Bldg 1 Full Set_Part6.pdf
PDF 4 MB Posted
- Attached to
- Y1DA--PN: 636A6-26-201, Patient Ceiling Lifts Install & Testing (DM) AP: 36C263-26-AP-1431
- Solicitation number
- 36C26326R0059
About this file
These are architectural "As Bid Drawings" for a healthcare facility addition and remodel, specifically detailing electrical lighting and reflected ceiling plans for a medical work area. The drawings are prepared by MSH Architects with consulting support from West Plains Engineering, Inc.
The project encompasses approximately 1,550 square feet of main work area plus 150 square feet of corridor space, containing multiple patient recovery and treatment rooms (numbered 4300-4322) including GI recovery, nourishment, electrolyte, restroom, outpatient recovery, and nurse station areas. The drawings specify detailed electrical lighting fixtures with a comprehensive lighting fixture schedule, ceiling panel specifications including 1-hour fire-rated panels, and patient lift support systems with structural attachment requirements for ceiling-mounted medical equipment. Key technical specifications include patient lift parallel rails, traverse rails, gypsum board wall assemblies with metal stud framing, diagonal bracing requirements, and wedge anchor installation protocols with minimum spacing requirements. The reflected ceiling plan shows fixture placement, coordination marks cross-referenced to floor plans, and detailed notes regarding lift system contractor responsibilities for installation. Architectural details reference coordination with existing structure conditions, pan joist integration, and compliance with Horcher Medical System standards for all fastening and anchoring components.
View the file
Other files
Show all 42
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
GYPSUM WALL BOARD SOFFIT
CEILING MOUNTED TRACK FURNISHED AND INSTALLED BY CONTRACTOR,
HOSPITAL CUBICAL CURTAIN FURNISHED AND INSTALLED BY OWNER.
SWING ARM TV MOUNT ATTACHED TO FLOOR DECK ABOVE (INSTALL BLOCKING
SUPPORT AS REQUIRED) (MOUNTING SYSTEM TO BE CAPABLE OF ACCEPTING A 22"
FLAT SCREEN TV)
CROWN MOLDING3
TRAVERSE PATIENT LIFT SYSTEM (REUSE SALVAGED MATERIALS) (NEW
MATERIALS TO BE FROM SAME MANUFACTURER AS SALVAGED MATERIALS)4
NEW 2' X 2' SUSPENDED ACOUSTICAL CEILING IN CORRIDOR. MATCH
EXISTING. REPLACE ALL EXISTING MECHANICAL, ELECTRICAL AND OTHER
ACCESSORIES BACK IN ORIGINAL LOCATIONS.
IF PHASE 1 CEILING WORK IN CORRIDOR EXTENDS BEYOND 2 WEEKS,
TEMPORARY 1 HR RATED CEILING PANELS WILL NEED TO BE INSTALLED UNTIL
PHASE 1 WORK IS COMPLETED. (SEE DETAIL 3 ON THIS SHEET)
8 PROVIDE BLOCKING ABOVE CEILING FOR FUTURE CEILING MOUNTED PRIVACY
CURTAINS AT THIS LOCATION.
A. THE REFLECTED CEILING PLAN MAY NOT INDICATE ALL CEILING MOUNTED DEVICES
AND FIXTURES
B. PROVIDE METAL BLOCKING AND BACKING AS REQUIRED FOR EQUIPMENT AND
ACCESSORIES.
C. VERIFY ALL UTILITIES AND CONNECTION LOCATIONS.
D. CONTRACTOR SHALL VERIFY ALL EXISTING CONDITIONS AND DIMENSIONS PRIOR TO
BEGINNING CONSTRUCTION.
E. ALL SURFACES OR FINISHES DAMAGED DURING CONSTRUCTION SHALL BE REPAIRED
OR REPLACED AND SHALL MATCH ADJACENT MATERIALS.
F. ALL WORK MUST BE COORDINATED CLOSELY WITH THE VA PROJECT ENGINEER TO
NOT ADVERSELY AFFECT OPERATIONS OUTSIDE OF THE WORK AREA. TEMPORARY 1-
HOUR FIRE RATED CONSTRUCTION PARTITIONS WILL BE REQUIRED WHEN THE WORK
AREA EXTENDS INTO AND AFFECTS EXISTING AREAS AND OPERATIONS.
G. IT SHALL BE THE CONTRACTOR'S RESPONSIBILITY TO PROVIDE ADEQUATE SHORING
AND BRACING OF THE STRUCTURE FOR ALL LOADS THAT MAY BE IMPOSED DURING
CONSTRUCTION.
H. IT IS NOT ACCEPTABLE TO CUT NEW OPENINGS IN EXISTING CONCRETE OR MASONRY
WITHOUT PRIOR APPROVAL FROM THE ARCHITECT. A LAYOUT OF ANY PROPOSED
OPENINGS SHALL BE REVIEWED BY THE ENGINEER PRIOR TO SAWCUTTING.
I. CONTRACTOR MUST FIELD VERIFY THE EXACT LOCATION OF JOIST GRIDS AS PAN WIDTH
MAY VARY.
J. IT IS NOT PERMISSIBLE TO DRILL HOLES IN ANY CONCRETE SLABS TO SUPPORT THE
PATIENT LIFTS.
K . DOUBLED UP ANGLES WILL NOT BE ACCEPTABLE FOR CEILING GRIDS AT WALL EDGES.
CEILING LEGEND
2' x 2' SUSPENDED ACOUSTICAL CEILING GRID
GYPSUM BOARD CEILING / SOFFIT
DIFFUSER
RETURN
2 x 4 LIGHT FIXTURE
2 x 2 LIGHT FIXTURE
DOWN LIGHT FIXTURE
CEILING MOUNTED EXIT LIGHT FIXTURE
EXHAUST FAN
FIRE SPRINKLER
AI-501
AI-501
AI-501
AI-501
9' - 4"
8' - 5"
7' - 8"8' - 8"
7'
8"
7'
8"
8'
5"
8'
5"
9' - 4"
7' - 7"
GI RECOVERY
4314B
NOURISH.
ELECT.
REST.
REST.
4316A
REST.
LOCKER
OUTPATIENT
RECOVERY
AREA
1'-8 5/8" 1'-0 1/2"
2'-5"
AI-103
2' -5
1' -0
3/
4"
2' -5
1'
-0
6' -8
/4
6'
-6
4' -9
/2
2'
-5
1' -0
7' -1
0"
1'-0" 2'-5"
4314A
NURSE
7'
7"
7' - 7"
8'
5"
7' - 1"
9' - 2"
9' - 2"
4' -2
4'
-1 1"8'-0"
9' - 4"6 6
AI-101
7'-9 1/16"
7' -4
7' -3
/2
7'-9" 5'-8" 5'-8" 5'-8"
6' -3
/4
5'
-9
/4
5' -9
/4
5'
-9
/4
AI-101
AI-103
7'-8 15/16"
AI-501
2'-4" 2'-3"
4'- 0 7
/16
3'- 7 3/4"
2'-10" 2'-10"
5' -0
3'-1" 2'-7"
1' -7
3' -4
2' -1
1/
4"
4'-5" 3'-4"
(MATCH EXISTING)
7 7
2' -3
1'-0"
1' -6
AI-103
TYP.
CORRIDOR
43034305
416B
43174319
NORTH
4TH FLOOR
BUILDING 1
AREA OF WORK
OUTPATIENT
SURGERY/GI AREA
7" MIN.
VARIES
FIELD VERIFY
1'-8"TYP.
FIELD VERIFY
5"TYP.
4" M
IN
FASTEN ALL ANGLES, BOLTS,
UNISTRUT, AND WEDGE. ALL
ANCHORS PER HORCHER
MEDICAL SYSTEM
UNISTRUT AS REQUIRED FOR SPANNING
TO STRUCTURE SUPPORTS (BY LIFT
MANUFACTURER)
THREADED ROD (BY LIFT
MANUFACTURER)
PATIENT LIFT PARALLEL RAIL
PATIENT LIFT TRAVERSE RAIL
* DO NOT INSTALL WEDGE ANCHORS IN THE SAME JOIST BACK TO
BACK. ONE JOIST SPACING MINIMUM MUST OCCUR BETWEEN
PATIENT LIFT SUPPORTS TO STRUCTURE.
* MINIMUM CENTER TO CENTER SPACING OF 2'-0" MUST OCCUR
BETWEEN STEEL SUPPORTS ON THE SAME JOIST.
* LIFT MANUFACTURER TO USE UNISTRUT TO SPAN TO
SUPPORTS AS REQUIRED
LINE OF FINISHED CEILING
(FIELD VERIFY)
DIAGONAL BRACE PERPENDICULAR TO
JOIST BEARING AS NEEDED BY
MANUFACTURER'S RECOMMENDATIONS
* WHEN DIAGONAL BRACE IS PARALLEL TO JOIST SPAN DO NOT BRACE TO
ADJACENT INTERIOR STUD WALLS, ANGLE BRACE TOWARDS CLOSEST
JOIST SIDE FOR ATTACHMENT.
* LIFT SUPPLIER TO COORDINATE AND REVIEW EXISTING PROJECT
CONDITIONS FOR INSTALLATION OF NEW COMPONENTS TO TIE INTO
EXISTING SALVAGED COMPONENTS. LIFT SUPPLIER TO INSTALL ENTIRE
SYSTEM.
EXISTING CONCRETE
FLOOR & PAN JOIST
7" MIN.
VARIES
FIELD VERIFY
1'-8"TYP.
FIELD VERIFY
5"TYP.
4" M
IN
FASTEN ALL ANGLES, BOLTS,
UNISTRUT, AND WEDGE. ALL
ANCHORS PER HORCHER
MEDICAL SYSTEM
UNISTRUT AS REQUIRED FOR
SPANNING TO STRUCTURE
SUPPORTS (BY LIFT
MANUFACTURER)
THREADED ROD (BY LIFT
MANUFACTURER)
PATIENT LIFT PARALLEL RAIL
PATIENT LIFT TRAVERSE RAIL
LINE OF FINISHED CEILING
(FIELD VERIFY)
THREADED ROD (BY LIFT
MANUFACTURER)
EXISTING CONCRETE
FLOOR AND PAN JOIST
5/8" TYPE X GYPSUM BOARD EACH
SIDE APPLIED PARALLEL TO 2 1/2"
METAL STUD FRAME.
5/8" TYPE X GYPSUM BOARD EACH
SIDE METAL STUD PANEL SYSTEM
SIMILAR TO GA FILE NUMBER WP 6130
FIELD VERIFY
1'-8"TYP.
FIELD VERIFY
5"TYP.
4" M
IN
FASTEN ALL ANGLES, BOLTS,
UNISTRUT, AND WEDGE. ALL
ANCHORS AS RECOMMENDED BY
MANUFACTURER
UNISTRUT AS REQUIRED FOR SPANNING TO
STRUCTURE SUPPORTS
LINE OF FINISHED CEILING
(FIELD VERIFY)
MANUFACTURER'S OUTER SUPPORT ARM
AND PLATE SYSTEM WITH DIAGONAL
BRACING.
EXISTING CONCRETE
FLOOR & PAN JOIST
MOUNT MANUFACTURE'S STEEL PLATE BASE
FOR SUPPORT ARM TO UNISTRUT FRAMING AS
REQUIRED BY MANUFACTURER
STEEL ANGLE TO JOISTS AS NEEDED BASED
ON LOCATION OF TV SWING ARM.
SWIVEL COLLAR AND TV
MOUNT INNER ARM PER
MANUFACTURER
22" FLAT SCREEN TV (BY OWNER)
WEST
PLAINS
ENGINEERING. INC.
RAPID CITY, SD SIOUX FALLS, SD CASPER, WY CEDAR RAPIDS, IA
www.westplainsengineering.com
4609 SOUTH TECHLINK CIRCLE SIOUX FALLS, SD 57106
PHONE: (605) 362-3753 FAX: (605) 362-3759
Office of Construction and Facilities Management
Department of Veterans Affairs
Project Number
Building Number
Drawing Number
Dwg. of
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: Project Director
ARCHITECTS/ ENGINEERS:CONSULTANTS:
Revisions: Date
VA FORM 08 -- 6231
on e e ig ht h in ch on e fo ot on e q ua rte r th re ei gh th s ha lf i nc f oo t th q ua rte rs o ne on e i nc f oo t on e a nd o ne ha lf es th es
6"
6"
6"
6"
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
TELE: (605) 332-7850 FAX: (605) 332-3539
SIOUX FALLS, SOUTH DAKOTA 57104-4873
625 S. MINNESOTA AVENUE, SUITE 204
KEYNOTES:
GENERAL NOTES:
KEY PLAN:
JANUARY 29, 2014
636A6−12−909
RTGDGS
REFLECTED CEILING PLAN
AI−103
AS BID DRAWINGS
6 32
BLDG 1 − FLR 4
DES MOINES IA.
SCALE:
PLAN
NORTH
1/4" = 1'-0"
REFLECTED CEILING PLAN
SCALE: 1 1/2" = 1'-0" AI-103
1PATIENT LIFT SUPPORT
SCALE: 1 1/2" = 1'-0" AI-103
2PATIENT LIFT SUPPORT
SCALE: 1 1/2" = 1'-0"
31 HR. CEILING PANELS
(FOR REFERENCE ONLY)
(FOR REFERENCE ONLY)
* LIFT SYSTEM CONTRACTOR SHALL BE RESPONSIBLE FOR
THE INSTALLATION OR THEIR LIFT SYSTEM AND SUPPORTS
SCALE: 1 1/2" = 1'-0" AI-103
4TYPICAL TV SWING ARM DETAIL
WORK AREA - MAIN AREA 1550 SF.
- CORRIDOR 150 SF.