Summary of Technical Specifications for Multi-Slice Spiral CT Scanner Procurement

This document contains technical specifications and requirements for the procurement of a multi-slice spiral computed tomography (CT) scanner system with associated equipment and services. The specifications detail a whole-body CT scanner requiring minimum 32 physical slices per rotation and 64 reconstructed slices, with a gantry tunnel diameter of at least 70 cm and a 360-degree rotation time not exceeding 0.35 seconds. The X-ray generator must deliver maximum power of 55 kW with tube current strength of 460 mA maximum and voltage range of 80–140 kV, while the anode must have heat capacity of 7.0 million thermal units and cooling rate of 1,070 thousand thermal units per minute. The detector system requires solid-state technology with minimum 54,000 detector elements, 32 physical rows, and 40 mm anatomical coverage per rotation, with capability to produce maximum 64 reconstructed slices at 0.625 mm minimum collimated slice thickness.

The procurement includes a motorized patient table with 1,730 mm scanning range and 225 kg load capacity, operator console with color LCD monitor (minimum 19 inches, 1280x1024 resolution), and comprehensive software packages supporting multiplanar reconstruction, 3D volumetric rendering, cerebral perfusion analysis, vessel analysis, and cardiac imaging with ECG synchronization. Additional equipment requirements include uninterruptible power supply, voltage stabilizer, lead glass shielding (800 x 1200 mm minimum), automatic bolus injector, and medical film printer. The procurement also specifies a 100–120 kW diesel generator with automatic start and protective cover. Required services include equipment delivery, commissioning, specialist-led training for medical professionals, and 12-month warranty from commissioning date. All equipment must comply with Ukrainian technical regulations, with Ukrainian-language software documentation and generator serviceability in Ukraine required.

View the file

Other files

Show all 13

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SBU - CONTRACTING AND ACQUISITIONS

№ Specific requirements Declaration of the participant

Links to the attached documents

1. Equipment type

1.1 Multi-slice spiral computed tomography scanner for whole body scanning

Not less than 32 physical slices per rotation and 64 reconstructed slices

2. Gantry

2.1 Gantry tunnel diameter (gantry aperture), not less than, (cm) 70

2.2 Minimum time for 360° rotation of the X-ray tube, no more than, (sec)

0,35

3. Generator and X-ray tube

3.1 Maximum power of the generator, not less than, (kW) 55

3.2 Maximum current strength on the X-ray tube, not less than, (mA)

3.3 Minimum voltage on the X-ray tube, no more than, (kV) 80

3.4 Maximum voltage on the X-ray tube, not less than, (kV) 140

3.5 The availability of two focal spots Compliance

3.6 Heat capacity of the anode of the X-ray tube, not less than, (million thermal units, MTU)

7,0

3.7 Anode cooling rate of the X-ray tube, not less than, (thousand thermal units/min, TTO)

1 070

3.8 Minimum current value on the tube, no more than (mA) 10

4. Detector

4.1 Detector type Solid state

4.2 Number of detector elements, not less than 54 000

4.3 Number of physical rows of the detector, not less than 32

4.4 Maximum anatomical coverage of the detector per one gantry rotation in the axial scanning mode (width of the detector along the Z-axis), not less than, (mm)

Declaration of the participant

Links to the attached documents

4.5 Maximum number of reconstructed slices per one rotation of the

X-ray tube, not less than

4.6 Minimum thickness of the collimated slice, no more than (mm) 0,625

5. Patient table

5.1 Maximum scanning range, not less than, (mm) 1730

5.2 Maximum load on the table with accurate positioning, not less than, (kg)

5.3 Motorized table drive Compliance

6. Scanning and reconstruction of images

6.1 Maximum field of view (FOV) of reconstruction, not less than, (mm)

6.2 Maximum image reconstruction speed with iterative

reconstruction not less than, (frames/sec)

Not less than 28

6.3 Maximum reconstruction matrix, not less than 1024×1024

7. Image quality, dose values

7.1 Maximum spatial resolution, not less than 15 pairs of lines per cm.

8. Operator console

8.1 Availability of operator console with a color LCD monitor, at least 19 inches in size and with resolution of at least 1280x1024

Compliance

8.2 Ability to record CT scan data on digital media (DVDs and/or

USB-flash or other)

Compliance

8.3 DICOM standards support Compliance

8.4 Providing image archiving Compliance

8.5 Ensuring the image transfer to the network Compliance

8.6 Providing printing of images on thermographic film Compliance

8.7 The patient-operator intercom system in the scan room Compliance

8.8 Ethernet network support Compliance

Declaration of the participant

Links to the attached documents

8.9 Cardiac imaging program with retrospective and prospective

ECG synchronization

Compliance

8.10 Availability of ECG monitor Compliance

9. Software and hardware packages and clinical applications on the operator console

9.1 Multiplanar reconstruction (MPR) Compliance

9.2 Software for volumetric imaging of anatomical structures Compliance

9.3 Maximum Intensity Projections and Minimum Intensity

Projections (MIP and minIP)

Compliance

9.4 Filming Compliance

9.5 Real-time beam load modulation software Compliance

10. Data post-processing workstation

10.1 Specialized workstation for image post-processing with a medical monitor

Compliance

10.2 Diagonal of a medical monitor, not less than 19’’

10.3 Software for reconstruction and editing of three-dimensional volume (3D Volume rendering), obtaining maximum intensity projections (MIP), obtaining minimum intensity projection

(minIP), automatic removal of bones, measurements - angles, density, dimensions, reporting, printing

Compliance

10.4 Specialized cerebral perfusion visualization software including creation of quantitative color maps of cerebral blood flow, cerebral blood volume, average transit time, time to peak.

The possibility of conducting an assessment of a dynamic volumetric brain scanning

Compliance

10.5 General vessel analysis software package with the ability to remove or edit bone structures, as well as select and segment vessels for faster standard measurements such as vessel internal diameter, vessel cross-sectional area, vessel segment length and curvature, vessel angle

Compliance

Declaration of the participant

Links to the attached documents

10.6 Software package for imaging and analysis of coronary vessels Compliance

10.7 Instructions for the software in Ukrainian Compliance

11. Additional equipment

11.1 Uninterruptible power supply for the digital part of a CT scanner Compliance

11.2 Voltage stabilizer for the entire CT scanner Compliance

11.3 Power distribution board Compliance

11.4 Lead glass, at least 800 x 1200 mm in size Compliance

11.5 Automatic bolus injector for injection of contrast agent and saline

Compliance

11.6 Medical printer for printing on film Compliance

12. Diesel generator for computed tomography scanner

100-120 kW, automatic start, silent, with protective cover 100-120 kW

Required by the Customer:

- certificate or declaration of compliance with technical regulations for equipment in accordance with the legislation of Ukraine

- services should include:

o delivery of equipment to the end Customer o commissioning of equipment o arrival of a specialist to train medical professionals o warranty service of the equipment for 12 months from the date of its commissioning

- generator must be serviceable in Ukraine

- generator installation on a concrete pad and commissioning required

- a protective cover for the generator is required

- warranty according to the manufacturer's warranty