANNUAL TESTING ANALYSIS
MEDICAL GAS TYPE
SCOPE OF
TESTING/VERIFICATION
NFPA 99 2024 Compliant Medical Gas Annual Inspection for a Category 1 Medical Facility. Includes testing of all medical gas source equipment, medical gas zone valves, medical gas master alarm panels, medical gas area alarm panels, medical gas control panels, medical gas flexible terminals and medical gas terminals and medical gas labeling.
ADDITIONAL COMMENTS
EXECUTIVE SUMMARY
SIGNALS
PASSED
SIGNALS
FAILED
EXECUTIVE OVERALL SUMMARY-SOURCE
SOURCE EQUIPMENT
GAS TYPE SOURCE TYPE TOTAL COUNT TOTAL PASSED TOTAL FAILED
MEDICAL AIR RECIPROCATING 1 1 0
OXYGEN BULK OXYGEN 1 1 0
CLAWVACUUM 1 1 0
NITROUS OXIDE GAS X GAS MANIFOLD 1 0 1
NITROGEN GAS X GAS MANIFOLD 1 0 1
2 2 0
SOURCE EQUIPMENT MASTER ALARMS
EQUIPMENT
EQUIPMENT MAKE &
MODEL
ALARM PANEL
COUNT
ALARM SIGNAL
COUNT
TOTAL 7 5 2
MASTER ALARM PANEL ALARMS
MASTER ALARM PANEL AMICO ALERT 1 20 20 0
MASTER ALARM PANEL AMICO ALERT 1 20 20
TOTALS 2 40 40 0
PASSED FAILED PASSED FAILED
PATIENT OUTLET TERMINALS
EXECUTIVE OVERALL SUMMARY-PIPELINE
ZONE VALVES AREA ALARMS
TOTAL VALVES TOTAL ALARMS
66 65 1 42 42 0
MISSING VALVES MISSING ALARMS
0 0
OXYGEN 378 369 9 0
GAS TYPE TOTAL COUNT TOTAL PASSED TOTAL FAILED TOTAL UNAVALIBLE
MEDICAL AIR 255 250 5 0
VACUUM 423 411 12 0
NITROGEN 25 25 0 0
NITROUS OXIDE 21 21 0 0
WAGD 22 22 0 0
PERCENTAGE 97.7% 2.3% 0.0%
TOTAL 1152 1126 26 0
TOTAL PERCENTAGE OF
OUTLETS TESTED
100.0%
TOTAL OUTLET
FAILURES
TOTAL OUTLETS
REPAIRED
26 0
COMPRESSED AIR 28 28 0 0
MGI TECHNICIAN: BRAYDEN SEGER
1 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
RIGHT VACUUM
2 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
MEDICAL AIR
3 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
RIGHT OXYGEN
4 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
OXYGEN
5 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
OXYGEN
6 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
RIGHT SIDE RIGHT OXYGEN
7 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
LEFT SIDE MEDICAL AIR
8 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
LEFT SIDE BOTTOM RIGHT VACUUM
9 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
LEFT SIDE BOTTOM LEFT VACUUM
10 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
RIGHT SIDE RIGHT OXYGEN
11 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
RIGHT SIDE BOTTOM LEFT VACUUM
12 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
RIGHT SIDE BOTTOM RIGHT VACUUM
13 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
RIGHT SIDE LEFT OXYGEN
14 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
LEFT SIDE MEDICAL AIR
15 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
LEFT SIDE BOTTOM RIGHT VACUUM
16 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
RIGHT SIDE BOTTOM LEFT VACUUM
17 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
RIGHT SIDE BOTTOM LEFT VACUUM
18 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
LEFT SIDE BOTTOM RIGHT VACUUM
19 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
LEFT SIDE LEFT OXYGEN
20 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
OXYGEN
21 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
2ND VACUUM
22 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
1ST MEDICAL AIR
23 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
2ND VACUUM
24 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
LEFT OXYGEN
25 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
3RD VACUUM
26 SPECIFIC OUTLET REPAIRED BY REPAIR DATE
2ND MEDICAL AIR
ROOM OUTLET SPECS ISSUE
6455 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
6454 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
6454 BEACON - DC - CH LEAKS W/ ADAPTER INSERTED
ROOM OUTLET SPECS ISSUE
6449 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
6449 BEACON - DC - CH OUTLET DAMAGED
4111 ALLIED - 500 - CH LEAKS W/ ADAPTER INSERTED
ROOM OUTLET SPECS
ROOM OUTLET SPECS ISSUE
4336 BED C BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
4336 BED B BEACON - DC - CH OUTLET DAMAGED
ISSUE
4206 ALLIED - 500 - CH LEAKS W/ AND W/OUT ADAPTER INSERTED
ROOM OUTLET SPECS ISSUE
ROOM OUTLET SPECS ISSUE
4110 ALLIED - 500 - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
OUTLET ISSUES BIRMINGHAM VAMC
ROOM OUTLET SPECS ISSUE
3106 PROCEDURE 6
WALL
ALLIED - 500 - CH LEAKS W/OUT ADAPTER INSERTED
4207 ALLIED - 500 - CH LEAKS W/ AND W/OUT ADAPTER INSERTED
ROOM OUTLET SPECS ISSUE
4329 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
4333 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
4333 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
4334 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
4334 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
6447 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS
END OF OUTLET ISSUES
4334 BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
ROOM OUTLET SPECS ISSUE
4336 BED B BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
4336 BED B BEACON - DC - CH OUTLET DAMAGED
ISSUE
4336 BED D BEACON - DC - CH OUTLET DAMAGED
ROOM OUTLET SPECS ISSUE
ISSUE
4336 BED D BEACON - DC - CH
ROOM OUTLET SPECS
LEAKS W/OUT ADAPTER INSERTED
ROOM OUTLET SPECS
ISSUE
STORAGE 6317 ALLIED - 500 - CH LEAKS W/ AND W/OUT ADAPTER INSERTED
ROOM OUTLET SPECS ISSUE
4336 BED A BEACON - DC - CH OUTLET DAMAGED
4334 BEACON - DC - CH OUTLET DAMAGED
BIRMINGHAM VAMC
CORRECTED
CORRECTED
CORRECTED
NOTE: CORRECTIVE ACTIONS ARE COMPONENTS OF THE MEDICAL GAS SYSTEM THAT DO NOT MEET THE NFPA 99 2024 CODE AND SHOULD BE UPGRADED OR REPAIRED TO MEET THE
NFPA 99 2024 CODE.
4435 OR-1
THE MEDICAL AIR ZONE VALVE GAUGE READING
DOES NOT FALL IN THE MIDDLE THIRD OF THE
GAUGE.
The scale range of positive pressure analog indicators shall be such that the normal operating pressure is within the middle third of the total range.
(NFPA 99 2024 5.1.8.1.3)
CORRECTIVE
ACTION
INSTALL A 0-100 PSI GAUGE.
CORRECTIVE
ACTION
UPDATE THE LABELS ON THE AREA ALARMS. SEE INDIVIDUAL AREA ALARM PAGES FOR SPECIFIC AREA ALARMS THAT NEED UPDATED.
AREA ISSUE
CORRECTIVE ACTIONS
AREA ISSUE NFPA 99 2024 CODE
ZONE VALVES
MULTIPLE ZONE VALVES NEED LABELING UPDATED
TO REFLECT THE ROOMS THE SERVE.
Zone valves shall be labeled in accordance with
5.1.11.2. (NFPA 99 2024 5.1.4.1.3)
CORRECTIVE
ACTION
UPDATE THE LABELS ON THE ZONE VALVES. SEE INDIVIDUAL ZONE VALVE PAGES FOR SPECIFIC ZONE VALVES THAT NEED UPDATED.
NFPA 99 2024 CODE
AREA ISSUE NFPA 99 2024 CODE
AREA ALARMS
MULTIPLE AREA ALARMS NEED LABELING UPDATED
TO REFLECT THE ROOMS THEY MONITOR.
Area alarm systems used for medical gas and vacuum systems shall include the following:
(NFPA 99 2024 5.1.9.1)
(7) Labeling of each alarm panel for its area of surveillance.
NOTE: THESE ARE COMPONENTS OF THE MEDICAL GAS SYSTEM THAT DO NOT MEET THE 2024 NFPA 99 CODE, HOWEVER THEY ARE PERMITTED TO
REMAIN IN USE UNTIL THE AREA(S) ARE UPGRADED OR RENOVATED DUE TO THE ABOVE NOT AFFECTING PATIENT CARE.
END OF FUTURE CORRECTIONS
INSTALL MECHANICAL VENTILATION TO GET THE APPROPRIATE AIR FLOW.
FUTURE CORRECTIONS
AREA ISSUE NFPA 99 2024 CODE
NITROUS OXIDE / NITROGEN MANIFOLD
ROOM
THERE IS NO VENTILATION TO CIRCULATE AIR FLOW.
Indoor storage or manifold areas and storage or manifold buildings for medical gases and cryogenic fluids shall be provided with natural ventilation or mechanical exhaust ventilation in accordance with
9.3.6.5.1 through 9.3.6.8.
(NFPA 99 2024 9.3.6.5)
FUTURE
CORRECTION
MEDICAL GAS MASTER
ALARM ANALYSIS
MASTER ALARM CODE REQUIREMENTS
Master alarm panels for medical gas and vacuum systems shall each include the following signals: (NFPA
99 2024 5.1.9.2.4)
(1) Alarm indication when, or just before, changeover occurs in a medical gas system that is supplied by a manifold or other alternating-type bulk system that has as a part of its normal operation a changeover from one portion of the operating supply to another.
(2) Alarm indication for a bulk cryogenic liquid system when the main supply reaches an average day’s supply, indicating low contents.
(3) Alarm indication when, or just before, the changeover to the reserve supply occurs in a medical gas system that consists of one or more units that continuously supply the piping system while another unit remains as the reserve supply and operates only in the case of an emergency.
(4) Alarm indication for cylinder reserve pressure low when the content of a cylinder reserve header is reduced below one average day’s supply.
(5) For bulk cryogenic liquid systems, an alarm when or at a predetermined set point before the reserve supply contents fall to one day’s average supply, indicating reserve low.
(6) Where a cryogenic liquid storage unit is used as a reserve for a bulk supply system, an alarm indication when the gas pressure available in the reserve unit is below that required for the medical gas system to function.
(7) Alarm indication when the pressure in the main line of each separate medical gas system increases 20 percent or decreases 20 percent from the normal operating pressure.
(8) Alarm indication when the medical–surgical vacuum pressure in the main line of each vacuum system drops.
(9) Alarm indication(s) from the local alarm panel(s) as described in 5.1.9.5.2 to indicate when one or more of the conditions being monitored at a site is in alarm.
(10) Medical air dew point high alarm from each compressor site to indicate when the line pressure dew point is greated than +2°C (+35°F)
(11) WAGD low alarm when the WAGD vacuum level or flow is below effective operating limits.
(12) An instrument air dew point high alarm from each compressor site to indicate when the line pressure dew point is greater than −30°C (−22°F).
(13) Alarm indication if the primary or reserve production stops on a proportioning system.
BIRMINGHAM VAMC MGI TECHNICIAN: BRAYDEN SEGER
BIRMINGHAM, AL 35233 TESTING DATE: 7/15/2024
VISUAL
STATUS
AUDIBLE
STATUS
LABELED
CORRECTLY
COMMENTS
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
ALARM MAKE AND MODEL: CALL CENTER
COMMENTS
MEDICAL AIR COMP. HIGH TEMP. SHUTDOWN PASS
OXYGEN LINE PRESSURE HIGH PASS
OXYGEN LINE PRESSURE LOW PASS
MEDICAL AIR LAG COMPRESSOR IN USE PASS
OXYGEN SECONDARY LIQUID SUPPLY IN USE PASS
OXYGEN SECONDARY LIQUID LEVEL LOW PASS
OXYGEN SECONDARY LIQUID LOW HEAD PRESSURE PASS
MASTER ALARM ANALYSIS
MASTER ALARM SIGNALS PASS/FAIL
OXYGEN PRIMARY LIQUID LEVEL LOW PASS
MEDICAL AIR HIGH DEW POINT PASS
MEDICAL AIR CO HIGH PASS
MEDICAL AIR LINE PRESSURE HIGH PASS
MEDICAL AIR LINE PRESSURE LOW PASS
NITROGEN LINE PRESSURE LOW PASS
NITROGEN RESERVE IN USE PASS
NITROGEN LINE PRESSURE HIGH PASS
NITROUS OXIDE RESERVE IN USE PASS
MEDICAL VACUUM PUMPS MAINT. REQUIRED PASS
MEDICAL VACUUM LINE PRESSURE LOW PASS
NITROUS OXIDE LINE PRESSURE HIGH PASS
NITROUS OXIDE LINE PRESSURE LOW PASS
AMICO ALERT LOCATION:
PER NFPA 99 2024 WHERE IS THE MASTER ALARM LOCATED OF THE FOLLOWING LOCATIONS LOCATION
(1) One master alarm panel shall be located in the office or work space of the on-site individual responsible for the maintenance of the medical gas and vacuum piping systems.
(2) In order to ensure continuous surveillance of the medical gas and vacuum systems while the facility is in operation, the second master alarm panel shall be located in an area of continuous observation (e.g., the telephone switchboard, security office, or other continuously staffed location).
PASS/FAIL
PASS
BIRMINGHAM VAMC MGI TECHNICIAN: BRAYDEN SEGER
BIRMINGHAM, AL 35233 TESTING DATE: 7/15/2024
VISUAL
STATUS
AUDIBLE
STATUS
LABELED
CORRECTLY
COMMENTS
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
P P P
ALARM MAKE AND MODEL: EMERGENCY DEPARTMENT
COMMENTSPER NFPA 99 2024 WHERE IS THE MASTER ALARM LOCATED OF THE FOLLOWING LOCATIONS LOCATION
(1) One master alarm panel shall be located in the office or work space of the on-site individual responsible for the maintenance of the medical gas and vacuum piping systems.
(2) In order to ensure continuous surveillance of the medical gas and vacuum systems while the facility is in operation, the second master alarm panel shall be located in an area of continuous observation (e.g., the telephone switchboard, security office, or other continuously staffed location).
PASS/FAIL
PASS
AMICO ALERT LOCATION:
NITROUS OXIDE RESERVE IN USE PASS
NITROUS OXIDE LINE PRESSURE HIGH PASS
NITROUS OXIDE LINE PRESSURE LOW PASS
NITROGEN RESERVE IN USE PASS
NITROGEN LINE PRESSURE HIGH PASS
NITROGEN LINE PRESSURE LOW PASS
MEDICAL AIR LINE PRESSURE LOW PASS
MEDICAL VACUUM PUMPS MAINT. REQUIRED PASS
MEDICAL VACUUM LINE PRESSURE LOW PASS
MEDICAL AIR HIGH DEW POINT PASS
MEDICAL AIR CO HIGH PASS
MEDICAL AIR LINE PRESSURE HIGH PASS
OXYGEN LINE PRESSURE LOW PASS
MEDICAL AIR LAG COMPRESSOR IN USE PASS
MEDICAL AIR COMP. HIGH TEMP. SHUTDOWN PASS
OXYGEN SECONDARY LIQUID LEVEL LOW PASS
OXYGEN SECONDARY LIQUID LOW HEAD PRESSURE PASS
OXYGEN LINE PRESSURE HIGH PASS
OXYGEN SECONDARY LIQUID SUPPLY IN USE PASS
MASTER ALARM ANALYSIS
MASTER ALARM SIGNALS PASS/FAIL
OXYGEN PRIMARY LIQUID LEVEL LOW PASS
MEDICAL GAS AREA
ALARM ANALYSIS
(2) *Critical Care Areas
AREA ALARM CODE REQUIREMENTS
(1) Separate visual indicators for each condition monitored, except as permitted in 5.1.9.5.2 for local alarms that are displayed on master alarm panels.
(2) Visual indicators that remain in alarm until the situation that has caused the alarm is resolved.
(3) Cancelable audible indication of each alarm condition that produces a sound with a minimum level of
80 dBA at 0.92m (3 ft)
(4) Means to visually indicate a lamp or LED failure
Area alarm systems used for medical gas and vacuum systems shall include the following:
(NFPA 99 2024 5.1.9.1)
(2)*Anesthetizing locations where moderate sedation, deep sedation, or general anesthesia is administered shall have the sensors installed either on the source side of any of the individual room zone valve box assemblies or on the patient or use side of each of the individual zone valve box assemblies.
(5) Visual and audible indication that the communication with an alarm-initiating devices is disconnected.
(6) Labeling of each indicator, indication the condition monitored.
(7) Labeling of each alarm panel for its area of surveillance.
(8) Reinitiating of the audible signal if another alarm condition occurs while the audible alarm is silenced.
(9) Power for master, area alarms, sensors, and switches from the life safety branch of the emergency electrical system as described in Chapter 6.
Area Alarms. Area alarm panels shall be provided to monitor all medical gas, medical–surgical vacuum, and piped WAGD systems supplying the following: (NFPA 99 2024 5.1.9.4)
(1) Anesthetizing locations where moderate sedation, deep sedation, or general anesthesia is administered.
Area alarm panels for medical gas systems shall indicate if the pressure in the lines in the area being monitored increases or decreases by 20 percent from the normal line pressure. (NFPA 99 2024 5.1.9.4.2)
Area alarm panels for medical–surgical vacuum systems shall indicate if the vacuum in the area drops to or below 300 mm (12 in.) gauge HgV. (NFPA 99 2024 5.1.9.4.3)
Alarm sensors for area alarms shall be located as follows: (NFPA 99 2024 5.1.9.4.4)
(1)*Critical care areas shall have the alarm sensors installed on the patient or use side of each individual zone valve box assemblies.
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 53 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 20" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: G ALARM LOCATION:
AREA ALARM ANALYSIS
GROUND FLOOR
OUTSIDE G109R ALARM MAKE AND MODEL: AMICO ALERT
END OF GROUND FLOOR AREA ALARM ANALYSIS
AREA MONITORED: COMMENTS
G109R, G109M, G109N, G109L
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 50 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 1ST ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 53 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 47 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 17" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 1ST ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 53 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 1ST ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 55 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 47 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 20" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 1ST ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 19" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 1ST ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
N2O 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 22" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 1ST ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 1ST ALARM LOCATION:
DENTAL CHECK IN ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
DENTAL ROOMS 1615, 1616,
1620, 1621, 1622, 1623, 1624, 1626, 1627, 1628, 1629
DENTAL CHECK IN ALARM MAKE AND MODEL: AMICO ALERT
DENTAL ROOMS 1610, 1611,
1612, 1613, 1614
NURSES STATION ALARM MAKE AND MODEL: CHEMETRON DIGITAL ALARM 2
AREA MONITORED: COMMENTS
DIALYSIS BAYS 1, 2, 3, 4, 5, 6, 7,
8, 9, 10, EXAM, TREATMENT,
EQUIPMENT REPROCESSING
OUTSIDE DIALYSIS TREATMENT ROOM ALARM MAKE AND MODEL: POWEREX MED TOUCH
AREA MONITORED: COMMENTS
AREA MONITORED: COMMENTS
ACROSS 1435L ALARM MAKE AND MODEL: CHEMETRON DIGITAL ALARM 2
AREA MONITORED: COMMENTS
END OF 1ST FLOOR AREA ALARM ANALYSIS
AREA ALARM ANALYSIS
1ST FLOOR
AREA MONITORED: COMMENTS
ER ROOMS 1435K, 1435L,
1435M, 1435P, 1435O
OLD ER ROOMS 14, 15, 16, 17,
ER NURSES STATION ALARM MAKE AND MODEL: CHEMETRON DIGITAL ALARM 2
AREA MONITORED: COMMENTS
ER ROOM 1435I
ONCOLOGY NURSES STATION ALARM MAKE AND MODEL: AMICO ALERT
ONCOLOGY BAYS 1, 2, 3, 4, 5, 6,
7, 8, 9, 10
AREA ALARM NEEDS
LABELED TO REFLECT
ROOMS SERVED
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 55 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 22" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 2ND ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 55 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 19" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 2ND ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 56 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 22" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 2ND ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR N/A PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 18" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 2ND ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 56 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 20" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 2ND ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 55 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
FLOOR: 2ND ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 55 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 19" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 2ND ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 53 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 2ND ALARM LOCATION:
END OF 2ND FLOOR AREA ALARM ANALYSIS
AREA MONITORED: COMMENTS
2367 X-RAY 6, 2365 X-RAY 4
ALARM MAKE AND MODEL: AMICO
ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
2603A, 2603B, 2606, 2607, 2608, 2611, 2612, 2613, 2615, 2616, 2617, 2618
AREA ALARM NEEDS
LABELED TO REFLECT
ROOMS SERVED
DENTAL CHECK IN
AREA MONITORED: COMMENTS
OUTSIDE 2343
AREA MONITORED: COMMENTS
NOT IN USE
ACROSS 2345 ALARM MAKE AND MODEL: CHEMETRON MODULAR
AREA MONITORED: COMMENTS
MRI 2115D, 2115
OUTSIDE 2370 ALARM MAKE AND MODEL: AMICO ALERT
INSIDE MRI SUITE ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
OUTSIDE 2121 ALARM MAKE AND MODEL: AMICO ALERT
ROOM 2121, 2116, MRI
HOLDING
AREA MONITORED: COMMENTS
CT 2123
OUTSIDE 2123 ALARM MAKE AND MODEL: AMICO ALERT
OUTSIDE 2125 ALARM MAKE AND MODEL: AMICO ALERT
AREA ALARM ANALYSIS
2ND FLOOR
AREA MONITORED: COMMENTS
CT/GANTY 2125
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 19" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 20" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 55 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 19" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 53 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 22" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 55 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 47 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 19" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
AREA MONITORED: COMMENTS
3A WING ROOMS 3106,
3104, 3102, 3101
ACROSS 3102 ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
3501, 3502, 3503, 3504, 3505, 3506, 3507, 3508, 3509, 3510, 3511, 3512, AMBULATORY NURSES STATION ALARM MAKE AND MODEL: AMICO ALERT
AMBULATORY NURSES STATION ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
AREA ALARM ANALYSIS
3RD FLOOR
AREA MONITORED: COMMENTS
SPECIAL PROCEDURE 3528
OUTSIDE 3811 ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
OUTSIDE 3810 ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
OUTSIDE 3809 ALARM MAKE AND MODEL: AMICO ALERT
AREA ALARM ANALYSIS
3RD FLOOR
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 46 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 53 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 19" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 53 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 49 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 22" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 55 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 50 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 19" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 49 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 3RD ALARM LOCATION:
END OF 3RD FLOOR AREA ALARM ANALYSIS
OUTSIDE 3808 ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
3114 PROCEDURE ROOM 1
AREA MONITORED: COMMENTS
3117 PROCEDURE ROOM 2,
3116 SCOPE CLEAN
OUTSIDE 3117 ALARM MAKE AND MODEL:
AREA MONITORED: COMMENTS
OUTSIDE 3114 ALARM MAKE AND MODEL: AMICO ALERT
AMICO ALERT
AREA MONITORED: COMMENTS
3119 PROCEDURE ROOM 3,
3120 SCOPE REPROCESSING
OUTSIDE 3119 ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
3121 PROCEDURE ROOM 4
OUTSIDE 3121 ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
3122 PROCEDURE ROOM 5
OUTSIDE 3122 ALARM MAKE AND MODEL: AMICO ALERT
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 55 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 22" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 4TH ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 47 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 20" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 4TH ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 52 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 49 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 20" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 4TH ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 52 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 4TH ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 53 PSI PASS 60 PSI 40 PSI PASS PASS MAINLINE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS MAINLINE P
VACUUM 18" HG PASS N/A 12" HG PASS PASS MAINLINE P
N2O 49 PSI PASS 60 PSI 40 PSI PASS PASS MAINLINE P
N2 147 PSI PASS 195 PSI 140 PSI PASS PASS MAINLINE P
FLOOR: 4TH ALARM LOCATION:
PACU NURSES STATION ALARM MAKE AND MODEL: CHEMETRON MODULAR
END OF 4TH FLOOR AREA ALARM ANALYSIS
SICU NURSES STATION ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
PACU BAYS 1, 2, 3, 4, 5, 6,
ISOLATION
AREA MONITORED: COMMENTS
OR'S 1, 2, 3, 4, 5, 6, 7, 8, 9,
10, 11
ACROSS PRE-OP
ALARM NEEDS LABELED
TO REFLECT ROOMS
SERVED
4B NURSES STATION ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
ALARM MAKE AND MODEL: AMICO ALERT
AREA ALARM ANALYSIS
4TH FLOOR
AREA MONITORED: COMMENTS
4A WING ROOMS 4102,
4103, 4104, 4110, 4111, 4112, 4113, 4119, 4120, 4121, 4122, 4123, 4124, 4127, 4128
ALARM NEEDS LABELED
TO REFLECT ROOMS
SERVED
SICU ROOMS 4329, 4330,
4331, 4332, 4333, 4334, 4335, 4336
ALARM NEEDS LABELED
TO REFLECT ROOMS
SERVED
4A NURSES STATION ALARM MAKE AND MODEL: CHEMETRON DIGITAL ALARM 2
AREA MONITORED: COMMENTS
4B WING ROOMS 4202,
4203, 4206, 4207, 4208, 4210, 4211, 4217, 4218, 4219, 4220, 4226, 4227, GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 52 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 48 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 5TH ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN N/A N/A N/A N/A N/A N/A N/A
MED AIR N/A N/A N/A N/A N/A N/A N/A
VACUUM N/A N/A N/A N/A N/A N/A N/A
FLOOR: N/A ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 50 PSI PASS 60 PSI 40 PSI PASS PASS ZONE
MED AIR 50 PSI PASS 60 PSI 40 PSI PASS PASS ZONE
VACUUM 20" HG PASS N/A 12" HG PASS PASS ZONE
FLOOR: 5TH ALARM LOCATION:
AREA MONITORED: COMMENTS
END OF 5TH FLOOR AREA ALARM ANALYSIS
NOT IN USE
ALARM NO LONGER IN
USE
ACROSS 5312 ALARM MAKE AND MODEL: CHEMETRON MODULAR
AREA MONITORED: COMMENTS
B-WING
UNIT IS UNDER
CONSTRUCTION
B-WING ALARM MAKE AND MODEL: N/A
5A NURSES STATION ALARM MAKE AND MODEL: AMICO ALERT
AREA ALARM ANALYSIS
5TH FLOOR
AREA MONITORED: COMMENTS
A-WING ROOMS 5102,
5103, 5104, 5110, 5111, 5112, 5113, 5119, 5120, 5121, 5122, 5123, 5124, ALARM NEEDS LABELED
TO REFLECT ROOMS
SERVED
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 50 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 20" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 6TH ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 49 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 22" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 6TH ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 49 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 6TH ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 52 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 47 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 17" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 6TH ALARM LOCATION:
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
MED AIR 46 PSI PASS 60 PSI 40 PSI PASS PASS ZONE P
VACUUM 21" HG PASS N/A 12" HG PASS PASS ZONE P
FLOOR: 6TH ALARM LOCATION:
6B WING ROOMS 6204,
6206, 6207, 6208, 6210, 6211, 6217, 6218, 6219, 6220, 6226, 6227, 6228
ALARM NEEDS LABELED
TO REFLECT ROOMS
SERVED
6B NURSES STATION ALARM MAKE AND MODEL: AMICO ALERT
ALARM NEEDS LABELED
TO REFLECT ROOMS
SERVED
6A NURSES STATION ALARM MAKE AND MODEL: AMICO ALERT
AREA MONITORED: COMMENTS
END OF 6TH FLOOR AREA ALARM ANALYSIS
AREA MONITORED: COMMENTS
CCU 6407, 6408, 6409, 6410,
6411, 6414, 6415, 6416
ALARM NEEDS LABELED
TO REFLECT ROOMS
SERVED
AREA MONITORED: COMMENTS
CATH LAB 6430, 6432, CATH
LAB HOLDING, GXT, ECHO
1, ECHO EXAM
ALARM NEEDS LABELED
TO REFLECT ROOMS
SERVED
ACROSS 6424 ALARM MAKE AND MODEL: CHEMETRON MODULAR
AREA MONITORED:
AREA MONITORED: COMMENTS
6A WING ROOMS 6105,
6111, 6114, 6116, 6119, 6124, 6125, 6128, 6130, 6132, 6133, 6136
MICU NURSES STATION ALARM MAKE AND MODEL: CHEMETRON MODULAR
AREA ALARM ANALYSIS
6TH FLOOR
COMMENTS
MICU ROOMS 6447, 6448,
6449, 6450, 6451, 6453, 6454, 6455, 6456, 6459
ALARM NEEDS LABELED
TO REFLECT ROOMS
SERVED
CCU NURSES STATION ALARM MAKE AND MODEL: CHEMETRON MODULAR
GAS MONITORED
PRESSURE
READING
LABELING
CORRECT
HIGH
ALARM
SETPOINT
LOW ALARM
SETPOINT
VISUAL
ALARM
FUNCTION
AUDIBLE
ALARM
FUNCTION
MAINLINE
OR ZONE
SENSOR
PASS FAIL
OXYGEN 54 PSI PASS 60 PSI 40 PSI PASS PASS MAINLINE P
VACUUM 20" HG PASS N/A 12" HG PASS PASS MAINLINE P
FLOOR: 7TH ALARM LOCATION:
END OF 7TH FLOOR AREA ALARM ANALYSIS
AREA ALARM ANALYSIS
7TH FLOOR
AREA MONITORED: COMMENTS
SLEEP LAB 7119, 7120, 7121,
7123, 7124, 7125
ACROSS 7118 ALARM MAKE AND MODEL: AMICO ALERT
MEDICAL GAS
ZONE VALVE ANALYSIS
Zone valves shall be readily operable from a standing position in the corridor on the same floor they serve. (NFPA 99 2024 5.1.4.6.1)
ZONE VALVE CODE REQUIREMENTS
Zone Valves. All station outlets/inlets shall be supplied through a zone valve as follows:
(NFPA 99 2024 5.1.4.6.1)
(2) The zone valves shall serve only outlets/inlets located on that same story.
(3) The zone valve shall not be located in a room with station outlets/inlets that it controls.
(1) The zone valve shall be placed such that a wall intervenes between the valve and outlets/inlets that it controls.
Zone valves shall be labeled in accordance with 5.1.11.2. (NFPA 99 2024 5.1.4.1.3)
Zone valves shall be so arranged that shutting off the supply of medical gas or vacuum to one zone will not affect the supply of medical gas or vacuum to another zone or the rest of the system.
(NFPA 99 2024 5.1.4.6.3)
A pressure/vacuum indicator shall be provided on the station outlet/inlet side of each zone valve.
(NFPA 99 2024 5.1.4.6.4)
Zone valve boxes shall not be installed behind normally open or normally closed doors or otherwise hidden from plain view. (NFPA 99 2024 5.1.4.6.1)
Zone valve boxes shall not be located in closed or locked rooms, areas, or closets. (NFPA 99 2024 5.1.4.6.1)
Zone valves shall be so arranged that shutting off the supply of gas to any one operating room or anesthetizing location will not affect the others. (NFPA 99 2024 5.1.4.6.3)
Zone valve boxes shall be installed where they are visible and accessible at all times.
(NFPA 99 2024 5.1.4.6.1 )
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 22" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: G VALVE LOCATION:
END OF GROUND FLOOR ZONE VALVE ANALYSIS
OUTSIDE G109R VALVE MANUFACTURER: AMICO
ZONE VALVE ANALYSIS
GROUND FLOOR
AREA SERVED: COMMENTS
G109R, G109M, G109N, G109L
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 50 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 21" HG PASS PASS 1-1/2" BALL PASS PASS P
FLOOR: 1ST VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR 59 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 20" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 1ST VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 55 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 22" HG PASS PASS 1" BALL PASS PASS P
FLOOR: 1ST VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 55 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 49 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 20" HG PASS PASS 1-1/4" BALL PASS PASS P
FLOOR: 1ST VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 55 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 48 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 22" HG PASS PASS 1-1/4" BALL PASS PASS P
FLOOR: 1ST VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
N2O 49 PSI PASS PASS 1/2" BALL PASS PASS P
FLOOR: 1ST VALVE LOCATION:
NURSES STATION VALVE MANUFACTURER: CHEMETRON
AREA SERVED: COMMENTS
DIALYSIS BAYS 1, 2, 3, 4, 5, 6, 7, 8, 9, 10,
EXAM, TREATMENT, EQUIPMENT
REPROCESSING
OUTSIDE DIALYSIS VALVE MANUFACTURER: POWEREX
AREA SERVED: COMMENTS
OLD ER ROOMS 14, 15, 16, 17, 18
ZONE NOT IN USE
ZONE NEEDS HANDLES
INSTALLED OR
DEPRESSURIZED
ER NURSES STATION VALVE MANUFACTURER: CHEMETRON
AREA SERVED: COMMENTS
ER ROOM 1435I
MEDICAL AIR GAUGE IS
NOT ACCURATE
ACROSS 1435L VALVE MANUFACTURER: CHEMETRON
AREA SERVED: COMMENTS
ZONE VALVE ANALYSIS
1ST FLOOR
AREA SERVED: COMMENTS
ER ROOMS 1435K, 1435L, 1435M, 1435P,
1435O
ONCOLOGY BAYS 1, 2, 3, 4, 5, 6, 7, 8, 9,
DENTAL ROOMS 1610, 1611, 1612, 1613
AREA SERVED:
ZONE VALVE NEEDS
LABELED TO REFLECT
ROOM NUMBERS
SERVED
OUTSIDE DENTAL CLINIC VALVE MANUFACTURER: AMICO
OUTSIDE ONCOLOGY VALVE MANUFACTURER: AMICO
COMMENTS
ZONE VALVE ANALYSIS
1ST FLOOR
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
COMP AIR 4 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 20" HG PASS PASS 1" BALL PASS PASS P
FLOOR: 1ST VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
COMP AIR 7 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 21" HG PASS PASS 1-1/4" BALL PASS PASS P
FLOOR: 1ST VALVE LOCATION:
COMMENTS
END OF 1ST FLOOR ZONE VALVE ANALYSIS
AREA SERVED: COMMENTS
ACROSS 1616 VALVE MANUFACTURER: AMICO
AMICO
DENTAL ROOMS 1615, 1616, 1620, 1621,
1622, 1623, 1624, 1626, 1627, 1628, 1629
DENTAL ROOMS 1610, 1611, 1612, 1613,
ACROSS 1613 VALVE MANUFACTURER:
AREA SERVED:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR N/A PASS PASS 1/2" BALL PASS PASS P
VACUUM 21" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 22" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 21" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 58 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR N/A PASS PASS 1/2" BALL PASS PASS P
VACUUM 21" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 19" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN N/A PASS PASS 1/2" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
AREA SERVED: COMMENTS
MRI 2115D, 2115 MEDICAL AIR NOT IN USE
INSIDE MRI SUITE VALVE MANUFACTURER: AMICO
OUTSIDE RADIOLOGY CHECK IN VALVE MANUFACTURER: AMICO
2367 X-RAY 6, 2365 X-RAY 4
ZONE VALVE ANALYSIS
2ND FLOOR
AREA SERVED: COMMENTS
CT/GANTY2125 MEDICAL AIR NOT IN USE
OUTSIDE 2125 VALVE MANUFACTURER: AMICO
2116, 2121, MRI HOLDING
AREA SERVED: COMMENTS
OUTSIDE 2121 VALVE MANUFACTURER: POWEREX
AREA SERVED: COMMENTS
CT 2123
AREA SERVED: COMMENTS
OUTSIDE 2123 VALVE MANUFACTURER: AMICO
OUTSIDE 2370 VALVE MANUFACTURER: AMICO
AREA SERVED: COMMENTS
2340A, 2341
AREA SERVED: COMMENTS
NOT IN USE
ACROSS 2345 VALVE MANUFACTURER: CHEMETRON
ZONE VALVE ANALYSIS
2ND FLOOR
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 21" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 55 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 21" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 0 PSI PASS PASS 1/2" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 55 PSI PASS PASS 1" BALL PASS PASS P
COMP AIR 6 PSI PASS PASS 1" BALL PASS PASS P
VACUUM 22" HG PASS PASS 1-1/2" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
COMP AIR 6 PSI PASS PASS 1" BALL PASS PASS P
FLOOR: 2ND VALVE LOCATION:
AREA SERVED: COMMENTS
ACROSS 2615 VALVE MANUFACTURER: AMICO
ACROSS 2352A VALVE MANUFACTURER: POWEREX
AREA SERVED: COMMENTS
2344, 2345, 2345B, 2346
AREA SERVED: COMMENTS
OUTSIDE 2343 VALVE MANUFACTURER: AMICO
AREA SERVED: COMMENTS
VALVE MANUFACTURER:
END OF 2ND FLOOR ZONE VALVE ANALYSIS
OUTSIDE 2345
NOT IN USE
AMICO
2603A, 2603B, 2606, 2607, 2608, 2611, 2612, 2613, 2615, 2616, 2617, 2618
ZONE VALVE NEEDS
LABELED TO REFLECT
ROOMS SERVED
AREA SERVED: COMMENTS
UNKNOWN
ZONE VALVE IS FOR HOSE
BARB OUTLET CONNECTION
ACROSS 2615 VALVE MANUFACTURER: AMICO
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 48 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 19" HG PASS PASS 1-1/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR 48 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 19" HG PASS PASS 1-1/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 50 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 22" HG PASS PASS 1-1/2" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 53 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR 48 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 22" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 55 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR 50 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 20" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR 48 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 22" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
AREA SERVED: COMMENTS
3A WING ROOMS 3106, 3104, 3102, 3101
ZONE VALVE NEEDS
LABELED TO REFLECT
ROOMS SERVED.
ACROSS 3102 VALVE MANUFACTURER: CHEMETRON
AREA SERVED: COMMENTS
ZONE VALVE ANALYSIS
3RD FLOOR
AREA SERVED: COMMENTS
3501, 3502, 3503, 3504, 3505, 3506, 3507, 3508, 3509, 3510, 3511, 3512, 3513
OUTSIDE AMBULATORY UNIT VALVE MANUFACTURER: AMICO
AREA SERVED: COMMENTS
SPECIAL PROCEDURE 3528
OUTSIDE ROOM 3528 VALVE MANUFACTURER: AMICO
OUTSIDE 3810 VALVE MANUFACTURER: AMICO
OUTSIDE 3811 VALVE MANUFACTURER: AMICO
AREA SERVED: COMMENTS
AREA SERVED: COMMENTS
OUTSIDE 3809 VALVE MANUFACTURER: AMICO
ZONE VALVE ANALYSIS
3RD FLOOR
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 49 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 21" HG PASS PASS 1" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
VACUUM 21" HG PASS PASS 1" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 49 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 20" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 55 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 49 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 21" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 50 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 22" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
AREA SERVED: COMMENTS
AREA SERVED: COMMENTS
ENT 3207, 3208, 3209, 3212, 3222, 3223,
3224, 3225, 3226, 3227
OUTSIDE 3231 VALVE MANUFACTURER: POWEREX
3119 PROCEDURE ROOM 3, 3120 SCOPE
REPROCESSING
SCOPE ROOM NEEDS
ADDED TO THE ZONE
VALVE LABEL
OUTSIDE 3119 VALVE MANUFACTURER: AMICO
OUTSIDE 3808 VALVE MANUFACTURER: AMICO
AREA SERVED: COMMENTS
3114 PROCEDURE ROOM 1
OUTSIDE OF 3114 VALVE MANUFACTURER: AMICO
AREA SERVED: COMMENTS
3117 PROCEDURE ROOM 2, 3116 SCOPE
CLEAN
SCOPE ROOM NEEDS
ADDED TO THE ZONE
VALVE LABEL
OUTSIDE 3117 VALVE MANUFACTURER: AMICO
ZONE VALVE ANALYSIS
3RD FLOOR
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 49 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 22" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 48 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 19" HG PASS PASS 3/4" BALL PASS PASS P
FLOOR: 3RD VALVE LOCATION:
END OF 3RD FLOOR ZONE VALVE ANALYSIS
AREA SERVED: COMMENTS
OUTSIDE 3121 VALVE MANUFACTURER: AMICO
3121 PROCEDURE ROOM 4
3122 PROCEDURE ROOM 5
OUTSIDE 3122 VALVE MANUFACTURER: AMICO
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 1" BALL PASS PASS P
MED AIR 50 PSI PASS PASS 1" BALL PASS PASS P
VACUUM 22" HG PASS PASS 1-1/2" BALL PASS PASS P
FLOOR: 4TH VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1" BALL PASS PASS P
MED AIR 49 PSI PASS PASS 1" BALL PASS PASS P
VACUUM 21" HG PASS PASS 1-1/2" BALL PASS PASS P
FLOOR: 4TH VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 53 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 48 PSI PASS PASS 3/4" BALL PASS PASS P
VACUUM 21" HG PASS PASS 2" BALL PASS PASS P
FLOOR: 4TH VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 75 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR 50 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 22" HG PASS PASS 1" BALL PASS PASS P
FLOOR: 4TH VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 3/4" BALL PASS PASS P
MED AIR 48 PSI PASS PASS 3/4" BALL PASS PASS P
N2O 50 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 22" HG PASS PASS 2" BALL PASS PASS P
FLOOR: 4TH VALVE LOCATION:
AREA SERVED: COMMENTS
SICU ROOMS 4329, 4330, 4331, 4332,
4333, 4334, 4335, 4336
ZONE VALVE NEEDS
LABELED TO REFLECT
ROOMS SERVED
SICU HALLWAY VALVE MANUFACTURER: CHEMETRON
AREA SERVED: COMMENTS
PRE-OP
OXYGEN GAUGE IS NOT
ACCURATE
OUTSIDE PRE-OP VALVE MANUFACTURER: CHEMETRON
AREA SERVED: COMMENTS
PACU BAYS 1, 2, 3, 4, 5, 6, ISOLATION N2O NOT IN USE
ACROSS OR-2 VALVE MANUFACTURER: CHEMETRON
ZONE VALVE ANALYSIS
4TH FLOOR
AREA SERVED: COMMENTS
4A WING ROOMS 4102, 4103, 4104,
4110, 4111, 4112, 4113, 4119, 4120, 4121, 4122, 4123, 4124, 4127, 4128
ZONE VALVE NEEDS
LABELED TO REFLECT
ROOMS SERVED
4A NURSES STATION VALVE MANUFACTURER: CHEMETRON
AREA SERVED: COMMENTS
4B WING ROOMS 4202, 4203, 4206,
4207, 4208, 4210, 4211, 4217, 4218, 4219, 4220, 4226, 4227, 4228
ZONE VALVE NEEDS
LABELED TO REFLECT
4B NURSES STATION VALVE MANUFACTURER: AMICO
ZONE VALVE ANALYSIS
4TH FLOOR
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 54 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR 48 PSI PASS PASS 1/2" BALL PASS PASS
N2O 50 PSI PASS PASS 1/2" BALL PASS PASS P
VACUUM 22" HG PASS PASS 3/4" BALL PASS PASS P
N2 150 PSI PASS PASS 1/2" BALL PASS PASS P
FLOOR: 4TH VALVE LOCATION:
GAS
GAUGE
READING
GAS LABELED
CORRECTLY
ZONE LABELED
CORRECTLY
VALVE SIZE &
TYPE
PROPER
LOCATION
LEAKAGE PASS FAIL
OXYGEN 56 PSI PASS PASS 1/2" BALL PASS PASS P
MED AIR 50 PSI PASS PASS 1/2" BALL PASS PASS P
N2O 50 PSI…