The Department of Veterans Affairs, Veterans Integrated Service Network 15, is issuing a sources sought/request for information for dental imaging equipment to be installed at the Rolla Community Based Outpatient Clinic in Rolla, Missouri, with responses due by July 10, 2026. This is a pre-solicitation notice only and does not constitute a binding solicitation; however, the VA encourages submission of pricing, capabilities, and market information for planning purposes in accordance with FAR Part 15.201(e).
The VA seeks one dental imaging software license expansion; one cone beam computed tomography (CBCT) 3D X-ray unit capable of 2D panoramic, 3D CBCT, and optional cephalometric imaging; seven wall-mounted dental radiographic units with 7 mA tube current and extended arm reach of at least 81 inches; and seven dental imaging sensor kits in pediatric (Size 1) and adult posterior (Size 2) configurations. The CBCT system must deliver high-resolution voxels of 75–150 microns or better, field-of-view options from approximately 5x5 cm to 16x17 cm, metal artifact reduction technology, and DICOM 3.0 compliance with VA PACS integration capability. Wall-mounted radiographic units require high-frequency DC X-ray generators at 60–70 kVp with rectangular collimation and FDA/NCRP radiation safety compliance. Intraoral sensors must use CMOS technology with minimum 20 line pairs/mm resolution and TWAIN and DICOM compliance with USB connectivity.
Responses should include company information, business size classifications, manufacturer or distributor status, GSA schedule or federal contract vehicle availability, manufacturing location, and UEI number. NAICS code 339114 applies with a 750-employee size standard. Contact Tracie Raggs at tracie.raggs@va.gov.
Notice text
THIS REQUEST FOR INFORMATION (RFI)/SOURCES SOUGHT
DISCLAIMER
This RFI is issued solely for information and planning purposes only and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. Responses to this notice are not offers and cannot be accepted by the Government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.
The submission of pricing, capabilities for planning purposes, and other market information is highly encouraged and allowed under this RFI In Accordance With (IAW) FAR Part 15.201(e).
SOURCES SOUGHT DESCRIPTION
This is NOT a solicitation announcement. This is a RFI/sources sought only. The purpose of this sources sought/RFI is to gain knowledge of potential qualified sources and their size classification relative to NAICS 339114 (size standard of 750 Employees). Responses to this sources sought will be used by the Government to make appropriate acquisition decisions. After review of the responses to this sources sought, a solicitation announcement may be published. Responses to this sources sought synopsis are not considered adequate responses for a solicitation announcement.
The Department of Veterans Affairs (VA), VISN 15 Network Contracting Office, is seeking sources that can provide the following,:
Requirements:
JSN
Nomenclature
Quantity
U0022
Dental Imaging Software License Expansion
1
X6500
X-Ray Unit, Dental, 3D
1
X6600
Radiographic Unit, Dental, Wall Mounted, 7 mA
7
X6610
Sensor Kit, Dental Imaging System, Size 1 & Size 2
7
Specifications:
Dental Imaging Software License Expansion:
Software License must be included for 3D imaging machine to have the ability to communicate between the machine and VA computers.
X-Ray Unit, Dental, 3D:
The system shall be a Cone Beam Computed Tomography (CBCT) imaging system capable of performing 2D panoramic, 3D CBCT, and optional cephalometric imaging for dental, oral surgery, endodontic, orthodontic, and maxillofacial applications. System shall be new, current production model.
Imaging Capabilities
Must provide:
2D digital panoramic imaging
3D CBCT imaging
Optional cephalometric imaging capability (expandable or integrated)
Field of View (FOV) options ranging from small focused volumes (~5x5 cm) up to large craniofacial volumes (~16x17 cm or equivalent).
High-resolution imaging with voxel sizes as small as 75-150 µm (or better).
Adjustable FOV selection to limit patient radiation exposure.
Metal artifact reduction (MAR) or equivalent technology.
Detector & Image Quality
Flat panel digital detector technology.
High-contrast resolution suitable for:
Implant planning
Endodontic diagnosis
TMJ evaluation
Airway analysis
Grayscale depth minimum 14-bit or equivalent.
Automatic exposure control or programmable exposure settings.
Software & Workflow
Manufacturer-provided imaging software included.
Software shall support:
Implant planning tools
Nerve tracing
Airway measurement
TMJ analysis
Endodontic assessment
DICOM 3.0 compliant.
Integration capability with PACS and third-party practice management systems.
3D rendering and multiplanar reconstruction (MPR).
Export capability in STL format for surgical guides.
Patient Positioning & Ergonomics
Open-face design or patient-friendly configuration.
Positioning guides (minimum three planes).
Adjustable chin rest, head supports, and temple supports.
Motorized height adjustment.
Accommodates standing and seated (wheelchair accessible) patients.
Touchscreen interface mounted on unit.
Power: 110-240V, 50/60Hz.
Compatible with standard dental operatory electrical infrastructure.
Network connectivity via Ethernet
Radiographic Unit, Dental, Wall Mounted, 7mA
The system shall be a wall-mounted intraoral dental X-ray unit with extended reach (long arm configuration) for use in general dentistry, oral surgery, and specialty dental clinics.
Imaging Performance
High-frequency DC X-ray generator.
Constant potential output for consistent image quality.
Tube voltage approximately 60-70 kVp.
Tube current approximately 7 mA (or equivalent).
Exposure time range suitable for digital sensors and phosphor plates.
Compatible with:
Digital intraoral sensors
Traditional film
Rectangular collimation capability (preferred) to reduce patient dose.
Focal spot size approximately 0.7 mm or smaller.
Long Arm Configuration
Extended arm reach minimum 81 inches (or equivalent).
Double-articulating arm for flexible positioning.
Smooth, drift-free positioning.
Stable tube head that maintains selected position without sag.
Wall-mounted installation with appropriate structural support requirements provided.
Long arm shall accommodate large operatories and specialty treatment rooms.
Control & Interface
Wall-mounted control panel.
Programmable exposure presets for:
Adult anterior/posterior
Pediatric anterior/posterior
Digital display for exposure time.
Manual exposure adjustment capability.
Error codes or diagnostic display.
Safety & Radiation Protection
Compliance with FDA and NCRP radiation standards.
Minimum total filtration equivalent to 2.5 mm aluminum at 70 kVp.
Dead-man type exposure switch.
Minimum source-to-skin distance compliant with federal standards.
Audible and visual exposure indicators.
Leakage radiation within regulatory limits.
Mechanical & Ergonomic Features
Lightweight tube head.
360-degree rotational capability at arm and tube head joints.
Smooth, low-resistance movement.
Sealed, oil-cooled X-ray tube housing.
Easy-to-clean surfaces suitable for clinical infection control.
Electrical Requirements
Standard 110-120V electrical service.
Compatible with standard dental operatory electrical infrastructure.
Line voltage regulation to ensure stable output.
Sensor Kit, Dental Imaging System, Size 1 & Size 2
The system shall consist of direct digital intraoral radiography sensors, Size 1 and Size 2, compatible with standard intraoral dental X-ray generators. Sensors shall be new, current production models designed for high-resolution diagnostic imaging in general dentistry and specialty dental clinics.
Sensor Technology
CMOS or equivalent digital sensor technology.
True resolution minimum 20 line pairs/mm (or equivalent high-resolution performance).
Theoretical resolution up to 24 lp/mm (or equivalent).
Pixel size approximately 19-20 µm (or better).
Active imaging area optimized for diagnostic accuracy.
Must include:
Size 1 (Pediatric/Anterior)
Active area approximately 20 x 30 mm (or equivalent).
Slim profile for patient comfort.
Rounded corners for intraoral comfort.
Size 2 (Adult/Posterior Standard)
Active area approximately 26 x 36 mm (or equivalent).
Standard adult posterior imaging size.
Ergonomic shape for ease of positioning.
Image Quality & Performance
High contrast resolution suitable for:
Caries detection
Endodontic evaluation
Periapical imaging
Periodontal assessment
Wide dynamic range.
Noise reduction capability.
Automatic exposure detection.
Compatible with standard 60-70 kVp intraoral X-ray units.
Durability & Construction
Shock-resistant housing.
Reinforced cable attachment.
Cable length minimum 2 meters (or equivalent).
Water-resistant and disinfectant-resistant outer casing.
IP rating suitable for clinical dental environments.
Sensor thickness designed to enhance patient comfort.
Software & Integration
Manufacturer-provided imaging software included or compatible.
TWAIN and DICOM compliant.
Compatible with:
PACS systems
Dental imaging software platforms
Practice management systems
USB connectivity.
Plug-and-play configuration.
Infection Control
Compatible with standard disposable sensor barriers.
Smooth, sealed housing for disinfection.
Manufacturer-recommended cleaning protocols provided.
Electrical & Technical Requirements
USB 2.0 or higher connectivity.
Compatible with Windows-based acquisition workstations.
The information identified above are intended to be descriptive, not restrictive, and to indicate the quality of the supplies that will be satisfactory. It is the responsibility of the interested source to demonstrate to the government that the interested parties can provide the supplies/services that fulfill the required specifications.
If you are interested and are capable of providing the sought-out supplies/services, please provide the requested information as well as the information indicated below. Response to this notice should include company name, address, point of contact, size of business pursuant to the following questions:
Please indicate the size status and representations of your business, such as but not limited to: Service Disabled Veteran Owned Small Business (SDVOSB), Veteran Owned Small Business (VOSB), Hubzone, Woman Owned Small Business (WOSB), Large Business, etc.)?
Is your company considered small under the NAICS code identified under this RFI?
Are you the manufacturer, an authorized distributor, or an equivalent solution to the items being referenced above? If an authorized distributor must provide proof.
If you re a small business and you are an authorized distributor/reseller for the items identified above or an equivalent solution, are you providing the equipment from a small business manufacturer?
Do you alter, assemble, or modify; the items requested in any way? If you do, state how and what is altered; assembled; modified?
Does your company have an FSS contract with GSA, VA NAC, NASA SEWP, or any other federal contract, that can be utilized to procure the requirement listed above? If so, please provide the contract number.
Where is the product manufactured?
Please submit your capabilities that show clear, compelling, and convincing evidence that you can meet the requirement described in the attached SOW.
Please provide your UEI number.
Responses to this notice shall be submitted via email to Tracie Raggs at tracie.raggs@va.gov. Telephone responses shall not be accepted. Responses must be received no later than Friday, July 10, 2026, at 0800 CST. If a solicitation is issued it shall be announced at a later date, and all interested parties must respond to that solicitation announcement separately from the responses to this sources sought. Responses to this sources sought notice are not a request to be added to a prospective bidders list or to receive a copy of the solicitation
Files
| File | Type | Posted |
|---|---|---|
| 36C25526Q0487.docx | DOCX document |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| J065--Imaging Service & Preventative Maintenance EHCC & MTV | Veterans Integrated Service Network 15 | No set-aside | |
| 6525--SEP 2026 Equipment Only Consolidation | National Acquisition Center | No set-aside | |
| 3D Image Reconstruction and Case Planning Services | Veterans Integrated Service Network 8 | Total Small Business | |
| 6520--NX EQ Dental Handpieces | VA Strategic Acquisition Center Frederick | No set-aside | |
| DA10--RFI VHA Radiology and Cardiology Enterprise Imaging | VA Technology Acquisition Center | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity