HC/HT Specifications
Radiography-Fluoroscopy Unit
TABLE CHARACTERISITCS
➢ Table must be able to lower to at least 30”
➢ Table must withstand at least 550 pounds when stationary
➢ Tilting range from +/- 90 degrees and able to withstand 500 pounds when tilted
GENERAL CHARACTERISTICS
➢ Bi-directional HL7 interface with VistA
➢ Digital storage hard drive size 320-500 GB minimum ideally 1 TB as max
➢ Exam presets/protocols - customizable
➢ VA VISTA Imaging compatible
➢ Interface with Carestream cPACS
➢ Windows 7 or proprietary operating system
➢ Ability to have individual user accounts, system administrator
➢ Ability to disable USB ports
➢ TIMS unit compatible
➢ FDA 510(k) clearance
➢ Stitching capability, i.e. orthopedic and scoliosis
➢ Dual-energy capable for chest x-rays
GENERAL RADIOGRAPHY
➢ Fixed wall buckey detector
➢ Wireless detectors that be IEEE802.1 compliant:
o 10”x12” o At least 14”x17”
➢ Wall buckey at least 15”x15”
➢ Wall buckey must be able to tilt 90 degrees
➢ Ability to burn images at acquisition station
➢ Digital detector(s) weighing no more than 7 pounds
➢ Automatic tracking of tube when height is adjusted during exams
➢ Have protective cover over detector able to withstand 500 pounds
FLUOROSCOPY
➢ At least 80kW generator
➢ Ability to digitally produce dose report digitally per exam and push to CPRS
TRAINING
Clinical training:
➢ 40 hours onsite basic clinical system training
Technical Training:
➢ Technical Biomedical Engineering Training – 2 reserved o Quote to include all prerequisite courses necessary for modality service training
➢ Technical Biomedical Engineering Training Travel Package (Lodging/Meals/Transportation)
WARRANTY
➢ 2 year warranty
SHIP TO: WAREHOUSE B81006
V.A. Medical Center
400 VETERANS AVENUE
BUILDING 26
BILOXI, MS 39531- 2410
DILVERY LOCATION WILL BE:
PO#: 520-B81006
Pensacola, Florida
Support and other Documentation to Provide:
➢ Provide DICOM Conformance Statement
➢ Provide approved FIPS 140-2 compliance document if detectors are wireless
➢ Provide current nationally approved memorandum of understanding/interagency security agreement
(MOU/ISA)
➢ Provide completed Pre-procurement Assessment form (6550)
➢ Provide information about your company’s applications and technical support structure during the warranty period (i.e. a listing of Field Service Engineer locations and availability, support 800 phone number(s), remote support, etc.)