D.1 Evaluated Salient Characteristics.xlsx

XLSX spreadsheet 35 KB Posted

Attached to
6515--Surgical Lights-Booms-Video Integration BHH
Solicitation number
36C26325Q0777
Issued by

About this file

The file is an evaluated salient characteristics spreadsheet detailing technical specifications for surgical lights, booms, and video integration systems for the Black Hills VA Healthcare System. The solicitation (36C26325Q0777) is a Service-Disabled Veteran-Owned Small Business (SDVOSB) set-aside procurement using FAR 13.5 procedures. The equipment requirements include six surgical lights with specific LED characteristics such as 6 lightheads, 110/120V power, color temperature adjustable between 3500-5000K, and a color rendering index over 90. The video integration system must support unified routing of multiple video signals, high-definition output, remote viewing capabilities, touchscreen controls, and storage of digital video and images.

The procurement requires vendors to submit comprehensive documentation, including a signed SF1449, price schedules, authorization letters from original equipment manufacturers, Buy American Certificate, specification sheets, and a pre-recorded demonstration of the video integration system. The system must integrate with existing VA electronic medical records, provide live video streaming, support multiple simultaneous video sources, and include features like wall-mounted cameras, touch screen controls, and the ability to manage operating room resources such as video routing, camera positions, and light intensities. The total site quantities include 6 surgical lights, 3 equipment booms, 3 anesthesia booms, and 3 video integration systems for the Black Hills VA Healthcare System.

View the file

Other files

File Type Posted
D.9 Room dimensions.pdf PDF
36C26325Q0777 A0001.pdf PDF
36C26325Q0777 A0001_1.docx DOCX document
36C26325Q0777_1.docx DOCX document
36C26325Q0777.pdf PDF
D.5 ICRA Introductory Information and Instruction.docx DOCX document
D.2 Supply Line Item Template.xlsx XLSX spreadsheet
D.4 Surgery structure.pdf PDF
D.3 Floor plans.pdf PDF
D.8 BUSINESS ASSOCIATE AGREEMENT.docx DOCX document
D.7 VA Directive 6550 App A Fillable.pdf PDF
D.6 PCRA Introductory Information and Instructions.docx DOCX document
Show all 12

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Surgical Lights

ItemDescriptionBlack Hills-OR 1-3Explanation of how vendor proposal meets this criteriaTotal
1LIGHT SOURCE, TYPELED
1aNumber of lightheads66
1bNumber of bulbs or elementsNo minimum requirements
1cVolts110/120
1dLife, hrGreater than 30,000 LED
2COLOR TEMPERATURE, KAdjustable from 3500-5000
2aAdjustableRequired-Touch at light and control display
3COLOR RENDERING INDEX>90
3aR9 valueminimum 90
4FIELD SIZE, cm (in)
4aDiameterNo minimum requirements
5FOCAL LENGTH, cm (in)≥65
6ILLUMINATION LEVEL, maximum lux at 1 m≥160,000
7MOUNTINGBoom only
8CONTROLS
8aDimmerRequired, At light and control display. 5 -100%
8bFocusRequired, At light and control display
8biAutomaticRequired
8cField sizeNo minimum requirements
8dOn sterile handleRequired
9ROTATION, °The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.
10VERTICAL ADJUSTMENT RANGE, cm (in)≥80 cm
11HEAT
11aMaximum irradiance at 1 m, W/m2No minimum requirements
11bHeat-to-light ratio, mW/m2.luxNo minimum requirements
12LIGHTHEAD MATERIALNo minimum requirements
13LENS COVER MATERIALNo minimum requirements
14FILTER MATERIALNo minimum requirements
15REFLECTOR MATERIALNo minimum requirements
16SHADOW CONTROL
16aAutomatic shadow dilution adjustmentAutomatically by in-light control
17SATELLITESNo minimum requirements
18AMBIENT LIGHTINGRequired
19BATTERY OR EMERGENCY BACKUPRequired
20STERILIZABLE HANDLERequired
21VIDEO CAMERA
21aCompatible (type) or integrated-separate video source incorporated into one of the surgical lights brought to one of the surgical fields streaming live videoYes, HD minimum
21bLocation on lightheadLight
22MINIMUM CEILING HEIGHT, cm (in)10 ft
23DIAMETER, cm (in)
23aLightheadCURRENTLY 24", NOT REQURED BUT SOMETHING CLOSE
24WEIGHT, kg (lb)No minimum requirements
25SUPPLY VOLTAGE, VACNo minimum requirements
26POWER, WNo minimum requirements
27PURCHASE INFORMATION
27aWarrantyMinimum of 1 year
28ADDITIONAL SPECIFICATIONS
28aAbility to turn light on/off at lightNo minimum requirements
28bRecent Upgrades

Booms

ItemDescriptionBlack Hiills-OR 1-3Explanation of how vendor proposal meets this criteriaTotal
1Number of Anesthesia Booms Per Room33
2Number of Equipment Booms Per Room33
3SWIVEL RADIUS, cm (in)No minimum requirement
4HEIGHT ADJUSTMENT, cm (in)No minimum requirement
5ELECTRICAL FEATURES
5aCircuit breakerNo minimum requirement
5bElectrical power receptacles10
5cEquipotential ground busNo minimum requirement
5dGrounding jacksNo minimum requirement
5eTelephone jackNo minimum requirement
5fIsolated power systemRequired
5gCode alarm buttonNot required
5hNurse call systemNot required
5iElapsed timer/clockNot required
6MEDICAL GAS SERVICES
6aOxygen outlets2
6bOutlet typeDiamond
6cVacuum inlets3
6dMedical compressed air1
6eNitrous oxide1
6fWAG evacuation1
6gMedical gas pipingNo minimum requirement
6hVacuum bottle slidesNo minimum requirement
6iVacuum bottle storageNo minimum requirement
7ACCESSORIES
7aMonitor bracketREQUIRED
7bBed bumperNo minimum requirement
7cIV supportsNo minimum requirement
7dCatheter basketNo minimum requirement
8H x W x D, cm (in)No minimum requirement
9WEIGHT, kg (lb)No minimum requirement
10LOAD CAPACITY, kg (lb)350 lbs or greater per arm
11PURCHASE INFORMATION
11aWarrantyMinimum of 1 year
12ADDITIONAL SPECIFICATIONS
12aNitrogenMust have regulated Nitrogen outlet/w gauge
12bNetwork Jacks4 plex or more
12cKeyboard/mouse1
12dAdustable Equipment ShelvesRequired
12eCarbon Dioxide on equipment boomRequired
12fMedical Gas ConnectionsEach room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging 12g Universal connections for video inputs •Equipment boom shall have 6 universal connections for video inputs •Anesthesia boom shall have 3 universal connections for video inputs

12hVideo integration Foot Switch / Sync cableNo minimum requirement
12iSmoke EvacuatorNo minimum requirement
12jStorage space for ESU, electrosurgical unit, foot switchesNo minimum requirement
12kMounts for display monitorsMust have mounts for at least 3 display monitors
12lCO2 Cylinder mountNo minimum requirement
12mArm SupportNo minimum requirement
12nVideo Integrtion overall needsThe system must be capable of capturing any video signal and sending it to the Integration storage system, It then must be able to send this information to the nurses control display for them to push to the room displays for viewing during the case. It must be able to record the video from the case and transfer it to a medium such as a disk for the patient for example to take or be sent to another facility for physician viewing if follow up care is needed. It also must be able to provide the playing of music for the surgeons and or staff for listening to before, during and after the case. The integration needs to be able to send images to the current eMR, patient electronic medical records system.
12mRecent Upgrades

Video Integration

ItemSpecification from Previous ProcurementBlack Hills-OR 1-3Explanation of how vendor proposal meets this criteriaTotal
1Qty of Video Integration System33
2Unified routing of disparate video signals from VA furnished equipment to desired display destinations. Video sources will include, but not limited to mobile and stationary endoscopic video systems, lighthead camera systems, urology systems, mobile c-arms, PACS, vital signs monitor, and the OR desktop computer.Required
3System will provide wall mounted pan/tilt/zoom camera articulation and viewing.Required
4System will improve ability to utilize OR protocols such as a timeout.Required
5Touch screen control of video source selection and video routing.Required
6Improve OR workflow and efficiency.Required
7Manipulation of VA information systems on the same workstation controlling video routing.Required
8Viewing of video feed from another active video integration solution within the hospital.Required
9High definition or better video output to all display monitors integrated to the system.Required
10Features that improve the ability to view and manipulate video feeds.Required
11Multiple video sources displayed simultaneously on a single display.Required
12Storage of digital video and digital images.Required
13Live video feed provided remotely to aid OR scheduling and workflow.Required
14System permits users to interact with surgery staff and video via remote connection.Required
15Contractor shall provide a description of site-preparation responsibilities identifying utilities, conduit, bandwidth and all other activities related to preparing the room for installation.Required
16Contractor shall furnish all cabling and hardware associated with the integration solution and provide a plan that addresses installation responsibilities of VA and contractor.Required
17Contractor must provide monitors and mounting for viewing case data both in the sterile field and outside of it.Required
18Specified monitors must be able to withstand routine cleaning practices typical to sterile operating room settings.Required
19Specified monitors must be medical grade (electrically shielded and NFPA 99 compliant).Required
20Router must be able to route multiple video sources simultaneously in various configurations as a split screen on a single display.Required
21Routed video signals should maintain native resolution from source to destination with minimal degradation.Required
22Router shall accept, process, and permit routing of at minimum 8 simultaneously broadcasted input signals in various combinations of different formats (i.e. 8x8 inputs/outputs).Required
23Contractor must provide acceptable furniture that houses their integration system.Required
24Integration control consoles (ICC) must be managed with touch screen displays for configuring, routing and controlling information and devices managed by the integration systemRequired
25The ICC must have preset capability. Presets must control the following operating room resources: Video routing from an imaging device to a preferred display. Systems will also manage in room camera positions from presets, room light levels, and intensities generated by surgical lights.Required
26Users must be able to preview any signal data prior to sending it to a monitor or anywhere else in the operating room.Required
27The system must accommodate all major audio and video input connection types and signal formats.Required
28To accommodate various signal types the system must utilize a universal connection port.Required
29System must be able to convert and scale various signal types so signals will be recognized by displays if necessary.Required
30Universal connection points must be able to be deployed on medical equipment booms. Universal connection points must also be able to be deployed in wall or ceiling locations as specified by the VA.Required
31For devices that the integration system recognizes and can interface with, the connection must be automatic and not require any configuration or troubleshooting on behalf of the clinical user. Device recognition must occur with 120 seconds of connecting the device to the integration system.Required
32ADDITIONAL SPECIFICATIONSNo minimum requirement
32aMonitors - Up to single 55 inch wall mount and single 32 inch or dual 26 inch displays.Required
32bSync CablesRequired
32cNetwork Connection for Video InputRequired
32dStorage of Images; scalable at 5 TB increments, so minimum is 5 TB. With ability to push and pull Endosoft data.Required
32eWall video/capture/data inputs. Imaging from wall inputs mapped to video integration, to include C-ARM.Required
32fRecent Upgrades

Total Site Quantities

SiteSurgical LightsEquipment BoomsAnesthesia BoomsVideo Integration Systems
Black Hills6333
Total6333