J065--Imaging Service & Maintenance EHCC & MTV
Open Pre-Solicitation Posted
Not the latest version. See the latest solicitation from .
- Solicitation number
- 36C25527Q0053
- Agency
-
Veterans Integrated Service Network 15
Veterans Health Administration, Department of Veterans Affairs
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 811210 Electronic and Precision Equipment Repair and Maintenance
- Place of performance
- Evansville Health Care Center Evansville, Indiana 47715, United States
- Points of contact
Notice details come from SAM.gov. Updated .
About this opportunity
The Department of Veterans Affairs, Veterans Integrated Service Network 15, is preparing to procure preventive maintenance, repair, and parts-replacement services for Philips imaging equipment at the Evansville Health Care Center, with responses due December 2, 2026, under solicitation 36C25527Q0053. No set-aside is used. The VA Network Contracting Office in Leavenworth, Kansas, issues the requirement, and the place of performance is Evansville Health Care Center in Evansville, Indiana. Work covers radiology and ultrasound equipment, with NAICS 811210 and PSC J065. The contractor must respond by telephone within 30 minutes of a service call and arrive on-site within 4 hours. Required uptime is 97% or greater for radiology equipment and 95% or greater for ultrasound equipment. The contractor must provide all replacement parts, including tubes, ultrasound probes, detectors, grids, tables, and batteries, at no additional charge, and deliver replaced parts on-site by 10:30 AM the following day. Services are performed during VA business hours, 8:00 AM to 4:30 PM Monday through Friday, excluding federal holidays, with after-hours work available by COR arrangement at no additional charge. The contractor must obtain all operational and technical documentation, because the VA does not provide service manuals or diagnostic software. Compliance with NFPA-99, OSHA, and CDRH standards is required. The source does not state a period of performance, contract value, or incumbent. Questions go to Contract Specialist Tracie Raggs at tracie.raggs@va.gov.
Notice text
This is for preventative maintenance and service of Philips Imaging equipment. See the solicitation that follows for full details and requirements.
SERVICES TO BE PROVIDED:
A. Repairs
Service Hours:Â
All repairs will be performed during normal VA business hours (8:00 AM 4:30 PM, Monday-Friday), excluding Federal Holidays unless otherwise specified. Contractor may work outside normal business hours by arrangement with the COR if such services are provided without additional charge to the government. Any overtime charges must be approved by the COR or designee prior to the initiation of overtime work.
Response Time:Â
A technical support representative will respond by telephone to the VA within thirty (30) minutes of the original service call. The contractor will arrive on-site for repair within four (4) hours of the original service call. Radiology equipment shall have a guaranteed uptime of 97% or greater. Ultrasound Equipment shall have an uptime of 95% or greater, ensuring minimal downtime for critical diagnostic operations.
Documentation and Resources:Â
The VA shall not provide service manuals or service diagnostic software to the contractor. The contractor shall obtain, have on file, and make available to its FSE(s) all operational and technical documentation (e.g., operational and service manuals, schematics, parts lists) necessary to meet the performance requirements of this contract. The location and listing of the service data manuals, by name, and/or the manuals shall be provided to the CO upon request. Any charges for parts, manuals, tools, or necessary software for maintenance/repairs are included within this contract and its agreed-upon price unless otherwise stated in writing.
B. Parts
Replacement Parts:Â
The contractor shall provide all necessary replacement parts to ensure the equipment performs within the manufacturer's specifications. This includes, but is not limited to, coverage for tubes, ultrasound probes, detectors, grids, tables, and batteries. All replacement parts must meet or exceed the factory specifications to maintain compatibility with existing systems and support future performance and reliability upgrades. Any parts removed and replaced by the contractor will become the property of the contractor. Replaced parts must be delivered on-site no later than 10:30 AM the following day.
C. Software Updates/Upgrades
Software Maintenance:Â
The contractor will furnish, free of charge, all OEM software or hardware upgrades to the equipment purchased by the VA, which are intended to correct safety risks. Software updates offering enhancements to previously purchased software features shall also be covered under this service agreement if they do not require hardware modifications or additions.
D. Scheduled Maintenance
Preventive Maintenance (PM):Â
The contract will include comprehensive scheduled preventive maintenance inspections (PM) as required by manufacturer specifications.
Scheduling:Â
A mutually agreeable time for the inspections will be scheduled through the COR, Healthcare Technology Management (HTM) staff, or the Radiology supervisor/CT Lead.
Inspection and Maintenance Procedures:Â
The contractor will perform PM Services to ensure that equipment listed above functions in conformance with the latest published editions of NFPA-99, OSHA, and CDRH standards. The contractor shall provide and utilize procedures and checklists with worksheet originals indicating work performed and actual values obtained (as applicable) provided to the COR or HTM staff upon completion of the PM. Preventive maintenance procedures will be submitted to the COR for approval prior to initiation of this service and kept on file. PM services shall include, but are not limited to:
Cleaning of equipment
Reviewing operating system diagnostics to ensure system operation meets manufacturer's specifications
Calibrating and lubricating the equipment
Performing non-emergent remedial maintenance
Testing and replacing faulty or worn parts and/or parts likely to become faulty, fail, or become worn
Inspecting and replacing necessary cables and bushings
Measuring, adjusting, and calibrating for optimal image quality
Inspecting all mechanical components including, but not limited to, probes, cables, mounting hardware, chains, belts, bearings, tracks, interlocks, clutches, motors, monitors, keyboards for mechanical integrity, safety, performance
Returning equipment to full and proper operating condition per manufacturer specifications
Conducting an electrical safety inspection (as applicable) and documenting the results.
Compliance:Â
The contractor, their personnel, and their subcontractors shall be subject to Federal laws, regulations, standards, and VA Directives and Handbooks regarding information and information system security as delineated in this contract.
Files
| File | Type | Posted |
|---|---|---|
| 36C25527Q0053.docx | DOCX document |
Notice history
| Notice | Type | Posted |
|---|---|---|
| J065--Imaging Service & Maintenance EHCC & MTV | Solicitation | |
| This notice | Pre-Solicitation |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| J065--Fort Harrison GE Precision Radiology Preventive Maintenance | Veterans Integrated Service Network 19 | No set-aside | |
| 6525--SEP 2026 Equipment Only Consolidation | National Acquisition Center | No set-aside | |
| J059--Imaging UPSs Service Contract | Veterans Integrated Service Network 22 | Service-Disabled Veteran-Owned Small Business | |
| J065-- Bayer Radiology Injector Maint. - White River Junction VAMC | Veterans Integrated Service Network 1 | Service-Disabled Veteran-Owned Small Business | |
| J065--New Base 4 PM and Emergency Service for Hologic Mammograph systems for James Lovell Federal Health Care Center 3001 Green Bay Rd, North Chicago, IL 60064-3048 | Veterans Integrated Service Network 12 | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity