About this file
This document is a technical equipment specification file containing detailed equipment drawings and utility requirements for STERIS Corporation washer/disinfector units. The file includes comprehensive specifications for two primary equipment models: the AMSCO® 7052HP Single-Chamber Washer/Disinfector (electric heated with ACU-rinse reservoir, unit offset from wall) and the Innowave PCF (internal drain cooling model).
The AMSCO 7052HP specifications detail physical dimensions (42" width x 80.75" height x 32" depth), operating weight of 1461 lbs when fully loaded, and chamber volume of 10.3 cubic feet. Utility requirements include hot water at 110-150°F with consumption of 3.5 US gallons per instrument cycle, cold water at 30-50 psig with 11.1 US gallons per cycle for sump filling, pure water at 5-20 psig dynamic pressure with 12.0 US gallons per cycle, and compressed air at minimum 80 psig. Electrical requirements are 460-480V (60 Hz, 3-phase) with 34.5A nominal amperage and 50A recommended circuit protection. The unit requires a minimum 4-inch floor drain with 8x8-inch floor sink and includes seismic mount provisions. Optional ventilation exhaust requires 3-inch diameter ducting at 60 SCFM. Total heat loss ranges from 8,504 to 10,314 BTU/hour depending on vented or non-vented configuration. The Innowave PCF model specifications include external height of 42.6" (closed lid) and 58.8" (open lid), operating weight of 456 lbs with standard basket and filled tank, and water consumption of 45 gallons per complete wash cycle (28.9 gallons cold, 16.1 gallons hot). Electrical supply options include 208V AC 3-phase (10kW requirement, 32A circuit breaker minimum) or 380-415V AC 3-phase with neutral (13kW requirement, 25A circuit breaker minimum). Operating environment requirements for both units specify indoor use only, altitude up to 6,562 feet, ambient temperature 41-104°F, and maximum relative humidity of 80% at 88°F decreasing to 50% at 104°F.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| S06-36C26126R0023 0001.pdf | ||
| 36C26126R0023_2.docx | DOCX document | |
| Attachment Request for Information RFI Form.pdf | ||
| SECTION 01 00 00.pdf | ||
| SECTION 22 67 19.16.pdf | ||
| Appendix A.pdf | ||
| WAGE DETERMINATION CA20260018 01-30-2026.pdf | ||
| Utility water sampling data.pdf | ||
| Basis of Design.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
09-01-15
22 67 19.16 - 1
SECTION 22 67 19.16
REVERSE-OSMOSIS WATER EQUIPMENT
Appendix B
Equipment list for reference purposes.
42 [1067]
UNIT WIDTH
80-3/4 [2051]
UNIT
HEIGHT
31 [787]
LOADING
HEIGHT
26-1/4 [667]
LOADING
CLEARANCE
HEIGHT
8-1/2 [216]
[381]
91 [2311]
RAISED DOOR
HEIGHT
34 [864]
MAX LENGTH
32 [813]
UNIT LENGTH
90 [2286]
SERVICE
AREA
HEIGHT
45 [1143]
WALL OPENING
*(NOTE F)
1-1/2 [38]
MINIMUM
94 [2388] *(NOTE G)
RECOMMENDED
98 [2489] *(NOTE H)
MAX 1 [25]
ADJUSTMENT WITH
WASHER LEGS
(UNIT HEIGHT 81-3/4 [2077] MAX)
MINIMUM
90 [2286] *(NOTE G)
RECOMMENDED
96 [2438] *(NOTE H)
81-7/8 [2080]
MAX
HEIGHT
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SERVICE AREA
REQUIRED FOR
DRYING MOTOR
REMOVAL
TYPICAL INSTALLATION
NOTES:
(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].
(B) SERVICE ACCESS SIDE OF UNIT IS RECOGNIZED BY THE TOUCH SCREEN AT THE RIGHT OF
WASH CHAMBER. ELECTRICAL AND CONTROL COMPONENTS, AS WELL AS MOST OF
SERVICABLE PARTS ARE ACCESSIBLE FROM THAT SIDE OF UNIT. NON-SERVICE SIDE IS
EQUIPPED WITH THE TOUCH SCREEN ON THE LEFT OF THE WASH CHAMBER.
(C) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.
(D) WEIGHTED EQUIVALENT SURFACE SOUND PRESSURE LEVEL: 64.0 dB(A) CALCULATED AS DESCRIBED IN ISO 3746.
(E) VERTICAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 3-1/2 in [89 mm] WIDTH X 80 in [2032 mm] LONG.
(F) HORIZONTAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 49 in [1245 mm] WIDTH X 16 in [406 mm] HEIGHT.
(G) MINIMUM REQUIRED DIMENSION FOR DOOR REMOVAL WITH LIMITED ADJUSTMENT AVAILABLE: +0 in [0 mm] THE STANDARD BARRIER WALL FLANGES CAN'T BE USED AT THAT DIMENSION. -3/4 in [19.1 mm]
(H) RECOMMENDED DIMENSION FOR DOOR REMOVAL WITH FULL LEVELING ADJUSTMENT AVAILABLE: +1 in [25.4 mm] THE STANDARD BARRIER WALL FLANGES CAN BE USED AT THAT DIMENSION. *(NOTE E AND F) -3/4 in [19.1 mm]
FRONT VIEW
SERVICE ACCESS
SIDE OF UNIT
*(NOTE B)
TOP VIEW
PASS THROUGH
DOOR
SIDE VIEWLEVEL FLOOR
*(NOTE C)
W A
LL
O
P E
N
IN
G H
E
IG
H T
C E
IL
IN
G H
E
IG
H T
FO
R
D O
O R
R E
M O
V A
L
[102]
Page 1 of 10 10292223.E
Released
2X 33-7/16 [849]
2X 27-11/16 [704]
5-11/16 [144]
7-1/8 [181]
4X 15 [381]
1-7/8 [47]
17-3/8 [441]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) CONCRETE FLOOR MUST REMAIN FREE OF DEFECTS IN THESE AREA, AS PER ESR-1917 & ESR-3187.
FOR INSTALLATION OF SEISMIC ANCHORS (NOT PROVIDED BY STERIS).
(B) THE FLOOR SINK AND DRAIN SHALL NOT OCCUPY THESE AREAS. 5 in x 5 in [127 mm x 127 mm] FLOOR SINK WITH Ø4 in [102 mm] DRAIN IS RECOMMENDED WHEN USING SEISMIC MOUNTS.
SEISMIC MOUNTS
SEISMIC ANCHORAGE
FOR DETAIL SEE FD354
QTY 4 ANCHORS
FINISHED WALL
PREFERENCE
LOCATION OF
FLOOR SINK
*(NOTE A AND B)
Page 2 of 10 10292223.E
80-3/4 [2051]
7-1/8 [181]
[381]
31-1/8 [791]
[432]
19-1/32 [483]
16-17/32 [420]
15-1/2 [394]
3-7/8 [99]
HW HOT WATER
CW COLD WATER
PW PURE WATER
A COMPRESSED AIR
D FLOOR DRAIN LOCATION
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (PLUMBING)
CONNECTION POSITION
FRONT VIEW
A
D
PW
HW
CW
SIDE VIEW
HW
PW
CW
A
D
TOP VIEW
HW
CW
PW
A
D
Page 3 of 10 10292223.E
79-1/4 [2013]
35-13/16 [910]
18-3/8 [467]
[534]
38-1/8 [967]
39-5/8 [1007]
E ELECTRICAL
N NETWORK
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (ELECTRICITY)
CONNECTION POSITION
FRONT VIEW SIDE VIEW
E
N
E
N
TOP VIEW
E
N
Page 4 of 10 10292223.E
6-5/32 [156]3
[76]
81-29/32 [2080]
EX VENTILATION EXHAUST
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (VENTILATION)
CONNECTION POSITION (OPTIONAL)
FRONT VIEW SIDE VIEW
EX
SUGGESTED DUCTWORK
(BY OTHERS)
EX
VAPOR CONDENSATE
COLLECTING MANAGED
BY THE UNIT
TOP VIEW
EX
VENTILATION CONNECTION (OPTIONAL)
(SEE PAGE 9 *(NOTE 1)
DRY CONTACT (SEE PAGE 10 *(NOTE 3))
EX
BY OTHERS
Page 5 of 10 10292223.E
60 [1524]
REF
8-1/2 [216]
132 [3350]
MAX.
17 [432]
INSTRUMENT
(With PUC HP)
UTENSIL
(With PUC HP)
CHEMICAL
(CH)
ENZYME CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
DETERGENT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
LUBRICANT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.2 oz [6 ml] N/A
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) DOSING STATION AND CHEMICAL CONTAINERS MUST BE ON THE SAME FLOOR OF THE BUILDING AS THE WASHER/DISINFECTOR.
CHEMICAL CONTAINER MUST BE LOCATED AT A LOWER HEIGHT THAN DOSING STATION. MAXIMUM CONTAINER INCLINATION
IS 15 DEG. FROM FLOOR. AMBIENT TEMPERATURE SHALL NOT EXCEED 32 °C [90 °F]
(B) HP PICKUP TUBE SYSTEM CAN PROVIDE UP TO TWO WASHERS WITH CHEMISTRY FROM THE SAME CONTAINER.
DOSING STATION
CEILING
ROUTE CHEMICAL LINES
AND SIGNAL CABLE
THROUGH PIPE TO REACH
BOTTOM OF UNIT
CHEMICAL CONTAINERS
*(NOTE A)
12 ft [3.6 m] OF INJECTION LINE
(FROM CONTAINERS TO
DOSING STATION)
*(NOTE B)
DOSING STATION
WEIGHT : 15 lb [6.8 kg]
2X FD349 SHOWN
UP TO 5 TUBES
100 ft [30 m] OF INJECTION LINE
AND SIGNAL CABLE
DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.
STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND
SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.
EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO
WASHER/STATION TO AVOID FORMING BUNDLES THAT COULD
AFFECT THE CHEMICAL INJECTION SYSTEM'S PERFORMANCE
(DO NOT CUT SIGNAL CABLE).
Page 6 of 10 10292223.E
UTILITY REQUIREMENTS CHART (PLUMBING)
UTILITY PARAMETER REQUIREMENT NOTES
HOT WATER
(HW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
1. STERIS REQUIRES TO INSTALL WATER HAMMER
ARRESTOR, UNION & PRESSURE GAUGE.
2. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
*STERIS OFFERS FLEXIBLE HOSES
ACCESSORY (FD088 OR FD037)
WITH INSTRUCTIONS.
3. SAMPLING VALVES INSTALLATION AS CLOSE AS
POSSIBLE TO WASHER WATER CONNECTIONS IS
RECOMMENDED.
PRESSURE RANGE
15 - 50 psig DYNAMIC
[103-345 kPa] MAX. 90 psig STATIC [620 kPA]
OPERATING FLOW RATE 6.8 US GPM [25.7 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
3.5 US GAL [13.3 L]
WITH ACU-WASH RESERVOIR:
0.0 US GAL [0.0 L]
TEMPERATURE RANGE 110 °F [43 °C] TO 150 °F [65.5 °C]
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
COLD WATER
(CW)
*(NOTES 6)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
4. CONSUMPTION BASED ON INSTRUMENT CYCLE WITH
PROLYSTICA ULTRA CONCENTRATE HIGH
PERFORMANCE CHEMICALS (PUC HP) WITH
THE FOLLOWING UTILITY CONDITIONS:
THREE-LEVEL MANIFOLD RACK
(CW) COLD WATER PRESSURE: 30 psig DYNAMIC [207 kPa]
(CW) COLD WATER TEMPERATURE: 70 °F [21 °C]
(HW) HOT WATER TEMPERATURE: 150 °F [65 °C]
(PW) PURE WATER TEMPERATURE 70 °F [21 °C]
UTENSIL CYCLE CONSUMPTION:
TWO-LEVEL MANIFOLD RACK
CW: 11.0 US GAL [41.6 L]
HW: 3.4 US GAL [12.9 L]
PW: 12.0 US GAL [45.4 L]
5. STERIS RECOMMENDS THE FOLLOWING WATER
QUALITY PARAMETERS:
TOTAL DISSOLVED SOLIDS: 100-200 ppm (MAX. 500 ppm) TOTAL ALKALINITY (CaCO3) : 70-120 ppm (MAX. 180 ppm) pH : 6.8-7 (MAX. 7.5) TOTAL SILICA : 0.1-0.5 ppm (MAX. 1 ppm)
PRESSURE RANGE
30 - 50 psig DYNAMIC
[206-345 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 10.7 US GPM [40.5 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
FOR SUMP FILLING: 11.1 US GAL [42.0 L]
(ADD THE FOLLOWING CONSUMPTION
FOR OPTIONS/CONFIGURATIONS)
NON-VENTED SYSTEM:
15.2 US GAL [57.5 L]
DRAIN COOLDOWN:
18.9 US GAL [71.5 L]
DRAIN ACU-COOLDOWN:
13.0 US GAL [49.2 L]
TEMPERATURE RANGE 70 °F [21 °C] MAXIMUM
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
PURE WATER
(PW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
6. BACK FLOW PREVENTER AGAINST BACK
SIPHONAGE MAY BE REQUIRED (LOCAL
PLUMBING CODE)
PRESSURE RANGE 5 - 20 psig DYNAMIC [30 - 138 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 5.5 US GPM [21 L/min] TO
10.3 US GPM [39 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
* (NOTE 4 )
12.0 US GAL [45.4 L]
TEMPERATURE RANGE 70 °F [21.1 °C] TO 150 °F [65.5 °C]
RESISTIVITY
(CONDUCTIVITY)
0.1-0.5 M -cm (2-10 µS/cm) pH 6.8 - 7.5
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 7 of 10 10292223.E
Released
Hly Highlight
UTILITY REQUIREMENTS CHART (PLUMBING)
COMPRESSED
AIR
(A)
CONNECTION SIZE / TYPE 1/8 in / FEMALE NPT [1/8 in / MALE BSPT]
1. AIR PURITY PER ISO-8573-1: 2010 [7:4:4]
MAX. PARTICLE DENSITY: 10 mg/m³
MAX. DEW POINT FOR WATER CONTENT:
37 °F [3 °C]
MAX. OIL CONCENTRATION FOR THE OIL
CONTENT: 5 mg/m³
2. STERIS RECOMMENDS A REFRIGERATED AIR
DRYER WHERE ENVIRONMENTAL DEW POINT
CONDITIONS ARE HIGHER THAN
RECOMMENDED.
3. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
PRESSURE RANGE MIN 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]
OPERATING FLOW RATE 1.75 SCFM [50 L/min]
DRAIN
(D)
FLOOR DRAIN SIZE MIN 4 in [102 mm] O.D. WITH 8 in x 8 in [204 mm x 204 mm] FLOOR SINK
4. WASHER DRAIN OUTLET SIZE 1-1/2 in / MALE NPT
5. MAX OFFSET OF 3/4 in [19 mm] FROM
OUTLET (PAGE 3, D).
6. FOR DRAIN REQUIREMENTS WHEN USING
SEISMIC MOUNTS, SEE PAGE 2.
OPERATING FLOW RATE GRAVITY DRAIN : 50 US GPM
[189.3 L/min]
CHEMICAL
DOSING
SYSTEM
CEILING ROUTING
CONDUIT 3 in CONDUIT
11. STERIS RECOMMENDS TO ROUTE CHEMICAL
LINES AND SIGNAL CABLE THROUGH A
3 in [76 mm] ID CONDUIT (BY OTHERS).
100 ft [30 m] OF INJECTION LINE AND
SIGNAL CABLE ARE PROVIDED WITH THE UNIT.
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 8 of 10 10292223.E
Released dverdoor Rectangle
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
36-1/4 [921]
20 [508]
25-5/8 [651]
30-3/8 [772]
27-1/2 [699]
15-5/8 [397]
UTILITY REQUIREMENTS CHART (ELECTRIC)
UTILITY
PARAMETER AND REQUIREMENT NOTES
CONNECTION
SIZE/TYPE
NOMINAL
AMPERAGE
RECOMMENDED
CIRCUIT
ELECTRICITY
(EL)
208V 60 Hz, 3-Ph
CONVERTED TO
460-480V
THROUGH TRANSFORMER
(OPTIONAL) FD787
N/A N/A
(NOTE 7)
N/A
(NOTE 7)
1. ALL CONNECTIONS SHOULD BE IN
ACCORDANCE WITH LOCAL CODES.
2. NEUTRAL WIRE NOT REQUIRED.
3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES
AND SAFETY-RELATED WORK PRATICES TO
DETERMINE THE VALUE OF PROTECTION.
4. WIRING ON THE EQUIPMENT TERMINATES AT
A JUNCTION BOX. WIRING BETWEEN JUNCTION
BOX AND BUILDING SERVICE LINES ARE NOT
SUPPLIED BY STERIS.
5. DISCONNECT SWITCHES WITH OFF POSITION
LOCKOUT (SUPPLIED BY OTHERS) MUST BE
INSTALLED IN ELECTRIC SUPPLY LINES NEAR
THE EQUIPMENT.
6. STERIS RECOMMENDS A LIGHT IN SERVICE
AREA.
7. IT IS CUSTOMER RESPONSABILITY TO
CALCULATE NOMINAL AMPERAGE REQUIRED AT
TRANSFORMER INPUT. NOMINAL AMPERAGE
REQUIRED AT TRANSFORMER OUTPUT TO
SUPPLY THE WASHER CAN BE FOUND IN THIS
TABLE FOR THE SPECIFIC ELECTRICAL
CONFIGURATION OF THE WASHER.
200V 50 Hz, 3-Ph
CONVERTED TO
380-400-415V
THROUGH TRANSFORMER
(OPTIONAL) FD78-800
N/A N/A
(NOTE 7)
N/A
(NOTE 7)
460-480V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
34.5 A 50 A
380-400V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
39.5 A 50 A
380-400-415V 50 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
39 A 50 A
NETWORK
(N)
OPTIONAL
100/1000 BASE T SUPPORTING
TCP/IP PROTOCOL
RJ45F CONNECTION
8. NETWORK CONNECTION (IF APPLICABLE)
- ACTIVE NETWORK DROP WITH CAT 5/6 CABLE
WITHIN 10 ft [3 m] OF UNIT
- STATIC OR DHCP RESERVED IP ADDRESS
(FOR EACH UNIT)
- NETWORK CABLE (SUPPLIED BY OTHER)
208V, 60Hz 200V, 50Hz POWER (kVA) 45
INPUT (HV) (V 3, 60Hz) 208/240/600 208
OUTPUT (LV) (V 3, 60Hz) 480Y / 277 231 / 400
WEIGHT LB (KG) 600 (272) 343 (156)
NOISE LEVEL (dBa) 45
NO LOAD LOSS (WATTS) 200 319
LOAD LOSS
338°f (170°C)
(WATTS) 2600 2461
TOTAL LOSSES (WATTS) 2800 2780
EXCITATION
CURRENT (%) 1.1 3.26
INSTALLATION FLOOR
SIZE IN (MM) SEE RIGHT
PRIMARY CONNECTORS
PER PHASE
(MCM) 1x 250 MCM
SECONDARY CONNECTORS
PER PHASE
(AWG) 1x #2 AWG-#14 AWG
TEMPERATURE
CLASS
°F (°C) 428° (220°)
WINDING (HV) (LV) COPPER
ENGINEERING DATA-RECOMMENDED
TRANSFORMER
WEIGHT AND DIMENSIONS MAY VARY
TRANSFORMER
Page 9 of 10 10292223.E
Released dverdoor Rectangle
UTILITY REQUIREMENTS CHART
VENTILATION
EXHAUST
(EX)
OPTIONAL
CONNECTION SIZE 3 in [76 mm]
OUTSIDE DIAMETER
1. NOT REQUIRED WITH NON VENTED UNIT.
2. STERIS REQUIRES CORROSION-RESISTANT AND
WATER-TIGHT DUCT:
- MIN OPERATING TEMPERATURE 210 °F [99 °C]
- SLOPED TOWARDS THE WASHER AND NO DEAD LEGS
- INSTALL DRAIN PART AT LOWEST POINT NEAR UNIT
3. EXHAUST FAN NORMALLY OPEN-AUXILIARY CONTACTS
IS AVAILABLE TO SUPPLY A SIGNAL TO THE FACILITY HVAC
CONTROL SYSTEM WHENEVER THE WASHER VENTILATION
EXHAUST FAN IS IN OPERATION.
PRESSURE RANGE 1/2 in [50 mm]
VACUUM AT CONNECTION
OPERATING FLOW RATE 60 SCFM [1.7 m³/min]
TEMPERATURE RANGE MAX. 240 °F [116 °C]
PHYSICAL SPECIFICATIONS
SHIPPING DIMENSION W x H x L 50 in x 91 in x 42 in [1270 mm x 2311 mm x 1067 mm]
1. LEVELING FEET ARE PROVIDED FOR PROPER INSTALLATION.
2. PERMISSIBLE ENVIRONMENTAL CONDITIONS.
THIS EQUIPMENT IS DESIGNED TO GIVE OPTIMAL RESULTS
UNDER THE FOLLOWING CONDITIONS:
- INDOOR USE ONLY;
- ALTITUDE OF OPERATION UP TO 6562 ft [2000 m];
- TEMPERATURE OF 41 TO 104 °F [5 to 40 °C];
- MAXIMUM RELATIVE HUMIDITY IS 80% FOR TEMPERATURE UP
TO 88 °F [31 °C], DECREASING LINEARLY TO 50% RELATIVE
HUMIDITY AT 104 °F [40 °C];
- POLLUTION DEGREE 2: EQUIPMENT MUST BE INSTALLED IN AN
ENVIRONMENT WHERE NORMALLY ONLY NON-CONDUCTIVE
POLLUTION OCCURS BUT WHERE OCCASIONAL, TEMPORARY
CONDUCTIVITY CAUSED BY CONDENSATION CAN BE EXPECTED
(AS DETERMINED ACCORDING TO INTERNATIONAL STANDARD
EN/IEC 61010-1).
3. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE TO,
OR FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT
DRAWING AND NOTE PAGES. SPECIFICATIONS AND
DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.
SHIPPING WEIGHT 1372 lb [622 kg]
OPERATING DIMENSION W x H x L 42 in x 80.75 in x 32 in [1067 mm x 2051 mm x 813 mm]
OPERATING WEIGHT
(WITH WATER AND FULLY LOADED) 1461 lb [663 kg]
AVERAGE FLOOR LOAD
OVER WASHER FOOTPRINT 1.09 psi [7.49 kPa]
MAX. FLOOR LOAD
AT WASHER FOOT PADS
(3 in [76.2 mm] DIAMETER PADS) 75 psi [517 kPa]
NOISE LEVEL:
A-WEIGHTED EQUIVALENT
SURFACE SOUND PRESSURE LEVEL
60.6 dB(A)
CALCULATED AS DESCRIBED
IN ISO 3746.
OPENING AREA FROM
CLEAN SIDE TO SOILED SIDE
151.8 in² [979 cm²]
*(NOTE 4)
4. OPENING AREA CAN BE REDUCED TO 26 in² [168 cm²]
WITH INSTALLATION OF AIR MANAGEMENT PANELS
ACCESSORY (FD352)
CHAMBER DIMENSION W x H x L
VOLUME
26.5 in x 26.25 in x 25.5 in [673 mm x 667 mm x 648 mm]
10.3 ft³ [0.29 m³]
TYPICAL HEAT LOSS AT 75 °F [24 °C]
CONTINUOUS OPERATION
NON-VENTED UNIT
WITH ACU-RINSE RES.
VENTED UNIT (OPTIONAL)
WITH ACU-RINSE RES.
CLEAN SIDE 2049 Btu/h [600 W]
2049 Btu/h [600 W]
SOILED SIDE 8265 Btu/h [2420 W]
6455 Btu/h [1890 W]
TOTAL 10314± 854 Btu/h [3020 ± 250 W]
8504 ± 854 Btu/h [2490 ± 250 W]
VALUES INCLUDE HEAT LOSS DURING DOOR OPENING FOR LOADING AND UNLOADING OPERATIONS. "TOTAL" WILL VARY DEPENDING ON DISINFECTION TIME AND TEMPERATURE, DRYING TIME AND TEMPERATURE AND USE RATE OF UNIT.
AN ADDITIONAL HEAT LOSS FOR LOAD COOL DOWN IN CLEAN SIDE CAN BE ADDED TO THE TOTAL HEAT LOSS OF THE WASHER:
MAXIMUM 1713 Btu [1807 kJ]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/ ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Michel Blouin 10292223
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 10 of 10 10292223.E
2.
6" (1
3m m
6" (150mm) MIN BOTH SIDES
34.5" (877mm)
8.
8" (1
3m m
27.8" (706mm)
11" (279mm) MIN
R
EC
O M
M
EN
D
ED
C
EI
LI
N
G H
EI
G
HT
0"
(2
6m m
) M
IN
EXTERNAL DIMENSIONS AND OPERATING CLEARANCES
(RECOMMENDED CLEARENCE ALL ROUND FOR SERVICABILITY = 18")
FOR EXTERNAL DRAIN COOLING MODEL SEE - 10997956, FOR PRE 0219 MODELS SEE - 764336970
PHYSICAL DIMENSIONS AND WEIGHT
EXTERNAL HEIGHT (LID CLOSED) 42.6" 1083mm
EXTERNAL HEIGHT (LID OPEN) 58.8" 1493mm
EXTERNAL WIDTH 34.5" 877mm
EXTERNAL DEPTH (LID OPEN) 27.8" 706mm
NET WEIGHT (NO BASKET, HOSES ETC) 311lb 141kg
OPERATING WEIGHT (STANDARD BASKET + HOSES, FILLED TANK & 2X CHEMICAL) 456lb 207kg
MAX OPERATING WEIGHT (STANDARD BASKET + HOSES, FILLED TANK & 2X
CHEMICAL + INSTRUMENTS) 500lb 227kg
TITLE:
DWG NO.
DRAWN BY: LC SHEET 1 OF 8
A4
REVISION: A
13464836
PCF (INTERNAL DRAIN COOLING, UNITS BUILT FROM SN
11210713301 AND CONVERTED UNITS) EQUIPMENT
DRAWING
DATE: ECN NUMBER:107745 ECN DESCRIPTION: PRODUCTION RELEASE
This document contains confidential and proprietary information of STERIS Corporation.
Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
28/03/2022
A A
B B
C C
D D
E E
F F
4" (102mm)
8.75" (223mm)
18.5" (470mm)
15.75" (400mm)
4" (102mm)
2m m
OVERFLOW FLEXIBLE
PLASTIC HOSE
PRINTER
CONNECTION
POWER MAINS
ENTRY
DRAIN 3/4"
HOSE TAIL
HOT WATER 3/4" BSP
COLD WATER 3/4" BSP
RO WATER 3/4" BSP
2m m
.7
5" (7
0m m
1.5" (38mm) MIN
7.75" (197mm) MIN
A
IR
G A
P
RE
Q
UI
RE
D
4" (102mm) MIN
OPEN FLOOR SINK
DRAIN FLEXIBLE
PLASTIC HOSING
STANDARD DISTANCES + DRAIN CONFIGURATION OPTION 1
(UNIT MUST NOT BE HARD-PIPED TO ENABLE SERVICABILITY)
DRAIN COMPONENTS AND INSTALLATION BY OTHERS
TITLE:
DWG NO.
DRAWN BY: LC SHEET 2 OF 8
A4
REVISION: A
13464836
PCF (INTERNAL DRAIN COOLING, UNITS BUILT FROM SN
11210713301 AND CONVERTED UNITS) EQUIPMENT
DRAWING
DATE: ECN NUMBER:107745 ECN DESCRIPTION: PRODUCTION RELEASE
This document contains confidential and proprietary information of STERIS Corporation.
Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
28/03/2022
A A
B B
C C
D D
E E
F F
4" (102mm) MIN
A
IR
G A
P
IF
R
EQ
UI
RE
D
BY
L
O C
A L
RE
G
UL
A
TIO
N
S
9.
0m m ) M
X
OVERFLOW FLEXIBLE
PLASTIC HOSE
A
IR
G A
P
RE
Q
UI
7.75" (197mm) MIN
DRAIN FLEXIBLE
PLASTIC HOSE
DRAIN CONFIGURATION OPTION 2
(UNIT MUST NOT BE HARD-PIPED TO ENABLE SERVICABILITY)
DRAIN COMPONENTS AND INSTALLATION BY OTHERS
MATERIALS FOR PLUMBING
PARTS IDENTIFIED MUST HOLD SHORT TERM TEMPERATURES (UP TO 2 MINUTES) OF
93 DEG C (200 DEG F) WATER. SEE MATERIAL RECOMMENDATIONS BELOW:
MATERIAL STANDARD
STAINLESS STEEL PIPE (304 AND 316L)
ASME A112.3.1
COPPER OR COPPER-ALLOY TUBING (TYPE K,WK, L,WL, M OR WM)
ASTM: B42, B302, B43, B75, B88, B251, B447
GALVANIZED STEEL PIPE
ASTM A53
CHLORINATED POLYVINYL CHLORIDE (CPVC) PLASTIC PIPE TUBING
ASTM: D2846; F441, F442
CSA B137.6
TITLE:
DWG NO.
DRAWN BY: LC SHEET 3 OF 8
A4
REVISION: A
13464836
PCF (INTERNAL DRAIN COOLING, UNITS BUILT FROM SN
11210713301 AND CONVERTED UNITS) EQUIPMENT
DRAWING
DATE: ECN NUMBER:107745 ECN DESCRIPTION: PRODUCTION RELEASE
This document contains confidential and proprietary information of STERIS Corporation.
Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
28/03/2022
A A
B B
C C
D D
E E
F F
4" (102mm) MIN
1.
5" (3
8m m
SU
PP
O
RT
U
SI
N
G
W
LL
P
IP
E
BR
A
C
KE
TS
A
IR
G A
P
RE
Q
UI
IR
G
P
IF
R
EQ
UI
RE
D
BY
L
O C
A L
RE
G
UL
A
TIO
N
S
-2
D
RA
IN
F
LL
9.
0m m ) M
X
OVERFLOW FLEXIBLE
PLASTIC HOSE
7.75" (197mm) MIN
DRAIN FLEXIBLE
PLASTIC HOSE
SUITABLE
WALL
BRACKETS
DRAIN CONFIGURATION OPTION 3
(UNIT MUST NOT BE HARD-PIPED TO ENABLE SERVICABILITY)
DRAIN COMPONENTS AND INSTALLATION BY OTHERS
MATERIALS FOR PLUMBING
PARTS IDENTIFIED MUST HOLD SHORT TERM TEMPERATURES (UP TO 2 MINUTES) OF
93 DEG C (200 DEG F) WATER. SEE MATERIAL RECOMMENDATIONS BELOW:
MATERIAL STANDARD
STAINLESS STEEL PIPE (304 AND 316L)
ASME A112.3.1
COPPER OR COPPER-ALLOY TUBING (TYPE K,WK, L,WL, M OR WM)
ASTM: B42, B302, B43, B75, B88, B251, B447
GALVANIZED STEEL PIPE
ASTM A53
CHLORINATED POLYVINYL CHLORIDE (CPVC) PLASTIC PIPE TUBING
ASTM: D2846; F441, F442
CSA B137.6
TITLE:
DWG NO.
DRAWN BY: LC SHEET 4 OF 8
A4
REVISION: A
13464836
PCF (INTERNAL DRAIN COOLING, UNITS BUILT FROM SN
11210713301 AND CONVERTED UNITS) EQUIPMENT
DRAWING
DATE: ECN NUMBER:107745 ECN DESCRIPTION: PRODUCTION RELEASE
This document contains confidential and proprietary information of STERIS Corporation.
Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
28/03/2022
A A
B B
C C
D D
E E
F F
*UNIT IS ABLE TO OPERATE WITHOUT HOT WATER CONNECTION OR WITH HOT WATER SUPPLY COOLER THAN RECOMMENDED MINIMUM TEMPERATURE. USING ONLY COLD WATER OR LOW-TEMPERATURE HOT
WATER WILL RESULT IN LONGER CYCLE OPERATION AS WATER TAKES LONGER TO HEAT UP.
TYPICAL WATER CONSUMPTION PER WASH
WATER CONSUMPTION WILL VARY DEPENDING ON VOLUME OF INSTRUMENTS LOADED AND
TEMPERATURE OF INLET WATER SUPPLY. WATER CONSUMPTION BELOW DESCRIBES A DA VINCI
DISINFECTION WASH PROGRAM WITH NO INSTRUMENTS LOADED.
WATER SUPPLY REQUIREMENTS
WATER HARDNESS (CaCO3), QUALITY
AND PURITY, HOT AND COLD SUPPLIES.
TREATED WATER SPECIFICATIONS
<125 PPM WITH WATER QUALITY AND PURITY TESTED AND
APPROVED FOR DISINFECTION AS PER ANSI/AAMI ST
15883-1:2009 (ITEM 6.4.1 AND 6.4.2) OR AAMI TIR 34
<30μS/cm (RO) <10μS/cm (DI) <30μS/cm (SOFT) IF
ABOVE THE RECOMMENDATION MORE CLEANING WILL
BE REQUIRED.
PRESSURE (HOT, COLD AND RO) 30 TO 70 PSI (2.0 TO 5.0 BAR)
FLOW RATE (HOT, COLD AND RO) 3 TO 8 US GAL/MIN (11 TO 30 LITRES/MIN)
TEMPERATURE (HOT, COLD AND RO) COLD: 41 F TO 86 F (5 C TO 30 C) HOT: 122 F TO 176 F
(50 C TO 80 C)* RO: 41 F TO 86 F (5 C TO 30 C)
CONNECTIONS
3/4" BSP MALE THREAD. RUBBER GASKET IS REQUIRED. DO
NOT USE TAPE OR ANY SEALENT. 3/4" NPT FACILITY
CONNECTIONS REQUIRED. HOSE CONNECTIONS WILL
NOT WORK
POSITION (HOT, COLD AND RO) WITHIN 3FT (1m) OF PROPOSED UNIT LOCATION
BACK SIPHONAGE PROTECTION INTEGRAL AB TYPE AIR GAP AND OVERFLOW
SITE DRAIN SHOULD MEET THESE REQUIREMENTS
DRAIN CONNECTION
3/4" HOSE TAIL OUTLET FROM UNIT. DRAIN MUST BE FITTED WITH SUITABLE TRAPS. NO LOOPS
IN DRAIN HOSE, CUT HOSE TO LENGTH REQUIRED. DRAIN LINE CANNOT BE TIED DIRECTLY
TO THE DRAIN LINE OF ANOTHER DEVICE. UNIT SHOULD BE INDEPENDENTLY PIPED TO A
DEDICATED DRAIN CONNECTION OR FLOOR SINK. AN AIR GAP MAY BE REQUIRED BY
LOCAL CODE.
POSTITION FLOOR SINK OR FLOOR DRAIN IS PREFERRED. STANDPIPE AND WALL DRAINS ARE OPTIONS
FOR CONNECTION TO DRAIN. ADHERE TO LOCAL CODE.
CAPACITY 9.0 US GAL/MIN. (34 LITRES/MIN) MAX DISCHARGE TO DRAIN INCLUDING OVERFLOW
OVERFLOW DRAIN
OVERFLOW HOSE CONNECTION TO MAIN DRAINAGE SYSTEM MUST NOT BE AIRTIGHT. AN
AIR TIGHT SEAL OF OVERFLOW TO DRAIN MAY CAUSE INCORRECT UNIT OPERATION. AN
AIR GAP MUST BE FITTED IF REQUIRED BY CODE.
OVERFLOW DRAIN
ROUTING
OVERFLOW HOSE MUST NOT BE POSITIONED OR ROUTED SUCH THAT IT CAN TRAP OR
HOLD WATER IN ANY CAPACITY.
EFFLUENT TEMPERATURE UP TO 140 ºF (60 ºC)
TYPICAL WATER CONSUMPTION FOR COMPLETE WASH CYCLE
STAGE COLD WATER - US GAL
(LITRES)
HOT WATER - US GAL
(LITRES)
TOTAL FOR STAGE - US
GAL (LITRES)
PRE-WASH 5.3 (20) 0 5.3 (20)
WASH 6.1 (23) 5.5 (21) 11.6 (44)
RINSE 0 5.3 (20) 5.3 (20)
DISINFECTION 0 5.3 (20) 5.3 (20)
INTERNAL DRAIN COOLING 16.4 (62) 0 16.4 (62)
TANK COOLING 1.1 (4) 0 1.1 (4)
TOTAL WATER CONSUMPTION 28.9 (109) 16.1 (61) 45 (170)
TITLE:
DWG NO.
DRAWN BY: LC SHEET 5 OF 8
A4
REVISION: A
13464836
PCF (INTERNAL DRAIN COOLING, UNITS BUILT FROM SN
11210713301 AND CONVERTED UNITS) EQUIPMENT
DRAWING
DATE: ECN NUMBER:107745 ECN DESCRIPTION: PRODUCTION RELEASE
This document contains confidential and proprietary information of STERIS Corporation.
Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
28/03/2022
A A
B B
C C
D D
E E
F F
Hly Highlight
ELECTRICAL CONNECTIONS
MODEL CONDUCTOR COLOUR AND MARKING
IWPCF, IW531066 & IW531065
(200-220V AC 3P+PE) SUPPLIED
WITH 9.84FT (3m) 4-CORE
POWER CABLE
PROTECTIVE EARTH GREEN/YELLOW
LIVE 1 BLACK 1
LIVE 2 BLACK 2
LIVE 3 BLACK 3
IW531055 & IW531056 (380 -
415V AC 3P+N+PE) SUPPLIED
WITH 9.84 FT (3m) 5-CORE
POWER CABLE
PROTECTIVE EARTH GREEN/YELLOW
LIVE 1 BROWN
LIVE 2 BLACK
LIVE 3 GREY
NEUTRAL BLUE
ELECTRICAL POWER SUPPLY REQUIREMENTS
REQUIREMENTS
MODEL
IWPCF IW531065 & IW531066 IW531055 & IW531056
ELECTRICAL SUPPLY
VOLTAGE
208V AC 3P+PE
(3-PHASE AND PROTECTIVE EARTH)
FREQUENCY: 60 Hz
200V AC 3P+PE
(3-phase and protective earth) IW531065 FREQUENCY: 50 Hz IW531066 FREQUENCY: 60Hz
380-415V AC 3P+N+PE (3-
PHASE, NEUTRAL AND
PROTECTIVE EARTH)
IW531055 FREQUENCY: 50Hz IW531056 FREQUENCY: 60Hz
SUPPLY CURRENT DEDICATED CIRCUIT IS REQUIRED WITH A CIRCUIT BREAKER RATING
OF 32A MIN. REFER TO LOCAL CODES.
DEDICATED CIRCUIT IS
REQUIRED WITH A CIRCUIT
BREAKER RATING OF 25A MIN.
REFER TO LOCAL CODES.
ELECTRICAL POWER
REQUIREMENT 10kW 13kW
ALLOWABLE SUPPLY
VOLTAGE
FLUCTUATION
10% OF NOMINAL VOLTAGE
OVER-VOLTAGE
CATEGORY II TRANSIENT OVER-VOLTAGES TYPICALLY PRESENT ON MAINS SUPPLY
TITLE:
DWG NO.
DRAWN BY: LC SHEET 6 OF 8
A4
REVISION: A
13464836
PCF (INTERNAL DRAIN COOLING, UNITS BUILT FROM SN
11210713301 AND CONVERTED UNITS) EQUIPMENT
DRAWING
DATE: ECN NUMBER:107745 ECN DESCRIPTION: PRODUCTION RELEASE
This document contains confidential and proprietary information of STERIS Corporation.
Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
28/03/2022
A A
B B
C C
D D
E E
F F
THE DISCONNECT SWITCH MUST:
ISOLATE ALL POLES SIMULTANEOUSLY•
BE LOCKABLE IN THE OFF POSITION•
COMPLY WITH STANDARDS IEC/EN 61058-1 AND IEC 60447 (CREEPAGE AND •
CLEARANCES FOR 4KV TRANSIENT VOLTAGE)
BE SITUATED CLOSE TO THE EQUIPMENT FOR EASE OF ACCESSIBILITY, AND IN A POSITION •
WHERE THE OPERATOR CAN TURN IT OFF IN AN EMERGENCY WITHOUT LEANING OVER THE
MACHINE
MARKED AS THE DISCONNECTION DEVICE FOR THIS EQUIPMENT•
DISCONNECT DEVICE OFF-POSITION SHOULD BE CLEARLY MARKED•
CABLES SHALL BE RATED FOR MAXIMUM CURRENT OF THE EQUIPMENT, MEETING THE
REQUIREMENTS OF IEC 60227 OR IEC 60245. THE INNOWAVE PCF IS SUPPLIED WITH 6FT OF
SCREENED POWER CABLE. THE POWER CABLE CAN BE HARD WIRED TO SUITABLE RATED
DISCONNECT SWITCH OR WIRED USING AN UL APPROVED PLUG AND SOCKET, FOLLOWING ALL
LOCAL ELECTRICAL CODES. PLUG AND SOCKET MUST BE WATERPROOF, AND RATED EQUAL TO
OR HIGHER THAN RECOMMENDED CIRCUIT PROTECTION. BECAUSE OF DIFFERING LOCAL CODES,
THE PLUG AND SOCKET ARE NOT INCLUDED WITH THE MACHINE.
OPERATING/ WORKING ENVIRONMENT LOCATION REQUIREMENTS
DESCRIPTION REQUIREMENTS
OPERATING SURFACE ON LEVEL FLOOR, CAPABLE OF SUPPORTING A WEIGHT OF AT
LEAST 530LB (240KG)
LOCATION REQUIREMENT ALTITUDE INDOORS ONLY, ALTITUDE UP TO 6562 FT (2000m)
AMBIENT OPERATING TEMPERATURE
RANGE 41 F TO 104 F (5 C TO 40 C)
MAXIMUM HUMIDITY 80% RELATIVE HUMIDITY FOR UP TO 88 ºF (31 ºC) DECREASING
LINEARLY TO 50% RELATIVE HUMIDITY AT 104 ºF (40 ºC)
SITE SERVICES
MACHINE CONNECTION POINTS WITHIN:
6 FT (1.8 m) OF ELECTRICAL ISOLATOR
3 FT (1.0M) OF HOT, COLD WATER CONNECTIONS
6FT (1.8M) OF EXTERNAL SITE DRAIN FOR EFFLUENT
AMBIENT LIGHTING 500 LUX (46.5 FOOT-CANDLE) MINIMUM ROOM LIGHTING
NOISE GENERATED DURING
OPERATION 68 dBA AT DISTANCE OF 3ft (1m) FROM FRONT OF UNIT
HEAT LOSS NON-DISINFECTION
CYCLE 2.6K BTU/HR
HEAT LOSS DISINFECTION CYCLE 5.8K BTU/HR
TITLE:
DWG NO.
DRAWN BY: LC SHEET 7 OF 8
A4
REVISION: A
13464836
PCF (INTERNAL DRAIN COOLING, UNITS BUILT FROM SN
11210713301 AND CONVERTED UNITS) EQUIPMENT
DRAWING
DATE: ECN NUMBER:107745 ECN DESCRIPTION: PRODUCTION RELEASE
This document contains confidential and proprietary information of STERIS Corporation.
Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
28/03/2022
A A
B B
C C
D D
E E
F F
TRANSPORTATION AND STORAGE
MAKE SURE PACKAGE MARKING INSTRUCTIONS AND ENVIRONMENTAL CONDTIONS ARE
ALWAYS FOLLOWED DURING TRANSPORT AND STORAGE OF INNOWAVE PCF.
LIFTING
LIFTING AND MANEUVERING THE INNOWAVE PCF IN SHIPPING CRATE MUST BE DONE WITH
PALLET TRUCK OR FORKLIFT TRUCK ABLE TO CARRY LOAD IN EXCESS OF 469.5lb (213kg)
STORAGE MARKINGS
THE MARKINGS ON SHIPPING CRATE STIPULATE UNIT'S STORAGE CONDITIONS AND HANDLING
INFORMATION.
SHIPPING INFORMATION
SHIPPING CRATE DIMENSIONS AND WEIGHT
SHIPPING CRATE HEIGHT 50.4" 1280mm
SHIPPING CRATE WIDTH 41.7" 1059mm
SHIPPING CRATE DEPTH 34.6" 879mm
SHIPPING CRATE GROSS WEIGHT 469.5lb 213kg
TITLE:
DWG NO.
DRAWN BY: LC SHEET 8 OF 8
A4
REVISION: A
13464836
PCF (INTERNAL DRAIN COOLING, UNITS BUILT FROM SN
11210713301 AND CONVERTED UNITS) EQUIPMENT
DRAWING
DATE: ECN NUMBER:107745 ECN DESCRIPTION: PRODUCTION RELEASE
This document contains confidential and proprietary information of STERIS Corporation.
Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
28/03/2022
A A
B B
C C
D D
E E
F F
| SECTION 22 67 19.16 |
| REVERSE-OSMOSIS WATER EQUIPMENT |
| INNOWAVE_PCF_EQUIPMENT_DRAWING_-_INTERNAL_DRAIN_COOLING__UNITS_BUILT_FROM_SN_11210713301_AND_CONVERTED_UNITS.pdf |
| Sheet1 |
| Drawing View1 |
| Drawing View2 |
| Drawing View3 |
| Sheet2 |
| Drawing View7 |
| Drawing View12 |
| Drawing View13 |
| Sheet3 |
| Drawing View14 |
| Drawing View15 |
| Drawing View16 |
| Sheet4 |
| Drawing View17 |
| Drawing View18 |
| Drawing View19 |
| Sheet5 |
| Sheet6 |
| Sheet7 |
| Sheet8 |
| AMSCO 7052HP SINGLE-CHAMBER WASHER-DISINFECTOR.pdf |
| Feuille 1 |
| Vues |
| new_view_35 |
| new_view_37 |
| new_view_57 |
| new_view_66 |
| Zones |
| A1 |
| Feuille 2 |
| Vues |
| new_view_55 |
| Zones |
| A1 |
| Feuille 3 |
| Vues |
| new_view_38 |
| new_view_39 |
| new_view_40 |
| Zones |
| A1 |
| Feuille 4 |
| Vues |
| new_view_44 |
| new_view_45 |
| new_view_46 |
| Zones |
| A1 |
| Feuille 5 |
| Vues |
| new_view_41 |
| new_view_42 |
| new_view_43 |
| new_view_65 |
| Zones |
| A1 |