This document is a Calibration Standards Specification (CSS) for a Magnification Check Kit (MCK) for the Department of the Navy's Metrology and Calibration (METCAL) Program. The specification details technical requirements for a precision measurement tool designed for calibrating optical comparators at magnifications ranging from 5X to 250X. Key technical specifications include a reticle between 2-6 inches long with 0.02-inch graduations, and a scale at least 8 inches long with 0.1-inch graduations. The MCK must have a combined measurement uncertainty not exceeding ±100 microinches.
The document prescribes strict requirements for equipment components, including material restrictions (no mercury, radioactive materials, or carcinogens), environmental performance (operating at 60°F to 85°F with 20-60% humidity), and durability standards. The kit must withstand random vibration testing and transit drop tests, have a 12-month calibration interval with 85% of units remaining in tolerance, and include an identification plate with specific marking requirements. The NAVAIR METCAL Program Office, Team 3, serves as the Cognizant Design Activity, with maximum use of commercial off-the-shelf (COTS) components desired.
View the file
Other files
Show all 12
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
CSS FY21-JAX-0007
CALIBRATION STANDARDS SPECIFICATION
PROCUREMENT PACKAGE
MAGNIFICATION CHECK KIT
DISTRIBUTION STATEMENT A: APPROVED FOR PUBLIC RELEASE, DISTRIBUTION IS UNLIMITED.
DEPARTMENT OF THE NAVY
METROLOGY AND CALIBRATION PROGRAM
NOVEMBER 2022
CALIBRATION STANDARDS SPECIFICATION
PROCUREMENT PACKAGE
MAGNIFICATION CHECK KIT
Prepared for:
NAVAIR Metrology and Calibration (METCAL) Program
Prepared by:
Essence Sugar
METCAL Branch Fleet Readiness Center Southeast
Enterprise Avenue, Bldg. 6G 1st Floor Jacksonville FL 32212-0016
DEPARTMENT OF THE NAVY
METROLOGY AND CALIBRATION PROGRAM
MAGNIFICATION CHECK KIT CSS FY21-JAX-0007
i
TABLE OF CONTENTS
1.0 SCOPE
1.1 General
1.2 Usage
1.3 Design activity
1.4 Enclosure
1.5 Color
1.6 Configuration
2.0 APPLICABLE DOCUMENTS
2.1 Reference documents
2.2 Order of precedence
3.0 REQUIREMENTS
3.1 Performance requirements
3.1.1 Components
3.1.1.1 Reticle
3.1.1.1.1 Reticle length
3.1.1.1.2 Reticle graduations
3.1.1.2 Scale
3.1.1.2.1 Scale length
3.1.1.2.2 Scale graduations
3.1.2 MCK Uncertainty
3.1.3 Range
3.2 Workmanship
3.3 Equipment components
3.3.1 Interchangeability
3.3.2 Restricted materials
3.3.2.1 Material restricted for Navy use
3.3.2.2 Other restricted materials and gases
3.3.2.3 Organic materials
3.3.2.4 Carcinogens
3.3.2.5 Combination of materials
3.4 Enclosure requirements
3.4.1 Stacking provisions
3.4.1.1 Accessory stowage
3.4.1.2 Corners
3.5 Environmental requirements
ii
3.6 Vibration, random
3.7 Transit drop
3.8 Marking and identification
3.8.1 Marking
3.8.1.1 Warning markings
3.8.2 Equipment identification
3.8.3 Identification plate
3.8.3.1 Identification plate, material
3.8.3.2 Identification plate, marking
3.8.3.3 Item Unique Identification (IUID)
3.10 Maintainability
3.10.1 Fault isolation
3.10.2 Preventive maintenance
3.10.3 Maintenance provisions
1.0 SCOPE
1.1 General. This document specifies the minimum requirements for a Magnification Check Kit. This equipment may be of commercial design. It is intended to be used by Navy personnel in shore-based laboratories and for on-site, in-place calibration. The equipment will be used to calibrate traditional optical comparators, both horizontal and vertical configurations. For the purpose of this Calibration Standards Specification (CSS), the Magnification Check Kit herein shall be referred to as the MCK.
1.2 Usage. Equipment covered by this CSS is intended for use in controlled environments in which Military equipment is operated. The requirements of this CSS are tailored to the intended use of the equipment.
1.3 Design activity. This MCK product is specific to Navy calibration requirements.
Maximum use of commercial-off-the-shelf (COTS) items is desired. Should COTS items be unavailable, the overall product design, packaging, and life-cycle configuration management is under the cognizance of the Navy. The Cognizant Design Activity (CDA) for documentation and configuration management shall be the NAVAIR METCAL Program Office, Team 3.
1.4 Enclosure. For the purpose of this CSS, the enclosure refers to the case for each individual major component. The enclosure provides protection for the individual system components during storage, handling, and use.
1.5 Color. The color of the equipment and the transit cases shall be the color normally provided by the manufacturer but shall not be green, magenta, or red.
1.6 Configuration. The MCK shall have one configuration. The configuration shall allow each component to be replaces if damaged without need to replace the entire kit.
2.0 APPLICABLE DOCUMENTS
2.1 Reference documents. The following documents form a part of this specification to the extent specified herein. Unless otherwise indicated, the issues of these documents shall be those in effect on the date of this solicitation.
DEPARTMENT OF DEFENSE SPECIFICATIONS
MIL-PRF-28800G - Test Equipment for use with Electrical and Electronic Equipment, General Specification for
(Copies of DoD specifications are available online at http://assist.daps.dla.mil/quicksearch/ or from the Standardization Document Order Desk, 700 Robbins Avenue, Building 4D, Philadelphia, PA 19111-5094.)
DEPARTMENT OF DEFENSE STANDARDS
MIL-STD-130 - Standard Practice of Identification Marking of U.S. Military Property
(Copies of DoD Standards are available online at http://assist.daps.dla.mil/quicksearch/ or from the Standardization Document Order Desk, 700 Robbins Avenue, Building 4D, Philadelphia, PA 19111-5094.)
2.2 Order of precedence. In the event of conflict between the text of this CSS and the references cited herein, the text of this specification shall take precedence.
Nothing in this CSS, however, supersedes applicable laws and regulations unless a specific exemption has been obtained.
3.0 REQUIREMENTS
3.1 Performance requirements. The MCK shall adhere to the performance capabilities listed below.
3.1.1 Components. The MCK shall include at a minimum the following components.
3.1.1.1 Reticle. The reticle shall include a small scale on transparent material, which can be placed on an optical comparator work table. Only the reticle scale shall be visible when projected onto the optical comparator screen. Specifications for the reticle are as described in 3.1.1.1.1 Reticle length and 3.1.1.1.2 Reticle graduations.
3.1.1.1.1 Reticle length. The reticle shall be at least 2 inches in length, but not exceed 6 inches.
3.1.1.1.2 Reticle graduations. The reticle scale graduations shall be 0.02 inches at a minimum.
3.1.1.2 Scale. The MCK scale shall utilize the same materials in the reticle. These materials shall allow uniform thermal expansion between the reticle and scale.
Specifications for the MCK scale are as described in 3.1.1.2.1 Scale length and
3.1.1.2.2 Scale graduations.
3.1.1.2.1 Scale length. The scale shall be at least 8 inches in length.
3.1.1.2.2 Scale graduations. The scale graduations shall be 0.1 inches at a minimum.
3.1.2 MCK Uncertainty. The combined uncertainty of the MCK reticle and scale shall not exceed ±100 in.
3.1.3 Range. The MCK shall provide graduations on the reticle and scale capable of supporting 5X to 250X magnification, including, but not limited to, the following magnifications: 5X, 10X, 20X, 25X, 31.25X, 50X, 62.5X, 100X, 125X, 200X, and 250X.
3.2 Workmanship. The manufacture, assembly and packaging of the items covered by this CSS shall be representative of the best commercial practices and conform to the requirements of this CSS. All items shall remain securely held in place by the enclosure during transit. Adequate protection will be afforded for the equipment on all sides to avoid transient damage.
3.3 Equipment components. Parts, materials, and process shall be as specified below in 3.3.1 Interchangeability through 3.3.2.5 Combination of Materials.
3.3.1 Interchangeability. A model number shall be designated for the unit and all components provided within the MCK. All units and the replaceable parts comprising the unit shall be interchangeable on a noninterference basis.
3.3.2 Restricted materials. Restricted materials shall be as specified below in 3.3.2.1 Material restricted for Navy use through 3.3.2.5 Combination of materials.
3.3.2.1 Material restricted for Navy use. Mercury and Radioactive Materials shall not be used in equipment for the Navy.
3.3.2.2 Other restricted materials and gases. Test equipment shall not contain materials that promote the premature failure of the equipment or that present a health hazard under normal use and storage conditions.
3.3.2.3 Organic materials. Test equipment shall not contain materials prone to organic decay or that act as a nutrient for the growth of multi-celled organisms.
3.3.2.4 Carcinogens. Certain chemicals have been identified in the Occupational Safety and Health Act (OSHA) as cancer-producing substances (carcinogens).
Materials that might contain these chemicals shall not be used unless specifically required and approved by the NAVAIR METCAL Program Office, Team 3.
3.3.2.5 Combination of materials. The equipment shall contain no combination of materials that cause deterioration of any material contained in the equipment in normal storage and operating conditions throughout normal life expectancy of the equipment.
3.4 Enclosure requirements. The enclosure shall also serve as the transit case, and adequately protects the MCK components from dirt, moisture, or contaminates during shipping and storage. The physical characteristics of the enclosure shall be as specified in 3.4.1 Stacking provisions through 3.4.1.2 Corners.
3.4.1 Stacking provisions. The MCK enclosure shall have a geometric configuration that permits stacking without harm to the enclosure or its contents.
3.4.1.1 Accessory stowage. Enclosures that have provisions for stowage of accessories, operating instructions, and designated spares shall be contained in a manner that prevents damage to the accessories or MCK components during transit.
3.4.1.2 Corners. All corners shall be designed to preclude injury to personnel or damage to material. All corners shall be adequately reinforced to protect the components from damage during shipping, handling and storage.
3.5 Environmental requirements. The MCK shall conform to the specified performance and accuracy while being stored, or operated at temperatures within 60°F to 85°F and relative humidity 20% to 60%.
3.6 Vibration, random. The MCK shall conform to the performance and accuracy requirements after the random vibration conditions, specified in MIL-PRF- 28800G Table 3, when tested in accordance with MIL-PRF-28800G paragraph
4.5.5.3.1 Random vibration test.
3.7 Transit drop. The MCK shall conform to the specified performance and accuracy requirements after being tested in accordance with MIL-PRF-28800G paragraph 4.5.5.4.2 Transit drop test.
3.8 Marking and identification. Marking and identification of test equipment shall be as specified below in 3.8.1 Marking through 3.8.3.3 Item Unique Identification
(IUID).
3.8.1 Marking. Test equipment shall be marked in a manner that does not adversely affect safety or performance.
3.8.1.1 Warning markings. Warning markings shall warn of the location, the nature, and the extent of a hazard. Letters of warning markings shall be clearly legible gothic capitals. Warning markings shall have high contrast between the letter and background colors. The markings shall be as permanent as normal life expectancy of the MCK, and located as close as possible to the point of danger.
3.8.2 Equipment identification. All components shall be identified with the items specified in a through d.
a. Manufacture name
b. Equipment model number
c. Equipment nomenclature
d. Serial number
3.8.3 Identification plate. An equipment identification plate shall be affixed to the MCK enclosure. The MCK shall be contained within a single case. The identification plate shall contain the information provided in a through e.
a. Configuration model number or equipment model number.
b. Configuration or equipment nomenclature.
c. Configuration or manufacturer CAGE.
d. Configuration serial number.
e. Consists of: Part number of all components comprising the equipment configuration.
3.8.3.1 Identification plate, material. The following materials with minimum thickness of 1/32 inch are recommended.
a. Brass
b. Corrosion resistant steel
c. Aluminum alloy
3.8.3.2 Identification plate, marking. The plate shall have a minimum character size of 1/8 inch. All characters shall be permanent using best commercial practices for laser engraving, etching, or photosensitive marking of material.
3.8.3.3 Item Unique Identification (IUID). DFARS clause 252-211.7003 is invoked. A two dimensional Item Unique Identification (IUID) data matrix conforming to MIL-STD-130 shall be included on the identification plate.
3.9 Calibration Interval. For the purpose of this CSS, calibration refers to the process of comparing a test instrument with a calibration standard to determine the indicated error at specified test points. The minimum interval between calibrations shall be 12 months. At the end of this interval, a minimum of 85 percent of the units shall remain in tolerance.
3.10 Maintainability. Maintainability shall be as specified in 3.10.1 Fault isolation through 3.10.3 Maintenance provisions.
3.10.1 Fault isolation. The MCK design shall permit isolation of faults, and replacement of the faulty component.
3.10.2 Preventive maintenance. Preventive maintenance shall not require more than 15 minutes per a 30-day period. Preventive maintenance shall not require breaking of the equipment seams where calibration seals would normally be placed.
3.10.3 Maintenance provisions. Equipment shall be as specified in a through c to facilitate maintenance:
a. Accessibility for maintenance. The equipment shall be designed to ensure that any necessary maintenance actions can be accomplished utilizing non-proprietary tools.
b. Maintenance and calibration aids. Special tools required for maintenance or calibration shall be provided. They shall be identified as a set with a unique part number and national stock number (NSN).
Stowage shall be in a reusable pouch or container.
c. Accessibility for repair. The MCK design shall allow individual components to be replaced with an identical unit if the component becomes damaged.