534B52004-XR RAD VAMS CHARLESTON SC_Tech Specification.docx

DOCX document 266 KB Posted

Attached to
6525--DEC 2024 Equipment Only Consolidation Federal contract opportunity
Solicitation number
36A79725Q0003
Issued by
Department of Veterans Affairs National Acquisition Center

About this file

This document is a technical specification for a digital radiographic system for the Ralph H. Johnson VA Medical Center in Charleston, SC, specifically for the Brunswick VA Clinic in Brunswick, GA. The specification outlines requirements for a digital x-ray system to be installed in an existing renovated room, with key technical details including a wireless wall detector configuration. The system will support expanded services for Veterans, with a 14x17 inch minimum wall detector size and specifications for wireless detector weight. The related federal contract opportunity (Solicitation Number: 36A79725Q0003) is a December 2024 Equipment Only Consolidation issued by the Department of Veterans Affairs National Acquisition Center, involving multiple documents including vendor workup spreadsheets, VA directives, and equipment consolidation schedules.

View the file

Other files for this federal contract opportunity

Other files attached to 6525--DEC 2024 Equipment Only Consolidation, newest first.
File Type Posted
36A79725Q0003 0008.docx DOCX document
534B52004-XR RAD VAMS CHARLESTON SC_Compliance Matrix.xlsx XLSX spreadsheet
36A79725Q0003 0007.docx DOCX document
36A79725Q0003 0006.docx DOCX document
Final Updated Tech Specs- Compliance Matrix.zip ZIP file
DEC 2024 EQ Consolidation - Schedule of Line Items w QA.xlsx XLSX spreadsheet
36A79725Q0003 0005.docx DOCX document
36A79725Q0003 0004.docx DOCX document
630B50014 _ XR US VAMC NEW YORK NY_ Tech Specification.docx DOCX document
36A79725Q0003 0003.docx DOCX document
36A79725Q0003 0002.docx DOCX document
DEC 2024 EQ Consolidation - RFO- 2-6-2025.pdf PDF
DEC 2024 EQ Consolidation - RFO Update.pdf PDF
36A79725Q0003 0001.docx DOCX document
Vendor Folders.zip ZIP file
36A79725Q0003_1.docx DOCX document
Vendor Workup Spreadsheet - 36A79725Q0003.xlsx XLSX spreadsheet
589B51001 _ 612B59000.zip ZIP file
436B50004 _ 589B51000.zip ZIP file
DEC 2024 EQ Consolidation _Schedule of Line Items.xlsx XLSX spreadsheet
DEC 2024 Equipment Only Consolidation Schedule.pdf PDF
Memorandum for Training.pdf PDF
VA Directive 6550 - New Appendix A - published Jun. 2023.pdf PDF
612B59001 _ 691B50047.zip ZIP file
DEC 2024 EQ Consolidation - RFO.pdf PDF
Show all 25

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Equipment Specifications Radiographic Systems: Digital (XR RAD)

V07 CHARLESTON SC

534B52004 v19.02

REQUIREMENT OVERVIEW

Ralph H. Johnson in Charleston, SC recently acquired the Brunswick, GA CBOC, formerly under Dublin’s healthcare system. Additional services will be offered to Veterans, including x-ray. An existing room will be renovated to house a digital radiographic system.

☒
1. Extended / Turnkey installation services are not required
Facility Name
Brunswick VA Clinic
Delivery Address
1111 Glynco Parkway, Bldg. 2, Suite 200, Brunswick, GA 31525
Requesting Department(s)
Radiology

technical requirements System digital detector configuration

a. Wall detector

☒
a. Wall detector
Wireless
☒
b. Minimum wall detector size [in x in]
14x17
☒
c. Maximum wireless wall detector weight, with battery installed [lbs]
<10

b. Table detector

☒
a. Table detector
Wireless
☒
b. Minimum table detector size [in x in]
14x17
☒
c. Maximum wireless table detector weight, with battery installed [lbs]
<10

c. Additional wireless detector(s)

☒
a. Quantity of additional wireless detectors
1
☒
b. Minimum 1st wireless detector size [in x in]
14x17
☒
c. Maximum 1st wireless detector weight, with battery installed [lbs]
<10

Unit physical specifications

a. Minimum generator output power [kW]
80
b. Minimum generator voltage range [kV]
40-150
c. Minimum spatial resolution [lp/mm]
5
d. Maximum pixel size [µm]
140
e. Maximum time for image display [s]
5
f. Minimum tube height from floor when used with wall stand [in]
11
g. Minimum image plate height from floor [in]
11
h. Minimum patient table load capacity (static) [lbs]
675
i. Minimum patient table load capacity (dynamic) [lbs]
675
j. Minimum table height from floor [in]
22-33
k. Wall stand bucky tilt range [deg]
90

Additional specifications

☒
a. High-frequency generator with automatic dose rate control
☒
b. Protective weight-bearing cover for wireless detector
☒
c. Charging station for wireless detector (as required by vendor)
☒
d. Grid attachments for wireless detector (as required by vendor)
☒
e. Table holder for wireless detector
☒
f. Mobile holder for wireless detector
☒
g. Tube mount
Overhead
☒
h. Bucky tracking for the table and wall stand
☒
i. In-room protocol adjustment via tube head or other in-room mounted system
☒
j. Ability to change between tabletop, upright bucky, or table bucky from tube head
☒
k. Free-floating manual tabletop
☒
l. Automatic parameter selection
☒
m. Foot controls on each side of table
☒
n. Auto Tracking
☒
o. Auto Positioning

Workstation Requirements

☒
a. Minimum acquisition workstation monitor size [in]
23
☒
b. Minimum acquisition workstation hard drive size [GB]
1
☒
c. UPS for the acquisition workstation

Software Requirements

☒
a. Stitching – ability to stitch multiple long-axis images with options for automated and manual process
☒
b. Bone suppression – ability to suppress overlaying bone such as ribs on chest images
☒
c. Ability to apply multiple image processing algorithms both pre- and post-acquisition to allow for soft tissue and/or bone enhancement
☒
d. Dose reduction – facilitate regular protocol optimization and reduced radiation dose to the patient
☒
e. Repeat rate – ability to track repeat/retake data to include such items as technologist (required unique identifier), reason for repeat, patient dose, exam type, etc. The data should be exportable to Excel or other databases for tracking, trending, and combining with data from other imaging sources within the facility.

Security/Connectivity Requirements

☒
a. OEM-supported operating system
☒
b. Latest DICOM print, store, commit, structured report (SR) and modality worklist
☒
c. Encrypted hard drive
☒
d. PACS compatibility – Philips Vue PACs
☒
e. PACS compatibility – VA VistA Imaging
☒
f. Integration to Dose Tracking System – Nexodose

training requirements Clinical Training

☒
a. On-site clinical applications training for 2 technologists during go-live
☒
b. On-site follow-up clinical applications training for 2 technologists once technologists have hands-on experience with the system
☒
c. Technologists who complete the clinical applications training shall receive continuing education credits (CMEs).
☒
d. Vendors shall be responsible for accommodating different personnel shifts for clinical applications training during go-live.

Biomedical Technician Training Please reference the “Instructions to Offers” for further information about the type of information to provide by equipment type not by specific request. Please also reference the “Instructions to Offers” for response format.

Technical training information to include detailed information about the curriculum and length of the biomedical technical training required for each equipment type.

Although the NAC will not award this training along with the equipment, it is imperative that the customer is informed that this training is available. Vendors must demonstrate that they can provide any required off-site training, therefore off-site training should be quoted as an optional item. Off-site training will be purchased at the time of need via a modification (if the original order remains open) or via a separate order. No travel expenses for any VA employees will be included in any HTME equipment or training order.

service requirements

1. VPN/Remote Access – The vendor shall provide any and all equipment service programs, such as remote diagnostics, during the warranty period. The vendor shall provide post-warranty remote diagnostic service program as an “Add Option” with the offer. The system shall provide vendor remote diagnostics via VPN. The vendor shall either utilize the VA national site-to-site VPN or work with the Office of Cyber and Information Security and the VAMC Information Systems Security Officer to establish a client-based VPN.

2. Service and Operator Manuals – The vendor shall provide the following documentation for the proposed systems:

a. Two (2) copies of operator instruction manuals (one (1) electronic and one (1) physical copy)

b. Two (2) copies of a service manuals (one (1) electronic and one (1) physical copy) *Vendors can include the physical copy as a priced line item in their quote as applicable.

3. Minimum Warranty – The system and accessories shall be covered under the manufacturer’s warranty and shall include all parts and labor for one year following acceptance by the VAMC. This warranty must include PMs as required by the manufacturer. The manufacturer’s factory-trained field service representatives shall perform installation and maintenance during the warranty period.

Vendors are encouraged to include any offerings for service, warranty, and training that may exceed the minimum requirements, to include information on their service support structure during and after the warranty period. Vendors who do not include any added value offerings for service, warranty, and training will not be docked or excluded from consideration. However, any such offerings will be evaluated based on added value.

Added Value Specifications listed below are not required, but preferred. Vendors who do not include the below specifications in the submitted offer will not be docked or excluded from consideration. Specifications listed below will be evaluated based on added value.

☒
a. Additional year(s) of warranty
☒
b. Post-warranty remote diagnostic service program
☒
c. Version/platform long-range plan
☒
d. Device compatible with Cerner CareAware MultiMedia (CAMM) Release
☒
e. Device compatible with Cerner CareAware MultiMedia (CAMM) Release 7:

Device receives Patient ID (0010, 0020), Issuer of Patient ID (0010, 0021), Other Patient IDs Sequence Patient ID (0010, 1002 (0010, 0020)), and Other Patient IDs Sequence Issuer of Patient ID (0010, 1002 (0010, 0021)) from worklist provider.

☒
f. Device compatible with Cerner CareAware MultiMedia (CAMM) Release 7:

Device sends Patient ID (0010, 0020), Issuer of Patient ID (0010, 0021), Other Patient IDs Sequence Patient ID (0010, 1002 (0010, 0020))and Other Patient IDs Sequence Issuer of Patient ID (0010, 1002 (0010, 0021)) to the storage location.

☒
g. Integrated camera for patient positioning

other information/documentation requested Please reference the “Instructions to Offers” for further information about the type of information to provide by equipment type not by specific request. Please also reference the “Instructions to Offers” for response format.

1. Completed pre-procurement assessment form (6550 Appendix A)

2. Completed Manufacture Disclosure Statement for Medical Device Security (MDS2) form

3. Federal Information Processing Standard (FIPS) 140-2 or (FIPS) 140-3 certification

4. Product brochures

5. Technical specification sheets, to include dimensions and weight of the system

6. Typical drawings (pdf version of the CAD drawings)

7. Technical training- Biomedical: information to include detailed information about the curriculum and length of the biomedical technical training required for each equipment type.

· Although the NAC will not award this training along with the equipment, it is imperative that the customer is informed that this training is available. Vendors must demonstrate that they can provide any required off-site training, therefore off-site training should be quoted as an optional item. Off-site training will be purchased at the time of need via a modification (if the original order remains open) or via a separate order. No travel expenses for any VA employees will be included in any HTME equipment or training order.

8. Support information to include your company’s support structure during and after the warranty period

· On-line or telephonic applications support and availability (include third party coverage)

· A listing of field service engineer locations and availability

· A listing of part depots trade-in

☒
1. VA has no trade-in units to offer.