534B52004-XR RAD VAMS CHARLESTON SC_Compliance Matrix.xlsx

XLSX spreadsheet 28 KB Posted

Attached to
6525--DEC 2024 Equipment Only Consolidation
Solicitation number
36A79725Q0003
Issued by

About this file

This document is a Compliance Matrix for a digital radiographic system procurement for the Ralph H. Johnson VA Medical Center in Charleston, SC, specifically targeting the Brunswick, GA Community-Based Outpatient Clinic (CBOC). The matrix provides detailed technical requirements for a digital x-ray system, focusing on wall-mounted wireless detector specifications, with dimensions of 14x17 inches and weight requirements for the wireless detector. The procurement is associated with Solicitation Number 36A79725Q0003, part of the Department of Veterans Affairs National Acquisition Center's December 2024 Equipment Only Consolidation, which involves equipment procurement for VA healthcare facilities. Vendors are instructed to complete the compliance matrix by indicating how their proposed system meets each specification, with options to respond as "exceeds," "meets," "partially meets," or "does not meet," and to provide supporting documentation for their responses.

View the file

Other files

File Type Posted
534B52004-XR RAD VAMS CHARLESTON SC_Tech Specification.docx DOCX document
36A79725Q0003 0008.docx DOCX document
36A79725Q0003 0007.docx DOCX document
36A79725Q0003 0006.docx DOCX document
DEC 2024 EQ Consolidation - Schedule of Line Items w QA.xlsx XLSX spreadsheet
Final Updated Tech Specs- Compliance Matrix.zip ZIP file
36A79725Q0003 0005.docx DOCX document
36A79725Q0003 0004.docx DOCX document
630B50014 _ XR US VAMC NEW YORK NY_ Tech Specification.docx DOCX document
36A79725Q0003 0003.docx DOCX document
36A79725Q0003 0002.docx DOCX document
DEC 2024 EQ Consolidation - RFO- 2-6-2025.pdf PDF
DEC 2024 EQ Consolidation - RFO Update.pdf PDF
36A79725Q0003 0001.docx DOCX document
DEC 2024 Equipment Only Consolidation Schedule.pdf PDF
36A79725Q0003_1.docx DOCX document
589B51001 _ 612B59000.zip ZIP file
DEC 2024 EQ Consolidation _Schedule of Line Items.xlsx XLSX spreadsheet
Vendor Folders.zip ZIP file
Vendor Workup Spreadsheet - 36A79725Q0003.xlsx XLSX spreadsheet
436B50004 _ 589B51000.zip ZIP file
Memorandum for Training.pdf PDF
VA Directive 6550 - New Appendix A - published Jun. 2023.pdf PDF
DEC 2024 EQ Consolidation - RFO.pdf PDF
612B59001 _ 691B50047.zip ZIP file
Show all 25

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Compliance Matrix

Radiographic Systems: DigitalCompliance Matrix Created:
V074/14/25
CHARLESTON SC
534B52004
OFFEROR
MANUFACTURER
MODEL
QUOTE NUMBER

VENDOR INSTRUCTIONS:

Please fill out the table above with the offeror name (if different than the device manufacturer), manufacturer name, model name/number, and the quote number. All required specifications are listed below. Please review the required specifications and provide supporting information about the proposed model. The "Vendor Response" column is limited to the following entries: "exceeds", "meets," "partially meets," and "does not meet." In the "Documentation" column, please indicate the location within the attached documentation (file name, page number) for information to support the vendor response. Please provide any additional information/explanations in the "Vendor Explanation" column. **NOTE THAT THIS SPREADSHEET HAS BEEN LOCKED TO ALLOW ONLY FOR FILLING OUT REQUIRED FIELDS (THE TABLE ABOVE AND COLUMNS E, F, & G). DO NOT MAKE ANY FORMATTING OR CONTENT CHANGES TO THE COMPLIANCE MATRIX.**

REQUIREMENT OVERVIEWVENDOR RESPONSEVENDOR DOCUMENTATIONVENDOR EXPLANATION
Ralph H. Johnson in Charleston, SC recently acquired the Brunswick, GA CBOC, formerly under Dublin’s healthcare system. Additional services will be offered to Veterans, including x-ray. An existing room will be renovated to house a digital radiographic system.
☒ a. Extended / Turnkey installation services are not required
Facility Name
Brunswick VA Clinic
Delivery Address
1111 Glynco Parkway, Bldg. 2, Suite 200, Brunswick, GA 31525
Requesting Department(s)
Radiology
TECHNICAL REQUIREMENTSVENDOR RESPONSEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. System digital detector configuration
a. Wall detector
☒ a. Wall detector
Wireless
☒ b. Minimum wall detector size [in x in]
14x17
☒ c. Maximum wireless wall detector weight, with battery installed [lbs]
<10
b. Table detector
☒ a. Table detector
Wireless
☒ b. Minimum table detector size [in x in]
14x17
☒ c. Maximum wireless table detector weight, with battery installed [lbs]
<10
c. Additional wireless detector(s)
☒ a. Quantity of additional wireless detectors
1
☒ b. Minimum 1st wireless detector size [in x in]
14x17
☒ c. Maximum 1st wireless detector weight, with battery installed [lbs]
<10
2. Unit physical specifications
a. Minimum generator output power [kW]
80
b. Minimum generator voltage range [kV]
40-150
c. Minimum spatial resolution [lp/mm]
5
d. Maximum pixel size [µm]
140
e. Maximum time for image display [s]
5
f. Minimum tube height from floor when used with wall stand [in]
11
g. Minimum image plate height from floor [in]
11
h. Minimum patient table load capacity (static) [lbs]
675
i. Minimum patient table load capacity (dynamic) [lbs]
675
j. Minimum table height from floor [in]
22-33
k. Wall stand bucky tilt range [deg]
90
3. Additional specifications
☒ a. High-frequency generator with automatic dose rate control
☒ b. Protective weight-bearing cover for wireless detector
☒ c. Charging station for wireless detector (as required by vendor)
☒ d. Grid attachments for wireless detector (as required by vendor)
☒ e. Table holder for wireless detector
☒ f. Mobile holder for wireless detector
☒ g. Tube mount
Overhead
☒ h. Bucky tracking for the table and wall stand
☒ i. In-room protocol adjustment via tube head or other in-room mounted system
☒ j. Ability to change between tabletop, upright bucky, or table bucky from tube head
☒ k. Free-floating manual tabletop
☒ l. Automatic parameter selection
☒ m. Foot controls on each side of table
☒ n. Auto Tracking
☒ o. Auto Positioning
4. Workstation Requirements
☒ a. Minimum acquisition workstation monitor size [in]
23
☒ b. Minimum acquisition workstation hard drive size [GB]
1
☒ c. UPS for the acquisition workstation
5. Software Requirements
☒ a. Stitching – ability to stitch multiple long-axis images with options for automated and manual process
☒ b. Bone suppression – ability to suppress overlaying bone such as ribs on chest images
☒ c. Ability to apply multiple image processing algorithms both pre- and post-acquisition to allow for soft tissue and/or bone enhancement
☒ d. Dose reduction – facilitate regular protocol optimization and reduced radiation dose to the patient
☒ e. Repeat rate – ability to track repeat/retake data to include such items as technologist (required unique identifier), reason for repeat, patient dose, exam type, etc. The data should be exportable to Excel or other databases for tracking, trending, and combining with data from other imaging sources within the facility.
6. Security/Connectivity Requirements
☒ a. OEM-supported operating system
☒ b. Latest DICOM print, store, commit, structured report (SR) and modality worklist
☒ c. Encrypted hard drive
☒ d. PACS compatibility – Philips Vue PACs
☒ e. PACS compatibility – VA VistA Imaging
☒ f. Integration to Dose Tracking System – Nexodose
TRAINING REQUIREMENTSVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. Clinical Training
☒ a. On-site clinical applications training for 2 technologists during go-live
☒ b. On-site follow-up clinical applications training for 2 technologists once technologists have hands-on experience with the system
☒ c. Technologists who complete the clinical applications training shall receive continuing education credits (CMEs).
☒ d. Vendors shall be responsible for accommodating different personnel shifts for clinical applications training during go-live.
2. Biomedical Technician Training
Please reference the “Instructions to Offers” for further information about the type of information to provide by equipment type not by specific request. Please also reference the “Instructions to Offers” for response format.
Technical training information to include detailed information about the curriculum and length of the biomedical technical training required for each equipment type.
Although the NAC will not award this training along with the equipment, it is imperative that the customer is informed that this training is available. Vendors must demonstrate that they can provide any required off-site training, therefore off-site training should be quoted as an optional item. Off-site training will be purchased at the time of need via a modification (if the original order remains open) or via a separate order. No travel expenses for any VA employees will be included in any HTME equipment or training order.
SERVICE REQUIREMENTSVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. VPN/Remote Access – The vendor shall provide any and all equipment service programs, such as remote diagnostics, during the warranty period. The vendor shall provide post-warranty remote diagnostic service program as an “Add Option” with the offer. The system shall provide vendor remote diagnostics via VPN. The vendor shall either utilize the VA national site-to-site VPN or work with the Office of Cyber and Information Security and the VAMC Information Systems Security Officer to establish a client-based VPN.
2. Service and Operator Manuals – The vendor shall provide the following documentation for the proposed systems:
a. Two (2) copies of operator instruction manuals (one (1) electronic and one (1) physical copy)
b. Two (2) copies of a service manuals (one (1) electronic and one (1) physical copy)
*Vendors can include the physical copy as a priced line item in their quote as applicable.
3. Minimum Warranty – The system and accessories shall be covered under the manufacturer’s warranty and shall include all parts and labor for one year following acceptance by the VAMC. This warranty must include PMs as required by the manufacturer. The manufacturer’s factory-trained field service representatives shall perform installation and maintenance during the warranty period.
Vendors are encouraged to include any offerings for service, warranty, and training that may exceed the minimum requirements, to include information on their service support structure during and after the warranty period. Vendors who do not include any added value offerings for service, warranty, and training will not be docked or excluded from consideration. However, any such offerings will be evaluated based on added value.
ADDED VALUEVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
Specifications listed below are not required, but preferred. Vendors who do not include the below specifications in the submitted offer will not be docked or excluded from consideration. Specifications listed below will be evaluated based on added value.
☒ a. Additional year(s) of warranty
☒ b. Post-warranty remote diagnostic service program
☒ c. Version/platform long-range plan
☒ d. Device compatible with Cerner CareAware MultiMedia (CAMM) Release
☒ e. Device compatible with Cerner CareAware MultiMedia (CAMM) Release 7:

Device receives Patient ID (0010, 0020), Issuer of Patient ID (0010, 0021), Other Patient IDs Sequence Patient ID (0010, 1002 (0010, 0020)), and Other Patient IDs Sequence Issuer of Patient ID (0010, 1002 (0010, 0021)) from worklist provider.

☒ f. Device compatible with Cerner CareAware MultiMedia (CAMM) Release 7:

Device sends Patient ID (0010, 0020), Issuer of Patient ID (0010, 0021), Other Patient IDs Sequence Patient ID (0010, 1002 (0010, 0020))and Other Patient IDs Sequence Issuer of Patient ID (0010, 1002 (0010, 0021)) to the storage location.

☒ g. Integrated camera for patient positioning
OTHER INFORMATION/DOCUMENTATION REQUESTEDVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
Please reference the “Instructions to Offers” for further information about the type of information to provide by equipment type not by specific request. Please also reference the “Instructions to Offers” for response format.
1. Completed pre-procurement assessment form (6550 Appendix A)
2. Completed Manufacture Disclosure Statement for Medical Device Security (MDS2) form
3. Federal Information Processing Standard (FIPS) 140-2 or (FIPS) 140-3 certification
4. Product brochures
5. Technical specification sheets, to include dimensions and weight of the system
6. Typical drawings (pdf version of the CAD drawings)
7. Technical training- Biomedical: information to include detailed information about the curriculum and length of the biomedical technical training required for each equipment type.
Although the NAC will not award this training along with the equipment, it is imperative that the customer is informed that this training is available. Vendors must demonstrate that they can provide any required off-site training, therefore off-site training should be quoted as an optional item. Off-site training will be purchased at the time of need via a modification (if the original order remains open) or via a separate order. No travel expenses for any VA employees will be included in any HTME equipment or training order.
8. Support information to include your company’s support structure during and after the warranty period
On-line or telephonic applications support and availability (include third party coverage)
A listing of field service engineer locations and availability
A listing of part depots
TRADE-INVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
☒ a. VA has no trade-in units to offer.

File details come from the government source that posted it. Updated .