This document is a Calibration Certificate issued by Fluke Calibration for a mass carrying bell (Model MB-7002-0.8) with Serial Number 1202, calibrated on 25 Jun 2018 for Portsmouth Naval Shipyard. The calibration was performed at Fluke Calibration's Phoenix Primary Pressure and Flow Laboratory, with the masses inspected and compared to working standards using substitution or direct weighing methods. The bell is composed of 6 AL-4V titanium, low carbon steel with nickel plate and AISI 304L non-magnetic stainless steel, with an adjustment tolerance of ±15 mg. The measurement result was 0.8000028 kg, with an expanded uncertainty of 2.0 mg against a specification of 8.0 mg, using method B. The calibration is traceable to the International System of Units (SI) and complies with ISO 17025:2005 and ANSI/NCSL Z540.1 standards.