558-CSI-400 Replace MRI 1.5T Drawings.pdf
PDF 3 MB Posted
- Attached to
- Z2DA--Replace MRI 1. 5T Federal contract opportunity
- Solicitation number
- 36C24626B0001
About this file
These are 100% construction documents for a VA Medical Center project to replace a 1.5T MRI machine at the Durham VA Health Care System located at 508 Fulton Street in Durham, NC. The project scope involves demolishing existing exterior walls, an RF shielding enclosure, and wood stud walls, and installing a new Siemens Magnetom Sola MRI machine with associated equipment and supporting infrastructure. The total area of work is 763 square feet, and the project will be fully sprinklered with construction occurring under the 2021 International Building Code and 2021 NFPA 101 Life Safety Code. The project includes comprehensive architectural, mechanical, electrical, plumbing, and fire protection plans for the MRI equipment room, exam room, control room, and surrounding spaces, with detailed specifications for RF shielding, power requirements, cable routing, and environmental conditions specific to MRI system installation.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C24626B0001 0006.docx | DOCX document | |
| Bid Abstract 36C24626B0001.pdf | ||
| Base Bid Cost Breakdown.xlsx | XLSX spreadsheet | |
| 36C24626B0001 0005.docx | DOCX document | |
| 08 34 46 RADIO-FREQUENCY-INTERFERENCE SHIELDING DOORS_r1.pdf | ||
| 36C24626B0001 0004.docx | DOCX document | |
| 36C24626B0001 0003.docx | DOCX document | |
| 36C24626B0001 0002.docx | DOCX document | |
| A101 - ARCHITECTURAL PLANS DETAILS AND SECTIONS.pdf | ||
| Pre-Bid Conference-Site Vist_Replace MRI 1.5T_Attendees.pdf | ||
| 08 34 46 RADIO-FREQUENCY-INTERFERENCE SHIELDING DOORS.pdf | ||
| 36C24626B0001 0001.docx | DOCX document | |
| Addendum 1 Narrative.pdf | ||
| 558-CSI-400 Replace MRI 1.5T Specs.pdf | ||
| 558-CSI-400 Replace MRI 1.5T Specs.docx | DOCX document | |
| 36C24626B0001_2.docx | DOCX document |
Show all 16
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
Watson Engineering, P.C.
782 Johnnie Dodds Blvd, Suite I-2 Mount Pleasant, SC 29464 (607)223-4334
LOCATION MAP DRAWING LIST
Approved by Medical Center Director
Approved by Project COR
Approved by Infection Control
Approved by Director of Facility Management Service
Approved by Medical Director
Approved by Safety Officer
Approved by Project CO
APPLICABLE CODES PROJECT CONTACT DIRECTORY
CONTRACTING OFFICER (CO)
VANGIE MILLER
(757) 315-3999 Vangie.Miller@VA.gov
CONTRACTING OFFICER REPRESENTATIVE (COR)
BRIAN RICHARDSON
(919) 286-0411 ext. 176421 Brian.Richardson4@va.gov
ARCHITECT
RDC/JPA ARCHITECTURE
524 Fernwood Drive Altamonte Springs, FL 32701
(800) 362-1523 www.rdcjpa.com
MEP/FP
WATSON ENGINEERING, PC
782 Johnnie Dodds Blvd, Suite I-2 Mount Pleasant, SC 29464
(607) 223-4334 www.watson-engineer.com
Au to de sk
D oc s:
//5
-C
SI
-4
R ep la ce M
R I 1
.5 T in B ui ld in g
1/ D ur ha m M
R I_
R
.rv t
/2
0/
:1 0:
P
M
T-100
TITLE SHEET
Replace MRI 1.5T 508 Fulton Street
Durham, NC 27705
100% CONSTRUCTION DOCUMENTS
Replace MRI 1.5T
508 Fulton Street Durham, NC 27705
12/20/2024 MPL JWC
558-CSI-400
100% CONSTRUCTION
DOCUMENTS
T-100 TITLE SHEET
G-100 GENERAL NOTES &
SYMBOLS
A001 CODE PLAN
A101 ARCHITECTURAL PLANS,
DETAILS, AND SECTIONS
A102 FINISH PLANS,
ELEVATIONS, AND
SCHEDULES
MD101 MECHANICAL
DEMOLITION PLAN
MD102 MECHANICAL
REFLECTED CEILING
DEMOLITION
MD103 MECHANICAL ROOF
DEMOLITION
M-101 MECHANICAL
REFLECTED CEILING
PLAN
M-102 MECHANICAL PLAN
M-501 MECHANICAL DETAILS
M-701 HVAC CONTROLS
DETAILS
524 Fernwood Drive, Altamonte Springs, Florida 32701 |www.rdcjpa.com| 800.362.1523
PROJECT LOCATION
Revision Number Revisions: Date:
ED101 ELECTRICAL POWER
DEMOLITION
ED102 ELECTRICAL LIGHTING
DEMOLITION
E-101 ELECTRICAL POWER
PLAN
E-102 ELECTRICAL LIGHTING
PLAN
E-501 ELECTRICAL DETAILS
E-502 ELECTRICAL DETAILS
E-601 ELECTRICAL
SCHEDULES
E-701 ELECTRICAL SINGLE
LINE
EC101 ELECTRICAL RACEWAY
PLAN
EC801 ELECTRICAL RACEWAY
ISOMETRIC
PD101 PLUMBING DEMOLITION
P-101 PLUMBING PLAN
FP-101 FIRE PROTECTION PLAN
GAS METER GM
WATER METER WM
BACKFLOW PREVENTER BFP
ELBOW UP
ELBOW DOWN
TEE UP
TEE DOWN
CONCENTRIC REDUCER
ECCENTRICAL REDUCER
CAP
UNION
WYE STRAINER
FLANGED CONNECTION
FLOW DIRECTION
PIPE ANCHOR
PIPE SLEEVE
PRESSURE SENSOR
SHUT OFF VALVE
CHECK VALVE
CONTROL VALVE
THREE-WAY CONTROL VALVE
TEMPERATURE/PRESSURE RELIEF
VALVE
FLOW CONTROL VALVE FCV
PRESSURE REDUCING VALVE
MAKEUP WATER VALVE
THROTTLING VALVE
HUMIDISTAT H
THERMOSTAT T
IN-LINE TEMPERATURE SENSOR
T
RETURN DUCT TURNING TOWARD
VIEWER
EXHAUST DUCT TURNING TOWARD
VIEWER
SUPPLY DUCT TURNING AWAY FROM
VIEWER
RETURN DUCT TURNING AWAY FROM
VIEWER
EXHAUST DUCT TURNING AWAY FROM
VIEWER
DUCT SIZE, SEE SPECIFICATIONS FOR
INSULATION TYPE AND THICKNESS
CHANGE IN DUCT ELEVATION - RISE
INDICATED BY "R"
R
CHANGE IN DUCT ELEVATION - DROP
INDICATED BY "D"
D
SUPPLY DUCT TURNING TOWARD
VIEWER
DUCT SECTION, RETURN
DUCT SECTION, EXHAUST
DUCT SECTION, SUPPLY
BUTTERFLY VALVE
BALL VALVE
PUMP
CHILLER
COOLING TOWER
HEAT EXCHANGER, SHELL & TUBE
HEAT EXCHANGER, PLATE
COOLING COIL
HEATING COIL
ELECTRIC COIL
UNIT HEATER
CENTRIFUGAL FANCHILLED WATER SUPPLY CHWS
DRAIN D
DUAL TEMP SUPPLY H/C S
CHILLED WATER RETURN CHWR
DUAL TEMP RETURN H/C R
GLYCOL RETURN GR
CONDENSER WATER SUPPLY CWS
GLYCOL SUPPLY GS
CONDENSER WATER RETURN CWR
COOLING TOWER WATER RETURN
HEAT PUMP WATER SUPPLY HPWS
COOLING TOWER WATER SUPPLY CTWS
HEAT PUMP WATER RETURN HPWR
CTWR
DOMESTIC COLD WATER DCW
DOMESTIC HOT WATER RETURN DHWR
SANITARY WASTE SAN
DOMESTIC HOT WATER DHW
HIGH PRESSURE STEAM HPS
LOW PRESSURE STEAM LPS
VACUUM SERVICE VAC
MEDIUM PRESSURE STEAM MPS
REFRIGERANT SUCTION RS
GAS
HOT WATER SUPPLY (HEATING) HWS
REFRIGERANT DISCHARGE (HOT GAS) RD
HOT WATER RETURN (HEATING) HWR
G
MOTORIZED DAMPER M
VARIABLE FREQUENCY DRIVE VFD
MANUAL VOLUME DAMPER
RETURN DUCT TURNING TOWARD
VIEWER
EXHAUST DUCT TURNING TOWARD
VIEWER
SUPPLY DUCT TURNING AWAY FROM
VIEWER
RETURN DUCT TURNING AWAY FROM
VIEWER
EXHAUST DUCT TURNING AWAY FROM
VIEWER
DUCT SIZE, SEE SPECIFICATIONS FOR
INSULATION TYPE AND THICKNESS
24 x 12
CHANGE IN DUCT ELEVATION - RISE
INDICATED BY "R"
R
CHANGE IN DUCT ELEVATION - DROP
INDICATED BY "D"
D
SUPPLY DUCT TURNING TOWARD
VIEWER
DUCT SECTION, RETURN
DUCT SECTION, EXHAUST
DUCT SECTION, SUPPLY
BUTTERFLY VALVE
BALL VALVE
PUMP
DRAWING SYMBOLS
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
Watson Engineering, P.C.
782 Johnnie Dodds Blvd, Suite I-2 Mount Pleasant, SC 29464 (607)223-4334
Au to de sk
D oc s:
//5
-C
SI
-4
R ep la ce M
R I 1
.5 T in B ui ld in g
1/ D ur ha m M
R I_
R
.rv t
/2
0/
:1 0:
P
M
G-100
GENERAL NOTES & SYMBOLS Replace MRI 1.5T
508 Fulton Street Durham, NC 27705
12/20/2024 MPL JWC
558-CSI-400
100% CONSTRUCTION
DOCUMENTS
1. UNLESS SPECIFICALLY NOTED AS EXISTING, ALL ITEMS, EQUIPMENT, AND WORK SHOW IS NEW AND SHALL BE FURNISHED AND INSTALLED BY
THE CONTRACTOR.
2. DURING THE EXECUTION OF THIS CONTRACT, THE SITE AND BUILDINGS WILL BE OCCUPIED AND IN USE. WORK DONE UNDER THIS CONTRACT SHALL BE PERFORMED WITHOUT INCONVENIENCE TO THE OWNER INSOFAR AS POSSIBLE. ALL WORK SHALL BE COORDINATED WITH THE
CONTRACTING OFFICER’S REPRESENTATIVE AND IS SUBJECT TO HIS CONTINUOUS APPROVAL.
3. THE CONTRACTOR IS COMPLETELY RESPONSIBLE FOR ALL CHARACTERISTICS OF THE EQUIPMENT THAT HE INTENDS TO INSTALL. IF THERE ARE ANY DIFFERENCES IN THE EQUIPMENT SHOWN ON THE DESIGN DRAWINGS AND THE EQUIPMENT TO BE INSTALLED, THE CONTRACTOR SHALL MAKE ALL CHANGES AND ADJUSTMENTS TO ACCOMMODATE THE PROPOSED EQUIPMENT AT NO ADDITIONAL COST TO THE OWNER. THIS APPLIES TO ALL SUPPORT SERVICES AND FACILITIES NECESSARY TO PROPERLY INSTALL THE EQUIPMENT TO FUNCTION ADEQUATELY AS PART OF THE SYSTEMS FOR WHICH THEY WERE DESIGNED AND SELECTED. THESE CHARACTERISTICS INCLUDE SUCH THINGS AS WEIGHT, DIMENSIONS, ELECTRICAL, MECHANICAL, PLUMBING, AND UTILITY SERVICES, ETC. ACCOMMODATING SERVICES INCLUDE SUCH THINGS AS DIMENSIONS, LOCATIONS, AND SIZES OF SUPPORTING MEMBERS, PIPE, CONNECTION SIZES, ELECTRICAL SERVICE SIZE AND PROTECTION, ELECTRICAL CONNECTION LOCATION, UTILITY AND PLUMBING CONNECTION SIZES AND LOCATIONS, ETC. ALL REQUIRED CHANGES TO THE CONTRACT DRAWINGS AND SPECIFICATIONS SHALL BE DESIGNED AND STAMPED BY REGISTERED ENGINEERS IN THE APPROPRIATE DISCIPLINE EMPLOYED BY THE CONTRACTOR AND SHALL BE INCLUDED AS PART OF THE EQUIPMENT SUBMITTALS. ALL CHANGES SHALL BE SUBJECT TO
THE APPROVAL OF THE PROJECT A/E AND THE CONTRACTING OFFICER’S REPRESENTATIVE.
4. PRELIMINARY AIR BALANCE REPORT. AT THE BEGINNING OF THE WORK IMMEDIATELY AFTER RECEIVING NOTICE TO PROCEED, THE CONTRACTOR SHALL MEASURE AND SUBMIT A PRELIMINARY BALANCE REPORT IN COMPLIANCE WITH THE SPECIFICATIONS FOR THE TERMINAL UNITS IN THE AREA OF WORK OF THIS PROJECT. MEASURE ALL SUPPLY, RETURN, AND EXHAUST AIR QUANTITIES. MEASURE AHU IN FULL COOLING, MAX AIR FLOW, AT THE AHU, AND MAX AIR FLOW AT ALL ZONES AND ALL DIFFUSERS; MEASURE THE DISCHARGE STATIC PRESSURE AT THE AHU AND THE SUCTION STATIC PRESSURE AT THE AHU. THEN EACH ZONE SHALL BE SET TO FULL REHEAT AND THE AIR QUANTITY MEASURED AT EACH OUTLET AND THE WATER FLOWS AND TEMPERATURES MEASURED AT EACH REHEAT BOX. THE EXISTING MECHANICAL DRAWINGS WILL BE PROVIDED BY THE FACILITY ENGINEERING DEPARTMENT AT THE VA. SUBMIT A BLANK AIR BALANCE REPORT FOR APPROVAL
PRIOR TO BEGINNING THE PRELIMINARY AIR BALANCE.
5. BEFORE ANY DEMOLITION OR CONSTRUCTION WORK IS BEGUN, THE CONTRACTOR AND THE CONTRACTING OFFICER’S REPRESENTATIVE SHALL SURVEY THE EXISTING CEILINGS TO BE REMOVED AND REPLACED FOR THEIR EXISTING CONDITION. THIS SURVEY WILL BE USED BY THE CONTRACTING OFFICER’S REPRESENTATIVE AT JOB COMPLETION TO DETERMINE WHICH RE-INSTALLED CEILING COMPONENTS MUST BE
REPLACED WITH NEW.
6. REQUEST APPROVAL FROM THE CONTRACTING OFFICER’S REPRESENTATIVE SEVEN DAYS IN ADVANCE FOR THE SHUTDOWN OF ANY EXISTING
SYSTEM.
7. FINAL ADVANCE NOTICE OF 48 HOURS SHALL BE GIVEN TO THE CONTRACTING OFFICER’S REPRESENTATIVE PRIOR TO THE PROPOSED
SHUTDOWN OF ANY SYSTEM.
8. THE CLOSING AND OPENING OF VALVES SHALL BE WITNESSED BY THE CONTRACTING OFFICER’S REPRESENTATIVE; NOTIFY LOCAL
AUTHORITIES AS REQUIRED.
9. PROVIDE ADEQUATE FIRE PROTECTION IN THE ENTIRE CONSTRUCTION AREA FOR THE DURATION OF THE CONSTRUCTION PERIOD. REFER TO
SPECIFICATION DIVISION 1, “GENERAL REQUIREMENTS”, “FIRE SAFETY REQUIREMENTS”.
10. THE ENTIRE EXISTING FIRE PROTECTION SYSTEM SHALL BE RESTORED TO FULL OPERATION AT THE END OF EACH WORKING DAY. ALL EXISTING FIRE ALARMS AND FIRE EXITS AND ANY OTHER ASPECT OF THE EXISTING FIRE ALARM SYSTEM SHALL REMAIN IN FULL OPERATION. THE COST OF RESPONSE TO ANY FALSE ALARM AS CHARGED BY ANY LOCAL FIRE DEPARTMENT SHALL BE PAID BY THE CONTRACTOR.
11. THE GENERAL NOTES ON THIS SHEET APPLY TO ALL THE DRAWING DISCIPLINES.
12. THE DRAWINGS SHALL BE FOLLOWED AS CLOSELY AS POSSIBLE, ALTHOUGH NO ATTEMPT WAS MADE TO SHOW EVERY OFFSET, FITTING, BEAM, PIPE SUPPORT, ETC. CHANGES REQUIRED TO FIT EQUIPMENT, PIPING, AND DUCTWORK SHALL BE PROVIDED AT NO ADDITIONAL COST TO THE
OWNER. ANY MAJOR DISCREPANCIES SHALL BE CLARIFIED PRIOR TO SUBMITTAL OF BID.
13. THE WORK INCLUDES ALL ITEMS OF THE SPECIFICATIONS.
14. VISIT THE SITE AND BE FULLY COGNIZANT OF ALL CONDITIONS PRIOR TO SUBMITTING BID. REPORT ANY DISCREPANCIES TO THE ENGINEER
PRIOR TO SUBMITTAL OF BID.
15. DO NOT BEGIN FABRICATION, CUT ANY MEMBERS, OR BEGIN INSTALLATION UNTIL THE SUBMITTALS AND SHOP DRAWINGS HAVE BEEN
APPROVED. ANY WORK CONSTRUCTED OR INSTALLED PRIOR TO THE APPROVAL OF SUBMITTALS AND SHOP DRAWINGS SHALL BE AT THE CONTRACTOR’S RISK AND SHALL BE REMOVED AND REINSTALLED OR RECONSTRUCTED AT THE CONTRACTOR’S EXPENSE AT THE REQUEST OF THE CONTRACTING OFFICER’S REPRESENTATIVE TO THE COMPLETE SATISFACTION OF THE CONTRACTING OFFICER’S REPRESENTATIVE.
16. DO NOT PENETRATE OR CUT ANY STRUCTURAL MEMBERS NOT SPECIFICALLY SHOWN ON THE DRAWINGS (E.G., JOISTS, BEAMS, ETC.). IF STRUCTURAL MEMBERS CONFLICT WITH ROUTING, CONTACT THE CONTRACTING OFFICER’S REPRESENTATIVE FOR DIRECTION BEFORE PROCEEDING. EXERCISE CAUTION IN CUTTING CONCRETE FLOORS AND WALLS TO AVOID EXISTING UTILITIES WHICH MAY BE LOCATED IN AND
UNDER CONCRETE SURFACES.
17. THE CONTRACTOR SHALL PROVIDE NEW OPENINGS IN WALLS, FLOOR, AND ROOF, AND PATCH EXISTING OPENINGS IN CONJUNCTION WITH THE DEMOLITION AND INSTALLATION OF THE NEW BUILDING COMPONENTS, INCLUDING MECHANICAL, PLUMBING, ELECTRICAL, AND GENERAL CONTRACTOR’S ITEMS. ALL NECESSARY FRAMING, COUNTER-FLASHING, ETC. SHALL BE INCLUDED.
18. FOR ALL FLOOR, WALL, AND ROOF PENETRATION, VERIFY THE LOCATION OF THE (E) STRUCTURAL MEMBERS PRIOR TO DRILLING AND CUTTING, RELOCATE OPENINGS AS REQUIRED AND OFFSET DUCTS, PIPES, CONDUIT, ETC. AS APPROPRIATE. REVIEW THE STRUCTURAL DRAWINGS IN THE
ENGINEERING OFFICE.
19. WHERE THERE IS A CONFLICT ON THE PLANS, SPECIFICATIONS, NOTES, GENERAL NOTES, REFERENCES, PUBLICATIONS, CODES, ETC. WITH RESPECT TO MATERIALS, PRODUCTS, METHODS, WORKMANSHIP, PROCEDURE, EXECUTION, ETC. THE MOST STRINGENT OR BENEFICIAL TO THE OWNER AS DETERMINED BY THE CONTRACTING OFFICER’S REPRESENTATIVE SHALL BE UTILIZED OR INSTALLED. NO RELIEF OR EXTRA PAYMENT SHALL BE MADE TO THE CONTRACTOR BECAUSE OF THE INTERPRETATION OF THE CONTRACTING OFFICER’S REPRESENTATIVE.
20. ALL WORK SHALL BE DONE WITH COMPLETE COMPLIANCE WITH THE PUBLISHED EDITIONS OF THE APPLICABLE CODES AND STANDARDS WHETHER STATUTORY OR NOT. ALL WORK SHALL BE INSTALLED TO COMPLY WITH THE UNIFORM BUILDING MECHANICAL, PLUMBING, ELECTRICAL, AND FIRE CODES, NATIONAL ELECTRICAL CODE, UNDERWRITER’S LABORATORY, INC., AMERICAN NATIONAL STANDARD INSTITUTE, INC., NATIONAL ELECTRICAL MANUFACTURERS ASSOCIATION, BASIC ELECTRICAL REGULATIONS, AND CODES AND STANDARDS OF THE NATIONAL FIRE PROTECTION ASSOCIATION (NFPA). IF ANY PART OF THE INSTALLATION DOES NOT MEET CODE, NOTIFY THE CONTRACTING OFFICER’S REPRESENTATIVE PRIOR TO BEGINNING THE INSTALLATION. IF ANY PART OF THE PROJECT CANNOT BE INSTALLED TO CONFORM WITH THE PLANS AND/OR SPECIFICATIONS, NOTIFY THE CONTRACTING OFFICER’S REPRESENTATIVE PRIOR TO BEGINNING THE INSTALLATION.
21. ALL EXISTING DUCTWORK, PIPING, CONDUIT, AND MECHANICAL, PLUMBING, AND ELECTRICAL EQUIPMENT NOT TO BE REUSED OR REMAIN IN SERVICE SHALL BE REMOVED FROM THE FACILITY. THE CONTRACTING OFFICER’S REPRESENTATIVE HAS THE OPTION TO RECLAIM ANY DEMOLISHED MATERIALS OR EQUIPMENT FOR THE GOVERNMENT, IN WHICH CASE IT SHALL BE DELIVERED TO THE LOCATION DESIGNATED BY
THE CONTRACTING OFFICER’S REPRESENTATIVE.
22. FOR EACH PHASE AND BEFORE INTERRUPTING EACH (E) UTILITY, DETERMINE, IN ADVANCE, IF ANY DOWNSTREAM SERVICE BEYOND THE CONSTRUCTION AREA OF THE PHASE, WILL BE INTERRUPTED AND PROVIDE A TEMPORARY SERVICE BYPASSING THE CONSTRUCTION AREA.
DETERMINE ALL SUCH TEMPORARY SERVICES PRIOR TO SUBMITTING BID. THERE WILL BE NO FUTURE CONSIDERATION TO THE CONTRACTOR FOR ANY TEMPORARY SERVICES REQUIRED. PREPARE A DETAILED SHOP DRAWING SHOWING ALL OF THE EXISTING UTILITIES IN THEIR
RELOCATED LOCATION AND ALL THE NEW HVAC SYSTEMS.
23. THE FLOOR AREAS BELOW OVERHEAD WORK, WHERE WELDING “HOT WORK” IS TO BE PERFORMED, SHALL BE CLEARED TO A MINIMUM DISTANCE OF 36 FEET BEYOND THE POINT DIRECTLY BELOW THE OVERHEAD “HOT WORK”. ALL FLAMMABLE MATERIALS SHALL BE CLEARED TO A
DISTANCE 50 FEET BEYOND THE POINT DIRECTLY BELOW THE OVERHEAD “HOT WORK”.
24. EXISTING DUCTWORK, DOMESTIC COLD WATER PIPING, DOMESTIC HOT WATER PIPING, CHILLED AND HEATING HOT WATER PIPING, WASTE, VENT, FIRE SPRINKLER LINES, FIRE ALARM EQUIPMENT, ELECTRICAL CONDUIT, TELEPHONE CONDUIT, COMMUNICATION CONDUIT, PNEUMATIC TUBING, ETC., MAY HAVE TO BE RELOCATED TO FACILITATE THE INSTALLATION OF THE NEW WORK. THE CONTRACTOR SHALL DETERMINE ALL EXISTING SERVICES THAT REQUIRE RELOCATION PRIOR TO SUBMITTING HIS BID AND INCLUDE ALL COSTS IN THE BID. THERE WILL BE NO EXTRA
COMPENSATION TO THE CONTRACTOR FOR THIS WORK.
25. THE CONTRACTOR IS HEREBY NOTIFIED THAT SOME WORK MAY BE REQUIRED OUTSIDE OF THE AREAS SHOWN ON THESE DRAWINGS. THIS COULD INCLUDE CONTROL MODIFICATIONS OR CONNECTIONS AT REMOTE LOCATIONS, CONTROL PANELS, CPU’S, ETC.
26. BECAUSE OF THE SMALL SCALE OF THE DRAWINGS, IT IS NOT POSSIBLE TO INDICATE ALL OFFSETS, TRANSITIONS, FITTINGS, AND ACCESSORIES THAT MAY BE REQUIRED. COORDINATE ALL THE WORK OF ALL THE TRADES TO INSURE ALL DUCTS, PIPES, CONDUITS, ETC. DO NOT CONFLICT
AND CLEAR STRUCTURAL AND ARCHITECTURAL MEMBERS.
27. THE CONTRACTOR SHALL PAY FOR ALL DAMAGE, INCLUDING WATER DAMAGE, TO PROPERTY AND EQUIPMENT CAUSED BY HIS WORK
THROUGHOUT THE CONSTRUCTION AND WARRANTEE PERIOD.
28. ALL CEILING DEMOLITION WORK SHALL BE DONE AFTER NORMAL WORKING HOURS. THE INSTALLATION OF THE NEW CEILING SHALL BE DONE AFTER NORMAL WORKING HOURS. THE WORK AREA SHALL BE CLEANED BEFORE THE START OF EACH WORKING DAY.
29. REMOVE T-BARS AND 2X4 CEILING PANELS AS REQUIRED FOR THE DEMOLITION AND THE INSTALLATION OF THE NEW WORK. REINSTALL EXISTING OR REPLACE WITH NEW AS REQUIRED TO THE SATISFACTION OF THE CONTRACTING OFFICER’S REPRESENTATIVE. FOR ANY HARD CEILING AREA, CUT AND DEMOLISH ENOUGH OF THE HARD CEILING TO DO THE WORK REQUIRED. REBUILD AND PATCH THE DEMOLISHED.
FINISH AND PAINT TO MATCH EXISTING TO THE SATISFACTION OF THE CONTRACTING OFFICER’S REPRESENTATIVE.
30. ALL VALVES, VOLUME DAMPERS, CONTROL DEVICES, VAR, ETC. SHALL BE INSTALLED IN AN ACCESSIBLE LOCATION.
31. ALL UL LISTED EQUIPMENT SHALL BE INSTALLED AS PER LISTING OR LABELING. MAXIMUM FUSE SIZES MEANS FUSE PROTECTION REQUIRED.
32. PATCH AND PAINT TO MATCH EXISTING, ALL SURFACES DAMAGED BY THE DEMOLITION AND ALL OPENINGS NOT TO BE REUSED.
33. REPAIR DAMAGE TO CEILING, WALLS, ETC. TO THE SATISFACTION OF THE CONTRACTING OFFICER’S REPRESENTATIVE.
34. WHEN A DETAIL OR NOTE IS IDENTIFIED AS TYPICAL, THE CONTRACTOR IS TO APPLY THE DETAIL OR NOTE TO EVERY SIMILAR CONDITION, WHETHER OR NOT THE REFERENCED DETAIL OR NOTE IS REPEATED OR REFERENCED.
35. WHERE NO CONSTRUCTION DETAIL IS INDICATED, SUCH WORK SHALL BE SIMILAR TO COMPARABLE WORK DETAILED OR NOTED ELSEWHERE OR
TO MATCH EXISTING CONDITIONS, SUBJECT TO THE APPROVAL OF THE CONTRACTING OFFICER’S REPRESENTATIVE.
36. ALL HOLES, VOIDS, AND DAMAGED OR UNEVEN SURFACES WHICH ARE CAUSED BY THE CONTRACTOR’S WORK AND WHICH REMAIN EXPOSED IN
THE WALL, FLOOR, CEILING, OR OTHER SURFACES SHALL BE PATCHED, REPAIRED, AND BROUGHT FLUSH WITH THE ADJACENT CONSTRUCTION.
SUCH SURFACES, NOT CONCEALED BY THE NEW CONSTRUCTION, SHALL MATCH IN EVERY RESPECT THE EXISTING NON-DAMAGED CONSTRUCTION. REMOVE EXISTING UTILITIES BACK BEHIND EXISTING SURFACES AND CAP SO AS NOT TO INTERFERE WITH NEW PATCHES.
37. WHEREVER NEW WALL, FLOOR, OR CEILING FINISHES ARE INDICATED, THE CONTRACTOR SHALL BRING EXISTING ELECTRICAL, MECHANICAL, PLUMBING, AND OTHER BUILDING COMPONENTS FLUSH WITH THE NEW FINISHED SURFACES IF APPROPRIATE.
38. THE DEMOLITION SHALL INCLUDE ALL ASSOCIATED ANCHORAGES, HANGERS, SUPPORTS, BRACES, FASTENERS, MASTIC, MORTAR, LATH, AND RELATED ITEMS. THE EXISTING SUBSTRATE, INCLUDING CONCRETE, MASONRY, METAL FRAMING AND FURRING SHALL BE REPAIRED WHERE
DAMAGE BY THE CONTRACTOR’S WORK.
39. PROVIDE ALL VALVES, INSTRUMENTS, FITTINGS, ETC. AS SHOWN ON THE PIPING DIAGRAMS, SCHEMATICS, ETC. IT IS NOT POSSIBLE TO SHOW ALL VALVES, INSTRUMENTS, ETC. ON THE PLANS, HOWEVER, SELECTED VALVES HAVE BEEN SHOWN FOR REFERENCE. ALL VALVES SHALL BE
ACCESSIBLE FROM THE FLOOR OR FROM FIXED PLATFORMS.
40. ALL NEW FIRE ALARM DEVICES INSTALLED IN HVAC EQUIPMENT SUCH AS SMOKE DETECTORS, SMOKE DAMPERS, FIRE/SMOKE DAMPERS, ETC.
SHALL BE CONNECTED TO THE MAIN FIRE ALARM PANEL.
41. ALL EQUIPMENT AND MATERIALS SUBJECT TO THE WEATHER THAT ARE NOT FACTORY PAINTED SHALL BE FIELD PAINTED AND WEATHER PROOFED WITH PRIMER, AND TWO COATS OF EXTERIOR PAINT. SEE THE SPECIFICATIONS. MATCH THE COLOR OF THE EXISTING BUILDING EXTERIOR. THE COLOR AND QUANTITY OF THE PAINT IS SUBJECT TO THE APPROVAL OF THE CONTRACTING OFFICER’S REPRESENTATIVE.
SUBMITTAL REQUIRED.
42. THE CONTROLS CONTRACTOR SHALL BE FULLY RESPONSIBLE FOR MAKING THE CONTROL SYSTEM COMPLETE AND FULLY OPERABLE. THE CONNECTION TO THE EXISTING CONTROL SYSTEM SHALL BE COMPLETE SUCH THAT ALL NEW MONITORING AND CONTROL POINTS ARE RECOGNIZED BY THE EXISTING CONTROL SYSTEM AND THAT ALL NEW POINTS CAN BE READ, ACTUATED, AND RESET FROM THE EXISTING FRONT END, AND VICE-VERSA. THE ENGINEER SHALL BE NOTIFIED WHEN EACH PHASE OF THE SYSTEM HAS BEEN COMPLETELY CHECKED OUT, ADJUSTED, CALIBRATED, AND PUT INTO FINAL WORKING CONDITION WITH ALL THE TROUBLESHOOTING COMPLETED. A SUBSEQUENT JOBSITE MEETING WILL BE SCHEDULED WITH TWO WEEKS NOTICE AT WHICH TIME THE CONTROLS CONTRACTOR WILL DEMONSTRATE THE OPERATION OF THE CONTROL SYSTEMS TO ALL CONCERNED. IF ANY ITEMS OF THE CONTROL SEQUENCE CANNOT BE SUCCESSFULLY DEMONSTRATED OR REQUIRE FURTHER TROUBLESHOOTING, A SUBSEQUENT CONTROLS INSPECTION WILL BE REQUIRED AND WILL BE AT THE CONTRACTOR’S EXPENSE. THIS PROVISION SHALL BE A CONDITION OF FINAL ACCEPTANCE. THE CONTRACTOR SHALL BE PREPARED TO DEMONSTRATE ALL ITEMS IN THE SEQUENCE OF OPERATION. SUBSEQUENT CONTROLS DEMONSTRATIONS AND VERIFICATIONS WILL BE REQUIRED AT THE
COMPLETION OF THE PROJECT.
43. THE DRAWINGS HEREIN AND CORRESPONDING REFERENCED EDITED PROJECT SPECIFICATIONS SHALL BE VALID FOR CONSTRUCTION NOT EXCEEDING TWO (2) YEARS FROM THE DATE OF STAMP/SEAL. IF CONSTRUCTION IS REQUIRED AFTER TWO YEARS FROM THE STAMP/SEAL DATE, ADDITIONAL SERVICES SHALL BE REQUIRED FOR THE DESIGNER OF RECORD (OR THEIR REPLACEMENT) TO VERIFY ANY CHANGES TO SITE CONDITIONS, ALONG WITH ANY RELEVANT CHANGES DUE TO CODE CYCLE REVISIONS AND/OR UPDATED LOCAL JURISDICTION REQUIREMENTS.
UPON COMPLETION, DRAWINGS AND CORRESPONDING SPECIFICATIONS SHALL BE RESEALED FOR CONSTRUCTION.
44. CONTRACTOR TO PROVIDE COORDINATED SHOP DRAWINGS FOR INTEGRATING FINAL EQUIPMENT SELECTED.
GENERAL NOTES
1. DURING THE EXECUTION OF THIS CONTRACT, THE PROJECT AREA WILL BE OCCUPIED AND IN USE. WORK DONE UNDER THIS CONTRACT SHALL
BE PERFORMED WITHOUT INCONVENIENCE TO THE OWNER, STAFF, OR OTHER BUILDING OCCUPANTS. ALL WORK SHALL BE COORDINATED WITH THE CONTRACTING OFFICER'S REPRESENTATIVE AND IS SUBJECT TO THEIR CONTINUOUS APPROVAL.
2. "NORMAL" WORKING HOURS AT THE SITE ARE FROM 8:00 AM TO 12:00 AM, MONDAY THROUGH FRIDAY, AND 8:00 AM TO 5:30 PM ON SATURDAY AND SUNDAY; HOURS OUTSIDE OF "NORMAL" WORKING HOURS ARE DEFINED AS "OFF HOURS". THE CONTRACTOR SHALL DEVELOP A SCHEDULE OF DEMOLITION, NEW INSTALLATION, AND CHANGEOVER OF THE MECHANICAL, PLUMBING, AND UTILITY SYSTEMS THAT WILL ALLOW FOR
MINIMUM DOWNTIME.
3. THE CONTRACTOR SHALL DEVELOP AND PRESENT A CONSTRUCTION PLAN AND WORK SCHEDULE WITH ALL PHASING, STAGING, AND SEQUENCING OF WORK FOR APPROVAL PRIOR TO THE COMMENCEMENT AND BEGINNING IMPLEMENTATION OF THEIR WORK. THE CONTRACTOR SHALL DETERMINE PRIOR TO SUBMITTING THEIR WORK SCHEDULE, THE SPECIFIC TIME AVAILABILITY OF ALL AREAS AND SPACES
AND DETERMINE WHAT WORK MUST BE DONE DURING "OFF HOURS".
4. ALL WORK IN OCCUPIED SPACES SHALL BE DONE DURING "OFF HOURS". ALL WORK AS REQUIRED TO RETURN EXISTING SYSTEMS TO FUNCTIONALITY, REPLACE CEILING TILES, AND CONSTRUCTION CLEANUP SHALL BE FULLY COMPLETE AND APPROVED BY THE COR PRIOR TO THE END-OF-SHIFT AND BEFORE THE OPENING OF THE AREA TO STAFF AND OTHER BUILDING OCCUPANTS.
5. WITH THE EXCEPTION OF SIGNIFICANT NOISE GENERATING ACTIVITIES SUCH AS CORE DRILLING, ETC., WORK OUTSIDE OF THE OCCUPIED
SPACES, SUCH AS ON THE ROOF, ETC., MAY BE PERFORMED DURING NORMAL WORKING HOURS.
PHASING
524 Fernwood Drive, Altamonte Springs, Florida 32701 |www.rdcjpa.com| 800.362.1523 Revision Number Revisions: Date:
Area of Work
N
MRI Scan Room 2
FG184B
MRI Exam
FG184A
Equipment
FG184C
Vestibule
C02
Control
FG184
Dressing
FG174
Dressing
FG173
Office
FG172
Alcove
C01
Sedation
FG168
Toilet
Toilet
FG169
Storage
FG170
Reception
FG171
MRI Zone IV
FG181
Equipment
FG178
HAC
FG179
Soiled Utility
FG180
Control
FG182-1
Elec
FG183
Break Room
FG182A
Dressing
FG182C
Dressing
FG182D
Storage
FG182B
Travel Distance to Exit:
Max Allowable = 200' Actual = 8'
Distance to Common Path of Travel:
Max Allowable = 100' Actual = 27'
Project Data Project Name: Replace MRI 1.5T Project Number: VA# 558-22-155 / RDCJPA# 24121.00 Project Address: Durham VA Health Care System
508 Fulton Street Durham, NC 23667
Project Description: Demolish existing exterior EIFS knockout panel, stud partition, RF Shielding enclosure and wood stud walls. Install new MRI machine. Replace exterior EIFS knockout wall and interior stud partition. Rebuild with finishes, RF Shielding enclosure.
Area of Work: 763 Square Feet
Applicable Building Codes Authority Having Jurisdiction: Durham VA Health Care System Building Code: NFPA 101, Life Safety Code, 2021 Edition
International Building Code, 2021 Edition Accessibility Code: Architectural Barriers Act Accessibility Standards w. VA supplement PG 18-13 Energy Code: ASHRAE 90.1-2013 (target 30% better than baseline) Mechanical Code: 2024 International Mechanical Code Plumbing Code: 2021 International Plumbing Code Electrical Code: 2023 NFPA 70 National Electrical Code Fire Code: 2022 NFPA 72 National Fire Alarm Code
Life Safety Summary Occupancy: Existing Health Care (existing - unchanged) Construction Type: I (222) (existing - unchanged) Hazard of Content: Ordinary Hazard Contents Sprinkler System: Fully suppressed (NFPA 13) Seismic Design Category: Category B Occupancy Load: Unchanged Travel Distance Restrictions: Dead-End Corridor Limit: 50 ft.
Common Path of Travel: 100 ft.
Travel Distance to Exits: 200 ft.
0-HR Smoke Partition - [OBC 710] / NFPA 101-8.4
1-HR Smoke Barrier - [OBC 709] / NFPA 101-8.5
1-HR Fire Barrier - [OBC 706] / NFPA 101-8.3
2-HR Fire Barrier - [OBC 706] / NFPA 101-8.3
Means of Egress
Existing Wall to Remain
New Wall
CODE LEGEND
S T A
T E
OF FLO
R
I D A
R E G
IS
TERED ARC
H
IT
E C T
J A S
ON
H. RO
S S
AR101907
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
GENERAL NOTES:
SYMBOL LEGEND:
Watson Engineering, P.C.
782 Johnnie Dodds Blvd, Suite I-2
Mount Pleasant, SC 29464
(607)223-4334 Durham VA Health Care System
KEYED NOTES:
KEY PLAN:
FULLY SPRINKLERED
A u to d e s k D o c s /5
-C S
I-
R e p la c e M
R I
.5 T i n B u ild in g
/2
.0
D u rh a m
M R
I_ R
R
D C
J P
A .r v t
/2 /2
:4
:5
P M
A001
CODE PLAN
BRJR12/20/2024
Replace MRI 1.5T 558-CSI-400
524 Fernwood Drive, Altamonte Springs, Florida 32701 |www.rdcjpa.com| 800.362.1523
N Scale: 1/8" = 1'-0"1 Code Plan
Revisions: Date:
100% CONSTRUCTION
DOCUMENTS
R Q P
5.6
MRI Exam
FG184A
MRI Scan Room 2
FG184B
Control
FG184
Vestibule
C02
Alcove
C01
Corridor
A
A102
A101
A101
184BB 184BA
Equipment
FG184C
3.09
3.09
A
A101
Sim
A101
Sim
R Q P
5.6
MRI Exam
FG184A
MRI Scan Room 2
FG184B
Control
FG184
Vestibule
C02
Alcove
C01
Corridor
V .I
.F
/8
W x
H
2.52
2.04
2.04
2.03
2.03
Align
2.59
2.55 2.55
2.51
2.51
2.58
2.56
2.572.57
2.07 2.53
2.03
2.03
2.06
Area of Work
N
2.04
9.21
5.07
5.10
6.19
7.17
9.24
2.08
7.04
7.81
7.247.24
7 1/4" V.I.F.
2 1/2"
3 1/8" V.I.F.
7.82
7.14
9.219.21
9.02 7.35
7.82
9.21
6.01
6.09
7.35
7.16
6.01
109'-0"
9.419.41
13.01
9.13
7.82
Construct Metal Stud wall only at wall opening used to bring MRI equipment in.
112'-0"
13.02
13.0313.03
13.01
6.01
4 7/8"
2" Isolation Air Space 1 3/4" RF Wall Panel
4 1/8"
EXTERIOR ELEVATION
INTERIOR ELEVATION
SECTION/DETAIL NUMBER
DRAWING NUMBER
EQUIPMENT TAG
COLUMN GRID
NOTE
WINDOW OPENING
DOOR OPENING
ROOM NAME
ROOM NUMBER
WALL TYPE
CEILING HEIGHT
FINISH DESIGNATION101
X
10'-8"
X
X
XX.X-X
XX
X
1/A101
X
A101
X
X
A101
SIM
DEMOLITION LEGEND
Wall to be Demolished
Existing Wall to Remain
Door and Frame to be Demolished
SEE SCHED.
S E
E S
C H
E D
U L E
F
SEE PLAN
H1 J1
13.01
6.01
7.359.21
6.01
6.10
7.82
9.21
9.02 7.35
9.08
7.82
MRI Side
SEE PLAN
13.01
6.01 7.35
9.21
7.82
9.21
9.02 7.35
9.09
7.82
MRI Side
6.14
S1
2.01
9.208.14
9.20
13.02
7.16
8.10 8.10
7.16 9.16
9.17
13.0513.05
13.04
9.19
9.20
MRI Side
13.01
6.01 7.35
9.21
6.01
6.10
7.84
8.09
9.219.21
9.02 7.35
9.08
6.03
7.84
MRI Side
13.01
9.21
6.01 7.35
6.14
7.84
8.09
9.219.21
9.02 7.35
9.09
7.84
6.03
MRI Side
8.09
7.84
7.84
9.21
6.01
6.01 7.35
9.21
13.01
9.21
7.35 6.03
6.03
E x is ti n g
V .I
.F
R
.O
V.I.F.
4'-0" R.O.A
9.419.41
13.0313.03
9.44
9.43 9.42
5.17
5.18
S T A
T E
OF FLO
R
I D A
R E G
IS
TERED ARC
H
IT
E C T
J A S
ON
H. RO
S S
AR101907
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
GENERAL NOTES:
SYMBOL LEGEND:
Watson Engineering, P.C.
782 Johnnie Dodds Blvd, Suite I-2
Mount Pleasant, SC 29464
(607)223-4334 Durham VA Health Care System
KEYED NOTES:
FULLY SPRINKLERED
A u to d e s k D o c s /5
-C S
I-
R e p la c e M
R I
.5 T i n B u ild in g
/2
.0
D u rh a m
M R
I_ R
R
D C
J P
A .r v t
/2 /2
:4
:0
P M
A101
ARCHITECTURAL PLANS, DETAILS,
AND SECTIONS
BRJR
Replace MRI 1.5T 558-CSI-400
N Scale: 1/8" = 1'-0"2 New Work Plan
A. All dimensions are taken to face of gypsum wallboard or unit masonry.
B. Edge of all door jambs at hinge side not otherwise indicated are to be 6" from the face of intersecting walls.
Edge of all door jambs at strike side not otherwise located are to be 18" minimum from the face of intersecting walls.
C. All wood blocking to be fire retardent treated.
D. First floor level 100'-0" equals XXXXX.
E. All partitions shall be type 'S1.0' unless noted otherwise.
F. All walls to be P-1 unless noted otherwise.
G. All flooring to be installed underneath all casework and storage cabinets.
H. Where new finishes would match existing finishes, or new finishes are not required, maintain existing finishes where possible during demolition and constuction phase for reuse. Patch, repair, and match existing finishes where disturbed by construction.
2.01 Existing construction to remain.
2.03 Existing gypsum board on metal stud wall to remain. Patch and install gypsum board wall, as necessary, for 'Like New' appearance before erection of new metal stud wall or installation of new finishes.
2.04 Existing masonry on metal stud wall to remain.
2.06 Existing exterior door to remain.
2.07 Existing ACT Ceiling, and Lighting Fixtures to remain. Protect ACT Tiles and Tracks, as needed, from demolition and construction damage. Coordinate storage of existing ACT tiles and tracks with the VA, as needed.
2.08 Existing steel lintel to remain.
2.51 Remove existing Gypsum Board, Wood Stud, and RF Shielding wall.
2.52 Remove existing EIFS, weather barriers, gypsum substrate, metal stud, insulation, and gypsum wall board.
Protect adjacent masonry, masonry lintels, metal stud, insulation, metal studs, wall finishes from damage during demolition of existing wall and construction of new wall. See New Work Plan.
2.53 Remove existing Vinyl Tile Flooring and prepare substrate to receive new finishes.
2.55 Remove existing RF Shielded Wood Door and Frame.
2.56 Remove existing window. Prepare wall opening to receive new RF Shielded hollow metal window and frame.
2.57 Remove existing ACT Ceiling System, lighting fixtures, mechanical diffusers, and RF Shielding.
2.58 Remove existing Casework.
2.59 Remove existing Gypsum on Metal Stud wall.
3.09 Slope up finish floor with high-strength concrete, to be be level with top of modular RF Shielding floor. Max slope: 1/20.
5.07 6" cold formed metal framing @ 16" O.C. Verify in Field.
5.10 Attach metal stud to existing metal stud structure. Allow for movement between structurers. Min. 1/4", Max 1/2".
5.17 All-thread, as required by Radio-Frequency (RF) ceiling panel manufacturer.
5.18 Fasten unistrut to structure above.
6.01 2" x 4" wood stud
6.03 2" x ___" wood blocking.
6.09 1-3/4" x 1-3/4" preserative treated wood blocking continuous mounting rail.
6.10 2" x 4" wood header
6.14 2" x 4" wood jack and kings studs, as needed.
6.19 5/8" exterior glass matt gypsum board sheathing. Seal all joints.
7.04 Two-piece break metal sill.
7.14 Batt thermal wall insulation. R-19 min.
7.16 Vapor retarder. Seal all joints and penetratons.
7.17 Air and moisture barrier. Lap air and moisture barrier over break metal sill flashing and adjacent air and moisture barriers to manufacturer's recommendations.
7.24 2-1/2" Exterior Insulation Finish System (EIFS) with Finish Coat on Base Coat with Reinforcing Mesh, on approved Insulation Board. Adhere approved Insulation Board to the Air and Moisture Barrier. Complete EIFS install to be to Manufacturer's Recommendations.
7.35 3-1/2" Acoustical batt insulation.
7.81 Continuous sealant over backer rod.
7.82 Continuous acoustic sealant @ non-rated walls / fire-rated sealant @ rated walls each side.
7.84 Acoustic sealant.
8.09 Radio-Frequency (RF) Shielding safety glass window and aluminum framing.
8.10 Aluminum door jamb, provided by RF door manufacturer.
8.14 Flush aluminum sill.
9.02 3 5/8" non-structural metal framing @ 16" o.c.
9.08 Metal framing box header.
9.09 Double metal framing at jambs, typ.
9.13 Attach to Structure above.
9.16 1/8" Hardboard Leveling Layer
9.17 1/8" Sub-Floor
9.19 1/8" Vinly Tile Filler
9.20 Scheduled finish flooring. See Finish Plan.
9.21 5/8" Type X gypsum board.
9.24 Gypsum Control Joint along perimeter of new infill wall.
9.41 Suspended acoustic ceiling tile/grid. See Finish Legend.
9.42 Aluminum wire hangers, as required by ACT Ceiling Manufacturer.
9.43 Aluminum angle
9.44 Aluminum screw.
13.01 1-3/4" thick continuous modular Radio-Frequency (RF) copper shielding wall panel with 1-1/2" acoustical batt insulation. Install to Manufacturer's Specifications.
13.02 1-1/8" total thickness modular Radio-Frequency (RF) Shielding floor system. Install to Manufacturer's Specifications.
13.03 1-3/4" thick continuous modular Radio-Frequency (RF) copper shielding ceiling paneling with 1-1/2" acoustical batt insulation. Install to Manufacturer's Specifications.
13.04 Bronze sprayed structural and electrical seam clamp with flat head self-threading screw. Install to Manufacturer's Specifications.
13.05 3/4" Cell panel with 3oz. copper on top side and polybak on bottom side.
N Scale: 1/8" = 1'-0"1 Demolition Plan
524 Fernwood Drive, Altamonte Springs, Florida 32701 |www.rdcjpa.com| 800.362.1523
GENERALCONSTRUCTION NOTES
A. Remove all existing partitions shown with dashed line. Repair wall and floor finishes to match adjacent surfaces at areas of removal or scheduled new surfaces as required.
B. Remove all existing borrowed lights, doors and hollow metal frames where indicated with dashed line. Unused hardware shall be turned over to Owner. Box hardware before giving to Owner.
C. Remove wall-hung toilet room accessories and medical equipment as required and turn over to Owner.
D. See Plumbing, HVAC and Electrical Drawings for removal work associated with those trades.
E. Computer and communication cables and/or other wiring, piping, tubing, etc. serving other areas and functions located above ceiling or in walls of area to be remodeled shall be maintained during demolition/construction process. Coordinate any required changes to these systems with the C.O.R.
F. Repair existing walls to match adjacent undisturbed wall where cabinets, countertops, etc. are shown to be removed.
G. Prepare walls and floor slabs as required to receive new fininsh materials where existing finish is indicated to be removed.
H. Remove existing corner guards in areas to be remodeled.
GENERAL DEMOLITION NOTES
Key Plan Scale: 3/4" = 1'-0"4 Wall Section - Exterior
Scale: 1 1/2" = 1'-0"3 Wall Section - Interior
DOOR SCHEDULE
DOOR
BETWEEN DOOR DESCRIPTION FRAME DESCRIPTION
FIRE
RATING REMARKSRM RM TYPE MAT'L
SIZE
TYPE MAT'L HEAD JAMB Sill Detail# W H T
184BA C02 FG184B F WD 1 4'-0" 7'-0" 1 3/4" 1 HM H1 J1 S1 - RF Shielding / Access Control
184BB FG184C FG184B F WD 1 3'-0" 7'-0" 1 3/4" 1 HM H1 J1 S1 - RF Shielding / Access Control
N.T.S.Scale:
Door Types N.T.S.Scale:
H.M. Door Frame Details
Scale: 3" = 1'-0"5 Section Detail - RF Flooring
N.T.S.Scale:
Window Detail - Header
N.T.S.Scale:
Window Detail - Jamb N.T.S.Scale:
Window Detail - Sill
N.T.S.Scale:
Window Type
EQUIPMENT SCHEDULE
TAG Description Manufacturer Model
Supplied By Installed By
VA Contractor VA Contractor
1 MRI Imaging Machine Siemens Magnetom Sola X X
2 MRI Electronics Cabinet Siemens X X
3 MRI SEP Cabinet Siemens X X
4 MRI UPS with Battery Siemens X X
Section Detail - Ceiling
Revisions: Date:
100% CONSTRUCTION
DOCUMENTS
12/20/2024
RB-1
PL-1
SS-1
PL-1
P-1
RB-1
6'-1"
3'-0" 3'-0"
B1
B1
W3W3
B1
PL-1
3'-0" 3'-0"
S E
E I N
T E
R
IO
R E
L E
V A
T
IO
N S o r
2'-1"
/4
9.60
6.27
6.29
6.30
6.38
12.20
12.19
6.36
Drawer
B1
3"
1"
12.10
W3
26.01
6.29
6.38
6.30
6.27
6.36
6.37
1'-2"
S E
E E
L E
V A
T
IO
N
Area of Work
N
ACT-1
2.07
9'-0"
9'-0"
MRI Exam
FG184A
MRI Scan Room 2
FG184B
Control
FG184
Vestibule
C02
Alcove
C01
Corridor
WSF-1Equipment
FG184C
CG-1
VCT-1 VCT-1
RB-1
CG-1
REFLECTED CEILING PLAN LEGEND
Recessed downlight fixture
Supply Air Diffuser
Return Air Grille
2' x 4' light fixture
2' x 2' acoustical tile ceiling
Gypsum board / portland cement plaster ceiling / soffit - painted
Pendant light fixture
Chain-hung light fixture
Access panel
Recessed wall-wash fixture
Surface-mounted strip light fixture
EXTERIOR ELEVATION
INTERIOR ELEVATION
SECTION/DETAIL NUMBER
DRAWING NUMBER
EQUIPMENT TAG
COLUMN GRID
NOTE
WINDOW OPENING
DOOR OPENING
ROOM NAME
ROOM NUMBER
WALL TYPE
CEILING HEIGHT
FINISH DESIGNATION101
X
10'-8"
X
X
XX.X-X
XX
X
1/A101
X
A101
X
X
A101
SIM
S T A
T E
OF FLO
R
I D A
R E G
IS
TERED ARC
H
IT
E C T
J A S
ON
H. RO
S S
AR101907
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
GENERAL NOTES:
SYMBOL LEGEND:
Watson Engineering, P.C.
782 Johnnie Dodds Blvd, Suite I-2
Mount Pleasant, SC 29464
(607)223-4334 Durham VA Health Care System
KEYED NOTES:
KEY PLAN:
FULLY SPRINKLERED
A u to d e s k D o c s /5
-C S
I-
R e p la c e M
R I
.5 T i n B u ild in g
/2
.0
D u rh a m
M R
I_ R
R
D C
J P
A .r v t
/2 /2
:4
:0
P M
A102
FINISH PLANS, ELEVATIONS, AND
SCHEDULES
100% CONSTRUCTION
DOCUMENTS
SMJR
Replace MRI 1.5T 558-CSI-400
A. Contractor verify dimensions prior to casework fabrication
B. Verify all conditions prior to fabrication, erection, installation.
C. See Electrical Drawings for lighting specifications.
D. See P, M, E and FP drawings for additional information regarding final information and quantites of fixtures/devices to be installed in ceiling.
E. All walls to be P-1 unless noted otherwise.
2.07 Existing ACT Ceiling, and Lighting Fixtures to remain. Protect ACT Tiles and Tracks, as needed, from demolition and construction damage. Coordinate storage of existing ACT tiles and tracks with the VA, as needed.
6.27 Concealed hinge.
6.29 Adjustable shelf. Plastic laminate over 3/4" particle board.
6.30 4" wire pull.
6.36 Plastic laminate over 3/4" particle board.
6.37 Sloped top. Plastic laminate over 3/4" particle board. Finish to match the casework laminate.
6.38 Self-latching, key-operated lock.
9.60 Resilient base. See Finish Legend.
12.10 Applied solid surface sidesplash.
12.19 Applied solid surface backsplash.
12.20 Solid surface material over 3/4" plywood, with 1/4" eased edge.
26.01 Light fixture. See Electrical Drawings.
GENERAL INT. ELEVATION NOTES
FINISH LEGEND
CODE FINISH TYPE MANUFACTURER PRODUCT / COLOR COMMENTS
06 41 16 PLASTIC LAMINATE CLAD ARCHITECTURAL CABINETS
PL-1 PLASTIC LAMINATE WILSONART TBD
09 51 00 ACOUSTICAL CEILINGS
ACT-1 ACOUSTICAL CEILING TILES ARMSTRONG Clean Room FL 2x2 Tile with 15/16" Aluminium Grid
09 65 16 RESILIENT SHEET FLOORING
RB-1 RUBBER BASE MANNINGTON Ideal Base, Color:TBD
WSF-1 WELDED SHEET FLOORING MANNINGTON BioSpe Armor Color: TBD
09 65 19 RESILIENT TILE FLOORING
VCT-1 VINYL COMPOSITE TILE Match Existing VCT Interior finish
09 91 00 PAINTING
P-1 INTERIOR PAINT PPG TBD
10 26 00 WALL AND DOOR PROTECTION
CG-1 CORNER GUARD INPRO 130 High Impact Surface Mount Corner Guard - 3" Wing, Full Height color TBD
12 36 00 COUNTERTOPS
SS-1 SOLID SURFACE WILSONART TBD
Scale: 1/4" = 1'-0"A Casework Elevation
Scale: 1" = 1'-0"
Sections - Casework
524 Fernwood Drive, Altamonte Springs, Florida 32701 |www.rdcjpa.com| 800.362.1523
N Scale: 1/8" = 1'-0"1 Reflected Ceiling Plan
N Scale: 1/8" = 1'-0"2 Finish Plan
Manufacturers' trade names and numbers are provided in Finish Legend to identify design intent. Equivalent products from other manufacturers that meet requirments of technical specifications will be acceptable upon approval in writing by contracting officer.
A. All wood blocking shall be fire retardant.
B. Light fixtures, sprinkler heads, speakers, and HVAC devices shall be located as close as possible to the center of acoustical ceiling panels. Confirm any deviation from this with the Project Engineer.
C. See P, M, E and FP drawings for additional information regarding final information and quantites of fixtures/devices to be installed in ceiling.
D. See Electrical Drawings for lighting specifications.
F. Ceiling height to be 9'-0" AFF, unless noted otherwise.
GENERAL RCP NOTES
Revisions: Date:
All finishes subject to approval of VA Interior Designer.
12/20/2024
T
T
T
T
T
T
T
T
T
T
T
T
T
R Q P N M
5.6
6.7
MRI Scan Room 2
FG184B
Equipment
FG184C
MRI Exam
FG184A
Control
FG184
Vestibule C02
Equipment
FG178
HAC
FG179
Soiled Utility
FG180
MRI Zone IV
FG181
Control
FG182-1
Elec
FG183
Break Room
FG182A
Storage
FG182B
Dressing
FG182D
Dressing
FG182C
Sedation
FG168
Toilet
Toilet
FG169
Reception
FG171
Storage
FG170
Office
FG172
Dressing
FG173
Dressing
FG174
Alcove C01
CH-1 CH-2 HP-1 HP-2
EF-2 EF-1
A. CONTRACTOR SHALL VERIFY ALL DIMENSIONS AND CONDITIONS PRIOR TO BEGINNING WORK.
B. ALL AREAS OF THE BUILDING AFFECTED BY MECHANICAL DEMOLITION WORK SHALL BE PATCHED AND REPAIRED TO MATCH ADJACENT FINAL FINISHES AND REQUIRED FIRE RATING. THE REPAIR WORK SHALL BE
DONE TO THE OWNER'S SATISFACTION.
C. DEMOLITION WORK SHALL INCLUDE REMOVAL OF EXISTING BUILDING CONSTRUCTION TO EXTENT REQUIRED
TO PERFORM NEW CONSTRUCTION ACTIVITIES INDICATED.
D. REMOVE ALL EXISTING MECHANICAL SYSTEM COMPONENTS, MATERIALS, APPARATUS, AND APPLIANCES WITHIN THE DESIGNATED PROJECT AREA, UNLESS OTHERWISE NOTED. COORDINATE WITH ARCHITECTURAL, MECHANICAL, ELECTRICAL, AND FIRE PROTECTION CONTRACT DOCUMENTS FOR ADDITIONAL SCOPE OF
WORK FOR SYSTEMS TO BE REMOVED OR RETAINED TO EXTENT REQUIRED TO PERFORM NEW
CONSTRUCTION ACTIVITIES INDICATED.
E. DEMOLITION PROCEDURES SHALL PROVIDE FOR SAFE CONDUCT OF THE WORK, PROTECTION OF PERSONNEL, CAREFUL REMOVAL AND OTHER DISPOSITION OF MATERIALS SPECIFIED TO BE SALVAGED.
PROTECTION OF PROPERTY TO REMAIN UNDISTURBED, COORDINATION WITH OTHER WORK IN PROGRESS, AND TIMELY DISCONNECTION OF UTILITY SERVICES.
F. EXISTING WORK SHALL BE CUT, DRILLED, ALTERED, REMOVED, OR TEMPORARILY REMOVED AND REPLACED FOR PERFORMANCE OF WORK UNDER THE CONTRACT. WORK REPLACED SHALL MATCH SIMILAR EXISTING WORK. STRUCTURAL MEMBERS OF CONCRETE OR STRUCTURAL STEEL SHALL NOT BE CUT OR ALTERED, EXCEPT AS SHOWN, WITHOUT AUTHORIZATION OF THE ENGINEER. WORK REMAINING IN PLACE DAMAGED OR DEFACED DURING THIS CONTRACT SHALL BE RESTORED TO THE CONDITION AT TIME OF AWARD OF CONTRACT. CUT, ALTER, REMOVE, OR TEMPORARILY REMOVE AND REPLACE EXISTING WORK FOR
INSTALLATION OF ELECTRICAL WORK.
G. UNLESS SPECIFIED OTHERWISE, ALL PIPING, WIRING, CONDUIT, AND OTHER DEVICES ASSOCIATED WITH THE EQUIPMENT NOT RE-USED IN THE NEW WORK AREA SHALL BE COMPLETELY DEMOLISHED AND REMOVED FROM THE OWNER'S PROPERTY. THIS INCLUDES ALL CONCRETE PADS, PIPE, VALVES, FITTINGS, INSULATION, AND ALL HANGERS, INCLUDING THE TOP CONNECTION AND ANY FASTENINGS TO BUILDING STRUCTURAL SYSTEMS. ALL OPENINGS SHALL BE SEALED AFTER REMOVAL OF EQUIPMENT, PIPES, DUCTS, AND OTHER
PENETRATIONS IN ROOF, WALLS, FLOORS, IN AN APPROVED MANNER. STRUCTURAL
H. REMOVAL AND DISPOSAL OF ALL EXISTING REFRIGERANT SHALL BE COMPLETED IN ACCORDANCE WITH SECTION 608 OF THE CLEAN AIR ACT (CAA), AND THE ENVIRONMENTAL PROTECTION AGENCY (EPA)
REGULATIONS LISTED IN 40 CFR PART 82, SUBPARTS A AND F.
AREA OF WORK
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
GENERAL NOTES:
SYMBOL LEGEND:
Watson Engineering, P.C.
782 Johnnie Dodds Blvd, Suite I-2 Mount Pleasant, SC 29464 (607)223-4334
KEYED NOTES:
KEY PLAN:
Au to de sk
D oc s:
//5
-C
SI
-4
R ep la ce M
R I 1
.5 T in B ui ld in g
1/ D ur ha m M
R I_
R
.rv t
/2
0/
:1 0:
P
M
MD101
MECHANICAL DEMOLITION PLAN Replace MRI 1.5T
508 Fulton Street Durham, NC 27705
12/20/2024 MPL JWC
558-CSI-400
100% CONSTRUCTION
DOCUMENTS
1/4" = 1'-0"1 Mech - Demo
NOTE
1 EXISTING MRI EQUIPMENT TO BE REMOVED.
2 EXISTING UPS TO BE REMOVED.
3 EXISTING MINI SPLIT AIR CONDITIONER TO BE TEMPORARILY REMOVED AND STORED FOR REINSTALLATION.
4 EXISTING CHILLER FOR MRI EQUIPMENT TO BE REMOVED.
5 EXISTING VAV BOX TO BE REMOVED FOR REPLACEMENT.
524 Fernwood Drive, Altamonte Springs, Florida 32701 |www.rdcjpa.com| 800.362.1523 Revision Number Revisions: Date:
N 1" = 100'-0"
KEY PLAN - BUILDING 1
SA
RA RA RA
R A
EXH
SA
SA
SA
SA
SA
RA
RA
R A
RA
R A
RA
R A
EX
H
EXH
EXHEXH
R A
SA
250 CFM
SA
300 CFM
SA
250 CFM
RA
800 CFM
SA
250 CFM
SA
300 CFM
RA
800 CFM
SA
250 CFM
QUENCH
PIPE FROM
MRI TO ROOF
12"ø EXH
9"ø SA
12"ø SA
14"x14" RA
16"ø SA9" ø
SA
12"x8" EXH
EF-2 EF-1
"x
8" E
XH
"x
8" E
XH
12"x8" EXH
9" ø
SA
"ø
S A
"x
" R
A
12"x12" RA10"ø SA
12"ø EXH
9" ø
SA
12"ø SA 16 "ø
S A
"x
" R
A
TYP.
TYP.
R Q P N M
5.6
6.7
Equipment
FG184C
MRI Scan Room 2
FG184B
MRI Exam
FG184A
Vestibule C02
Control
FG184
Equipment
FG178
HAC
FG179
Soiled Utility
FG180
MRI Zone IV
FG181
Control
FG182-1
Elec
FG183
Break Room
FG182A
Storage
FG182B
Dressing
FG182C
Dressing
FG182D
Alcove C01
Sedation
FG168
Toilet
Toilet
FG169
Reception
FG171
Storage
FG170
Alcove
Dressing
FG173
Dressing
FG174
Office
FG172
VAV-01(E)
A. CONTRACTOR SHALL VERIFY ALL DIMENSIONS AND CONDITIONS PRIOR TO BEGINNING WORK.
B. ALL AREAS OF THE BUILDING AFFECTED BY MECHANICAL DEMOLITION WORK SHALL BE PATCHED AND REPAIRED TO MATCH ADJACENT FINAL FINISHES AND REQUIRED FIRE RATING. THE REPAIR WORK SHALL BE
DONE TO THE OWNER'S SATISFACTION.
C. DEMOLITION WORK SHALL INCLUDE REMOVAL OF EXISTING BUILDING CONSTRUCTION TO EXTENT REQUIRED
TO PERFORM NEW CONSTRUCTION ACTIVITIES INDICATED.
D. REMOVE ALL EXISTING MECHANICAL SYSTEM COMPONENTS, MATERIALS, APPARATUS, AND APPLIANCES WITHIN THE DESIGNATED PROJECT AREA, UNLESS OTHERWISE NOTED. COORDINATE WITH ARCHITECTURAL, MECHANICAL, ELECTRICAL, AND FIRE PROTECTION CONTRACT DOCUMENTS FOR ADDITIONAL SCOPE OF
WORK FOR SYSTEMS TO BE REMOVED OR RETAINED TO EXTENT REQUIRED TO PERFORM NEW
CONSTRUCTION ACTIVITIES INDICATED.
E. DEMOLITION PROCEDURES SHALL PROVIDE FOR SAFE CONDUCT OF THE WORK, PROTECTION OF PERSONNEL, CAREFUL REMOVAL AND OTHER DISPOSITION OF MATERIALS SPECIFIED TO BE SALVAGED.
PROTECTION OF PROPERTY TO REMAIN UNDISTURBED, COORDINATION WITH OTHER WORK IN PROGRESS, AND TIMELY DISCONNECTION OF UTILITY SERVICES.
F. EXISTING WORK SHALL BE CUT, DRILLED, ALTERED, REMOVED, OR TEMPORARILY REMOVED AND REPLACED FOR PERFORMANCE OF WORK UNDER THE CONTRACT. WORK REPLACED SHALL MATCH SIMILAR EXISTING WORK. STRUCTURAL MEMBERS OF CONCRETE OR STRUCTURAL STEEL SHALL NOT BE CUT OR ALTERED, EXCEPT AS SHOWN, WITHOUT AUTHORIZATION OF THE ENGINEER. WORK REMAINING IN PLACE DAMAGED OR DEFACED DURING THIS CONTRACT SHALL BE RESTORED TO THE CONDITION AT TIME OF AWARD OF CONTRACT. CUT, ALTER, REMOVE, OR TEMPORARILY REMOVE AND REPLACE EXISTING WORK FOR
INSTALLATION OF ELECTRICAL WORK.
G. UNLESS SPECIFIED OTHERWISE, ALL PIPING, WIRING, CONDUIT, AND OTHER DEVICES ASSOCIATED WITH THE EQUIPMENT NOT RE-USED IN THE NEW WORK AREA SHALL BE COMPLETELY DEMOLISHED AND REMOVED FROM THE OWNER'S PROPERTY. THIS INCLUDES ALL CONCRETE PADS, PIPE, VALVES, FITTINGS, INSULATION, AND ALL HANGERS, INCLUDING THE TOP CONNECTION AND ANY FASTENINGS TO BUILDING STRUCTURAL SYSTEMS. ALL OPENINGS SHALL BE SEALED AFTER REMOVAL OF EQUIPMENT, PIPES, DUCTS, AND OTHER
PENETRATIONS IN ROOF, WALLS, FLOORS, IN AN APPROVED MANNER. STRUCTURAL
H. REMOVAL AND DISPOSAL OF ALL EXISTING REFRIGERANT SHALL BE COMPLETED IN ACCORDANCE WITH SECTION 608 OF THE CLEAN AIR ACT (CAA), AND THE ENVIRONMENTAL PROTECTION AGENCY (EPA)
REGULATIONS LISTED IN 40 CFR PART 82, SUBPARTS A AND F.
AREA OF WORK
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
GENERAL NOTES:
SYMBOL LEGEND:
Watson Engineering, P.C.
782 Johnnie Dodds Blvd, Suite I-2 Mount Pleasant, SC 29464 (607)223-4334
KEYED NOTES:
KEY PLAN:
Au to de sk
D oc s:
//5
-C
SI
-4
R ep la ce M
R I 1
.5 T in B ui ld in g
1/ D ur ha m M
R I_
R
.rv t
/2
0/
:1 0:
P
M
MD102
MECHANICAL REFLECTED CEILING
DEMOLITION
Replace MRI 1.5T
508 Fulton Street Durham, NC 27705
12/20/2024 MPL JWC
558-CSI-400
100% CONSTRUCTION
DOCUMENTS
1/4" = 1'-0"1 Mechanical Ceiling Plan - Demo
NOTE
1 CONTRACTOR TO DISCONNECT AND DEMOLISH THE EXISTING QUENCH PIPE FROM THE EXISTING MRI UNIT TO THE ROOF. DISCONNECT AND DEMOLISH THE EMERGENCY EXHAUST DUCTWORK FROM THE EXISTING MRI UNIT
TO DIRECTLY ABOVE THE CEILING FOR FUTURE CONNECTION TO NEW MRI UNIT.
2 CONTRACTOR TO DEMOLISH THE EXISTING SUPPLY DIFFUSERS AND RETURN GRILLES.
3 CONTRACTOR TO REMOVE ALL EXISTING DUCTWORK ABOVE THE CEILING OF MRI EXAM 2 (RM 103) TO THE
EXTENT INDICATED ON THIS PLAN.
4 EXISTING MINI SPLIT AIR CONDITIONER TO BE TEMPORARILY REMOVED AND STORED FOR REINSTALLATION.
5 EXISTING VAV BOX TO BE REMOVED FOR REPLACEMENT.
524 Fernwood Drive, Altamonte Springs, Florida 32701 |www.rdcjpa.com| 800.362.1523 Revision Number Revisions: Date:
RA
R Q P N M
5.6
6.7
A. CONTRACTOR SHALL VERIFY ALL DIMENSIONS AND CONDITIONS PRIOR TO BEGINNING WORK.
B. ALL AREAS OF THE BUILDING AFFECTED BY MECHANICAL DEMOLITION WORK SHALL BE PATCHED AND REPAIRED TO MATCH ADJACENT FINAL FINISHES AND REQUIRED FIRE RATING. THE REPAIR WORK SHALL BE
DONE TO THE OWNER'S SATISFACTION.
C. DEMOLITION WORK SHALL INCLUDE REMOVAL OF EXISTING BUILDING CONSTRUCTION TO EXTENT REQUIRED
TO PERFORM NEW CONSTRUCTION ACTIVITIES INDICATED.
D. REMOVE ALL EXISTING MECHANICAL SYSTEM COMPONENTS, MATERIALS, APPARATUS, AND APPLIANCES WITHIN THE DESIGNATED PROJECT AREA, UNLESS OTHERWISE NOTED. COORDINATE WITH ARCHITECTURAL, MECHANICAL, ELECTRICAL, AND FIRE PROTECTION CONTRACT DOCUMENTS FOR ADDITIONAL SCOPE OF
WORK FOR SYSTEMS TO BE REMOVED OR RETAINED TO EXTENT REQUIRED TO PERFORM NEW
CONSTRUCTION ACTIVITIES INDICATED.
E. DEMOLITION PROCEDURES SHALL PROVIDE FOR SAFE CONDUCT OF THE WORK, PROTECTION OF PERSONNEL, CAREFUL REMOVAL AND OTHER DISPOSITION OF MATERIALS SPECIFIED TO BE SALVAGED.
PROTECTION OF PROPERTY TO REMAIN UNDISTURBED, COORDINATION WITH OTHER WORK IN PROGRESS, AND TIMELY DISCONNECTION OF UTILITY SERVICES.
F. EXISTING WORK SHALL BE CUT, DRILLED, ALTERED, REMOVED, OR TEMPORARILY REMOVED AND REPLACED FOR PERFORMANCE OF WORK UNDER THE CONTRACT. WORK REPLACED SHALL MATCH SIMILAR EXISTING WORK. STRUCTURAL MEMBERS OF CONCRETE OR STRUCTURAL STEEL SHALL NOT BE CUT OR ALTERED, EXCEPT AS SHOWN, WITHOUT AUTHORIZATION OF THE ENGINEER. WORK REMAINING IN PLACE DAMAGED OR DEFACED DURING THIS CONTRACT SHALL BE RESTORED TO THE CONDITION AT TIME OF AWARD OF CONTRACT. CUT, ALTER, REMOVE, OR TEMPORARILY REMOVE AND REPLACE EXISTING WORK FOR
INSTALLATION OF ELECTRICAL WORK.
G. UNLESS SPECIFIED OTHERWISE, ALL PIPING, WIRING, CONDUIT, AND OTHER DEVICES ASSOCIATED WITH THE EQUIPMENT NOT RE-USED IN THE NEW WORK AREA SHALL BE COMPLETELY DEMOLISHED AND REMOVED FROM THE OWNER'S PROPERTY. THIS INCLUDES ALL CONCRETE PADS, PIPE, VALVES, FITTINGS, INSULATION, AND ALL HANGERS, INCLUDING THE TOP CONNECTION AND ANY FASTENINGS TO BUILDING STRUCTURAL SYSTEMS. ALL OPENINGS SHALL BE SEALED AFTER REMOVAL OF EQUIPMENT, PIPES, DUCTS, AND OTHER
PENETRATIONS IN ROOF, WALLS, FLOORS, IN AN APPROVED MANNER. STRUCTURAL
H. REMOVAL AND DISPOSAL OF ALL EXISTING REFRIGERANT SHALL BE COMPLETED IN ACCORDANCE WITH SECTION 608 OF THE CLEAN AIR ACT (CAA), AND THE ENVIRONMENTAL PROTECTION AGENCY (EPA)
REGULATIONS LISTED IN 40 CFR PART 82, SUBPARTS A AND F.
AREA OF WORK
Project Number
Building Number
Drawing Number
Project…
This is the start of the file's text. The full file is on GovTribe.