589-B80016___VISN_15_KCVA_Medical_Center,_589-B80016revised.xlsx

XLSX spreadsheet 19 KB Posted

Attached to
SEPTEMBER 2018 CONSOLIDATION
Solicitation number
36A79719Q0002
Issued by

About this file

36A79719Q0002 0004 589-B80016 _ VISN 15 KCVA Medical Center, 589-B80016revised.xlsx

View the file

Other files

File Type Posted
589_B82017revised.pdf PDF
36A79719Q0002_0004.docx DOCX document
589-B80016revised.pdf PDF
589-B82017___VISN_15_Eastern_Kansas_VAMC_Topeka_Campus,_589_B82017revised.xlsx XLSX spreadsheet
SEP_FY18_Consolidation_List_Amend_0004.xlsx XLSX spreadsheet
AMENDMENT_0003_-_Continuation_page.pdf PDF
Sept_18_Consolidation_Schedule_-_Revised_12122018.pdf PDF
36A79719Q0002_0003.docx DOCX document
537-B82059_-_XR_PORT_RAD,_VAMC_CHICAGO,_IL_Redacted.pdf PDF
549-B80428_-_XR_US,_VAMC_DALLAS,_TX_redacted.pdf PDF
36A79719Q0002_0002.docx DOCX document
523-B81008_B81009_B81010___VISN1_VA_Boston_Healthcare_System-_West_Roxbury_Campus,_523-B81008B81009B81010revisedafterquestions.xlsx XLSX spreadsheet
668-B80015revised_V2.pdf PDF
608-B83003revisedafterquestions.pdf PDF
521-B72027_-_XR_RF,_VAMC_BIRMINGHAM,_AL_redacted.pdf PDF
523-B81008B81009B81010revisedafterquestions.pdf PDF
689-B81009___VISN_1_VA_Connecticut_Healthcare_System,_689-B81009revisedafterquestions.xlsx XLSX spreadsheet
608-B83003___Manchester_VA_Medical_Center,_608-B83003revisedafterquestions.xlsx XLSX spreadsheet
AMENDMENT_0002_-_Continuation_page.pdf PDF
612-B86025_-_XR_US_VAMC_MARTINEZ_CA_redacted.pdf PDF
537-B82069_-_XR_US,_VAMC_CHICAGO,_IL_redacted.pdf PDF
689-B81009revisedafterquestions.pdf PDF
636-B84043_-_Specifications_Radiology_Ultrasound_-_Iowa_City_-_REVISED.pdf PDF
668-B80015___20_Mann-Grandstaff_VA_,_668-B80015revised_V2.xlsx XLSX spreadsheet
SEP_FY18_Consolidation_List_Amend_0002.xlsx XLSX spreadsheet
36A79719Q0002-0001014.pdf PDF
36A79719Q0002-0001015.pdf PDF
36A79719Q0002-0001035.zip ZIP file
36A79719Q0002-0001013.pdf PDF
36A79719Q0002-0001012.docx DOCX document
36A79719Q0002-0001020.zip ZIP file
36A79719Q0002-0001017.pdf PDF
36A79719Q0002-0001021.zip ZIP file
36A79719Q0002-0001016.pdf PDF
36A79719Q0002-0001019.zip ZIP file
36A79719Q0002-0001026.zip ZIP file
36A79719Q0002-0001031.zip ZIP file
36A79719Q0002-0001022.zip ZIP file
36A79719Q0002-0001036.pdf PDF
36A79719Q0002-0001032.zip ZIP file
36A79719Q0002-0001023.zip ZIP file
36A79719Q0002-0001018.zip ZIP file
36A79719Q0002-0001028.zip ZIP file
36A79719Q0002-0001037.xlsx XLSX spreadsheet
36A79719Q0002-0001033.zip ZIP file
36A79719Q0002-0001024.zip ZIP file
36A79719Q0002-0001025.zip ZIP file
36A79719Q0002-0001029.zip ZIP file
36A79719Q0002-0001027.zip ZIP file
36A79719Q0002-0001034.pdf PDF
Show 40 more

Latest 50 of 141 listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Compliance Matrix

Computed Tomography Scanner
VISN 15/KCVA Medical Center
589-B80016
OFFEROR
MANUFACTURER
MODEL
QUOTE NUMBER

VENDOR INSTRUCTIONS:

Please fill out the table above with the offeror name (if different than the device manufacturer), manufacturer name, model name/number, and the quote number. All required specifications are listed below. Please review the required specifications and provide supporting information about the proposed model. The "Vendor Response" column is limited to the following entries: "exceeds", "meets," "partially meets," and "does not meet." In the "Documentation" column, please indicate the location within the attached documentation (file name, page number) for information to support the vendor response. Please provide any additional information/explanations in the "Vendor Explanation" column.

REQUIREMENT OVERVIEWVENDOR RESPONSEVENDOR DOCUMENTATIONVENDOR EXPLANATION
The KCVA Radiology Department is seeking to procure a Computed Tomography X-ray system. It is intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, Spectral, and Gated acquisitions for all patient ages. The system is indicated for head, whole body, cardiac and vascular x-ray Computed Tomography applications. The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
TECHNICAL REQUIREMENTSVENDOR RESPONSEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. Unit physical specifications
a. Minimum number of slices
320
b. Minimum gantry aperture [cm]
50
c. Minimum detector width [mm]
0.05
d. Minimum detector field of view [cm]
50
e. Table height range from floor [cm]
45-90
f. Maximum patient weight supported [kg]
300
g. Minimum 360° rotation time [s] – Vendors are encouraged to propose the fastest 360° rotation available.
0.4
h. Maximum system dimensions [cm]
Fit properly into room dimensions 8’ height, 20’ length, 14’ width
i. Maximum system weight [kg]
2500
2. Additional specifications
☒ a. The CT scanner shall provide the ability for the user to independently operate the table, gantry, and CT fluoroscopy mode from the control room.
☒ b. The CT scanner shall provide the ability for the user to independently operate the table, gantry, and CT fluoroscopy mode while performing biopsy or patient-related exams.
☒ c. Ability to display CTDlvol (dose) before scan
☒ d. Ability to display DLP (dose) after exposure
☒ e. Modeled iterative reconstruction/Noise reduction. Vendors are encouraged to propose the greatest available noise reduction.
☒ f. Longest z-axis acquisition per rotation and, while achieving that – maximum temporal resolution, maximum spatial resolution in mm and in line pair equivalents, maximum construction rate (frames/sec) with maximum available noise reduction applied, and smallest focal spot size for largest z-axis coverage
☒ g. Triggered studies using an ROI gate place in the appropriate vessel for CTA and CCTA
☒ h. Gantry- or ceiling-mounted contrast injector with proven reliable synchronicity.
☒ i. Full UPS
☒ j. Power conditioning (as defined by vendor)
☒ k. Ability to lock protocols
☒ l. Ability to facilitate regular protocol optimization and reduced radiation dose to the patient, including but not limited to iterative reconstruction technology
3. Workstation Requirements
☒ a. Primary operating/scanning workstation with a UPS
☒ b. Secondary reconstruction workstation with a UPS
4. Advanced Applications
☒ a. CT Perfusion
☒ b. Lung CAD
☒ c. Dose Tracking
☒ d. Integration to Advanced Visualization System – VITREA
Each vendor is to respond with advanced applications that meet the criteria listed above. Please include all other advanced applications offered by your company in the optional section on the quotes.
5. Security/Connectivity requirements
☒ a. OEM-supported operating system
☒ b. DICOM 3.0 print, store, commit, radiation dose structure report and modality worklist
☒ c. HL7 integration (HIS/RIS)
☒ d. Encrypted hard drive
☒ e. PACS compatibility – Philips
6. Added Value
Specifications listed below are not required, but preferred. Vendors who do not include the below specifications in the submitted offer will not be docked or excluded from consideration. Specifications listed below will be evaluated based on added value.
☒ a. Gantry tilt
☒ b. Ceiling-mounted contrast injector
☒ c. Version/platform long-range plan
☒ d. Post-warranty remote diagnostic service program
☒ e. Additional year(s) of warranty
TRAINING REQUIREMENTSVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
Description
No. of Personnel
1. On-site clinical applications training during go-live for technologists, to cover each of the software packages
4
2. On-site follow-up clinical applications training, to be provided after technologists/physicians have hands-on experience with the system
4
3. Off-site clinical applications training for technologists (to include tuition)
4
4. On-site clinical applications training for physicians during go-live
4
5. Biomedical technician training package (to include tuition)
3
Biomedical technician training shall include any prerequisites required prior to the training and shall be equivalent to the training received by OEM field service representatives. Technicians shall be given all service manuals, schematics, diagrams, diagnostic software, other special tools, and keys equivalent to what OEM field service representatives have available to diagnose, troubleshoot, repair, and maintain the equipment.
Technologists who complete the clinical applications training shall receive continuing education credits (CMEs).
Off-site training will not be purchased at the time of award. Vendors must demonstrate that they can provide any required off-site training listed above, therefore off-site training should be quoted as an optional item. Travel for VA employees is not authorized under the HTME contracts. In no case should any training include expenses for travel or travel for VA personnel at no cost.
SERVICE REQUIREMENTSVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. VPN/Remote Access – The vendor shall provide any and all equipment service programs, such as remote diagnostics, during the warranty period. The system shall provide vendor remote diagnostics via VPN. The vendor shall either utilize the VA national site-to-site VPN or work with the Office of Cyber and Information Security and the VAMC Information Systems Security Officer to establish a client-based VPN.
2. Service and Operator Manuals – The vendor shall provide the following documentation for the proposed systems:
a. Two (2) copies of operator instruction manuals (one (1) electronic and one (1) physical copy)
b. Two (2) copies a service manual (one (1) electronic and one (1) physical copy).
3. Minimum Warranty – The system and accessories shall be covered under the manufacturer’s warranty and shall include all parts and labor for one year following acceptance by the VAMC. This warranty must include PMs as required by the manufacturer. The manufacturer’s factory-trained field service representatives shall perform installation and maintenance during the warranty period.
Vendors are encouraged to include any offerings for service, warranty, and training that may exceed the requirements with their proposals. Vendors who do not include any added value offerings for service, warranty, and training will not be docked or excluded from consideration. However, any such offerings will be evaluated based on added value.
OTHER INFORMATION/DOCUMENTATION REQUESTEDVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. Product brochures
2. Technical specification sheets, to include dimensions and weight of the system
3. Longest z-axis acquisition per rotation, and while achieving that, the
a. Maximum temporal resolution,
b. Maximum spatial resolution in mm and in line pair equivalents,
c. Maximum reconstruction rate (frames/sec) with maximum available noise reduction applied, and
d. Smallest focal spot size
4. Information on the functionality each workstation provides (e.g. same functionality, if they operate independently)
5. Method used for cooling (i.e. water-cooled vs air cooled)
a. If water-cooled, specify the method of connection (e.g. simple access to cold water vs intercooler)
b. If air-cooled, specify the sound levels and BTU production into the room.
6. DICOM Conformance Statement
7. IHE integration statement
8. Completed pre-procurement assessment form (6550)
9. Completed MDS2 form
10. Detailed information about the curriculum and length of the biomedical technical training
11. Details on any off-site training offered for technologists
12. Information about your company’s support structure during the warranty period.
a. Describe on-line or telephonic applications support and availability.
b. Provide a listing of field service engineer locations and availability
c. Provide a listing of part depots
13. Information about your company’s support options following the warranty period, including a description of on-line or telephonic applications support and availability
14. Information on any FDA safety recalls associated with the proposed equipment
TRADE-INVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
☒ a. In instances where sanitization of ePHI compromises the OS and/or application software, or requires the removal of internal storage media, the vendor accepts the equipment “as is” and can elect at their own discretion to contract with the original equipment manufacturer (OEM) to restore the system.
The following equipment is available for trade-in. Please reflect any credits provided for trade-in equipment in the proposal.
Station
589
Manufacturer
GE
Model
VCT
EE/Asset Number
98223
Serial Number
6525-439454