Manual: 4Q1.7 Procedure: 204
Revision: 9 WASHER OPERATION Effective Date: 10/24/2016
Type-Class: Tech-Ref Page: 1 of 6
1.0 INTRODUCTION
1.1 Purpose
To provide instructions for operating the respirator washer at the Satellite Respiratory Equipment Facility (SREF) located in the 766-H Training Facility and at the Respiratory Equipment Facility (REF) located in Building 705-C.
1.2 Scope
This procedure applies to the use of the respirator washer in the REF and SREF when washing and sanitizing respirator equipment.
This is a Reference procedure. The user is not required to have this procedure present while performing the activity.
2.0 GENERAL INFORMATION
2.1 Definitions and Abbreviations
GPM - gallons per minute
MSA - Mine Safety Appliance
SDS – Safety Data Sheet
2.2 Responsibilities
1. REF Personnel are responsible for:
Operating the respirator washer
Completing qualification training as specified in the Job Performance Standard or approved equal prior to performing this procedure.
2.3 Equipment Description
1. The washer is front-loading and automatic. A revolving wash arm sprays the wash and rinse solutions. A 45-gpm capacity pump circulates the wash and rinse solutions, and empties the solutions during the drain cycles. The water temperature entering the washer should not be greater than 120°F. The total time for a complete cycle is 4-1/2 minutes. Special rack(s) are used to hold respirators in the washer.
2. The manufacturer’s recommendation is not to exceed 120°F. This is achieved by setting thermostat on the hot water heater to 120°F and running the water in the sink until it is hot. The temperature on the front of the washer should not read greater than noted temperature and no additional checks are necessary.
Washer Operation Revision: 9
Page: 2 of 6
2.3 Equipment Description, (cont.)
3. MSA cleaner-disinfectant is used for cleaning and sanitizing respirators.
3.0 PRECAUTIONS AND LIMITATIONS
1. A hazard analysis has been performed on this task. (INF-19991)
2. SDS 29354-1, Elimstaph #2 (Germicidal Cleaner Concentrate), Confidence Plus Quat Disinfectant or MSA Confidence Plus One-Step Germicidal Cleaner is used in the performance of this procedure.
3. Rubber gloves and safety glasses with side shields are required when handling MSA cleaner/disinfectant.
4. Comply with all facility safety rules.
5. Do not open washer door during wash cycle unless absolutely necessary.
6. Mats have been placed at washer to aid in prevention of slip and trip hazards.
4.0 PREREQUISITES
None
5.0 PERFORMANCE
NOTE
Rubber gloves and safety glasses with side shields are required when handling MSA cleaner-disinfectant.
5.1 Washer Operation
5.1.1 705-C Hobart Washers – Model Numbers WM-5C and WM-5
1. OPEN door, AND
PULL rack(s) out of washer onto door.
NOTE
1. Respirator facepieces and facepiece nosecups are not to be comingled with other respiratory equipment or items during washing operation.
2. DO NOT overload rack(s).
3. DO NOT place more than nine facepieces in each rack.
4. Racks shall be stacked no more than two high.
Page: 3 of 6
5.1.1 705-C Hobart Washers – Model Numbers WM-5C and WM-5, (cont.)
2. WHEN washing respirators, THEN
A. PLACE respirator facepieces in the rack(s) with the facepiece opening in the down position so that wash and rinse solutions can freely enter facepieces.
B. SLIDE the rack(s) into washer.
NOTE
1. DO NOT overload rack(s).
2. Racks shall be stacked no more than two high.
3. WHEN washing other respiratory equipment or approved miscellaneous items, THEN
A. PLACE respiratory equipment or items in the rack(s) with the item positioned so that wash and rinse solutions can effectively wash and rinse the respiratory equipment item(s).
B. SLIDE the rack(s) into washer.
4. IF using MSA cleaner-disinfectant, THEN
MEASURE AND ADD one (1) capful of MSA cleaner-disinfectant onto the interior of washer door after loading respiratory equipment/item(s).
5. CLOSE door, THEN
PRESS handle down.
NOTE
The manufacturer’s recommendation is not to exceed 120°F. This is achieved by setting thermostat on the hot water heater to 120°F and running the water in the sink until it is hot.
6. RUN water in the adjacent sink until water temperature is hot.
7. MOVE toggle switch to ON to start washer operation.
NOTE
1. If more than one load is being washed, the toggle switch may remain in the ON position until the last load is complete.
2. Washer will not operate if handle is not in the DOWN position.
8. OPEN washer door upon wash completion.
Page: 4 of 6
5.1.1 705-C Hobart Washers – Model Numbers WM-5C and WM-5, (cont.)
9. REMOVE respirator facepieces, respiratory equipment or other items.
10. SHAKE respiratory equipment/items to remove trapped rinse water.
11. CLOSE washer door, AND
MOVE toggle switch to OFF position.
5.1.2 766-H Hobart Washer – Model Number LX18C
1. RUN water in the adjacent sink until water temperature is hot.
2. PRESS ‘ON’ button on front of washer to run one (1) minute rinse cycle.
WHEN completed, THEN
TURN adjacent sink water OFF.
3. OPEN door, AND
PULL rack(s) out of washer onto door.
NOTE
1. Respirator facepieces and facepiece nosecups are not to be comingled with other respiratory equipment or items during washing operation.
2. DO NOT overload rack(s).
3. DO NOT place more than nine facepieces in each rack.
4. Racks shall be stacked no more than two high.
4. WHEN washing respirators, THEN
A. PLACE respirator facepieces in the rack(s) with the facepiece opening in the down position so that wash and rinse solutions can freely enter facepieces.
B. SLIDE the rack(s) into washer.
NOTE
1. DO NOT overload rack(s).
2. Racks shall be stacked no more than two high.
Page: 5 of 6
5.1.2 766-H Hobart Washer – Model Number LX18C, (cont.)
5. WHEN washing other respiratory equipment or approved miscellaneous items, THEN
A. PLACE respiratory equipment or items in the rack(s) with the item positioned so that wash and rinse solutions can effectively wash and rinse the respiratory equipment item(s).
B. SLIDE the rack(s) into washer.
6. IF using MSA cleaner-disinfectant, THEN
MEASURE AND ADD one (1) capful of MSA cleaner-disinfectant onto the interior of washer door after loading respiratory equipment/item(s).
NOTE
The manufacturer’s recommendation is not to exceed 120°F. This is achieved by setting thermostat on the hot water heater to 120°F and running the water in the sink until it is hot. The temperature on the front of the washer should not read greater than noted temperature and no additional checks are necessary.
7. CLOSE door, THEN
A. RUN water in the adjacent sink.
B. PRESS WASH button to start the washer operation.
C. OBSERVE temperature on front display of washer to ensure temperature reaches 120°F.
D. WHEN temperature is reached, THEN
TURN water to adjacent sink OFF.
8. OPEN washer door upon wash completion.
9. REMOVE respirator facepieces, respiratory equipment or other items.
10. SHAKE respiratory equipment/items to remove trapped rinse water.
11. CLOSE washer door
Page: 6 of 6
6.0 REFERENCES
Hazard Analysis (INF-19991)
SDS 29354-1, Elimstaph #2 (Germicidal Cleaner Concentrate), Confidence Plus Quat Disinfectant or MSA Confidence Plus One-Step Germicidal Cleaner
7.0 RECORDS
Records generated as a result of implementing this procedure are maintained in accordance with Manual 1B, Procedure 3.31, Records Management.