About this file
Written by GovTribe AI
· Dec 21, 2023
This spreadsheet details power requirements for various laboratories and instruments within the Fire Debris Lab, XRD Lab, Explosives Lab, Explosives Lab 2, Trace Lab, and related uninterrupted power supply needs. Locations, instruments, voltage and amperage needs, circuit breakers, existing uninterruptible power supplies, and required upgrades are specified. A related federal contract opportunity from the Department of Justice Bureau of Alcohol Tobacco Firearms and Explosives seeks offers for laboratory dual in-line uninterruptible power supply installation to address the needs documented in the spreadsheet, with response due by the listed solicitation number and closing date. Relevant details include federal agency, required products and services, and incumbent equipment to be disposed of or upgraded.
Other files
File
Type
Posted
Site Visit Q and A Part 2 - 15A00023R00000085.docx
DOCX document
Sep 5, 2023
Attachment 1 - Statement of Work Update 2.docx
DOCX document
Sep 5, 2023
ATF Load Study.pdf
PDF
Sep 5, 2023
Site Visit Q and A Part 1 - 15A00023R00000085.docx
DOCX document
Aug 24, 2023
Attachment 1 - Statement of Work Update 1.docx
DOCX document
Aug 24, 2023
Questions and Answers 1 - 15A00023R00000085.docx
DOCX document
Aug 11, 2023
Site Visit Instructions - 15A00023R00000085.docx
DOCX document
Aug 11, 2023
15A00023R00000085 - CO Signed.pdf
PDF
Aug 9, 2023
Attachment 4 - Past Performance Survey.docx
DOCX document
Aug 9, 2023
Attachment 3 - Evaluation Criteria.docx
DOCX document
Aug 9, 2023
Attachment 1 - Statement of Work.docx
DOCX document
Aug 9, 2023
Attachment 2 - Quotation Instructions.docx
DOCX document
Aug 9, 2023
Show all 12
On GovTribe
Work with this file
Download the original file
Contacts named in this file
Similar government files
Ask GovTribe AI about this file
Text version
The file's text, extracted by GovTribe without its formatting.
Sheet1
Location Room Name Volts Amps VA Hz Circuit Breakers UPS?/comments
Fire Debris Lab 460 FD1 GC 120 2250 47.5-63 LD4 9/11 Toshiba 6KVA-to be disposed
need two quad boxes MS 120 1100 60
run one new circuit to LD panel ~40 feet
2 ovens verify the power pole is on LD panel
Fire Debris Lab 460 FD4 GC 220 2950 47.5-63 LD5 10/12 none
need two quad boxes MS 120 1100 60 LD1 45 none
run one new circuit to LD ~40 feet has a 220 already
Fire Debris Lab 460 FD5 GC 120 2250 47.5-63 LD4 2/4 Toshiba 6KVA-to be disposed
need to desginate a 220 outlet (has power) MS 220 1100 50-60
has a 220 already
Fire Debris Lab 460 FD6 GC 120 2250 47.5-63 LD2 78 none
need two quad boxes (has power) MS 120 1100 60 ? ? none
Fire Debris Lab 460 FD7 GC 120 2350 50-60 LD2 69 Eaton 2KVA
need two quad boxes (has power) MS 120 1100 60 LD1 47 Eaton 1.5KVA
Fire Debris Lab 460 ISQ GC 120 2000 50-60 LD4 5/7 Toshiba 6KVA-to be disposed
need one quad box (has power) MS 120 15 50-60
XRD Lab 456 Panylitical XRD 230 38 60 LD2 44 none
has all outlets Haskirs Chiller 208 15 60 LD2 43 none
Explosives Lab 454 FTIR 120 10 LD2 71 Eaton 3KVA
need one quad box (has power)
Explosives Lab 454 HPLC DAD 120 255 50-60 LD2 70 Eaton 3KVA
need one quad box (has power) Sampler 120 550 50-60
Pump 120 255 50-60
Column 120 330 50-60
Explosives Lab2 450 IC Sampler 120 4 50-60 LD4 1/3 Toshiba 8KVA-to be disposed
need one quad box (has power) DC 120 10 50-60
EG 120 2
Pump 120 2 50-60
Peak Generator 120 12 60
ISQEC MS 120 15 50-60 ? Eaton 2KVA
Explosives Lab2 450 HPLC DAD 120 150 50-60 LD2 76/77 Powervar (new)
need three 220 outlets (has power) Sampler 120 550 50-60
need one quad box (has power) Pump 120 550 50-60
Column 120 330 50-60
TSQ MS 230 10 50-60
Peak Generator 230 7 50-60
Sogevac Pump 230 12 50-60
441 Bruker XRD 230 ? ? ? LD5 39/41 Integrity (new)
has all outlets
Trace Lab 418 EDAX XRF 120 13 L1 59 Eaton 2KVA
need one quad box (has power)
run one new circuit to LD panel ~30 feet
Trace Lab 418 FTIR Bench 120 10 50-60 LD5 4 Eaton 3KVA
need one quad box (has power) Scope 120 3 50-60
Trace Lab 418 PyroGC GC 120 2350 50-60 L1 55 none
Need one quad box (has power) MS 120 1100 60 LD5 30
Pyrolysis 120 10 LD5 30
four extra Toshiba 6KVA-to be disposed