DI-QCIC-80798C.pdf

Data Item Description Rev. C PDF 53 KB Posted

Attached to
Remanufacture of Air Launched Cruise Missile (ALCM) Air Cycle Machine (ACM)
Solicitation number
FA8119-24-R-0001_Solicitation
Issued by

About this file

This document provides a Data Item Description for calibration certificates or reports. It specifies the required format and content for calibration certificates or reports accompanying calibrations performed on Department of Defense test, measurement, and diagnostic equipment. The calibration certificate or report must include the title, laboratory name and address, unique identification number, client name and address, calibration method used, item description and identification, date of receipt and calibration, reference to any sampling plans or procedures, test or calibration results with measurement units, names and signatures of authorizing personnel, a statement about results only applying to calibrated items, environmental conditions affecting measurements, measurement uncertainty or statement of compliance, and evidence of traceable measurements.

The related federal contract opportunity is a solicitation from the Department of the Air Force Materiel Command Air Force Sustainment Center to remanufacture Air Launched Cruise Missile air cycle machines.

View the file

Other versions and copies of DI-QCIC-80798C

Show all 17

Who attaches DI-QCIC-80798C

Other files

File Type Posted
FA811924R0001_______0001.pdf PDF
DI-SESS-80639E.pdf PDF
FA8119-24-R-0001.pdf PDF
DI-MISC-81832.pdf PDF
DI-PSSS-81995A.pdf PDF
DI-MGMT-81838.pdf PDF
DI-MGMT-80441D.pdf PDF
DI-MGMT-80368A.pdf PDF
Performance Work Statement.pdf PDF
DI-NDTI-80809B.pdf PDF
AFMC 158_Packaging_Requirements.pdf PDF
DI-MGMT-82256A.pdf PDF
DI-MISC-80711A.pdf PDF
DD 1653_Transportation_Data.pdf PDF
DI-ADMN-81373.pdf PDF
DI-ADMN-81249C.pdf PDF
DI-PSSS-81656B.pdf PDF
DI-ADMN-81250C.pdf PDF
DI-MGMT-80501.pdf PDF
Contract Data Requirements List (CDRLS).pdf PDF
Wage Determination.pdf PDF
Quality Assurance Surveillance Plan (QASP).pdf PDF
Show all 22

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DATA ITEM DESCRIPTION

Title: CALIBRATION CERTIFICATE/REPORT

Number: DI-QCIC-80798C Approved Date: 20170123 AMSC Number: F9766 Limitation: N/A DTIC Applicable: No GIDEP Applicable: No Preparing Activity: 36 Project Number: QCIC-2017-003 Applicable Forms: N/A Use/Relationship: The Calibration Certificate/Report documents calibration certificates or reports accompanying the calibrations performed on Department of Defense (DoD) test, measurement, and diagnostic equipment.

a. This Data Item Description (DID) contains the format and content preparation instructions for the data product generated by the specific and discrete task requirement as delineated in the contract.

b. This DID is applicable to contracts requiring calibration of test equipment prior to delivery.

This DID is consistent with ISO/IEC 17025, General requirements for the competence of testing and calibration laboratories, requirements for calibration reports or certificates.

(Copies of this document are available online at http://www.iso.org.)

c. This DID supersedes DI-QCIC-80798B.

Requirements:

1. Reference documents. The applicable issue of the documents cited herein, including their approval dates and dates of any applicable amendments, notices, and revisions, shall be as specified in the contract.

2. Format. Contractor format is acceptable.

3. Content. The calibration certificate or report shall include:

a. Title (e.g., calibration certificate or calibration report);

b. The name and address of the laboratory and the location where the calibrations were carried out, if different from the address of the laboratory;

c. Unique identification of the calibration certificate or test report (such as the serial number), and on each page an identification in order to ensure that the page is recognized as a part of the test report or calibration certificate, and a clear identification of the end of the calibration certificate or test report;

d. The name and address of the client or customer;

e. Identification of the method used;

f. A description of, the condition of, and the unambiguous identification of the item(s) tested or calibrated;

g. The date of receipt of the test or calibration item(s) where this is critical to the validity and application of the results, and the date(s) of performance of the test or calibration;

DISTRIBUTION STATEMENT A. Approved for public release. Distribution is unlimited.

Source: https://assist.dla.mil -- Downloaded: 2017-03-13T13:39Z Check the source to verify that this is the current version before use.

DI-QCIC-80798C

h. Reference to the sampling plan and procedures used by the laboratory or other bodies where these are relevant to the validity or application of the results;

i. The test or calibration results with, where appropriate, the units of measurement;

j. The name(s), function(s) and signature(s) or equivalent identification of person(s) authorizing the calibration certificate or test report;

k. Where relevant, a statement to the effect that the results relate only to the items tested or calibrated;

l. The conditions (e.g., environmental) under which the calibrations were made that have an influence on the measurement results;

m. The uncertainty of measurement or a statement of compliance with an identified metrological specification or clauses thereof;

n. Evidence that the measurements are traceable.

End of DI-QCIC-80798C.