This document is a Site Planning and Pre-Installation Guide for the Selenia Dimensions Digital Mammography System and Tomosynthesis System. It provides detailed specifications and requirements for preparing the site to properly install and operate this medical equipment.
The key details include:
- Description of the Tubestand, Acquisition Workstations, and other system components with measurements and weight requirements
- Electrical input specifications for the Tubestand and Acquisition Workstations
- Environmental and operating requirements for temperature, humidity, and storage
- Clearance, cabling, wireway, and mounting requirements for the system components
- Radiation shielding and regulatory compliance information
This guide is intended to assist the Installation Coordinator in planning and preparing the site for the proper installation of the Selenia Dimensions system. It does not appear to be a solicitation or contract document, but rather an informational technical reference.
View the file
Other files
Show all 36
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Selenia® Dimensions® Digital Mammography System
Digital Tomosynthesis System
Site Planning and Pre-Installation Guide
Part Number MAN-03169 Revision 005 December 2017
© 2012–2017 Hologic, Inc. Printed in the USA. This manual was originally written in English.
Hologic, 3Dimensions, Dimensions, Selenia, Affirm, FAST Paddle, SecurView, SecurXchange, and associated logos are trademarks and/or registered trademarks of Hologic, Inc., and/or its subsidiaries in the United States and/or other countries. All other trademarks, registered trademarks, and product names are the property of their respective owners.
This product may be protected by one or more U.S. or foreign patents as identified at www.Hologic.com/patents
Technical Support
USA: +1.877.371.4372
Europe: +32 2 711 4690 Asia: +852 37487700 Australia: +1 800 264 073 All Other: +1 781 999 7750
Selenia Dimensions System Site Planning and Pre-Installation Guide Table of Contents
MAN-03169 Revision 005 v
List of Figures ________________________________________________________________ vii
List of Tables __________________________________________________________________ ix
1: Introduction and General Information ___________________________________________1
1.1 Introduction
1.2 System Description
1.2.1 Tubestand Overview
1.2.2 Acquisition Workstation Overview
1.3 System Location Considerations
1.4 Safety
1.4.1 Isolation Integrity
1.4.2 Shielding
1.4.3 Remote X-ray On/Power On Indicators
1.4.4 The X-ray Interlock
1.5 Compliance
1.5.1 Compliance Requirements
1.5.2 Compliance Statements
1.5.3 Electromagnetic Compatibility
2: Specifications and Reference Information ______________________________________15
2.1 Product Measurements
2.1.1 Tubestand (Gantry with C-Arm)
2.1.2 Universal Acquisition Workstation
2.1.3 Premium Acquisition Workstation
2.1.4 Standard Acquisition Workstation
2.2 Electrical Input
2.2.1 Tubestand
2.2.2 Acquisition Workstation
2.3 Operation and Storage Environment
2.3.1 General Conditions for Operation
2.3.2 Storage Environment
2.4 Center of Gravity Reference
3: Site Planning Checklist _______________________________________________________25
4: Room Layout Worksheet ______________________________________________________44
MAN-03169 Revision 005 vii
List of Figures
Figure 1: Tubestand for Selenia® Dimensions® System Figure 2: Acquisition Workstations Figure 3: Tubestand (Gantry with C-arm) Measurements Figure 4: Universal Acquisition Workstation Measurements Figure 5: Premium Acquisition Workstation Measurements Figure 6: Standard Acquisition Workstation Measurements Figure 7: Gantry Center-of-Gravity Reference Figure 8: Premium Acquisition Workstation Center-of-Gravity Reference Figure 9: Standard Acquisition Workstation Center-of-Gravity Reference Figure 10: Universal Acquisition Workstation Center-of-Gravity Reference
MAN-03169 Revision 005 ix
List of Tables
Table 1: Electronic Emissions Table 2: Electromagnetic Immunity Part 1 Table 3: Electromagnetic Immunity Part 2 Table 4: Separation Distances for RF Equipment Table 5: Room Requirements Table 6: Clearance Requirements Table 7: Power Requirements Table 8: Environmental Requirements Table 9: Cable Requirements Table 10: Wireway and Threshold Requirements Table 11: X-Ray Shielding Requirements Table 12: Mounting Requirements
Chapter 1: Introduction and General Information
MAN-03169 Revision 005 Page 1
1: Introduction and General Information
1.1 Introduction
This guide, an aid for the Installation Coordinator responsible for site planning and preparation, contains all product information, specifications, and directions necessary for determining the installation requirements. All information in this guide is important and relevant to the planning process.
Note Refer to MAN-03170 for Mobile applications.
The mounting diagrams provided in this manual are recommendations only; the final responsibility for proper installation belongs to the Installation Coordinator.
Ensure all installations meet local regulations. A licensed electrician must perform the necessary electrical service.
Chapter 1
Page 2 MAN-03169 Revision 005
1.2 System Description
1.2.1 Tubestand Overview
Figure 1: Tubestand for Selenia® Dimensions® System
Figure Legend
1. Tubestand (Gantry and C-arm)
2. Gantry
3. C-arm (Tube Arm and Compression Arm)
4. Tube Arm
5. Compression Arm
MAN-03169 Revision 005 Page 3
1.2.2 Acquisition Workstation Overview
Figure 2: Acquisition Workstations
1. Universal Acquisition Workstation
2. Premium Acquisition Workstation
3. Mobile Acquisition Workstation
4. Standard Acquisition Workstation
Page 4 MAN-03169 Revision 005
1.3 System Location Considerations
The Site Planning Checklist on page 25 is set up to assist you with the following topics when selecting a location for the system:
1. System component sizes and weight
• Flooring type for mounting and weight considerations
• Doorway clearance
• Installation space
• Relocation of equipment
2. Room size
• Movement clearance—Allocate space for patient and technologist movement.
Avoid obstructions in the room that hinder access to the unit controls or the patient.
• Storage—Provide convenient storage for system accessories. If it is not possible to store accessories within the exam room, arrange for safe storage close by.
3. Location for patient throughput
4. Power source requirements
• Interlocks (room, door, lights, and so on)
• Service access
5. Networking requirements (DICOM, PACS, and so on)
6. Physical and environmental requirements
7. Shielding requirements
8. Cabling and wireways
MAN-03169 Revision 005 Page 5
1.4 Safety
1.4.1 Isolation Integrity
WARNING!
To correctly isolate the system, attach only approved accessories or options to the system. Only approved personnel can change the connections.
WARNING!
Keep a 1.5 meter safe distance between the patient and any non-patient devices.
Do not install non-patient system components (like the Workflow Manager, a diagnostic review workstation, or a hard copy printer) in the Patient Area.
1.4.2 Shielding
Structural Shielding A Medical Physicist should review the room walls in which this system is used to ensure that the room meets local guidelines for radiation shielding. Refer to the table X-ray Shielding Requirements in Chapter 3 of this document.
The Acquisition Workstation (AWS) Shield The operator radiation shield attached to the Acquisition Workstation, provides a minimum of 0.5 mm of lead equivalent attenuation.
1.4.3 Remote X-ray On/Power On Indicators
The system has provisions for remote lights which indicate when the system is On and when x rays are being taken. These lights are usually installed outside the exam room, above the door. Have a certified electrician install these lights.
The relay contacts are rated:
• 10 A, 250 VAC (normally open)
• 10 A, 30 VDC (normally open)
Page 6 MAN-03169 Revision 005
1.4.4 The X-ray Interlock
The X-ray Interlock connector at the rear of the Gantry provides a normally closed (NC) contact (5V 10mA) for the Acquisition Workstation. When an external interlock is used (for example a door or shield switch), a contact opens and an exposure cannot take place.
A switch closure on the contact must occur for an exposure to occur.
If the X-ray Interlock is utilized, the state of the interlock contacts should be displayed by visual means to the operator per IEC 60601-2-45:2011 (3rd Edition), Clause 203.6.2.1.101.
MAN-03169 Revision 005 Page 7
1.5 Compliance
This section describes the mammography system compliance requirements and the responsibilities of the manufacturer.
1.5.1 Compliance Requirements
The manufacturer has the responsibility for the safety, reliability, and performance of this equipment with the following provisions:
• The electrical installation of the room meets all requirements.
• The equipment is used according to the User Guide.
• The assembly operations, extensions, adjustments, changes, or repairs are performed only by authorized persons.
• The network and communication equipment is installed to meet IEC Standards. The complete system (network and communications equipment and the mammography system) must be in compliance with IEC 60601-1 and IEC 60601-1-1.
Caution:
Medical Electrical Equipment needs special precautions about EMC and must be installed, put into service and used according to the EMC information provided.
Portable and mobile RF communications can affect medical electrical equipment.
The use of unauthorized accessories and cables can result in increased emissions or decreased immunity. To keep the isolation quality for the system, attach only approved Hologic accessories or options to the system.
The Medical Electrical (ME) Equipment or ME System should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, make sure that the ME Equipment or ME System operates correctly in this configuration.
This system is intended for use by healthcare professionals only. This system may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the equipment or shielding the location.
Page 8 MAN-03169 Revision 005
Changes or modifications not expressly approved by Hologic could void your authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
1.5.2 Compliance Statements
The manufacturer states this device is made to meet the following requirements:
• CAN/CSA - ISO 13485-03 Medical Devices – Quality Management Systems – Requirements for Regulatory Purposes (Adopted ISO 13485:2003 second edition, 2003-07-15)
• CAN/CSA C22.2 NO. 60601-1-08 Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance (Adopted IEC 60601-1:2005, third edition, 2005-12), includes Corrigendum 1:2011; also CAN/CSA C22.2 NO.
601.1-M90 (R2005) Medical Electrical Equipment Part 1: General Requirements for Safety
• EN 60601-1:2006 Medical Electrical Equipment. General Requirements for Basic Safety and Essential Performance; also EN 60601-1:1990 +A1+A11+A12+A2+A13 Medical Electrical Equipment—General Requirements for Safety
• ETSI EN 300 330-1: V1.3.1, and ETSI EN 300 330-2: V1.5.1: 2006—Electromagnetic compatibility and Radio spectrum Matters (ERM); Short Range Devices (SRD); Radio equipment in the frequency range 9 kHz to 25 MHz and inductive loop systems in the frequency range 9 kHz to 30 MHz
• ETSI EN 301 489-1: V1.6.1, and ETSI EN 301 489-3: V1.8.1: 2008—Electromagnetic compatibility and Radio spectrum Matters (ERM); ElectroMagnetic Compatibility (EMC) standard for radio equipment and services
• FCC, 47 CFR Part 15, Subpart C, Section 15.225: 2009
• FDA, 21 CFR [Parts 820, 900 and 1020]
• IEC 60601-1 Ed. 3.0:2005 Medical Electrical Equipment – Part 1: General
Requirements for Basic Safety and Essential Performance; also IEC 60601-1 Ed.
2.0:1988 +A1+A2:1995 Medical Electrical Equipment—General Requirements for Safety
MAN-03169 Revision 005 Page 9
• IEC 60601-1-1Ed. 2.0:2000 Medical Electrical Equipment - Part 1-1: General Requirements for Safety - Collateral Standard: Safety Requirements for Medical Electrical Systems
• IEC 60601-1-2 Ed. 3.0:2007 Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard:
Electromagnetic Compatibility – Requirements and Tests
• IEC 60601-1-3 Ed. 2.0:2008 Medical Electrical Equipment – Part 1-3: General Requirements for Basic Safety and Essential Performance –Collateral Standard:
Radiation Protection in Diagnostic X-Ray Equipment; also IEC 60601-1-3 Ed. 1.0:1994 Medical Electrical Equipment – Part 1: General Requirement for Safety -3. Collateral Standard: Requirements for Radiation Protection in Diagnostic X-ray Equipment
• IEC 60601-1-4 Ed. 1.1:2000 Medical Electrical Equipment – Part 1-4: General Requirements for Safety - Collateral Standard: Programmable Electrical Medical Systems
• IEC 60601-2-28 Ed. 2.0:2010 Medical Electrical Equipment - Part 2-28: Particular Requirements for the Basic Safety and Essential Performance of X-ray Tube Assemblies for Medical Devices; also IEC 60601-2-28 Ed. 1.0:1993 Medical Electrical Equipment – Part 2: Particular Requirements for the Safety of X-Ray Source Assemblies and X-ray Tube Assemblies for Medical Diagnosis
• IEC 60601-2-32 Ed. 1.0:1994 Medical Electrical Equipment – Part 2: Particular Requirements for the Safety of Associated Equipment of X-ray Equipment
• IEC 60601-2-45 Ed. 3.0:2011 Medical Electrical Equipment – Part 2-45: Particular Requirements for Basic Safety and Essential Performance of Mammographic X-Ray Equipment and Mammographic Stereotactic Devices; also IEC 60601-2-45 Ed.
2.0:2001 Medical Electrical Equipment Part 2-45: Particular Requirements for the Safety of Mammographic X-ray Equipment and Mammographic Stereotactic Devices
• RSS-210: Issue 7, 2007 Radio Standards Specification Low-power License-exempt Radiocommunication Devices: Category I Equipment
• ANSI/AAMI ES60601-1:2005 (IEC 60601-1:2005, MOD) Medical Electrical Equipment, Part 1: General Requirements for Basic Safety and Essential Performance, includes amendment (2010); also UL 60601-1 1st Edition: Medical Electrical Equipment, Part 1—General Requirements for Safety
Page 10 MAN-03169 Revision 005
1.5.3 Electromagnetic Compatibility
This section provides information about the electromagnetic compatibility of system per
IEC 60601-1-2.
Table 1: Electronic Emissions Electromagnetic Emissions
The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A Meets Class A Compliance.
The system is suitable for use in all establishments other than domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies
Table 2: Electromagnetic Immunity Part 1 Electromagnetic Immunity – Part 1
The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV Contact ±8 kV Air
±6 kV Contact ±8 kV Air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/ burst
IEC 61000-4-4
±2 kV for power supply lines ±1 kV for input/output lines
±2 kV for power supply lines ±1 kV for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital environment.
MAN-03169 Revision 005 Page 11
Table 2: Electromagnetic Immunity Part 1 Electromagnetic Immunity – Part 1
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
<5% Ut (>95 % dip in Ut) for 0,5 cycle
40% Ut (60 % dip in Ut) for 5 cycles
70% Ut (30 % dip in Ut) for 25 cycles
<5% Ut (>95 % dip in Ut) for 5 s
<5% Ut (>95 % dip in Ut) for 0,5 cycle
40% Ut (60 % dip in Ut) for 5 cycles
70% Ut (30 % dip in Ut) for 25 cycles
<5% Ut (>95 % dip in Ut) for 5 s
Mains power quality should be that of a typical commercial or hospital environment. If the user of the system requires continued operation during power mains interruptions, it is recommended that the system be powered from an uninterruptible power supply or battery.
Power Frequency (50/60Hz) Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at l4evels characteristic of a typical location in a typical commercial or hospital environment.
NOTE Ut is the a.c. mains voltage prior to application of the test level.
Page 12 MAN-03169 Revision 005
Table 3: Electromagnetic Immunity Part 2
The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.
Immunity Test EN/IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 kHz to 80MHz
3 V/m 80 MHz to 2.5 GHz
[V1] = 3 V
[E1] = 3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of the system, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
where P is the maximum output power rating of the transmitter in watts (W) and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya, should be less than the compliance level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
MAN-03169 Revision 005 Page 13
Table 3: Electromagnetic Immunity Part 2
The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.
Immunity Test EN/IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the system is used exceeds the applicable RF compliance level above, the system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the system.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 4: Separation Distances for RF Equipment
Recommended Separation Distances for Portable and Mobile RF Communications Equipment and the system
The system is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the system as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.7 3.7 7.38
100 11.7 11.7 23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Chapter 2: Specifications and Reference Information
MAN-03169 Revision 005 Page 15
2: Specifications and Reference Information
2.1 Product Measurements
2.1.1 Tubestand (Gantry with C-Arm)
Figure 3: Tubestand (Gantry with C-arm) Measurements
A. Height 223 cm (87.8 inches)
B. Width 66 cm (26 inches)
C. Width 173 cm (68.0 inches)
D. Depth 138 cm (54.3 inches)
Weight Maximum of 400 kg (882 pounds)
Chapter 2
Page 16 MAN-03169 Revision 005
2.1.2 Universal Acquisition Workstation
Figure 4: Universal Acquisition Workstation Measurements
A. Width (maximum) with optional articulated display arm extended
136 cm (53.4 inches) - series I UAWS
128 cm (50.3 inches) - series II UAWS
Width (maximum) with standard display arm
94.0 cm (36.9 inches) - series I UAWS
107 cm (42.0 inches) - series II UAWS
B. Depth (maximum) with keyboard tray extended and optional articulated display monitor arm
122 cm (48.4 inches) - series I UAWS, rotated to the side
115 cm (45.1 inches) - series II UAWS, rotated to the side
Depth (maximum) with keyboard tray extended and standard display arm
83.6 cm (32.9 inches) - series I and II UAWS
C. Height (nominal) 219 cm (86.1 inches) after August 2017
204 cm (80.3 inches) before September 2017
Weight (maximum) 209 kg (460 pounds)
MAN-03169 Revision 005 Page 17
2.1.3 Premium Acquisition Workstation
Figure 5: Premium Acquisition Workstation Measurements
A. Height 202 cm (79.8 inches)
B. Width 92.7 cm (36.5 inches)
C. Depth 58.5 cm (23.0 inches)
Weight 154 kg (340 pounds)
Page 18 MAN-03169 Revision 005
2.1.4 Standard Acquisition Workstation
Figure 6: Standard Acquisition Workstation Measurements
A. Height 192 cm (75.3 inches)
B. Width 107 cm (42.1 inches)
C. Depth 76.2 cm (30.0 inches)
Weight 219 kg (462 pounds)
2.2 Electrical Input
2.2.1 Tubestand
Mains Voltage 200/208/220/230/ 240 VAC ±10%
Mains Impedance Maximum line impedance not to exceed 0.20 ohms for 208/220/230/240 VAC, 0.16 ohms for 200 VAC
Mains Frequency 50/60 Hz ±5%
Average Current over 24 Hours < 5 A
Peak Line Current 4 A (65 A maximum for < 5 seconds)
MAN-03169 Revision 005 Page 19
2.2.2 Acquisition Workstation
Mains Voltage 100/120/200/208/220/230/240 VAC ±10%
Mains Frequency 50/60 Hz ±5%
Power Consumption < 1000 watts
Duty Cycle (Standard Acquisition Workstation)
10% ~ 6 minutes per hour or 2 minutes on, 18 minutes off
Overcurrent Protection 8A
2.3 Operation and Storage Environment
2.3.1 General Conditions for Operation
Temperature Range 20 °C (68 °F) to 30 °C (86 °F)
Relative Humidity Range 20% to 80% without condensing moisture
2.3.2 Storage Environment
Gantry Temperature Range –10 °C (14 °F) to 40 °C (104 °F)
Relative Humidity Range 0% to 95% without condensing moisture
(Put in a package for storage in a building.)
X-ray Detector Temperature Range 10 °C (50 °F)to 30 °C (86°F) indefinitely
10 °C (50 °F) to 35°C (95 °F) for a maximum of 12 hours
Maximum rate of temperature change Less than 10 °C (50 °F) per hour
Relative Humidity Range 10% to 80% without condensing moisture
Acquisition Workstation Temperature Range –10 °C (14 °F) to 40 °C (104 °F)
Relative Humidity Range 0% to 95% without condensing moisture
Page 20 MAN-03169 Revision 005
2.4 Center of Gravity Reference
Figure 7: Gantry Center-of-Gravity Reference
A. 31.8 cm (12.5 inches)
B. 95.5 cm (37.6 inches)
MAN-03169 Revision 005 Page 21
Figure 8: Premium Acquisition Workstation Center-of-Gravity
Reference
A. 62.9 cm (24.8 inches)
B. 10.5 cm (4.13 inches)
Page 22 MAN-03169 Revision 005
Figure 9: Standard Acquisition Workstation Center-of-Gravity Reference
A. 64.3 cm (25.3 inches)
B. 11.4 cm (4.5 inches)
C. 18.8 cm (7.4 inches)
MAN-03169 Revision 005 Page 23
Figure 10: Universal Acquisition Workstation Center-of-Gravity Reference
A. 61.7 cm (24.3 inches)
B. 29.0 cm (11.4 inches)
C. 14.2 cm (5.6 inches)
Chapter 3: Site Planning Checklist
MAN-03169 Revision 005 Page 25
3: Site Planning Checklist
Chapter 3
Page 26 MAN-03169 Revision 005
Table 5: Room Requirements
Item Requirements Actual Corrective Action Done
(Init/Date)
Room Size
Length 365 cm (12 feet)
Width 275 cm (9 feet)
Ceiling 244 cm (8 feet)
Door 214 cm x 92 cm (84 in. x 36 in.)
Notes:
The exam room layout should be pre-planned before the arrival of the system. Refer to Room Layout Worksheet on page
44. A typical room layout is shown here. Check specific local or hospital requirements for additional data. Check the route from the loading dock to the room.
Maintain a clearance of 50 cm between the tube head at any position of rotation and any stationary object (excluding the floor). If this clearance cannot be maintained due to space constraints, provide as much clearance as possible. The standard operation of the controls for C-arm rotation should not be altered at installations that do not meet the 50 cm clearance.
1. Universal Acquisition Workstation
2. Tubestand/Gantry
3. Minimum clearance of 50 cm (20 in.)
4. Consult local regulations for minimum clearance
5. Minimum doorway opening of 92 cm (36 in.)
6. Patient Area
MAN-03169 Revision 005 Page 27
Additional Notes
Note # Topic Notes
Page 28 MAN-03169 Revision 005
Table 6: Clearance Requirements
Item Minimum Requirements Actual Corrective Action Done (Init/Date)
Access Clearance
Acquisition Workstation
61 cm (2.0 feet) on 3 sides
Gantry Rear 30 cm (1 feet)
Gantry Front 76 cm (2.5 feet)
Work Area 76 x 76 cm diameter (2.5 feet x 2.5 feet)
Working Clearance
Gantry with C- Arm
50 cm (20 inches) each side of C-Arm (for complete C-Arm rotation)
Work and Patient Area
76.2 cm (30 inches) diameter (minimum) between Tubestand and Acquisition
Note: Refer to the specifications.
MAN-03169 Revision 005 Page 29
Page 30 MAN-03169 Revision 005
Table 7: Power Requirements
Done
(Init/Date)
Power Requirements
Dedicated Power Line
Dedicated disconnect box with lockout capability, circuit breaker or fuse (per local codes), and remote emergency OFF trip switch if required (the location is determined by local codes).
Acquisition
Refer to Electrical Input on page 18.
Light Indicator Relay Contact Ratings*
10 A, 250 VAC (N.O.)
10 A, 30 VDC (N.O.)
*There are provisions in the system to accommodate local regulations that require an X-ray System Power- On, and X-ray On Indicators at the door. These lights are normally installed above the door to the exam room.
MAN-03169 Revision 005 Page 31
Page 32 MAN-03169 Revision 005
Table 8: Environmental Requirements
Done
(Init/Date)
Environmental Requirements: Operating
Temperature Range
20° C to 30° C (68° F to 86°F)
Relative Humidity Range
20% to 80% non condensing
MAN-03169 Revision 005 Page 33
Page 34 MAN-03169 Revision 005
Table 9: Cable Requirements
Item Recommendations Actual Corrective Action Done
(Init/Date)
Cables (Length)
Gantry Power Conduit
7.5 m (25 feet) If Mains box is over 15 m (50 feet) away, custom lengths must drop 2 gauges when length is doubled.
Optional, purchased separately
(Optional)
Workstation to Gantry Interconnect Cable
12.2 m (40 feet)
Ground Wire 12.2 m (40 feet)
Workstation to Gantry Interconnect (Fiber Optic Cable)
13 m (42.65 feet)
Power Cord
3.05 m (10 feet)
If flex cable is used, do not permanently affix to wall, floor, or place in wireway that is permanently attached to the structure. Conduit is recommended with use of NEC wiring.
Power cables must not be enclosed in wireways or permanently affixed to the building structure.
Hologic does not provide the Gantry power cable for some countries. If the power cable is not provided, the installed cable must meet the following requirements and all local codes that apply: 3 conductor, 8 AWG (10 mm2) copper not more than 25 feet (7.62 meters) in length.
MAN-03169 Revision 005 Page 35
Page 36 MAN-03169 Revision 005
Table 10: Wireway and Threshold Requirements
Done
(Init/Date)
Interconnects
Wireways
Horizontal surface wireway, 1-3/4 inches x 5-1/4 inches divided with removable covers.
Vertical surface flush-mounted wireway, 1-3/4 inches x 5-3/4 inches divided with removable covers.
Horizontal surface mounted floor duct wireway, 1-3/4 inches x 5-1/4 inches divided with removable covers, grommeted on open end.
Thresholds (if required)
Length depends on cable run.
MAN-03169 Revision 005 Page 37
Page 38 MAN-03169 Revision 005
Table 11: X-Ray Shielding Requirements
Done
(Init/Date)
X-ray Shielding
Room Per state and local codes based on Medical Physicist’s test results.
Shielding refers to the protection from leakage radiation that usually occurs during mammography exams.
Leakage Radiation All related x ray safety features meet 21 CFR, Chapter I, 1020.31, IEC 60601-1-3, and IEC 60601-2-45.
Typical values of x ray leakage shown are at a distance of 1 meter from the tube focal spot, at maximum kVp, and adjusted for a maximum in use duty cycle of 0.027.
Duty cycle is derived from a maximum throughput of 12 patients per hour x 4 exposures, or 48 exposures per hour with a maximum exposure time of 2 seconds.
Therefore; 48x2 = 96 exposure seconds per hour, or 96 exposure seconds/3600 seconds per hour = 0.027.
The meter measures x ray leakage in units of mR/Hour.
The meter indication is multiplied by the duty cycle to get the correct leakage value (N x 0.027) and this is compared with the maximum allowed criteria 90 mR/Hour.
MAN-03169 Revision 005 Page 39
Page 40 MAN-03169 Revision 005
Table 12: Mounting Requirements
Done
(Init/Date) Mounting Requirements
Gantry (The Gantry base has four 11/16 inch mounting holes.)
Anchors embedded 2- 1/2 inches minimum and appropriate bolts/hardware. The floor must be able to support a minimum of 210 lb./ft2.
(The AWS base has four (two additional holes are available, if needed) 5/8 inch mounting holes.)
Anchors embedded 2- 1/2 inches minimum and appropriate bolts/hardware. The floor must be able to support a minimum of 210 lb./ft2.
Seismic (Refer to 'Center-of- Gravity' section in Chapter 2.)
(Seismic installations are dependent on local regulations.)
Strongly recommended that the customer consults with a professional structural engineer familiar with seismic requirements.
Notes: The mounting diagrams provided in this document are recommendations only; the final responsibility for proper installation belongs to the site field engineer. Ensure all installations meet local regulations.
MAN-03169 Revision 005 Page 41
Page 42 MAN-03169 Revision 005
MAN-03169 Revision 005 Page 43
Room Planning Templates
Chapter 3: Room Layout Worksheet
Page 44 MAN-03169 Revision 005
4: Room Layout Worksheet These templates can be used to establish a functional room layout.
Instructions: Make copies of the grid and the system component cutouts. Outline the designated room size (to scale) on the grid, and then cut out the system components from the copy. Position the component cutouts on the grid to layout the desired work space.
Allow sufficient and convenient storage for accessories. Be sure to include access clearances for service personnel. Avoid areas that may hinder access to the equipment and patient. Also avoid equipment positioning near heat ducts or air-conditioning vents.
To protect from radiation exposure, locate the Acquisition Workstation so the radiation shield provides complete protection for the operator. If possible, include extra space and power outlets for future expansion of clinical services.
Chapter 3: Room Layout Worksheet
MAN-03169 Revision 005 Page 45
| Cover |
| Title Page |
| Copyright Page |
| Table of Contents |
| List of Figures |
| List of Tables |
| 1: Introduction and General Information |
| 1.1 Introduction |
| 1.2 System Description |
| 1.2.1 Tubestand Overview |
| 1.2.2 Acquisition Workstation Overview |
| 1.3 System Location Considerations |
| 1.4 Safety |
| 1.4.1 Isolation Integrity |
| 1.4.2 Shielding |
| Structural Shielding |
| The Acquisition Workstation (AWS) Shield |
| 1.4.3 Remote X-ray On/Power On Indicators |
| 1.4.4 The X-ray Interlock |
| 1.5 Compliance |
| 1.5.1 Compliance Requirements |
| 1.5.2 Compliance Statements |
| 1.5.3 Electromagnetic Compatibility |
| 2: Specifications and Reference Information |
| 2.1 Product Measurements |
| 2.1.1 Tubestand (Gantry with C-Arm) |
| 2.1.2 Universal Acquisition Workstation |
| 2.1.3 Premium Acquisition Workstation |
| 2.1.4 Standard Acquisition Workstation |
| 2.2 Electrical Input |
| 2.2.1 Tubestand |
| 2.2.2 Acquisition Workstation |
| 2.3 Operation and Storage Environment |
| 2.3.1 General Conditions for Operation |
| 2.3.2 Storage Environment |
| Gantry |
| X-ray Detector |
| Acquisition Workstation |
2.4 Center of Gravity Reference
| 3: Site Planning Checklist |
| 4: Room Layout Worksheet |
/ASCII85EncodePages false /AllowTransparency false /AutoPositionEPSFiles true /AutoRotatePages /None /Binding /Left /CalGrayProfile (Dot Gain 20%) /CalRGBProfile (sRGB IEC61966-2.1) /CalCMYKProfile (U.S. Web Coated \050SWOP\051 v2) /sRGBProfile (sRGB IEC61966-2.1) /CannotEmbedFontPolicy /Error /CompatibilityLevel 1.4 /CompressObjects /Tags /CompressPages true /ConvertImagesToIndexed true /PassThroughJPEGImages true /CreateJobTicket false /DefaultRenderingIntent /Default /DetectBlends true /DetectCurves 0.0000 /ColorConversionStrategy /CMYK /DoThumbnails false /EmbedAllFonts true /EmbedOpenType false /ParseICCProfilesInComments true /EmbedJobOptions true /DSCReportingLevel 0 /EmitDSCWarnings false /EndPage -1 /ImageMemory 1048576 /LockDistillerParams false /MaxSubsetPct 100 /Optimize true
/OPM 1
/ParseDSCComments true /ParseDSCCommentsForDocInfo true /PreserveCopyPage true /PreserveDICMYKValues true /PreserveEPSInfo true /PreserveFlatness true /PreserveHalftoneInfo false /PreserveOPIComments true /PreserveOverprintSettings true /StartPage 1 /SubsetFonts true /TransferFunctionInfo /Apply /UCRandBGInfo /Preserve /UsePrologue false /ColorSettingsFile () /AlwaysEmbed [ true /NeverEmbed [ true /AntiAliasColorImages false /CropColorImages true /ColorImageMinResolution 300 /ColorImageMinResolutionPolicy /OK /DownsampleColorImages true /ColorImageDownsampleType /Bicubic /ColorImageResolution 300 /ColorImageDepth -1 /ColorImageMinDownsampleDepth 1 /ColorImageDownsampleThreshold 1.50000 /EncodeColorImages true /ColorImageFilter /DCTEncode /AutoFilterColorImages true /ColorImageAutoFilterStrategy /JPEG /ColorACSImageDict << /QFactor 0.15 /HSamples [1 1 1 1] /VSamples [1 1 1 1] /ColorImageDict << /QFactor 0.15 /HSamples [1 1 1 1] /VSamples [1 1 1 1] /JPEG2000ColorACSImageDict << /TileWidth 256 /TileHeight 256 /Quality 30 /JPEG2000ColorImageDict << /TileWidth 256 /TileHeight 256 /Quality 30 /AntiAliasGrayImages false /CropGrayImages true /GrayImageMinResolution 300 /GrayImageMinResolutionPolicy /OK /DownsampleGrayImages true /GrayImageDownsampleType /Bicubic /GrayImageResolution 300 /GrayImageDepth -1 /GrayImageMinDownsampleDepth 2 /GrayImageDownsampleThreshold 1.50000 /EncodeGrayImages true /GrayImageFilter /DCTEncode /AutoFilterGrayImages true /GrayImageAutoFilterStrategy /JPEG /GrayACSImageDict << /QFactor 0.15 /HSamples [1 1 1 1] /VSamples [1 1 1 1] /GrayImageDict << /QFactor 0.15 /HSamples [1 1 1 1] /VSamples [1 1 1 1] /JPEG2000GrayACSImageDict << /TileWidth 256 /TileHeight 256 /Quality 30 /JPEG2000GrayImageDict << /TileWidth 256 /TileHeight 256 /Quality 30 /AntiAliasMonoImages false /CropMonoImages true /MonoImageMinResolution 1200 /MonoImageMinResolutionPolicy /OK /DownsampleMonoImages true /MonoImageDownsampleType /Bicubic /MonoImageResolution 1200 /MonoImageDepth -1 /MonoImageDownsampleThreshold 1.50000 /EncodeMonoImages true /MonoImageFilter /CCITTFaxEncode /MonoImageDict << /K -1 /AllowPSXObjects false /CheckCompliance [ /None /PDFX1aCheck false /PDFX3Check false /PDFXCompliantPDFOnly false /PDFXNoTrimBoxError true /PDFXTrimBoxToMediaBoxOffset [ 0.00000 0.00000 0.00000 0.00000 /PDFXSetBleedBoxToMediaBox true /PDFXBleedBoxToTrimBoxOffset [ 0.00000 0.00000 0.00000 0.00000 /PDFXOutputIntentProfile () /PDFXOutputConditionIdentifier () /PDFXOutputCondition () /PDFXRegistryName () /PDFXTrapped /False
/CreateJDFFile false /Description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> /CHS <FEFF4f7f75288fd94e9b8bbe5b9a521b5efa7684002000410064006f006200650020005000440046002065876863900275284e8e9ad88d2891cf76845370524d53705237300260a853ef4ee54f7f75280020004100630072006f0062006100740020548c002000410064006f00620065002000520065006100640065007200200035002e003000204ee553ca66f49ad87248672c676562535f00521b5efa768400200050004400460020658768633002> /CHT <FEFF4f7f752890194e9b8a2d7f6e5efa7acb7684002000410064006f006200650020005000440046002065874ef69069752865bc9ad854c18cea76845370524d5370523786557406300260a853ef4ee54f7f75280020004100630072006f0062006100740020548c002000410064006f00620065002000520065006100640065007200200035002e003000204ee553ca66f49ad87248672c4f86958b555f5df25efa7acb76840020005000440046002065874ef63002> /CZE <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> /DAN <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> /DEU <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> /ESP <FEFF005500740069006c0069006300650020006500730074006100200063006f006e0066006900670075007200610063006900f3006e0020007000610072006100200063007200650061007200200064006f00630075006d0065006e0074006f00730020005000440046002000640065002000410064006f0062006500200061006400650063007500610064006f00730020007000610072006100200069006d0070007200650073006900f3006e0020007000720065002d0065006400690074006f007200690061006c00200064006500200061006c00740061002000630061006c0069006400610064002e002000530065002000700075006500640065006e00200061006200720069007200200064006f00630075006d0065006e0074006f00730020005000440046002000630072006500610064006f007300200063006f006e0020004100630072006f006200610074002c002000410064006f00620065002000520065006100640065007200200035002e003000200079002000760065007200730069006f006e0065007300200070006f00730074006500720069006f007200650073002e> /ETI <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> /FRA <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> /GRE <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a stvaranje Adobe PDF dokumenata najpogodnijih za visokokvalitetni ispis prije tiskanja koristite ove postavke. Stvoreni PDF dokumenti mogu se otvoriti Acrobat i Adobe Reader 5.0 i kasnijim verzijama.)
/HUN /ITA /JPN /KOR /LTH /LVI /NLD (Gebruik deze instellingen om Adobe PDF-documenten te maken die zijn geoptimaliseerd voor prepress-afdrukken van hoge kwaliteit. De gemaakte PDF-documenten kunnen worden geopend met Acrobat en Adobe Reader 5.0 en…
This is the start of the file's text. The full file is on GovTribe.