Amendment 0002 Site Planning and Utility Connection Guide.pdf
PDF 10 MB Posted
- Solicitation number
- 36C25626R0008
About this file
This document is a Site Planning and Utility Connections Guide for the STERIS Mobile Endoscope Reprocessing Unit, prepared for the VA Medical Center-Fayetteville (Revision A, dated 02/10/2025).
The guide provides comprehensive technical specifications and requirements for deploying a 53-foot expandable mobile endoscope reprocessing unit and accompanying 28-foot support unit at healthcare facilities. Key infrastructure requirements include 480VAC 3-phase 60Hz electrical service (350 AMP MOP for the mobile unit, 100 AMP MOP for the support unit) with disconnects located within 30 feet of the units, cold water supply (1.5" FNPT connection, 50-80 psi dynamic pressure, 700 GPH average, 40-70°F temperature), 4-inch wastewater connection (700 GPH average, 140°F discharge temperature), 2-inch Victaulic fire sprinkler coupling (143 GPM at 46 psi minimum residual pressure), and instrument air lines connecting the support unit to the mobile unit (maximum 120-foot separation). The facility must provide grounding rod installation per NEC Article 250, HVAC condensate drainage with air gaps, IT network extension to an internal patch panel, and all utility line protection. Customer responsibilities include obtaining permits, making all utility connections, providing a connecting corridor or structure (constructed only after unit arrival), installing protective barriers around the unit, weather-sealing interfaces, and providing fire extinguishers. The mobile unit requires minimum 45-inch threshold height (up to 54 inches with dunnage), operates within a 75-foot by 30-foot footprint, and must be anchored with tie-down points installed post-delivery. Delivery slopes cannot exceed specified pitch/rise ratios, and minimum support pads are required based on site conditions. The document establishes a two-week commissioning schedule with deployment occurring on Tuesdays and turnover on Fridays of Week 2, with parallel responsibilities allocated between STERIS and the customer throughout pre-deployment, installation, operation, and decommissioning phases.
View the file
Other files
Show all 18
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Mobile Endoscope Reprocessing Unit Site Planning and Utility Connections Guide sterismobilesolutions.com 800.548.4873
VA M
ed ica l C en ter -Fay ett ev ille
MOBILE ENDOSCOPE REPROCESSING UNIT
SITE PLANNING & UTILITY GUIDE
REV A 02/10/2025
Introduction and Disclaimer
The purpose of this document is to provide the basic information needed for site planning and implementation of the STERIS Mobile Endoscope Reprocessing Unit at your facility. Please note that STERIS mobile units may have minor variations in dimensions, equipment layout, product portfolio, ancillary cooling packages and means of egress.
Once a specific Mobile Endoscope Reprocessing Unit has been assigned to your facility, STERIS will provide guidance regarding any requirements that may be different from this site guide.
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Symbol Definitions
Important note that must be followed
May result in additional charges or delays
May result in personal injury or damage to the equipment
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Important Notes
1. Prior to deployment, STERIS will perform a site visit and/or teleconference to review mechanical, electrical, and plumbing (MEP) requirements with all parties involved to ensure a successful project.
2. Customer is responsible for all licenses, permits, and government approvals.
3. Customer is responsible to make all utility connections.
4. Power must be readily available within 30’ of the front/nose of the Mobile Endoscope Reprocessing Unit and the Support Unit upon delivery. All utilities must be stubbed out to the deployment locations.
5. Customer is responsible for providing a connecting structure or corridor to transport instruments into and out of the mobile unit. This structure can only be built AFTER the mobile unit arrives on site. Substantial costs and project delays will result if the structure is built prior to mobile unit arrival and placement.
6. STERIS will provide a staircase for the Mobile Endoscope Reprocessing Unit emergency exit door and the Support Unit only. If the provided stairs do not meet local code or preference requirements, the Customer is required to provide alternate stairs.
7. Anchor tie down points must only be constructed or installed AFTER the mobile unit arrives on site.
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
General Project Progression
Contract Signed
• Solution Planning and Proposal
MEP
Complete
• Clinical and MEP Planning
Customer Operation
• Mobile Unit Delivery
• Mobile Unit Setup / Corridor Built
• Training and Validation
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
ACTIVITY STERIS CUSTOMER NOTES
Pre-Planning
Concept Site Visit ✓ ✓ PDM/PS Capacity Study ✓ Wheels Out Report Required Proposal Submittal ✓ Executed Contract ✓ ✓
Post Agreement Planning STERIS Professional Services Visit ✓ ✓ MEP Page Flip ✓ ✓
Pre-Commissioning Electrical Ready ✓ 2 Weeks Prior to Delivery Date Drain Lines Stubbed Out ✓ 2 Weeks Prior to Delivery Date Water Lines Stubbed Out ✓ 2 Weeks Prior to Delivery Date Outline Mobile Unit and Support Placement ✓ 2 Weeks Prior to Delivery Date Install Road Plates ✓ 2 Weeks Prior to Delivery Date Final Site Readiness Confirmation ✓ 2 Weeks Prior to Delivery Date Coordinate Mobile Unit Delivery ✓ ✓ 2 Weeks Prior to Delivery Date
Commissioning - Unit Setup (Week 1) Delivery of Mobile Unit ✓ Units Deliver on Tuesdays Mobile Unit Expanded and Leveled ✓ Connect Electrical Supplies ✓ Priority for Day 1 Mobile Unit Interior Setup ✓ Install Tie Down Connection Points ✓ Connect Plumbing ✓ STERIS Equipment Start-Up ✓ Facilities / BioMed In-Services ✓ ✓ Install Fire Supression ✓ Install IT ✓
Commissioning - Clinical Activities (Week 2) STERIS Equipment In-Services ✓ ✓ IPT (3) Cycles Confirmation ✓ ✓ Final Sign-Off ✓ ✓ EVS Terminal Cleaning ✓
Mobile Unit Turn-Over to Customer ✓ Friday of Week 2 is Turn-Over Day
VA M
ed ica l C en ter -Fay ett
REV A 02/10/2025
Project Activities and Responsibilities
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
PROJECT ACTIVITIES and RESPONSIBILITIES
PROJECT PHASE ACTION ITEM STERIS CUSTOMER
Inquiry into a STERIS mobile unit solution ✓ ✓
Collaborative site visit to determine unit placement and feasibility ✓ ✓
Perform capacity study to validate mobile capacity vs. facility need ✓ ✓
Provide facility drawings for project ✓
Identify and create mobile unit placement options ✓ ✓
Provide all documentation to Customer (drawings, facility requirements, layout, etc.) ✓
Provide quote for mobile unit solution ✓
PROCUREMENT Sign Rental Agreement and issue Purchase Order at least 2 months in advance of project ✓ ✓
Project kickoff call (STERIS Mobile Solutions will lead) ✓ ✓
Provide clinical operations data to STERIS Professional Services team ✓
Create and distribute a commission schedule ✓
Provide mobile unit project drawings ✓
Meeting with Customer's Planning and MEP teams to review drawings, utilities, and commission ✓ ✓
Perform or confirm a capacity study and wheels out report ✓
Perform a water sample test to ensure water meets requirements ✓
Review consumables needed for start-up ✓ ✓
Confirm site readiness for receiving mobile unit ✓ ✓
Pre-Deployment Coordinate logistics and resources for delivery ✓
Solution Planning and Proposal
Clinical + MEP Planning and
Execution
Responsibility
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Project Activities and Responsibilities
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
PROJECT ACTIVITIES and RESPONSIBILITIES
PROJECT PHASE ACTION ITEM STERIS CUSTOMER
Place, level, and expand mobile unit upon arrival ✓
Perform general commission activities (configuration, etc.) ✓
Complete utilities and corridor connections ✓
Review unit setup and emergency protocols with Customer Facilities and BioMed ✓ ✓
Inspect all STERIS equipment for damage during transport ✓
Perform STERIS equipment start-up ✓
Perform testing and validation cycles ✓ ✓
Perform equipment in-service activities to all users for all shifts ✓ ✓
Review Customer Service Protocol with Customer (1-800 number) ✓ ✓
Obtain signatures for individuals who received in-service training ✓ ✓
Change management/transition into mobile unit ✓ ✓
Review current processes and remove bottlenecks (scheduling, throughput, etc.) ✓
Determine Go-Live date ✓
Perform terminal cleaning operations ✓
Go-Live ✓
Check-in calls to discuss project status and schedule (STERIS Mobile Solutions to lead) ✓ ✓
Perform equipment preventative maintenance ✓
Act as the "eyes and ears" on site for Mobile Solutions Team (issues, feedback, etc.) ✓
Notify changes of schedule and if a contract extension is needed ✓
On-Site Implementation
Customer Operation
Responsibility
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Project Activities and Responsibilities
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
PROJECT ACTIVITIES and RESPONSIBILITIES
PROJECT PHASE ACTION ITEM STERIS CUSTOMER
Project call to develop decommission schedule (STERIS Mobile Solutions to lead) ✓ ✓
Publish decommission schedule ✓
Coordinate logistics and resources for removal of unit ✓ ✓
Perform terminal cleaning activities ✓
Perform equipment shutdown and winterization activities (documentation will be provided) ✓
Remove roll around equipment and configure unit for transport ✓
Perform secondary terminal cleaning for areas of removed equipment ✓
Remove all utility connections and associated connecting structure ✓
Obtain Customer signatures to trigger unit departure ✓ ✓
Complete survey with STERIS Mobile Solutions Project Manager ✓ ✓
Project Completion
Responsibility
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Mobile Endoscope Reprocessing Unit and Support Unit At-A-Glance
• Mobile Endoscope Reprocessing Unit o A 53’ expandable mobile unit that is operated as the endoscope reprocessing department.
o Features essential equipment necessary for endoscope reprocessing including multi-basin sinks, disinfection, drying and storage equipment, and a multi-stage water filtration system.
o Requires 480V power, cold water, waste drain, and fire sprinkler connections to customer facility, a facility-provided access corridor or structure, IT connection, as well as an air hose connection (provided) to the support unit.
• Support Unit o A 28’ unit that houses the compressor and filtration system that supplies instrument air to the Mobile Endoscope Reprocessing Unit and fulfills additional storage and transport needs.
o Requires 480V power connection to customer facility.
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Support Unit
Mobile Endoscope Reprocessing Unit
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Site Planning and Preparation
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Floorplan and Facility Access
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Customer is responsible for providing a facility access structure (deck or corridor) to transport instruments into and out of the mobile unit.
②
① ① ①
② ②
③ ③ ③ ③ ③
⑤
④ ④
⑤
⑥
⑦ ⑦ ⑦ ⑦
⑦ ⑦
⑦
⑧
⑨ ⑨ ⑩
⑯ ⑪
⑫
⑬
⑭⑮⑰
⑱
⑲
⑳
IN OUT FACILITY ACCESS CONNECTING STRUCTURE
- BUILD AFTER MOBILE UNIT ARRIVES ON SITE -
DECONTAMINATION AREA
CLEAN WORK ROOM
UTILITY ROOM
VESTIBULE VESTIBULE
EMERGENCY
USE ONLY
A connecting structure must only be built AFTER the mobile unit arrives on site. Failure to meet this delivery requirement will result in additional project costs and delays.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Top View
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Passenger Side View
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Driver Side View
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Front View
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Rear View
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Weight Distribution
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
4’ x 8’ x 1” (minimum size) steel road plates are needed to support and distribute weight accordingly.
STERIS can advise if these plates are NOT needed once a formal site visit is performed.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Top View - Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
LIFT GATE
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Driver Side - Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Passenger Side - Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Front and Rear Views - Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT VIEW REAR VIEWVA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Operational Envelope – Weight Distribution – Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
4’ x 8’ x 1” (minimum size) steel road plates are needed to support and distribute weight accordingly.
STERIS can advise if these plates are NOT needed once a formal site visit is performed.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Touch Points and Weight Distribution Images
FRONT UNIT STABILIZERS REAR UNIT STABILIZERSREAR AXLE STABILIZERS
If necessary, and based on individual site conditions, three 4’ x 8’ x 1” (minimum size) steel road plates may be needed.
Plates are not provided by
STERIS.
Total unit weight is approximately
70,000 lbs.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Touch Points and Weight Distribution Images – Support Unit
If necessary, and based on individual site conditions, three 4’ x 8’ x 1” (minimum size) steel road plates may be needed.
Plates are not provided by
STERIS.
Total unit weight is approximately
20,000 lbs.
FRONT UNIT STABILIZERS REAR AXLE
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Anchoring Images and Capacities Unit Anchoring / Tie Downs
• The Mobile Endoscope Reprocessing Unit has 2 tie down rings at each corner and 4 additional per side (16 total) at 9,200 lb. capacity each (6,500 lb horizontal/vertical capacity).
• At minimum, corners must be secured - final tie down configuration is field specific.
• Anchor points to be located approximately 45 degrees from the ground.
• STERIS will provide straps needed to secure the unit at each corner.
• The Support Unit has 3 tie rings on each side to be used as needed.
1. Ground penetrations are the responsibility of the Customer.
2. Tie down requirements are at the discretion of the Authority Having Jurisdiction (AHJ).
Tie Down Rings Located at Each Corner
6,500 lbs
6, lb s
45°
NOTE: Mobile SPD Unit shown for reference only. Mobile Endoscope Reprocessing Unit design will differ from photo.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Anchor Point Locations
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
It is required that anchor tie downs are placed after the Mobile Endoscope
Reprocessing Unit arrives. Although STERIS does not penetrate the ground, Installation Specialists will be on site and can provide guidance regarding tie down locations.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Support Pad Recommendations
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
The Customer is responsible for providing a pad based on ground conditions, unit weight, unit dimensions, and/or local AHJ requirements.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Support Pad Recommendations
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
The Customer is responsible for providing a pad based on ground conditions, unit weight, unit dimensions, and/or local AHJ requirements.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Maximum Pitch and Slope for Delivery
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
The Customer is responsible for providing an appropriate delivery path and ramp to accommodate the mobile unit.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Maximum Pitch and Slope for Deployment
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
The Customer is responsible for providing an appropriate delivery path and ramp to accommodate the mobile unit.
Failure to meet STERIS delivery requirements will result in project delays and additional delivery charges.
FRONT
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Leveling and Threshold Heights
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Assuming level ground, the minimum threshold height is 45” and the maximum height is 54” with dunnage.
Additional height requirements need to be approved by
STERIS.
Dunnage Blocks
FRONT
MOBILE UNIT LEVELING ACCOMODATIONS
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Maximum Pitch and Slope for Deployment – Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
The Customer is responsible for providing an appropriate delivery path and ramp to accommodate the mobile unit.
Failure to meet STERIS delivery requirements will result in project delays and additional delivery charges.
FRONT
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Connecting Corridor / Structure
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Step 3: Mobile Unit Delivery Step 4:
Start Connecting Corridor Construction
A connecting structure must only be built AFTER the mobile unit arrives on site. Failure to meet this delivery requirement will result in additional project costs and delays.
Step 1: STERIS Site Walk
Step 2: Document Sharing (Customer to provide elevation drawings)
Access Requirements
• Facility must provide a connecting corridor or structure for transport of instruments into and out of the Mobile Endoscope Reprocessing Unit. This structure can only be built once the unit is delivered and in place.
• Entry and Exit doors are located on the driver side of the Mobile Endoscope Reprocessing Unit and the Emergency Exit door is located at the rear of the passenger side expandable section.
NOTE: Mobile SPD Unit shown for reference only. Mobile Endoscope Reprocessing Unit design will differ from photo.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Site Preparation Procedure
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
The mobile unit can raise and lower to meet various height challenges, however there are limitations on the minimum and maximum height of the unit.
STERIS requires that deck elevations be sent for approval prior to unit delivery. Customer must not build the connecting structure until AFTER the unit arrives on-site.
The connecting corridor must be built to fit the settled position of the mobile unit to ensure the best possible alignment between these structures. STERIS will not manipulate the unit to align with a preexisting structure.
STERIS takes great care in setting up the mobile unit to ensure that it properly leveled and expanded. The final height and horizontal placement of access doors will shift during this process. Once set in place, it is not practical to further manipulate the mobile unit without risking significant delay.
NOTE: Mobile SPD Unit shown for reference only. Mobile Endoscope Reprocessing Unit design will differ from photo.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Connecting Corridor Protective Bumper
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Customer is responsible for any damage resulting from transfer carts or improper installation of bumpers/rails that fail to protect the mobile unit.
Customer is responsible for ensuring the protection of the exterior of the mobile unit. When installing a connecting corridor or structure, please incorporate a protective bumper, rail, or barrier to prevent the unit from being damaged by transfer carts.
NOTE: Mobile SPD Unit shown for reference only. Mobile Endoscope Reprocessing Unit design will differ from photo.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
WEATHER SEAL
Connecting Corridor and Weather Sealing
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
It is the responsibility of the Customer or contractor to ensure that all connecting corridors are installed with a sealer such as expansion foam. Seals shall be installed to prevent water leaks and insects from entering the transition space between the corridor and the mobile unit.
EXTERIOR WALL OF
MOBILE UNIT
EXTERIOR DOOR OF
MOBILE UNIT
INTERIOR WALL OF
CORRIDOR
Sealing of Access Area
• The connecting corridor must be sealed to the exterior wall of the mobile unit around the access doors.
• STERIS recommends the use of expansion foam or a similar sealer to prevent water leaks and insects from entering the corridor or mobile unit.
Example of Proper
Weather Seal
NOTE: Mobile SPD Unit shown for reference only.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Connecting Corridor Roof Extension
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Damage resulting from construction of the connecting corridor will be the responsibility of the Customer.
The design of the connecting corridor must incorporate a weatherproof seal where the connecting corridor and the mobile unit meet. The structure’s roof should be designed to extend over the interface area for best protection.
Structure Roof Extends Over Mobile Unit for Best Weather Protection
NOTE: Mobile SPD Unit shown for reference only. Mobile Endoscope Reprocessing Unit design will differ from photo.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Customer/contractor is responsible for all utility connections. STERIS Mobile Installation Specialists will be on site to provide guidance as needed.
Water Waste DrainPower Fire
Protection
Utility Connection Drawings and Connection Points
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Utility Connection Points – Top View
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Utility Connection Points – Driver Side
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Utility Connection Points – Passenger Side
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Utility Connection Points – Front and Rear
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Utility Connection Points – Top View - Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Utility Connection Points – Passenger Side View - Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Utility Connection Points – Front View - Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Power Connection Images and Requirements
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Power Requirements
• Electrical Service: 480 VAC, 3-phase, 60 Hz
• Total Load: 350 AMP MOP (262 AMP FLA) with parallel feed through (2) provided 40’ shoreline cables having Meltric DR250 connectors. Facility to provide (2) female mating connectors or purchase from STERIS.
• Configuration: 4 wire (L1, L2, L3, G) NOTE: A Neutral is NOT required.
• Disconnects must be within 30’ of the front of unit.
• A single point ground stud is located adjacent to power connections on the front of the unit.
Disconnects must be within 30’ of the front/nose of the unit.
Male Connection
(2) On Front of UnitParallel Feed Power Connectors on Front of Mobile Unit
480VAC / 3-Phase / 60 Hz Feed 350 AMP MOP (262 AMP FLA) Parallel Feed
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Male Connection
(2) On Front of Unit
Facility/Contractor is responsible for making the final power connection from the disconnects to the mobile unit.
Power connections utilize a 4-wire configuration;
(L1, L2, L3, G).
A Neutral is NOT needed in this configuration.
SUPPLIED BY FACILITY
(2) Required Connectors
MELTRIC DR250 Receptacles 39-24243-K04
Power Connection Images and Requirements
Facility to provide (2) mating connectors or purchase from STERIS.
Connection Requirements
• Shoreline cables having male and female connectors will be provided. The facility must provide disconnect(s) having (2) female mating connectors. The required connectors may be purchased from STERIS.
• Mating Connectors may be mounted to the bottom of the disconnect. If mounted to a vertical surface on the disconnect, an angled mount must be used.
Power Connection Shown w/Angled Mount and Strain Relief on Shoreline Cable
SUPPLIED BY FACILITY
(2) Required if Connectors are Mounted on Vertical Surface
MELTRIC 60° Angle 596M6
(2) 40’ Shoreline Cables Provided*
Female Connector to Unit
Male Connector to Disconnect
* Support Unit Shoreline cable shown for reference. Mobile unit cables may differ from photo.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Power Connection Images and Requirements – Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Power Requirements
• Electrical Service: 480 VAC, 3-phase, 60 Hz
• Total Load: 100 AMP MOP (52 AMP FLA) through (1) provided 40’ shoreline cable with
Meltric DR100 connectors. Facility to provide disconnect with mating female connector.
• Configuration: 4-wire (L1, L2, L3, G) NOTE: A Neutral is NOT required.
• Disconnect within 30’ of the front of unit.
• A single point ground stud is located adjacent to power connection on the front of the unit.
Disconnect must be within 30’ of the front/nose of the unit.
SUPPLIED BY FACILITY
(1) Required Connector
MELTRIC DR100 Receptacle 39-64243-K04
Power connection utilizes a 4-wire configuration; (L1, L2, L3, G). A neutral is NOT needed in this configuration.
480VAC / 3-Phase / 60 Hz Feed
100 AMP MOP (52 AMP FLA)
Facility/Contractor is responsible for making the final power connection from the disconnects to the support unit.
SUPPLIED BY FACILITY
(1) Required if Connector is
Mounted on Vertical Surface MELTRIC 70° Angle 514M7
(1) 40’ Shoreline Cable Provided
Female Connector to Unit
Male Connector to Disconnect
Male Connection
(1) On Front of Unit
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Ground Point Images and Requirements Mobile Endoscope Reprocessing Unit and Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Failure to install the ground rod and cable could result in damage to the mobile unit and included equipment.
Earth grounding points and grounding kits are provided for Mobile Endoscope Reprocessing Unit and Support Unit. Each kit includes:
- (1) Single point ground stud
- (1) 20’ ground cable
- (2) Clamps
(2) Ground rods to be Supplied by Facility
Installation by facility per NEC article 250
Support Unit - Front
Mobile Endoscope Reprocessing Unit - Front
(2) Ground rods are to be supplied by the facility.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Utility Connections Detail and Images
COLD WATERWASTE DRAIN
INSTRUMENT AIR
Connections are located inside the main driver side belly box
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Cold Water Supply Images and Requirements
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
1. Connection sizes indicate the equipment termination sizes only.
2. Size of supply piping is dependent on length of pipe run to ensure adequate supply service pressure and demand flow at equipment terminals.
Cold Water Supply Requirements
• 1-½” FNPT connection inside driver side belly box
• Dynamic pressure 50-80 psi at mobile unit input
• Facility to provide pressure reducer / backflow prevention if required.
• Temperature: 40° F Min, 70° F Max
• 700 GPH average (30 GPM peak)
IMPORTANT: Failure to meet STERIS water quality requirements may result in equipment performance problems or additional charges for equipment maintenance.
Water Manifold (Customer Connection Point) 1-1/2-inch NPT Fresh Water Inlet
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Wastewater Images and Requirements
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Waste flow will vary depending on the frequency of equipment operation in the unit.
It is highly recommended that a P-Trap be installed on the waste line to minimize the potential for foul odors.
Drain can’t be connected to a grease trap or grease interceptor line.
Waste Water Access Point Inside Main Driver Side Belly Box
Wastewater Discharge Requirements
• 4-inch wastewater connection
• 700 GPH average (60 GPM peak flow)
• Wastewater discharge temperature 140° F
• Wastewater line must be facility protected.
• Waste connection can be made through the side of the belly box door or through the floor of the belly box.
• Facility connection to mobile unit should be made using a non-glued connector such as Fernco or similar.
4” Wastewater Connection
(Belly Box Floor Option)
4” Wastewater Connection (Belly Box Door Option)
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Instrument Air Connection The Mobile Endoscope Reprocessing Unit receives instrument air from the Support Unit.
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
The Support Unit must be located within 120’ of the Mobile Endoscope
Reprocessing Unit.
Instrument air lines must be facility protected from freezing and physical damage.
Support Unit: Instrument Air feed connection is located just forward of the passenger side belly box.
(3) 50’ Instrument Air Hoses Provided
Mobile Unit: Instrument Air inlet connection is in the rear compartment of the main driver side belly box.
Support Unit Mobile Unit
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Fire Sprinkler Images and Requirements
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
2-inch Victaulic Coupling
Consider connecting to a nearby water source such as a fire hydrant if allowed by local municipalities.
Always check and follow local laws and codes prior to making this connection.
Fire Sprinkler – Wet System Requirements
• Sprinkler supply: 2” Victaulic coupling (rear of unit)
• External lines must be facility protected.
• Supply required:
o 143 GPM at mobile unit o 46 psi minimum residual pressure at mobile unit
Supply Line Runs Up Through Floor of Belly Box
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Fire Alarm System
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Fire Evacuation Plan
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Aluminum stairs are provided by STERIS for use at the emergency exit door.
If the provided stairs do not meet local code or preference requirements, the
Customer is responsible for providing stairs for the emergency exit door.
EMERGENCY EXIT FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Fire Extinguisher Locations
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
NOTE: Customer is Responsible for Providing Fire Extinguishers ABC Type Dry Chemical - Shown for Reference Only
There are two fire extinguisher brackets installed in the Mobile Endoscope Reprocessing Unit, one at each door.
FRONT
It is the Customer’s responsibility to provide fire extinguishers and for any incoming and routine inspections required by local code or policies.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Fire Alarm Panel Drawing
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Example of fire panel shown, actual fire panel may vary
The Fire Panel is on the rear wall of the Utility Room
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Fire Alarm Images and Requirements
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
The fire alarm system has no capability to repeat or react to external alarms or external PA systems.
Fire Alarm Description
• Conventional analog fire alarm system with local audible/visual horn-strobe alarm devices for occupant warning.
• Unit alarm system configured with:
o 2 zones with detection and pull stations o 1 zone with supervisory tamper switch on sprinkler water control valve o 1 zone for water flow switch on sprinkler system
• System exports trouble, supervisory and alarm conditions to three dry contacts available at the rear of the unit.
Fire Alarm Connection - Customer Connection Point Fire Alarm Remote Dry Contacts
Fire Extinguisher Shown for Reference
Only – See Page 57VA M ed ica l C en ter
-Fay ett
REV A 02/10/2025
Customer Data Connection
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
The Customer data patch panel is located on the rear interior wall of the decontamination zone
IT Patch Panel - TOR Series Part # TOR-4-20RP Interior Connections (shown below)
Customer must extend their IT network to the mobile unit for their IT provided equipment.
If fiber optic cable is required, STERIS coordination for cable routing will be required.
Customer data connection is routed up through a conduit inside the rear passenger side belly box to a patch panel
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Data connection is the responsibility of the customer. STERIS will not connect the mobile unit to the hospital IT network.
Locations and QTY of RJ45 connections may vary based on unit availability.
Data Connections
RJ45 Connections Provided Throughout the Unit
CUSTOMER PATCH
PANEL LOCATION
Wall Mount Ceiling MountRJ45 data connections are available at several locations throughout the unit to support network, voice and
POTS (phone).
FRONT
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
HVAC Condensate Images and Connections Mobile Endoscope Reprocessing Unit and Support Unit
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
If condensate from the HVAC units is not captured, the condensate will drain to ground. HVAC drains must be run with a
Customer installed air gap.
Mobile Endoscope Reprocessing Unit - Front Mobile Endoscope Reprocessing Unit - Rear
Support Unit - Front
HVAC Condensate Drain Terminations: All Drains ¾” Nominal PVC
¾” PVC
Typical
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Ancillary Cooling Package
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
Power Requirements o Electrical Service: 480 VAC, 3-phase, 60hz o (1) 30 AMP breakers
The ancillary cooling package is required in locations where the heat index consistently exceeds 100 degrees.
Air inlets are 6”x24”.
Unit Information STERIS will supply HVAC pre-cooling capability o HVAC Pre-cool unit dimensions and weight:
- Width = 32”
- Length = 48”
- Height = 52”
- Weight = 685 lbs
NOTE: Mobile SPD Unit shown for reference only. Mobile Endoscope Reprocessing Unit design will differ from photo.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
NOTES
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR
WRITTEN PERMISSION OF STERIS CORPORATION.
VA M
ed ica l C en ter
-Fay ett
REV A 02/10/2025
Site Guide Revision Log
NOTICE: THE MATERIAL AND INFORMATION CONTAINED HEREIN IS CONFIDENTIAL AND IS THE PROPERTY OF STERIS CORPORATION AND IS NOT TO BE USED, DISCLOSED, COPIED, TRANSFFERED OR REPRODUCED WITHOUT THE PRIOR