J065--DC Ceiling Lift Replacement RFQ Ammendment 5

Closed Solicitation Posted

This opportunity was awarded. See the award notice from .

Solicitation number
36C24526Q0269
Agency
Veterans Integrated Service Network 5 Veterans Health Administration, Department of Veterans Affairs
Responses due
Set-aside
No set-aside

Opportunity facts

NAICS code
339112 Surgical and Medical Instrument Manufacturing
PSC
J065 Maint/Repair/Rebuild Of Equipment- Medical, Dental, And Veterinary Equipment And Supplies
Place of performance
Department of Veterans Affairs washington Dc Vamc Washington, District of Columbia 20422, United States
Points of contact

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 5 seeks a complete fleet replacement solution for all ceiling lifts at Washington DC VAMC, with proposals due June 26, 2026, and a mandatory site visit on May 4, 2026 at 10:00 AM. This is a firm fixed-price RFQ with no set-aside.

Place of performance is 50 Irving Street NW, Washington, DC 20422. The procurement requires supply and installation of 276 standard ceiling lift motors (600+ lb capacity), 6 bariatric ceiling lift motors (1,000+ lb capacity), associated track systems up to 18 x 18 feet, hardware, trolleys, rails, and accessories across approximately 282 installation locations. Seven rooms require new XY track systems with specified dimensions ranging from 8 feet 9 inches by 6 inches to 18 feet by 18 feet. The vendor must provide comprehensive labor, deinstallation, installation, structural assessment and certification, electrical work, ICRA (Infection Control Risk Assessment) and HEPA containment per ICRA/PCRA team requirements, HVAC and lighting modifications as needed, Ground Penetrating Radar scanning for embedded conduits, seismic bracing, comprehensive drawings, training for three shifts (minimum five business days clinical training, one to three days biomedical/technical training), and one year of preventative maintenance and repair service. All ceiling lifts and rail systems must be fully operational within one year of contract award under at least three-year warranty. ICRA carts and HEPA containment are required in all adaptable rooms regardless of above-ceiling access; above-ceiling tiles will likely be Level 2 containment, with new installations Level 3 or 4 based on work requirements. The NAICS code is 339112 (Surgical and Medical Instrument Manufacturing), and the PSC is J065 (Medical Equipment). Vendors must submit all technical and implementation documents, 2025 manufacturer-signed Letters of Authorization, certificates, past performance references from the past two years, and detailed pricing in Excel and PDF formats; non-compliant submissions will be rejected. Evaluation weighs technical compliance with SOW requirements, past performance, implementation capability, and pricing.

Notice text

5 versions

Update #5 · Latest ·

The purpose of modification number 5 is to answer question.

Are ICRA carts and/or HEPA containment required in all adaptable rooms if no above-ceiling access is necessary?

Yes, per ICRA/PCRA team above ceiling tiles will likely always be level 2 and for the new installations/others can be 3 or 4 based on work requirements.

Update #4 ·

The purpose of amendment 4 is to extend RFQ close date.

RFQ close date is changed from 6/19/26 @12PM EST to 6/26/26 @12PM EST

Attached RFQ Amendment 4

Update #3 ·

The purpose of this amendment is to respond to the additional requests for information and to attach drawings..

Update #2 ·

The purpose of amendment 1 is to post site visit attendee list, answer pre site visit questions, and extend RFQ close date.

Question & Answers- Pre site visit questions

Can you provide a list of rooms where existing tracks are in acceptable condition and eligible for retrofit?

Vendor to provide a solution for up to 282 ceiling motors retrofit is determined based on the equipment/implementation/solution vendor provides as long as vendor can support motor and rail service at least one year /equipment/rails under warranty at least 3 years upon installation completion. Â Refer to rail picture and lift xls attachment in RFQ.

For adaptable rooms, can you provide the charging method for each room to ensure compatibility with continuous charging requirements?

Current charging system for each is continuous charging for the three manufacturers. Vendor must provide a ceiling lift/rail motor solution that is capable to be charged upon installation completion.

For rooms where tracks are not adaptable, should vendors assume full replacement with an 18' x 18' XY system? If so, can you confirm that this configuration fits all applicable rooms?

For HTM we only aim for functionality and would prefer standardization of rails. Patient rooms are various sizes. Adaptability is to be addressed in vendor solution. Scaled drawings to follow. Refer to rail picture attachment in RFQ. Scaled room drawings to follow.

For the seven rooms requiring new track systems, can you provide detailed reflected ceiling plans to confirm layout and installation requirements?

Here are the 3 areas:

2A-104: 8 x9 6 XY

2B-130: 10' x 10' XY

2J-153: 18'x18' XY

2K-108: 8 2 X 13 6 XY

2K-213A: 12'x10' XY

2K-213B: 8'x10' XY

GI Bay Area (x2) TBD

Shared space Floor/Area 3A TBD

Are slings or harnesses required as part of the proposal? If so, please specify type and quantity per room (e.g., reusable seated sling).

Refer to SOW solicitation:

xlix & lx Slings and Accessories:

Compatibility with various reusable and disposable slings. If specific models are required to function, please provide part numbers/prices for disposable and reusable slings and accessories.

How many days of clinical training are required?

Clinical has 3 shifts of patients between multiple units. To accommodate this how is training provided?

In person

Online

How many days are offered for the standard?

Ideally a week (5 business days) if not more to allow shift changes for all teams. Vendor RFQ solution to address training plan.

How many days of biomedical or technical training are required?

Biomed- 1 day if software/computerized steps are not involved and it is only mechanical/physical. If software/computerized steps training should be 2-3 days in this case.

Is an extended warranty on lift motors (e.g., 3 years) required or preferred?

At least 3 years per SOW; vendor provide RFQ solution to address warranty.

II. Site visit completed 5/4/26

III. RFQ close date extended:

RFQ close date is extended from 5/28/26 12PM EST to 6/16/26 12PM EST.

Update #1 ·

The purpose of this requirement is to provide a Fleet Replacement Acquisition solution for all Ceiling Lifts at Washington DC VAMC that does not void warranties of new lift motors or track system. This Vendor shall provide all labor, personnel, equipment 276 standard ceiling lift motors, 6 bariatric ceiling lift motors, hardware, caps, trolley, parts, rails, accessories, and track systems up to 18 x18 , tools, materials, training, to identify and move conflicting infrastructure sprinkle movement, specifications, identify structural deficiencies, comprehensive drawings, electrical, ICRA, Hepa cart, HVAC supply return, lighting, deinstallation, installation, one year of service, at least 3 year warranty, supervision and other items and implementation, services necessary for a fully installed and fully operational patient ceiling lift system within one year of contract award.

Missing one or more technical/implementation documents, required responses to SOW,

incomplete certificates, 2025 manufacturer signed LOA, to include incomplete vendor

RFQ technical/past performance, implementation, pricing XLS/PDF responses will be

considered non-compliant/non-conforming. Government determines acceptability

See attached

RFQ

Price template Xls

lift inv-room list Xls

Appendix B

Existing rail picture

Checklist (2): Appendix A, Appendix C

Site Visit 5/4/26 @1000AM- see solicitation for instructions

Attachments

Files attached to this notice, newest first
File Type Posted
36C24526Q0269_4.docx DOCX document
36C24526Q0269 AMD 4 RFQ.pdf PDF
36C24526Q0269_3.docx DOCX document
36C24526Q0269 0001.pdf PDF
Drawings.zip ZIP file
lift inv- AM1.xlsx XLSX spreadsheet
36C24526Q0269 AM1.pdf PDF
36C24526Q0269_2.docx DOCX document
AL14-07 Checklist APPENDIX A.pdf PDF
AL14-07PSA Checklist APPENDIX C.pdf PDF
CEILING PRICE LIST .xlsx XLSX spreadsheet
lift inv-.xlsx XLSX spreadsheet
RAIL PICTURE.jpg JPG image
AL17-07 CEILING APPENDIX B.pdf PDF
36C24526Q0269 RFQ.pdf PDF
36C24526Q0269_1.docx DOCX document
Show all 16

Notice history

Notices posted for this opportunity, newest first
Notice Type Posted
DC ceiling lift Award Award Notice
J065--DC Ceiling Lift Replacement RFQ Ammendment 5 This notice · Latest solicitation Solicitation

On GovTribe

Work this opportunity on GovTribe

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity