Exhibit D Analyzer, Calibrator and Training Specifications.pdf

PDF 142 KB Posted

Attached to
Air Monitoring Instrumentation, Replacement Parts and Servic
Solicitation number
RFP FAAA 2026000164 2
Issued by
Arapahoe County, Denver City, Colorado

About this file

This document is an exhibit of specifications for air monitoring instrumentation, replacement parts, and services required by the Colorado Department of Public Health & Environment Air Pollution Control Division. The specification exhibit outlines detailed requirements for Federal Reference Method (FRM) or Federal Equivalency Method (FEM) designated equipment that must be listed on the EPA's List of Designated Reference and Equivalent Methods as of December 15, 2025. The required equipment includes seventeen categories of analyzers and calibrators for measuring ambient air pollutants such as ozone, carbon monoxide, oxides of nitrogen, sulfur dioxide, and particulate matter (PM10 and PM2.5), along with zero air generators meeting specified purity standards and filter samplers with FRM designation. Additionally, the specifications call for other equipment including multi-port carbonyl samplers meeting PAMS specifications, a solar-powered remote air quality monitoring platform with meteorological capabilities and cellular data collection, and portable battery-powered mass flow controllers and meters for remote quality control environments.

Bidders must provide available replacement parts specific to each manufacturer's equipment being proposed and maintain full capability to repair all equipment in their own facilities. The contract includes comprehensive training requirements that bidders must deliver, specifically factory-level operation, maintenance, and advanced troubleshooting and repair training covering hands-on instruction for complete equipment operation with reference documentation, routine maintenance procedures at weekly, monthly, quarterly, and annual intervals, troubleshooting and diagnosis of equipment faults, repair of all equipment components, and certification documentation for all training completed. The specifications reference the EPA's List of Designated Reference and Equivalent Methods and the EPA's TAD R3 guidance documents to establish the technical baseline for all required equipment and quality control standards.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Exhibit D

SUMMARY OF ANALYZER, CALIBRATOR and TRAINING SPECIFICATIONS

FOR

COLORADO DEPARTMENT OF PUBLIC HEALTH & ENVIRONMENT

AIR POLLUTION CONTROL DIVISION

The items on the following list must have a Federal Reference Method (FRM) or Federal Equivalency Method

(FEM) designation, and be listed on the EPA’s List of Designated Reference and Equivalent Methods, most recently published on December 15, 2025. Equipment must be able to operate in a quality control environment that meets FEM standards.

FRM/FEM Equipment:

1. Ambient UV Adsorption Ozone Analyzer

2. Ambient Chemiluminescence Ozone Analyzer

3. Ambient Trace Level Carbon Monoxide Analyzer

4. Ambient Chemiluminescence Oxides of Nitrogen (NO/NO2/NOx) Analyzer

5. Ambient Trace Level Chemiluminescence Oxides of Nitrogen (NO/NO2/NOx) Analyzer

6. Ambient Trace Level “True” Nitrogen Dioxide (NO2) Analyzer with or without NO

7. Ambient Trace Level Total Oxides of Nitrogen (NO/NOy-NO/NOy) Analyzer

8. Ambient Trace Level Sulfur Dioxide Analyzer

9. Ambient Continuous PM10, PM2.5, PM10-2.5 Analyzer

10. Zero Air Generator meeting the following requirements:

a. less than 0.1 ppb SO2

b. less than 0.1 ppb H2S

c. less than 0.1 ppb NO

d. less than 0.1 ppb NO2

e. less than 0.1 ppb NH3

f. less than 0.4 ppb O3

g. less than 20 ppb CO

h. less than 5 ppb Hydrocarbons

11. Trace Level Zero Air Generator meeting the following requirements:

a. less than 0.025 ppb SO2

b. less than 0.025 ppb H2S

c. less than 0.025 ppb NO

d. less than 0.025 ppb NO2

e. less than 0.025 ppb NH3

f. less than 0.30 ppb O3

g. less than 10 ppb CO

h. less than 0.25 ppb Hydrocarbons

12. Ozone Calibrator

13. Trace Level Ozone Calibrator

14. Multi-Gas Dynamic Dilution Calibrator without Photometer (2 MFC) https://www.epa.gov/system/files/documents/2025-12/amtic-list-december-2025_508-compliant.pdf

Exhibit D

15. Multi-Gas Dynamic Dilution Calibrator with Photometer (2 MFC)

16. Trace Level Multi-Gas Dynamic Dilution Calibrator (3 MFC)

17. Ambient filter samplers (with FRM designation) for PM10 and PM2.5

Other Equipment:

1. Multi-port carbonyl sampler meeting PAMS sampling specifications.

2. Solar-powered, remote air quality monitoring platform offering meteorological monitoring, cellular service with online data collection/storage, and sensors, including, but not limited to: CO, NO, NO2, O3, SO2, PM10, and PM2.5.

3. Portable, battery-powered, mass flow controllers and mass flow meters designed for remote use in a quality control environment, at a variety of different ranges.

Parts:

Available replacement parts specific to each manufacturer’s piece of equipment being bid on.

Repairs:

Bidders must be fully capable of repairing their own equipment, in their own facility, when the need for such repairs arises.

Training:

1. Factory Level Operation, Maintenance, and Advanced Troubleshooting and Repair Training to include at a minimum:

a. Hands-on training for the complete operation of all provided equipment, to include reference documentation

b. Hands-on training for all routine (weekly, monthly, quarterly, and annual) maintenance procedures, to include reference documentation

c. Hands-on training for the troubleshooting and diagnosis of common and difficult equipment faults, to include reference documentation

d. Hands-on training for the repair of all equipment components, to include reference documentation

e. Certification documentation for all above mentioned training https://www.epa.gov/system/files/documents/2023-05/TAD%20R3%20May%202023.pdf