DS-09243-Rev005_Brevera_DataSheet_US.pdf

PDF 2 MB Posted

Attached to
Z1DA--CON Project #667-24-155 Mammography Site Prep (VA-25-00001572)
Solicitation number
36C25624R0098
Issued by

About this file

This document appears to be a data sheet for the Brevera Breast Biopsy System, a medical device manufactured by Hologic. The Brevera system is designed to streamline the breast biopsy process with real-time imaging and automated specimen handling. Key technical specifications are provided, including dimensions, weight, imaging capabilities, and electrical requirements.

The related federal contract opportunity is a solicitation from the Department of Veterans Affairs for a mammography site preparation project. The solicitation number is 36C25624R0098, and the project title is "Z1DA--CON Project #667-24-155 Mammography Site Prep (VA-25-00001572)". No additional details about the specific requirements or timeline for this opportunity are provided in the given information.

View the file

Other files

File Type Posted
Dimensions Site Planning Guide MAN-03169 EN Rev_005 12-2017.pdf PDF
A103 3rd Floor Plan.pdf PDF
1_Revised RECORD DRAWINGS-FULL SET.pdf PDF
18 A1.2 SECOND FLOOR PLAN.pdf PDF
36C25624R0098 RFI Register 08212024.docx DOCX document
36C25624R0098 0006.docx DOCX document
01 00 00 - General Requirements.docx DOCX document
36C25624R0098 0005.docx DOCX document
36C25624R0098 0004.docx DOCX document
36C25624R0098 Mammography Layout-1B .pdf PDF
36C25624R0098 0003.docx DOCX document
667-24-155 Site Visit Sign in Sheet.pdf PDF
36C25624R0098 0002.docx DOCX document
36C25624R0098 0001.docx DOCX document
36C25624R0098 EXHIBIT B.docx DOCX document
36C25624R0098 ICRA Precautions Mammography Site Prep.docx DOCX document
36C25624R0098 Mammography Site Prep - SOW 05292024.docx DOCX document
36C25624R0098 26 05 11 - Requirements for Electrical Installations.docx DOCX document
36C25624R0098 RFP.docx DOCX document
36C25624R0098 EXHIBIT C.docx DOCX document
36C25624R0098 WD LA20240031 Dated 01052024.pdf PDF
36C25624R0098 26 05 33 - Raceway and Boxes for Electrical Systems.docx DOCX document
36C25624R0098 01 00 00 - General Requirements.docx DOCX document
36C25624R0098 23 05 93 - Testing Adjusting and Balancing for HVAC.docx DOCX document
36C25624R0098 EXHIBIT F.docx DOCX document
36C25624R0098 01 33 23 - Shop Drawings-Product Data and Samples .docx DOCX document
36C25624R0098 26 05 19 - Low-Voltage Electrical Power Conductors and Cables.docx DOCX document
36C25624R0098 02 41 00 - Demolition.docx DOCX document
36C25624R0098 ICRA Preliminary.pdf PDF
36C25624R0098 01 45 00 - Quality Control.docx DOCX document
36C25624R0098 EXHIBIT A.docx DOCX document
36C25624R0098 01 35 26 - Safety Requirements.docx DOCX document
36C25624R0098 07 84 00 - Firestopping.docx DOCX document
36C25624R0098 EXHIBIT D.docx DOCX document
36C25624R0098 23 05 11 - Common Work Results for HVAC .docx DOCX document
36C25624R0098 EXHIBIT E.docx DOCX document
Show all 36

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

The Brevera system is designed to streamline the entire breast biopsy process from start to finish – with real-time imaging for instant verification and automated specimen handling.

Explore the technical details here.

Brevera® Breast Biopsy System with CorLumina® Imaging Technology www.hologic.com | BSHInsideSales@Hologic.com | +1.800.442.9892

DS-09243 Rev.005 (8/20) ©2020 Hologic, Inc. All rights reserved. Printed in USA. Specifications are subject to change without prior notice.

Hologic, Brevera, CorLumina, The Science of Sure and associated logos are trademarks and/or registered trademarks of Hologic, Inc. and/ or its subsidiaries, in the United States and/or other countries. All other trademarks and registered trademarks are the property of their respective owners. This information is intended for medical professionals in the U.S. and other markets and is not intended as a product solicitation or promotion where such activities are prohibited. Because Hologic materials are distributed through websites, eBroadcasts and tradeshows, it is not always possible to control where such materials appear. For specific information on what products are available for sale in a particular country, please contact your local Hologic representative or write to womenshealth@hologic.com.

Hologic is a global company using The Science of Sure to improve lives by innovating and bringing to market the best diagnostic, detection and surgical technology in the world. We integrate The Science of Sure into everything we do to help healthcare professionals discover, diagnose and treat their patients with progressive certainty to deliver ever-greater peace of mind.

Increase accuracy.

Real-time imaging delivers a wealth of information at the point of care – so you can make informed clinical decisions with confidence.

Enhance workflow.

An intuitive user interface, real-time imaging, and automated specimen collection and separation work together to save facilities an average of 12 minutes per procedure.1

Improve the patient experience.

Fast, accurate and streamlined procedures mean less time under compression and can result in a more positive and compassionate biopsy experience for you and your patients.

REFERECE: 1. Brevera Pulse Survey, Quantitative report. Inspired Insights. April 2019.

Hologic Headquarters United States / Latin America 250 Campus Drive Marlborough, MA 01752 USA Tel: +1.508.263.2900 Sales: +1.781.999.7453 Fax: +1.781.280.0668 Email: info@hologic.com

Hologic Europe The Corporate Village Da Vincilaan 5 Building Caprese, 8th floor 1935 Zaventem Belgium Tel: +32.2.711.4680

Hologic Asia Pacific One Temasek Avenue 21-04 Millenia Tower Singapore 03192 Tel: +65 6206 5833

Hologic Australia Suite 302, Level 3 2 Lyon Park Road Macquarie Park NSW 2113 Australia Tel: +61.2.9888.8000 Fax: +61.2.9870.7555

Technical Specifications

Min 53.25 in (135.3 cm) Max 65.25 in (165.7 cm)

22.75 in (57.8 cm)

22.75 in (57.8 cm)

The Brevera System Transforming breast biopsy in real time.

Dedicated Technologist Screen

Intuitive Interface

Real-time Core Imaging

Automated Core Handling

System-guided Setup

Flexible Dual Monitors

The Brevera Biopsy Device The control is in your hands.

Physician-controlled Sampling

Disposable and Reusable Design

Integrated Site Marking

Variable Aperture

Biopsy Device Components

Reusable driver dimensions 7. 8 L x 2.9 W x 1.7 in H (19.8 x 7.4 x 4.3 cm)

Reusable driver weight 17 oz (500g)

Disposable needle specifications 14 L x 2 W x 1.25 in H (35.6 x 5.1 x 3.2 cm)

Disposable needle weight with tubing set 9.5 oz (269.3g)

Tissue filter diameter 2.4 in (6.1 cm)

Acquisition Workstation/Specimen Imaging Display

CPU Multi-core Intel

Memory 4GB RAM

Solid state disk 500 GB (approx. 10,000 patient files)

Storage media External USB flash drive

Display monitor 15 in (38.1 cm) WXGA (1366 x 768) color LCD display, touchscreen

Operating system Windows 10 loT Enterprise 2019

LTSC

Network connection 10, 100, 1000-Base-T; wireless

802.11 b/g/n/ac

DICOM services Worklist, print, storage

User interface On screen keyboard

Visualization enhancement 5 levels of image sharpening

Magnification Continuous image zoom .075X-20X

Annotations Circle

Console Control Display

Display monitor 7 in (18 cm) LED display

Operating system` Embedded RTOS

X-ray Source

Energy range 20-35 kV

Anode type Tungsten

Tube current 1 mA

Exposure 10 mAs maximum

Focal spot size 50 μm

Filtration 125 μm beryllium

Pixel size and output image 20 μm 12bit

Active imaging area 3.3 x 2.5 cm

Cabinet

X-ray features Drawer interlock, key switch, fully shielded

Indicators Power, door closed, ready, X-ray on

Console

Console weight 200 lbs (91kg)

Voltage range Brevera 100 100 –120 V

Voltage range Brevera 200 220 –240 V

Operating frequency 50Hz–60Hz

Peak line current 9 Amps peak, 4 Amps typical

Power consumption 800W

Power cord length 20 ft (6.1 m)

Vacuum generated 26.5 in Hg (66 cm)

Cycle time 8 seconds for acquisition, imaging and verification