Attachment 1 Tomography System Technical Specifications.pdf
PDF 211 KB Posted
- Solicitation number
- W564KV26QA014
About this file
This is a technical specifications document for the procurement of a multi-slice spiral computed tomography (CT) scanner for whole-body scanning. The document is structured as a declaration form where potential bidders confirm compliance with the listed technical requirements.
The equipment specifications include: a minimum 32-slice physical (64 reconstructed) CT scanner with a 70 cm gantry tunnel diameter and 0.35-second rotation time; a generator with minimum 55 kW power and X-ray tube capabilities of 460 mA maximum current and 140 kV maximum voltage; a solid-state detector with at least 54,000 elements across 32 physical rows providing 40 mm anatomical coverage per rotation; a motorized patient table supporting 225 kg with 1,730 mm scanning range; image reconstruction capabilities including 1024×1024 matrix with at least 28 frames/second iterative reconstruction and 15 pairs of lines per cm spatial resolution; and operator console with 19-inch color LCD monitor, DICOM support, image archiving, network transfer, ECG synchronization, and cardiac imaging capabilities. The system must include specialized software packages for multiplanar reconstruction, 3D volume rendering, cerebral perfusion analysis, vessel analysis, and coronary vessel imaging with Ukrainian-language instructions. Additional required equipment includes uninterruptible power supply, voltage stabilizer, power distribution board, lead glass shielding (800×1200 mm minimum), automatic bolus injector, and medical film printer. The supplier must provide equipment delivery, commissioning, specialist training for medical professionals, and 12-month warranty service from commissioning date. A certificate or declaration of compliance with Ukrainian technical regulations is required.
View the file
Other files
Show all 13
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SBU - CONTRACTING AND ACQUISITIONS
№ Specific requirements Declaration of the participant
Links to the attached documents
1. Equipment type
1.1 Multi-slice spiral computed tomography scanner for whole body scanning Not less than 32 physical slices per rotation and 64 reconstructed slices
2. Gantry
2.1 Gantry tunnel diameter (gantry aperture), not less than, (cm) 70
2.2 Minimum time for 360° rotation of the X-ray tube, no more than, (sec) 0,35
3. Generator and X-ray tube
3.1 Maximum power of the generator, not less than, (kW) 55
3.2 Maximum current strength on the X-ray tube, not less than, (mA)
3.3 Minimum voltage on the X-ray tube, no more than, (kV) 80
3.4 Maximum voltage on the X-ray tube, not less than, (kV) 140
3.5 The availability of two focal spots Compliance
3.6 Heat capacity of the anode of the X-ray tube, not less than, (million thermal units, MTU) 7,0
3.7 Anode cooling rate of the X-ray tube, not less than, (thousand thermal units/min, TTO)
1 070
3.8 Minimum current value on the tube, no more than (mA) 10
4. Detector
4.1 Detector type Solid state
4.2 Number of detector elements, not less than 54 000
4.3 Number of physical rows of the detector, not less than 32
4.4 Maximum anatomical coverage of the detector per one gantry rotation in the axial scanning mode (width of the detector along the Z-axis), not less than, (mm) participant
Links to the attached documents
4.5 Maximum number of reconstructed slices per one rotation of the
X-ray tube, not less than
4.6 Minimum thickness of the collimated slice, no more than (mm) 0,625
5. Patient table
5.1 Maximum scanning range, not less than, (mm) 1730
5.2 Maximum load on the table with accurate positioning, not less than, (kg)
5.3 Motorized table drive Compliance
6. Scanning and reconstruction of images
6.1 Maximum field of view (FOV) of reconstruction, not less than, (mm)
6.2 Maximum image reconstruction speed with iterative
reconstruction not less than, ( frames/sec)
Not less than 28
6.3 Maximum reconstruction matrix, not less than 1024×1024
7. Image quality, dose values
7.1 Maximum spatial resolution, not less than 15 pairs of lines per cm.
8. Operator console
8.1 Availability of operator console with a color LCD monitor, at least 19 inches in size and with resolution of at least 1280x1024 Compliance
8.2 Ability to record CT scan data on digital media (DVDs and/or USB-flash or other)
Compliance
8.3 DICOM standards support Compliance
8.4 Providing image archiving Compliance
8.5 Ensuring the image transfer to the network Compliance
8.6 Providing printing of images on thermographic film Compliance
8.7 The patient-operator intercom system in the scan room Compliance
8.8 Ethernet network support Compliance
participant
Links to the attached documents
8.9 Cardiac imaging program with retrospective and prospective
ECG synchronization Compliance
8.10 Availability of ECG monitor Compliance
9. Software and hardware packages and clinical applications on the operator console
9.1 Multiplanar reconstruction (MPR) Compliance
9.2 Software for volumetric imaging of anatomical structures Compliance
9.3 Maximum Intensity Projections and Minimum Intensity
Projections (MIP and minIP) Compliance
9.4 Filming Compliance
9.5 Real-time beam load modulation software Compliance
10. Data post-processing workstation
10.1 Specialized workstation for image post-processing with a medical monitor Compliance
10.2 Diagonal of a medical monitor, not less than 19’’
10.3 Software for reconstruction and editing of three-dimensional volume (3D Volume rendering), obtaining maximum intensity projections (MIP), obtaining minimum intensity projection (minIP), automatic removal of bones, measurements - angles, density, dimensions, reporting, printing
Compliance
10.4 Specialized cerebral perfusion visualization software including creation of quantitative color maps of cerebral blood flow, cerebral blood volume, average transit time, time to peak.
The possibility of conducting an assessment of a dynamic volumetric brain scanning
Compliance
10.5 General vessel analysis software package with the ability to remove or edit bone structures, as well as select and segment vessels for faster standard measurements such as vessel internal diameter, vessel cross-sectional area, vessel segment length and curvature, vessel angle
Compliance participant
Links to the attached documents
10.6 Software package for imaging and analysis of coronary vessels Compliance
10.7 Instructions for the software in Ukrainian Compliance
11. Additional equipment
11.1 Uninterruptible power supply for the digital part of a CT scanner Compliance
11.2 Voltage stabilizer for the entire CT scanner Compliance
11.3 Power distribution board Compliance
11.4 Lead glass, at least 800 x 1200 mm in size Compliance
11.5 Automatic bolus injector for injection of contrast agent and saline Compliance
11.6 Medical printer for printing on film Compliance
Required by the Customer:
- certificate or declaration of compliance with technical regulations for equipment in accordance with the legislation of Ukraine
- services should include:
o delivery of equipment to the end Customer o commissioning of equipment o arrival of a specialist to train medical professionals o warranty service of the equipment for 12 months from the date of its commissioning