This solicitation is for a Building 5 Pharmacy Cleanroom Installation project located at the W. G. Hefner VAMC in Salisbury, North Carolina. The Department of Veterans Affairs requires a contractor to provide all labor, material, equipment and supervision to construct a pre-manufactured ISO class 7 cleanroom and separate compounding areas in accordance with USP chapters 797 and 800 for non-hazardous and hazardous sterile compounding. A site visit will take place on February 9th and questions are due by February 18th. Bids for this 100% SDVOSB set-aside are due on March 4th and will be publicly opened virtually on March 10th. The construction cost is estimated between $500,000 and $1 million. The NAICS code is 236220 for commercial and institutional building construction with a size standard of $39.5 million.

View the file

Other files

Show all 14

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SHEET NAMING CONVENTION

NOTE:

WIRE SIZE FOR ALL 277V, 20A CIRCUITS,

UIO:

FOR ALL ONE-WAY CIRCUITS OF LENGTH OF LESS THAN 170 FT,

PROVIDE 2#12 & 1#12G, 3/4"C.

FOR ALL ONE-WAY CIRCUITS OF LENGTH OF LESS THAN 280 FT

AND GREATER THAN OR EQUAL TO 170 FT, PROVIDE 2#10 & 1#10G,

3/4"C.

FOR ALL ONE-WAY CIRCUITS OF LENGTH OF LESS THAN 440 FT

AND GREATER THAN OR EQUAL TO 280 FT, PROVIDE 2#8 & 1#8G,

3/4"C.

FOR ALL ONE-WAY CIRCUITS OF LENGTH OF LESS THAN 700 FT

AND GREATER THAN OR EQUAL TO 440 FT, PROVIDE 2#6 & 1#6G,

3/4"C.

WIRE SIZE FOR ALL 120V, 20A CIRCUITS,

UIO:

FOR ALL ONE-WAY CIRCUITS OF LENGTH OF LESS THAN 75 FT,

PROVIDE 2#12 & 1#12G, 3/4"C.

FOR ALL ONE-WAY CIRCUITS OF LENGTH OF LESS THAN 125 FT

AND GREATER THAN OR EQUAL TO 75 FT, PROVIDE 2#10 & 1#10G,

3/4"C.

FOR ALL ONE-WAY CIRCUITS OF LENGTH OF LESS THAN 190 FT

AND GREATER THAN OR EQUAL TO 125 FT, PROVIDE 2#8 & 1#8G,

3/4"C.

FOR ALL ONE-WAY CIRCUITS OF LENGTH OF LESS THAN 300 FT

AND GREATER THAN OR EQUAL TO 190 FT, PROVIDE 2#6 & 1#6G,

3/4"C.

EXISTING EQUIPMENT

NEW OR RELOCATED EQUIPMENT

HATCH INDICATES EQUIPMENT TO BE DEMOLISHED

ELECTRICAL GENERAL NOTES:

1. THE WORK SHALL CONFORM WITH ALL REQUIREMENTS OF:

A. NFPA 70-2020 (NATIONAL ELECTRICAL CODE)

B. ANSI C2-2017 (NATIONAL ELECTRICAL SAFETY CODE)

C. APPLICABLE LOCAL CODES AND FEDERAL AND STATE LAWS.

D. PG-18-10 VA ELECTRICAL DESIGN MANUAL

2. MINIMUM RACEWAY SIZE SHALL BE 3/4". INCREASE RACEWAY SIZE AS

REQUIRED TO LIMIT RACEWAY FILL RATIO TO LESS THAN 40% FULL.

3. CONTRACTOR SHALL CAREFULLY COORDINATE WORK WITH OTHER

TRADES THROUGH THE GENERAL CONTRACTOR AND SHALL BE

RESPONSIBLE FOR SECURING SPACE REQUIREMENTS FOR ELECTRICAL

EQUIPMENT CLEARANCE FOR RECESSED LUMINAIRES AND CORRECT

ROUGH-IN LOCATIONS OF ELECTRICAL CONNECTIONS.

4. CONTRACTOR SHALL BE RESPONSIBLE FOR VERIFYING CATALOG

NUMBERS ON THESE DRAWINGS TO MATCH WITH MATERIAL DESCRIPTIONS

INDICATED.

5. VERIFY EXACT HEIGHT OF EACH COUNTERTOP AND BACKSPLASH ON

ARCHITECTURAL DETAILS AND/OR CASE WORK SHOP DRAWINGS AND

ADJUST SPECIFIED MOUNTING HEIGHT OF WALL OUTLETS TO LOCATE

BOTTOM OF OUTLET BOX 4" ABOVE TOP OF BACKSPLASH. IF NO

BACKSPLASH IS USED, LOCATE BOTTOM OF OUTLET BOX 6" ABOVE

COUNTERTOP.

6. VERIFY DOOR SWINGS WITH ARCHITECTURAL DRAWINGS BEFORE

ROUGHING IN WALL SWITCHES. SWITCHES IN THE SAME LOCATION SHALL

BE GANGED TOGETHER IN ONE COMMON BACKBOX AND SHALL HAVE ONE

COMMON FACE PLATE.

7. ALL FEEDERS AND BRANCH CIRCUITS SHALL INCLUDE A GREEN INSULATED

GROUND CONDUCTOR, SIZE PER NATIONAL ELECTRICAL CODE, OR AS

SHOWN, CONNECTED TO EACH DEVICE AND OUTLET BOX ON THE CIRCUIT

AND TO THE PANELBOARD GROUND BUS. PROVIDE A DEDICATED NEUTRAL

CONDUCTOR FOR EACH BRANCH CIRCUIT. MULTIPLE BRANCH CIRCUITS IN

ONE RACEWAY REQUIRE ONLY ONE GROUND CONDUCTOR.

8. VERIFY LUMINAIRE, CEILING MOUNTED OCCUPANCY SENSOR LOCATIONS

WITH ARCHITECTURAL REFLECTED CEILING PLANS FOR EXACT LOCATIONS

AND DIMENSIONS PRIOR TO INSTALLATION. VERIFY EXACT LOCATIONS OF

MOTORS AND EQUIPMENT BEFORE ROUGHING-IN.

9. EXISTING ELECTRICAL WORK (NOT SHOWN) SHALL REMAIN, UNLESS

INDICATED OTHERWISE. SHOULD ANY EXISTING ELECTRICAL POWER,

LIGHTING OR AUXILIARY CIRCUIT, FEEDER OR EQUIPMENT BE SEVERED,

DISCONNECTED OR DELETED IN THE PROCESS OF CONSTRUCTION OR

REMODELING WHICH IS DONE AS A RESULT OF CONTRACT PLANS AND

SPECIFICATIONS, AND UNLESS SPECIFICALLY DESIGNATED BY THE

DRAWINGS TO BE DELETED, THEN SAID CIRCUIT OR FEEDER SHALL BE

RESTORED TO WORKING CONDITION. THE RESTORATION SHALL INCLUDE

RE-ROUTING, RELOCATION, RECONNECTION OR REPLACEMENT AS MAY BE

REQUIRED BY THE NEW WORK. ANY SUCH WORK REQUIRED SHALL BE

INCLUDED IN THE CONTRACT AND NO EXTRA COMPENSATION WILL BE

GRANTED.

10. NEW WORK SHALL BE MADE TO TIE INTO THE EXISTING IN A UNIFORM

MANNER, SIMILAR ITEMS OF NEW WORK SHALL BE CHECKED AGAINST

EXISTING WORK FOR TYPE MOUNTING, MOUNTING HEIGHTS, ETC. ITEMS

SHOWN IN NEW WORK AT VARIANCE FROM THE EXISTING SHALL BE

REFERRED TO THE ARCHITECT FOR DECISION BEFORE ROUGH-IN.

11. REFER TO ONE-LINE DIAGRAMS, SCHEDULES AND RISER DIAGRAMS FOR

CONDUCTOR AND CONDUIT SIZES NOT SHOWN ON PLANS.

12. PROVIDE IS AN INCLUSIVE TERM USED TO DESCRIBE ASPECTS OF THE

WORK TO BE ACCOMPLISHED, AND IS HEREBY DEFINED TO REQUIRE TO

STORE, FURNISH, INSTALL, MOUNT, CONNECT, CONTROL AND POWER

EQUIPMENT INDICATED, AS WELL AS ALL APPURTENANCES REQUIRED TO

MAKE ELECTRICAL SYSTEMS OPERATE AS INDICATED WITHIN THESE

DRAWINGS AND SPECIFICATIONS AND TO FULFILL THE SCOPE OF WORK.

13. DEMOLISH IS AN INCLUSIVE TERM USED TO DESCRIBE ASPECTS OF THE

WORK TO BE ACCOMPLISHED, AND IS HEREBY DEFINED TO REQUIRE

CONTRACTOR TO DISCONNECT EQUIPMENT FROM ALL CONNECTIONS,

REMOVE FROM THE OWNER'S SITE, AND DISPOSE OF IN ACCORDANCE

WITH FEDERAL, STATE AND LOCAL LAWS AND ORDINANCES. COST OF

DISPOSAL IS ENTIRELY THE CONTRACTOR'S RESPONSIBILITY.

14. ALL CONDUCTORS ARE COPPER UNLESS SPECIFICALLY NOTED AS

ALUMINUM.

15. CONTRACTOR SHALL FIELD VERIFY ALL EXISTING CONDITIONS,

DIMENSIONS AND ELEVATIONS BEFORE PROCUREMENT OF ANY MATERIALS

AND DEVELOPMENT OF ANY SHOP DRAWINGS OR SUBMITTALS.

16. PROVIDE LABELS ON ALL RECEPTACLES, WALL MOUNTED LIGHT

SWITCHES/OCCUPANCY SENSORS AND JUNCTION BOXES INDICATING THE

SOURCE PANEL & CIRCUIT(S). TEXT SHALL BE BLACK, 1/8” IN HEIGHT

PRINTED ON A CLEAR LABEL. HANDWRITTEN LABELS ARE NOT PERMITTED

EXCEPT FOR JUNCTION BOXES LOCATED ABOVE FINISHED CEILING WHICH

CAN BE HANDWRITTEN WITH AN INDELIBLE MARKER.

17. ALL EQUIPMENT PADS FOR INTERIOR ELECTRICAL EQUIPMENT SHALL

EXTEND 2 ” BEYOND THE FOOTPRINT OF THE EQUIPMENT.

18. EMT SHALL COMPLY WITH ANSI C80.3 AND UL 797. ALL EMT SHALL BE

FACTORY PAINTED.

19. ALL JUNCTION BOXES, ELECTRICAL DEVICES, AND FACEPLATES SHALL BE

SEALED FOR AIR LEAKAGE USING APPROVED SEALANT.

ELECTRICAL ABBREVIATIONS

A OR AMP

ACC

AC

AC

AF

AFF

AFG

AHU

AIC

AL

AM

ASYM

AT

ATS

AUTO

AWG

B

BC

BCW

BFF

BFG

BLDG

C

CAT

CB

CCT

CCTV

CFCI

CH

CHWCP

CHWP

CHWSP

CKT

CMH

CP

CPT

CRAC

CRI

CT

CT

CU

CWP

D

DC

DF

DH or DUH

DHWCP

DISC

DOAS

DPST

DPDT

DSCU

DSHP

DSS

DWBP

DX

EC

EF

ELEC

EMH

EMT

E or EMER

EPA

EQUIP

ERV

ERU

EXIST or EX

EUH

EXP

EWC

EWH

F

FACP

FCU

FLEX

FWE

G OR GND

GFGI

GFI

GRS

GUH

GWH

H or HT

HAC

HOA

HP

HP

HT

HVU

HWCP

HWP

AMPERE(S)

AIR COOLED CHILLER

ALTERNATING CURRENT

AIR COMPRESSOR / AIR CURTAIN

AMP FRAME

ABOVE FINISHED FLOOR

ABOVE FINISHED GRADE

AIR HANDLING UNIT

AMPERE INTERRUPTING CAPACITY

ALUMINUM

AMMETER

ASYMMETRICAL

AMP TRIP

AUTOMATIC TRANSFER SWITCH

AUTOMATIC

AMERICAN WIRE GAUGE

BOILER

BRANCH CONTROLLER

BARE COPPER WIRE

BELOW FINISHED FLOOR

BELOW FINISHED GRADE

BUILDING

CONDUIT

CATEGORY

CIRCUIT BREAKER

CORRELATED COLOR TEMPERATURE

CLOSED CIRCUIT TELEVISION

CHILLER

CHILLED WATER PUMP

CIRCUIT

COMMUNICATIONS MANHOLE

CONDENSATE PUMP

CONTROL POWER TRANSFORMER

COMPUTER ROOM AC UNIT

COLOR RENDERING INDEX

CURRENT TRANSFORMER

COOLING TOWER

COPPER OR CONDENSING UNIT

CONDENSER WATER PUMP

DEPTH

DIRECT CURRENT

DESTRATIFICATION FAN

DUCT HEATERS

DOMESTIC HOT WATER CIRCULATING PUMP

DISCONNECT SWITCH

DEDICATED OUTSIDE AIR SYSTEM (UNIT)

DOUBLE POLE SINGLE THROW

DOUBLE POLE DOUBLE THROW

DUCTLESS SPLIT SYSTEM (INDOOR UNIT)

DOMESTIC WATER BOOSTER PUMP

DIRECT EXPANSION (DX) COOLING COIL

EMPTY CONDUIT

EXHAUST FAN

ELECTRICAL

ELECTRICAL MANHOLE

ELECTRICAL METALLIC TUBING

EMERGENCY

EFFECTIVE PROJECTED AREA

EQUIPMENT

ENERGY RECOVERY VENTILATOR

ENERGY RECOVERY UNIT

EXISTING

ELECTRIC UNIT HEATER

EXPLOSION PROOF

ELECTRIC WATER COOLER

ELECTRIC WATER HEATER

FUSE

FIRE ALARM CONTROL PANEL

FAN COIL UNIT

FLEXIBLE

FURNISHED WITH EQUIPMENT

GROUND

GROUND FAULT INTERRUPTER

GALVANIZED RIGID STEEL CONDUIT

GAS UNIT HEATER

GAS WATER HEATER

HEIGHT

HEATED AIR CURTAIN

HAND-OFF AUTOMATIC

HORSE POWER

HEAT TRACKING

HEATING/VENTILATING UNIT

HWRP

HWSP

HWUH

HZ

IDS

IMC

IRH

J OR JB

K

KAIC

KCM OR KCMIL

KEF

KH

KV

KVA

KW

KWHR

L

LA

LAN

LAV

LTG

MAU

MAX

MCA

MCB or MB

MCC

MGB

MH

MIN

MLO

MOD

MT or MTD

MRS

N

NEC

NEMA

NFPA

NC

NESC

NL

NO

NTS

P

PA

PDU

PF

PH

PIU

PMT

PNL

PTAC

PVC

RECEPT or RECP

RF

RH

RMC

RMS

RTU

RVNR

SA

SCCR

SD

SF

S/N

SP

SPD

SPDT

SPEC

SPST

SWBD

SWGR

TMGB

TOL

TP

TYP

U

UH

UG

UIO

UL

UTP

V

VA

VAV

VEF

VM

VRF

W

WC

WEF

WH

WHDM

WSHP

WP

XFMR

Z

HOT WATER RECIRCULATION PUMP

HOT WATER UNIT HEATER

HERTZ

INTRUSION DETECTION SYSTEM

INTERMEDIATE METAL CONDUIT

INFRARED HEATER

JUNCTION BOX

KILO

THOUSAND AMPERE INTERRUPTING CAPACITY

THOUSAND OF CIRCULAR MILS

KITCHEN EXHAUST FAN

KITCHEN HOOD

KILOVOLT

KILOVOLT-AMPERES

KILOWATT

KILOWATT-HOUR

LENGTH

LIGHTNING ARRESTOR

LOCAL AREA NETWORK

LAVATORY

LIGHTING

MAKE-UP AIR UNIT

MAXIMUM

MINIMUM CIRCUIT AMPACITY

MAIN CIRCUIT BREAKER

MOTOR CONTROL CENTER

MAIN GROUND BAR

MANHOLE or MOUNTING HEIGHT

MINIMUM

MAIN LUGS ONLY

MOTOR OPERATED DAMPER

MOUNT or MOUNTED

MOTOR RATED SWITCH

NEUTRAL

NATIONAL ELECTRICAL CODE

NATIONAL FIRE PROTECTION ASSOCIATION

NORMALLY CLOSED

NATIONAL ELECTRICAL SAFETY CODE

NIGHT LIGHT

NORMALLY OPEN

NOT TO SCALE

POLE(S) or PUMP

PUBLIC ADDRESS

POWER DISTRIBUTION UNIT

POWER FACTOR

PHASE

PAD MOUNTED TRANSFORMER

PANEL OR PANELBOARD

PACKAGED THRU-WALL AIR CONDITIONER

POLYVINYL CHLORIDE

RECEPTACLE

RETURN FAN

RADIANT HEATERS (ELECTRIC)

RIGID METAL CONDUIT

ROOT MEAN SQUARE

ROOF TOP UNIT

REDUCED VOLTAGE NON-REVERSING

SURGE ARRESTOR

SHORT CIRCUIT CURRENT RATING

SMOKE DAMPER

SUPPLY FAN

SOLID NEUTRAL

SUMP PUMP

SURGE PROTECTIVE DEVICE

SINGLE POLE DOUBLE THROW

SPECIFICATIONS

SINGLE POLE SINGLE THROW

SWITCHBOARD

SWITCHGEAR

THERMAL OVERLOAD

TWISTED PAIR

TYPICAL

URINAL

UNIT HEATER

UNDERGROUND

UNLESS INDICATED OTHERWISE

UNDERWRITERS LABORATORY

UNSHIELDED TWISTED PAIR

VOLTS

VOLT AMPERES

VAV TERMINAL UNIT

VEHICLE EXHAUST FAN

VOLTMETER

VARIABLE REFRIGERANT FLOW SYSTEM

WATTS or WIRE or WIDTH

WATER CLOSET

WELDING EXHAUST FAN

WATER HEATER

WATTHOUR DEMAND METER

WATER SOURCE HEAT PUMP

WEATHERPROOF

TRANSFORMER

IMPEDANCE

1. NOT ALL ABBREVIATIONS SHOWN WILL BE USED ON THIS

PROJECT.

2. ALL EXISTING EQUIPMENT IS SHOWN IN HALFTONE LINEWORK.

ALL DEMOLISHED EQUIPMENT IS SHOWN IN BOLD LINEWORK,

DASHED AND HATCHED. ALL NEW OR RELOCATED EQUIPMENT

IS SHOWN IN BOLD LINEWORK. BELOW IS AN EXAMPLE OF

EACH:

TELECOMMUNICATIONS MAIN GROUNDING

BUSBAR

VAV TERMINAL UNIT WITH PARALLEL OR

SERIES FAN (PRIMARY INDUCTION UNIT)

NATIONAL ELECTRICAL MANUFACTURER'S

ASSOCIATION

HEATING WATER SUPPLY PUMP IN A PRIMARY-

SECONDARY SYSTEM

CONTRACTOR FURNISHED CONTRACTOR

INSTALLED

CHILLED WATER CIRCULATING PUMP IN A

PRIMARY-SECONDARY SYSTEM

CHILLED WATER SUPPLY PUMP IN A PRIMARY-

SECONDARY SYSTEM

DUCTLESS SPLIT CONDENSING UNIT (POWERS

THE DSS INDOOR UNIT)

DUCTLESS SPLIT HEAT PUMP (POWERS THE

DSS INDOOR UNIT)

GOVERNMENT FURNISHED GOVERNMENT

INSTALLED

HEAT PUMP (OUTDOOR PORTION OF SPLIT

SYSTEM HEAT PUMP, USE CU FOR

STANDARD DX OUTDOOR UNIT)

HEATING WATER CIRCULATING PUMP IN A

PRIMARY-SECONDARY SYSTEM

HEATING WATER SUPPLY IN A PRIMARY-ONLY

SYSTEM

FFU FAN FILTER UNIT

LED LIGHT EMITTING DIODE

POE POWER OVER ETHERNET

IES ILLUMINATING ENGINEERING SOCIETY

WAP WIRELESS ACCESS POINT

ES ELECTRONIC STRIKE

CR CARD READER

AV AUDIO/VISUAL

TV TELEVISION

T-STAT THERMOSTAT

PB PANIC BUTTON

E-001A

2 DIGIT DISCIPLINE

DESIGNATOR

G GENERAL

A ARCHITECTURAL

S STRUCTURAL

F FIRE PROTECTION

E ELECTRICAL

ED ELECTRICAL DEMOLITION

M MECHANICAL

MD MECHANICAL DEMOLITION

MH MECHANICAL HVAC

P PLUMBING

1 DIGIT SHEET TYPE

DESIGNATOR

0 - GENERAL

1 - PLANS

2 - ELEVATIONS

3 - SECTIONS

4 - ENLARGED PLANS

5 - DETAILS

6 - SCHEDULES AND DIAGRAMS

7 - VARIES

8 - VARIES

9 - 3D VIEWS (ISO, PERSPECTIVES)

2 DIGIT SEQUENTIAL # (01-99)

1 DIGIT AREA

DESIGNATOR

(AREA DESIGNATOR

ONLY USED WHEN

PLANS ARE SUBDIVIDED

INTO AREAS.)

DOCUMENTS ARE PROPERTY OF THE U.S. GOVERNMENT.

Phone: (804) 726-9540 https//raymond.global

1601 Summit Avenue Richmond, VA 23230

DATE DESCRIPTION

PROJECT NO.: RVA1033.003

OWNER:

WG HEFNER VAMC

BUILDING 5

PHARMACY CLEAN ROOM

1601 BRENNER AVE

SALISBURY, NC, 28144

SENSITIVE UNCLASSIFIED

DESIGNED BY:

DRAWN BY:

CHECKED BY:

ISSUES / REVISIONS

09/15/20 15% SD SUBMISSION

10/29/20 35% SD SUBMISSION

DATE VA COR SIGN OFF

09/15/20 ANDRE L. NOBS

SUB

15% SD

11/18/20 ANDRE L. NOBS35% SD

12/19/20 65% DD SUBMISSION

02/25/21 95% DD SUBMISSION

2/2/21 ERIC STEWART65% DD

3/12/21 ERIC STEWART95% DD

03/26/21 100% DD SUBMISSION

05/09/21 FINAL BID DOCUMENTS

4/16/21 ANDRE HAYLES100% DD

09/29/21 DILANDRO COMMENTS1

ADRIAN LOZA

ADRIAN LOZA

FLOYD KEELS

E-001

GENERAL ELECTRICAL

NOTES AND ABBREVIATIONS1

1/2

B

W

SB

M

M

2/60/40/1

60/40/3/3R

40/3/3RCB

T

M

EM

J J

PB

G

G4

A

AE

M

X X

C

D a b

LUMINAIRES

LUMINAIRE AND OUTLET BOX. LETTER INDICATES

LUMINAIRE TYPE. SEE LUMINAIRE SCHEDULE.

LUMINAIRE AND OUTLET BOX WIRED FOR

MULTILEVEL SWITCHING, LOWER CASE LETTER

INDICATES SWITCHLEG DESTINATION.

LUMINAIRE AND OUTLET BOX, WITH PROVISIONS

FOR EMERGENCY LIGHTING. LETTER INDICATES

LUMINAIRE TYPE. SEE LUMINAIRE SCHEDULE.

WALL MOUNTED LUMINAIRE AND OUTLET BOX.

LETTER INDICATES LUMINAIRE TYPE. SEE

LUMINAIRE SCHEDULE.

CEILING OR WALL MOUNTED EXIT SIGN AND

OUTLET BOX. PROVIDE NUMBER OF ARROWS

AND FACES INDICATED. LETTER INDICATES

LUMINAIRE TYPE. SEE LUMINAIRE SCHEDULE.

WALL MOUNT AT 96" ABOVE FINISHED FLOOR.

RECESSED OR PENDANT MOUNTED LUMINAIRE

AND OUTLET BOX. LETTER INDICATES LUMINAIRE

TYPE. SEE LUMINAIRE SCHEDULE.

WALL MOUNTED LUMINAIRE AND OUTLET BOX.

LETTER INDICATES LUMINAIRE TYPE. SEE

LUMINAIRE SCHEDULE.

CEILING OR WALL MOUNTED LUMINAIRE AND

OUTLET BOX WITH PROVISIONS FOR

EMERGENCY LIGHTING. LETTER INDICATES

LUMINAIRE TYPE. SEE LUMINAIRE SCHEDULE.

EMERGENCY LIGHT UNIT. LETTER INDICATES

LUMINAIRE TYPE. SEE LUMINAIRE SCHEDULE.

REMOTE HEAD OR FLOOD LUMINAIRE. LETTER

INDICATES LUMINAIRE TYPE. SEE LUMINAIRE

SCHEDULE.

WALL WASH DOWNLIGHT. LETTER INDICATES

LUMINAIRE TYPE. SEE LUMINAIRE SCHEDULE.

POLE MOUNTED LUMINAIRE. LETTER INDICATES

LUMINAIRE TYPE. SEE LUMINAIRE SCHEDULE.

POST-TOP OR BOLLARD LUMINAIRE. LETTER

INDICATES LUMINAIRE TYPE. SEE LUMINAIRE

SCHEDULE.

EQUIPMENT

MOTOR, HORSEPOWER AS INDICATED

MOTORIZED DAMPER

3-POLE COMBINATION MAGNETIC MOTOR

STARTER/ DISCONNECT (NEMA SIZE/ FUSE SIZE/

NEMA ENCLOSURE) NF = NONFUSED

DISCONNECT SWITCH, (SWITCH AMPS/ FUSE

SIZE/ POLES/ NEMA ENCLOSURE) NF = NON-

FUSED

INDIVIDUALLY MOUNTED CIRCUIT BREAKER,

(CIRCUIT BREAKER SIZE/ POLES/ NEMA

ENCLOSURE)

DRY TYPE TRANSFORMER, KVA AS INDICATED

METER

ELECTRONIC METER

PANELBOARD, SEE SCHEDULES

DISTRIBUTION PANELBOARD, SEE SCHEDULES

EQUIPMENT AS INDICATED

CEILING OR WALL MOUNTED JUNCTION BOX

PULL BOX

CEILING MOUNTED BLUE LIGHT SYSTEM

ROTATING BEACON

WHITE NOISE PLENUM SPEAKER; MOUNT ABOVE

DROP CEILING

BLUE LIGHT LOCAL CONTROL SWITCH; MOUNT

48" ABOVE FINISHED FLOOR, UNLESS INDICATED

OTHERWISE

WHITE NOISE ROTARY VOLUME CONTROL;

MOUNT 48" ABOVE FINISHED FLOOR, UNLESS

INDICATED OTHERWISE.

LA-1,3,5

WIRING DEVICES

LOW VOLTAGE ON/OFF WALL SWITCH, WALL

MOUNT 48" ABOVE FINISHED FLOOR. PROVIDE

DIMMER CONTROL WITH DIMMABLE FIXTURES

INTEGRAL WITH SWITCH.

CEILING MOUNTED DUAL TECHNOLOGY

OCCUPANCY SENSOR

CEILING MOUNTED DUAL TECHNOLOGY

VACANCY SENSOR

CEILING MOUNTED DAYLIGHT SENSOR

PHOTOELECTRIC CELL

WIRING

CIRCUIT HOMERUN TO PANELBOARD, LA-1,3,5

ADJACENT TO ARROW INDICATES HOMERUN OF

CIRCUITS, 1,3,5 TO PANEL LA. MARKS ACROSS

RACEWAY INDICATE THE NUMBER OF PHASE

CONDUCTORS AND NEUTRAL IN RACEWAY.

GROUND CONDUCTORS ARE INDICATED BY

LONGER HASHMARKS. NO MARKS ACROSS

RACEWAY INDICATES 2#12 CONDUCTORS AND 1

#12 GROUND CONDUCTOR. CONDUCTOR SIZE #

12 UNLESS INDICATED OTHERWISE. MINIMUM

RACEWAY SIZE SHALL BE 3/4"

RACEWAY EXPOSED TO VIEW

CONCEALED RACEWAY, LOCATED IN WALL OR

ABOVE FINISHED CEILING

UNDERGROUND RACEWAY, LOCATED BELOW

GRADE OR CONCRETE SLAB

FLEXIBLE RACEWAY

RACEWAY TURNED TOWARD VIEWER

RACEWAY TURNED AWAY FROM VIEWER

RACEWAY TERMINATION, STUB-OUT AND CAP

PLUG-IN BUSWAY, UNLESS INDICATED

OTHERWISE

SURFACE METAL RACEWAY OR MULTIOUTLET

ASSEMBLY AS INDICATED

LADDER TYPE CABLE TRAY (SIZE AS INDICATED

ON PLAN)

DUCTBANK, TEXT INDICATES QUANTITY AND

SIZE OF DUCTS (I.E. 2W4" = TWO 4" DUCTS)

LIGHTNING PROTECTION

3/4" x 10'-0" COPPERCLAD GROUND ROD, 18" BFG

GROUND ROD TEST WELL

BARE COPPER GROUND CONDUCTOR, 1/0

UNLESS INDICATED OTHERWISE.

4/0 BARE COPPER GROUND CONDUCTOR

LIGHTNING PROTECTION SYSTEM AIR TERMINAL

LIGHTNING PROTECTION ROOF CONDUCTOR

ELECTRICAL CONNECTION

SITE

ELECTRICAL MANHOLE

ELECTRICAL HANDHOLE

PAD MOUNTED TRANSFORMER

POWER POLE

DOWN GUY AND ANCHOR (QUANTITY AS

INDICATED)

2W4"

WIRING DEVICES

SIMPLEX RECEPTACLE NEMA 5-20R, MOUNT 18"

ABOVE FINISHED FLOOR, UNLESS INDICATED

OTHERWISE.

DUPLEX RECEPTACLE NEMA 5-20R, MOUNT 18"

ABOVE FINISHED FLOOR, UNLESS INDICATED

OTHERWISE

DUPLEX RECEPTACLE, NEMA 5-20R, MOUNT 48"

ABOVE FINISHED FLOOR OR ABOVE COUNTER

TOP/BACKSPLASH

DUPLEX GFI RECEPTACLE NEMA 5-20R, MOUNT

18" ABOVE FINISHED FLOOR, UNLESS INDICATED

OTHERWISE

DUPLEX GFI RECEPTACLE NEMA 5-20R, MOUNT

48" ABOVE FINISHED FLOOR OR ABOVE

COUNTER TOP/BACKSPLASH

DOUBLE-DUPLEX RECEPTACLE NEMA 5-20R,

MOUNT 18" ABOVE FINISHED FLOOR, UNLESS

INDICATED OTHERWISE

SWITCHED DUPLEX RECEPTACLE NEMA 5-20R,

MOUNT 18" ABOVE FINISHED FLOOR, UNLESS

INDICATED OTHERWISE

SWITCHED DUPLEX RECEPTACLE NEMA 5-20R,

MOUNT 48" ABOVE FINISHED FLOOR OR ABOVE

COUNTER TOP/BACKSPLASH

SWITCHED DUPLEX GFI RECEPTACLE NEMA

5-20R, MOUNT 18" ABOVE FINISHED FLOOR,

UNLESS INDICATED OTHERWISE

SWITCHED DUPLEX GFI RECEPTACLE NEMA

5-20R, MOUNT 48" ABOVE FINISHED FLOOR OR

ABOVE COUNTER TOP/BACKSPLASH

DOUBLE-DUPLEX GFI RECEPTACLE, NEMA

5-20R, MOUNTED CONCEALED BEHIND ELECTRIC

WATER COOLER

FLOOR BOX WITH DUPLEX RECEPTACLE NEMA

5-20R. PROVIDE FIRE RATED POKE THRU

DEVICES ON ALL FLOORS ABOVE GROUND

LEVEL.

SPECIAL RECEPTACLE, NEMA TYPE AS

INDICATED, MOUNT 18" ABOVE FINISHED FLOOR,

UNLESS INDICATED OTHERWISE

CEILING BOX WITH DUPLEX RECEPTACLE, NEMA

5-20R

RECEPTACLE WP = IN-USE WEATHER PROOF

SUBSCRIPTS TR = TAMPER RESISTANT

TV = TELEVISION, MOUNT AT 60" ABOVE

FINISHED FLOOR

AFCI = ARC FAULT CIRCUIT INTERRUPTER

FLOOR BOX SUITABLE FOR CONCRETE FLOOR

POUR; TWO COMPARTMENTS - POWER AND

TELECOMMUNICATIONS; FLUSH COVER/FLANGE

WITH HINGED ACCESS TO RECEPTACLES AND

JACKS RECESSED BELOW COVER IN FLOOR

BOX. PROVIDE 8" FIRE RATED POKE THRU

ABOVE GROUND FLOOR.

WALL SWITCH, AC TYPE, SPST, MOUNT 48"

ABOVE FINISHED FLOOR

WALL SWITCH, AC TYPE, DPDT, MOUNT 48"

ABOVE FINISHED FLOOR

3-WAY WALL SWITCH, MOUNT 48" ABOVE

FINISHED FLOOR

4-WAY WALL SWITCH, MOUNT 48" ABOVE

FINISHED FLOOR

SLIDE DIMMER, MOUNT 48" ABOVE FINISHED

FLOOR

MOTOR RATED DISCONNECT SWITCH WITH

THERMAL OVERLOADS, SINGLE POLE SINGLE

THROW, MOUNT ON UNIT, UNLESS INDICATED

OTHERWISE.

WALL MOUNTED, DUAL TECHNOLOGY VACANCY

SWITCH, MOUNT 48" ABOVE FINISHED FLOOR

WALL MOUNTED, DUAL TECHNOLOGY

OCCUPANCY SWITCH, MOUNT 48" ABOVE

FINISHED FLOOR

ELECTRICAL LEGEND

D

M

V

O

LV

OS

DS

PE

5-30R

EWC

SW

S

S

S

S

S

S

S

S

S

VS

D

F

G

H

P

J

LP

TV

COMMUNICATIONS

COAXIAL CABLE TELEVISION OUTLET WALL MOUNTED

60" ABOVE FINISHED FLOOR, UNLESS INDICATED

OTHERWISE

COMMUNICATIONS COMBINATION VOICE/DATA OUTLET

(TWO VOICE & TWO DATA JACKS) MOUNT 18" ABOVE

FINISHED FLOOR, UNLESS INDICATED OTHERWISE

COMMUNICATIONS VOICE OUTLET MOUNT 48" ABOVE

FINISHED FLOOR, UNLESS INDICATED OTHERWISE

COMMUNICATIONS DATA OUTLET, MOUNT 18" ABOVE

FINISHED FLOOR, UNLESS INDICATED OTHERWISE

FLOOR BOX SUITABLE FOR CONCRETE FLOOR POUR;

TWO COMPARTMENTS - POWER AND

TELECOMMUNICATIONS; FLUSH COVER/FLANGE WITH

HINGED ACCESS TO RECEPTACLES AND JACKS

RECESSED BELOW COVER IN FLOOR BOX. PROVIDE 8"

FIRE RATED POKE THRU ABOVE GROUND FLOOR.

CEILING MOUNTED PA SPEAKER, ONE WAY TYPE

WALL MOUNTED PA SPEAKER, ONE WAY TYPE; MOUNT

120" ABOVE FINISHED FLOOR, UNLESS INDICATED

OTHERWISE

CLOSED CIRCUIT TELEVISION CAMERA

CONTROL STATION

NOTE: NOT ALL SYMBOLS WILL BE USED DURING THIS PROJECT

FIRE PROTECTION

FACP

BATT

ANN

SIMPLEX 4100ES FIRE ALARM CONTROL PANEL

BATTERY CABINET

FIRE ALARM ANNUNCIATOR

FIRE ALARM CLEAR VISUAL STROBE AND

SPEAKER. WALL MOUNT AT 80" AFF TO BOTTOM

OF LENS OR 6" BELOW CEILING, WHICHEVER IS

LOWER AND NOT GREATER THAN 96".

WP FIRE ALARM CLEAR VISUAL STROBE. WALL

MOUNT AT 80" AFF TO BOTTOM OF LENS OR 6"

BELOW CEILING, WHICHEVER IS LOWER AND NOT

GREATER THAN 96". WP=WEATHER PROOF

F

MANUAL FIRE ALARM PULL STATION,

MOUNT 42" - 48" AFF TO TOP OF

HANDLE

AC

DC

S

CR

EL

EH

M

IDS

ACS

IR

KP

ML

DR

BR

HSS

EPT

ELECTRONIC SECURITY

ACCESS CONTROL JUNCTION BOX

BALANCED MAGNETIC DOOR CONTACT

ELECTRIC STRIKE

SEISMIC VIBRATION SENSOR

CARD READER

ELECTRIC LOCKSET

ELECTRIC HINGE

MOTION DETECTION SENSOR

INTRUSION DETECTION SYSTEM PANEL

ACCESS CONTROL SYSTEM PANEL

INFRARED REQUEST TO EXIT

IDS KEYPAD

MAGLOCK

DOOR RELEASE

BIOMETRIC READER

HIGH SECURITY SWITCH

ELECTRIC POWER TRANSFER

S

FIRE ALARM DUCT MOUNTED SMOKE

DETECTOR

PA

PA

KEYED REMOTE TEST SWITCH FOR DUCT MOUNTED

SMOKE DETECTOR, MOUNT 48" ABOVE FINISHED

FLOOR UNLESS INDICATED OTHERWISE

DDS

CEILING MOUNTED FIRE ALARM

CLEAR VISUAL STROBE

COAXIAL CABLE TELEVISION OUTLET CEILING MOUNTED

WAP WIRELESS ACCESS POINT

ES

PB PANIC BUTTON

2 WAY COMMUNICATION CAMERA 1

DOCUMENTS ARE PROPERTY OF THE U.S. GOVERNMENT.

Phone: (804) 726-9540 https//raymond.global

1601 Summit Avenue Richmond, VA 23230

DATE DESCRIPTION

PROJECT NO.: RVA1033.003

OWNER:

WG HEFNER VAMC

BUILDING 5

PHARMACY CLEAN ROOM

1601 BRENNER AVE

SALISBURY, NC, 28144

SENSITIVE UNCLASSIFIED

DESIGNED BY:

DRAWN BY:

CHECKED BY:

ISSUES / REVISIONS

09/15/20 15% SD SUBMISSION

10/29/20 35% SD SUBMISSION

DATE VA COR SIGN OFF

09/15/20 ANDRE L. NOBS

SUB

15% SD

11/18/20 ANDRE L. NOBS35% SD

12/19/20 65% DD SUBMISSION

02/25/21 95% DD SUBMISSION

2/2/21 ERIC STEWART65% DD

3/12/21 ERIC STEWART95% DD

03/26/21 100% DD SUBMISSION

05/09/21 FINAL BID DOCUMENTS

4/16/21 ANDRE HAYLES100% DD

09/29/21 DILANDRO COMMENTS1

ADRIAN LOZA

ADRIAN LOZA

FLOYD KEELS

E-002

GENERAL ELECTRICAL

SYMBOL LEGEND1

VENDING

DARK RM

CORRIDOR

1021E

OFFICE

CLOSET

CLOSET

BIOMED SHOP

CLOSET

CORRIDOR

STAIRS

OFFICE

MOP

RECEIV. VEST.

CART CHAR.

WASH RM

GARB. REFR

SUPPLY REFRIG

DRY STORAGE

REFRIG

VEG PREP

1021B

OFFICE

1021C

OFFICE

CLOSET

TOILET

OFFICE

OFFICE

CLOSET

TOILET

1012A

TELE. ROOM

FAC

STAIRS

MEN'S TOILET

WOMEN'S TOILET

CLOSET

M/F TOILET

TRAVEL

AGENT CASH

CORRIDOR

WAITING

STORAGE

STORAGE

MEDICAL MEDIA

MECH. ROOM

MECH. ROOM

FREEZOR

REFRIG

STORAGE

PROJECT AREA PANEL 5F

PANEL 5D

PANEL 5DA

1062 1061

1055 1056

GYNECOLOGY CLINIC

WAITING ROOM

DOCUMENTS ARE PROPERTY OF THE U.S. GOVERNMENT.

Phone: (804) 726-9540 https//raymond.global

1601 Summit Avenue Richmond, VA 23230

DATE DESCRIPTION

PROJECT NO.: RVA1033.003

OWNER:

WG HEFNER VAMC

BUILDING 5

PHARMACY CLEAN ROOM

1601 BRENNER AVE

SALISBURY, NC, 28144

SENSITIVE UNCLASSIFIED

DESIGNED BY:

DRAWN BY:

CHECKED BY:

ISSUES / REVISIONS

09/15/20 15% SD SUBMISSION

10/29/20 35% SD SUBMISSION

DATE VA COR SIGN OFF

09/15/20 ANDRE L. NOBS

SUB

15% SD

11/18/20 ANDRE L. NOBS35% SD

12/19/20 65% DD SUBMISSION

02/25/21 95% DD SUBMISSION

2/2/21 ERIC STEWART65% DD

3/12/21 ERIC STEWART95% DD

03/26/21 100% DD SUBMISSION

05/09/21 FINAL BID DOCUMENTS

4/16/21 ANDRE HAYLES100% DD

09/29/21 DILANDRO COMMENTS1

ADRIAN LOZA

ADRIAN LOZA

FLOYD KEELS

E-003

GENERAL ELECTRICAL KEY PLAN

SCALE: 1/16" = 1'-0"A FIRST FLOOR KEY PLAN

PROJECT NORTH

A. PRIOR TO SUBMITTING A PROPOSAL THE ELECTRICAL CONTRACTOR SHALL VISIT THE SITE AND CAREFULLY INVESTIGATE THE EXISTING AREAS AFFECTED BY THIS WORK IN ORDER TO BECOME FAMILIAR WITH EXISTING CONDITIONS AND THE DIFFICULTIES THAT WILL AFFECT THE EXECUTION OF THIS WORK. THE CONTRACTOR SHALL BASE HIS BID ACCORDINGLY.

SUBMISSION OF PROPOSAL WILL BE CONSTRUED AS CONFIRMATION THAT A THOROUGH EXAMINATION OF THE SITE HAS BEEN MADE BY THE CONTRACTOR. LATER CLAIMS FOR UNFORESEEN EXTRA LABOR, EQUIPMENT, OR MATERIALS WILL NOT BE ACCEPTED IF SAID CLAIMS COULD HAVE BEEN FORESEEN DURING THE SITE INVESTIGATION.

B. CONTRACTORS WORK SHALL COMPLY WITH ALL SAFETY RELATED WORK PRACTICES DETAILED IN THE LASTEST EDITION OF

NFPA 70E.

C. THE ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR REMOVAL OF ALL ELECTRICAL FACILITIES RENDERED INACTIVE OR OBSOLETE BY THE SCOPE OF THE ARCHITECTURAL, MECHANICAL, AND PLUMBING DEMOLITION WORK. THE CONTRACT IS ALTERED THAT SUCH ELECTRICAL FACILITIES MAY NOT BE EXPLICITLY SHOWN ON THE ARCHITECTURAL DEMOLITION DRAWINGS. THE CONTRACTOR SHALL VISIT THE SITE, REFER TO THE ARCHITECTURAL MECHANICAL, ELECTRICAL, AND PLUMBING DRAWINGS, AND COORDINATE WITH THE GENERAL, MECHANICAL, PLUMBING CONTRACTORS. THE CONTRACT INCLUDES ALL ELECTRICAL DEMOLITION WORK, HOWEVER NOT NECESSARILY LIMITED TO, THE COMPLETE ELECTRICAL AND CONTROL WIRE, CONDUIT, JUNCTION BOXES, ENCLOSURE, AND RACEWAY SUPPORT SYSTEM, DISCONNECT SWITCHES, CIRCUIT BREAKERS, ETC INTENDED FOR REMOVAL AS REQUIRED TO COMPLETE THE WORK.

D. THE INTENT OF THE DEMOLITION WORK REQUIREMENTS INCLUDE ALL LABOR AND EQUIPMENT REQUIRED TO ATTAIN THE FINAL CONDITION AS SHOWN ON THE ARCHITECTURAL , MECHANICAL, PLUMBING, STRUCTURAL, AND ELECTRICAL NEW WORK PLANS. THE CONTRACTOR SHALL REVIEW ALL OTHER TRADE DRAWINGS, AND COORDINATE WITH ALL OTHER TRADE

CONTRACTORS, AND BASE HIS BID ACCORDINGLY.

E. WHERE THE REMOVAL OF EXISTING FACILITIES RESULT IN THE DE-ENERGIZATION OF EXISTING REMAINING FACILITIES, THE CONTRACTOR SHALL INSTALL JUNCTION BOXES AND OR OTHER DEVICES, PROVIDE BYPASS CONNECTIONS NECESSARY TO MAKE THE EFFECTED CIRCUITS CONTINUOUS AND READY FOR OPERATION OR AS OTHERWISE INDICATED OR INTENDED.

F. THE CONTRACTOR SHALL REMOVE AND/OR RELOCATE ALL EXISTING ELECTRICAL FACILITIES THAT INTERFERE WITH THE NEW ARCHITECTURAL, MECHANICAL, PLUMBING, STRUCTURAL, ELECTRICAL LAYOUTS AND SCHEMES. REFER TO THE ARCHITECTURAL, MECHANICAL, PLUMBING, STRUCTURAL, AND ELECTRICAL DRAWINGS AND COORDINATE WITH THE

GENERAL, MECHANICAL, PLUMBING, STRUCTURAL CONTRACTORS.

G. PRIOR TO THE COMMENCEMENT OF WORK, ELECTRICAL POWER SHALL BE DISCONNECTED AND OR SWITCHED OFF AS REQUIRED. TEMPORARY LIGHTING AND POWER FOR ALL TRADES, REMAINING BUILDING OCCUPANTS, AND CRITICAL BUILDING SERVICES SHALL BE PROVIDED AS REQUIRED WITHOUT EXCEPTION. UNDER NO CIRCUMSTANCES SHALL POWER BE INTERRUPTED TO ANY AREA WITHOUT PRIOR WRITTEN APPROVAL FROM BUILDING MANAGEMENT.

H. TEMPORARY LIGHT STRINGERS WHERE SPLICED ARE TO EMPLOY COMPRESSION TYPE FITTINGS OR SOLDERED

CONNECTIONS AND MADE UP NEATLY AND SAFELY AS REQUIRED.

I. MAINTAIN CONTINUOUS ELECTRICAL SERVICE TO ALL ACTIVE AREAS AT ALL TIMES EXCEPT WHERE GIVEN WRITTEN PERMISSION BY BUILDING PROPERTY MANAGEMENT FOR A SCHEDULED OUTAGE FOR A DECLARED OUTAGE DURATION.

J. REMOVE EXPOSED CONDUITS, WIREWAYS, OUTLET BOXES, HANGERS, SUPPORTS, AND DEVICES MADE OBSOLETE BY THIS WORK UNLESS BEING REUTILIZED FOR THE NEW INSTALLATION. THE REUTILIZATION OF EXISTING FACILITIES BY THE CONTRACTOR SHALL RENDER A FINISHED INSTALLATION IN STRICT ACCORDANCE WITH THE NEC WITHOUT EXCEPTION.

K. ALL ADJACENT FACILITIES IMPACTED OR TEMPORARILY DISCONNECTED TO FACILITATE DEMOLITION WORK SHALL BE RECONNECTED AND RESTORED TO A CONDITION EQUAL TO OR BETTER THAN ORIGINALLY FOUND PRIOR TO THE COMMENCEMENT OF WORK. WHERE EXISTING CONDITIONS DO NOT MEET NEC REQUIREMENTS, THE CONTRACTOR SHALL

NOTIFY THE CONSTRUCTION MANAGER PRIOR TO THE COMMENCEMENT OF WORK.

L. PRIOR TO THE COMMENCEMENT OF WORK THE CONTRACTOR SHALL TRACE EXISTING CIRCUITS AND CONFIRM OR IDENTIFY

TYPE AND LOCATION OF LOAD SERVED.

M. WHERE EXISTING ELECTRICAL PANELS ARE CALLED FOR REMOVAL AND OR REPLACEMENT, THE CONTRACTOR SHALL REMOVE ALL SERVING FEEDER WIRE AND CONDUIT BACK TO ITS RESPECTIVE SOURCE OF POWER OR AS OTHERWISE

INDICATED.

N. ALL OPEN FLOOR OUTLETS SHALL BE PLUGGED. ANY AND ALL PANEL ENCLOSURE OR EQUIPMENT TERMINAL BOX COMPARTMENTS WITH UNOCCUPIED KNOCKOUTS, RACEWAY OR CABLE ASSEMBLY PENETRATIONS SHALL BE PLUGGED WITH A LISTED DEVICE. ANY AND ALL UNOCCUPIED BREAKER SPACES IN EXISTING AND OR NEW PANELS SHALL BE PLUGGED WITH A LISTED DEVICE OR FURNISHED WITH A SPARE, OPERABLE CIRCUIT BREAKER LISTED FOR USE IN THE PANEL. PROVIDE

LISTED PLATES ON ALL UNUSED OUTLET AND JUNCTION BOXES.

O. ELECTRICAL PANEL COVERS ARE NOT TO BE LEFT OFF AT ANY TIME UNLESS MEN ARE WORKING ON, ADEQUATELY PROTECTING AND BARRICADING THEM AS REQUIRED. COVERS SHALL BE REPLACED EACH NIGHT BEFORE LEAVING JOB SITE.

P. REMOVE AND PROPERLY DISPOSE OF ALL REMOVALS IN ACCORDANCE WITH APPLICABLE LAWS AND ENVIRONMENTAL RESTRICTIONS UNLESS OTHERWISE REQUESTED OR DIRECTED BY THE BUILDING PROPERTY MANAGER.

Q. THE REMOVAL AND OR RELOCATION OF ALL FIRE ALARM COMMUNICATIONS, DATA AND SECURITY EQUIPMENT AND ASSOCIATED CABLING SHALL BE COORDINATED WITH BUILDING OPERATIONAL PERSONNEL. EXISTING BASE BUILDING FIRE ALARM SYSTEM (AND FLOOR SECURITY SYSTEM WHERE APPLICABLE) INTEGRITY SHALL BE MAINTAINED AT ALL TIMES

BEFORE, DURING, AND AFTER DEMOLITION.

R. CONTRACTOR SHALL PERFORM THE FOLLOWING PRIOR TO START ANY TRADES DEMOLITION:

R.A. TRACE ALL INCOMING AND OUTGOING FEEDS (LOW VOLTAGE AND HIGH VOLTAGE) AS WELL AS LIGHTING CIRCUITS BACK

TO THEIR ORIGINAL SOURCE OF POWER (I.E. SWITCHBOARDS, PANELBOARDS, ETC.)

R.B. IDENTIFY ALL CONDUITS, CABLES, ETC. WITH CLEARLY IDENTIFIABLE MARKINGS. IDENTIFICATION SHALL BE PROVIDED

FROM EACH LOAD TO THE MOST UPSTREAM SOURCE PANEL. IDENTIFICATION SHALL BE PROVIDED AT ALL JUNCTION

AND PULL BOXES, TERMINATIONS/SPLICES AND PANEL LOCATIONS AS REQUIRED.

R.C. REVIEW LOCATIONS, CONFIGURATION, AND SERVICE LOCATIONS OF ALL FEEDER AND BRANCH CIRCUIT SOURCES WITH CONSTRUCTION MANAGER/GENERAL CONTRACTOR SO AS TO ENSURE SERVICES WILL REMAIN SERVICES OR INTERRUPTED AS ANTICIPATED. REFER TO EXISTING BUILDING RECORD DRAWINGS PRIOR TO COMMENCEMENT OF

WORK.

R.D. COORDINATE ALL WORK WITH BUILDING PROPERTY MANAGEMENT.

R.E. EXISTING ELECTRICAL EQUIPMENT FACILITIES WHEN REUSED, SHALL BE THOROUGHLY CLEANED AND REFURBISHED

PRIOR TO RE-USE. ANY AND ALL EXISTING RE-USED SET SCREW TYPE CONDUIT CONNECTORS SHALL BE THOROUGHLY

TIGHTENED.

ELECTRICAL DEMOLITION NOTES:

PANEL 5D

SS

S S

1TYP.

TYP.

2TYP.

1 TYP.

PANEL 5DA

4WAP

J

S1

DOCUMENTS ARE PROPERTY OF THE U.S. GOVERNMENT.

Phone: (804) 726-9540 https//raymond.global

1601 Summit Avenue Richmond, VA 23230

DATE DESCRIPTION

PROJECT NO.: RVA1033.003

OWNER:

WG HEFNER VAMC

BUILDING 5

PHARMACY CLEAN ROOM

1601 BRENNER AVE

SALISBURY, NC, 28144

SENSITIVE UNCLASSIFIED

DESIGNED BY:

DRAWN BY:

CHECKED BY:

ISSUES / REVISIONS

09/15/20 15% SD SUBMISSION

10/29/20 35% SD SUBMISSION

DATE VA COR SIGN OFF

09/15/20 ANDRE L. NOBS

SUB

15% SD

11/18/20 ANDRE L. NOBS35% SD

12/19/20 65% DD SUBMISSION

02/25/21 95% DD SUBMISSION

2/2/21 ERIC STEWART65% DD

3/12/21 ERIC STEWART95% DD

03/26/21 100% DD SUBMISSION

05/09/21 FINAL BID DOCUMENTS

4/16/21 ANDRE HAYLES100% DD

09/29/21 DILANDRO COMMENTS1

ADRIAN LOZA

ADRIAN LOZA

FLOYD KEELS

ED101

ELECTRICAL DEMOLITION PLAN

SCALE: 1/8" = 1'-0"1 ELECTRICAL DEMOLITION PLAN

PROJECT NORTH

1. THESE DRAWINGS SHOW THE INTENT OF THE

DEMOLITION AND THE NEW CIRCUITING DESIGN FOR THE

AREAS INDICATED. IT IS THE ELECTRICAL CONTRACTOR'S

RESPONSIBILITY TO FOLLOW THIS INTENT WHILE

ADAPTING TO THE EXISTING CONDITIONS FOUND IN THE

FIELD.

2. DURING EACH PHASE OF CONSTRUCTION, THE ELECTRICAL

CONTRACTOR SHALL TAKE CARE TO MAINTAIN

CONTINUITY TO ALL CIRCUITS NOT PART OF THE PHASE

INDICATED OR THE FINAL CONSTRUCTION.

3. IF OUTAGES ARE REQUIRED IN AREAS NOT BEING

RENOVATED, THE CONTRACTOR SHALL COORDINATE WITH

THE COR AND PROVIDE AT LEAST 72 HOUR NOTICE

BEFORE THE OUTAGE TAKES PLACE. SHOULD AN

UNPLANNED OUTAGE OCCUR, RESTORING POWER WILL

BE THE PRIORITY.

4. CARE MUST BE TAKEN TO ASSURE EMERGENCY LIGHTING

AND/OR SYSTEMS ARE MAINTAINED THROUGHOUT

CONSTRUCTION.

5. ELECTRICAL CONTRACTOR SHALL FIELD VERIFY EXISTING

PANEL CAPACITY PRIOR TO INSTALLATION OF NEW LOADS.

DO NOT EXCEED 80% OF MAIN CIRCUIT BREAKER

AMPACITY RATING. PROVIDE NEW WIRING, ACCESSORIES,

AND CIRCUIT BREAKERS NECESSARY FOR COMPLETE

INSTALLATION.

6. ELECTRICAL CONTRACTOR SHALL PROVIDE UPDATED

PANEL DIRECTORIES FOR ALL PANELS AFFECTED DURING

CONSTRUCTION.

7. ELECTRICAL CONTRACTOR SHALL CIRCUIT ELECTRICAL

DEVICES TO SPARE CIRCUIT BREAKERS & CIRCUITS THAT

BECOME AVAILABLE DURING DEMOLITION.

8. FIRE ALARM RELATED WORK SHALL BE PERFORMED BY A

LICENSED FIRE ALARM CONTRACTOR.

1. DISCONNECT AND REMOVE EXISTING ELECTRICAL DEVICE.

PULLBACK ALL CONDUCTORS TO THE NEAREST JUNCTION

BOX WHERE CIRCUIT ENTERS THE ROOM. MODIFY

EXISTING WIRING TO MAINTAIN CONTINUITY OF ANY

ADJACENT DEVICES AS REQUIRED.

2. DISCONNECT AND REMOVE EXISTING LIGHTING FIXTURE.

PULLBACK ALL CONDUCTORS TO THE NEAREST JUNCTION

BOX INSIDE CEILING. MODIFY EXISTING WIRING TO

MAINTAIN CONTINUITY OF ANY ADJACENT DEVICES AS

REQUIRED.

3. DISCONNECT AND REMOVE EXISTING JUNCTION BOX.

MODIFY EXISTING WIRING TO MAINTAIN CONTINUITY OF

ANY ADJACENT DEVICES AS REQUIRED.

4. DISCONNECT AND REMOVE EXISTING FIRE ALARM STROBE

DEVICE. PULL BACK ALL CONDUCTORS TO THE NEAREST

JUNCTION BOX AND FEED TO NEW FIRE ALARM DEVICES

AS SHOWN ON NEW ELECTRICAL WORK PLAN.

5. EXISTING WIRELESS ACCESS POINT SHALL BE

DISCONNECTED AND REMOVED DURING DEMOLITION

AND RECONNECTED AFTER CONSTRUCTION. SEE NEW

ELECTRICAL CEILING PLAN FOR NEW LOCATION (E-102).

ELECTRICAL CONTRACTOR SHALL STORE AND PROTECT

WAP IN A SECURE LOCATION DURING CONSTRUCTION.

ENSURE WAP IS IN GOOD WORKING CONDITION AFTER

REINSTALLATION.

GENERAL NOTES

KEY NOTES

PANEL 5D

5F-42

J

J

5F-41

5D-41

5F-41

5F-6

5D-45

5D-36

J 5D-45

5D-32

5F-41

5F-41

5D-47

5D-47

5D-47

J 5D-36

5F-42

5F-42

5F-6

2SDD

CR

5D-34

PANEL 5DA

TYP.

TYP.

TYP.

TYP.

J 5D-36

TYP.

J

ES

F

1TYP.

1TYP.

5D-39

5F-41

5D-36

5D-36

5F-41

5F-42

5F-42

5F-42

5F-6

5F-6

5F-6

5F-6

5F-6

PB

8 TYP.

1. THESE DRAWINGS SHOW THE INTENT OF THE

DEMOLITION AND THE NEW CIRCUITING DESIGN FOR THE

AREAS INDICATED. IT IS THE ELECTRICAL CONTRACTOR'S

RESPONSIBILITY TO FOLLOW THIS INTENT WHILE

ADAPTING TO THE EXISTING CONDITIONS FOUND IN THE

FIELD.

2. DURING EACH PHASE OF CONSTRUCTION, THE ELECTRICAL

CONTRACTOR SHALL TAKE CARE TO MAINTAIN

CONTINUITY TO ALL CIRCUITS NOT PART OF THE PHASE

INDICATED OR THE FINAL CONSTRUCTION.

3. IF OUTAGES ARE REQUIRED IN AREAS NOT BEING

RENOVATED, THE CONTRACTOR SHALL COORDINATE WITH

THE COR AND PROVIDE AT LEAST 72 HOUR NOTICE

BEFORE THE OUTAGE TAKES PLACE. SHOULD AN

UNPLANNED OUTAGE OCCUR, RESTORING POWER WILL

BE THE PRIORITY.

4. CARE MUST BE TAKEN TO ASSURE EMERGENCY LIGHTING

AND/OR SYSTEMS ARE MAINTAINED THROUGHOUT

CONSTRUCTION.

5. ELECTRICAL CONTRACTOR SHALL FIELD VERIFY EXISTING

PANEL CAPACITY PRIOR TO INSTALLATION OF NEW LOADS.

DO NOT EXCEED 80% OF MAIN CIRCUIT BREAKER

AMPACITY RATING. PROVIDE NEW WIRING, ACCESSORIES,

AND CIRCUIT BREAKERS NECESSARY FOR COMPLETE

INSTALLATION.

6. ELECTRICAL CONTRACTOR SHALL PROVIDE UPDATED

PANEL DIRECTORIES FOR ALL PANELS AFFECTED DURING

CONSTRUCTION.

7. ELECTRICAL CONTRACTOR SHALL CIRCUIT ELECTRICAL

DEVICES TO SPARE CIRCUIT BREAKERS & CIRCUITS THAT

BECOME AVAILABLE DURING DEMOLITION.

8. FIRE ALARM RELATED WORK SHALL BE PERFORMED BY A

LICENSED FIRE ALARM CONTRACTOR.

GENERAL NOTES

KEY NOTES

1. INSTALL NEW CLEAN ROOM GRADE JUNCTION BOX AND

DUPLEX RECEPTACLE. TYPICAL FOR ALL RECEPTACLES. SEE

E-501 - DETAIL 1 & DETAIL 4.

2. INSTALL NEW CLEAN ROOM GRADE JUNCTION BOX AND

DUPLEX RECEPTACLE FOR HOOD/WORKSTATION OR

REFRIGERATOR AND CIRCUIT BACK TO PANEL. SEE E-501 -

DETAIL 1 & DETAIL 4.

3. INSTALL THREE NEW DUPLEX RECEPTACLES FOR QAS OR

EQUAL REAL TIME MONITORING SYSTEM AND CIRCUIT

BACK TO PANEL. SEE E-501 - DETAIL 1 AND DETAIL 4.

4. INSTALL NEW WALL MOUNTED CLEAN ROOM GRADE

JUNCTION BOX WITH STAINLESS STEEL COVER PLATE FOR

HARD WIRED CONNECTION TO LAVATORY. CIRCUIT TO GFI

RECEPTACLE (LINE-SIDE) AND BACK TO PANEL. SEE E-501 -

DETAIL 4.

5. INSTALL NEW SECURITY CARD READER AND FAIL-SECURE

ELECTRONIC STRIKE AND TIE INTO EXISTING PHYSICAL

ACCESS CONTROL SYSTEM. REFER TO ARCHITECTURAL

SHEET A-002 FOR DOOR HARDWARE SPECIFICATIONS.

6. INSTALL NEW JUNCTION BOX IN CEILING FOR HARD WIRED

CONNECTION TO MOTORIZED AUTOMATIC DOOR

OPERATOR AND MOTION SENSOR. SEE E-501 - DETAIL 4.

REFER TO ARCHITECTURAL SHEET A-002 FOR DOOR

HARDWARE SPECIFICATIONS.

7. INSTALL NEW CLEAN ROOM GRADE TELEPHONE & DATA

COMBINATION OUTLET AND PROVIDE 3 4" CONDUIT TO

CEILING SPACE. SEE E-501 - DETAIL 3.

8. INSTALL NEW TWO WAY COMMUNICATION CAMERA AND

TIE INTO EXISTING COMMUNICATION SYSTEM.

COORDINATE INSTALLATION AND FINAL LOCATION WITH

COR.

9. INSTALL NEW CCTV SURVEILLANCE CAMERA AND TIE INTO

EXISTING CCTV VIDEO SURVEILLANCE SYSTEM.

COORDINATE INSTALLATION AND FINAL LOCATION WITH

COR.

10. INSTALL NEW FIRE ALARM PULL HANDLE AND TIE INTO

EXISTING FIRE ALARM CONTROL PANEL NOTIFICATION

CIRCUIT. INSTALL NEW PANIC BUTTON AND TIE INTO

EXISTING SECURITY NOTIFICATION SYSTEM.

11. INSTALL NEW FIRE ALARM SPEAKER AND/OR STROBE

DEVICE AND TIE INTO EXISTING FIRE ALARM CONTROL

PANEL NOTIFICATION CIRCUIT.

12. INSTALL NEW SMOKE DUCT DETECTOR (SEE MECHANICAL

SHEET MH101 FOR LOCATION) AND KEYED REMOTE TEST

SWITCH AND TIE INTO EXISTING FIRE ALARM CONTROL

PANEL NOTIFICATION CIRCUIT.

DOCUMENTS ARE PROPERTY OF THE U.S. GOVERNMENT.

Phone: (804) 726-9540 https//raymond.global

1601 Summit Avenue Richmond, VA 23230

DATE DESCRIPTION

PROJECT NO.: RVA1033.003

OWNER:

WG HEFNER VAMC

BUILDING 5

PHARMACY CLEAN ROOM

1601 BRENNER AVE

SALISBURY, NC, 28144

SENSITIVE UNCLASSIFIED

DESIGNED BY:

DRAWN BY:

CHECKED BY:

ISSUES / REVISIONS

09/15/20 15% SD SUBMISSION

10/29/20 35% SD SUBMISSION

DATE VA COR SIGN OFF

09/15/20 ANDRE L. NOBS

SUB

15% SD

11/18/20 ANDRE L. NOBS35% SD

12/19/20 65% DD SUBMISSION

02/25/21 95% DD SUBMISSION

2/2/21 ERIC STEWART65% DD

3/12/21 ERIC STEWART95% DD

03/26/21 100% DD SUBMISSION

05/09/21 FINAL BID DOCUMENTS

4/16/21 ANDRE HAYLES100% DD

09/29/21 DILANDRO COMMENTS1

ADRIAN LOZA

ADRIAN LOZA

FLOYD KEELS

E-101SCALE: 1/4" = 1'-0"2 ELECTRICAL FLOOR PLAN

NEW ELECTRICAL FLOOR PLAN

PROJECT NORTH

LEDLED

LED

LED

LED

5D-37

PANEL 5D

C

PA

PA

PANEL 5DA

WAP

RELOCATED WAP

SEE ED101 - KEY NOTE 5

OS

OS

SD

a aS3 a

FFU

LED

LED

5D-38 dB

5D-37 dB

5D-37

7TYP.

5TYP.

3TYP.

LED

SD

b bB

5D-37 aA

5D-37 aA

5D-37 aA

5D-37 aA

5D-37 aA

5D-37 a

OS

d

SD

cOS c

OS

e

TYP.

TYP.

TYP.

TYP.

7TYP.

5TYP.

5TYP.

5 TYP.

5D-37 C

5D-37 C

SD

e

SD

d

TYP.

FFU

5D-42

LED

bB

5D-37

OS

b

FFU

5D-38

FFU

5D-42

FFU

5D-38

FFU

5D-38

LED

dB

5D-37

LED

dB

5D-37

FFU

5D-38

FFU

5D-42

FFU

5D-42

FFU

5D-42

LED

cB

5D-37

LED

cB

5D-37

LED

cB

5D-37

FFU

5D-40

FFU

5D-40

FFU

5D-40

FFU

5D-40

LED

eB

5D-37

LED

eB

5D-37

LED

eB

5D-37

LED

eB

5D-37

LED

eB

5D-37

1. THESE DRAWINGS SHOW THE INTENT OF THE

DEMOLITION AND THE NEW CIRCUITING DESIGN FOR THE

AREAS INDICATED. IT IS THE ELECTRICAL CONTRACTOR'S

RESPONSIBILITY TO FOLLOW THIS INTENT WHILE

ADAPTING TO THE EXISTING CONDITIONS FOUND IN THE

FIELD.

2. DURING EACH PHASE OF CONSTRUCTION, THE ELECTRICAL

CONTRACTOR SHALL TAKE CARE TO MAINTAIN

CONTINUITY TO ALL CIRCUITS NOT PART OF THE PHASE

INDICATED OR THE FINAL CONSTRUCTION.

3. IF OUTAGES ARE REQUIRED IN AREAS NOT BEING

RENOVATED, THE CONTRACTOR SHALL COORDINATE WITH

THE COR AND PROVIDE AT LEAST 72 HOUR NOTICE

BEFORE THE OUTAGE TAKES PLACE. SHOULD AN

UNPLANNED OUTAGE OCCUR, RESTORING POWER WILL

BE THE PRIORITY.

4. CARE MUST BE TAKEN TO ASSURE EMERGENCY LIGHTING

AND/OR SYSTEMS ARE MAINTAINED THROUGHOUT

CONSTRUCTION.

5. ELECTRICAL CONTRACTOR SHALL FIELD VERIFY EXISTING

PANEL CAPACITY PRIOR TO INSTALLATION OF NEW LOADS.

DO NOT EXCEED 80% OF MAIN CIRCUIT BREAKER

AMPACITY RATING. PROVIDE NEW WIRING, ACCESSORIES,

AND CIRCUIT BREAKERS NECESSARY FOR COMPLETE

INSTALLATION.

6. ELECTRICAL CONTRACTOR SHALL PROVIDE UPDATED

PANEL DIRECTORIES FOR ALL PANELS AFFECTED DURING

CONSTRUCTION.

7. ELECTRICAL CONTRACTOR SHALL CIRCUIT ELECTRICAL

DEVICES TO SPARE CIRCUIT BREAKERS & CIRCUITS THAT

BECOME AVAILABLE DURING DEMOLITION.

8. FIRE ALARM RELATED WORK SHALL BE PERFORMED BY A

LICENSED FIRE ALARM CONTRACTOR.

GENERAL NOTES

KEY NOTES

1. INSTALL NEW 2' X 2' LAY-IN PA SPEAKER AND TIE INTO

EXISTING PUBLIC ADDRESS SYSTEM.

2. INSTALL NEW CLEAN ROOM GRADE JUNCTION BOX AND

WALL MOUNTED SWITCH WITH STAINLESS STEEL COVER

PLATE AND CIRCUIT TO NEW LIGHT FIXTURES AND

OCCUPANCY SENSORS AS DESIGNATED. SEE E-501 - DETAIL

2 AND DETAIL 4.

3. INSTALL NEW CEILING MOUNTED OCCUPANCY SENSOR

SWITCH AND CIRCUIT TO NEW LIGHT FIXTURES AS

DESIGNATED.

4. INSTALL NEW 2' X 2' LAY-IN TROFFER LED. SEE E-601 -

LUMINAIRE SCHEDULE AND E-501 - DETAIL 5.

5. CIRCUIT NEW 2' X 2' LAY-IN TROFFER LED (PROVIDED AND

INSTALLED BY OTHERS) AS DESIGNATED. SEE E-601 -

LUMINAIRE SCHEDULE.

6. INSTALL NEW WALL OR CEILING MOUNTED LED EXIT SIGN.

SEE E-601 - LUMINAIRE SCHEDULE AND E-501 - DETAIL 6.

7. CIRCUIT NEW FAN FILTER UNIT (PROVIDED AND INSTALLED

BY OTHERS) AS DESIGNATED.

DOCUMENTS ARE PROPERTY OF THE U.S. GOVERNMENT.

Phone: (804) 726-9540 https//raymond.global

1601 Summit Avenue Richmond, VA 23230

DATE DESCRIPTION

PROJECT NO.: RVA1033.003

OWNER:

WG HEFNER VAMC

BUILDING 5

PHARMACY CLEAN ROOM

1601 BRENNER AVE

SALISBURY, NC, 28144

SENSITIVE UNCLASSIFIED

DESIGNED BY:

DRAWN BY:

CHECKED BY:

ISSUES / REVISIONS

09/15/20 15% SD SUBMISSION

10/29/20 35% SD SUBMISSION

DATE VA COR SIGN OFF

09/15/20 ANDRE L. NOBS

SUB

15% SD

11/18/20 ANDRE L. NOBS35% SD

12/19/20 65% DD SUBMISSION

02/25/21 95% DD SUBMISSION

2/2/21 ERIC STEWART65% DD

3/12/21 ERIC STEWART95% DD

03/26/21 100% DD SUBMISSION

05/09/21 FINAL BID DOCUMENTS

4/16/21 ANDRE HAYLES100% DD

09/29/21 DILANDRO COMMENTS1

ADRIAN LOZA

ADRIAN LOZA

FLOYD KEELS

E-102SCALE: 1/4" = 1'-0"3 ELECTRICAL CEILING PLAN

NEW ELECTRICAL CEILING PLAN

PROJECT NORTH

AHU-49

BMEF-1

30/NF/3/3R

30/NF/3/3R5D- 31, 33, 35

5D- 44, 46, 48

EXISTING TO REMAIN

1 1

1. THESE DRAWINGS SHOW THE INTENT OF THE

DEMOLITION AND THE NEW CIRCUITING DESIGN FOR THE

AREAS INDICATED. IT IS THE ELECTRICAL CONTRACTOR'S

RESPONSIBILITY TO FOLLOW THIS INTENT WHILE

ADAPTING TO THE EXISTING CONDITIONS FOUND IN THE

FIELD.

2. DURING EACH PHASE OF CONSTRUCTION, THE ELECTRICAL

CONTRACTOR SHALL TAKE CARE TO MAINTAIN

CONTINUITY TO ALL CIRCUITS NOT PART OF THE PHASE

INDICATED OR THE FINAL CONSTRUCTION.

3. IF OUTAGES ARE REQUIRED IN AREAS NOT BEING

RENOVATED, THE CONTRACTOR SHALL COORDINATE WITH

THE COR AND PROVIDE AT LEAST 72 HOUR NOTICE

BEFORE THE OUTAGE TAKES PLACE. SHOULD AN

UNPLANNED OUTAGE OCCUR, RESTORING POWER WILL

BE THE PRIORITY.

4. CARE MUST BE TAKEN TO ASSURE EMERGENCY LIGHTING

AND/OR SYSTEMS ARE MAINTAINED THROUGHOUT

CONSTRUCTION.

5. ELECTRICAL CONTRACTOR SHALL FIELD VERIFY EXISTING

PANEL CAPACITY PRIOR TO INSTALLATION OF NEW LOADS.

DO NOT EXCEED 80% OF MAIN CIRCUIT BREAKER

AMPACITY RATING. PROVIDE NEW WIRING, ACCESSORIES,

AND CIRCUIT BREAKERS NECESSARY FOR COMPLETE

INSTALLATION.

6. ELECTRICAL CONTRACTOR SHALL PROVIDE UPDATED

PANEL DIRECTORIES FOR ALL PANELS AFFECTED DURING

CONSTRUCTION.

7. ELECTRICAL CONTRACTOR SHALL CIRCUIT ELECTRICAL

DEVICES TO SPARE CIRCUIT BREAKERS & CIRCUITS THAT

BECOME AVAILABLE DURING DEMOLITION.

8. FIRE ALARM RELATED WORK SHALL BE PERFORMED BY A

LICENSED FIRE ALARM CONTRACTOR.

GENERAL NOTES

KEY NOTES

1. INSTALL NEW 3Φ, 208V, 30A, NON-FUSED ELECTRICAL

DISCONNECT WITH NEMA 3R RATED ENCLOSURE FOR

NEW HVAC EQUIPMENT SPECIFIED ON MECHANICAL

DRAWINGS.

DOCUMENTS ARE PROPERTY OF THE U.S. GOVERNMENT.

Phone: (804) 726-9540 https//raymond.global

1601 Summit Avenue Richmond, VA 23230

DATE DESCRIPTION

PROJECT NO.: RVA1033.003

OWNER:

WG HEFNER VAMC

BUILDING 5

PHARMACY CLEAN ROOM

1601 BRENNER AVE

SALISBURY, NC, 28144

SENSITIVE UNCLASSIFIED

DESIGNED BY:

DRAWN BY:

CHECKED BY:

ISSUES / REVISIONS

09/15/20 15% SD SUBMISSION

10/29/20 35% SD SUBMISSION

DATE VA COR SIGN OFF

09/15/20 ANDRE L. NOBS

SUB

15% SD

11/18/20 ANDRE L. NOBS35% SD

12/19/20 65% DD SUBMISSION

02/25/21 95% DD SUBMISSION

2/2/21 ERIC STEWART65% DD

3/12/21 ERIC STEWART95% DD

03/26/21 100% DD SUBMISSION

05/09/21 FINAL BID DOCUMENTS

4/16/21 ANDRE HAYLES100% DD

09/29/21 DILANDRO COMMENTS1

ADRIAN LOZA

ADRIAN LOZA

FLOYD KEELS

E-103SCALE: 1/4" = 1'-0"4 ELECTRICAL ROOF PLAN

NEW ELECTRICAL ROOF PLAN

PROJECT NORTH

EXIT SIGN REQUIREMENTS:

1. HOUSING - DIE-CAST ALUMINUM, STEEL, OR HIGH-IMPACT, UV-STABILIZED,

INJECTION- MOLDED THERMOPLASTIC. SINGLE OR DOUBLE-FACED AS

INDICATED.

2. FINISH (ON CAST ALUMINUM HOUSING ONLY) - TEXTURED POWDER COAT

FINISH OPTIONS INCLUDE WHITE, WHITE WITH BRUSHED ALUMINUM FACE,

BLACK, OR BLACK WITH BRUSHED ALUMINUM FACE.

3. LETTERS/CHEVRONS - MINIMUM 6" HIGH WITH 3/4" STROKE. RED OR GREEN

LETTERS AS INDICATED. PROVIDE CHEVRONS AS INDICATED EITHER LEFT,

RIGHT OR BOTH DIRECTIONS AS INDICATED. CHEVRONS PUNCHED OUT

THROUGH HOUSING AS REQUIRED.

4. MOUNTING - UNIVERSAL MOUNTING KIT FOR CEILING, WALL OR

END-OF-FIXTURE MOUNTING.

5. ILLUMINATION - PROVIDED BY RED, GREEN OR WHITE HIGH-OUTPUT LEDS

INSIDE OF FIXTURE HOUSING. PROVIDE POLYSTYRENE DIFFUSER IN COLOR

INDICATED WITH FREQUENCY-MATCHED SILKSCREEN COATING FOR MAXIMUM

LED LIGHT OUTPUT.

6. CERTIFICATION - UL LISTED AND CERTIFIED FOR DAMP LOCATIONS.

7. FIXTURE TYPE: C

NOTE: SEE SPECIFICATION 26 51 00 - INTERIOR LIGHTING

FOR ADDITIONAL LUMINAIRE REQUIREMENTS.

WHERE DRAWINGS AND SPECIFICATIONS CONFLICT, THE MOST STRINGENT

REQUIREMENT SHALL GOVERN.

LUMINAIRE REQUIREMENTS:

1. HOUSING & FINISH - 2’ X 2’ TROFFER BODY TO HAVE SMOOTH HEMMED SIDES

AND SMOOTH INWARD FORMED END FLANGES.

2. REFLECTORS - HIGHLY TRANSMISSIVE ACRYLIC PATTERN #12 (.125") LENS

DIFFUSE TO ELIMINATE LED IMAGE FROM BELOW.

3. LAMPS - LED LAMP TO HAVE 3500K KELVIN COLOR TEMPERATURE OF LIGHT

WITH A CRI OF 80 AND AN L70 OF 50,000 HOURS OR GREATER.

4. DRIVER - LED DRIVER TO BE AUTOSENSING 120-277 VOLT AND MUST BE

ACCESSIBLE FROM BELOW THE CEILING. FIXTURE IS TO BE DIMMABLE DOWN

TO 1 PERCENT USING 0 TO 10 VOLT DRIVER BEING FLICKER FREE DURING

DIMMING PROCESS. PROVIDE DIGITALLY ADDRESSABLE DRIVER OR LOW

VOLTAGE DIGITAL ADDRESSABLE CONTROL DEVICE.

5. CERTIFICATION - FIXTURE MUST BE DLC QUALIFIED (DESIGN LIGHT

CONSORTIUM) AND UL LISTED. FIXTURE PERFORMANCE SPECIFICATIONS

MUST MEET OR EXCEED LINE 2 AS TESTED BY DLC.

6. WARRANTY - PROVIDE A MINIMUM 5-YEAR WARRANTY BY MANUFACTURER.

PROVIDE IES FILE AND PROVIDE SPECIFICATION SHEET FOR REVIEW.

7. FIXTURE TYPE: A

NOTE: SEE SPECIFICATION 26 51 00 - INTERIOR LIGHTING

FOR ADDITIONAL LUMINAIRE REQUIREMENTS.

WHERE DRAWINGS AND SPECIFICATIONS CONFLICT, THE MOST STRINGENT

REQUIREMENT SHALL GOVERN.

13" NOMINAL

8" NOMINAL

PNL-CKT

NOTE: REFER TO SPECIFICATION 26 27 26 - WIRING DEVICES

FOR ADDITIONAL LIGHT SWITCH REQUIREMENTS

INSTALL APPROVED

MACHINE SCREWS.

BOND LIGHT SWITCH TO

BOX WITH GREEN

GROUNDING SCREW.

STAINLESS STEEL

FACEPLATE

PROVIDE DURABLE

WIRE MARKERS OR

TAGS INSIDE THE

OUTLET BOX.

WHITE LABEL TAPE WITH

TYPED, BLACK LETTERING

DESIGNATING PANEL AND

CIRCUIT NUMBER.

PROVIDE LIGHT SWITCH IN

WHITE COLOR.

SEAL FACEPLATE AND LIGHT

SWITCH FOR AIR TIGHTNESS

WITH SEALANT.

PNL-CKT

NOTE: REFER TO SPECIFICATION 26 27 26 - WIRING DEVICES

FOR ADDITIONAL RECEPTACLE REQUIREMENTS

INSTALL GROUND

PIN ON TOP.

INSTALL APPROVED

MACHINE SCREWS.

BOND RECEPTACLE TO

BOX WITH GREEN

GROUNDING SCREW

STAINLESS STEEL

FACEPLATE

PROVIDE DURABLE

WIRE MARKERS OR

TAGS INSIDE THE

OUTLET BOX.

WHITE LABEL TAPE WITH

TYPED, BLACK LETTERING

DESIGNATING PANEL AND

CIRCUIT NUMBER.

PROVIDE RECEPTACLES IN

THE FOLLOWING COLORS:

• WHITE: NORMAL POWER

• RED: EMERGENCY POWER

SEAL FACEPLATE AND

RECEPTACLE FOR AIR

TIGHTNESS WITH

SEALANT.

SWITCH-PORT

NOTE: REFER TO SPECIFICATION 27 15 00 - COMMUNICATIONS STRUCTURED CABLING

FOR ADDITIONAL DATA/VOICE OUTLET REQUIREMENTS

INSTALL APPROVED

MACHINE SCREWS.

STAINLESS STEEL

FACEPLATE

PROVIDE 3 4" CONDUIT

TO CEILING SPACE.

COORDINATE CABLING

WITH COR..

WHITE LABEL TAPE WITH

TYPED, BLACK LETTERING

DESIGNATING WORKSTATION

AND PATCH PANEL #.

COORINATE LABELING WITH

COR.

PROVIDE TWO RJ-45 CAT5E

DATA OUTLETS AND TWO

RJ-11 VOICE TELEPHONE

COMMUNICATION OUTLETS.

SEAL FACEPLATE AND

OUTLET FOR AIR

TIGHTNESS WITH

SEALANT.

SEAL ALL UNUSED KNOCKOUTS AND

OPENINGS FOR AIR TIGHTNESS.

PROVIDE STAINLESS STEEL

CLEANROOM GRADE JUNCTION BOX.

NOT TO EXCEED BOX FILL

CALCULATIONS PER NEC 314.16(B).

INSTALL APPROVED MACHINE

SCREWS AND GREEN GROUNDING

SCREWS AS REQUIRED. BOND ALL

METAL BOXES AND ELECTRICAL

DEVICES.

NOTE: REFER TO SPECIFICATION 26 05 33 - RACEWAY AND BOXES FOR ELECTRICAL SYSTEMS

FOR ADDITIONAL BOX REQUIREMENTS

TV

TV TV

�6"

MAX.

�6"

MAX.

AV

�12"

MAX.

FI

R

E

FI

R

E

�12"

MAX.

T-STAT

T $

�6"

MAX.

�12"

MAX.

�18"

MAX.

P

�5'-0" MAX.

(EITHER SIDE OF DOOR)

" M

EA

SU

R

ED

T O

H

AN

D

LE

" M

IN

" A

FF

M

IN

BO

TT

O

M O

F

LE

N S

" A

FF

M

AX

TO

P

O F

LE

N

S

FIRE ALARM

PULL STATION

LIGHT

SWITCH

VOICE

OUTLET

PROJECTOR

MOMENTARY

TOGGLE

SWITCH

DATA

OUTLET

DUPLEX

RECEPTACLE

DOUBLE DUPLEX

RECEPTACLE

FIRE

ALARM

M

AR

KE

R

BO

AR

D 6"

MAX.

6"

MAX.

6"

MAX.

6"

MAX.

DOCUMENTS ARE PROPERTY OF THE U.S. GOVERNMENT.

Phone: (804) 726-9540 https//raymond.global

1601 Summit Avenue Richmond, VA 23230

DATE DESCRIPTION

PROJECT NO.: RVA1033.003

OWNER:

WG HEFNER VAMC

BUILDING 5

PHARMACY CLEAN ROOM

1601 BRENNER AVE

SALISBURY, NC, 28144

SENSITIVE UNCLASSIFIED

DESIGNED BY:

DRAWN BY:

CHECKED BY:

ISSUES / REVISIONS

09/15/20 15% SD SUBMISSION

10/29/20 35% SD SUBMISSION

DATE VA COR SIGN OFF

09/15/20 ANDRE L. NOBS

SUB

15% SD

11/18/20 ANDRE L. NOBS35% SD

12/19/20 65% DD SUBMISSION

02/25/21 95% DD SUBMISSION

2/2/21 ERIC STEWART65% DD

3/12/21 ERIC STEWART95% DD

03/26/21 100% DD SUBMISSION

05/09/21 FINAL BID DOCUMENTS

4/16/21 ANDRE HAYLES100% DD

09/29/21 DILANDRO COMMENTS1

ADRIAN LOZA

ADRIAN LOZA

FLOYD KEELS

E-501

ELECTRICAL DETAILS

NO SCALE1

RECEPTACLE DETAIL (TYP.)

NO SCALE2

LIGHT SWITCH DETAIL (TYP.)

NO SCALE5

2' X 2' LED LAY-IN TROFFER (TYP.)

NO SCALE6

LED EXIT LIGHT FIXTURE (TYP.)

NO SCALE3

COMM. OUTLET DETAIL (TYP.)

NO SCALE4

JUNCTION BOX DETAIL (TYP.)

NOT TO SCALE7 ELECTRICAL MOUNTING HEIGHT DETAIL

DOCUMENTS ARE PROPERTY OF THE U.S. GOVERNMENT.

Phone: (804) 726-9540 https//raymond.global

1601 Summit Avenue Richmond, VA 23230

DATE DESCRIPTION

PROJECT NO.: RVA1033.003

OWNER:

WG HEFNER VAMC

BUILDING 5

PHARMACY CLEAN ROOM

1601 BRENNER AVE

SALISBURY, NC, 28144

SENSITIVE UNCLASSIFIED

DESIGNED BY:

DRAWN BY:

CHECKED BY:

ISSUES / REVISIONS

09/15/20 15% SD SUBMISSION

10/29/20 35% SD SUBMISSION

DATE VA COR SIGN OFF

09/15/20 ANDRE L. NOBS

SUB

15% SD

11/18/20 ANDRE L. NOBS35% SD

12/19/20 65% DD SUBMISSION

02/25/21 95% DD SUBMISSION

2/2/21 ERIC STEWART65% DD

3/12/21 ERIC STEWART95% DD

03/26/21 100% DD SUBMISSION

05/09/21 FINAL BID DOCUMENTS

4/16/21 ANDRE HAYLES100% DD

09/29/21 DILANDRO COMMENTS1

ADRIAN LOZA

ADRIAN LOZA

FLOYD KEELS

E-601

ELECTRICAL SCHEDULES

1. ELECTRICAL CONTRACTOR SHALL FIELD VERIFY EXISTING

PANEL CAPACITY PRIOR TO INSTALLATION OF NEW LOADS.

DO NOT EXCEED 80% OF MAIN CIRCUIT BREAKER

AMPACITY RATING. PROVIDE NEW WIRING, ACCESSORIES,

AND CIRCUIT BREAKERS NECESSARY FOR COMPLETE

INSTALLATION. PROVIDE NEW PANELBOARD AND CIRCUIT

BREAKERS SIZED AS NEEDED FOR THE COMPLETE

INSTALLATION OF NEW LOADS AND REFEEDING OF

EXISTING LOADS.

2. ELECTRICAL CONTRACTOR SHALL PROVIDE UPDATED

PANEL DIRECTORIES, TYPED AND PRINTED, FOR ALL

PANELS AFFECTED DURING CONSTRUCTION.

3. ELECTRICAL CONTRACTOR SHALL CIRCUIT ELECTRICAL

DEVICES TO SPARE CIRCUIT BREAKERS & CIRCUITS THAT

BECOME AVAILABLE DURING DEMOLITION.

GENERAL NOTES

LUMINAIRE SCHEDULE

TYPE MANUFACTURER DESCRIPTION LAMP TYPE VOLTAGE INPUT WATTS LOCATION

A KENALL

OR EQUAL

2' X 2' RECESSED MOUNTED LED SIMPLESEAL FIXTURE

LED 120 V 72 W STAGING AREA

CSEDO-22-67L-35K8-DIM1-DV-2F-2H-SYM-AMF

B

QAS

OR EQUIAL

2' X 2' RECESSED MOUNTED LED FIXTURE

LED 120 V 60 W CLEANROOM

PROVIDED BY CLEANROOM MANUFACTURER

C COOPER

LIGHTING OR EQUAL

SURE-LITES WALL MOUNTED LED EXIT SIGN

LED 120 V 1 W CLEANROOM

SLX7

EXISTING PANELBOARD SCHEDULE 5D

LOCATION: BUILDING 5 MAINS RATING: 225A

SUPPLY FROM: 5X VOLTAGE 208/120 Y

MOUNTING: SURFACE PHASES: 3

ENCLOSURE: NEMA 1 WIRES: 4

CONNECTED LOAD (VA)

CKT CIRCUIT DESCRIPTION: TRIP POLE A B C POLE TRIP CIRCUIT DESCRIPTION CKT

1 EXISTING LOAD - - - - - - EXISTING LOAD 2

3 EXISTING LOAD - - - - - - EXISTING LOAD 4

5 EXISTING LOAD - - - - - - EXISTING LOAD 6

7 EXISTING LOAD - - - - - - EXISTING LOAD 8

9 EXISTING LOAD - - - - - - EXISTING LOAD 10

11 EXISTING LOAD - - - - - - EXISTING LOAD 12

13 EXISTING LOAD - - - - - - EXISTING LOAD 14

15 EXISTING LOAD - - - - - - EXISTING LOAD 16

17 EXISTING LOAD - - - - - - EXISTING LOAD 18

19 EXISTING LOAD - - - - - - EXISTING LOAD 20

21 EXISTING LOAD - - - - - - EXISTING LOAD 22

23 EXISTING LOAD - - - - - - EXISTING LOAD 24

25 EXISTING LOAD - - - - - - EXISTING LOAD 26

27 EXISTING LOAD - - - - - - EXISTING LOAD 28

29 EXISTING LOAD - - - - - - EXISTING LOAD 30

AHU-49 15 A 3

888 1000 1 20 A HAZ. STORAGE FRIDGE 32

33 888 1000 1 20 A PHARM. STAGING FRIDGE 34

35 888 1000 1 20 A CLEAN ROOM DOORS 36

37 CLEAN ROOM LTG. 20 A 1 1500 1680 1 20 A HAZ. COMP. ROOM FFU 38

39 BIOSAFE HOOD 20 A 1 1500 1344 1 20 A COMP. ROOM FFU 40

41 LAMINAR WORKSTATION 20 A 1 1500 1680 1 20 A ANTEROOM, STORAGE, HAZ. COMP ROOM FFU 42

43 EXISTING LOAD - - - 1272

3 30 A BMEF-1

45 ANTEROOM SINK 20 A 1 195…