B01 - W519TC26QCALI_Gage Lab Calibration_QAs_11MAR2026.xlsx
XLSX spreadsheet 40 KB Posted
- Attached to
- GAGE LAB CALIBRATION SERVICES & REPAIRS
- Solicitation number
- W519TC-26-Q-CALI
About this file
This is a Questions and Answers document addressing inquiries from potential offerors regarding solicitation W519TC-26-Q-CALI for Gage Lab Calibration & Repair Services. The Q&A clarifies key requirements and expectations for bid submission by the March 12, 2026 deadline, with all questions due 72 hours prior to that date. This is a recompete of incumbent contract W519TC-25-P-2005, and offerors should direct inquiries to Bridget Garnica (bridget.m.garnica.civ@army.mil) or Bryce Willett (bryce.t.willett.civ@army.mil) at the Department of the Army Materiel Command Contracting Command Rock Island Arsenal.
Offerors must quote only the cells highlighted in yellow on the price matrix tabs and should include all line items on the equipment list, with on-site calibration preferred for stationary items and laboratory calibration acceptable for portable equipment. The document specifies that there are 15 feeler gage sets (primarily 31-leaf) and 2 gage block sets requiring pricing; make and model information is largely unnecessary as calibration standards are uniform across brands for standard items. Travel costs and mileage may be added to applicable CLINs, and pricing for scales should reflect the most common capacity where breakdown is not feasible. Subcontracting is permitted with requirements outlined in Section 12.0 of the Performance Work Statement; if items are sent to different labs or third parties, offerors must provide subcontractor accreditations and notify if delays exceeding one week are anticipated. NIST certification pricing should be included where applicable, and previous certifications are available upon request.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| B01 - W519TC26QCALI_Gage Lab Calibration_QAs_12MAR2026.xlsx | XLSX spreadsheet | |
| B01 - Attachment 0005 - Price Matrix 12MAR2026.xlsx | XLSX spreadsheet | |
| B01 - W519TC26QCALI_Gage Lab Calibration_QAs v2_09MAR2026.xlsx | XLSX spreadsheet | |
| B01 - W519TC26QCALI_Gage Lab Calibration_QAs_09MAR2026.xlsx | XLSX spreadsheet | |
| B01 - W519TC26QCALI_Gage Lab Calibration_QAs_05MAR2026.xlsx | XLSX spreadsheet | |
| B01 - Attachment 0005 - Price Matrix 26FEB2026.xlsx | XLSX spreadsheet | |
| B01 - Attachment 0004 - DD Forms 1423 (CDRLS).pdf | ||
| B01 - Attachment 0001 - PWS Gage Calibration 26FEB2026.pdf | ||
| B01 - Attachment 0002 - Gage Calibration Item List 26FEB2026.xlsx | XLSX spreadsheet | |
| B01 - Attachment 0003 - WD No. 2015-4821 Revision 31; dtd 02DEC2026.pdf | ||
| B01 - Attachment 0007 - Provisions and Clauses - Gage Calibration.pdf | ||
| B01 - Attachment 0006 - Gage Calibration Questionnaire.docx | DOCX document | |
| B01 - Attachment 0005 - Price Matrix 26FEB2026.xlsx | XLSX spreadsheet |
Show all 13
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sheet1
| W519TC26QCALI GAGE LAB CALIBRATION | ||
| # | QUESTION | ANSWER |
| 1 | Is there any incumbent/ recompete or this is a brand-new project? If yes, could you please provide us with their award number or even if the customer bought similar services in the past. | Incumbent contract under W519TC-25-P-2005 |
| 2 | I would like to confirm if we should field any questions regarding this bid to you, or if there is another inquiry process that we need to follow? I also would like to confirm if there is an end date for inquiries prior to the bid response deadline of March 12th? | Please submit any questions to the following: Bridget Garnica at bridget.m.garnica.civ@army.mil and Bryce Willett at bryce.t.willett.civ@army.mil. Per the solicitation, all questions shall be submitted no later than 72 hours prior to 12 March 2026. |
| 3 | I don't see a listed manufacturer or model for these but we will do the best we can given this is due 3/12. For the CMM's and Keyence items, I would need a model for those. | The Keyence themselves do not get calibrated, we use an OP-88275 VR series calibration standard on them. For CMM manufacturer/ model, see chart below: |
| ERROR:#VALUE! | ||
| 4 | What are the names of the QA manager and/or the site lead for this bid? | At this time we are unable to provide this information. Please list contract specialist, Bridget Garnica, in the solicitation. |
| 5 | Am I quoting the gauge calibration item list (excel file) or the lists in the tabs for the price matrix? | Please quote the tabs in the price matrix, but only those cells highlighted in yellow. |
| 6 | There are a number of feeler gage sets and gage block sets that do not have a quantity included, can this be provided? I am looking for the quantities of gages in the Feeler gage sets, and quantity of blocks in the gage block sets. There is a difference in price for these items based on quantities. | There are 15 sets of feller gages and 2 gage block sets. The majority of them are 31 leaf. 31 leaf is the largest we have. |
| 7 | Should we include NIST certification pricing in the bid for items that we cannot calibrate to ISO 17025 or do not require ISO 17025? | Yes, these should be included, however, any reputable calibration company will/ should be certified to the ISO standards |
| 8 | Can previous Certs be provided if requested? | Yes |
| 9 | We noticed that more than half of the items on the equipment list do not include the make and model. Would it be possible to obtain a list that includes the available make and model information for these items? | This information is largely unknown and unneccessary for most items. For example all brands of 1/4" plug gage will be calibrated the same. |
| 10 | Do you want all these items quoted as an on-site calibration, or just the highlighted ones? We would not pick up all these items, take them back to the lab, calibrate/certify them, and them deliver them back, although you can send them all in to the lab if you desire. Our objective would be to do an on-site calibration which would be much faster and more effective for you. | All items on the list shall be quoted. If the contractor can bring everything in to calibrate each item, that would be fine. We will not pack items and send them. Per the PWS, the contractor will pick-up, deliver, and provide containers. Per Attachment 0002 Calibration Item List, if the stationary column is marked "Y" the plan would be to calibrate onsite. Other than some scales, most of these items can't be moved without equipment. Items were this column is blank are usually shipped to the contractor. |
| 11 | Could you please clarify whether subcontracting is permitted for this specific notice? I would appreciate any information you can provide regarding the subcontracting requirements or limitations for this solicitation. | Subcontracting is permitted for this requirement. See Sections 12.0 Security Considerations of the PWS for subcontractor requirements |
| 12 | Would you like me to add the travel cost and mileage to the line item on the excel? Or just place calibration cost? | Yes, you can add this cost to whatever CLIN it applies to. |
| 13 | Scales, we have different prices for different capacities. I chose the most common cost and divided it to fill in the pricing. Is this acceptable? | Yes, this is acceptable as we don’t provide an opportunity to breakdown the cost for each type of scale. |
| 14 | If there is an item on your list and my location doesn’t have the capability of calibrating it, can I send to another location of ours? And do I still fill in the pricing for the bid? | Yes, per the PWS, you can send an item to a different lab or 3rd party (subcontractor). We would need to see the subcontractors’ accreditations to ensure they are qualified. If a different lab/ 3rd party is utilized, please let us know if this would cause a delay longer than the 1 week that is anticipated. Yes, the price for these items need to be priced out in the Price Matrix. |