This document is a technical data sheet for the Fluke Calibration P3000 Series Hydraulic Deadweight Testers, specifically Models P3100 and P3200. The testers are precision instruments used for pressure measurement, featuring pressure ranges up to 20,000 psi (1,400 bar) with accuracy better than 0.015% of reading (with an optional 0.008% accuracy). Available in oil and distilled water operated versions, these testers utilize a vertically mounted, precision-lapped piston and cylinder assembly with calibrated weight masses to measure pressure by balancing forces.

Key specifications include multiple model variants for different pressure ranges (single and dual piston configurations), test port adaptors (M14 x 1.5, M20 x 1.5, and various NPT/BSP sizes), and standard weight materials of non-magnetic austenitic stainless steel. Features include a built-in lever-style hand pump, mounted spirit level, adjustable feet, high-quality screw press for fine pressure control, and an acrylic reservoir for fluid level visibility. The instrument weighs 16 kg, has dimensions of 440 mm x 300 mm x 215 mm, and comes with optional accessories like improved accuracy software (PressCal) and fine increment weights for more precise pressure measurements.

View the file

Other files

Show all 15

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Technical Data

P3000 Series Hydraulic Deadweight Testers

Models P3100 and P3200

Features

• Pressure ranges to

20,000 psi (1,400 bar)

• Accuracy better than

0.015 % of reading (Increased accuracy option of 0.008 %)

• Mineral oil and distilled water models

• Dual piston models allow calibration over a wide range

• Psi, bar, kgf/cm2, kPa and MPa ranges available

• Dual piston models for calibration over a wide pressure range

• Built-in lever style hand pump now standard on all hydraulic models

• Mounted spirit level and adjustable feet

• High quality screw press for fine pressure control

• Test station design with O-rings eliminates the need for PTFE tape or wrenches

• Acrylic reservoir provides visibility of fluid level and quality

• Improved case with spring-loaded latches for hood

• Built-in drain plug to remove old fluid

• Sturdy weight box with hinged lid and side han-dles for easy transport

The P3000 Series is the culmination of over 50 years experience in the production and design of primary pres-sure standards. With features designed to improve accuracy and performance, increase reliability and simplify opera-tion these deadweight testers can be used to calibrate virtually any pressure sensing device, including transduc-ers, transmitters, gauges or pressure switches.

Each instrument is supplied with a detachable lid which makes it neat, compact and easily portable.

The weight masses are stored in a high-quality case with a self locking mechanism to protect them during transit.

All units are provided with an accredited calibration report, weight mass details, M14 x 1.5, M20 X 1.5, 1/8 inch, 1/4 inch, 3/8 inch and 1/2 inch NPT and BSP female adaptors, operating fluid (where applicable) and spare seals.

The piston/cylinder assemblies are manufactured to the very highest standards and are provided standard with ISO/IEC 17025 accredited calibra-tion reports.

1.888.610.7664 sales@GlobalTestSupply.com Fluke-Direct.com

2 Fluke Calibration P3000 Series Hydraulic Deadweight Testers Models P3100 and P3200

Operating principle Deadweight testers are the primary standard for pressure measurement. Utilizing the well proven piston-gauge system consisting of a vertically mounted, precision lapped piston and cylinder assembly, accurately calibrated weight masses (Force) are loaded on the piston (Area), which rises freely within its cylinder. These weights balance the upward force created by the pres-sure within the system.

The pressure is measured when placed on a correctly spinning and floating piston. The total pressure measured is the summation of the weights plus the piston weight carrier assembly.

Instrument base There are three basic variations in the hydrau-lic offering; single low pressure, single high pressure and dual piston models. Pressure is generated and controlled by means of a high quality screw press located on the front of the tester. A built-in hand pump is included as standard for all hydraulic models to prime the system and accommodate large volume requirements.

Piston/cylinder assemblies The piston/cylinder assembly is the heart of each deadweight tester. They are manufactured from materials that provide stability, durability, and low thermal coefficients and distortion. Our experience and knowledge of piston/cylinder production and calibration ensure the precision and performance required for today’s demand-ing calibration requirements.

Weight masses Standard weight masses are series 3 non-magnetic austenitic stainless steel. Each mass is marked with the serial number of the instru-ment and the nominal pressure value relative to the high or low-pressure piston, when appli-cable. Optional fractional weights are stainless steel and/or solution heat treated aluminum.

Gravity correction Gravity varies significantly with geographical location and this variation has a direct effect on the force of the weights and the accuracy of the deadweight tester. Each instrument can be trimmed to local gravity at no extra cost. If unspecified, instruments will be calibrated to Standard Gravity at 980.665 cm/s2.

Force

Force

Area

Area

Pressure

Pressure =

Dual piston design Priming pump

Hand tighten adapters

Floatation indicator

Acrylic reservoir

Fine control screw press

Leveling feet

Spirit level

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com

3 Fluke Calibration P3000 Series Hydraulic Deadweight Testers Models P3100 and P3200

Pressure ranges

Oil operated Ranges to 20,000 psi (1,400 bar)

Distilled water operated Ranges to 10,000 psi (700 bar)

Accuracy ± 0.015 % of reading (± 0.008 % optional) Note: Accuracy based on % of reading from 10 % to 100 % of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10 % ± (accuracy class) x 10 % of the piston range.

Materials of construction

Standard weight material Series 3 non-magnetic austenitic stainless steel Density: 7.8 g/cm3

Optional fractional weights Solution heat treated aluminum Density: 2.7 g/cm3

Piston material Tungsten carbide with nickel binder Density: 15.0 g/cm3

Cylinder material Hardened martensitic steel Tungsten carbide (water above 500 psi, 35 bar)

Thermal coefficients of expansion Oil piston/cylinder 16.5 ppm/°C Water (above 500 psi, 35 bar) 11 ppm/°C

General

Test port adaptors M14 X 1.5, M20 X 1.5, 1/8 in, 1/4 in, 3/8 in and 1/2 in NPT and BSP

Weight 16 kg (36 lb)

Dimensions (W x D x H) 440 mm x 300 mm x 215 mm (17.5 in x 12 in x 8.5 in)

Mass set weight (typical) 36 kg (80 lb)

Reservoir volume 150 cm3 (9.2 in3)

Screw press displacement 5.5 cm3 (0.34 in3)

Pump displacement 4.7 cm3 (0.29 in3) per stroke

O-ring seal materials Viton®

Weight increments

Minimum standard weight increments

Up to 500 psi (35 bar) 1 psi (0.1 bar)

500 psi to 5 000 psi (35 bar to 350 bar) 10 psi (1 bar)

5 000 psi to 20 000 psi (350 bar to 1 400 bar) 20 psi (2 bar)

Model P3112 1 psi (0.1 bar)

Optional fractional weights

Up to 500 psi (35 bar) 0.1 psi (0.01 bar)

500 psi to 5 000 psi (35 bar to 350 bar) 1 psi (0.1 bar)

5 000 psi to 20 000 psi (350 bar to 1400 bar) 2 psi (0.2 bar)

Operating fluids

Mineral oil Shell spindle oil 22, our reference 55-655, viscosity 40 cs@ 20 °C (68 °F)

Distilled or deionized water

Specifications

4 Fluke Calibration P3000 Series Hydraulic Deadweight Testers Models P3100 and P3200

Oil operated - single PCU Model P3111-BAR 1 to 35 bar P3111-KGCM2 1 to 35 kgf/cm2

P3111-PSI 10 to 500 psi P3111-KPA 100 to 3,500 kPa P3111-MPA 0.1 to 3.5 MPa P3112-BAR 4 to 140 bar P3112-KGCM2 4 to 140 kgf/cm2

P3112-PSI 40 to 2,000 psi P3112-MPA 0.4 to 14 MPa P3113-BAR 10 to 350 bar P3113-KGCM2 10 to 350 kgf/cm2

P3113-PSI 100 to 5,000 psi P3113-MPA 1 to 35 MPa P3114-BAR 20 to 700 bar P3114-KGCM2 20 to 700 kgf/cm2

P3114-PSI 200 to 10,000 psi P3114-MPA 2 to 70 MPa P3115-BAR 20 to 1,100 bar P3115-KGCM2 20 to 1,100 kgf/cm2

P3115-PSI 200 to 16,000 psi P3115-MPA 2 to 110 MPa P3116-BAR 20 to 1,400 bar P3116-KGCM2 20 to 1,400 kgf/cm2

P3116-PSI 200 to 20,000 psi P3116-MPA 2 to 140 MPa

Water operated - single PCU Model P3211-BAR 1 to 35 bar P3211-KGCM2 1 to 35 kgf/cm2

P3211-PSI 10 to 500 psi P3211-KPA 100 to 3,500 kPa P3211-MPA 0.1 to 3.5 MPa P3213-BAR 10 to 350 bar P3213-KGCM2 10 to 350 kgf/cm2

P3213-PSI 100 to 5,000 psi P3213-MPA 1 to 35 MPa P3214-BAR 20 to 700 bar P3214-KGCM2 20 to 700 kgf/cm2

P3214-PSI 200 to 10,000 psi P3214-MPA 2 to 70 MPa

Water operated - dual PCU Model P3223-BAR 1 to 350 bar P3223-KGCM2 1 to 350 kgf/cm2

P3223-PSI 10 to 5,000 psi P3223-MPA 0.1 to 35 MPa P3224-BAR 1 to 700 bar P3224-KGCM2 1 to 700 kgf/cm2

P3224-PSI 10 to 10,000 psi P3224-MPA 0.1 to 70 MPa

Ordering information

Options A. Improved Accuracy Option with Software - PressCal Software is a Windows based software program that allows users to easily apply all neces-sary corrections, allowing for improved accuracy of 0.008%.

PressCal is provided as standard with all 0.008 % instruments.

B. Fine Increment Weights

- Additional small weights that allow for finer pressure increments

• PPA9159-CAL For use with kPa, MPa, bar, or kgf/cm2 weight sets

• PPA9608-CAL For use with psi weight sets

C. Conversion Weights - Allows for an instrument to be used in a pressure unit other than the pressure unit the instrument was ordered for.

Oil operated - dual PCU Model P3123-BAR 1 to 350 bar P3123-KGCM2 1 to 350 kgf/cm2

P3123-PSI 10 to 5,000 psi P3123-MPA 0.1 to 35 MPa P3124-BAR 1 to 700 bar P3124-KGCM2 1 to 700 kgf/cm2

P3124-PSI 10 to 10,000 psi P3124-MPA 0.1 to 70 MPa P3125-BAR 1 to 1,100 bar P3125-KGCM2 1 to 1,100 kgf/cm2

P3125-PSI 10 to 16,000 psi P3125-MPA 0.1 to 110 MPa P3125XT-BAR 1 to 1,200 bar P3125XT-KGCM2 1 to 1,200 kgf/cm2

P3125XT-MPA 1 to 120 MPa

©2011-2013 Fluke Calibration. Specifications subject to change without notice.

Printed in U.S.A. 6/2013 3633536C_EN Pub-ID 11745-eng

Modification of this document is not permitted without written permission from Fluke Calibration.

Fluke Calibration. Precision, performance, confidence.™