This document is a specification sheet for a Sulfur Dioxide Analyzer requested by the Texas Commission on Environmental Quality (TCEQ). The analyzer must meet strict requirements, including having a U.S. EPA federal reference or equivalent method designation, being free-standing and no larger than 7"H x 17"W x 24"D, and weighing no more than 40 pounds. It must have standard 115-volt 60 Hz AC operation, an internal sample pump, analog and digital outputs, and remote access capabilities. The analyzer must also have specific performance criteria, such as a minimum detection limit of less than 0.4 parts per billion, zero and span noise limits, and response time requirements. The required delivery time is 120 calendar days after receipt of order, and the warranty period is a minimum of two years.

The specification sheet is related to a state and local contract opportunity for replacement equipment and parts to continue operating stationary air monitoring sites in Texas to meet federal requirements. No additional pricing terms, set asides, incumbents, funding sources, or other salient information is provided.

View the file

Other files

File Type Posted
ESBD_370407_1722528717355_582-24-02706 Attachment F.pdf PDF
ESBD_370407_1722528604326_RFB 582-24-02706 Non-Fillable Form.pdf PDF
ESBD_370407_1722528620382_582-24-02706 Attachment A.pdf PDF
ESBD_370407_1722528738846_TCEQ RFB Terms and Conditions Final 6-22-2023.pdf PDF
ESBD_370407_1722528639799_582-24-02706 Attachment B.pdf PDF
ESBD_370407_1722528699916_582-24-02706 Attachment E.pdf PDF
ESBD_370407_1722528563785_RFB 582-24-02706 Fillable Form.pdf PDF
ESBD_370407_1722528679153_582-24-02706 Attachment D.pdf PDF
ESBD_370407_1722528662858_582-24-02706 Attachment C.pdf PDF
ESBD_370407_1722528604326_RFB 582-24-02706 Non-Fillable Form.pdf PDF
ESBD_370407_1722528620382_582-24-02706 Attachment A.pdf PDF
ESBD_370407_1722528738846_TCEQ RFB Terms and Conditions Final 6-22-2023.pdf PDF
ESBD_370407_1722528639799_582-24-02706 Attachment B.pdf PDF
ESBD_370407_1722528717355_582-24-02706 Attachment F.pdf PDF
ESBD_370407_1722528662858_582-24-02706 Attachment C.pdf PDF
ESBD_370407_1722528699916_582-24-02706 Attachment E.pdf PDF
ESBD_370407_1722528679153_582-24-02706 Attachment D.pdf PDF
ESBD_370407_1722528563785_RFB 582-24-02706 Fillable Form.pdf PDF
ESBD_370407_1722528738846_TCEQ RFB Terms and Conditions Final 6-22-2023.pdf PDF
ESBD_370407_1722528662858_582-24-02706 Attachment C.pdf PDF
ESBD_370407_1722528639799_582-24-02706 Attachment B.pdf PDF
ESBD_370407_1722528604326_RFB 582-24-02706 Non-Fillable Form.pdf PDF
ESBD_370407_1722528679153_582-24-02706 Attachment D.pdf PDF
ESBD_370407_1722528717355_582-24-02706 Attachment F.pdf PDF
ESBD_370407_1722528563785_RFB 582-24-02706 Fillable Form.pdf PDF
ESBD_370407_1722528620382_582-24-02706 Attachment A.pdf PDF
Show all 26

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Solicitation Name: Sulfur Dioxide Analyzer Solicitation Number:

Attachment E Specifications for a Sulfur Dioxide Analyzer

Description: Sulfur Dioxide (SO2) Monitor for detecting and measuring SO2 in ambient air. The instrument shall be microprocessor controlled and use the Ultra-Violet (UV) fluorescence method of detection.

MINIMUM GENERAL SPECIFICATIONS

The SO2 analyzer must meet the following requirements:

A. Must have a U.S. Environmental Protection Agency (EPA) federal reference or federal equivalent method designation EQSA-0495-100.

B. The system must contain the following components:

i. Must be free-standing no larger than 7”H x 17”W x 24”D, and weight no more than 40 pounds.

ii. Must have standard 115-volt 60 Hz AC operation;

iii. The sample pump must be internal to the analyzer and flow rate controlled;

iv. Analog output to be 0-1 volt direct current;

v. Front panel USB connections for firmware upgrades;

vi. The Photomultiplier Tube (PMT) and all sockets and high voltage circuitry thereof, shall be of the highest quality;

vii. PMT temperature controlled for reduced noise;

viii. The sample inlet shall be a 0.25-inch Swagelok® compatible compression fitting;

ix. Analyzer must have easily viewed digital display with a touch screen interface;

x. The operational temperature range must be between 5 and 40ºC;

xi. The display should allow the operator to monitor levels, flows, alarms, temperature and pressure, trend data and voltages;

xii. Setup including zeroing, spanning, etc. should be accomplished using pushbuttons and the display;

xiii. Must have a RS-232 interface; and

xiv. Software and non-volatile random-access memory (RAM) storage shall allow the internal storage of a large number of 1-minute to 5-minute average SO2 samples and a record of alarms and out of limits conditions.

C. Must have user selectable dual ranges of 0-50 ppb and 0-20,000 ppb;

D. Minimum detection limit of less than 0.4 parts per billion;

E. Zero noise of less than 0.2 ppb;

F. Span noise of less than 0.5% of reading (RMS) above 50 ppb;

G. Precision of less than 0.5% of reading above 50 ppb;

H. Linearity should be ±1 percent of full scale;

I. Zero drift should be less than 0.5 ppb over 24 hours;

Solicitation Name: Sulfur Dioxide Analyzer Solicitation Number:

J. Span drift should be less than 0.5% of full scale over 24 hours;

K. Response time less than 120 seconds to 95% of scale;

L. Sample flow rate of 0.6 to 0.7 SLPM;

M. Measurements should be temperature and pressure compensated;

N. Span and zero adjustable and off-setable; and O. The operating system shall be capable of being updated.

P. Must have hardware specific remote access software that allows for remote connection with virtual interface and data downloading capabilities.

ADDITIONAL TERMS AND CONDITIONS

1. All items must be bid new, of current production, and include the manufacturer’s standard parts and accessories.

2. Routine delivery response to TCEQ will be within 120 calendar days after receipt of order.

3. Equipment must be shipped Freight on Board/Free on Board (FOB) Destination. The seller shall be responsible for the cost of transporting merchandise to the TCEQ delivery address provided in #4.

4. The warranty period should consist of complete coverage of system onsite, parts, and labor for a minimum of two years from the date of acceptance.

5. At least one manual or CD must accompany each instrument. The manual will include complete operating instructions, circuit descriptions, theory of operation, location of controls and switches, software commands, maintenance, and service instructions. The manual will also include schematic diagrams, pneumatic flow charts, lists of commonly replaced parts, warnings, and safety precautions. Electronic manuals are acceptable.

6. Technical support must be available for TCEQ staff between the hours of 8 a.m. and 5 p.m. Central Standard Time (CST) during regular business days, Monday through Friday, by telephone and electronic mail (e-mail).

Attachment E
Specifications for a Sulfur Dioxide Analyzer