This document is a Statement of Work (SOW) for a federal contract opportunity to prepare the site at the Department of Veterans Affairs (VA) Palo Alto Health Care System (VAPAHCS) for the installation of a new Inpatient Pharmacy Robot. The key details are:

The project involves removing an existing outdated robot and installing a new 3-bay Omnicell Model XR-2 Robot, along with a new Yuyama packager. The work will be executed in multiple phases, including decommissioning and removing the existing equipment, site preparation, providing utility connections, and installing the new equipment. The contractor will be responsible for providing all labor, materials, equipment, and supervision required to complete the construction services. The work must comply with applicable federal and local codes, standards, and VA requirements, including interim life safety measures and infection control procedures. The anticipated duration of the work is 9 weeks for the robot and 1 week for the packager, to be completed prior to the scheduled go-live of the new equipment.

View the file

Other files

File Type Posted
Wage Decision Number CA20240018 12.20.2024.pdf PDF
36C26125R0014 0007.docx DOCX document
36C26125R0014 0006.docx DOCX document
36C26125R0014 0005.docx DOCX document
12.20.2024 Response to RFIs.docx DOCX document
12.13.2024 Site Visit Log.pdf PDF
36C26125R0014 0004.docx DOCX document
Updated Enclosure 9_Inpatient Pharmacy Robot_ICRA.pdf PDF
36C26125R0014 0003 SF30.docx DOCX document
36C26125R0014 0001.docx DOCX document
VA Palo Alto Site Plan.pdf PDF
36C26125R0014 0002.docx DOCX document
S02 RFI FORM.docx DOCX document
P07 DOL WAGE CA20240018 10.18.2024.pdf PDF
Enclosure 4_AIO Readiness Guide.pdf PDF
S02 - Past Performance Questionnaire ROBOT .docx DOCX document
Enclosure 11_B100 E1-120 GPR Report.pdf PDF
Enclosure 13_Exist Air Compressor-Air Dryer location.pdf PDF
Enclosure 12_Yuyama Equipment Specs.pdf PDF
Enclosure 3_Blister Packager Specs.pdf PDF
Enclosure 9_Inpatient Pharmacy Robot_ICRA.pdf PDF
S02 RFI FORM.docx DOCX document
SPECS.zip ZIP file
Enclosure 8_AIO Layout.pdf PDF
Enclosure 7_Reflected Ceiling Plan above Robot.pdf PDF
Enclosure 10_Pharmacy Robot Replacement B100 Structural Drawings.pdf PDF
Enclosure 2_Yuyama Packager Specs.pdf PDF
852.219-75 VA NOTICE OF LIMITATIONS SUBCONTRACTING-CERTIFICATE OF COMPLIANCE.docx DOCX document
Enclosure 5_Scanner Specifications.pdf PDF
Enclosure 14_New Patch Panel Location.pdf PDF
36C26125R0014.docx DOCX document
Cost Breakdown.docx DOCX document
Enclosure 6_Fit-Test Plan.pdf PDF
Show all 33

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

http://www.omnicell.com/ http://www.omnicell.com/ https://hcai.ca.gov/document/OPM-0619/

VHAPALGargM Rectangle

VHAPALGargM Text Box 19'L x 12'W x 7'-6"H

VHAPALGargM Rectangle

Arrow

Rectangle

Text Box For auto packager

VHAPALGargM Arrow

VHAPALGargM Callout 3 phase for robot

VHAPALGargM

Need 130 PSI at wall http://myomnicell.com/

Callout 208 VAC WYE 3 Phase, 30 Amp, 4pole 5 wire grounding service drop on e-power.

VHAPALGargM Callout 120 VAC 20A For Auto Packager

VHAPALGargM

4 Network drops http://myomnicell.com/

File details come from the government source that posted it. Updated .