About this file

This document is a space plan and equipment specifications related to a federal contract opportunity for the Core Lab electrical upgrade at the Washington DC Veterans Affairs Medical Center (DC VAMC).

The key details are:

  • The Core Lab is installing new processing equipment on the basement floor in Suite GB 200.
  • A 40 KVA Uninterruptable Power Supply (UPS) will be used to provide emergency power to the new equipment.
  • The electrical contractor is required to install the power distribution and interface the equipment to the 40 KVA UPS located in room GB 220.
  • Detailed power and utility requirements are provided for each piece of Roche Diagnostics equipment, including power voltage, amperage, phases, receptacle types, UPS ratings, and heat dissipation.
  • The solicitation number is 36C24524Q0670 and the agency is the Department of Veterans Affairs Veterans Health Administration.

View the file

Other files for this federal contract opportunity

Other files attached to J059--Posting Answers to Vendor Questions - Core Lab Electrical Upgrade Install - DC VAMC 688-24-3-459-0307, newest first.
File Type Posted
PowerVar 3300 users manaul A01-00134 3300 REV F 4-5-21.pdf PDF
ANSWERS TO VENDOR QUESTIONS - 36C24524Q0670.pdf PDF
3300 Series Footprint and Clearance Requirements.pdf PDF
UPS Specs.pdf PDF
36C24524Q0670 0001_1.docx DOCX document
Power Conditioning Distribution Unit.pdf PDF
36C24524Q0670 0001.pdf PDF
Wage Determination Washington D.C.pdf PDF
36C24524Q0670 -.pdf PDF
36C24524Q0670_1.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

NOT FOR CONSTRUCTION

KEY LEGEND

INSTALL

x mm.

x’-x”

REFERENCE

x’-x”

WATER/DRAIN

x’-x”

POWER

NEW FLOOR DRAIN

EXISTING FLOOR DRAIN

# UTILITY LOCATION

DATA

POWER

WATER LOOP

W/SHUT OFF

CCM PSU

AIR

ROCHE DIAGNOSTICS

9115 HAGUE ROAD

INDIANAPOLIS, IN 46256

SCALE: ¼” = 1'-0" SHEET 1 OF 3

WASHINGTON VA

MEDICAL CENTER

WASHINGTON, DC

VA Washington DC_8100 Space Plan111422MS-1

LDA: MARK SIMPSON

SWC: HARRY LANZONI

GENERAL NOTES:

1.SIZE AND CONFIGURATION OF SYMBOLS

SHOWN ON THE DRAWING ARE USED TO

REPRESENT THE EXISTENCE OF, OR NEED FOR

ADDITIONAL UTILITIES.

2.CUSTOMER IS RESPONSIBLE FOR MEETING ALL

LOCAL CODES.

3.CUSTOMER IS RESPONSIBLE FOR CONTACTING

EACH 3

RD

PARTY VENDOR TO OBTAIN TECHNICAL

EQUIPMENT SPECIFICATIONS RELATED TO

INSTALLATION OF INSTRUMENTS AND REQUIRED

UTILITIES. ROCHE WILL PROVIDE TECHNICAL

SPECIFICATIONS AND GENERAL LOCATIONS FOR

UTILITIES OF ROCHE EQUIPMENT ONLY.

CONNECTION TO 3

RD

PARTY SYSTEMS TO BE

VERIFIED BY CUSTOMER.

4.THIS DRAWING IS SUBJECT TO CHANGE.

DATE: DESCRIPTION:

XX/XX/XX TEXT

PROJECT HISTORY:

UTILITIES – SPACE PLAN

Lo ck e rs

Lo ck e rs

Fr zr

-2 c

Refrig Fisher Refrig

S a fe ty s h o w e r

H2O system

Shred

Box

H2O system

S ta in e r

R e fr ig e ra to rs a n d S u p p li e s

Exist ing drain

Exist ing drain

Exist ing drain

Fr ig

Exist ing drain

R ef ri g

Refrig

-8 c Fr ee ze r

RefrigRefrig

D xH

D xH

M anuals

S T

A -C o m p a c t

Fi le

Fi le

Fi le

Shelf

Sh el f

Shelf S hred

Box

S T

A -C o m p a c t

PTS

H20

S ta in e r

L

IS

Heme Archive Table Heme Input Table

New +4c

N ew

4c

D I C o n tro lle r

8100 Control Unit & Report Printer

R S

F A

O B

O B

S c o b a s o u tp u t s ta tio n

DSPACU with ACBIPB

HS

cobas 8100 input station

ACU with ACB BCL AQM cobas 8100 aliquot station

CRW

c o b a s

TLJ

T L J

TLJ

TLJ

S L L cobas 8100

B R

F c o b a s p ro c o b a s p ro

B R

F c o b a s p ro c o b a s p ro c o b a s p ro c o b a s L in k

R e p o rt

P ri n te r c o b a s p ro c o b a s L in k

R e p o rt

P ri n te r c o b a s

A

T L

SUGGESTED AREA FOR NEW DRAINS.

IF REUSING EXISTING DRAINS, THEY SHOULD BE WITHIN 16

FT OF THE INSTRUMENT. IF NOT, SEE ROCHE INSTALLATION

SITE REQUIREMENTS FOR ADDITIONAL SPECIFICATIONS.

SUGGESTED AREA

FOR UTILITY DROPS.

TYPICAL OF 4

Data Power Water Loop

Power

#1

U P

S

.0 k V

A

S e ri e s

B a tt e ry C a b in e t

See UPS/Battery Cabinet Specs.

Power

Off-site location TBD. Max distance is 250ft.

#0

Firewall c o b a s p ro

IS

E -5

-I S

E -5

-8 c o b a s p ro

IS

E -5

-8

Virtual Server Primary

Virtual Server Shadow

Off-site location TBD.

#0

Data

Water Loop

FirewallFirewall

#2

#4

Data

#3

KEY LEGEND

INSTALL

x mm.

x’-x”

REFERENCE

x’-x”

WATER/DRAIN

x’-x”

POWER

NEW FLOOR DRAIN

EXISTING FLOOR DRAIN

# UTILITY LOCATION

DATA

POWER

WATER LOOP

W/SHUT OFF

CCM PSU

AIR

ROCHE DIAGNOSTICS

9115 HAGUE ROAD

INDIANAPOLIS, IN 46256

WASHINGTON VA

MEDICAL CENTER

WASHINGTON, DC

LDA: MARK SIMPSON

SWC: HARRY LANZONI

GENERAL NOTES:

1.SIZE AND CONFIGURATION OF SYMBOLS

SHOWN ON THE DRAWING ARE USED TO

REPRESENT THE EXISTENCE OF, OR NEED FOR

ADDITIONAL UTILITIES.

2.CUSTOMER IS RESPONSIBLE FOR MEETING ALL

LOCAL CODES.

3.CUSTOMER IS RESPONSIBLE FOR CONTACTING

EACH 3

RD

PARTY VENDOR TO OBTAIN TECHNICAL

EQUIPMENT SPECIFICATIONS RELATED TO

INSTALLATION OF INSTRUMENTS AND REQUIRED

UTILITIES. ROCHE WILL PROVIDE TECHNICAL

SPECIFICATIONS AND GENERAL LOCATIONS FOR

UTILITIES OF ROCHE EQUIPMENT ONLY.

CONNECTION TO 3

RD

PARTY SYSTEMS TO BE

VERIFIED BY CUSTOMER.

4.THIS DRAWING IS SUBJECT TO CHANGE.

DATE: DESCRIPTION:

XX/XX/XX TEXT

PROJECT HISTORY:

CUSTOMER PROVIDED INTERFACE CONNECTIONS PROVIDED ON CONSTRUCTION PLAN

EQUIPMENT SPECIFICATIONS

NOT TO SCALE SHEET 2 OF 3

VA Washington DC_8100 Space Plan111422MS-1

TOTAL SYSTEM HEAT LOAD

BTU/HR (VALUES ROUNDED)

103,808

AT UTILITY

LOCATION ROCHE SUPPLIED EQUIPMENT POWERED FROM

NUMBER OF DEVICE

POWER CABLES

PRIMARY POWER,

VOLTS

POWER PROTECTION,

AMPS PHASE RECEPTACLE TYPE

DEVICE'S UPS

RATING, KVA

HEAT

DISSIPATION,

BTU/HR

0 UPS 40.0 kVA WITH BATTERY CABINET FACILITY 0 208 150 3 PHASE 4 CONDUCTOR WITH GROUND HARDWIRED FROM CUSTOMER'S FACILITIES DEPARTMENT 40.0 12,270

0 DI - SHADOW CONTROL UNIT (PC & MONITOR) UPS 40.0 kVA MAIN LUG PANEL (MLP) 2 120 20 SINGLE PHASE NEMA 5-20R 0.6 1,025

0 DI - PRIMARY CONTROL UNIT (PC & MONITOR) UPS 40.0 kVA MAIN LUG PANEL (MLP) 2 120 20 SINGLE PHASE NEMA 5-20R 1.4 1,655

1 REPORT PRINTER (from COBAS PRO) FACILITY 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 1,877

1 COBAS LINK 2 UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 0

1 FORTIGATE FIREWALL UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 61

1 COBAS_PRO <ISE/503/ISE/503/801> UPS 40.0 kVA MAIN LUG PANEL (MLP) 2 208 50 SINGLE PHASE HUBBELL 3762C (CORD) OR 3771 (WALL MOUNT) 10.0 28,661

2 REPORT PRINTER (from COBAS PRO) FACILITY 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 1,877

2 COBAS LINK 2 UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 0

2 FORTIGATE FIREWALL UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 61

2 COBAS_PRO <ISE/503/801> UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 208 50 SINGLE PHASE HUBBELL 3762C (CORD) OR 3771 (WALL MOUNT) 10.0 19,107

3 REPORT PRINTER (from COBAS 8100) FACILITY 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 2,048

3 COBAS 8100 CONTROL UNIT (PC & MONITOR) UPS 40.0 kVA MAIN LUG PANEL (MLP) 2 120 20 SINGLE PHASE NEMA 5-20R 0.8 1,550

3 FORTIGATE FIREWALL UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 61

3 COBAS 8100 SYSTEM UPS 40.0 kVA MAIN LUG PANEL (MLP) 5 208 30 SINGLE PHASE NEMA L6-30R 5.2 33,385

4 AUTOMATED TIP LOADER FACILITY 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 170

AT UTILITY

LOCATION CUSTOMER SUPPLIED DATA CONNECTION CONNECTION TYPE

1 DATA UTILITY SEE CUSTOMER PROVIDED INTERFACE CONNECTIONS

2 DATA UTILITY SEE CUSTOMER PROVIDED INTERFACE CONNECTIONS

3 DATA UTILITY SEE CUSTOMER PROVIDED INTERFACE CONNECTIONS

AT UTILITY

LOCATION CUSTOMER SUPPLIED WATER CONNECTION CONNECTION SIZE

DEIONIZED WATER

SUPPLY RATE,

LITERS/HR

MINIMUM FLOOR DRAIN

INSIDE DIAMETER,

INCHES

1 WATER UTILITY 5/8" BARBED HOSE FITTING W 1/2" ID 98 2

2 WATER UTILITY 5/8" BARBED HOSE FITTING W 1/2" ID 64 2

CUSTOMER'S WATER/WASTE RESPONSIBILITY

NOTE: TYPE 1 CLRW, UTILIZING A CONTINUOUS RECIRCULATION /POLISHING LOOP, WITH A MINIMAL VELOCITY OF 5 FT/SEC. AND A 0.2 MICRON OUTPUT FILTER. PLEASE REFER TO THE SPECIFIC INSTALLATION SITE REQUIREMENTS DOCUMENTATION.

CUSTOMER'S POWER RESPONSIBILITY

CUSTOMER'S DATA RESPONSIBILITY

NOT TO SCALE SHEET 3 OF 3

ROCHE DIAGNOSTICS

9115 HAGUE ROAD

INDIANAPOLIS, IN 46256

POWERVAR 3300 SERIES UPS CIRCUIT DIAGRAM CLEARANCE SPECIFICATIONS

RECOMMENDED INSTALLATION OPTIONAL PCDU OPTIONAL PCDU INSTALLATION

VA Washington DC_8100 Space Plan111422MS-1.vsdx
1/4" - SPACE PLAN
1/4" - SPECS
1/4" - UPS/PCDU SPECS