This Data Item Description (DID) provides detailed requirements for calibration certificates used in Department of Defense (DoD) test, measurement, and diagnostic equipment contracts. Specifically, the document outlines the mandatory content and format for calibration reports or certificates, which must be consistent with ISO 17025 requirements. The calibration certificate must include comprehensive details such as the laboratory's name and address, unique report identification, client information, calibration method, item description, dates of receipt and performance, sampling procedures, test results, measurement uncertainties, environmental conditions, and evidence of measurement traceability. This version (DI-QCIC-80798B) supersedes the previous version DI-QCIC-80798A and is applicable to contracts requiring test equipment calibration prior to delivery.

View the file

Other files

Show all 12

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DATA ITEM DESCRIPTION

Title: CALIBRATION CERTIFICATE

Number: DI-QCIC-80798B Approval Date: 20050328 AMSC Number: F7554 Limitation:

DTIC Applicable: GIDEP Applicable:

Preparing Activity: AFMETCAL DET 1/MLE Applicable Forms: None Use, Relationships:

This Data Item Description (DID) contains the format and content requirements for calibration reports or certificates accompanying calibrations performed on DoD test, measurement, and diagnostic equipment.

This DID is applicable to contracts requiring calibration of test equipment prior to delivery. This DID is consistent with ISO 17025 requirements for calibration reports or certificate.

This DID supersedes DI-QCIC-80798A.

Requirements:

1. Format. Contractor format is acceptable.

2. Content. The calibration certificate or report shall include:

a. title (e.g., calibration report or calibration certificate);

b. the name and address of the laboratory, and the location where the calibrations were carried out, if different from the address of the laboratory;

c. unique identification of the test report or calibration certificate (such as the serial number), and on each page an identification in order to ensure that the page is recognized as a part of the test report or calibration certificate, and a clear identification of the end of the test report or calibration certificate;

d. the name and address of the client or customer;

e. identification of the method used;

f. a description of, the condition of, and unambiguous identification of the item(s) tested or calibrated;

g. the date of receipt of the test or calibration item(s) where this is critical to the validity and application of the results, and the date(s) of performance of the test or calibration;

Downloaded from http://www.everyspec.com

DI-QCIC-80798B

h. reference to the sampling plan and procedures used by the laboratory or other bodies where these are relevant to the validity or application of the results;

i. the test or calibration results with, where appropriate, the units of measurement;

j. the name(s), function(s) and signature(s) or equivalent identification of person(s) authorizing the test report or calibration certificate;

k. where relevant, a statement to the effect that the results relate only to the items tested or calibrated;

l. the conditions (e.g. environmental) under which the calibrations were made that have an influence on the measurement results;

m. the uncertainty of measurement or a statement of compliance with an identified metrological specification or clauses thereof;

n. evidence that the measurements are traceable.

End of DI-QCIC-80798B