ESBD_370407_1722528620382_582-24-02706 Attachment A.pdf

PDF 111 KB Posted

Attached to
Teledyne Equipment and Parts
Solicitation number
582-24-02706
Issued by
Texas

About this file

This document is a set of technical specifications for a Dynamic Dilution Calibrator to be provided under a State and Local contract opportunity in Texas. The calibrator must be capable of accurately diluting concentrated source gases with diluent gases, have multiple operating modes, generate ozone and nitrogen dioxide, and provide a range of gas flow and concentration capabilities. It must be integrated into an existing complex calibration and data logging system and provide remote access capabilities. The calibrator must meet specific physical, software, and documentation requirements.

The related State and Local contract opportunity is for the replacement of equipment and parts to continue operating stationary air monitoring sites to meet federal requirements. This appears to be an ongoing contract for Teledyne equipment and parts in the state of Texas.

View the file

Other files

File Type Posted
ESBD_370407_1722528717355_582-24-02706 Attachment F.pdf PDF
ESBD_370407_1722528604326_RFB 582-24-02706 Non-Fillable Form.pdf PDF
ESBD_370407_1722528620382_582-24-02706 Attachment A.pdf PDF
ESBD_370407_1722528738846_TCEQ RFB Terms and Conditions Final 6-22-2023.pdf PDF
ESBD_370407_1722528639799_582-24-02706 Attachment B.pdf PDF
ESBD_370407_1722528699916_582-24-02706 Attachment E.pdf PDF
ESBD_370407_1722528563785_RFB 582-24-02706 Fillable Form.pdf PDF
ESBD_370407_1722528679153_582-24-02706 Attachment D.pdf PDF
ESBD_370407_1722528662858_582-24-02706 Attachment C.pdf PDF
ESBD_370407_1722528604326_RFB 582-24-02706 Non-Fillable Form.pdf PDF
ESBD_370407_1722528620382_582-24-02706 Attachment A.pdf PDF
ESBD_370407_1722528738846_TCEQ RFB Terms and Conditions Final 6-22-2023.pdf PDF
ESBD_370407_1722528639799_582-24-02706 Attachment B.pdf PDF
ESBD_370407_1722528717355_582-24-02706 Attachment F.pdf PDF
ESBD_370407_1722528662858_582-24-02706 Attachment C.pdf PDF
ESBD_370407_1722528699916_582-24-02706 Attachment E.pdf PDF
ESBD_370407_1722528679153_582-24-02706 Attachment D.pdf PDF
ESBD_370407_1722528563785_RFB 582-24-02706 Fillable Form.pdf PDF
ESBD_370407_1722528738846_TCEQ RFB Terms and Conditions Final 6-22-2023.pdf PDF
ESBD_370407_1722528662858_582-24-02706 Attachment C.pdf PDF
ESBD_370407_1722528639799_582-24-02706 Attachment B.pdf PDF
ESBD_370407_1722528699916_582-24-02706 Attachment E.pdf PDF
ESBD_370407_1722528604326_RFB 582-24-02706 Non-Fillable Form.pdf PDF
ESBD_370407_1722528679153_582-24-02706 Attachment D.pdf PDF
ESBD_370407_1722528717355_582-24-02706 Attachment F.pdf PDF
ESBD_370407_1722528563785_RFB 582-24-02706 Fillable Form.pdf PDF
Show all 26

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

582-24-02706

Attachment A

Specifications for Dynamic Dilution Calibrator

Description: Dynamic dilution calibrator; microprocessor controlled for accurately diluting concentrated source gases with diluent gases. Source and diluent flows controlled and measured by separate microprocessor-controlled mass flow controllers.

Minimum General Specifications.

1. Shall dynamically mix a component gas with a balance gas (zero air) at total flow rates from two (2) to twenty (20) liters per minute (LPM) using automatic computation of component and balance gas flow rates.

2. The system shall have a minimum of four (4) operating modes as described hereafter.

a. Blend: By controlling the system from the front panel the user shall be able to specify the port at the rear of the chassis for cylinder gas (input gas port), the cylinder gas concentration, the required dilution and the concentration of the output gas. The system will automatically open the correct solenoid and allow cylinder gas to flow at the correct rate to mix with the correctly controlled dilution gas and provide the desired blend at the rear output at the desired concentration.

b. Blend with GPT. As in 2.A with appropriately mixed ozone to produce the desired level of nitrogen dioxide.

c. Manual. As in 2.A and 2.B with manual selection of all parameters.

d. Set-timer mode: As in 2.A and 2.B except by presetting levels and sequences of levels. Shall provide up to 20 sequences of up to 99 levels that may be set up on a repeating schedule. This permits the calibrator to run automatic, periodic calibrations of up to 6 different gases (including ozone and GPT produced nitrogen dioxide) of various blends. Calibrator should also provide I/O outputs for controlling external valves for routing of gas. These valves draw about 1.0 amps at 12 volts direct current (DC).

3. Shall have built-in stable ozone generator to provide precise and accurately known ozone concentrations (0-1 parts per million [ppm]) using an ultraviolet lamp for ozone generation. The system shall incorporate an internal ultraviolet (UV) optical sensor in a feedback circuit to control the generation of ozone to produce a constant, stable and repeatable ozone concentration at the specified flow rate.

4. Shall have gas phase titration of nitric oxide and ozone for the generation of nitrogen dioxide at precise levels.

5. Shall have a feature that simulates a gas phase titration mixing operation that uses the calibrator’s photometer to adjust the ozone generator drive voltage to match as closely as possible to the target concentration requested in any subsequent real GPT operation.

6. Shall be a freestanding bench top single cabinet model.

7. Shall have front panel alphanumeric display and keypad for manual data input.

8. Gas and air flows shall be controlled by microprocessor supported precision mass flow controllers (MFC) (gas flows 0-100 standard cubic centimeters per minute [SCCM], gas flows 0-20 SCCM, and air flow 0-20 standard liters per minute [SLPM]). Each MFC must be calibrated with nitrogen or air at 25 degrees Celsius. Other ranges shall be optional.

9. Total flow and individual source gas and diluent flows flow rates shall be available for monitoring via the display, which shall also provide gas concentrations, diluent and source gas pressures, ozone oven and chassis temperatures, calibration constants and various other operating parameters.

10. The calibrator shall have diluent air, calibration gas, ozone generator, and ozone photometer pressure sensors that can be verified and calibrated.

11. The calibrator shall include a battery backed-up electronic clock that includes 24-hourtime, days, months, and years. The clock shall be settable via the keypad (also remotely) and continue to operate during power outages. The clock will be time synched to the system one or more times weekly via RS-232 commands.

12. The calibrator shall restart itself after a power outage and shall not require readjustment or recalibration upon resumption of power.

13. Calibration data for the flow controllers shall be stored internally and be recallable. Mass flow controller (MFC) calibration data shall utilize a corrective algorithm with at least a ten-point comparison of commanded versus actual flow with linear interpolation between points.

14. Calibration data for the ozone generator shall be stored internally. The calibrator’s software shall be able to run an internal automatic ozone generator calibration program that uses the calibrated internal photometer as a reference. When calibrated the calibrator shall be able to produce ozone at any total flow between one (1) and five (5) SLPM at a concentration of 0.003 to 1 ppm with an accuracy of plus or minus 5%.

15. The system shall have both a reaction chamber for combining ozone and nitrogen oxide gas and a dilution/mixing chamber for mixing and or dilution of gas blends. The ozone generator’s constant flow output goes directly to the reaction chamber. The output of the reaction chamber flows directly to the mixing/dilution chamber and mixes with dilution gas.

16. The calibrator shall have at least one diluent gas input with a single 0-20 LPM MFC and at least four (4) source gas inputs each isolated and controlled by a separate normally closed solenoid valve. The gas valves feed common 0-100 ccm and 0-20 ccm MFCs. There shall be at least four output ports and one vent port. All ports must be located at the rear of the calibrator.

a. The diluent gas input shall be a 0.25 inch Swagelok® compatible compression fitting.

b. The source gas input fittings are to be 0.125 inch Swagelok® compatible compression fittings.

c. The output and vent ports shall have 0.25-inch Swagelok® compatible compression fittings.

17. The mass flow controllers accuracy and repeatability from 10 to 90 percent of full scale shall be at least plus or minus 0.5% with a response time to within two (2)% of set-point of 15 seconds or less.

18. All gas handling components shall be constructed of non-reactive materials; Teflon, stainless steel, glass or equivalent.

19. The system shall be able to operate between 10 and 50 degrees Celsius (C).

20. Power requirements 90-115 volts alternating current (VAC) 60 hertz (Hz).

21. The calibrator shall have an RS-232 serial data port (DB-25) to provide for data input and provide information about calibrations being performed, report various operating parameters, etc. All internal programming shall be addressable via these ports for modifications, upgrades, calibration etc.

22. The ozone generator shall be temperature controlled and be insensitive to ambient light and temperature variations.

Special requirements.

1. These instruments are to be integrated into an existing complex calibration and data logging system (LEADS) and must therefore perform and operate exactly as the existing calibrators already in place.

2. Must have hardware specific remote access software that allows for remote connection with virtual interface and data downloading capabilities.

Physical.

1. Freestanding cabinet with display and manual key entry at front. Gas and air inlets and outlets at the rear.

2. DB-25 connector (male) and solenoid driver output at the rear. The solenoid driver connector shall mate with a Weidmuller BL series plug 12601-6. The pin out of this connector shall match existing connectors and wiring. The drive for the solenoids shall be open collector type to pull +12 volt to ground with up to 1.5 amp load.

3. The calibrator must include an attached handle for carrying the instrument.

Operating system (internal software).

1. The calibrator shall work with providing levels of test gases and timed sequences of those levels. The levels are defined by two numbers, the first is keyed to the gas produced (1 = NO, 2 = SO2, 3 = CO, 4 = H2S, 5 = NO2, 6 = ozone, 7 = CH4, 8 = NOY). The second number indicates the gas concentration. Example: level 66 is ozone 400 parts per billion (ppb) (at 3 liters per minute [LPM]). Sequences are numbered one to twenty with no significance to the number. Levels are defined by flow and concentration.

Sequences consist of levels, and duration and valving. Gas inputs follow the level first number as consistently as possible i.e. inlet #1 is for NO, #2 is SO2, etc. Of course, since there are more gases used in our system than we have inputs, a single calibrator can only provide four test mixtures (plus ozone).

2. The calibrator can be setup to automatically time and program sequences and have sequences started externally. Levels and sequences can be programmed into the calibrator either remotely or manually and will be retained in the calibrator’s memory. The calibrator shall report the status of the running sequence and level when interrogated.

3. The nomenclature used with the levels and sequences includes additional information called status blocks which provides additional data to the data retrieval system.

4. The calibrator communicates with the data retrieval/control system via a system called “dot commands.” This is a system which involves a higher language for control/reporting.

5. The calibrator must have a built-in reference photometer to assay ozone and can be used to verify ozone production. The gas lines within the calibrator must have pressure sensors and must not have an automated flush mechanism.

6. All units will be delivered with two (2) sets of Operation and Service/Maintenance manuals. The manuals will include at minimum the following: complete operating instructions, circuit descriptions, theory of operation, location of controls and switches, software commands, maintenance, service instructions, schematic diagrams, pneumatic flow charts, lists of replacement parts, warnings, and safety precautions.

7. Warranty period: complete coverage of system, parts and labor for a minimum of two years from date of acceptance.

Attachment A
Specifications for Dynamic Dilution Calibrator
Minimum General Specifications.
Physical.
Operating system (internal software).

File details come from the government source that posted it. Updated .