Question 1:

Referring to description of current ACC type chamber in Answers #6, there is no way 3 – 1700 watt heaters can do the 3 hour transition with a 6000 pound dead load. Please elaborate.

Answer 1:

The Combined Synopsis/Solicitation (CSS) attachment titled "Answers 6" has an error we identified regarding the ACC heater question.

Total heating capacity in the existing ACC is ~15kW.

3 manually selectable banks of heaters, 3 EA 1700 Watt resistors per bank