668-B80015___20_Mann-Grandstaff_VA_,_668-B80015revised_V2.xlsx

XLSX spreadsheet 29 KB Posted

Attached to
SEPTEMBER 2018 CONSOLIDATION
Solicitation number
36A79719Q0002
Issued by

About this file

36A79719Q0002 0002 668-B80015 _ 20 Mann-Grandstaff VA , 668-B80015revised_V2.xlsx

View the file

Other files

File Type Posted
589_B82017revised.pdf PDF
589-B80016___VISN_15_KCVA_Medical_Center,_589-B80016revised.xlsx XLSX spreadsheet
36A79719Q0002_0004.docx DOCX document
589-B80016revised.pdf PDF
589-B82017___VISN_15_Eastern_Kansas_VAMC_Topeka_Campus,_589_B82017revised.xlsx XLSX spreadsheet
SEP_FY18_Consolidation_List_Amend_0004.xlsx XLSX spreadsheet
AMENDMENT_0003_-_Continuation_page.pdf PDF
Sept_18_Consolidation_Schedule_-_Revised_12122018.pdf PDF
36A79719Q0002_0003.docx DOCX document
537-B82059_-_XR_PORT_RAD,_VAMC_CHICAGO,_IL_Redacted.pdf PDF
549-B80428_-_XR_US,_VAMC_DALLAS,_TX_redacted.pdf PDF
36A79719Q0002_0002.docx DOCX document
523-B81008_B81009_B81010___VISN1_VA_Boston_Healthcare_System-_West_Roxbury_Campus,_523-B81008B81009B81010revisedafterquestions.xlsx XLSX spreadsheet
668-B80015revised_V2.pdf PDF
608-B83003revisedafterquestions.pdf PDF
521-B72027_-_XR_RF,_VAMC_BIRMINGHAM,_AL_redacted.pdf PDF
523-B81008B81009B81010revisedafterquestions.pdf PDF
689-B81009___VISN_1_VA_Connecticut_Healthcare_System,_689-B81009revisedafterquestions.xlsx XLSX spreadsheet
608-B83003___Manchester_VA_Medical_Center,_608-B83003revisedafterquestions.xlsx XLSX spreadsheet
AMENDMENT_0002_-_Continuation_page.pdf PDF
612-B86025_-_XR_US_VAMC_MARTINEZ_CA_redacted.pdf PDF
537-B82069_-_XR_US,_VAMC_CHICAGO,_IL_redacted.pdf PDF
689-B81009revisedafterquestions.pdf PDF
636-B84043_-_Specifications_Radiology_Ultrasound_-_Iowa_City_-_REVISED.pdf PDF
SEP_FY18_Consolidation_List_Amend_0002.xlsx XLSX spreadsheet
36A79719Q0002-0001014.pdf PDF
36A79719Q0002-0001015.pdf PDF
36A79719Q0002-0001035.zip ZIP file
36A79719Q0002-0001013.pdf PDF
36A79719Q0002-0001012.docx DOCX document
36A79719Q0002-0001020.zip ZIP file
36A79719Q0002-0001017.pdf PDF
36A79719Q0002-0001021.zip ZIP file
36A79719Q0002-0001016.pdf PDF
36A79719Q0002-0001019.zip ZIP file
36A79719Q0002-0001026.zip ZIP file
36A79719Q0002-0001031.zip ZIP file
36A79719Q0002-0001022.zip ZIP file
36A79719Q0002-0001036.pdf PDF
36A79719Q0002-0001032.zip ZIP file
36A79719Q0002-0001023.zip ZIP file
36A79719Q0002-0001018.zip ZIP file
36A79719Q0002-0001028.zip ZIP file
36A79719Q0002-0001037.xlsx XLSX spreadsheet
36A79719Q0002-0001033.zip ZIP file
36A79719Q0002-0001024.zip ZIP file
36A79719Q0002-0001025.zip ZIP file
36A79719Q0002-0001029.zip ZIP file
36A79719Q0002-0001027.zip ZIP file
36A79719Q0002-0001034.pdf PDF
Show 40 more

Latest 50 of 141 listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Compliance Matrix

Digital Radiography and Fluoroscopy
[20/Mann-Grandstaff VA]
[668-B80015]
OFFEROR
MANUFACTURER
MODEL
QUOTE NUMBER

VENDOR INSTRUCTIONS:

Please fill out the table above with the offeror name (if different than the device manufacturer), manufacturer name, model name/number, and the quote number. All required specifications are listed below. Please review the required specifications and provide supporting information about the proposed model. The "Vendor Response" column is limited to the following entries: "exceeds", "meets," "partially meets," and "does not meet." In the "Documentation" column, please indicate the location within the attached documentation (file name, page number) for information to support the vendor response. Please provide any additional information/explanations in the "Vendor Explanation" column.

REQUIREMENT OVERVIEWVENDOR RESPONSEVENDOR DOCUMENTATIONVENDOR EXPLANATION
The requirement for this procurement is for a single general-purpose R/F room for the Mann-Grandstaff VA (668) in Spokane, WA. The intended applications include but are not limited to general radiography; upper GI; lower GI; esophagrams; barium swallows; arthrograms; joint injections; nephrostomy tubes; colostomy tubes; port placement; and weight bearing knees, feet, hips.
Facility
Quantity
Mann-Grandstaff VA (668)
1
TECHNICAL REQUIREMENTSVENDOR RESPONSEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. Unit physical specifications
a. Minimum wall detector size [in x in]
14’’ x 17’’
b. Minimum table detector size [in x in]
17’’ x 17’’
c. Minimum detector resolution [lp/mm]
3.4
d. Maximum pixel size [µm]
148
e. Height range for wall bucky [in]
26”-68.1”
f. Maximum patient weight [lbs]
Table 660 lbs
g. Minimum table width [cm]
80
h. Minimum table tilt [deg]
+90°/-45°
i. Minimum generator power [kW]
80
j. Minimum generator phases
three
k. Radiographic kVp range [kVp]
40 kV to 150 kV
l. Radiographic mA range [mA @ kVp]
530 mA at 150 kV
m. Fluoroscopy kVp range [kVp]
110 kV/23 mA
n. Minimum pulse rate [frames/sec]
30,15,10,7.5, or 3 p/s 0.5,1,2,4 or 8 f/s
o. Minimum acquisition matrix size
12-bit Fluoro/14 bit Digital radiology
p. Minimum acquisition bit depth
16 bits
q. Minimum spatial resolution [lp/mm]
DQE 66% at 0.05 lp/mm to 17% at Nquist
r. Minimum SID range [in]
45.3
s. Heat unit x-ray tubes [kHU]
Anode heat storage 783,000 HU
t. Minimum heat dissipation rate [kHU/min]
162,000 HU/min.
u. Minimum control room monitor size [in]
19”
v. Minimum in-room monitor size [in]
19”
w. Minimum control room computer hard drive memory [GB]
4 GB
x. Minimum imaging tower expanded tabletop clearance[in]
45” to 59”
y. Minimum Imaging Tower field of view [in]
Fullformat 42cmX42.6cm, Zoom1 30cmX30cm, Zoom2 22cmX22cm, Zoom 3 15cmX15cm
z. Maximum system dimensions [cm]
210 cm to 250 cm
aa. Maximum system weight [kg]
1320 kg system
bb. Other physical specifications
Longitudinal table travel 62.9”
�
2. Additional specifications
Generator
☒ a. High-frequency generator with automatic dose rate control
☒ b. Continuous 30 p/s and pulsed fluoroscopy ma modes
Fluoroscopic Tube
☒ c. Configuration
● Overhead tube
☒ d. Pulsed/grid fluoroscopic tube
☒ e. Selectable filtration
☒ f. Dual focal spots (please specify) 0.6 and 1.0
☒ g. Collimation – fully adjustable manual and automatic exposure control with visible light field display.
☒ h. Radiation free collimation
Fluoroscopic spot device/Imaging tower
☒ i. Flat panel detector technology
☒ j. Automated image capture and save, to include last image hold
☒ k. Variable speed power assist controlled in all directions
☒ l. Footswitch and tower controls, to provide the following:

Control of both fluoroscopy and spot shots Automatic shut-off when footswitch or tower control is released

☒ m. Removable lead curtains with counter weights for when curtains are not in use
☒ n. Tabletop travel controls
☒ o. Collimation control
Radiographic/overhead tube crane system
☒ p. Patient alignment system (laser alignment/positioning lights)
☒ q. Table and upright auto tracking package
☒ r. Pre-programmed exposure settings located on tube or in control room
☒ s. Ability to change between table top, upright bucky, and table bucky from tube head
☒ t. Ability to lower overhead tube to touch the floor when used with the wall stand
☒ u. Automatic tube crane, protocol-based movement
☒ v. In room protocol adjustment via tube head or other in-room mounted system
Control room fluoroscopic/radiographic control panel
☒ w. Auto HIS/RIS refresh package
☒ x. Quality control tracking package
☒ y. Bar code reader to scan additional detectors (sharing)
☒ z. UPS for x-ray control/image memory
☒ aa. Ability to send images directly from the control panel (no separate workstation required)
☒ bb. Ability to capture live video
☒ cc. DVD recorder-option
Wall stand/fixed wall detector
☒ dd. Ability to lower center of imaging plate to touch the floor
☒ ee. Ability to tilt the imaging surface at least 90◦ and -20°
☒ ff. Automated movement for long axis stitching
☒ gg. Auto tracking of the tube and detector during vertical adjustment at the wall stand and/or tube.
☒ hh. Tilt-able detector holder with inherent grid (-20 degrees to 90 degrees)
☒ ii. Height minimum – low enough to complete standing knee exams while patient is standing on the floor
☒ jj. Height maximum – high enough to complete AP C-spine exams while patient is standing on the floor
In-room monitor
☒ kk. LCD in-room monitor
● Ceiling-mounted
☒ ll. In-room remote control to orientate
☒ mm. High contrast black and white
☒ nn. Anti-glare display
Removable table detector requirements
☒ oo. Wireless detector (not tethered to the system/table)14”x17”
☒ pp. Detector charger in bucky/holder
☒ qq. Additional battery
☒ rr. Charging station (if required)
Table requirements
☒ ss. Ability to tilt 90 degrees in both directions/full tilt option
☒ tt. Full articulation
☒ uu. Auto-centering option to exact middle while the table is in horizontal or vertical position
☒ vv. Table movement controls
● Trolley
☒ ww. Removable/adjustable patient handgrips
☒ xx. Removable/adjustable footrest
☒ yy. Removable/adjustable shoulder rests
☒ zz. Removable/adjustable stirrups- option
☒ aaa. Bariatric table
☒ bbb. Remote or manual table top movements
☒ ccc. Motor driven table top movement in the X and Y axis at stated maximum weight capacity.
☒ ddd. Auto tracking of bucky / tube for longitudinal travel
☒ eee. Ability to lower tabletop height to at least 18.9” from the floor in the horizontal position at stated maximum weight capacity.
☒ fff. Selectable automatic stop in horizontal position 0 °
☒ ggg. Longitudinal table and tube travel
3. Software Requirements
☒ a. Monitoring of Dosage- The system must be able to supply a technology that allows for monitoring and tracking of radiation dose provided to a patient; DICOM structured reporting is preferred. DAP accuracy must be 99% reproducible with <2% deviation.
☒ b. Bone suppression
☒ c. Repeat rate – ability to track repeat/retake data to include such items as technologist (required unique identifier), reason for repeat, patient dose, exam type, etc. The data should be exportable to Excel or other databases for tracking, trending, and combining with data from other imaging sources within the facility.
☒ d. Rapid image display, < 5 seconds preferred.
☒ e. Stitching – ability to stitch multiple long axis images with options for automated and manual process. Using tube tilt with the upright wall bucky and the table.
☒ f. Ability to apply multiple image processing algorithms both pre and post-acquisition to allow for soft tissue and/or bone enhancement.
☒ g. Dose Reduction- The system must have systems in place to facilitate regular protocol optimization and reduced radiation dose to the patient.
☒ h. Fluoroscopy Loop – Ability to record and store dynamic fluoroscopy sequences after the fluoroscopy has been completed. Storage capacity may be variable according to pulse per second settings (please specify). Please include storage and display options for recorded data review and archive.
☒ i. Automatic movement to position as defined in the programs by pressing a single button.
☒ j. Dose free collimation with the last image hold (side to side and top to bottom of the image)
4. Security/Connectivity Requirements
☒ a. OEM-supported operating system
☒ b. DICOM 3.0 print, store, commit, and modality worklist
☒ c. HL7 integration (HIS/RIS)
☒ d. Wireless connectivity to VA network – compatible with 802.11b/g/n and FIPS 140-2 compliant
☒ e. Encrypted hard drive
☒ f. PACS compatibility – [Agfa, EPIC, Carestream and the majority of all PACS]
☒ g. (NIST)FIPS document
5. Added Value
Specifications listed below are not required, but preferred. Vendors who do not include the below specifications in the submitted offer will not be docked or excluded from consideration. Specifications listed below will be evaluated based on added value.
☒ a. Oblique projections -45o/+45o
☒ b. Version/platform long-range plan
☒ c. Post-warranty remote diagnostic service program
☒ d. Additional year(s) of warranty
TRAINING REQUIREMENTSVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
Description
No. of Personnel
1. On-site clinical applications training for technologists during go-live
6+
2. Off-site clinical applications training for technologists (to include tuition)
n/a
3. On-site follow-up clinical applications training after technologists have hands-on experience with the system
6+
4. Biomedical technician training package (to include tuition)
2
Biomedical technician training shall include any prerequisites required prior to the training and shall be equivalent to the training received by OEM field service representatives. Technicians shall be given all service manuals, schematics, diagrams, diagnostic software, other special tools, and keys equivalent to what OEM field service representatives have available to diagnose, troubleshoot, repair, and maintain the equipment.
Technologists who complete the clinical applications training shall receive continuing education credits (CMEs).
SERVICE REQUIREMENTSVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. VPN/Remote Access – The vendor shall provide any and all equipment service programs, such as remote diagnostics, during the warranty period. The system shall provide vendor remote diagnostics via VPN. The vendor shall either utilize the VA national site-to-site VPN or work with the Office of Cyber and Information Security and the VAMC Information Systems Security Officer to establish a client-based VPN.
2. Service and Operator Manuals – The vendor shall provide the following documentation for the proposed systems:
a. Two (2) copies of operator instruction manuals (one (1) electronic and one (1) physical copy)
b. Two (2) copies a service manual (one (1) electronic and one (1) physical copy).
3. Minimum Warranty – The system and accessories shall be covered under the manufacturer’s warranty and shall include all parts and labor for one year following acceptance by the VAMC. This warranty must include PMs as required by the manufacturer. The manufacturer’s factory-trained field service representatives shall perform installation and maintenance during the warranty period.
Vendors are encouraged to include any offerings for service, warranty, and training that may exceed the requirements with their proposals. Vendors who do not include any added value offerings for service, warranty, and training will not be docked or excluded from consideration. However, any such offerings will be evaluated based on added value.
OTHER INFORMATION/DOCUMENTATION REQUESTEDVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. Product brochures
2. Technical specification sheets, to include dimensions and weight of the system
3. DICOM Conformance Statement
4. IHE integration statement
5. FIPS 140-2 certification
6. Completed pre-procurement assessment form (6550)
7. Completed MDS2 form
8. Detailed information about the curriculum and length of the biomedical technical training
9. Details on any off-site training offered for technologists
10. Information about your company’s support structure during the warranty period
a. Describe on-line or telephonic applications support and availability (include third party coverage)
b. Provide a listing of field service engineer locations and availability
c. Provide a listing of part depots
11. Information about your company’s support options following the warranty period, including a description of on-line or telephonic applications support and availability
TRADE-INVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
The VA will retain any hard drives containing electronic personal health information (ePHI). The following equipment is available for trade-in. Please reflect any credits provided for trade-in equipment in the proposal.
Station
668
Manufacturer
GE
Model
Precision RXI
EE/Asset Number
24643
Serial Number