This is a Capabilities Assertion Matrix that serves as an evaluation tool for the Portable Patient Transport System (PPTS) prototype solicitation. Offerors must complete this matrix and include it within their 15-page Solution Brief to demonstrate compliance with six threshold-level Force Protection requirements. The matrix requires offerors to assert their compliance status (Met, Partially Met, Not Met, or Exceeds) for each requirement and provide specific references to supporting evidence in their Supporting Data Package.

The six requirements address critical performance specifications for the PPTS enclosure: maintaining a negative pressure differential of at least 0.01 inches of water gauge, directing airflow from patient head to foot, providing a minimum of 12 air exchanges per hour, achieving 99.97% filtration efficiency for 0.3-micron particles, utilizing NIOSH HE certified filters, and containing at least 3 liters of liquid contamination without leaking. This matrix is part of the Army's evaluation criteria for selecting offerors to receive invitations to submit a full Request for Solution Proposal (RSP) under the PPTS Call for Solutions (CFS) issued under BAA authority. The Army intends to award multiple Prototype Other Transaction Agreements to the most advantageous offerors, with the Agreements Officer and Specialist contact information available for questions.

View the file

Other files

Show all 12

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

PPTS

Requirement IDFunctional AreaRequirement DescriptionThreshold / DesiredAssertion (Met, Partially Met, Not Met, Exceeds)Supporting Data Reference
1.1Force ProtectionNegative Pressure Differential: Maintains = 0.01 inches of water gauge.Threshold
1.2Force ProtectionAir Flow Direction: Airflow is directed from the patient's head to their foot.Threshold
1.3Force ProtectionAir Exchanges: Provides = 12 air exchanges per hour.Threshold
1.4Force ProtectionFiltration Efficiency: Exhaust air filtration is = 99.97% for 0.3-micron particles.Threshold
1.5Force ProtectionFilter Certification: Filters meet or exceed NIOSH HE certified filter standards.Threshold
1.6Force ProtectionLiquid Containment: The enclosure holds at least 3 liters of liquid contamination without leaking.Threshold

Instructions for the Offeror:

This Assertion Matrix must be completed and included within the 15-page Solution Brief.

Offerors are directed to the Call for Solutions (CFS) document for the complete downselect evaluation criteria and the specific requirements necessary to be considered for an invitation to submit a Request for Solution Proposal (RSP).

For each requirement listed below, assert your compliance in the "Assertion" column and provide a specific reference to the verifiable evidence located in your "Supporting Data Package" (File 2).