Guldmann_GH3_BayPines_VA_Submission_Package.pdf
PDF 3 MB Posted
- Attached to
- Z2DA--FY26: NRM (PROJ: 516-CSI-142) BB - Site Prep for Emergency Department CT Scanner Replacement
- Solicitation number
- 36C24826Q0004
About this file
This document is a detailed specification and installation guide for the Guldmann GH3 Ceiling Lift System, submitted by the Department of Veterans Affairs Bay Pines Medical Center. The Guldmann GH3 is a patient lifting system designed for healthcare environments, with a safe working load of 770-800 lbs and a vertical lifting distance of approximately 88 inches. The system features an electric hoist with smooth start/stop, rechargeable battery pack, hand controls with emergency stop, and multiple safety features including emergency lowering and overload protection.
The specification outlines three rail configurations: straight rail for single-room transfers, curved rail for navigating obstacles or walls, and room-to-room transfer capability for transitioning between doorways. Installation requirements include a minimum ceiling height of 7 feet, standard 120V AC power supply, and structural support capable of handling the maximum load. The system meets ISO 10535 standards and is designed for indoor use with a temperature range of +41°F to +104°F. The installation guide provides five key highlights: verifying structural integrity, installing the modular aluminum rail system, mounting the hoist unit, testing lift operations, and training staff on proper use and maintenance.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C24826Q0004 0005.pdf | ||
| WD-FL20260222 01-02-26.pdf | ||
| 36C24826Q0004 0004.pdf | ||
| Base Offer Price Breakdown - 1 page.xlsx | XLSX spreadsheet | |
| 36C24826Q0004 0003.docx | DOCX document | |
| 516-CSI-142 - Bay Pines ED CT_ADD 1.pdf | ||
| 03 30 00-CAST-IN-PLACE CONCRETE.pdf | ||
| 01 45 29-TESTING LABORATORY SERVICES.pdf | ||
| 36C24826Q0004 0002.docx | DOCX document | |
| RFI_516-CSI-142 Pre-Bid.docx | DOCX document | |
| PAST AND PRESENT PERFORMANCE QUESTIONNAIRE.docx | DOCX document | |
| ED CT scanner site walk sign-in sheet 10-07-25.pdf | ||
| 36C24826Q0004 0001.docx | DOCX document | |
| 516-CSI-142_Bay Pines_ED CT_ Specs.pdf | ||
| VHA Infection Control Risk Assessment.pdf | ||
| 36C24826Q0004_1.docx | DOCX document | |
| EMR 1 page.pdf | ||
| WD-FL20250222 05-23-2025.pdf | ||
| 516-CSI-142_Bay Pines_ED CT_ Drawings.pdf |
Show all 19
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Department of Veterans Affairs Bay Pines VA Medical Center
Guldmann GH3 – Submission Package
Page 1: Guldmann GH3 Spec Sheet
Manufacturer: Guldmann Model: GH3 Subcategory: Bariatric and Standard Patient Lifting
Technical Specifications:
• Safe Working Load (SWL): 770 lb (350 kg) or 800 lb (363 kg)
• Lift Range: Approx. 88 in (2,235 mm) vertical lifting distance
• Charging System: In-rail charging (GH3+ models)
• Motor Type: Electric hoist with smooth start/stop
• Power Supply: Rechargeable battery pack
• Controls: Hand control with up/down and emergency stop
• Safety Features: Emergency lowering, overload protection, manual lowering
• Optional Features: Integrated scale, Twin lift configuration, Room-to-room transfer
Dimensions: Unit Weight: Approx. 17 lb (8 kg); Rail System: Modular aluminum track
Compliance: Meets ISO 10535; Designed for healthcare environments
Page 2: Install Guide & Rail Configurations
Site Requirements:
• Structural Support: Ceiling or wall structure must support max load (770–800 lb plus system weight).
Reinforcement may be required.
• Power Supply: Standard 120V AC outlet for charging station or in-rail charging.
• Ceiling Rail Layout: Clear path for rail installation; minimum ceiling height 7 ft (2.1 m).
• Environmental Conditions: Indoor use only; Temperature +41°F to +104°F (+5°C to +40°C);
Non-condensing humidity.
Rail Configurations:
Straight Rail Basic single-room transfers
Curved Rail Navigates around obstacles or along walls
Room-to-Room Transfer Doorway/header transition between rooms
XY/H-Frame Full-room coverage with traversing bridge
Integrity | Commitment | Advocacy | Respect | Excellence NCO 8 – Project No. 516-CSI-142
File details come from the government source that posted it. Updated .