Manual: 4Q1.7 Procedure: 215
Revision: 10
OPERATION OF TSI FILTER TESTER 8127 AND TSI
8119 RESPIRATOR LEAK TESTER
Effective Date: 02/16/2017 Type-Class: Tech-REF Page: 1 of 26
1.0 INTRODUCTION
1.1 Purpose
To provide instructions for performing efficiency and resistance testing of respirator cartridges/filters and leakage testing on assembled respirators, using the Certitest Model 8127, TSI Automatic Filter Tester and the Model 8119, TSI Respirator Leak Tester.
1.2 Scope
This procedure provides guidelines for Respiratory Equipment Facility (REF) personnel to operate the TSI Filter Tester 8127 and TSI Respirator Leak Tester 8119.
This is a Reference procedure. The user is not required to have this procedure present while performing the activity.
The following data/round sheets may be utilized in the performance of this procedure:
OSR 4-510, Respiratory Equipment Record and Quality Assurance (QA)
ISF-REF-003, Voltage Value and Trigger Level Recording Sheet
All respirator repairs will be conducted in accordance with appropriate procedure.
2.0 GENERAL INFORMATION
2.1 Responsibilities
1. The REF Lead is responsible for:
Providing technical guidance on the use of the filter tester and respirator leak tester, Evaluating results of OSR 4-510 to determine if significant defective items are reported and notifying the Respiratory Protection Program Administrator
(RPPA),
Determining if issuance of a Nonconformance Report (NCR) is required and necessary, Initiating and responding to NCR’s in accordance with Manual 1Q, Procedure 15-1, Control of Nonconforming Items.
2. REF Personnel are responsible for:
Performing proper use of the filter and respirator leak tester, Completing qualification training as specified in the Job Performance Standard or approved equal, Ensuring Quality Assurance testing is completed as specified in Manual 4Q1.7, Procedure 104, Respiratory Equipment Facility Quality Verification.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 2 of 26
3.0 PRECAUTIONS AND LIMITATIONS
3.1 Safety
1. Workers shall follow applicable facility safety rules and observe personal protective equipment (PPE) postings.
2. Tools shall be inspected prior to use for obvious defects and that they are in good condition and good condition and safe for use.
3. Caution shall be used when installing the respirator on the headform to prevent hand injury.
4. The following chemicals may be used during this procedure to generate aerosol to perform leak test on respirators and to grease O-ring.
Emery 3004 synthetic Hydrocarbon 4 CST fluid (MSDS 31281-1), one gallon to be used annually
Dow Corning High Vacuum Grease (MSDS 6381-1)
5. Do NOT operate Certitest Model 8127, TSI Automatic Filter Tester and the Model 8119, TSI Respirator Leak Tester unless authorized and familiar with the contents of this procedure.
6. Gloves shall be worn when accessing front panel and when changing out 8119 respirator leak tester headform.
7. Proper lifting techniques shall be used per Manual 8Q, Procedure 12, General Site Safety Requirements.
3.2 Process Hazards
1. Do NOT overfill oil reservoir.
2. The “Emergency OFF” button is only to be used in an emergency, NOT for normal shut down.
3. The TSI 8127 contains an enclosed laser. This laser shall NOT be accessed by site personnel, but only by the manufacturer.
4.0 PREREQUISITES
None
5.0 PERFORMANCE
5.1 New P100 Cartridges Quality Verification at the REF
1. DETERMINE the number of new P100 cartridges to be inspected and placed in service each day. Those inspected and placed in service on each day are considered one lot.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 3 of 26
5.1 New P100 Cartridges Quality Verification at the REF, (cont.)
2. Using ANSI/ASQ Z1.4 tables listed in 4Q1.7, Procedure 104, DETERMINE the number of new P100 cartridges to be tested for Quality Verification.
3. PERFORM Quality Verification testing of new P100 cartridges per Section 5.3 and 5.4, AND
RECORD quantity on OSR 4-510.
5.2 Initial Start-Up
1. OPEN front door of the filter tester, AND
LOCATE the oil aerosol generator (shown as Figure 8.1).
CAUTION
Do NOT overfill oil reservoir.
2. CHECK level in oil aerosol generator. The red marker must be between the black rings. The top of the red marker should never extend above the top black ring.
A. IF the oil level is within proper range, THEN
PROCEED to Step 5.2.3.
B. IF the red marker is not between the black rings, THEN
1) REMOVE the fill cover.
2) FILL the oil reservoir until the liquid indicator level is between the black rings.
3) REPLACE the fill cover.
NOTE
The compressed air valve is located on the wall behind the filter tester.
3. ENSURE the valve is OPEN.
4. ENSURE that air pressure is within the 85 to 100 pounds psig range on the gauge.
A. IF psig is NOT within proper range, THEN
ADJUST the air pressure until appropriate psig is achieved.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 4 of 26
5.2 Initial Start-Up, (cont.)
Step 4, (cont.)
B. IF psig reading cannot be achieved, THEN
NOTIFY the REF Lead.
5. TURN ON the power switch (shown as Figure 8.2) which is located on the right front of TSI model 8127, as you face the machine. The main solenoid valve opens and compressed air is supplied to the tester. The filter holder should open.
A. IF the filter holder does not open, THEN
TURN power switch OFF, AND
NOTIFY the REF Lead.
6. LOCATE main pressure regulator adjustment knob and gauge (shown as Figure 8.3). The pressure regulator gauge shall read between 70 to 80 psig.
A. IF the gauge does NOT read between 70 to 80 psig, THEN
TURN the adjustment knob clockwise to increase the pressure, OR
TURN the adjustment knob counter-clockwise to decrease the pressure until the gauge reads between 70 to 80 psig.
B. IF psig reading cannot be achieved, THEN
NOTIFY the REF Lead.
7. LOCATE the filter holder pressure regulator, adjustment knob, and gauge (see Figure 8.4).
8. PULL the regulator knob up to release the knob lock, AND
ADJUST the filter holder pressure to read 20 psig on the gauge.
9. WHEN psig reading is achieved, THEN
PUSH the regulator knob down to relock.
A. IF psig reading cannot be achieved, THEN
NOTIFY the REF Lead.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 5 of 26
5.2 Initial Start-Up, (cont.)
10. LOCATE the aerosol generator pressure regulator, adjustment knob, and gauge (see Figure 8.4).
11. ADJUST the generator pressure to read 40 psig.
A. IF psig reading cannot be achieved, THEN
NOTIFY the REF Lead.
12. LOCATE make-up air flowmeter and adjustment knob on the left side of the cabinet (see Figure 8.5).
13. ADJUST the make-up air flowrate to 50 liters per minute (lpm).
14. READ the flowrate in the middle of the flowmeter float ball.
A. IF psig reading cannot be achieved, THEN
NOTIFY the REF Lead.
NOTE
If the tester has run for at least 15 minutes and then turned off for five minutes or less, no warm-up is required as noted in Step 5.2.15.
15. CLOSE the front door of the cabinet, AND
ALLOW filter tester to warm-up for 15 minutes.
5.3 Tester Set-Up for Testing Respirator Cartridges
1. SELECT OPTIONS from the main menu screen.
2. SELECT PROTOCOL from the Options menu.
3. SELECT SELECT TEST PROTOCOL from the Protocol menu.
4. WHEN testing MSA Ultratwin Cartridges and North Cartridges, THEN
ENSURE Protocol "0" is displayed.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 6 of 26
5.3 Tester Set-Up for Testing Respirator Cartridge, (cont.)
Step 4, (cont.)
A. IF Protocol "0" is displayed, THEN
1) PRESS INSTALL.
2) PRESS ESC until the main menu is displayed.
3) PROCEED to Step 5.3.6.
B. IF Protocol "0" is NOT displayed, THEN
1) PRESS NEXT until Protocol "0" is displayed.
2) PRESS INSTALL.
3) PRESS ESC until the main menu is displayed.
4) PROCEED to Step 5.3.6.
5. WHEN testing MSA MM2K PAPR filters, THEN
ENSURE Protocol "1" is displayed.
A. IF Protocol "1" is displayed, THEN
1) PRESS INSTALL.
2) PRESS ESC until the main menu is displayed.
3) PROCEED to Step 5.3.6.
B. IF Protocol "1" is not displayed, THEN
1) PRESS NEXT until Protocol "1" is displayed.
2) PRESS INSTALL.
3) PRESS ESC until the main menu is displayed.
6. ENSURE the filter holder is empty.
7. SELECT TESTER SET-UP option from the main menu screen, AND
WAIT until READY is displayed.
8. CLOSE the filter holder when READY is displayed on the screen by SIMULTANEOUSLY pressing both CLOSE buttons until the filter flow is displayed on the screen.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 7 of 26
5.3 Tester Set-Up for Testing Respirator Cartridge, (cont.)
9. SET flow for the filter test by using the flow control valve, located on the left front of the tester, as follows:
42.5 lpm for cartridges used in pairs, OR
85 lpm for cartridges used singularly
10. SELECT CONTINUE SET-UP from the screen display.
11. VERIFY the following messages appear on the screen:
SET-UP IN PROGRESS (This process includes the test finishing, the filter holder opens)
SET-UP COMPLETED
12. PERFORM filter testing.
5.4 Perform Filter Test
1. SELECT FILTER TEST on the main menu screen.
2. PLACE a filter in the lower half of the filter holder, AND
SIMULTANEOUSLY PRESS both CLOSE buttons. A TEST IN PROGRESS message will appear.
3. WHEN the test is complete, THEN
CONFIRM the filter holder OPENS and the results are shown on the display.
4. IF FAIL appears, THEN
A. DISCARD cartridge in the appropriate waste container
B. RECORD as a deficiency to be totaled for the day on OSR 4-510
C. CONTINUE testing the next cartridge.
5. IF PASS appears, THEN
PLACE the cartridge in the TESTED container for reuse.
6. REPEAT Steps 5.4.2 through 5.4.5 until all respirator cartridges have been tested.
7. PRESS ESC button to return to the main menu screen.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 8 of 26
5.5 Tester Set Up for Respirator Leak Testing
5.5.1 Mounting the Base Plate and Headform on the TSI 8127 Filter Tester for Testing Assembled Respirators
NOTE
A diagram on mounting the Model 8119 on the filter tester is found in Figure 8.6.
1. IF headform is already installed, THEN
PROCEED to Section 5.5.3.
2. REMOVE the bottom filter holder by loosening and removing the mounting screws using an Allen wrench, THEN
REMOVE the filter holder by lifting.
3. OBTAIN the headform base-plate.
4. APPLY silicone grease to the O-ring on the end of the insertion tube, on the bottom of the headform base plate, if needed.
5. NOTICE the large groove in the bottom of the plate.
6. With the headform facing toward the front of the tester, LOWER the headform base plate so the groove SLIDES over the filter-holder plate on the filter tester.
7. ALIGN the large hole in the headform plate with the hole in the filter-tester plate.
8. LOCATE the 5-1/2 inch diameter airflow distribution block.
9. PUSH the insertion tube on the flow distribution block DOWN into the hole in the base plate.
10. MOVE the base plate so the flow distribution block slips into the mounting hole.
11. ATTACH the flow distribution block using the two 2-inch button head screws.
12. PUSH the screws through the holes in the flow distribution block.
13. ROTATE the block so the screws align with the threaded holes in the filter tester plate.
14. TIGHTEN the mounting screws using the 1/8-inch hex wrench to secure both the mounting block and headform base-plate to the filter tester plate.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 9 of 26
5.5.2 Connecting the Aerosol Probe and Flow Tubing
1. CONNECT the probe to approximately 45 inches of 3/16-inch ID clear vinyl tubing.
2. FIND the free end of the tubing, AND
PUSH the end over the barbed aerosol probe connection on the flow distribution block.
NOTE
Aerosol probe and flow tubing can be left attached to head flow connection and flow distribution block when NOT in use.
3. CONNECT the 1/2-inch ID vinyl tubing to the back of the headform and to the head flow connection on the flow distribution block.
5.5.3 Setting Up the TSI 8119 Respirator Leak Tester
NOTE
Use caution when installing the respirator on the headform to prevent hand injury.
1. INSTALL a mask on the headform, AND
PULL the straps tight.
2. SELECT OPTIONS from the Main Menu to set the leak trigger level.
3. SELECT VOLTAGES from the Options Menu.
4. RECORD the Voltage value for the downstream photometer on ISF-REF-003, Voltage Value and Trigger Level Recording sheet.
A. RECORD daily for each machine in use. This is the background photometer signal. The photometer produces this signal when it samples clean air.
5. PRESS ESC to return to the Options Menu.
6. SELECT SYSTEM from Options Menu.
7. SELECT WATCH POINTS from the System Menu.
8. FIND the RESP option in the WATCH POINTS screen.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 10 of 26
5.5.3 Setting Up the TSI 8119 Respirator Leak Tester, (cont.)
9. IF the trigger level is .0004 volts greater than that recorded for the Voltage value, THEN
A. RECORD as the trigger level on ISF-REF-003 for each machine.
B. PRESS ESC to return to the Main Menu.
C. PROCEED to Step 5.5.3.15.
10. IF trigger level in NOT .0004 volts above the Voltage Level for the downstream photometer, THEN
PROCEED to Step 5.5.3.11 to set trigger level.
11. LOCATE AND SELECT the RESP option in the WATCH POINTS screen.
12. ENTER a number which will be the trigger level value for the respirator leak tests. The number entered must be larger than the background value recorded on ISF-REF-003.
A. START by entering a value that is about .0003 volts above the background.
Example:
Background .0003 Add .0003 Test adds .0001 After Saving Trigger Level .0007
B. IF the tester does not add .0001, THEN
ADD .0001 to make the total added to the background equal .0004.
Example:
Background .0003 Add .0004 Trigger Level .0007
13. WHEN entry is complete, THEN
A. RECORD trigger level on ISF-REF-003.
B. PRESS ENTER.
C. PRESS SAVE.
D. SELECT WATCH POINTS.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 11 of 26
5.5.3 Setting Up the TSI 8119 Respirator Leak Tester, (cont.)
Step 13, (cont.)
E. VERIFY machine displays trigger level input.
F. PRESS ESC three times to return to the Main Menu.
NOTE
If you entered the trigger level properly, the Main Menu has changed. "Respirator Test" option will appear in the Main Menu instead of the "Flow Adjustment" option.
14. RETURN to the WATCH POINTS screen in order to have the FLOW ADJUSTMENT option, AND
RESET the RESP to zero.
NOTE
If RESPIRATOR TEST option is not reset to zero main menu will always display
RESPIRATOR TEST.
15. SELECT the RESPIRATOR TEST Option from the Main Menu.
16. FOLLOW screen instructions to close the filter holder using the CLOSE buttons.
17. HOLD close buttons until the screen shown below appears.
A. IF previously in cartridge test configuration, THEN
CONFIRM the following:
Cartridge/Filter Test
Mask Flow 85 lpm for single cartridge or 42.5 lpm for dual cartridge
Probe Pressure 50 mm H2O
Signal Level 0.00033 (example) Volts
B. IF not in cartridge test configuration, THEN
CONFIRM the following:
Respirator Leak Test
Mask Flow 60 lpm
Probe Pressure 50 mm H2O
Signal Level 0.00033 (example) Volts
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 12 of 26
5.5.3 Setting Up the TSI 8119 Respirator Leak Tester, (cont.)
18. USE the flow control valve to adjust mask flow through the respirator to 60 lpm (± 3 lpm) once in the RESPIRATOR TEST mode.
5.5.4 Setting the Probe Pressure
1. ENSURE the tubing is attached to both the probe and the fitting on the flow distribution block (see Figure 8.6).
2. OPEN the front access door on the Filter Tester, AND
FIND the valve for the make-up air flowmeter.
NOTE
Probe pressure may be set lower or higher than 50 mm H2O with REF Lead approval.
3. TURN the make-up air valve until the probe pressure is 50 mm H2O on the display, not on the gauge (see Figure 8.5).
4. IF you have problems obtaining the correct probe pressure, THEN
A. CONTACT REF Lead.
B. REQUEST to add extra tubing to the exhaust to increase back pressure.
5.6 Positive Fit Check (MSA, North)
NOTE
When respirator processing is paused (i.e., breaks, meetings, lunch, etc.), the TSI machine does not have to be turned off. Turn off aerosol generator by pressing escape (esc.), options, aerosol off on the display. Restart respirator processing by pressing aerosol on, pressing escape (esc.) twice followed by respirator test then resume testing.
1. OPEN the filter holder
2. PLACE hand completely over the exhalation valve cover blocking all air from escaping the facepiece, AND
OBSERVE the facepiece lift slightly from the headform.
3. IF the facepiece lifts from the headform, the respirator has successfully passed the positive fit check, THEN
PROCEED to Section 5.7.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 13 of 26
5.6 Positive Fit Check (MSA, North), (cont.)
4. IF the facepiece does NOT lift from the headform, THEN
A. REMOVE the exhalation valve cover.
B. PLACE hand completely over the exhalation valve blocking all air from escaping the facepiece.
C. OBSERVE the facepiece to determine if it lifts from the headform.
D. RE-INSTALL the exhalation valve cover.
E. IF the facepiece lifts from the headform, THEN
PROCEED to Section 5.7.
5. IF the facepiece does NOT lift from headform, THEN
A. READJUST respirator on test head.
B. REPEAT Steps 5.6.2 and 5.6.3.
C. OBSERVE the facepiece to determine if it lifts from the headform.
6. IF respirator still does not lift, THEN
A. REMOVE respirator cartridges one at a time.
B. VISUALLY INSPECT that inhalation valves are in place.
7. IF either valve is missing, THEN
A. REPLACE valve(s), AND REPEAT Steps 5.6.2 and 5.6.3, OR
B. MARK with masking tape, AND
PLACE respirator in repair drum or container labeled “TO BE
REPAIRED”.
8. IF both inhalation valves are in place, THEN
ENSURE that the inhalation valves are NOT folded or warped.
9. IF a valve is folded or warped, THEN:
A. UNFOLD OR REPLACE valve, REINSTALL cartridge/filter, AND REPEAT Steps 5.6.2 and 5.6.3, OR
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 14 of 26
5.6 Positive Fit Check (MSA, North), (cont.)
Step9, (cont.)
B. MARK with masking tape, AND
PLACE respirator in repair drum or container labeled “TO BE
REPAIRED”.
10. IF valves are in place and do NOT appear to be folded or warped but the speaking diaphragm may be leaking, THEN
A. MARK the speaking diaphragm with masking tape.
B. PLACE in drum or container labeled “TO BE REPAIRED”.
11. IF facepiece lifts from the headform, THEN
A. CLOSE the filter holder.
B. PROCEED to Section 5.7.
5.7 Assembled Respirator Test
NOTE
The probe has a push-button valve on the handle which turns the aerosol flow from the probe ON and OFF. The push button must be continually pressed for the aerosol to flow.
1. GRASP the probe body, AND
PRESS the push-button to release the aerosol stream.
2. VERIFY aerosol is flowing through the probe by:
A. DIRECTING the probe under the respirator seal.
B. PRESSING the push button one to two seconds.
C. NOTE a signal level increase followed by a signal level decrease when the push button is released.
D. IF the increase – decrease is NOT noted, THEN
DISCONTINUE with processing respirators until diagnosis by the REF Lead is completed.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 15 of 26
5.7 Assembled Respirator Test, (cont.)
3. PLACE probe tip within 1/4-inch of mask surface, AND
MOVE probe around area being tested approximately one to two inches per second.
4. PROBE the sealing surface, lens, area behind cartridge receptacles, cartridge and cartridge sealing areas, speaking diaphragm, and exhalation valve using applicable sections.
5. IF no alarm sounds, THEN
A. OPEN AND REMOVE respirator.
B. PLACE the respirator in the “PASS” drum or container.
6. IF alarm sounds, THEN
PROCEED to the section below that is specific to the item alarm sounded for.
(Sections 5.7.1 through 5.7.6)
5.7.1 Sealing Surface
1. IF alarm sounds on sealing surface, THEN
A. RE-ADJUST the facepiece, OR
TIGHTEN the straps being careful not to over tighten.
B. RETEST the sealing surface of the entire facepiece.
2. IF alarm sounds, THEN
REMOVE facepiece, AND
INSPECT it for damage to the sealing surface.
3. IF damage is found, THEN
A. MARK with masking tape “sealing surface damaged”.
B. PLACE in designated area to be cannibalized and removed from
RITS.
C. MARK as deficiency to be totaled on OSR 4-510 at the end of day.
4. IF NO damage is found, THEN
NOTIFY the REF Lead for determination in how to proceed.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 16 of 26
5.7.2 Lens
1. IF alarm sounds on lens, THEN
READJUST respirator on test head, AND
SLOWLY REPROBE.
2. IF alarm sounds again, THEN
MARK the leak with masking tape, AND
PLACE respirator in drum or container labeled “TO BE REPAIRED”.
5.7.3 Area Behind Both Cartridges or Singular Cartridge
1. IF alarm sounds on area behind both cartridges or a singular cartridge, THEN
A. SLIGHTLY TIGHTEN cartridge receptacle by 1/4 turn while holding the inhalation valve seat.
B. RETEST.
2. IF alarm sounds, THEN
REMOVE cartridge receptacle, AND
INSPECT inhalation valve seat for damage.
A. IF inhalation valve seat damage is found, THEN
MARK with masking tape, AND
PLACE in drum or container labeled “TO BE REPAIRED”.
B. IF cartridge receptacle is damaged, THEN
REPLACE with a new one, AND
RETEST.
3. IF second alarm sounds, THEN
MARK the leak with masking tape, AND
PLACE respirator in drum or container labeled “TO BE REPAIRED”.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 17 of 26
5.7.4 Cartridge and Cartridge Sealing Area
1. IF alarm on cartridge or cartridge sealing area sounds, THEN
VERIFY cartridge is NOT cross-threaded.
2. IF cross-threaded, THEN
A. REMOVE cartridge, AND
CHECK the following for dents or damage while cartridge is off.
Inhalation Valve
Inhalation Valve seat
Cartridge
Gasket
Cartridge receptacle
B. RE-INSTALL AND RETEST cartridge.
3. IF second alarm sounds, THEN
REPLACE cartridge with new cartridge.
4. IF third alarm sounds, THEN
MARK cartridge/cartridge sealing area with masking tape, AND
PLACE in drum or container labeled “TO BE REPAIRED”.
5.7.5 Speaking Diaphragm
1. IF alarm sounds at the speaking diaphragm housing, THEN
CHECK for gaps in speaking diaphragm housing, AND
CHECK exhalation valve for dust, lent, and debris or damage.
2. IF alarm continues after checking, THEN
MARK with masking tape, AND
PLACE in drum or container labeled “TO BE REPAIRED”.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 18 of 26
5.7.6 Exhalation Valve
1. IF exhalation valve alarm sounds, THEN
A. REMOVE exhalation valve cover, AND
LIFT valve
B. ENSURE NO dirt, dust, or debris is causing the valve to remain open, AND
REPLACE cover, AND
RETEST.
2. IF alarm sounds on retest, THEN
A. REMOVE cover.
B. REMOVE valve.
C. INSPECT exhalation valve seat for damage.
D. IF seat is NOT damaged, THEN
1) REPLACE valve.
2) REINSTALL cover.
3) RETEST.
3. IF alarm continues after checking, THEN
PLACE masking tape on exhalation valve seat cover, AND
PLACE respirator in drum or container labeled “TO BE REPAIRED”.
4. IF no alarm or all alarms have been resolved, THEN
A. OPEN filter holder, AND
REMOVE respirator.
B. PLACE another respirator on the headform.
C. REPEAT Section 5.6.
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 19 of 26
5.8 Tester Shut Down
CAUTION
The filter holder will close when the power is turned off or terminated.
1. TURN the power switch to the OFF position. This will shut down the filter tester.
5.9 Post Performance
1. FORWARD completed OSR 4-510 forms to the REF Lead for evaluation and filing.
6.0 REFERENCES
Hazard Analysis INF-20083
1Q, Quality Assurance Manual, Procedure 15-1, Control of Nonconforming Items
4Q1.7, Procedure 104, Respiratory Equipment Facility Quality Verification
4Q1.7, Procedure 203, Inspection and Repair of Respiratory Equipment
4Q1.7, Procedure 216, Instructions for Inspection and Repair of the MSA Ultra Twin Respirator
4Q1.7, Procedure 218, Instructions for Inspection and Repair of the MSA OptimAir MM2K Respirator Equipped with the MSA Ultravue Facepiece
8Q, Procedure 12, General Site Safety Requirements
OSR 4-510, Respiratory Equipment Record and QA Form
ISF-REF-003, Voltage Value and Trigger Level Recording Sheet
7.0 RECORDS
Records generated as a result of implementing this procedure are maintained in accordance with Manual 1B, Procedure 3.31, Records Management.
OSR 4-510, Respiratory Equipment Record and QA Form
ISF-REF-003, Voltage Value and Trigger Level Recording Sheet
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 20 of 26
8.0 ATTACHMENTS
Figure 8.1 Diagram of the Aerosol Generator
Figure 8.2 Diagram of the Power Turn On/Off Switch
Figure 8.3 Diagram of the Main Pressure Regulator, Gauge, and Adjustment Knob
Figure 8.4 Diagram of the Filter Holder and Aerosol Generator Pressure Regulators
Figure 8.5 Diagram of the Make-Up Air Flowmeter and Valve
Figure 8.6 Diagram of Mounting the Model 8119 on the Filter Tester
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 21 of 26
FIGURE 8.1
Diagram of the Aerosol Generator
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 22 of 26
FIGURE 8.2
Diagram of the Power Turn On/Off Switch
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 23 of 26
FIGURE 8.3
Diagram of the Main Pressure Regulator, Gauge, and Adjustment Knob
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 24 of 26
FIGURE 8.4
Diagram of the Filter Holder and Aerosol Generator Pressure Regulators
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 25 of 26
FIGURE 8.5
Diagram of the Make-Up Air Flowmeter and Valve
Operation of TSI Filter Tester 8127 and TSI 8119 Respirator Leak Tester
Revision: 10 Page: 26 of 26
FIGURE 8.6
Diagram of Mounting the Model 8119 on the Filter Tester