BFSU Addendum 01 Drawings.pdf
PDF 2 MB Posted
- Attached to
- BLACKFEET HOSPITAL CT ROOM RENOVATION
- Solicitation number
- 75H70125R00016
About this file
These are architectural and engineering design drawings for the Blackfeet Hospital CT Room Renovation project, dated February 16, 2024 with an Addendum on April 22, 2025. The comprehensive set of technical drawings includes detailed plans for renovating an existing CT scan room and adjacent spaces, covering architectural, mechanical, plumbing, electrical, and power system modifications. The drawings specify comprehensive changes including demolition of existing walls and finishes, installation of new lead-lined walls, reconfiguration of HVAC and medical gas systems, new electrical and power infrastructure, and specialized equipment placement for a new 128-slice CT scanner.
The renovation focuses on creating an appropriately sized and equipped CT scan room that meets 2018 Facility Guidelines Institute (FGI) standards and manufacturer recommendations. Key design elements include lead shielding, specialized electrical and mechanical systems, new floor trench ducts, medical gas connections, and integration of equipment like a wall-mounted air conditioner, humidifier, and medical equipment. The drawings provide precise technical specifications for contractors, including detailed schedules for grilles, registers, diffusers, electrical equipment, and mechanical components, ensuring a comprehensive approach to the medical facility upgrade.
View the file
Other files
Show all 15
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
CODE COMPLIANCE LEGEND
1-HOUR FIRE RATED PARTITION - UL DES U419
2-HOUR FIRE RATED BARRIER - UL DES U419
SMOKE BARRIER - NFPA 101 - 8.3
ROOM NAME
XXX SF
ROOM NAME
ROOM NUMBER
ROOM AREA
CODE ROOM TAG INFORMATION
PATH OF EGRESS
EXIT
OCCUPANT LOAD PER EXIT
OCCUPANT LOAD
EGRESS CORRIDOR
ROOM OCCUPANT LOADXX
AREA NOT IN SCOPE
1/2 HOUR FIRE RESISTIVE WALL
519 SF
CT SCAN
92 SF
SECURITY
58 SF
1556B
CHANGING
83 SF
1556A
CONTROL ROOM
114 SF
FAMILY CONSULT
AREA NOT IN SCOPE
CODES & REGULATIONS
AUTHORITY
HAVING
JURISDICTION
CODES AND REGULATIONS
• INTERNATIONAL BUILDING CODE - 2021 EDITION
• NFPA 10, STANDARD FOR THE PORTABLE FIRE
EXTINGUISHERS - 2022 EDITION
• NFPA 13, STANDARD FOR THE INSTALLATION OF
SPRINKLER SYSTEMS - 2022 EDITION
• NFPA 70, NATIONAL ELECTRIC CODE - 2020
EDITION
• NFPA 72, NATIONAL FIRE ALARM CODE - 2022
EDITION
• NFPA 80, STANDARD FOR FIRE DOORS AND OTHER
OPENING PROTECTIVES - 2022 EDITION
• NFPA 101, LIFE SAFETY CODE - 2021 EDITION
• NFPA 90A, STANDARD FOR THE INSTALLATION OF
AIR-CONDITIONING AND VENTILATING SYSTEMS -
2021 EDITION
• NFPA 99, HEALTH CARE FACILITIES CODE - 2021
EDITION
• NFPA 220, STANDARD ON TYPES OF BUILDING
CONSTRUCTION - 2021 EDITION
GENERAL BUILDING LIMITATIONS
CODE/REGULATED
ITEM
IBC
OCCUPANCY
CONSTRUCTION
TYPE
AUTOMATIC FIRE
SPRINKLER
NFPA 101
I-2, B
II
YES YES
ACTUAL
BUILDING HEIGHT
15' 4" EXISTING NO CHANGE
1 STORY NO CHANGESTORIES
DEAD END
CORRIDOR
COMMON PATH
OF TRAVEL
100' 18.2.5.3
EXISTING
HEALTHCARE, BUSINESS
LIFE SAFETY CODE ANALYSIS
GENERAL PROJECT DESCRIPTION
1. THE PROJECT SCOPE CONSISTS OF UPGRADING CURRENT
ROOM FINISHES AND EQUIPMENT
INTERNATIONAL BUILDING CODE - 2021
OCCUPANCY CLASSIFICATION
1. EXISTING HEALTHCARE AREA, EXISTING BUSINESS
ALLOWABLE HEIGHT
1. TABULAR HEIGHT - UNLIMITED [TABLE 504.3]
2. TABULAR NUMBER OF STORIES - UNLIMITED [TABLE 504.4]
ALLOWABLE AREA
1. TABULAR AREA - UNLIMITED [TABLE 506.2]
STRUCTURAL FIRE RESISTIVE
RATING OF BUILDING ELEMENTS
1. REQUIRED FIRE RESISTANCE RATINGS STRUCTURAL
ELEMENTS BASED ON
IBC TYPE IA CONSTRUCTION [TABLE 601 & TABLE 602]
STRUCTURAL ELEMENT FIRE RESISTANCE
RATING (HOURS)
BEARING WALLS EXTERIOR 3
INTERIOR 3
FIRE BARRIERS EXIT STAIRWAYS CONNECTING
FEWER THAN FOUR STORIES
EXIT STAIRWAYS CONNECTING
FOUR OR MORE STORIES
SHAFTS (OTHER THAN EXITS) AND
ELEVATOR HOISTWAY
NON-LOAD
BEARING
WALLS
EXTERIOR ( < 30 FT. SEPARATION) 1
0 ( ≥ 30 FT. SEPARATION)
INTERIOR 0
PRIMARY STRUCTURAL MEMBERS 3
FLOOR CONSTRUCTION (INCLUDING BEAMS) 2
ROOF
CONSTRUCTION
(INCLUDING BEAMS,
TRUSSES, ARCHES
AND ROOF DECK)
LESS THAN 20 FT. IN HEIGHT
TO LOWEST MEMBER
1-1/2
20 FT. OR MORE IN HEIGHT
TO LOWEST MEMBER
a, b c c a
NOT LESS THAN THE FIRE SEPARATION DISTANCE (IBC TABLE 602)
b
NOT LESS THAN FIRE RESISTANCE RATING OF THE WALL SUPPORTED
c
RATING MAY BE REDUCED BY 1-HOUR IF SUPPORTING ROOF ONLY
NFPA 101, LIFE SAFETY CODE - 2021
OCCUPANT USE LOAD FACTORS [TABLE 7.3.1.2]
1. ASSEMBLY USE (LESS CONCENTRATED): 15 SF/PERSON
2. BUSINESS USE: 100 SF/PERSON
3. BUSINESS USE (CONCENTRATED): 50 SF/PERSON
TRAVEL DISTANCE WITH AUTOMATIC
FIRE SPRINKLER PROTECTION
1. MAXIMUM TRAVEL DISTANCE
1.1. NEW BUSINESS: 300 FT. [§38.2.6.3]
1.2. EXISTING BUSINESS: 300 FT. [§39.2.6.3]
2. MAXIMUM COMMON PATH OF TRAVEL
2.1. NEW BUSINESS: 100 FT. [§38.2.5.3.1]
2.2. EXISTING BUSINESS: 100 FT. [§39.2.5.3.1]
3. MAXIMUM DEAD END CORRIDOR DISTANCE
3.1. NEW BUSINESS: 50 FT. [§38.2.5.2.1]
3.2. EXISTING BUSINESS: 50 FT. [§39.2.5.2.1]
OCCUPANCY SEPARATION
1. SEPARATED OCCUPANCY (BUSINESS AND HEALTHCARE)
1.1. EACH PART OF THE BUILDING COMPRISING OF A DISTINCT
OCCUPANCY SHALL BE COMPLETELY SEPARATED FROM OTHER
OCCUPANCIES BY FIRE-RESISTIVE CONSTRUCTION [§6.1.14.4.1]
HAZARDOUS AREA PROTECTION
1. PROTECTION FROM ANY AREA HAVING A DEGREE OF HAZARD
GREATER THAN THAT NORMAL TO THE GENERAL OCCUPANCY OF THE
BUILDING OR STRUCTURE SHALL BE PROVIDED BY ONE OF THE
FOLLOWING MEANS [§8.7.1.1]
1.1. 1-HOUR FIRE BARRIER
1.2. AUTOMATIC SPRINKLER SYSTEM
1.3. APPLYING BOTH WHEN HAZARD IS SEVERE
2. IN NEW CONSTRUCTION, WHERE PROTECTION IS PROVIDED WITH
AUTOMATIC SPRINKLERS WITHOUT FIRE-RESISTIVE SEPARATION, THE
SPACE SHALL BE ENCLOSED WITH SMOKE PARTITIONS [§8.7.1.2]
3. SMOKE PARTITION - CONTINUOUS MEMBRANE THAT IS DESIGNED TO
FORM A BARRIER TO LIMIT THE TRANSFER OF SMOKE. IT IS NOT
REQUIRED TO HAVE A FIRE RESISTANCE RATING [§3.3.256]
4. NO SEVERE HAZARDS ARE ANTICIPATED WITHIN THE SCOPE OF THE
PROJECT
FIRE RESISTANCE RATED SEPARATION
1. CORRIDOR WALL
1.1. 0-HOUR [ §38.3.6.1, §39.3.6.1]
2. HAZARDOUS AREA (GENERAL STORAGE, BOILER, FURNACE ROOM)
2.1. NEW BUSINESS: SMOKE PARTITION [§38.3.2.1, §8.7.1.2]
2.2. EXISTING BUSINESS: 0-HOUR [ §39.3.2.1, §8.7.1.2]
3. HIGH HAZARD CONTENTS
3.1. 1-HOUR [§38.3.2.1, §39.3.2.1, §8.7.1.1]
OPENING PROTECTIVE
1. BARRIER WALLS - 1-HOUR
1.1. 3/4-HOUR [TABLE 8.3.4.2]
2. BARRIER WALLS - 2-HOUR
2.1. 1 1/2-HOUR [TABLE 8.3.4.2]
3. SMOKE PARTITIONS - 1/2-HOUR
3.1. 1/3-HOUR [TABLE 8.3.4.2]
REQUIRED FIRE RESISTANCE RATING
1. ENCLOSURES CONNECTING ≥ 4 STORIES
1.1. 2-HOUR
2. OTHER ENCLOSURES
1.2. 1-HOUR
INTERIOR FINISH WITH AUTOMATIC
SPRINKLER PROTECTION
1. EXIT ACCESS CORRIDORS AND OTHER EXIT WAYS
1.1. EXISTING HEALTHCARE: CLASS A or B [§19.3.3.2]
1.2. EXISTING BUSINESS: CLASS C [§39.3.3.2.1]
2. ROOMS AND ENCLOSED SPACES
2.1. EXISTING HEALTHCARE:
2.2. EXISTING BUSINESS: CLASS C [§39.3.3.2.2]
3. FLOOR FINISH
3.1. EXISTING HEALTHCARE: CLASS II [§38.3.3.3.2]
3.2. EXISTING BUSINESS: CLASS II [§39.3.3.3.2]
FIRE AND SMOKE DAMPERS
1. FIRE DAMPERS
1.1. FIRE DAMPERS SHALL BE PROVIDED WHERE DUCTS PENETRATE
OR TERMINATE AT OPENINGS IN WALLS OR PARTITIONS REQUIRED
TO HAVE A FIRRESISTANCE RATING OF 2-HOURS OR MORE [NFPA
90A §5.3.1.1]
1.2. FIRE DAMPERS SHALL BE INSTALLED AT EACH DIRECT OR
DUCTED OPENING INTO AND OUT OF SHAFT ENCLOSURES [NFPA
90A §5.3.4.6]
2. SMOKE DAMPERS
2.1. SMOKE DAMPERS SHALL BE INSTALLED IN SYSTEMS WITH
CAPACITY GREATER THAN 15,000 CFM TO ISOLATE THE AIR-
HANDLING EQUIPMENT FROM THE REMAINDER OF THE SYSTEM
ON BOTH THE BUILDING SUPPLY SIDE AND THE RETURN SIDE
[NFPA 90A §4.3.10.2]
MEANS OF EGRESS COMPONENTS
1. DOORS
1.1. MINIMUM CLEAR WIDTH: 32 INCHES [§7.2.1.2.4]
1.2. SWING IN THE DIRECTION OF EGRESS WHEN SERVING AN
OCCUPANT LOAD GREATER THAN 50 OCCUPANTS [§7.2.1.4.2]
1.3. SWING IN DIRECTION OF EGRESS TRAVEL WHEN USED IN AN
EXIT ENCLOSURE OR WHERE SERVING A HIGH HAZARD
CONTENTS AREA [§7.2.1.4.2]
2. MARKING OF MEANS OF EGRESS
2.1. EXITS SHALL BE MARKED BY AN APPROVED SIGN READILY
VISIBLE FROM ANY DIRECTION OF EXIT ACCESS [§7.10.1.2.1]
2.2. SIGN PLACEMENT SHALL BE SUCH THAT NO POINT IN AN EXIT
ACCESS CORRIDOR IS IN EXCESS OF THE RATED VIEWING
DISTANCE OR 100 FT, WHICHEVER IS LESS, FROM THE NEAREST
SIGN [§7.10.1.5.2]
2.3. A DIRECTIONAL SIGN READING "EXIT" WITH A DIRECTIONAL
INDICATOR (CHEVRON STYLE) SHOWING THE DIRECTION OF
TRAVEL SHALL BE PLACED IN EVERY LOCATION WHERE THE
DIRECTION OF TRAVEL TO REACH THE NEAREST EXIT IS NOT
APPARENT [§7.10.2.1/7.10.3.1]
2.4. NEW SIGN PLACEMENT SHALL BE SUCH THAT NO POINT IN
AN EXIT ACCESS CORRIDOR IS IN EXCESS OF THE RATED
VIEWING DISTANCE OR 100 FT. TO THENEAREST SIGN [§
7.10.1.5.2]
3. EXIT SIGN ILLUMINATION
3.1. EVERY EXIT SIGN SHALL BE SUITABLY ILLUMINATED BY A
RELIABLE LIGHT
SOURCE. EXTERNALLY AND INTERNALLY ILLUMINATED SIGNS
SHALL BE LEGIBLE IN BOTH THE NORMAL AND EMERGENCY
LIGHTING MODE [§7.10.5.1]
3.2. THE ILLUMINATION SHALL BE FOR 90 MINUTES IN THE CASE OF
PRIMARY POWER LOSS [§7.9.2.1]
3.3. EXIT SIGNS THAT ARE EXTERNALLY ILLUMINATED SHALL HAVE
AN INTENSITY NOT LESS THAN 5 FT-CANDLES [§7.10.6.3]
4. NUMBER AND ARRANGEMENT OF MEANS OF EGRESS
4.2. OCCUPANT LOAD OF 501 - 1,000: 3 EXITS [§7.4.1.2]
4.3. OCCUPANT LOAD OF MORE THAN 1,000: 4 EXITS [§7.4.1.2]
5. ARRANGEMENT OF EXITS
5.1. WHERE TWO OR MORE EXITS ARE REQUIRED, THEY MUST BE
SPACED APART A MINIMUM OF 1/3 THE LENGTH OF THE MAXIMUM
DIAGONAL OF THE BUILDING OR AREA BEING SERVED [§7.5.1.3.3]
4.1. OCCUPANT LOAD OF 1 - 500: 2 EXITS [§7.4.1.1]
4. HEALTH CARE USE (INPATIENT TREATMENT): 240 SF/PERSON
5. HEALTH CARE USE AMBULATORY): 150 SF/PERSON
6. STORAGE, MECH. AND ELEC. AREAS: 500 SF/PERSON
30' 18.2.5.3
BUILDING AREA EXISTING BLDG
MAIN LEVEL = 90,000 S.F.
CLASS A or B [§19.3.3.2]
SEISMIC
SEISMIC RISK CATEGORY: III
0 8' 16' 32'
600 E. Michigan Ave., Suite B
Kalamazoo, MI 49007
P: 269.927.0144 www.7GenAE.com
ARCHITECTURE + ENGINEERING
SEVEN
GENERATIONS
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
KEYPLAN
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
G201.1
FIRST FLOOR CODE
COMPLIANCE PLAN
BID DOCUMENTS02/16/24
PROJECT A
NORTH
SCALE: 1/16" = 1'-0"
FIRST FLOOR CODE
COMPLIANCE PLAN
04/22/25 ADDENDUM 01
A1
DEMOLITION KEY NOTES
REMOVE EXISTING WALL CONSTRUCTION
COMPLETE AS SHOWN. INCLUDING WALL
PROTECTION, WALL BASE, AND ALL ITEMS ATTACHED
TO THE WALLS (NOT ALL ITEMS ATTACHED TO THE
WALLS ARE SHOWN ON THIS PLAN). PATCH &
REPAIR ADJACENT SURFACES TO REMAIN AS
NECESSARY TO PREPARE FOR NEW CONSTRUCTION
AND / OR NEW FINISHES.
REMOVE EXISTING WALL CONSTRUCTION AND LEAD
LINED GYPSUM BOARD AS SHOWN. PATCH &
REPAIR ADJACENT SURFACES TO REMAIN AS
NECESSARY TO PREPARE FOR NEW CONSTRUCTION
AND/OR NEW FINISHES.
REMOVE & SALVAGE EXISTING DOORS(S), FRAME(S),
HARDWARE, ETC. TURN OVER TO COR. AND NOTIFY
COR UNLESS DIRECTED OTHERWISE.
REMOVE & SALVAGE EXISTING DOORS(S), FRAME(S),
HARDWARE, ETC FOR REINSTALLATION.
REMOVE WINDOW FRAME AND LEADED GLASS.
REMOVE EXISTING WALL FINISHES, WALL BASE, AND
ALL ITEMS ATTACHED TO WALLS. PATCH & REPAIR
EXPOSED SURFACES TO REMAIN AS REQUIRED FOR
NEW CONSTRUCTION AND/OR NEW FINISHES.
REMOVE EXISTING FLOORING AND FLOOR FINISHES
INCLUDING ALL ADHESIVES, SETTING MATERIALS
DOWN TO EXISTING SUBSTRATE. PREPARE CONCRETE
FLOORS AS REQUIRED TO PROVIDE A SMOOTH AND
UNIFORM SURFACE FOR NEW FINISHES.
REMOVE EXISTING CEILING TILE AND GRID SYSTEM.
REMOVE EXISTING CEILING MOUNTED LIGHT
FIXTURES (UNLESS NOTED IN CONJUNCTION),
SPEAKERS CAMERAS, FIRE ALARMS, MECHANICAL
DEVICES, SMOKE DETECTORS, SENSORS, WIRELESS
ACCESS POINT, EXIT SIGNAGE, SAFETY MIRRORS,
ETC. TURN OVER TO THE COR. OR SALVAGE TILE FOR
RECYCLING UNLESS DIRECTED OTHERWISE.
REMOVE EXISTING CASEWORK & COUNTERTOP
WITH WALL MOUNTED EQUIPMENT. SALVAGE
EQUIPMENT & TURN OVER TO COR. UNLESS
DIRECTED OTHERWISE. PATCH & REPAIR EXPOSED
SURFACES TO REMAIN AS REQUIRED TO RECEIVE
NEW CONSTRUCTION AND FINISHES.
REMOVE EXISTING MECHANICAL, PLUMBING &
ELECTRICAL ITEMS - REFER TO MECHANICAL,
PLUMBING & ELECTRICAL DEMOLITION PLANS FOR
FURTHER INFORMATION. PATCH AND REPAIR
EXPOSED SURFACES TO REMAIN AS REQUIRED TO
RECEIVE NEW CONSTRUCTION AND FINISHES.
THE CT VENDOR WILL DISASSEMBLE AND REMOVE
THE EXISTING CT EQUIPMENT AND ASSOCIATED
EQUIPMENT (GANTRY, CONSULT DESK, PATIENT
TABLE, POWER DISTRIBUTION UNIT, SERVICE
CABINET, PARTIAL UPS, REAR CABLE COVER).
REMOVE EPOXY PAD, CUT AND GRIND THE EXISTING
ANCHORS FLUSH WITH THE EXISTING FLOOR.
REMOVE EXISTING ISOLATION PODS. SALVAGE
EQUIPMENT & TURN OVER TO COR. REFER TO
MECHANICAL AND ELECTRICAL DEMOLITION
PLANS.
EXISTING WALL FRAMING TO REMAIN. REMOVE
GYPSUM BOARD AND PREPARE FOR NEW FINISHES.
EXISTING WALL FRAMING TO REMAIN. REMOVE
LEAD LINING AND GYPSUM BOARD LAYERS AS
NEEDED TO PREPARE FOR NEW FINISHES AND NEW
WORK. CONTRACTOR TO ASSUME EXISTING LEAD
SHIELDING EXTENDS ABOVE 7'-0" ABOVE FINISH
FLOOR AND IS ALSO CONTINUOUSLY LAMINATED
TO UNDERSIDE OF BAR JOIST. CONTRACTOR TO
PATCH AND INFILL ALL WALL AND CEILING LEAD
LINING INCLUDING AT UNDERSIDE OF EXISTING
STRUCTURE ABOVE.
REMOVE EXISTING HANDRAILS. PROVIDE BLOCKING
AS NEEDED FOR NEW HANDRAILS.
REMOVE AND SALVAGE INJECTOR, HEAD END
MONITOR, CONTROL ARM AND BRACING.
CONTRACTOR IS TO SALVAGE FOR REINSTALLATION
IN NEW LAYOUT, INCLUDING CABLING.
REMOVE ALL WALL MOUNTED ITEMS (SUCH AS PIN
BOARDS, HAND SANITIZERS, TVS, CLOCKS ETC.) TURN
OVER ALL ITEMS TO THE COR. PATCH AND REPAIR
EXPOSED SURFACES TO REMAIN AS REQUIRED TO
RECEIVE NEW CONSTRUCTION AND FINISHES.
REMOVE ALL LOOSE AND WALL MOUNTED
FURNITURE. TURN OVER ALL ITEMS TO COR. PATCH &
REPAIR ADJACENT SURFACES TO REMAIN AS
NECESSARY TO PREPARE FOR NEW CONSTRUCTION
AND/OR NEW FINISHES.
SAWCUT AND REMOVE A PORTION OF EXISTING
CONCRETE SLAB AND GRADE BASE IN PREPARATION
FOR NEW ELECTRICAL CONDUIT TRENCH.
SAWCUT AND REMOVE A PORTION OF EXISTING
CONCRETE SLAB AND GRADE BASE IN PREPARATION
FOR NEW FLOOR DRAIN AND SANITARY
LINE/CONNECTIONS. EXACT LOCATION TO BE
DETERMINED IN FIELD, THROUGH GROUND-
PENETRATING RADAR OR OTHER METHODS.
AREA NOT IN SCOPE
LINE OF EXISITING FOUNDATION WALLS
1A
WAITING
SECURITY
1556A
CONTROL 1556
CT SCAN
ORTHO
CORRIDOR
CORRIDOR
CORRIDOR
1A
1A
6 7
PORTION OF CEILING
TO BE REMOVED
AND REPLACED.
EXTENT OF CEILING
DEMOLITION ABOVE
TRENCH
1'-2" 2'-0"
/2
/2
/8
8'-9 1/4"
1'-2" VERIFY W/ CT MANUFACTURER
2'-7 1/2" VERIFY W/ CT MANUFACTURER
EXACT DIMENSIONS TO BE COORDINATED IN FIELD W/ CT VENDOR
30'-10 5/8" +/-
/4
14 13
6 76
ED.7D.2D
8 8
REMOVE DECORATIVE
COLUMN
REMOVE GLASS
BLOCK WALL
REMOVE COAT
HANGER ON BACK
OF DOOR, DOOR TO
REMAIN
REMOVE
ROOM SIGN
17 18
17 18
REMOVE ROOM
SIGN & ACCESS
BUTTON
/8
EXISTING MED GAS TO REMAIN.
CONTRACTOR TO REMOVE COVER
PLATES (STORE FOR REINSTALLATION)
TO FACILITATE GYP BOARD
DEMOLITION.
SLAB ON GRADE
EXISTING CRAWLSPACE
SLAB ON GRADE EXISTING CRAWLSPACE
SLAB ON GRADE
EXISTING CRAWLSPACE
EXISTING CEILING TO REMAIN
REWORK EXISTING GRID AND
TILES AS REQUIRED TO
ACCOMODATE NEW CEILING
REMOVE EXISTING
WALL PAPER BORDER
REMOVE "X-RAY
IN USE" LIGHT
PROTECT EXISTING ADJACENT
WALL TILE TO REMAIN
REMOVE EXISTING COILING
SECURITY WINDOW COMPLE TE
INCLUDING ALL TRACKS,
BRACKETS, WALL SUPPORT AND
HEADER
A1
A1
GENERAL DEMO NOTES
1. ALL DEMOLITION WORK REQUIRED IS NOT NECESSARILY LIMITED
TO WHAT IS SHOWN ON THE DEMOLITION PLANS. THE INTENT IS
TO REMOVE ALL MATERIALS AND ITEMS AS REQUIRED TO
FACILITATE NEW CONSTRUCTION.
2. CONTRACTORS ARE TO VERIFY FIELD CONDITIONS, IF
CONDITIONS DO NOT REFLECT WHAT IS NOTED IN THE
CONTRACT DOCUMENTS, THE CONTRACTOR IS TO NOTIFY THE
ARCHITECT.
3. THIS PLAN IS TO BE USED IN CONJUNCTION WITH THE ENTIRE SET
OF CONSTRUCTION DOCUMENTS. DO NOT REMOVE ANY ITEMS
WITHOUT VERIFYING AND COORDINATING WITH ALL GENERAL
TRADES AS TO HOW THEY RELATE TO THE OVERALL PROJECT.
4. CONTRACTOR IS RESPONSIBLE FOR PROVIDING DUMPSTERS.
LOCATION OF DUMPSTERS TO BE DETERMINED BY COR. WITH
CONTRACTORS INPUT.
5. CONTRACTOR SHALL REMOVE AND DISPOSE OF ALL
DEMOLISHED MATERIALS. DO NOT ALLOW MATERIALS AND
DEBRIS GENERATED BY DEMOLITION ACTIVITIES TO
ACCUMULATE ON THE JOBSITE. REMOVE DAILY AND DISPOSE
OF IN A LEGAL MANNER. NO ONSITE SALE OR BURNING OF
REMOVED ITEMS IS PERMITTED.
6. THE CONTRACTOR MUST TAKE ALL NECESSARY PRECAUTIONS
TO ENSURE THE SAFETY OF THE PUBLIC AND/OR WORKMEN ON
THE SITE TO PREVENT ACCIDENTS OR INJURY TO ANY PERSON
ON, ABOUT OR ADJACENT TO THE PREMISES. THE CONTRACTOR
SHALL COMPLY WITH ALL LAWS, ORDINANCES, CODES AND
REGULATIONS PERTAINING TO SAFETY AND THE PREVENTION OF
ACCIDENTS.
7. THE CONTRACTOR MUST MAINTAIN ADEQUATE SUPPORT,
INSULATION, WATERPROOFING, EMERGENCY LIGHTING,
SECURITY, ALARMS, ETC. FOR ALL OR PART OF ITEMS WHICH
ARE TO REMAIN.
8. THE COR. RESERVES THE RIGHTS TO ANY REMOVED MATERIALS
OR ITEMS. CONTRACTOR TO COORDINATE WITH COR.
9. DO NOT CUT AND/OR PATCH STRUCTURAL ELEMENTS IN A
MANNER THAT COULD CHANGE THEIR LOAD-CARRYING
CAPACITY OR LOAD-DEFLECTION RATIO.
10. PATCH/REPAIR OR REPLACE ALL AREAS DAMAGED BY
DEMOLITION, REWORK AS REQUIRED, PREPARE FOR NEW
CONSTRUCTION AS REQUIRED. BRICK OR BLOCK PATCHING
SHALL BE TOOTHED-IN TO MATCH EXISTING COURSING
PATTERN.
11. WHERE WALLS & CEILINGS ARE TO BE REMOVED, ALL ITEMS
THAT FALL WITH THE WALLS &/OR CEILINGS ARE TO BE
REMOVED OR REROUTED. REFER TO MECHANICAL, ELECTRICAL,
AND PLUMBING PLANS FOR ADDITIONAL DEMOLITION NOT
SHOWN ON THIS SHEET.
12. CONSTRUCTION AREAS TO BE COMPLETELY SEPARATED FROM
OCCUPIED AREAS WITH A ONE-HOUR FIRE RATED ASSEMBLY,
AND WITH SUITABLE DUST BARRIER. TEMPORARILY SEAL OFF ALL
RETURN & SUPPLY AIR DIFFUSERS IN CONSTRUCTION AREA.
PROVIDE RECIRCULATING HEPA FILTER IN CONSTRUCTION
AREA. ALL DEMOLISHED MATERIAL IS TO BE REMOVED FROM
SITE IN A COVERED CONTAINER. NEGATIVE AIR PRESSURE TO BE
MAINTAINED WITHIN CONSTRUCTION ZONE.
13. IF HAZARDOUS MATERIALS ARE ENCOUNTERED DURING
DEMOLITION, CONTRACTOR SHALL CONTACT COR. FOR
ABATEMENT. (LEAD GYP. BACKING IS TO BE REMOVED BY
CONTRACTOR AND DOES NOT REQUIRE COR. ABATEMENT).
14. REMOVE, REPLACE, PATCH, AND REPAIR MATERIALS AND
SURFACES CUT OR DAMAGED DURING DEMOLITION, BY
METHODS AND WITH MATERIALS SO AS NOT TO VOID EXISTING
WARRANTIES.
15. CLEAN AREAS AND SPACES WHERE CUTTING AND PATCHING
ARE PERFORMED.
16. CONTRACTOR SHALL PATCH, REPAIR, AND INFILL ALL APPLIED
FIREPROOFING MATERIAL IN AREAS LEFT UNPROTECTED DUE TO
REMOVAL OF EXISTING ELEMENTS OR AREAS DAMAGED
DURING NEW WORK AND DEMOLITION. MATCH EXISTING
FIREPROOFING MATERIAL TYPE AND THICKNESS TO MAINTAIN
REQUIRED LEVEL OF PROTECTION.
0 2' 4' 8'
600 E. Michigan Ave., Suite B
Kalamazoo, MI 49007
P: 269.927.0144 www.7GenAE.com
ARCHITECTURE + ENGINEERING
SEVEN
GENERATIONS
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
KEYPLAN
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
AD101.1
FIRST FLOOR DEMOLITION
PLAN
BID DOCUMENTS02/16/24
04/22/25 ADDENDUM 01
PROJECT A
NORTH
SCALE: 1/4" = 1'-0"
FIRST FLOOR DEMOLITION PLAN
04/22/25 ADDENDUM 01
SECURITY
13'-4"
FAMILY CONSULT
1556A
CONTROL ROOM
1556B
CHANGING
CT SCAN
/8
/2
LE
A
D G
L A
S S
/8
50'-5 15/16" +/-
A420.101
101A
104A420.1
A420.1
CORRIDOR
AREA NOT IN SCOPE
9'-0"
D4
E6
C4
A6
A4
A4
C4
10'-0"
A420.1
A420.1
A420.1 08
A420.1
A4
G6
F12
E6
G4
G4
ELEVATOR
ED.7D.2D
E6
A
B
A420.1
F12
D4
09 11
H4H4
H4
H4
H4 H4
H4
H4
EXSITING WALL FRAMING TO REMAIN
PROVIDE NEW FINISHES ON BOTH SIDES OF
WALL FRAMING. NEW FINISHES TOBE
CONSISTENT WITH WALL TAGS ON EITHER SIDE
1'-4" lOCATION FOR FUTURE SECURITY MONITORS
B6
9'-6"
CLEAR
3'-0"
A1
A1
A1
A420.1
A420.1
SIM
FLOOR PLAN LEGEND
EXISTING WALL
NE WALL
EXISTING DOOR
NEW DOOR
1-HOUR FIRE RATED PARTITION - UL DES U419
2-HOUR FIRE RATED BARRIER - UL DES U419
1-HOUR SMOKE BARRIER - NFPA 101 - 8.3
NEW CONCRETE SLAB, DEPTH TO MATCH
EXISTING, WITH WELDED WIRE FABRIC ON
NEW VAPOR BARRIER ON COMPACTED
GRANULAR FILL ON COMPACTED
SUBGRADE. REFER TO PLUMBING DRAWINGS.
ABBREVIATIONS:
A = ARCHITECT
C = CONTRACTOR
G = GOVERNMENT
O = OWNER
V = VENDOR
X = ITEM TO BE INCLUDED WITH ASSOCIATED EQUIPMENT
A420.101
A420.1
A420.1
04 19
A420.1
A420.1 0807
ED.7D.2
CT TABLE
CEILING MOUNTED
INJECTOR
BENCHMIR
RH1UPS
PTD1
SD1
PDU1
MDC1
FAMILY CONSULT
1556A
CONTROL ROOM
1556B
CHANGING
CT SCAN
50'-5 15/16" +/-
IN USE LIGHT
IN USE LIGHT
AR1
HS1
ST1
CLK1
SH1
GLB1
ISO POINT
9'-0 5/8"
BW
HS1
PB1
CB
LKR LKR
/2
PP PP
CG1
CG1
CG1
CG1
CG1
A420.1
CT FLOOR RACEWAY
SYSTEM NOTES
WALL RACEWAY FROM FLOOR TO ABOVE CEILING.
WALL RACEWAY SHALL CONNECT TO CEILING
RACEWAY AND FLOOR TRENCH DUCT. RACEWAY SHALL
BE SIZED BY CT MANUFACTURER WITH (1) DIVIDERS.
FLOOR TRENCH DUCT 3-1/2"H X 10"W WITH (1) DIVIDERS,
REMOVABLE COVERS AND MOISTURE GASKETS.
A
B
GENERAL NOTES
1. INTERSECTION BETWEEN ALL RATED WALL SYSTEM AND
DECK SYSTEM TO BE SEALED WITH APPROVED UL FIRE SPRAY
MATERIAL.
2. REFER TO CODE COMPLIANCE FOR RATED WALL LOCTIONS.
3. PROVIDE UL RATED WALLS WHERE SHOWN. REFER TO WALL
TYPES FOR WALL CONSTRUCTION.
4. ALL COUNTERTOP OUTSIDE CORNERS TO BE 2" RADIUS
UNLESS NOTED OTHERWISE.
5. PROVIDE METAL BLOCKING WITHIN WALL CAVITY AS
REQUIRED FOR WALL MOUNTED EQUIPMENT AS SHOWN ON
ENLARGED PLANS AND ON INTERIOR ELEVATIONS.
COORDINATE FINAL LOCATIONS WITH OWNER.
6. ALL WALL BOARD CARRYING CERAMIC TILE SHALL BE
CEMENT BOARD - REFER TO FINISH SCHEDULE.
7. ALL WALLS CARRYING PLUMBING FIXTURES SHALL BE 5/8"
MOISTURE RESISTANT FIRECORE GYP BOARD UNLESS NOTED
OTHERWISE.
8. CONTRACTOR SHALL PATCH, REPAIR, AND INFILL ALL
APPLIED FIREPROOFING MATERIAL IN AREAS LEFT
UNPROTECTED DUE TO REMOVAL OF EXISTING ELEMENTS
OR AREAS DAMAGED DURING NEW WORK AND
DEMOLITION. MATCH EXISTING FIREPROOFING MATERIAL
TYPE AND THICKNESS TO MAINTAIN REQUIRED LEVEL OR
PROTECTION.
9. CONTRACTOR TO INSTALL 1/16" LEAD EQUIVALENT
BACKING ON ALL VOIDS AND PENTRATIONS IN THE WALL
SHIELDING.
10. ALL INTERIOR DIMENSIONS ARE TO FACE OF DRYWALL.
REFER TO WALL TYPES ON A001 FOR ADDITIONAL WALL
INFORMATION.
11. THE CONCRETE FLOOR MUST BE LEVEL WITHIN 3/8"
MAXIMUM DEVIATION IN THE AREA AROUND THE BASE OF
THE GANTRY AND THE BASE OF THE PATIENT TABLE. PROVIDE
CEMENTITIOUS FLOOR LEVELER AS REQUIRED.
12. WHERE FIRE RATED SEPARATIONS ARE SHOWN IN CODE
PLANS PROVIDE STENCILING EVERY 20'-0'' HORIZONTALLY
WITHIN THE CEILING PLENUM SPACE WITH THE FOLLOWING
13. PROVICE GYP BOARD CONTROL JOINTS ON BOTH SIDES OF
THE DOOR. TYPICAL AT ALL NEW DOORS.
0 2' 4' 8'
600 E. Michigan Ave., Suite B
Kalamazoo, MI 49007
P: 269.927.0144 www.7GenAE.com
ARCHITECTURE + ENGINEERING
SEVEN
GENERATIONS
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
KEYPLAN
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
A101.1
FIRST FLOOR PLAN
BID DOCUMENTS02/16/24
PROJECT A
NORTH
SCALE: 1/4" = 1'-0"
FIRST FLOOR PLAN - CT ROOM
RENOVATION
EQUIPMENT SCHEDULE - A
ABBREV. EQUIPMENT DESCRIPTION S
P
E C
IF
IE
D B
Y
S U
P P
L
IE
D B
Y
IN
S T A
L L E D
B Y
E L E C
D A
T A
C A
B L E T
V
B L O
C K
IN
G
W A
T E R
D R
A
IN
BASIS OF DESIGN -
MANUFACTURER
BASIS OF DESIGN -
MODEL REMARKS
AR1 APRON RACK A C C X PHILLIPS SAFETY 7 APRON PEG RACK
BW BLANKET WARMER G G G X MAC MEDICAL MODEL DWC243074T
CB CODE BLUE CALL BUTTON A C C X REFER TO LOW VOLTAGE DRAWINGS
CG1 CORNER GUARD A C C CONSTRUCTION SPECIALTIES REFER TO FINISH KEY
CL1 BARIATRIC PATIENT CEILING LIFT A C C X X ARJO / GULDMANN
CLK1 CLOCK G G C X X X
GLB1 GLOVE BOX G G C X BOBRICK
HS1 HAND SANITIZER G G C X PURELL
MDC1 MAIN DISCONNECT CONTROL G C C GE
MIR MIRROR: 24"x60" A C C X BOBRICK B-167 SERIES
PB1 PATIENT TRANSFER BOARD A C C X GE
PDU1 POWER DISTRIBUTION UNIT G C C X X GE PROVIDED BY CT EQUIPMENT
SUPPLIER
PP BARRIER FREE PUSH PLATE A C C X REFER TO DOOR HARDWARE
PTD1 PAPER TOWEL DISPENSER G G C X
RH1 ROBE HOOK A C C X BRADLEY CORPORATION 9B1-110200 DOUBLE HOOK
SD1 SOAP DISPENSER G G C X
SH1 SHARPS CONTAINER G G C X BOBRICK 305017
ST1 STRETCHER G G G
NORTH
SCALE: 1/4" = 1'-0"
FIRST FLOOR EQUIPMENT PLAN
AR1
PDU1
PTD1 SD1GLB1
MDC1
A520.1
A520.1
A701.1
A701.1
A520.1
6'-0"
F.E.P.
1'-3"
F.E.P.
1'-6" 2'-1"
W P
-1
S E E S
C H
E D
U
LE
SCHD.
SEE
S E E S
C H
E D
U
LE
S E E S
C H
E D
U
LE
SCHD.
SEE
SCHD.
SEE
2"
SCHD.
SEE 2"
S E E S
C H
E D
U
LE
A B C D1
DOOR TYPES
SCALE: 1/4" = 1'-0"
FRAME TYPES
SCALE: 1/4" = 1'-0"
6" 4"
6'-0"
LEAD GLAZING
FLOOR
LINE
6"6"
SCHD.
SEE
S E E S
C H
E D
U
LE
MIRMIR
A520.1
A520.1
FILLER
1' - 0"
F.E.P.
1' - 0"
F.E.P.
1' - 11"
1'-6" 2'-0"
11 1/8" 6'-0"
HS1
LEAD GLAZING AND
LEAD LINED HOLLOW
METAL FRAME
A420.1
CL1
W P
-1
CG1
OAV
A701.1
A701.1 20
A420.1
F.E.P.
3'-0"FILLER
CG1
A520.1
BW
CG1
S2
LEAD BACKED GYP. BD. INSTALLED
CONTINUOUS FROM FLOOR TO 7'-0"
ABOVE FINISH FLOOR.
RADIATION
ROOM SIDE
CONTROL
ROOM SIDE
16 GA. STEEL FRAME, 2 PIECE
TELESCOPIC WINDOW FRAME AND
STOP, W/ MITER CORNERS.
2"
1/16" LEAD SHIELDING AT INSIDE
OF FRAME INCLUDING PROFILE
LEAD LINED GLASS, 1/16"
LEAD EQUIVALENT
FACTORY SET STOPS
3/8"
SPECIFICATION NOTES:
1. THE LEAD SHIELDING VALUE TO BE THE SAME
AS THE SURROUNDING WALL, PARTITION OR
CEILING. (SEE PLANS OR PHYSICIST RADIATION
SHIELDING REPORT FOR LEAD THICKNESS
VALUES).
2. ALL SHEET LEAD IS 99.9% PURE MEETING
FEDERAL SPC QQ-L-201F GRADE C AND ASTM
B749 TYPE L51121 AND ALL APPLICABLE NCRP
REPORTS.
3. ENSURE IF LEAD IS PENETRATED IT SHALL HAVE
A CONTINUOUS LEAD SHIELDING WRAP
CL1
PB1
SH1
CG1
HS1
OAV
W P
-1
A701.1
A701.1
MDC1
CL1
C4
CB
CG1
PDU1
A520.1
RAKKS BRACKET; 3'-0" O.C. TYP. FINAL
LOCATION TO BE COORDINATED WITH OWNER.
COUNTERTOP W.
GROMMET; COORD.
LOCATION W. OWNER 2
FILLER 2'-6"
F.E.P.
2'-6"
A520.1
A520.1
FILLER 2'-6"
F.E.P.
2'-6"
W P
-1
S1
BG-1; BUMPER GUARD
HR-1; HANDRAIL
S1
A701.1
A701.1
W P
-1
CG1
HR-1; HANDRAIL
BG-1; BUMPER GUARD
A701.1
A701.1
W P
-1
CG1
S1 HR-1; HANDRAIL
BG-1; BUMPER GAURD
A701.1
A701.1
A420.1
/2
8'-3 5/8"
LEAD GLAZING AND
LEAD LINED HOLLOW
METAL FRAME
A520.1
RAKKS BRACKET;
36" O.C. MAX
PAINTED WRAP-AROUND HOLLOW
METAL DRYWALL FRAME. LEAD
LINED HOLLOW METAL FRAME AT
LEAD LINED DOORS.
OPTIONAL SECURITY ANCHOR AT
HEAD FOR FRAMES WIDER THAN 3'-0"
SEALANT CONTINUOUS @
PERIMETER (BOTH SIDES)
LT. GA. METAL BOX HEADER
CLOSER AS REQ'D - (SEE SCHED.)
PAINTED WRAP-AROUND
HOLLOW
METAL DRYWALL FRAME. LEAD
LINED HOLLOW METAL FRAME
AT LEAD LINED DOORS.
JAMB ANCHORS (2 PER JAMB-
& ADJ. BASE ANCHORS - TYP.)
DOUBLE METAL STUD
@ EA. JAMB
SEALANT CONTINUOUS @
PERIMETER (BOTH SIDES)
ADJACENT WALL
(WHERE OCCURS)
CG1CG1
HR-1; HANDRAIL BG-1; BUMPER GUARD
W P
-1
S2 S2 S2
A701.1
A701.1
/2
V.I.F. 10" V.I.F.
EXISTING 5" CONC. SLAB. PATCH
AND REPAIR TO ALLOW A
UNIFORM TRANSITION TO THE
NEW TRENCH
ELECTRICAL TRENCH DUCT.
REFER TO ELECTRICAL AND
CT EQUIPMENT SUPPLIER
NEW CONCRETE
VAPOR BARRIER
COMPACTED GRANULAR FILL
ON COMPACTED SUBGRADE
#4 DOWEL (4" @ 1'-6" O.C.)
A520.1
RAKKS BRACKET; 3'-0" O.C. TYP.
FINAL LOCATION TO BE
COORDINATED WITH OWNER.
WELDED WIRE FABRIC
#4 DOWEL (4" @ 1'-6" O.C.)
NEW CONCRETE
VAPOR BARRIER
COMPACTED GRANUAL FILL
(NEW PLUMBING NOT SHOWN)
600 E. Michigan Ave., Suite B
Kalamazoo, MI 49007
P: 269.927.0144 www.7GenAE.com
ARCHITECTURE + ENGINEERING
SEVEN
GENERATIONS
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
KEYPLAN
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
A420.1
INTERIOR ELEVATIONS
AND DOOR SCHEDULE
BID DOCUMENTS02/16/24
04/17/25 ADDENDUM 01
PROJECT A
DOOR AND FRAME SCHEDULE BP1
DOOR
NUMBER SIZE
DOOR FRAME FIRE RATING
(MIN.) HARDWARE SET REMARKSTYPE MATERIAL TYPE MATERIAL
101 (2) 3'-0" D HM 1 HM 90 MIN. 04 LEAD LINED FRAME AND DOOR, AUTO OPERATOR, PANIC BARS.
101A 4'-0" X 7'-0" A 1 HM 01 OWNER PROVIDED LEAD LINED DOOR.CONTRACTOR INSTALLED.
103 3'-0" X 7'-0" B WD 1 HM 20 MIN. 02
104 3'-0" X 7'-0" A WD 1 HM 20 MIN. 03
105 (2) 3'-0" x 7'-0" B WD 1 HM 20 MIN. 05 AUTO OPERATOR, PANIC BARS, CARD READER ON CORRIDOR 1559 SIDE.
SCALE: 1/4" = 1'-0"A420.1
CT SCAN - 10101
SCALE: 1/4" = 1'-0"A420.1
LOCKER ROOM 10406
SCALE: 1/4" = 1'-0"A420.1
CT SCAN - 10103
SCALE: 1/4" = 1'-0"A420.1
CT SCAN - 10105
SCALE: 1 1/2" = 1'-0"A420.1
WINDOW FRAME DETAIL19
SCALE: 1/4" = 1'-0"A420.1
CT SCAN - 10102
SCALE: 1/4" = 1'-0"A420.1
CT SCAN - 10104
SCALE: 1/4" = 1'-0"A420.1
SECURITY ROOM 10610
SCALE: 1/4" = 1'-0"A420.1
CONTROL ROOM 10208
SCALE: 1/4" = 1'-0"A420.1
CORRIDOR 10515
SCALE: 1/4" = 1'-0"A420.1
CORRIDOR 10514
SCALE: 1/4" = 1'-0"A420.1
CORRIDOR 10512
SCALE: 1/4" = 1'-0"A420.1
CONTROL ROOM 10207
SCALE: 1 1/2" = 1'-0"A420.1
DOOR HEAD DETAIL17
SCALE: 1 1/2" = 1'-0"A420.1
DOOR JAMB DETAIL18
SCALE: 1/4" = 1'-0"A420.1
CORRIDOR 10513
SCALE: 1 1/2" = 1'-0"A420.1
TRENCH DETAIL16
SCALE: 1/4" = 1'-0"A420.1
SECURITY ROOM 10609
SCALE: 1/4" = 1'-0"A420.1
SECURITY ROOM 10611
04/22/25 ADDENDUM 01
A1
A1
SCALE: 1 1/2" = 1'-0"A420.1
SLAB INFILL DETAIL20
AREA NOT IN SCOPE
USS1
CONTROL RM
1556A
FAMILY CONSULT
SECURITY
CORRIDOR
CORRIDOR
Space
TREADMILL
2" FD-3
2" SS
1 1/2" V
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
P101.1
FIRST FLOOR SANITARY
AND VENT PLAN
BID DOCUMENTS02/16/24
PROJECT A
KEYNOTES
1 CONNECT NEW SANITARY PIPING TO EXISTING SANITARY
PIPING. 2" MINIMUM. FIELD VERIFY CONDITIONS.
2 CONNECT NEW VENT PIPING TO EXISTING VENT PIPING. 2"
MINIMUM. FIELD VERIFY CONDITIONS.
3 ROUTE DRAIN LINE FROM HUMIDIFER IN WALL DOWN TO
FLOOR DRAIN. TERMINATE DRAIN PIPE AT FLOOR DRAIN WITH
AIR GAP. ROUTE 2" DRAIN LINE FROM FLOOR DRAIN TO
NEAREST EXISTING SANITARY LINE.
PLUMBING NOTES
1 THIS PLAN IS DIAGRAMMATIC IN NATURE AND SHALL NOT BE
SCALED TO DETERMINE THE LOCATION OR DIMENSION OF THE
WORK. CONTRACTOR SHALL VERIFY EXACT LOCATION OF
PIPING AND PENETRATIONS.
2 CUTTING, BORING, SAWCUTTING OR DRILLING THROUGH THE
STRUCTURAL MEMBERS IS PROHIBITED.
3 WALL-MOUNTED ACCESS PANELS SHALL BE LOCKABLE TYPE.
4 CONTRACTOR SHALL BE RESPONSIBLE TO VERIFY THE LOCATION
OF EXISTING BELOW GRADE WASTE PIPING AND REFLECT ANY
DEVIATION GREATER THAN 1'-0" FROM THIS PLAN ON THE
AS-BUILT DRAWINGS.
SANITARY & VENT NOTES
1 ALL SANITARY DRAINAGE (SS) PIPING IS LOCATED BELOW SLAB,
UNLESS OTHERWISE NOTED.
2 ROUTE ALL PIPING INTENDED FOR ABOVE CEILING LEVEL
INSTALLATION AS HIGH AS POSSIBLE. COORDINATE ROUTING
WITH OTHER TRADES. SLOPED PIPING ELEVATIONS TAKE PRIORITY
OVER ALL OTHER SYSTEMS AND TRADES.
3 PROVIDE SLEEVED AND WATERTIGHT PENETRATIONS AT ALL PIPE
PENETRATIONS THROUGH: LOWER LEVEL FLOOR SLAB, UPPER
LEVEL FLOOR SLAB, MECHANICAL PENTHOUSE FLOOR SLAB,
ROOF STRUCTURES AND, PERIMETER WALLS BELOW GRADE.
4 PROVIDE WATERTIGHT INSTALLATIONS OF ALL FLOOR EMBEDDED
DEVICES INCLUDING BUT NOT LIMITED TO: FLOOR DRAINS,
FLOOR SINKS, SHOWER DRAINS, HUB DRAINS, FLOOR RECESSED
CLEANOUTS AND, FLUSH TO FLOOR ACCESS HOUSINGS.
5 COORDINATE WITH ARCHITECT TO MAINTAIN THE WATERTIGHT
INTEGRITY OF THE BUILDING ENVELOPE AND FLOOR SLABS.
NORTH
SCALE: 1/4" = 1'-0"
BP1 - FIRST FLOOR WASTE & VENT PLAN
USS11
CONTROL RM
1556A
FAMILY CONSULT
SECURITY
CORRIDOR
CORRIDOR
Space
TREADMILL
EWH-01 1/2" CW
KE-31
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
P201.1
FIRST FLOOR DOMESTIC
WATER AND GAS PLAN
BID DOCUMENTS02/16/24
PROJECT A
KEYNOTES
1 CONNECT COLD WATER TO EXISTING COLD WATER PIPING.
1/2" MINIMUM. FIELD VERIFY CONDITIONS.
2 CONNECT EXISTING HOT WATER TO UNDER COUNTER WATER
HEATER INPUT. ROUTE WATER HEATER OUTPUT TO EXTERNAL
ASSE 1070 COMPLIANT THERMOSTATIC MIXING VALVE.
3 CONNECT OXYGEN, MEDICAL AIR, AND MEDICAL VACUUM
TO NEW HEADWALL.
4 CONNECT UNDERFLOOR 1/2" COLD WATER LINE UP TO HM-01
HUMIDIFIER. PROVIDE WALL MOUNTED VALVE BOX.
PLUMBING NOTES
1 THIS PLAN IS DIAGRAMMATIC IN NATURE AND SHALL NOT BE
SCALED TO DETERMINE THE LOCATION OR DIMENSION OF THE
WORK. CONTRACTOR SHALL VERIFY EXACT LOCATION OF
PIPING AND PENETRATIONS.
2 CUTTING, BORING, SAWCUTTING OR DRILLING THROUGH THE
STRUCTURAL MEMBERS IS PROHIBITED.
3 WALL-MOUNTED ACCESS PANELS SHALL BE LOCKABLE TYPE.
4 CONTRACTOR SHALL BE RESPONSIBLE TO VERIFY THE LOCATION
OF EXISTING BELOW GRADE WASTE PIPING AND REFLECT ANY
DEVIATION GREATER THAN 1'-0" FROM THIS PLAN ON THE
AS-BUILT DRAWINGS.
DOMESTIC WATER NOTES
1 PROVIDE SLEEVED AND WATERTIGHT PENETRATIONS AT ALL PIPE
PENETRATIONS THROUGH: LOWER LEVEL FLOOR SLAB AND,
PERIMETER WALLS BELOW GRADE.
2 PROVIDE WATERTIGHT INSTALLATIONS OF ALL FLOOR EMBEDDED
DEVICES INCLUDING BUT NOT LIMITED TO: FLOOR DRAINS,
FLOOR RECESSED CLEANOUTS.
3 COORDINATE WITH ARCHITECT TO MAINTAIN THE WATERTIGHT
INTEGRITY OF THE BUILDING ENVELOPE AND FLOOR SLABS.
4 PROVIDE BALL VALVE FOR ISOLATION OF ALL DOMESTIC WATER
BRANCH LINES. TYPICAL UNLESS OTHERWISE NOTED.
NORTH
SCALE: 1/4" = 1'-0"
BP1 - FIRST FLOOR DOMESTIC WATER PLAN
04/22/25 ADDENDUM 01
BradenCox Callout at what elevation?
PIPE
HIGH DENSITY
INSULATION
INSERT, REFER TO
THE
SPECIFICATIONS
CLEVIS HANGER
INSULATION
PIPE RING HANGER
GALVANIZED
STEEL SHIELD
HOT WATER
TO FAUCET
1/2" HOT WATER
SHUT-OFF VALVE
COLD WATER
TO FAUCET
INSTANTANEOUS ELECTRIC WATER HEATER. MOUNT
ON BACK WALL UNDER SINK. LOCATE TO MAKE
ACCESSIBLE FOR REPLACEMENT AND NOT
OBSTRUCTING UNDER SINK PLUMBING OF STORAGE.
LAVATORY OR SINK
THERMOSTATIC
MIXING VALVE
1/2" COLD WATER
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
P500.1
PLUMBING DETAILS AND
SCHEDULES
BID DOCUMENTS02/16/24
PROJECT A
SCALE: NTSP500.1
INSULATED PIPE AT HANGER DETAIL1
PLUMBING FIXTURE SCHEDULE
MARK DESCRIPTION MANUFACTURER MODEL NO.
MATERIAL
DESCRIPTION FINISH
TRIM FLOW FIXTURE FLUSH FIXTURE
WASTE
ROUGH-IN
PIPE SIZE
INDIRECT
WASTE
PIPE SIZE
VENT
PIPE
SIZE
COLD WATER
ROUGH-IN
PIPE SIZE
HOT WATER
ROUGH-IN
PIPE SIZE DESCRIPTION NOTESMANUFACTURER MODEL NO. TYPE
MOTION
SENSOR
CONTROL WATER FLOW
MAX.
MWT
VOL. PER
FLUSH
MIN. VOL.
PER FLUSH
KE-31 OUTLET BOX SIOUX CHIEF 696-RG1010MF ABS PLASTIC WHITE No 0.5 GPM 105 °F 1/2" GALVANIZED METAL CONSTRUCTION-NO PLASTIC BOXES ALLOWED
USS1 UNDERMOUNT SOLID
SURFACE SINGLE BOWL SINK
WILSONART AK2015 SOLID SURFACE
SS-3
SOLID
SURFACE
ELKAY LK406GN04T4 MANUAL No 1.5 GPM 105 °F 2" 1 1/2" 1/2" 1/2" SINGLE BOWL, UNDER COUNTER MOUNTED, SOLID SURFACE. SINGLE LEVER SWIVEL FAUCET WITH AN EXTERNAL ASSE 1070 COMPLIANT THERMOSTATIC MIXING VALVE. ONE ELKAY MODEL NO. LK35 BASKET STRAINER, P-TRAP, TAILPIECES, SUPPLIES AND STOPS. REFER TO FINISH PLANS FOR FURTHER
SINK AND COUNTERTOP INFORMATION. REFER TO FINISH KEY IN ARCHITECT PLANS FOR
ADDITIONAL MATERIAL INFORMATION. INSTALL WITH INCLUDED 1.5 GPM AERATOR.
ELECTRIC WATER HEATER SCHEDULE
MARK MANUFACTURER MODEL NO. TYPE
ELECTRIC HEAT EXCHANGER
EF
UNIT
WEIGHT FLA MOCP VOLT PH NOTES
HEATING
CAPACITY
FLOW ENTERING
WATER
TEMPERATURE
LEAVING
WATER
TEMPERATURE
MAXIMUM
TEMPERATURE
RISEDESIGN MIN MAX
EWH-01 BOSCH TR4000C-6 TANKLESS 6.5 kW 1.3 GPM 0.5 GPM 1.5 GPM 70 °F 105 °F 35 °F 0.96 4 lb 30 A 40 A 208 V 1
SCALE: NTSP500.1
INSTANTANEOUS ELECTRIC BOOSTER WATER HEATER2
SCHEDULE NOTES:
1. INSULATE THE SUPPLY AND DRAIN PIPING WITH A WHITE INSULATION DEVICE EQUAL TO McGUIRE MODEL NO. PW-2150-WC AND PROVIDE A THERMOSTATIC MIXING VALVE ON THE SUPPLIES EQUAL TO WATTS MODEL NO. USG-B.
FLOOR DRAIN SCHEDULE
MARK DESCRIPTION MANUFACTURER MODEL NO.
MATERIAL
DESCRIPTION WASTE VENT
DESCRIPTION NOTESDRAIN BODY PIPE SIZE PIPE SIZE
FD-3 FLOOR DRAIN ZURN Z415S EPOXY COATED
CAST IRON
2" 1 1/2" DURA COATED CAST IRON FLOOR DRAIN WITH ANCHOR FLANGE, INVERTIBLE MEMBRANE CLAMP, TRAP PROTECTION INSERT, AND ADJUSTABLE COLLAR WITH SEEPAGE SLOTS, ADJUSTABLE HEAVY
DUTY SQUARE HEEL PROOF NICKEL BRONZE STRAINER, AND NO HUB OUTLET.
FPVAV-10.4
CC-1
VAV-XX
S6L1
S10L1
"x
S /A 8"Ø S/A8"Ø S/A
S16L1
S16L1
S12L1
"x
R /A
10"Ø S-O/A
20"x20" S/A16"Ø S/A
16"Ø S/A
S8L1
8"x8" S/A
R14L1
R14L1
(E) O/A
8"Ø S/A (E)
R8L1
R10L1
E R/A
14"Ø R/A
14"Ø R/A
8"Ø R/A
B
9 9
CORRIDOR
SECURITY
CORRIDOR
FAMILY CONSULT
CHANGING
1556B
CT SCAN
CT SCAN
CONTROL RM
1556A 18"x14" R/A10"Ø S-O/A
12"Ø S/A
20"x20" S/A
10"Ø R/A
12"Ø R/A
AC-1
CU-1
HM-01
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
M101.1
FIRST FLOOR
MECHANICAL PLAN
BID DOCUMENTS02/16/24
PROJECT A
KEYNOTES
1 REBALANCE EXISTING RETURN GRILLE TO 190 CFM.
2 RELOCATE EXISTING FAN COIL UNIT (FPVAV-10.4) AND
COOLING COIL (CC-1) AS SHOWN.
3 NEW LOCATION FOR THE RELOCATED SECONDARY DRAIN
PAN UNDER COOLING COIL. INSTALL WITH MOISTURE SENSOR
INTERLOCKED WITH FPVAV-10.4 TO SHUT-DOWN OPERATION
UPON DETECTION OF MOISTURE.
4 DUCT CONTINUES AND CONNECTS TO BRANCH DUCTWORK
FROM AHU-6.
5 CONNECT NEW DIFFUSERS AND DUCTWORK TO EXISTING
RETURN BRANCH DUCTWORK APPROXIMATELY 15 FEET.
6 PLACE FABRIC BLADE BACKDRAFT DAMPER IN VERTICAL
EXHAUST DUCTWORK AND TRANSITION DUCTWORK FROM
42x42 TO 18x18.
7 CONNECT NEW TO EXISTING AND REBALANCE TO 500 CFM.
BALANCING DAMPER LOCATED ON TAKE-OFF FROM MAIN
BRANCH.
8 REBALANCE EXISTING VAV-XX TO PRE-BALANCING REPORT
AIRFLOW.
9 INSTALL FILTERS AT RETURN GRILLES FOR MAINTAINABILITY.
10 REFRIGERANT LINE SET ROUTED DOWN TO CONDENSING UNIT
IN CRAWLSPACE. FIELD VERIFY ROUTE AND CONNECTIONS.
11 CONDENSING UNIT LOCATED IN CRAWLSPACE BENEATH THE
CT SCAN ROOM AND STRUCTURALLY MOUNTED ABOVE
GRADE. REFRIGERANT LINE SET ROUTED TO THE INDOOR COIL,
FIELD VERIFY THE ROUTE.
MECHANICAL SHEET METAL NOTES
1 THE INSTALLATION OF ALL DUCTWORK SHALL BE CLOSELY
COORDINATED WITH NEW PLUMBING, ELECTRICAL, AND
STRUCTURAL CONDITIONS. NOT ALL REQUIRED OFFSETS AND
FITTINGS ARE INDICATED ON DRAWINGS, BUT SHALL BE
PROVIDED. REFER TO ARCHITECTURAL AND STRUCTURAL PLANS
FOR CLEARANCES. ALTERNATE DUCT ROUTING SHALL BE
APPROVED BY ARCHITECT/ENGINEER BEFORE PROCEEDING IN
ORDER TO ENSURE THAT THE AVAILABLE STATIC PRESSURE
REMAINS ADEQUATE. DUCTWORK LOCATION SHALL TAKE
PRECEDENCE OVER HVAC AND FIRE PROTECTION PIPING AND
ELECTRICAL CONDUIT AND CABLE TRAY.
2 LOCATE MECHANICAL EQUIPMENT SUCH THAT THERE IS
UNOBSTRUCTED ACCESS TO UNIT ACCESS PANELS, CONTROLS
AND VALVING.
3 INSTALL, SUPPORT AND BRACE NEW DUCTWORK AND
ACCESSORIES PER SMACNA GUIDELINES.
4 UNLESS OTHERWISE NOTED, ALL DUCTWORK ABOVE CEILING OR
EXPOSED IS OVERHEAD AND AS HIGH AS POSSIBLE TO THE
UNDERSIDE OF THE STRUCTURE, WITH SPACE FOR INSULATION
WHERE REQUIRED.
5 DUCTS CONNECTED TO EQUIPMENT SHALL EQUAL EQUIPMENT
CONNECTION SIZE UNLESS NOTED OTHERWISE.
6 PROVIDE FLEXIBLE CONNECTIONS IN ALL DUCTWORK SYSTEMS
CONNECTED TO MECHANICAL EQUIPMENT THAT REQUIRE
VIBRATION ISOLATION. FLEXIBLE CONNECTIONS SHALL BE
PROVIDED AT THE POINT OF CONNECTION TO THE EQUIPMENT
UNLESS OTHERWISE NOTED.
7 REFER TO DUCT TAKEOFF DETAILS. SPIN-IN TYPE WITH SCOOPS
SHALL NOT BE ACCEPTED. A MINIMUM OF 2 FEET SHALL BE
PROVIDED BETWEEN RUNOUT TAKEOFFS FROM TRUNK DUCTS.
8 MAXIMUM LENGTH OF FLEXIBLE DUCT SHALL BE 5'-0",
9 RUNOUT BALANCING DAMPERS SHALL BE MOUNTED AS CLOSE
TO MAIN DUCT AS POSSIBLE.
10 ROUND DUCT ELBOWS SHALL BE LONG RADIUS ELBOWS, UNLESS
SHOWN OTHERWISE.
11 PROVIDE FLAT BLADE MANUAL VOLUME DAMPERS AT ALL
TERMINAL DUCT BRANCHES AND AS INDICATED.
12 TERMINAL UNITS SHALL BE MOUNTED TO NOT IMPAIR ACCESS TO
FILTERS, COILS AND CONTROLS.
14 TRANSFER DUCTWORK SHALL BE INTERNALLY LINED WITH A
MINIMUM 1/2" ACOUSTIC LINING.
MECHANICAL NOTES
1 CONTRACTOR SHALL PROVIDE A COMPLETE AND WORKING
SYSTEM.
2 INSTALL EQUIPMENT PER MANUFACTURER'S
RECOMMENDATIONS. ROOF MOUNTED EQUIPMENT SHALL BE
LOCATED A MINIMUM OF 10'-0" FROM THE ROOF EDGE.
3 COORDINATE ALL EXTERIOR PENETRATIONS INCLUDING ROOF
PENETRATIONS WITH OTHER TRADES TO PROVIDE A COMPLETE
AND FULLY WEATHER-PROOF INSTALLATION.
4 PRIMARY CONDENSATE DRAIN PIPING SHALL BE INSULATED AND
SHALL INCLUDE A VAPOR RETARDANT INSULATION JACKET.
SEAL ALL JOINTS AND PENETRATIONS.
5 CONTRACTOR SHALL ENGAGE A TESTING AND BALANCING
FIRM CERTIFIED BY AABC OR NEBB TO PERFORM TESTING AND
BALANCING PROCEDURES ON EACH SYSTEM ACCORDING TO
THE PROCEDURES CONTAINED IN AABC'S OR NEBB'S
STANDARDS FOR TESTING AND BALANCING HEATING,
VENTILATION, AND AIR CONDITIONING SYSTEMS.
6 THIS DRAWING IS DIAGRAMMATIC IN NATURE AND SHALL NOT
BE SCALED.
A CONTRACTOR SHALL BE RESPONSIBLE TO PROVIDE A
COMPLETE AND WORKING SYSTEM.
7 THIS DRAWING IS BASED ON VISUALLY OBSERVABLE EXISTING
CONDITIONS AS OF THE TIME OF DESIGN. CONTRACTOR SHALL
BE RESPONSIBLE TO FULLY VERIFY ALL EXISTING CONDITIONS,
COMPONENTS, ETC. PRIOR TO THE START OF THE WORK. ANY
DEVIATION FROM THIS DRAWING IN KIND, OR IN LOCATION
EXCEEDING 1'-0", SHALL BE BROUGHT TO THE ATTENTION OF THE
ENGINEER.
8 HVAC SYSTEMS IN PENTHOUSE #4 CONDITION THIS SPACE.
SYSTEM AC4B(1) SUPPLIES VENTILATION THROUGH THE
CRAWLSPACE DUCTWORK, CV BOXES WITH REHEAT, THEN UP
INTO THE ROOMS THROUGH THE ATU'S SELECTED CV'S ALSO
HAVE HUMIDIFIERS. SYSTEM AC4B(3) PROVIDES TEMPERATURE
CONTROL OVERHEAD THROUGH THE CEILING BY VAV BOXES
WITHOUT REHEAT. THE TWO SYSTEMS WORKED TOGETHER FOR
SPACE CONDITIONING. ONE THERMOSTAT CONTROLS BOTH THE
CRAWL SPACE VENTILATION SUPPLY AND ABOVE CEILING 4B(3)
VAV BOXES. BOTH AC4B(1) AND AC4B(3) HAVE A COMMON
PLENUM FOR RETURN.
NORTH
SCALE: 1/4" = 1'-0"
BP1 - FIRST FLOOR HVAC PLAN
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
M600.1
MECHANICAL SCHEDULES
BID DOCUMENTS02/16/24
PROJECT A
GRILLES, REGISTERS AND DIFFUSERS SCHEDULE
MARK DESCRIPTION MANUFACTURER MODEL
FACE
SIZE
MAX
AIRFLOW
CFM
TOTAL
PRESSURE
NECK THROW
DISTANCE
(A) NC (B) FINISH NOTESSIZE
R8L1 LAY-IN EGG-CRATE, ROUND-NECK
RETURN GRILLE
TITUS 50F 24x24 200 CFM 0.08 in-wg 8" 0" 0 WHITE ENAMEL 1
R10L1 LAY-IN EGG-CRATE, ROUND-NECK
RETURN GRILLE
TITUS 50F 24x24 200 CFM 0.06 in-wg 10" 0" 0 WHITE ENAMEL 1
R14L1 LAY-IN EGG-CRATE, ROUND-NECK
RETURN GRILLE
TITUS 50F 24x24 815 CFM 0.06 in-wg 14" 0" 0 WHITE ENAMEL 1
S6L1 LAY-IN LOUVERED FACE,
ROUND-NECK SUPPLY DIFFUSER
TITUS TDC-AA 24x24 100 CFM 0.08 in-wg 6" 4' - 0" 18 WHITE ENAMEL
S8L1 LAY-IN LOUVERED FACE,
ROUND-NECK SUPPLY DIFFUSER
TITUS TDC-AA 24x24 200 CFM 0.08 in-wg 8" 6' - 0" 21 WHITE ENAMEL
S10L1 LAY-IN LOUVERED FACE,
ROUND-NECK SUPPLY DIFFUSER
TITUS TDC-AA 24x24 200 CFM 0.09 in-wg 10" 7' - 0" 21 WHITE ENAMEL
S12L1 LAY-IN LOUVERED FACE,
ROUND-NECK SUPPLY DIFFUSER
TITUS TDC-AA 24x24 430 CFM 0.09 in-wg 12" 8' - 0" 22 WHITE ENAMEL
S16L1 LAY-IN LOUVERED FACE,
ROUND-NECK SUPPLY DIFFUSER
TITUS TDC-AA 24x24 850 CFM 0.09 in-wg 16" 8' - 0" 20 WHITE ENAMEL
GENERAL NOTES:
A. THROW AT 50 FEET PER MINUTE VELOCITY.
B. "-" NC RATING IS LOW ENOUGH TO NOT REGISTER IN THE MANUFACTURER'S DATA.
SCHEDULE NOTES:
1. FULL 24"x24" EGG CRATE FACE.
WALL MOUNTED AIR CONDITIONER SCHEDULE
MARK
BASIS OF DESIGN
MANUFACTURER
BASIS OF
DESIGN MODEL
NO. TYPE
FAN COOLING COIL
UNIT
WEIGHT MCA MOCP VOLT PH
INTERLOCK
ACCESSORIES NOTES
DESIGN
AIRFLOW
MOTOR NOMINAL
CAP
CAP
MARKQTY POWER ECM TOTAL
AC-1 BLUERIDGE ZKFN-90-8-1 WALL AC UNIT 500 CFM 1 0.13 hp No 4 ton 48000 Btu/h 88 lb 2 A 15 A 208 V 1 CU-1 CP 1
SPLIT SYSTEM OUTDOOR UNIT SCHEDULE
MARK DESCRIPTION MANUFACTURER MODEL NO. TYPE
COOLING
CAPACITY
COMPRESSOR
TYPE
REFRIGERANT
TYPE
LOW
AMBIENT
KIT
SUMMER
AMBIENT
DBT
WINTER
AMBIENT
DBT
EFFICIENCY
UNIT
WEIGHT MCA MOCP VOLT PH ACCESSORIES NOTESSEER2 EER2
CU-1 CONDENSING UNIT BLUERIDGE KTQ420D1UMU OUTDOOR 4.00 ton ROTARY R-410A Yes 75.0 °F 70.0 °F 18.9 9.3 219 lb 37 A 50 A 208 V 1 LA, T 1-3
ACCESSORIES:
CP - ASPEN CONDENSATE PUMP
SCHEDULE NOTES:
1. UNIT TO BE MOUNTED 7' ABOVE FINISHED FLOOR.
ACCESSORIES:
LA - LOW AMBIENT KIT
T - THERMOSTATIC INTERFACE CONNECTED TO EXISTING JOHNSON CONTROLS
SCHEDULE NOTES:
1. MANUFACTURER SUBMITTALS MUST INCLUDE REFRIGERANT PIPING DIAGRAM WITH PIPE DIAMETERS, LENGTHS, AND REFRIGERANT VOLUME.
2. SUBSTITUE MANUFACTER SHALL BE RESPONSIBLE FOR ADDITIONAL PIPING AND REFRIGERANT.
3. PROVIDE INSULATION ON REFRIGERANT PIPING.
ELECTRIC HUMIDIFIER SCHEDULE
MARK
LOCATION
BASIS OF DESIGN
MANUFACTURER
BASIS OF
DESIGN MODEL
NO. TYPE ARRANGEMENT
HEATING ELEMENT HUMIDIFIER
OUTPUT
(LBS/HR)
UNIT
WEIGHT FLA MCA MOCP VOLT PH NOTESNAME NO. QTY TYPE POWER SCR
HM-01 CT SCAN 1556 CONDAIR EL-010 ELECTRODE WALL-MOUNT 1 ELEC 4 kW Yes 55 lb 18 A 18 A 25 A 208 V 1 N/A
GENERAL NOTES:
1. PROVIDE WITH DUAL CHANNEL FACTORY CONTROL.
2. PROVIDE WITH DUCT MOUNTED STEAM DISTRIBUTOR ASD 18.
10.0
J
48"
J
48"
J 4
J
J4
J4 "J 9
J9
J
96"
CLNG
J
J
48"
4D A
C A
C 2D
A C
A C2D
84"
TP
TP
TPTP
TP
TP
TP
TP
A
C TP
TP
TP
TP
J
J
96"
TP
84" 1D
2D
4D
2D
2D
S
2D
J
48"
CLNG
J
M
CR
R
M L
ML
60" 60"
S
CLNG
J
J J
AREA NOT IN SCOPE
2 1
CHANGING
1556B
CT SCAN
CT SCAN
CONTROL RM
1556A
FAMILY CONSULT
SECURITY
Space
CORRIDOR
CORRIDOR
CT MDP
CT UPS
PDU
AC-1
E500.1
HM-01
GENERAL POWER NOTES
1 WHERE CONNECTED TO A 20 AMP BRANCH CIRCUIT
SUPPLYING AN INDIVIDUAL (SIMPLEX OR DUPLEX), THE
RECEPTACLE SHALL BE RATED AT 20 AMPS.
2 REFER TO SECTION 26 0519 FOR MINIMUM CONDUCTOR SIZE
ADJUSTMENTS FOR VOLTAGE DROP.
3 WIRE COUNTS FOR CIRCUIT CONDUCTORS ARE NOT SHOWN.
PROVIDE PROPER NUMBER OF CONDUCTORS TO ACHIEVE
CIRCUIT AND SWITCHING CONNECTIONS SHOWN.
4 CIRCUIT WIRING IS NOT SHOWN EXCEPT FOR SWITCHING
INTENT OF FIXTURES AND CONTROL OF DEVICES.
5 PROVIDE PROPER NUMBER OF CONDUCTORS TO ACHIEVE
CIRCUITING AND SWITCHING SHOWN.
6 SINGLE PHASE BRANCH CIRCUITS SHALL NOT SHARE
NEUTRALS.
7 THE CONTRACTOR SHALL PROVIDE THE DATA CONDUITS,
BOXES, AND DATA CABLING AS SHOWN.
8 THE CONTRACTOR SHALL FIELD VERIFY THE ROUGH-IN
DIMENSIONS OF ALL DATA OUTLETS, TV OUTLETS AND PRIOR
TO ANY INSTALLATION. ANY CHANGE OF LOCATION OF A
BOX THAT WAS NOT VERIFIED WITH THE OWNER PRIOR TO
INSTALLATION SHALL BE RELOCATED, TO THE OWNER’S
SATISFACTION. AT THE CONTRACTOR’S EXPENSE.
9 PROVIDE THE REQUIRED POWER FOR ALL MECHANICAL OR
PLUMBING EQUIPMENT SPECIFIED IN THE CONTRACT
DOCUMENTS, EVEN IF NOT SHOWN ON THE POWER PLANS.
10 THE CONTRACTOR SHALL FIELD VERIFY THE POWER
REQUIREMENTS OF ALL MECHANICAL AND PLUMBING
EQUIPMENT, WITH THE EQUIPMENT SUPPLIERS, PRIOR TO
PREPARING SUBMITTALS OR ORDERING ANY SWITCHGEAR
OR EQUIPMENT. SUPPLIED PLUMBING AND/OR MECHANICAL
EQUIPMENT MAY VARY FROM EQUIPMENT SPECIFIED IN THE
PLANS.
11 FOR EACH THERMOSTAT, TEMPERATURE SENSOR OR WALL
MOUNTED HVAC CONTROL DEVICE PROVIDE A JUNCTION
BOX, WITH A SINGLE GANG MUD RING, AT 48” ABOVE THE
FINISH FLOOR WITH A ¾” CONDUIT (WITH A PULL STRING) TO
ABOVE THE CEILING FOR THE HVAC CONTROLS. HVAC
CONTROLS WIRING AND DEVICES INSTALLED BY
MECHANICAL CONTROLS CONTRACTOR.
12 ALL PENETRATIONS MUST BE BACKED WITH SHEET LEAD OF THE
SAME THICKNESS (OR GREATER) AS REQUIRED FOR THE WALL
OR PARTITION THAT PENETRATION OCCURS IN.
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
259 E. Michigan Ave., Suite 407
Kalamazoo, MI 49007
P: 269.585.2255 steelheadengineeringco.com
THE REPRODUCTION, COPYING OR OTHER USE OF THIS
DRAWING WITHOUT WRITTEN AUTHORIZATION IS PROHIBITED.
c SEVEN GENERATIONS ARCHITECTURE & ENGINEERING, LLC
SHEET NO.
SHEET TITLE
JOB NUMBER:
ISSUE DATES:
BID DOCUMENTS
222112
IH
S B
F S U
M U
LT
IP
LE
P
R O
J E C
T D
E S
IG
N
IH
S B
LA
C
K F E E T
S E R
V
IC
E U
N
IT
B
LA
C K
W E A
S E L
R O
A D
B R
O W
N
IN
G M
O N
TA
N
A
E201.1
FIRST FLOOR POWER PLAN
BID DOCUMENTS02/16/24
04/22/25 ADDENDUM 01
PROJECT A
KEYNOTES
1 JUNCTION BOX AT 48" ABOVE THE FINISH FLOOR FOR THE
POWERED DOOR OPERATOR PUSH BUTTON. PROVIDE A 3/4"
CONDUIT TO THE JUNCTION BOX OVER THE DOOR.
2 JUNCTION BOX ABOVE THE CEILING OVER THE DOOR FOR THE
POWERED DOOR POWER CONNECTION, FIELD VERIFY THE
CONNECTION REQUIREMENTS WITH THE EQUIPMENT SUPPLIER.
3 JUNCTION BOX AT 48" ABOVE THE FINISH FLOOR, FOR THE
CODE BLUE BUTTON, WITH A 3/4" CONDUIT TO ABOVE THE
CEILING.
4 JUNCTION BOX IN THE CEILING FOR THE POWER CONNECTION
TO THE PATIENT LIFT, CONNECT NEW DEVICES TO EXISTING
HOMERUN CIRCUITRY FROM DEMOLISHED FIXTURES. WHERE
NEEDED ROUTE AND PROVIDE NEW CIRCUITRY FROM PANELS
NB OR NBB IN ROOM 1057 TO ACHIEVE APPROXIMATE
LAYOUT SHOWN. FIELD VERIFY THE REQUIREMENTS WITH THE
EQUIPMENT SUPPLIER.
5 JUNCTION BOX ABOVE DOOR TO PROVIDE POWER FOR
X-RAY. 'IN-USE' SIGNAGE PROVIDED BY OWNER, FIELD VERIFY
THE CONNECTION REQUIREMENTS.
6 SWITCH FOR CONTROL OF XRAY IN-USE LIGHTS.
7 TWIST LOCK EMERGENCY POWER OFF BUTTON FOR CT
SCANNER. CONNECT TO POWER DISTRIBUTION UNIT. FIELD
VERIFY THE CONNECTION REQUIREMENTS WITH THE
EQUIPMENT SUPPLIER.
8 SEE CT RISER DIAGRAM FOR ADDITIONAL CONNECTION
DETAIL ON SHEET E500.1.
9 EXISTING RECEPTACLE AND ASSOCIATED BRANCH CIRCUITRY
TO REMAIN.
10 CONNECT NEW DEVICES TO EXISTING HOMERUN CIRCUITRY
FROM DEMOLISHED FIXTURES. WHERE NEEDED ROUTE AND
PROVIDE NEW CIRCUITRY TO ACHIEVE APPROXIMATE LAYOUT
SHOWN. ROUTE ANY NEW HOMERUN CIRCUITRY FROM SPARE
BREAKERS IN PANEL NBB OR PANEL NB IN ROOM 1057. FIELD
VERIFY EXISTING CONDITIONS AND CONNECTION
REQUIREMENTS.
11 PROVIDE DEDICATED 20-AMP 1-POLE BREAKER IN PANEL NB
OR PANEL NBB IN ROOM 1057. ROUTE #12 COPPER
CONDUCTORS TO THE NEW BLANKET WARMER RECEPTACLE.
FIELD VERIFY THE CIRCUIT LOCATION DUE TO PANEL SPACE
AND AVAILABLE AMPACITY CONSTRAINTS.
12 COORDINATE SPEAKER WITH CT MANUFACTURER, PROVIDE
SPEAKER WIRING FROM CT TO SPEAKER PER MANUFACTURER'S
REQUIREMENTS.
13 NEW LOCATION FOR STAFF NURSE CALL IN CONTROL ROOM,
CONNECT TO EXISTING NURSE CALL SYSTEM, EXTEND AND
MODIFY CIRCUITRY AS REQUIRED.
14 APPROXIMATE LOCATION OF NEW UNDERFLOOR RACEWAY,
REFER TO ARCHITECTURAL FLOOR PLAN FOR MORE
INFORMATION.
15 ROUTE NEW FEEDER FOR GANTRY POWER FROM CT MDP TO
CT GANTRY VIA FLOOR TRENCH DUCT AND CONNECT PER
MANUFACTURER SPECIFICATIONS.
16 APPROXIMATE FLOOR TO CEILING RACEWAY LOCATION
REFER TO ARCHITECTURAL FLOOR PLAN FOR MORE
INFORMATION.
17 CONNECT NEW CT MAIN DISTRIBUTION PANEL TO NEW
FEEDERS FROM EXISTING CT DISCONNECT LOCATED IN THE
CRAWL SPACE. PROVIDE COMPATIBLE VERTICAL TRENCH
RACEWAY SECTION TO ROUTE THE CONDUCTORS FED
THROUGH THE FLOOR TRENCH UP TO THE CT MDP FOR
TERMINATION, PROVIDE COVERS AND OPENINGS FOR
TRENCH AS REQUIRED. SEE CT RISER DIAGRAM ON FOR
ADDITONAL DETAILS.
18 NEW 40-AMP 208V 1-PHASE BREAKER WITH #8 COPPER
CONDUCTORS FOR WATER HEATER FROM PANEL NB OR NBB,
FIELD VERIFY CONNECTION REQUIREMENTS.
19 DISCONNECT, EXTEND AND RECONNECT THE POWER
CIRCUITRY FOR RELOCATED VAV BOX.
20 JUNCTION BOX FOR CT INJECTOR CONTROL CONSOLE,
COORDINATE AND FIELD VERIFY MANUFACTURER
REQUIREMENTS PRIOR TO INSTALLATION.
21 CT MAIN DISTRIBUTION PANEL SHALL PROVIDE POWER FOR CT
INJECTOR. COORDINATE AND FIELD VERIFY MANUFACTURER
REQUIREMENTS PRIOR TO INSTALLATION.
22 AC UNIT POWERED BY CONDENSING UNIT PER
MANUFACTURER SPECIFICATIONS. PROVIDE 3-POLE MOTOR
SWITCH AS DISCONNECTING MEANS FOR INDOOR AC UNIT.
SEE MECHANICAL DRAWINGS.
23 PROVIDE 50A 2-POLE CIRCUIT BREAKER IN PANEL NBB OR NB
FOR CONNECTION TO NEW AC CONDENSING UNIT. PROVIDE
2 POLE, 50 AMP FUSED DISCONNECT AS DISCONNECTING
MEANS IN CRAWLSPACE. FIELD VERIFY ROUTE AND
CONNECTIONS. SEE MECHANICAL DRAWINGS.
24 PROVIDE POWER FOR DOOR ACCESS CARD READER AND
AUTOMATIC SWING OPERATOR, ROUTE NEW 20A SPARE
BREAKER AND ASSOCIATED BRANCH CIRCUITRY FROM PANEL
NBB OR NB IN ROOM 1057, FIELD VERIFY CONNECTION
REQUIREMENTS.
25 ROUTE NEW FEEDER FOR OPERATOR CONSOLE FROM CT MDP
THROUGH FLOOR TRENCH DUCT AND CONNECT PER
MANUFACTURER SPECIFICATIONS.
26 NEW 25-AMP 208V 1-PHASE BREAKER WITH #10 COPPER
CONDUCTORS FOR HUMIDIFER FROM PANEL NB OR NBB.
PROVIDE 2 POLE, 30-AMP DISCONNECT IN ACCESSIBLE
CEILING WITHIN 6 FT. OF EQUIPMENT. FIELD VERIFY
CONNECTION REQUIREMENTS AND COORDINATE WITH
MECHANICAL…