A5 4152VL07 Techincal Ordering Data Sheet.pdf
PDF 114 KB Posted
- Attached to
- NUCLIDE GAMMA CALIBRATION
- Solicitation number
- 4152-VL00
About this file
This document is an Ordering Data Sheet for a sealed beta reference source. It provides detailed specifications for the required source, including the size, activity level, and calibration requirements. The key details are:
The source is a sealed beta source in the form of a steel disk with a Co-60 active element, an aluminum ring, and an aluminized Mylar window. The source activity is 100 nCi (+/- 10%) and must meet exempt quantity requirements per 10 CFR 30. The source must have NIST-traceable calibration, with activity and uncertainty values reported. Surface contamination levels on the source and packaging must be below 450 pCi/100cm2. The vendor must ship the source to the Puget Sound Naval Shipyard and make prior contact with the specified point of contact.
This Ordering Data Sheet is related to a pre-solicitation for "NUCLIDE GAMMA CALIBRATION" requirements, posted by the Department of the Navy Naval Supply Systems Command. The solicitation number is 4152-VL00. The agency is interested in connecting with potential vendors for this requirement.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
ORDERING DATA SHEET – GENERAL
NOUN NAME: RADIOACTIVE SOURCE
DESCRIPTION:
1 ea Sealed beta reference source in the form of a steel disk with active element surrounded by an aluminum ring with 0.9 mg/cm^2 aluminized Mylar window. Overall diameter of 47 mm; Active diameter of 41 mm; Height of 3.18 mm.
-The source activity is 100 nCi (–10%, +10%) Co-60.
-The vendor must insure that the source activity is not greater than exempt quantity per 10 CFR 30.
-The reported radioactivity uncertainty of each nuclide shall not exceed ± 5% at the 99% confidence level.
-Surface contamination levels on the source and packing materials must be less than 450 pCi/100cm^2.
MATERIAL: N/A
MANUFACTURER/VENDOR P/N: N/A
QUANTITY: 1 each
ADDITIONAL NOTES:
CERTIFICATION
-A Certificate of Calibration is required for the source and shall be provided from the vendor that documents:
-The source shall be National Institute of Standards and Technology (NIST) traceable
-The total activity of the source
-The activity of each nuclide in the source
-The uncertainty value, reference time and date
-The activity of any gamma emitting impurities detected
-The vendor assigned source serial number
-ISO Classification
-The results and date of leakage testing
SHIPPING REQUIREMENTS
Source will be shipped by vendor to:
(N4523A) Puget Sound Naval Shipyard & IMF
1400 Farragut Avenue, Building 514
Bremerton, WA 98314-5001
Attn: C/2380.4R - R. Nicole Van Den Eijkel (360) 801-6048
POC: C/105.55 Marc Sobolesky (360) 340-4797
Prior to shipping, vendor will make voice contact via phone with Nicole Van Den Eijkel or their representative at (360) 476-
3520 with date of shipment and shipment tracking information (i.e. UPS/FEDEX tracking numbers). A phone message will not be considered voice contact.
Changes to specification or other technical requirements are not allowed without Shipyard approval via the Administrative
Contracting Officer (ACO).
JOB ORDER
991566SOUR
KEY OP
H01
KO ITEM
REV
IDN
4152VL07
IDN REV
- PAGE 1 OF 1
2305.3 Form; Rev 02/2020