This document provides details for a federal contract opportunity to prepare site for nuclear medicine support space at a Veterans Affairs medical center. The Department of Veterans Affairs, through its Southern Arizona Veterans Affairs Healthcare System network contracting office, intends to solicit fixed-price bids to provide construction services at the New Mexico VA Health Care System in Albuquerque. Work will include demolition, site preparation, and modifications to HVAC, plumbing, electrical, IT, fire suppression systems, and architectural renovation of existing rooms for new counters and cabinetry. The solicitation will be set aside for Service-Disabled Veteran Owned Small Businesses and released on July 20, 2020. Bids will be evaluated based on low price, and the contract awarded for a period of performance within 272 calendar days of the notice to proceed. The magnitude of the project is estimated between $100,000 to $250,000 under NAICS code 236220.

View the file

Other files

Show all 20

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

LBT Diagnostic Radiation Physics Consulting Ltd Phillip W Patton, PhD CHP DABR

New Mexico Veterans Affair Healthcare System

Raymond G. Murphy Medical Center

1501 San Pedro SE

Albuquerque, New Mexico 87108

Report of the Radiation Shielding Calculations

For the

Uptake Room for the PET CT Scanner

August 30, 2018

Phillip W Patton, PhD, CHP, DABR, DABSNM

LBT Radiation Physics Consulting, Ltd.

6367 Apple Orchard Ave

Las Vegas, NV 89142

702-336-3004

Discussion

In general, calculated shielding is often overestimated due to assumptions on workload, use factors and occupancy factors. I have attempted to provide more realistic factors based on information from the facility and modified occupancy estimates. The workload is set high enough to allow for any minor increases without requiring further shielding.

Radiation protection is based on three concepts: time, distance, and shielding. All three are used to maintain radiation exposures in the workplace at or below safe levels. The maximum weekly exposure currently allowed for occupationally exposed workers is 100 millirem (controlled areas) and for non-occupationally exposed workers (uncontrolled areas) is 2 millirem. Therefore to apply the ALARA concept, shielding design exposure limits are 100 uR/week for controlled areas and 20 uR/week for uncontrolled areas.

All lead sheeting should extend to a height of at least 7 feet above floor level. Nail heads do not need to be overlapped. Overlying drywall seams should be offset and great care should be exercised in taping and plastering the seams. No gaps or holes in drywall are acceptable.

Any questions concerning this report may be directed to Phillip Patton 702-336-3004

Performed by:

Phillip W Patton, PhD CHP, DABR, DABSNM

Diagnostic Medical Physicist, References

(1) American Association of Physicist in Medicine (AAPM) Task Group 108, PET & PET/CT Shielding

Requirements, AAPM, 2006

Patient Dose Rate

Constant (uSv-m /MBq-h)

Dose Rate

Constant (uSv-m /MBq-h)

Dose Reduction

Factor over time

Uptake

Dose Reduction Factor over time Imaging

FDG 0.092 0.143 0.83 0.91

Typical Dose (MBq) 573.5

Typical Patients/week 25

Uptake Time (hr) 1

Imaging Time (hr) 0.5

Important: see summary for final recommendations

Barrier T Distance (m) Weekly Dose Equiv. Allowed Exposure (uSv) Transmission Factor

A-B Control Room 1.000 4.870 46.2 100.0 2.166

B-C Injection Room 0.050 2.440 183.9 100.0 10.876

C-D Waiting 0.050 1.830 326.9 20.0 1.224

D-E Bone Density Room 1.000 3.050 117.7 100.0 0.850

A-E Film Request 1.000 2.286 209.5 20.0 0.095

Required Pb (inches)Required Pb (mm)

None

None

None

None

None

None

1/18

16.239 1/2

1.393

Based upon the above calculations and professional judgement, the following recommendations for shielding are made:

Barrier Recommended Shielding (Uptake)

A-B Control Room None

B-C Injection Room None

C-D Waiting Room None

D-E Bone Density Room 1/8

A-E Film Request Room 1/2

Summary