EXACTRAC
VARIAN
TRUEBEAM
LINACS
Version 6.x
Site Planning Manual Revision 1.0 Copyright 2016, Brainlab AG Germany. All rights reserved.
TABLE OF CONTENTS
GENERAL INFORMATION
Contact Data and Legal Information Contact Data Legal Information
Symbols Symbols Used in This Guide Hardware Symbols
Intended Use
Documentation
SYSTEM OVERVIEW
Introduction
Technical Key Features
SITE SPECIFICATIONS
Treatment Room Overview Treatment Room Overview
Cabling Overview Cable Conduits Cabling System Cable Specifications
Dual X-Ray Generator Cabinets HFE 601
ExacTrac Computer Cabinet
Heat Dissipation
FLOOR BOXES
12” Floor Boxes (Standard) Overview Recess Dimensions and Side Views for 12” Floor Boxes Installation of 12“ X-Ray Floor Boxes Customized Cable Outlet
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 3
On Floor Boxes (Optional) Overview External Dimensions of the On Floor Boxes Recommended Position of Above-Ground Floor Boxes
CEILING AND WALL MOUNTS
Introduction
Ceiling Mount Installation How to Mount the Flat Panel Detector Systems How to Mount the Ceiling Interface Plate
Ceiling-Mounted Touch Screen Monitor (Optional) Overview Ceiling Frame Installation
Wall-Mounted Touch Screen Monitor (Standard)
Ceiling Mounted In-Room Distribution Box
CONTROL ROOM
Control Room Distribution Box
Control Room Workspace
ELECTRICAL REQUIREMENTS
X-Ray Generator
Computer Cabinet Power Supply
System Mains Power & Protective Earth Cabling
SYSTEM INTERLOCKS AND STATUS SIGNALS
Overview
Emergency STOP Interlock General Information Varian Emergency STOP Interlock
System Power Emergency OFF - ETX Introduction Collision Detect Interlock and Motion Enable
ExacTrac Warning Lights
NETWORK REQUIREMENTS
Introduction
4 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
EXACTRAC ROBOTICS
Robotics Installation Requirements
External Bluetooth Module
COMPLIANCES
Statement of Compliance from Hospital/Contractor
Authorities
APPENDIX
Pre-Installation Kit
Standard Shipping List
Cable Classifications
Facility Pre-Installation Summary Overview of Mains Power Supply and Interlocks Overview of Wire Diameters in Grounding and Functional Cables Generator Power Supply (Distribution Box)
Types of Networks for Electricity Supply
IT Policy, Screen Saver
17” Floor Boxes for X-Ray Tubes
PREINSTALLATION CHECKLISTS
Floor Boxes
Ceiling Brackets
Electrical Infrastructure
Conduits
Logistics
INDEX
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 5
6 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
1 GENERAL INFORMATION
1.1 Contact Data and Legal Information
1.1.1 Contact Data
Support
If you cannot find information you need in this guide, or if you have questions or problems, contact Brainlab support:
Region Telephone and Fax Email
United States, Canada, Central and South America
Tel: +1 800 597 5911 Fax: +1 708 409 1619 us.support@brainlab.com
Brazil Tel: (0800) 892 1217 brazil.support@brain-lab.com
UK Tel: +44 1223 755 333 support@brainlab.com
Spain Tel: +34 900 649 115
France and French-speaking regions Tel: +33 800 676 030
Africa, Asia, Australia, Europe Tel: +49 89 991568 44 Fax: +49 89 991568 5811
Japan Tel: +81 3 3769 6900 Fax: +81 3 3769 6901
Project Site Coordination
The Project Site Coordinator team for pre-installation can be contacted at:
• rt.coordination@brainlab.com or
• us.rt.coordination@brainlab.com (for US customers only)
Feedback
Despite careful review, this manual may contain errors.
Please contact us at oncology.manuals@brainlab.com if you have suggestions as to how we can improve this manual.
Manufacturer
Brainlab AG Olof-Palme-Str. 9 81829 Munich Germany
GENERAL INFORMATION
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 7 mailto://us.support@brainlab.com mailto://brazil.support@brainlab.com mailto://brazil.support@brainlab.com mailto://support@brainlab.com mailto://rt.coordination@brainlab.com mailto:/us.rt.coordination@brainlab.com mailto://igs.manuals@brainlab.com
1.1.2 Legal Information
Copyright
This guide contains proprietary information protected by copyright. No part of this guide may be reproduced or translated without express written permission of Brainlab.
Brainlab Trademarks
• ExacTrac® is a registered trademark of Brainlab AG in Germany and/or the US.
• Brainlab® is a registered trademark of Brainlab AG in Germany and/or the US.
Non-Brainlab Trademarks
• Microsoft® and Windows® are registered trademarks of Microsoft Corporation in the United States and other countries.
• Exact®, Truebeam®, TrueBeamSTx® and PerfectPitch® are registered trademarks of Varian Medical Systems, Inc.
CE Label
• The CE label indicates that the Brainlab product complies with the essential re-quirements of European Council Directive 93/42/EEC, the Medical Device Di-rective (MDD).
• According to the rules established by the MDD, ExacTrac is a Class IIb prod-uct.
NOTE: The validity of the CE label can only be confirmed for products manufactured by Brainlab.
Disposal Instructions
The collimators contain lead and should only be disposed of in accordance with statutory regulations.
Only dispose of electrical and electronic equipment in accordance with statutory regu-lations. For information regarding the WEEE (Waste Electrical and Electronic Equip-ment) directive, visit:
www.brainlab.com/en/sustainability
Sales in the US
US federal law restricts this device to sale by or on the order of a physician.
Electronic Product Radiation Control (EPRC) Program (US Only)
According to the product classification for medical devices of the FDA Center for Devices and Radiological Health, ExacTrac is defined as a medical charged-particle radiation therapy system (product code IYE).
Based on 21 CFR (Code of Federal Regulations Title 21) of the CDRH, no performance standard is applicable and General Radiological Health requirements apply (21 CFR 1000 - 1005).
Contact Data and Legal Information
8 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x http://www.brainlab.com/en/sustainability/
1.2 Symbols
1.2.1 Symbols Used in This Guide
Warnings
Warnings are indicated by triangular warning symbols. They contain safety-critical information regarding possible injury, death or other serious consequences associated with equipment misuse.
Cautions
Cautions are indicated by circular caution symbols. They contain safety-critical information regarding possible problems with the device. Such problems include device malfunctions, device failure, damage to device or damage to property.
Notes
NOTE: Notes are formatted in italic type and indicate additional useful hints.
GENERAL INFORMATION
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 9
1.2.2 Hardware Symbols
Symbols on Hardware Components
Symbol Explanation
Type B Applied Part according to IEC 60601-1
Caution
Serial number
Article number
Date of manufacture
Manufacturer
Follow the instructions for use
Prescription only
Symbols
10 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
1.3 Intended Use
Indications for Use
ExacTrac is intended for placing patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures. Treatment includes lesions, tumors and conditions anywhere in the body where radiation therapy is indicated. ExacTrac is also used for monitoring the patient position during treatment.
Intended User
The intended users of ExacTrac are medical staff having received dedicated training for this application, such as:
Title Education Actions
Radiation Therapist Education, training Main workflow
Radiation Oncologist/Physician Physician with MD Fusion approval, patient initialization
Medical Physicist/Physicist PhD in medical phys-ics
System Quality Assurance, data handling
(This classification depends on country-specific requirements/regulations and might vary from the classifications above.)
Only trained medical staff may operate system components and accessory instrumentation.
Known Contra-Indications
To prevent injury to patients wearing pacemakers, contact the pacemaker manufacturer regarding possible hazards resulting from radiation exposure before commencing patient treatment.
Place of Use
The application is to be used in the treatment room.
Careful Hardware Handling
System components and accessory instrumentation comprise precise mechanical parts.
Handle them carefully.
Only trained medical personnel may operate system components and accessory instrumentation.
Plausibility Review
Before patient treatment, review the plausibility of all information input to and output from the system.
No modification of this equipment is allowed.
GENERAL INFORMATION
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 11
1.4 Documentation
Intended Audience
This Site Planning Manual is intended as a reference for the effective and correct implementation of ExacTrac in combination with a Varian TrueBeam Linear Accelerator. This guide is written for customers (in particular, customers who have purchased or intend to purchase an ExacTrac system), as well as all members of the clinical architectural, structural and consulting teams who are involved in the planning and implementation of an ExacTrac system or parts of it.
This guide is to be provided to all relevant parties at the beginning of the planning process and explained in appropriate detail. Read this user guide carefully and acquaint yourself sufficiently with the system before planning its implemenation.
This Site Planning Manual contains important information about the responsibilities and duties of all relevant parties as well as a detailed installation description. Pay special attention to the warnings and cautions provided in this manual.
Reading User Guides
The user guides describe complex medical devices and software that must be used with care.
It is important that all users of system, instruments and software:
• Read the user guides carefully before handling the equipment
• Have access to the user guides at all times
Pre-Installation Information
It is highly recommended to read this document prior to the main system installation. The customer is responsible for finishing all pre-working/pre-assembly steps to ensure a simple and timely system installation. It is also very important that you inform Brainlab of all information regarding the hospital’s electrical and spatial conditions.
At the end of the document is a checklist you can use to ensure that all listed requirements are completed and finalized.
If a subcontractor shall perform the required pre-working/pre-assembly steps, the customer is responsible for providing the subcontractor with this complete pre-installation information including all warnings.
Please note that not all components and items described in this document belong to each installed system. Please contact Brainlab support for details or if any constraints cannot be met.
Documentation
12 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
2 SYSTEM OVERVIEW
2.1 Introduction
Customer Responsibility
All structural preparations, such as reinforcement and/or strengthening for roof mounted items, ceilings, superstructures, or any related objects are the responsibility of the customer.
SYSTEM OVERVIEW
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 13
2.2 Technical Key Features
Infrared Components
The infrared components consist of the following:
• Ceiling-mounted infrared and video camera system.
• Wall-mounted touch screen monitor (standard).
• Control room workspace with monitor, keyboard and mouse.
• Compact-sized computer rack for computer, power supplies, electrical safety devices, control electronics and excess cables (ExacTrac Computer Cabinet).
X-Ray Components
The X-ray components consist of the following:
• Two X-ray tubes recessed into the linac room floor.
• Two ceiling-mounted amorphous silicon detectors (ExacTrac Flat Panels).
• ExacTrac Console.
• Dual 65kW X-ray generator with integrated transformer unit.
Optional Components
Some optional components are:
• Ceiling-mounted Touch Screen Monitor for convenient handling (optional, standard is wall-mount solution)
• Two X-ray Tubes mounted on-floor in the linac room (optional, standard is in-floor solution).
• Tilt alignments of targeting volume via ExacTrac Robotics.
Technical Key Features
14 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
3 SITE SPECIFICATIONS
3.1 Treatment Room
3.1.1 Overview
General Information
The ExacTrac system shall only be installed in a bunker for medical radiotherapy accelerators that has a stable foundation. Keep vibrations to a minimum since they influence the accuracy of the tracking system.
Exposure to high neutron radiation can lead to recurring malfunction and permanent damage to computer hardware and electronic components. For sites where patients will be treated with energies 10 MV or higher, do not place the computer cabinet in the treatment room as it is not designed for exposure to high neutron radiation.
The system must not be exposed to humid environment. Humidity can destroy the system and can lead to short circuits.
The ExacTrac system is not designed for use in areas where there is a danger of explosion.
Moreover, the system must not be placed close to flammable materials, such as volatile anesthetics, detergents or endogenous gases. Observe local regulations to avoid explosive gas mixtures forming during usage and cleaning.
Ensure that the tracking volume of the ExacTrac infrared camera system is not reduced/modified by other devices after the ExacTrac system has been accepted.
Do not install the ceiling mount and device mount in the direct airflow of an air conditioning or heating system.
Do not install any of the electronic components described in this site planning manual (e.g., monitors, cabinets, warning lights, distribution boxes) in the area of the linac’s primary beam.
SITE SPECIFICATIONS
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 15
3.1.2 Treatment Room Overview
Sample Set-Up
All major ExacTrac components and positions within the bunker have a dedicated numbered label for further reference.
The following image shows an overview of the main system components in a treatment room (diagram not meant to scale):
Floor Box with X-Ray Tube (left side)
2.1a
IR and Video Camera System
1.1
Flat Panel Detector (right side)
X-Ray Warning Light
Control-Room Workplace
X-Ray Warning Light
Control Room Distribution Box
Wall Mounted Touch Screen Monitor (standard)
Ceiling Mounted Touch Screen Monitor (optional)
Network Plug (provided by Hospital)
1.2
2.3a
2.5b
Treatment Couch
0.1
Primary Beam
Lasers
1.2
1.5
2.5a
Isocenter
2.4
ExacTrac Console
1.0
Computer Cabinet
Stand & Gantry
0.0
In-Room Distribution Box
1.3
X-Ray Dual Generator
2.0a 2.0b
4.0
2.2a
Floor Box with X-Ray Tube (right side)
2.1b 2.2b
Flat Panel Detector (left side)
2.3b
Control-Room Workspace
1.6
Emergency Off
Emergency Off
Emergency Off max. radius 2900 EXACT couchtop with foot extension
3.2
External Bluetooth Module
Power Plug (for Robotics Charger)
L
LL
Figure 1
Treatment Room
16 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Couch Clearance Specifications for Treatment Room (Measurements in mm)
Isocenter radius 1855 couch end
EXACT couchtop with foot extension radius 2670 EXACT couchtop w/o foot extension radius 2460 necessary distance for treating cranial cases (Couchmount+Headring+ Localizer or Frameless)
15° radius 2900
Figure 2
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 17
3.2 Cabling
3.2.1 Overview
General Information
Position all components so that you can use the shortest cable whenever possible.
Determine the correct length of X-ray tube power cable necessary to connect the X-ray generator to the X-ray tubes. Calculate an extra 3 m for the cabling within the X-ray generator to avoid too much tension between the cables and the circuit points.
In most hospitals, conduits are required for ExacTrac cables due to hospital and local electrical regulations and laws. Below is a table listing the requirements for these conduits.
NOTE: Brainlab does not require you to store ExacTrac cables within conduits. However, some of the supplied cables are not plenum rated. Therefore, it may be necessary for some customers to install all cables within conduits.
You are responsible for obtaining installation permits with respect to applicable local or regional standards. This might affect cable routes, mains power and data cables specifications, connection to potential equalization and where you install system components.
Be aware that fire protection engineering aspects (plenum rating) may be affected when installing cables and system components. Install cable conduits and cables so that they are not a tripping hazard and the cables will not be damaged.
Cabling
18 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
3.2.2 Cable Conduits
Sample Treatment Room Set-Up
The following image shows an overview of cable conduits (diagram not meant to scale):
3.2
In-Room Distribution Box
Control-Room Workplace
External Bluetooth Module
Stand & Gantry
0.0
Primary Beam
1.3
Control Room Distribution Box
1.5
Isocenter
2.4
ExacTrac Console
Control-Room Workspace
1.6
Tube 1 Tube 2
2.2a 2.2b
1.0
Computer Cabinet
Hospital Network
4.0
X-Ray Dual Generator
2.0a 2.0b
2.3b
2.3a
1.2
1.2 1.1
2.5a
2.5b
Warning Light
Warning Light
Flatpanel
Flatpanel
IR Camera
L
LL
A Ø75mm
D1 Ø75mm
G2 Ø50mm
J
T Ø25mm
M Ø130mm
O Ø25mm
R Ø50mm
I Ø25mm (optional)
C Ø75mm
H2 Ø25mm
P Ø25mm
B1 Ø75mmG1 Ø50mm (optional)
S Ø75mm
= Ø25mm
= Ø50mm
= >Ø75mm_
H1 Ø50mm
D2 Ø75mm
B2 Ø75mm
E Ø50mm
K Ø25mm
M Ø130mm
U Ø25mm
Grounding Spot
Figure 3
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 19
Treatment Room Sample Side View
The following image shows a side view of cable conduits (diagram not meant to scale):
3.
2.
0b
2.
2a
2.
2b
N et w or k Pl ug (p ro vi de d by H os pi ta l)
1.
2.
3b
0.
1.
1.
2.
5a
2.
1.
0.
1.
C on tr ol
-R oo m W or kp la ce
2.
0a
2.
5b
2.
1b
2.
1a
2.
3a
4.
1.
1.
Tr ea tm en t R oo m
V M
S C am er a
O R
B
H
C T
G
J G
M A
E D
S
Ø m m
Ø m m
>Ø m m
H op tio na l
D
B
U
Figure 4
20 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Description of Cable Conduits
# From To Cable Length
Cable Di-ameter
Largest Plug Diame-ter
Minimum Conduit Di-ameter
A Computer cabi-net (1.0)
Generator cabinet 1 (2.0a) 30 m 6 mm 35 mm 50 mm (low voltage cable)
B1 Computer cabi-net (1.0)
In-room distribution box (1.3) 30 m 9, 10 mm No plug 75 mm (low voltage cable)
B21) Computer cabi-net (1.0)
In-room distribution box (1.3) 30 m 5, 6 mm 35 mm 75, 100 mm (low volt-age cable)
C1) Computer cabi-net (1.0)
Flat panel detectors (2.3a/2.3b) 30 m 2x 8 mm, 2x 9 mm 52 mm 75, 100 mm (low volt-age cables)
D1 Computer cabi-net (1.0)
Control room distri-bution box (1.5) 30 m 9, 10 mm No plug 75 mm (low voltage cables)
D2 Computer cabi-net (1.0)
Control room distri-bution box (1.5) 30 m 5, 11 mm 53 mm 75 mm (low voltage cable)
E2) Computer cabi-net (1.0)
Linac gantry stand (0.0) for Varian 30 m 6 mm 35 mm 50 mm (low voltage cable)
G1 In-room distri-bution box (1.3)
Optional: Ceiling-mounted touch screen (1.2)
8 m 4, 5.9, 2x 7 mm 35 mm 50 mm (low voltage cable)
G2 In-room distri-bution Box (1.3)
Standard: Wall-mounted touch screen (1.2)
8 m 5.9, 2x 7 mm
35 mm 50 mm (low voltage cable)
H1 Computer cabi-net (1.0) Camera system (1.1) 30 m 7, 8 mm 35 mm 50 mm (low voltage cables)
H2 In-room distri-bution box (1.3)
Camera system (1.1) n/a 5 mm No plug 25 mm (see page 77)
I In-room distri-bution box (1.3)
Alternatively: PE grounding spot n/a 9 mm No plug 25 mm (see page
77)
J In-room distri-bution box (1.3)
External bluetooth module (3.2) 5 m 6 mm 10 mm n/a (low voltage ca-bles)
K Computer cabi-net (1.0) PE grounding spot n/a 9 mm No plug 25 mm (see page
77)
M3) Generator cab-inet 1 (2.0a)
Left side: Floor box 1 (2.1a)
11, 15, 18 mm
9, 12, 17 mm
Diameter: 70 mm Plug length: 180 mm
130 mm [trench:
100mm] (high voltage cables, no sharp turns)
M3) Generator cab-inet 2 (2.0b)
Right side: Floor box 2 (2.1b)
11, 15, 18 m
9, 12, 17 mm
Diameter: 70 mm Plug length: 180 mm
130 mm [trench:
100mm] (high voltage cables, no sharp turns)
O Computer cabi-net (1.0)
X-ray warning light 1 & 2 (2.5a/2.5b)
30 m + 30 m 4 mm No plug 25 mm (low voltage cables)
P Generator cab-inet 1 (2.0a) PE grounding spot n/a 9 mm No plug 25 mm (see page
77)
R Computer cabi-net (1.0)
External bluetooth module (3.2) 30 m 6 mm 35 mm 50 mm (low voltage cable)
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 21
# From To Cable Length
Cable Di-ameter
Largest Plug Diame-ter
Minimum Conduit Di-ameter
S Generator cab-inet 1 (2.0a)
Generator cabinet 2 (2.0b)
5 m (op-tional 20 m)
2x 6, 24 mm 35 mm
75 mm (low and high voltage cables togeth-er in the same conduit, no sharp turns)
T In-room distri-bution box (1.3)
Flat panel detectors (2.3a/2.3b) n/a 9 mm No plug 25 mm (see page
77)
U Computer cabi-net (1.0)
Hospital network (4.0) 4 m 5 mm 14 mm 25 mm
1) The cabling from the computer cabinet to the in-room power box and the flat panels can be placed together in a single 100 mm conduit.
2) Depending on system configuration, one or two cables are required.
3) A 130 mm conduit is required from generator 1 to floor box 1 (labeled “a”) and from generator 2 to floor box 2 (labeled “b”). If generators 1 and 2 are positioned directly next to each other, it is possible to place both X-ray tube cables into a single conduit. In this case, separate the high voltage cables from the data cables.
22 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
3.2.3 Cabling System
Functional Cabling for Varian TrueBeam Systems
The following image shows an overview of the functional cabling (excluding mains, PE and building installation) for Varian TrueBeam systems:
S in gl e
K V
M M od ul e
D ua l K
V M
M od ul e
In -r oo m da ta
/p ow er di st rib ut io n bo x K
V M re ce iv er
Ju nc tio n
In -r oo m m on ito r
R O
B O
TI
C
S B lu et oo th
C on tro l-r oo m da ta
/p ow er di st rib ut io n bo x
C am er a
S ys te m
C on tro l-r oo m m on ito r
IR
ca m er a Vi de o ca m er a
Fl at
P an el a
Fl at
P an el b
E xa cT ra c C om pu te r C ab in et
X-Ray Generator
W ar ni ng
Li gh t a
K V
M re ce iv er
X -R ay Tu be b
X -R ay Tu be a
2.
2a
2.
2b
2.
0b
1.
2.
5a
1.
3.
1.
2.
3a
1.
2.
3b
1.
1.
0.
8a
E xa cT ra c C on so le 2.
W ar ni ng
Li gh t b
2.
5b
U S
B A
Fl at
P an el fra m e gr ab be r
E xa cT ra c W or kS ta tio n
LA
N
G ra ph ic c ar d
U S
B B
U S
B C
U S
B D
C O
M 1 C
O M 2
C O
M 3 C
O M 4
C O
M P
W R
C O
M
IN
K V
M
Vi de o Fr am e
G ra bb er
H os pi ta l N et w or k
K ey bo ar d, V id eo
M ou se c ab le s
5, 5, X-Ray Generator
2.
0a
Junction
P ow er su pp ly P ow er su pp ly
Cable Extension 5m
Fl at
P an el fra m e gr ab be r
Fl at
P an el P ow er su pp ly
Fl at
P an el P ow er su pp ly
U S
B P ow er su pp ly
A
,1 2B
A
C
,1 2D
B
E T
D X
R E
T C
O U
E
T R
TM
E T
C O
M E
T
IL
K
In se rt
E T
P O
W In se rt
IT
U
In se rt In se rt
Figure 5
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 23
3.2.4 Cable Specifications
Cable Slack (Excess Length)
Cable slack is needed for each cable and varies from cable to cable. Contact the Brainlab site coordinator for more details, unless otherwise stated in this document.
ExacTrac Cables installed by Brainlab
The following table provides an overview of all ExacTrac cables that need to be pulled by Brainlab, their start and end point as well as their size and correct orientation. All other building installation cables for system interlocks are not listed here (for more information, see page 81).
R ef er en ce N r.
C ab le N am e
C on du it
R ef er en ce
ST
A
R T
ST
A
R T
(C om po ne nt , P lu g, C ab le
EN
D
EN
D
(C om po ne nt , P lu g, C ab le
N ot e C ab le
D ire ct io n
Pl ug
S
TA
R
T
Pl ug
E
N D
M ax
. C ab le Le ng th
[m
C ab le D ia m et er
[m
La rg es t P lu g
D im en si on s
[m m
Sh ip pe d w ith
Pr e-
In sa ll- K it
Sh ip pe d w ith
M ai n Sh ip m en t
M od ul e
C ab le T yp e
12A XR High Volt-age Cable M 2.0a HVT/(+) 2.2a Tube/(+)
Anode 11/15/1 8 17 180x70 x XR Power Tube
150kV
12B XR High Volt-age Cable M 2.0a HVT/(-) 2.2a Tube/(-)
Cathode 11/15/1 8 17 180x70 x XR Power Tube
150 kV
12C XR High Volt-age Cable M 2.0b HVT/(+) 2.2b Tube/(+)
Anode 11/15/1 8 17 180x70 x XR Power Tube
150 kV
12D XR High Volt-age Cable M 2.0b HVT/(-) 2.2b Tube/(-)
Cathode 11/15/1 8 17 180x70 x XR Power Tube
150 kV
13A XR Stator Ca-ble M 2.0a Sub-D./X6 2.2a Tube/
Stator x Attached to tube
11/15/1 8 12 x XR Power Stator
230 V
13B XR Stator Ca-ble M 2.0b Sub-D./X6 2.2b Tube/
Stator x Attached to tube
11/15/1 8 12 x XR Power Stator
230V
25 ET EXGI
CDOS Cable 1.0 LC x 4-pin 9-pin
DSUB 6 x Gating Data, Gating
27 Cat 5E Patch Cable U 1.0 WS/LAN
Conn. 4.0 Network Plug RJ45 RJ45 4 5 14x11 x Patch
38 PEQ, 16 mm2 M 2.0 2.2a 2) 9 N/A x XR PEQ
39 PEQ, 16 mm2 M 2.0 2.2b 2) 9 N/A x XR PEQ
ExacTrac Cables installed by the Contractor
The following table provides an overview of all ExacTrac cables that need to be pulled by the contractor, their start and end point as well as their size and correct orientation. All other building installation cables for system interlocks are not listed here (for more information, see page 81).
R ef er en ce N r.
C ab le N am e
C on du it
R ef er en ce
ST
A
R T
ST
A
R T
(C om po ne nt , P lu g, C ab le
EN
D
EN
D
(C om po ne nt , P lu g, C ab le
N ot e C ab le
D ire ct io n
Pl ug
S
TA
R
T
Pl ug
E
N D
M ax
. C ab le Le ng th
[m
C ab le D ia m et er
[m
La rg es t P lu g
D im en si on s
[m m
Sh ip pe d w ith
Pr e-
In sa ll- K it
Sh ip pe d w ith
M ai n Sh ip m en t
M od ul e
C ab le T yp e
1 ET Video Ca-ble H 1.0 Video Fr.
Grabber 1.1 Video Adapter C.
x 7-pin thin
7-pin thick 30 8 35 x IR Video
2 ET KVM Cable B1 1.0 Dual KVM 1.3 KVM (in-room) 30 5 14x11 x IR Patch
3 ET Serial Ex-tension B1 1.0 COM 3 1.3 Cable 16 x FE-
MALE MALE 30 9 31x15 x IR Touch Exten-sion
4 ET Infrared USB Adapter H 1.0 USB B 1.1 Spectra
Camera x Adapt er MALE 30 7 x IR Data/Power
5A ET Flatpanel Data Cable C 1.0 FG with lower ID 2.3a
FP1
(right)/ Data
30 8 50x34 x XR FP Data
5B ET Flatpanel Data Cable C 1.0 FG with higher ID 2.3b FP2 (left)/ Data 30 8 50x34 x XR FP Data
24 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
R ef er en ce N r.
C ab le N am e
C on du it
R ef er en ce
ST
A
R T
ST
A
R T
(C om po ne nt , P lu g, C ab le
EN
D
EN
D
(C om po ne nt , P lu g, C ab le
N ot e C ab le
D ire ct io n
Pl ug
S
TA
R
T
Pl ug
E
N D
M ax
. C ab le Le ng th
[m
C ab le D ia m et er
[m
La rg es t P lu g
D im en si on s
[m m
Sh ip pe d w ith
Pr e-
In sa ll- K it
Sh ip pe d w ith
M ai n Sh ip m en t
M od ul e
C ab le T yp e
6A ET Flatpanel DC Cable C 1.0 Power sup-ply 2.3a
FP1
(right)/ Power x Thick, “ring”
Thin end 30 9 x XR FP Power
6B ET Flatpanel DC Cable C 1.0 Power sup-ply 2.3b FP2 (left)/ Power x Thick
“ring” Thin end 30 9 x XR FP Power
7 ET KVM Cable D1 1.0 Single
KVM 1.5
KVM
(Control R.)
30 5 14x11 x IR Patch
8A ET Serial Ex-tension E 1.0 ILK 0.0 VMS In-terface x MALE FE- MALE 30 9 31x15 x ILK Serial
X-ray Control Panel Cable 30 m (UL)
A 1.0 DXR/XR
Generator 2.0a Dual In-terf./X1 x FE- MALE MALE 30 11 53 x XR Data, xr-sync
ET Dual Gen-erator Connec-tion Cable
S 2.0a Pre-Distri-bution 2.0b Sub-Dis-tribution Attached to Gen 2 20 25 x XR Power
11A ET Dual Gen-erator Connec-tion Cable
S 2.0a Dual Interf./X23 2.0b NAG2/
X11 x FE- MALE MALE 20 6 35 x XR Data
11B ET Dual Gen-erator Connec-tion Cable
S 2.0a Dual Interf./X25 2.0b BUS2/
X28 RJ45 RJ45 20 5 14x11 x XR Patch
15 ET Touch VGA Cable G 1.3 KVM (in-room) 1.2
VGA
Connec-tor x FE- MALE MALE 8 7 x IR Touch VGA
16 ET Touch Data Cable G 1.3 Cable 3 1.2
RS232
Connec-tor x FE- MALE MALE 8 7 x IR Touch Signal
ET Touch Power Exten-sion
G 1.3 Touch Pow.
Sup. 1.2
Power Connec-tor
8 6 x IR Touch Power 12V
19A ET Warning Light Cable O 1.0 DXR/XR
W. Light 2.5a Luster Terminal x MALE Cable 30 4 x XR Data
19B ET Warning Light Cable O 1.0/
2.5a
* DXR/XR
Warning Light
* Luster Terminal
2.5b Luster Terminal 30 4 x XR Data
20 ET Console Cable D1 1.0 ILK/
Console 1.5 Control Room Box x MALE FE- MALE 30 11 53 x AutP/
XR Data
23 RTM2 Cross Over Cable R 1.0 RTM2/
Robotics 3.2
RS232
Connec-tor x FE- MALE MALE 30 7.3 35x15 x RTM Data
Power, max.
3x 16 mm2 / 1.0 DP c c c c c c c c IR Power
Power, max.
5x 25 mm2 / 2.0 DP c c c c c c c c XR Power
Power, 3x 2.5 mm2 B2 1.0 1.3 1) 10 N/A x IR Power
Power, 3x 2.5 mm2 D2 1.0 1.5 1) 10 N/A x IR Power
34 PEQ, 16 mm2 K 1.0 PE 2) 9 N/A x IR PEQ
35 PEQ, 16 mm2 B2 1.0 1.3 2) 9 N/A x IR PEQ
36 PEQ, 16 mm2 D2 1.0 1.5 2) 9 N/A x IR PEQ
37 PEQ, 16 mm2 P 2.0 PE 2) 9 N/A x XR PEQ
40 PEQ, 2.5 mm2 H 1.3 1.1 3) 4 N/A x IR PEQ
41 PEQ, 2.5 mm2 G 1.3 1.2 3) 4 N/A x IR PEQ
42 PEQ, 2.5 mm2 T 1.3 2.3a 3) 4 N/A x XR PEQ
43 PEQ, 2.5 mm2 T 1.3 2.3b 3) 4 N/A x XR PEQ
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 25
Further Table Information
Abbreviations:
• W.: Warning
• c: Mains cables not provided by Brainlab
• 1) A total of 60 m of cable is included
• 2) A total of 70 m of 16 mm2 cable is included
• 3) A total of 50 m of 2.5 mm2 cable is included
• DP: Electric Distribution Panel
• PEG: Protective Earth Grounding spot
Additional Cable Information:
• Cable 3: Label cable 3 with “Monitor” on the computer cabinet side.
• Cable 4: The IR USB adapter cable is equipped with a device to combine power and data, which needs to be stored carefully and kept protected from dust and shocks.
• Cable 8a: Label cable 8a with “Linac” on the computer cabinet side.
• Cables 15, 16 and 17: If you ordered a ceiling-mounted touch screen monitor, cables 15, 16 and 17 come pre-installed in the monitor arm.
• Cable 22: Minimum bend radius is 20 mm.
• Cable 23: Minimum bend radius is 55 mm.
NOTE: In some cases, cable direction is important. Refer to the column, “Note Cable Direction”.
26 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
3.3 Dual X-Ray Generator Cabinets HFE 601
X-Ray Generator Cabinet Distances
The exact cable lengths need to be specified by the hospital.
• Possible cable length between the generator and X-ray tubes: 11, 15 and 18 m.
NOTE: Use the same cable length for both tubes.
• Possible cable length between the two generator cabinets: 5 m NOTE: Brainlab can provide an extension to 20 m on request only if there is absolutely no way to install the generators 5 m apart.
X-Ray Generator Cabinet Connections
Generator 1 (2.0a) G enerator 2 (2.0b)
Figure 6
• External interfaces (mains power and control data input) are only allowed at generator cabinet
1.
• X-ray tube 1 (left side of the gantry when facing the gantry) must be connected to generator cabinet 1 (labeled “a” in the diagrams).
• X-ray tube 2 (right side of the gantry when facing the gantry) must be connected to generator cabinet 2 (labeled “b” in the diagrams).
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 27
X-Ray Generator Cabinet Specifications
Component Description
Dimensions
550 mm x 550 mm x 1510 mm (each cabinet)
1100mm
480mm 480mm
450mm m m
0m m
80mm
Servicing area and cable access
For service purposes the cabinet is pulled to the front. Keep free a servicing area of 800 mm.
cable access from top or bottom side of cabinet (left / right optional)
450mm approx. 800mm
0m m
Weight (including wheels)
• Dual generator without pre-transformer: 320 kg
• Dual generator with pre-transformer 208 V - 400 V: 390 kg
• Dual generator with pre-transformer 420/440/480 V - 400 V: 370 kg
• Dual generator with pre-transformer Delta 200/210/220/208 V -star 400 V: 460 kg
• Single cabinet without pre-transformer: 160 kg
Dual X-Ray Generator Cabinets HFE 601
28 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Wall fixation of rear panel
Fix the rear panel of each X-ray generator cabinet to the wall ac-cording to the drilling pattern. Each rear panel weighs a maximum of 25 kg, depending on the configuration.
Use anchors according to wall conditions and local regulations (not provided by Brainlab).
214mm
1000mm
1454mm
180mm 180mm
Location
When selecting the distance between the two cabinets, consider the length of the cable connecting the two cabinets:
• Standard: 5 m
• Optional: 20 m
Do not place the cabinets in the primary beam of the linac.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 29
3.4 ExacTrac Computer Cabinet
General Information
The ExacTrac computer cabinet is a two-way accessible cabinet for storing the workstation, power supplies, electrical safety devices, interface electronics and excess cables.
The following image shows front and back views of the ExacTrac computer cabinet:
Figure 7
ExacTrac Computer Cabinet
30 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
ExacTrac Computer Cabinet Specifications cable access from bottom side of cabinet
610mm 810mm m m
880mm service access service access
Figure 8
Component Description
Dimensions Cabinet only: 810 mm x 610 mm x 1780 mm Cabinet plus door handles: 880 mm x 650 mm x 1780 mm
Weight 250 kg
Location The cabinet can be placed with any long side facing the wall. The door can be opened from either side, depending on how it is mounted.
Wall fixation
In earthquake endangered regions, the com-puter cabinet must be fixed to the wall. An an-gle bracket is supplied with the computer cabi-net.
Network connection Cat 5E (or higher) patch cable, 4 m; cable no.
27 (see ).
Avoid placing the cabinet in the treatment room. If this is unavoidable, ensure a maximum distance between the cabinet and the linac and ensure the cabinet is outside the primary beam of the linac.
Exposure to high neutron radiation can lead to recurring malfunction and permanent damage to computer hardware and electronic components. For sites where patients will be treated with energies 10 MV and above, do not place the computer cabinet in the treatment room as it is not designed for exposure to high neutron radiation.
NOTE: Ensure sufficient spatial clearance before rigging the computer cabinet.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 31
NOTE: If you have limited space, you may position the cabinet with the rear service door side next to a wall (100 mm clearance) after all cabling is finished.
ExacTrac Computer Cabinet
32 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
3.5 Heat Dissipation
Calculation
The calculations used to determine the heat dissipation are based on the following:
• 1000 BTU/h = 293 W
• 1 W = 3.412 BTU/h
X-Ray Dual Generator and Tubes
In stand-by mode, the heat dissipation for the X-ray dual generator and tubes is approximately 250 W (853 BTU/h).
For air condition layout, use 400 W (1365 BTU/h): 100 W for each generator (2 in total) plus 100 W for each tube (2 in total) = 400 W.
Computer Cabinet and All System Components
Measured: 406 W (1386 BTU/h) For air condition layout, use: 500 W (1706 BTU/h)
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 33
Heat Dissipation
34 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
4 FLOOR BOXES
4.1 12” Floor Boxes (Standard)
4.1.1 Overview
General Information
X-ray tubes are encased in specially designed floor boxes, which are delivered with leveling feet for fine adjustment.
For more information on 17” floor boxes, see page 116.
① ② No. Description
① 12” floor box with aluminum cover on top of the wooden cover.
② 12” floor box showing the wooden cover that supports the aluminum cover from under-neath.
FLOOR BOXES
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 35
Cable Outlet
Only use the cable outlet ① provided:
①
Figure 9
Top and Back Views
Ø130mm
8m m 6m m
668mm
407mm
380mm
Ø16mm
Ø16mm
Ø16mm
6m leveling screws (3x) lifting ancors (3x) fixation holes (3x)
"A"
4m distance rim
Detail A
Top View Back View
Figure 10
12” Floor Boxes (Standard)
36 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
4.1.2 Recess Dimensions and Side Views for 12” Floor Boxes
Recess Positions and Dimensions
45°
834+/-7mm mm
4+
/-7 m m
Cable conduit* Cable conduit*
Isocenter
1357+/-7mm m in
. 7 min. 780
Figure 11
• *Cable conduit should ideally be a non-flexible PVC pipe with a minimum inner diameter of 100 mm and a maximum outer diameter of 130 mm.
• See page 39 for worst case positioning.
• After placing the box, pour concrete into the hole so that the aluminum cover rests on a minimum of 10 mm stable concrete on all sides.
• The absolute minimum depth of the pit over the whole surface is 310 mm.
• The recess surface must be even for fine adjustment of the floor boxes.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 37
Recess Side View finished floor
0+2
-2 m m
3m m
Figure 12
Safety Notes
It is mandatory to get approval and/or clearance from a stress analyst before cutting the floor box recesses, as building statics will change.
In case the linac is already installed, ensure that the linac base frame is not damaged during recess construction.
Until the floor box is installed, the empty pit must be marked with warning signs or covered.
The aluminum cover must rest on stable (concrete) floor minimum with 10 mm.
All dimensions for positioning floor boxes are based on an isocenter height of 1295 mm +/- 20 mm (finished floor line – isocenter).
The exact value depends on the linac type. Please refer to the respective documentation. If the vertical isocenter position is not within the tolerance range, please contact Brainlab.
38 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
4.1.3 Installation of 12“ X-Ray Floor Boxes
Installation Top View
25mmmaximum
45°1240 +/-10mm
7+
/-7 m m
877+/-7mm
Ideal position for 12" Floor Box
Linac Baseframe
Worst Case position for 12" Floor Box
Figure 13 NOTE: Worst case position: Due to an incorrect recess set up, the floor box must be rotated within the recess to the correct position. Rotation within the recess is very limited. The maximal deviation of the line “floor box tip to midpoint rear wall of box” at the isocenter is 25 mm.
The determining measurement is the exact distance between isocenter and floor box tip at 45°.
Installation Steps
Steps
1 Ensure that all the openings are sealed (taped) before pouring concrete. Concrete should not seep into the boxes.
2 Place the floor boxes into the recesses.
finished floor
0m m
5m m
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 39
Steps
The 12” ET floor box is heavy (approximately 40 kg). Adhere to your specific health and safety procedures.
Use the hooks to lift the box. Never use the angle frame.
Use the adjustment screws ① to level the floor box into the correct horizontal position and to set the height of the floor box within the recess.
Screw the floor box to the floor at the fix-ation hole ② and if necessary to the walls.
Leave the plastic bags with the screws (for Sapphire tube housing) and the sealing tape in the box. The screws will be needed during installation.
NOTE: Small metal sheets can be placed below the box to avoid that the adjustment screws screw in the con-crete. The metal sheets can be found in-side the box.
②
①
4 After adjustment, the metal rim should be flush with the finished floor.
finished floor metal rim concrete
Place the wooden cover plate ① and the wooden insert ② onto the box.
NOTE: Leave the aluminum window in-sert ③ attached to the wooden cover as it will be needed during installation. The aluminum window insert is delivered in bubble wrap.
③
②
You must fasten the wooden cover into place with screws.
40 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Place first one half and then the second half of the cardboard protection cover on top of the wooden cover. Slide the edges under the metal rim ①.
①
7 Use duct tape to connect the two halves of the protection cover.
Pour quick drying concrete into the gap around the floor box up to the finished floor.
finished floor
0m m
> 10mm filled up with concrete
At least 10 mm of the rim of the aluminum cover must rest on stable (concrete) floor.
The ET floor box must be fixed to the pit with anchors (not supplied by Brainlab) and be set in concrete.
Use mineral material only. Do not use chemical filling.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 41
4.1.4 Customized Cable Outlet
It may be necessary (e.g., for updates) to widen the cable outlet on the 12” Floor Boxes.
The following image shows the allowed area for customizing the cable outlet in the rear wall of the floor boxes.
42 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
4.2 On Floor Boxes (Optional)
4.2.1 Overview
General Information
The on floor boxes are an optional system setup that can be used in case there are constructional or structural limitations that prevent the installation of the standard 12” floor boxes.
②
Figure 14
① A typical treatment room setup for on floor boxes.
②On floor box with a warning ramp and cable raceway.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 43
4.2.2 External Dimensions of the On Floor Boxes
Side View
6m m
Figure 15
Top View
2m m
840mm
Figure 16
On Floor Boxes (Optional)
44 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
4.2.3 Recommended Position of Above-Ground Floor Boxes
Recommended Position Relative to the Isocenter
• Mark the 45° lines and the distance of the leading edge of the tube holder (1215 mm) on the floor for both above ground floor boxes. These markings will be used during installation.
• Aluminum raceways (cable conduits) are available in multiple 1.5 m pieces. Please contact Brainlab if you require a total length of 9, 12 or 15m of raceway.
45°
Cable conduit* 65x30mm (to X-Ray
Generator)
Cable conduit* 65x30mm (to X-Ray
Generator)
1050±100
1250±100 12
*Aluminium raceway delivered and installed by Brainlab
Figure 17 *The raceways will be delivered and installed by Brainlab.
All dimensions for positioning floor boxes are based on isocenter height of 1295 ± 20 mm (finished floor line – isocenter). The exact value depends on the linac Type. Please refer to the respective documentation. If the vertical isocenter position is not within the tolerance range, please contact Brainlab.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 45
On Floor Boxes (Optional)
46 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
5 CEILING AND WALL
MOUNTS
5.1 Introduction
General Information
Brainlab provides all hardware parts, except bolts for the concrete ceiling, as local conditions and building construction varies. The hospital or contractor is responsible for:
• Installing all ceiling and wall mounts according to local work safety regulations.
• Providing appropriate fixation with bolts and anchors according to local regulations.
• Assuring correct installation by signing a compliance statement during the acceptance procedure. Ideally, the statement of compliance should be signed before starting installation.
Maximum Working Height
If you need to stand higher than 2 m above the floor, use a scaffold.
00 30
00 31
Max. working height without scaffold
Max. height for ETX interface plate and in-room distribution box
Figure 18
Types of Ceiling Mounts
List of ExacTrac ceiling mounts:
• Ceiling frame for touch screen monitor (optional)
• Two flat panel ceiling mounts
• One in-room distribution box
If immobile objects are in the way of the mounting positions, you may provide interfaces for the Brainlab equipment (see page 57).
CEILING AND WALL MOUNTS
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 47
Height Limitations
Recommended maximal ceiling heights (not meant to scale):
2.2a 2.2b
1.1
2.3b
0.1
0.0
2.1b2.1a
2.3a
1.2 suspended ceiling concrete ceiling
BL Camera VMS Camera
~1 m m m in
. 1 m m
(lo w er e dg e in te rfa ce pl at e) m ax
. 4
0m m m ax
. 3
0m mm ax . 5
0m m m in
. 1
0m m m in
. 2
0m m
(if c le ar an ce o f c ei lin g-m ou nt ed m on ito r a rm is
0m m
Figure 19
ExacTrac requires the same minimum ceiling height as the linac gantry. If the height of the concrete ceiling is lower than the minimum height required for the linac gantry, contact Brainlab support.
Special Height Requirements
For concrete ceilings higher than the allowed maximum height, additional construction is required to maintain the following maximum mounting heights (measured from the finished floor):
Component Maximum Height
Ceiling mount for monitor arm 4250 mm
Ceiling mount for flat panels 5270 mm
Component Minimum Height Maximum Height
Suspended ceiling 2700 mm 3100 mm
In case of closed mono block suspended ceilings, service openings to access ceiling mounts must be provided.
Suspended ceilings should be no higher than the recommended maximum height (3100 mm), otherwise you must cover all visible parts of the ceiling mounts.
The minimum construction height (clearance between concrete ceiling and interface plate) of the ExacTrac ceiling mounts is 150 mm. If any parts of the ceiling mounts are visible due to suspended ceiling setup, you are responsible for covering these.
Introduction
48 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
The minimum plenum height (clearance between concrete ceiling and suspended ceiling) of the in-room distribution box is 175 mm, although 240 mm are recommended for optimal service access.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 49
5.2 Ceiling Mount Installation
5.2.1 How to Mount the Flat Panel Detector Systems
General Installation Information
Component Description
Weight 50 kg
Factor of safety Four (construction must carry 200 kg)
Bolts required
Six (two for the ceiling bracket, four for support cables) Use the 17 mm hole to attach the angle to the ceiling.
60mm
0m m
Ø18mm
0m m m m
Ø17mm50 m m
88mm
Ceiling Bracket Rope Ceiling Angle
Mounting accuracy for Infrared Camera System
The interface plate has the same orientation as the vertical profile bar is. The slots are perpendicular to the center line.
Mounting accuracy for Flat Panel Detector System
If there are pipes, HVAC, med gas, or other fixed objects ob-structing the mounting position, the ExacTrac ceiling mount may not be moved closer to isocenter, but can be adjusted at a maximum of 30 mm further away from isocenter.
Furthermore the brackets (ceiling plate) may be rotated for 90°. Otherwise, items that impede the installation of the flat panel ceiling brackets must be re-routed or moved.
The Interface Plate must be attached to the profile at a 45° angle. The slots of the plate are perpendicular to the line to-wards the isocenter.
Ceiling Mount Installation
50 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Ceiling Bracket Positioning
The following image shows a top view of how ceiling brackets should be positioned for camera and flat panel suspension:
Flat Panel Ceiling Brackets
X-Ray Tube Floor Box
Camera Ceiling Brackets
C
B
A
C
Isocenter
Figure 20
• A = n/a
• B = 850 mm
• C = 600 mm
NOTE:
• All measurements are taken from the isocenter.
• Define a mounting spot on the floor first.
• Use a plumb bob or laser pointer to project the dimensions to the concrete ceiling.
• Varian defines the distance between the isocenter and the Varian camera mount.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 51
5.2.2 How to Mount the Ceiling Interface Plate
Standard Installation
The following image shows correct rope fixation of the upper interface plate to the ceiling mount:
Figure 21
Incorrect Rope Fixation to the Upper Interface Plate
The following image shows incorrect rope fixation of the upper interface plate to the ceiling mount:
Figure 22
52 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Do not use asymmetric ring patterns.
Ceiling Mount Configurations
The following image shows a side view of the ceiling mount dimensions for flat panels:
concrete ceiling suspended ceiling l1 l2 a x= va ria bl e y=150 mm
<45°
①
Figure 23
The clearance of y = 150 mm between the lower edge of the interface plate of the ExacTrac ceiling mount ① and the upper edge of the suspended ceiling must be reserved for the installation and adjustment of Brainlab devices.
The maximum distance between the interface plate and isocenter must not exceed 1950
mm. For an isocenter height of 1245 mm, the maximum height of the interface plate above the finished floor is 3245 mm.
Rope Tighteners
Use rope tighteners to apply proper tension to the ropes if vertical bars (l > 80 mm) are used for ceiling mount installation.
Figure 24 Ceiling mount configurations depending on the distance between the concrete ceiling and the suspended ceiling:
Config-uration a I1 I2 I = I1 + I2 Bars Ropes Rope Ceil-ing Angles
XS < 220 mm 50 mm n/a 50 mm 1 2 2
S 220 - 250 mm a - 170 mm n/a a - 170 mm 1 2 2
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 53
Config-uration a I1 I2 I = I1 + I2 Bars Ropes Rope Ceil-ing Angles
M 250 - 450 mm a - 170 mm n/a a - 170 mm 1 4 4
L 450 - 2170 mm
(a - 170 mm)/2
(a - 170 mm)/2 a - 170 mm 2 4 4
XL > 2170 mm 1000 mm 1000 mm 2000 mm 2 4 4
Installation Instructions for Different Configurations
Configura-tion
Instructions
XS and S concrete ceiling suspended ceiling l1
150mm
20 Nm (4xM8) a
• Only use the upper vertical bar.
• Cut the bar as straight as possible.
• Cut four M8 threads into the bar and mount the interface plate.
• Tighten the screws with the specified torque.
• It is possible to leave out the tension ropes. However, you must install ropes on either side as additional mechanical protection.
NOTE: Inform Brainlab if a is shorter than 150 mm because the construction could be visible.
Do not use rope tighteners or apply tension to the safety rope.
Fix the rope loop with at least three rope clamps.
54 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Configura-tion
Instructions
M concrete ceiling suspended ceiling l1 a
0m m
<45° z ≥ l + 140mm
20 Nm (4xM8)
• Only use the upper vertical bar.
• Cut the bar as straight as possible.
• Cut four M8 threads into the bar and mount the interface plate.
• Tighten the screws with the specified torque of 20 Nm.
• Mount the rope ceiling angles according to the drawing.
NOTE: The angle between the rope tightener and ceiling can be smaller than 45° if necessary (e.g., angle brackets further away from the ceiling bracket).
Fix the rope loop with at least three rope clamps.
Fix the rope tighteners only after applying the proper tension.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 55
Configura-tion
Instructions
L and XL concrete ceiling suspended ceiling l1 a
0m m
<45° z = l + 140mm
20 Nm (4xM8) l2
<1
0m m
Isocenter
• Use both vertical bars.
• Mount the ceiling angles according to the drawing.
• Assemble the bars’ four connection plates. Fix each connection plate with four tight screws.
• Tighten the screws with the specified torque of 20 Nm.
• L only: Cut each vertical profile bar accordingly at the half-length distance be-tween the interface plate and concrete ceiling.
Fix the rope loop with at least three rope clamps.
Fix the rope tighteners only after applying proper tension.
56 Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Requirements for Custom-Made Ceiling Mount
If it is impossible to install the ceiling mount at the required position, you can provide an interface plate and a custom-made ceiling mount according to the following requirements:
m m
Ceiling
= space for mounting screws-must be easy reachable
0m m
170mm
,5 m m
8m m Custom made ceiling plate
~6 0m m
~220mm
Custom made ceiling plate
Figure 25
You are responsible for adhering to safety factors, local regulations and mounting the custom made interface plate and custom-made ceiling mount.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 57
5.3 Ceiling-Mounted Touch Screen Monitor (Optional)
5.3.1 Overview
General Installation Information
A 17” touch screen monitor is mounted to the ceiling to the left or right of the couch area (see image below). The area of movement is given by the maximum arm extension of 1910 mm. The rotation can be limited by a stopper.
Figure 26
Component Description
Weight 99 kg
Safety factor Must be defined according to the local…