LINAC Vaults 1 & 2 Modifications FY19 (VA-19-00023131) Project 652-16-107
Solicitation number
36C24619B0021
Issued by

About this file

36C24619B0021 0004 ASK-01 Richmond LINAC Vaults 1 _ 2 Modifications.pdf

View the file

Other files

File Type Posted
36C24619B0021-0005000.docx DOCX document
36C24619B0021-0004006.pdf PDF
36C24619B0021-0004003.pdf PDF
36C24619B0021-0004004.pdf PDF
36C24619B0021-0004007.pdf PDF
36C24619B0021-0004005.pdf PDF
36C24619B0021-0004000.docx DOCX document
36C24619B0021-0004002.pdf PDF
36C24619B0021-0003000.docx DOCX document
36C24619B0021-0002002.pdf PDF
36C24619B0021-0002001.pdf PDF
36C24619B0021-0002000.docx DOCX document
36C24619B0021-0001002.pdf PDF
36C24619B0021-0001005.pdf PDF
36C24619B0021-0001000.docx DOCX document
36C24619B0021-0001006.pdf PDF
36C24619B0021-0001007.pdf PDF
36C24619B0021-0001004.pdf PDF
36C24619B0021-0001001.pdf PDF
36C24619B0021-0001003.pdf PDF
36C24619B0021-0001009.pdf PDF
36C24619B0021-0001008.pdf PDF
36C24619B0021-002.docx DOCX document
36C24619B0021-007.pdf PDF
36C24619B0021-004.pdf PDF
36C24619B0021-011.pdf PDF
36C24619B0021-001.docx DOCX document
36C24619B0021-003.pdf PDF
36C24619B0021-010.pdf PDF
36C24619B0021-008.pdf PDF
36C24619B0021-005.pdf PDF
36C24619B0021-006.pdf PDF
36C24619B0021-009.pdf PDF
36C24619B0021-000.docx DOCX document
Show all 34

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

1Z-133

1Z-134

1Z-132

1Z-135

C12

M3B

± 3 1/2"

1Z-133A

9' - 6" 12' - 0"

"M3B

M3B M3B

M3B

M3B

133A

6' - 0" 15' - 1 7/8"

M1B

M1B

M1B

± 25' - 3 3/8"± 21' - 1 7/8"

6'

3"

7'

4"

6'

1/

2"

1' - 11 1/4"

1' - 7 5/8"

M1B

M1BM1BM3

M3

5'

/4

/2

6'

3/

4"

1/

2"

Date

Project Number

PF&A DESIGN

ARCHITECTURE, PLANNING, INTERIORS

World Trade Center 101 West Main Street, Suite 7000 Norfolk, VA 23510

Phone: 757-471-0537 Fax: 757-471-4205 www.pfa-architect.com

AE114

HUNTER

HOLMES

MCGUIRE VAMC

RICHMOND

LINAC

VAULTS 1 & 2

MODIFICATIONS

08/20/18

652-16-107

Wall Type Locations

1/8" = 1'-0"

NEW WORK PLAN - FIRST LEVEL ASK-01

WALL TYPE SCHEDULE ASK-01

TYPE STC DESCRIPTION

FIRE

RATING UL REMARKS

C 12 N/A 12" CONCRETE TO 17' - 9" AFF.

SEAL CONCRETE FACES

EXPOSED TO EXTERIOR.

N/A N/A COORDINATE WITH

STRUCTURAL

DRAWINGS

M1B N/A 1 5/8" METAL STUDS AT 16" O.C.

MAX. 5/8" GWB ONE SIDE.

EXTEND WALL ASSEMBLY TO 4"

ABOVE FINISHED CEILING.

N/A N/A

M3 N/A 3 5/8" METAL STUDS AT 16" O.C.

MAX. 5/8" GWB BOTH SIDES.

EXTEND WALL ASSEMBLY TO 4"

ABOVE FINISHED CEILING.

N/A N/A

M3B N/A 3 5/8" METAL STUDS AT 16" O.C.

MAX. 5/8" GWB ONE SIDE.

EXTEND WALL ASSEMBLY TO 4"

ABOVE FINISHED CEILING.

N/A N/A

Sheets
AE114 - Wall Type Locations