The Department of Veterans Affairs Network Contracting Office 8 intends to sole source six GE Healthcare E14C4T triplane endo transducers and six reprocessing trays for the C.W. Bill Young Veterans Medical Center Surgery Service in Bay Pines, Florida, with responses due November 16, 2026 at 14:00 EDT. The Set Aside line lists no set-aside. The notice is a market research sources sought that targets SDVOSBs, VOSBs, and SBs, and states that responses will determine whether an SDVOSB set-aside is feasible. If SDVOSB responses are insufficient, the requirement may be solicited as a VOSB or SB set-aside, or as unrestricted if small business responses are also insufficient. The notice states it is not a solicitation, and it requests email replies with a capability statement and supporting documentation to Contracting Officer Michael A. Shook at michael.shook@va.gov. The transducers replace probes that Biomedical Engineering found non-functional, unrepairable, and beyond economical repair, and that the manufacturer no longer supports. The attached government file summary describes the requirement as firm, fixed-price and lists delivery to Building 36 Equipment Warehouse no later than December 19, 2026, for placement at the Surgery Service in Building 100. Required capabilities include triplane transrectal ultrasound transducers with integrated biopsy guidance, 4-14 MHz operation, compatibility with BK Medical ultrasound systems, specified imaging modes, and sterilization and safety requirements. The source does not name an incumbent or state an award value.

Notice text

Good day, SDVOSB s, VOSB s & SB's Sources Sought INTENT TO SOLE SOURCE Pre-Solicitation.

This Sources Sought is issued solely for informational and planning purposes only. It does not constitute the issuance of a solicitation or a promise to issue solicitation in the future. The purpose of this notice is to conduct market research and obtain information from qualified SDVOSBs VOSBs , or SBs . This is not a request for bids, quotes, offers, or proposals. The responses to this announcement will assist in determining the socio-economic status such as a SDVOSB set-aside if feasible. If insufficient SDVOSB responses are received to establish the basis for setting this requirement aside for SDVOSB s, the requirement may be solicited as a VOSB or a SB set aside, if none or not enough small business responses are received, this may be solicited as unrestricted, all interested parties responding shall be eligible to compete.

Please reply in the affirmative via email if you are interested and can provide the required product/services for this requirement. Please provide your capability statement, and any other supportive documentation. I will accept responses for this Sources Sought until the 16th of November 2026 14:00 EDT, thanks.

1. General: C.W. Bill Young Veterans Medical Center Surgery Service located at 10000 Bay Pines Blvd. Bay Pines, Fl. 33744 is seeking procurement of six transducers and six trays from GE Healthcare. The department requires replacement of ultrasound transducers that have reached the end of their operational life and can no longer be repaired or supported by the manufacturer.

2. Background: These transducers are vital to patient diagnostic and procedural imaging. The existing probes have been evaluated by Biomedical Engineering and determined to be non-functional, unrepairable, and beyond economical repair. Without replacements, the service cannot maintain adequate imaging capacity to support patient care.

2.Scope of Work: This requirement needs to provide six (P/N: 9018) E14C4T Triplane endo transducers and six (P/N:UA2511) trays.

3.Place of Performance: The product is requested to be delivered to Surgery Service, building 100, MOR. 10,000 Bay Pines Blvd, Bay Pines, Florida 33744, via the Warehouse Building 36 for Equipment.

4.Period of Performance: The product is requested to be delivered no later than December 19th.

5.Type of Contract: Supply: Firm, Fixed Price.

6.Requirement Overview: The Contractor shall provide the full requested quantity of products listed below.

6.1 Salient Characteristics

NAME BRAND - Equipment Requirement: GE Healthcare

QTY. 6 - (P/N: 9018) E14C4T Triplane Endo Transducer

QTY. 6 (P/N: UA2511) Reprocessing Trays for E14C4T

Equipment Requirement:

Minimum Required Features and Specifications

Type / Intended Use:

Must be a transrectal ultrasound transducer specifically designed for prostate imaging and biopsy guidance in urology applications.

Must provide multi-plane (triplane) imaging capability allowing visualization of transverse, sagittal, and end-fire planes without changing probes.

Frequency & Imaging Performance:

Must operate within a frequency range of approximately 4 MHz to 14 MHz to provide both high-resolution and deep-penetration imaging of the prostate.

Must support B-mode, M-mode, Color Doppler, and Power Doppler imaging modes (with optional elastography and contrast imaging if supported by the connected ultrasound system).

Biopsy and Procedural Capability:

Must include or support an integrated biopsy guide channel that allows both biplane and end-fire biopsy procedures without requiring probe exchange.

Must accommodate standard 17- or 18-gauge biopsy needles through compatible biopsy guide attachments.

Must permit real-time needle visualization during transrectal or transperineal prostate biopsy procedures.

4. Design & Ergonomics

Must be a compact, endocavity-style probe suitable for patient comfort and operator ergonomics during transrectal use.

Must include a control ( Smart ) button or equivalent for hands-free freezing, storing, or switching imaging planes directly from the probe handle.

Must be designed for one-handed operation and compatible with sterile probe covers.

5. Compatibility

Must be fully compatible with BK Medical ultrasound systems, including bk3000, bk5000, bkActiv, and bkSpecto platforms.

Must connect using the appropriate BK Medical proprietary connector interface without the need for modification or adapters.

6. Cleaning & Sterilization

Must be fully immersible and validated for standard high-level disinfection or sterilization processes used for endocavity probes.

Must be compatible with approved sterile covers and disinfection solutions specified in the manufacturer s reprocessing guide.

7. Construction & Safety

Must be constructed of biocompatible materials suitable for internal use (per ISO 10993 or equivalent standards).

Must comply with applicable IEC/EN 60601-1 electrical safety and ultrasound acoustic output limits.

Must be supplied new, manufacturer-certified, and covered under OEM warranty for clinical use.

Summary Table

Feature Category

Requirement

Type

Triplane transrectal prostate ultrasound transducer

Frequency

4 14 MHz

Imaging Modes

B, M, Color Doppler, Power Doppler (Elastography optional)

Biopsy Capability

Integrated dual-plane biopsy guide

Needle Compatibility

17 18 G

Operation

One-handed, Smart Button on probe

Sterilization

Fully immersible, high-level disinfectable

Compatibility

bk3000, bk5000, bkActiv, bkSpecto

Intended Use

Prostate imaging and biopsy guidance

Condition

New OEM product, OEM warranty

Files

File Type Posted
36C24827Q0127.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
Philips EPIQ CVx Ultrasound System Repair Palo Alto, California 94304, United States · NAICS 339112 Veterans Integrated Service Network 21 No set-aside
J065--Fort Harrison GE Precision Radiology Preventive Maintenance NAICS 811210 Veterans Integrated Service Network 19 No set-aside
NX EQ Microscope Spine United States · NAICS 339113 VA Strategic Acquisition Center Frederick No set-aside
6515-- TruClear Hysteroscopes Department of Veterans Affairs VA Northern California Health Care Martinez CA 94553, United States · NAICS 339112 Veterans Integrated Service Network 21 No set-aside
J065--Intent to Sole Source Hines VA Base 2 Carto 3 EP Cath Lab Service NAICS 811210 Veterans Integrated Service Network 12 No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors