Project 580-22-106 PDC Drawings dated 4-26-22.pdf

PDF 4 MB Posted

Attached to
Y1DA--580-22-106 Site Prep Extremity-MRI Federal contract opportunity
Solicitation number
36C25623R0087
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 16

About this file

This document contains a pre-solicitation notice for a federal construction contract opportunity with the Department of Veterans Affairs. The Michael E. DeBakey VA Medical Center in Houston, Texas is seeking a service-disabled veteran-owned small business to perform site preparation work for a replacement MRI-Extremity located on the first floor of Building 100. The work includes demolition, mechanical, electrical, plumbing, structural, fire protection, medical gas, and interior finishes construction per provided plans and specifications. The period of performance is 300 calendar days from notice to proceed. Interested parties should monitor the website on or after June 13, 2023 for the request for proposals solicitation and any amendments. Proposals will be accepted for 90 days and must include a bid bond. Performance and payment bonds will be required from the awarded contractor. The construction value is estimated between $1,000,000 to $2,000,000. Offerors must be registered in SAM, VetCert, and have completed online representations and certifications by the solicitation due date.

View the file

Other files for this federal contract opportunity

Other files attached to Y1DA--580-22-106 Site Prep Extremity-MRI, newest first.
File Type Posted
Site Visit Sign-In Sheet.jpg JPG image
36C25623R0087 0001.docx DOCX document
DOL Wage Determinations TX20230253 dated 5-5-23.pdf PDF
SF24 Bid Bond.pdf PDF
36C25623R0087_1.docx DOCX document
Project 580-22-106 Drawings.pdf PDF
Project 580-22-106 GE Drawings dated 10-5-22.pdf PDF
Project 580-22-106 Statement of Work.docx DOCX document
Project 580-22-106 Specifications.pdf PDF
36C25623R0087.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

PRJ DT

PROJECT

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

T100

TABLE OF CONTENTS

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

1 15

System:

PROJECT NUMBER

Drawing Reference:

GE Representative:

V. Kelly Martin phone:

email: Vinson.Martin@ge.com

Company:

Address :

Contact:

Facility:

Address :

2002 Holcombe Blvd

Houston, TX 77030

Contact:

Ravi Kannan, PE General Engineer

Firm:

Spur Design

Address :

312 SW 25th Street

Oklahoma City, OK 73109

Contact:

Jennifer Knight, AIA, LEED AP BD+C

Project Manager / Senior Architect

D 405.673.2140

O 405.842.6100

Company:

PDC Facilities, Inc.

Design/Coordination:

David Knuth, AIA email: david.knuth@pdcfacilities.com phone: 262-367-7950

Electronics:

Mike Slemin, Electrical Project Manager email: mslemin@pdcbiz.com phone: 262-367-7702

THESE DRAWINGS INDICATE THE INTERCONNECTION BUT NOT THE PLACEMENT OF THE LISTED EQUIPMENT COMPONENTS. THESE DRAWINGS ARE NOT CONSTRUCTION DRAWINGS NOR SITE PREPARATION DRAWINGS. CUSTOMER REMAINS ULTIMATELY RESPONSIBLE

FOR PREPARING THE SITE TO ACCOMMODATE THE INSTALLATION AND OPERATION OF SUCH EQUIPMENT IN COMPLIANCE WITH ALL APPLICABLE FEDERAL, STATE, AND/OR LOCAL REQUIREMENTS. THESE SHEETS ARE A DOCUMENT SET AND SHOULD NOT BE SEPARATED.

NOTE:

EVEN THOUGH THE CMRS SYSTEM CONTAINS AN AUDIO COMPONENT, IT IS NOT A PRODUCT USED FOR COMMUNICATIONS BETWEEN TECHNOLOGIST AND PATIENT. HOWEVER, THE PDC SUITE SYSTEM HAS BEEN DEVELOPED WITH THE CAPABILITY TO BE INTEGRATED WITH AN MRI VENDOR'S PATIENT COMMUNICATION SYSTEM. THEREFORE, PDC HIGHLY RECOMMENDS EITHER THE RELOCATION OF EXISTING COMMUNICATION EQUIPMENT OR PURCHASING THIS EQUIPMENT FROM YOUR MAGNET VENDOR.

Sheet

Number

TABLE OF CONTENTS

SHEET NAME Revision

Revision

Date

T100 TABLE OF CONTENTS

E001 EQUIPMENT SUPPLIED

E002 WIRE SCHEDULE

E201 SCAN ROOM: RFP EF30 PVD SPK

713-248-7733

Michael E. DeBakey VA Medical Center

MANUFACTURER SITE ARCHITECT GENERAL CONTRACTOR

GE Representative

E200 SCAN ROOM: RFP LP6 LPD

A100 FLOOR PLAN

A101 REFLECTED CEILING PLAN

E202

E300 CONTROL ROOM: AVB BRP PAS LC-W S-OO

SCAN ROOM: PSR CIT

Projects Planning, Design & Construction

Facilities Management Service Line

E101 EQUIPMENT ROOM: SCR PSF

E100 EQUIPMENT ROOM: SCR RFP EF30

E204 SCAN ROOM: CLF IWF-L

E203 SCAN ROOM: CLF IWF-R

E205 SCAN ROOM: IWF-L IWF-R

E102 EQUIPMENT ROOM: SCR IAP P-CF CAT6 S-OO

MEDVAMC Houston, Texas

RaviKumar.Kannan@va.govmailto:

email: jennifer.knight@spur-design.com spur-design.com

02/22/232

02/22/232

11/11/221

11/11/221

C1C-316B

CORRIDOR1C-394

EQUIPMENT

1C-396A

MRI

1C-396

CONTROL

LINE OF SOFFIT

(above)

(Right Image)

PORTABLE VIDEO

DISPLAY w/ CART by PDC (comes w/ 40' of cables)

PAS

BRP

IWF-R

IWF-L

PVD

NOTE:

PAD IS MOUNTED TO AVB

SCR

NOTE:

SCR TO HAVE (2) 115vac, 20A DEDICATED

CIRCUITS FOR INPUT POWER.

(both in same conduit)

LC-w

IWF-L

IWF-R

(Right Image)

(Left Image)

(Left Image)

LINE OF SOFFIT

(above)

ISOCENTER

AVB

C le ar an ce 3'

6"

C lr

1'

0"

3' - 0"

EO-d

P-cf

P-pvd

Cat6

EO-d

S-oo

1/4" = 1'-0"1 FLOOR PLAN

PRJ DT

PROJECT

CHK BY

SHEET TITLE

SHEET NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED THEREIN

ARE THE PROPERTY OF PDC FACILITIES, INC. HARTLAND

WISCONSIN.

RECIPIENT AGREES NOT TO:

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON IN

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER WITHOUT

THE EXPRESS WRITTEN PERMISSION OF PDC FACILITIES. INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN WHOLE

OR IN PART WITH OUT THE EXPRESS WRITTEN PERMISSION OF

PDC FACILITIES, INC.

REVISIONS

PROJECT

CARING MR SUITE

DESIGN AND

CONSULTATION

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN 53029

PH: (262) 367-7700

FAX: (262) 367-7744

SITE INSTALLATION DRAWINGS (SID)

THESE DRAWINGS ARE NOT CONSTRUCTION OR SITE

PREPARATION DRAWINGS

Facilities, Inc.

Caring MR Suite

CHK DT

A100

FLOOR PLAN

KJL

DK/MS

10/21/2022

Michael E. DeBakey VA Medical Center 2002 Holcombe Blvd Houston, TX 77030

10/21/2022

Revision Schedule

No. Description Date

D SCR: Additional Dedicated Circuit 18 June 2020 E LC-w replaces DTP 13 July 2020

ISOCENTER

SERVICE CLEARANCE:

2'-0" FRONT & BACK

NOTE:

CLEAR ACCESS TO RF FILTER PACK

(RFP) WIRE CONNECTIONS &

MOUNTING BOLTS; FROM BELOW,

MUST BE MAINTAINED FOR FUTURE

SERVICE, UPGRADES & MAINTENANCE.

SPK

CIT

RFP

EF30

CLF

LP6

LPD

SPK

CLF

PRJ DT

PROJECT

CHK BY

SHEET TITLE

SHEET NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED THEREIN

ARE THE PROPERTY OF PDC FACILITIES, INC. HARTLAND

WISCONSIN.

RECIPIENT AGREES NOT TO:

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON IN

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER WITHOUT

THE EXPRESS WRITTEN PERMISSION OF PDC FACILITIES. INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN WHOLE

OR IN PART WITH OUT THE EXPRESS WRITTEN PERMISSION OF

PDC FACILITIES, INC.

REVISIONS

PROJECT

CARING MR SUITE

DESIGN AND

CONSULTATION

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN 53029

PH: (262) 367-7700

FAX: (262) 367-7744

SITE INSTALLATION DRAWINGS (SID)

THESE DRAWINGS ARE NOT CONSTRUCTION OR SITE

PREPARATION DRAWINGS

Facilities, Inc.

Caring MR Suite

CHK DT

A110

REFLECTED CEILING

PLAN

KJL

DK/MS

10/21/2022

Michael E. DeBakey VA Medical Center 2002 Holcombe Blvd Houston, TX 77030

10/21/2022

1/4" = 1'-0"1 REFLECTED CEILING PLAN

Revision Schedule

No. Description Date

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E001

EQUIPMENT SUPPLIED

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

4 15

Mark

EQUIPMENT SUPPLIED

Count Description Part No.

Provided by

Installed by

Scope of Work

LP6

RFP

Mark

ADDITIONAL EQUIPMENT

Count Description Part No.

Provided by

Installed by

Scope of Work

Comments

EO-d

EO-q

S-oo

1 N/A

N/A ECEC

N/A ECEC

PDC

PDC

SHC

EC

PDC = PDC Facilities, Inc. GC = General Contractor EC = Electrical Contractor VEN = Magnet Vendor SHC = Shielding Contractor

1-01-000024

1-07-000132

Comments

RF Filter Pack

115vac, 15A Duplex Outlet

115vac, 20A Quad Outlet

On/Off Switch ECEC

Low Profile 6" Can Light

LPD 2 PDC EC1-07-000105Directional LED Can Light

SPK 4 Audio Speakers

LC-w

AVB 1 Audio Controller with Tablet

1 Lighting Controller - Wall Mounted

EF30 1 RF Filter - 30A

PAS 1 Patient Audio System

P-cf 1 Faceplate for AVB

1-06-000084

3-23-000309

BRP 1 Blu-Ray Player 3-23-000158

1-06-000076 PDC EC

3-01-000008

N/A

1-06-000094

VENVEN

PDC SHC

PDC SHC

PDC PDC

PDC EC

PDC = PDC Facilities, Inc. GC = General Contractor EC = Electrical Contractor VEN = Magnet Vendor SHC = Shielding Contractor

SCR 1 1-06-000074System Control Rack PDC EC

IAP 1

PDC PDC

Internet Access Point N/A ECEC

Cat6 1 1-06-000095Cat-6 Faceplate for LC-w PDC EC

PSF 1 1-05-000184MZPS Filter PCBA 16 Channel PDC

IWF-L 2 PDC1-07-000141Illuminated Wall Fixture - Left Image GC/EC

IWF-R 2 PDC1-07-000142Illuminated Wall Fixture - Right Image GC/EC

CIT 9 Ceiling Image Tile - 2' x 2' 1-07-000322 PDC PDC

CLF 2 Curved Cove Light Fixture 1-07-000011 PDC PDC

PVD 1 1-03-000188Portable Video Display - 32" PDC GC/EC

SHC

P-pvd 1 Faceplate for PVD 5-03-000200 PDC EC

Equipment Supplied

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E002

WIRE SCHEDULE

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

5 15

1 Remove "2-Wire" 11/11/22

Group

Types

Wire

Group

Wire Schedule

Conduit

Size

Provided by

Installed by

Termin. at device by

Wire

Length

Wire

Designation

Cable

Ends

PDC

Part #

CAT-6 (A) one (1) Cat-6 cable as per NEC as requiredEC ECECShielded Cat6 Cable; RJ45/RJ45

CAT-6 (B) one (1) Cat-6 cable n/a 10 ft.PDC PDCPDCShielded Cat6 Cable; RJ45/RJ453-23-000605

The PDC Facilities, Inc. provided cables are not plenum rated. The site architect or electrical contractor (EC) is responsible for determining if conduit must be utilized to meet local codes and regulations. Contact PDC Facilities, Inc. immediately if any substitutions are required.

The EC is responsible for determining if the wire types suggested by PDC Facilities, Inc. are sufficient to meet local codes and regulations. All electrical conduit and J-boxes are provided by and installed by EC. The EC will verify with MR vendor and architect whether all conduit, J-boxes and any other material in the Scan Room are to be non-ferrous. When cabling the CMRS equipment, the excess cable should be coiled in a figure eight, zip tied and stored. Do not route cables through cable exclusion areas above magnet or above rear pedestal.

CAT-5 (A) three (3) Cat-5 cables ECECShielded Cat5e Ethernet Cable; Snagless/Molded Boot3-23-000153

CAT-5 (B) one (1) Cat-5 cable as per NEC as requiredEC ECECShielded Cat5e Ethernet Cable; RJ45/RJ45

MDV (A) Assorted Audio Device Cables

AUX (A) Patient Audio System Cable as requiredEC ECECON/OFF two (2) #18 AWG, Cu as per NECNote: Leviton 1081 Momentary Contact Switch or equivalent

6 ft.PDC PDCPDC

6 ft.PDC PDCPDC one (1) "Speaker Cable", 18 awg

2-conductor, in-wall, cable, twisted pair

SPKR (A)

SPKR (B)

two (2) "Speaker Cable", 18 awg

2-conductor, in-wall, cable, twisted pair as requiredEC ECEC as requiredEC ECEC as per NEC as per NEC

25 ft.

3-23-000151

5-06-000068 n/a n/a

115vac, 15A

(power cord) two (2) #18 AWG, Cu & one (1) #18 AWG, Cu gnd.

n/a variesPDC ECECComes with Equipment

115vac, 20A

(ded. circuit) two (2) #12 AWG, Cu & one (1) #12 AWG, Cu gnd.

Power for the Suite System (hard/flex piped to System Control Rack - SCR) n/a as requiredEC ECEC

Varies

12VDC (A) as requiredsix (6) #12 AWG, Cu as per NEC EC ECEC208A (RED), 209A (GREEN), 210A (BLUE), +12A, +12B, +12C

12VDC (E) as requiredfour (4) #12 AWG, Cu as per NEC EC ECEC208A (RED), 209A (GREEN), 210A (BLUE), +12A

12VDC (F) as requiredfour (4) #12 AWG, Cu as per NEC EC ECEC208A (RED), 209A (GREEN), 210A (BLUE), +12B

12VDC (G) as requiredfour (4) #12 AWG, Cu as per NEC EC ECEC208A (RED), 209A (GREEN), 210A (BLUE), +12C

PDC

F/O (A) two (2) fiber optic cables: Duplex OM3

Multinode Armored PVC 3mm; ST/ST

3-23-000505 n/a 1 ft.PDC PDCPDCHDMI (A) one (1) "HDMI" Cable HDMI Cable

3-23-000371 ECECPDCn/a 25 ft.

Note:

Minimum 12" distance from vendor's gradient cables is required.

F/O (B) two (2) fiber optic cables: Duplex OM3

Multinode Armored PVC 3mm; ST/ST

3-23-000372 ECECPDCn/a 50 ft.

Note:

Minimum 12" distance from vendor's gradient cables is required.

42VDC (A) as requiredas per NEC EC ECEC+42VDC (Orange), -42VDC (Blue/White)two (2) #12 AWG, Cu

F/O (C) two (2) fiber optic cables: Duplex OM3

Multinode Armored PVC 3mm; LC/LC

3-23-000378 ECECPDCn/a 100 ft.1" n/a

F/O (D) two (2) fiber optic cables: Duplex OM3

Multinode Armored PVC 3mm; LC/LC

3-23-000604

CAT-6 (C) one (1) Cat-6 cable n/a 1 ft.PDC PDCPDCShielded Cat6 Cable; RJ45/RJ453-23-000108

ECECPDCn/a 10 ft.1"

My Device Cable

3.5mm to RCA Cable

AUX (A) Patient Audio System Cable 8 ft.PDC PDCPDC3-23-000581 n/a3.5mm to 3.5mm

3-23-000093 n/a 6 ft.PDC PDCPDCHDMI (B) one (1) "HDMI" Cable HDMI Cable

Wire Schedule

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E100

EQUIPMENT ROOM

SCR RFP EF30

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

6 15

CAT-5 (A)

9-12

5-8

1-4

RFP

SITE

NETWORK

DCLC

LINK

LEFT

SPEAKERS

RIGHT

SPEAKERS

LIGHTS

OVERRIDE

ACCESS

POINT

TABLET

AVB

LIGHTS

9-12

PDC

NETWORK

SERVICELIGHTS

5-8

LIGHTS

1-4

1A

1B

2A

2B

3A

3B

4A

4B

5A

5B

6A

6B

7A

7B

SCR

DC WIRE J-BOX

Flexible Conduit by EC

+SPKR_R

-SPKR_R

-SPKR_L

+SPKR_L

ACC-1

ACC-2

ACC-3

ACC-4

ACC-5

ACC-6

+12C

+12B

+12A

210A

209A

208A

CK6

CK5

CK4

CK3

CK2

CK1

AC WIRE J-BOX

Green

Black

To: Ground screw located in AC wire J-box.

UPS

12CH

RGB

Lights - Line

White

Lights - Neutral

White

A/V - Neutral

Black

A/V - Line

Flexible Conduit by EC

BLACK

WHITE

GREEN

BLACK

WHITE

To: Grounded 115 vac source equipped with 20 amp circuit breaker by EC

To: 115 vac source equipped with 20 amp circuit breaker by EC

115, 20A

(ded. circuit)

115, 20A

(ded. circuit)

L

V

V

EF30

Ground screw located in pen panel.

NPE

F/O (A)

SPKR (B)

42VDC (A)

Flexible conduit by EC

LEFT SPEAKER -

+SPKR_R

-SPKR_R

-SPKR_L

+SPKR_L

ACC-1

ACC-2

ACC-3

ACC-4

ACC-5

ACC-6

LEFT SPEAKER +

RIGHT SPEAKER -

RIGHT SPEAKER +

RGB

Black +12A

Common

Black +12B

Common

Black +12C

Common

Red

208A

Green

209A

Blue

210A

Gray

Gray

Blue/White

Blue/White

Blue

12VDC (A)

MCB

Blue/White

-42VDC

Orange

+42VDC

Electrical contractor is responsible for determining if conduit must be utilized to meet local codes and regulations.

Equipment Room

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E101

EQUIPMENT ROOM

SCR PSF

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

7 15

SITE

NETWORK

DCLC

LINK

LEFT

SPEAKERS

RIGHT

SPEAKERS

LIGHTS

OVERRIDE

ACCESS

POINT

TABLET

AVB

LIGHTS

9-12

PDC

NETWORK

SERVICELIGHTS

5-8

LIGHTS

1-4

1A

1B

2A

2B

3A

3B

4A

4B

5A

5B

6A

6B

7A

7B

SCR

PSF

Input Voltage: 100-240VAC

50/60Hz by EC

O r a n g e

Y e l l o w

Light Blue

B r o w n

G r e e n Y e l l o w

White

Black

CAT-5 (B)

115vac, 15A

(ded. circuit)

Red

LINE

NEUTRAL

GROUND

Equipment Room

Electrical contractor is responsible for determining if conduit must be utilized

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E102

EQUIPMENT ROOM

SCR IAP P-CF CAT6 S-OO

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

8 15

SITE

NETWORK

DCLC

LINK

LEFT

SPEAKERS

RIGHT

SPEAKERS

LIGHTS

OVERRIDE

ACCESS

POINT

TABLET

AVB

LIGHTS

9-12

PDC

NETWORK

SERVICELIGHTS

5-8

LIGHTS

1-4

1A

1B

2A

2B

3A

3B

4A

4B

5A

5B

6A

6B

7A

7B

Internet Access Point

Located Near SCR by EC

J-Box by EC

RACO 256 or equivalent

(3-3/4" box by 3-1/2" deep)

PAD and AVB Wall Plate

P-cf

IAP

ON/OFF

CAT-6 (B)

F/O (C)

CAT-6 (A)

CAT-6 (A)

A1 L1

Typical Momentary Switch

Leviton 1081 or Similar

B1

S-oo

Located in Control Room by ECLocated in Control Room by EC

J-Box by EC

RACO 256 or equivalent

(3-3/4" box by 3-1/2" deep)

SCR

Typical Cat-6 Wall Plate

Cat6

CAT-6 (A)

J-Box by EC

RACO 256 or equivalent

(3-3/4" box by 3-1/2" deep)

Located in Control Room Near

Lighting Control (LC-w) by EC

TABLET

ACCESS POINT

(When used)

Equipment Room

Electrical contractor is responsible for

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E200

SCAN ROOM

RFP LP6 LPD

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

9 15

OUTIN

CAT-5 (B)

LP6

IN

OUT

IN

OUT

CAT-5 (B)

LPD

3a 3b

PARALLEL

IN LPD

PARALLEL

OUT

PARALLEL

IN

PARALLEL

OUT

OUTIN

CAT-5 (B)

LP6

OUTIN OUTIN

CAT-5 (B)

OUTIN

CAT-5 (B)

LP6 LP6 LP6

9 10 12

CAT-5 (B)

OUTIN

CAT-5 (B)

OUTIN

CAT-5 (B)

LP6 LP6

6 8

OUTIN

CAT-5 (B)

LP6

CAT-5 (B)

RFP

CK6

CK5

+12C

+12B

+12A

CK4

210A

CK3

209A

CK2

208A

CK1

1-4

5-8

9-12

+SPKR_R

-SPKR_R

-SPKR_L

+SPKR_L

ACC-2

ACC-4

ACC-5

ACC-1

ACC-4

+SPKR_R

-SPKR_R

-SPKR_L

ACC-5

+SPKR_L

ACC-3

12VDC (A)

12VDC Distribution

J-Box by EC

RACO 256 or equivalent

(3-3/4" box by 3-1/2" deep)

To: Curved Cove Light Fixture (CLF) and

Illuminated Wall Fixture (IWF-R) on Right Side

12VDC (G)

12VDC (E)

12VDC (F)

OUTIN

CAT-5 (B)

LP6

OUTIN

CAT-5 (B)

LP6

Electrical contractor is responsible for determining if conduit must be utilized to meet local codes and regulations.

Scan Room

Ref: Drawing E203

To: Curved Cove Light Fixture (CLF) and

Illuminated Wall Fixture (IWF-L) on Left Side

Ref: Drawing E204

To: Illuminated Wall Fixture (IWF-L) on Right Side and Illuminated Wall Fixture (IWF-R) on Left Side

Ref: Drawing E205

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E201

SCAN ROOM

RFP EF30 SPK PVD

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

10 15

1 Font Size 11/11/22

SPKR (A)

SPKR (A)

SPKR (A)

SPKR (A)

Left FrontSPK SPK SPK SPKLeft Rear

Right Front Right Rear

Orange +42V

Blue/White -42V

J-Box by EC

RACO 260

(4-11/16" box by 3-1/4" deep) or equivalent

Included with PVD

EF30

42VDC (A)

P-pvd

Faceplate in Scan Room

F/O (B)

PVD

42 Volts

+42V

J-BOX BY EC

NL

-42V

RFP

+SPKR_R

-SPKR_R

-SPKR_L

+SPKR_L

ACC-1

ACC-2

ACC-3

ACC-4

ACC-5

ACC-6

SPKR (B)

+SPKR_R

-SPKR_R

-SPKR_L

+SPKR_L

ACC-1

ACC-2

ACC-3

ACC-4

ACC-5

ACC-6

RIGHT SPEAKER +

RIGHT SPEAKER -

LEFT SPEAKER -

LEFT SPEAKER +

Electrical contractor is responsible for determining if conduit must be utilized

Scan Room

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E202

SCAN ROOM

PSF CIT

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

11 15

O U

T

I N

O U

T

I N

CAT-5 (B) CAT-5 (B)

CAT-5 (B)

CAT-5 (B)

CAT-5 (B)

O U

T

I N

O U

T

I N

O U

T

I N

O U

T

I N

CAT-5 (B) CAT-5 (B)

O U

T

I N

O U

T

I N

CAT-5 (B) CAT-5 (B)

O U

T

I N

CIT

PSF

Electrical contractor is responsible for determining if conduit must be utilized

Scan Room

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E203

SCAN ROOM

CLF IWF-R

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

13 15

+12A

210A

209A

208A

Black

Blue

Green

Red

Black

Blue

Green

Red

+12A

210A

209A

208A

+12A

Black

Blue

Green

Red

210A

209A

208A

12VDC (E)

12VDC (E)

J-Box by Others

RACO 256 or equivalent

(3-3/4" box by 3-1/2" deep)

12VDC (E)

IWF-R

210A

209A

208A

+12A

Black

Blue

Red

Green

CLF

12VDC (E)

12VDC (E)

(Right Side of Room)

Electrical contractor is responsible for determining if conduit must be utilized to meet local codes and regulations.

Scan Room

Right Side of Room

From: 12VDC Distribution J-Box

Ref: Drawing E200

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E204

SCAN ROOM

CLF IWF-L

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

14 15

Electrical contractor is responsible for determining if conduit must be utilized to meet local codes and regulations.

Black

+12B

210A

209A

208A

Blue

Green

Red

J-Box by Others

RACO 256 or equivalent

(3-3/4" box by 3-1/2" deep)

12VDC (F)

+12B

210A

209A

208A

Black

Blue

Green

Red

+12B

Black

Blue

Green

Red

210A

209A

208A

12VDC (F)

12VDC (F)

IWF-L

210A

209A

208A

+12B

Black

Blue

Red

Green

12VDC (F)

CLF

12VDC (F)

(Left Side of Room)

Scan Room

Left Side of Room

From: 12VDC Distribution J-Box

Ref: Drawing E200

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E205

SCAN ROOM

IWF-L IWF-R

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

12 15

210A

209A

208A

+12C

Black

Blue

Red

Green

IWF-R

J-Box by Others

RACO 256 or equivalent

(3-3/4" box by 3-1/2" deep)

12VDC (G)

IWF-L

210A

209A

208A

+12C

Black

Blue

Red

Green

12VDC (G)

12VDC (G)

From: 12VDC Distribution J-Box

Ref: Drawing E200

Electrical contractor is responsible for determining if conduit must be utilized to meet local codes and regulations.

Scan Room

Right Side of Room

Left Side of Room

CHK BY

SHEET TITLE

DRAWING NUMBER

DRW BY

Copyright © 1993-2022

THIS DOCUMENT AND THE INFORMATION CONTAINED

THEREIN ARE THE PROPERTY OF PDC FACILITIES, INC.

HARTLAND, WISCONSIN.

RECIPIENT AGREES NOT TO:

REVISIONS

PROJECT

SITE INSTALLATION DRAWINGS (SID)

THESE ARE NOT CONSTRUCTION DRAWINGS

OR SITE PREPARATION DRAWINGS

Facilities, Inc.

Caring Suite

CHK DT

Revision Schedule

No. Description Date

(1) USE SUCH INFORMATION FOR PURPOSES OTHER THAN

THOSE EXPRESSLY AUTHORIZED BY PDC FACILITIES, INC.

(2) FURNISH SUCH INFORMATION TO ANY OTHER PERSON

WHOLE OR IN PART FOR ANY PURPOSE WHATSOEVER

WITHOUT THE EXPRESS WRITTEN PERMISSION OF PDC

FACILITIES, INC.

(3) REPRODUCE OR COPY THIS DOCUMENT EITHER IN

WHOLE OR IN PART WITHOUT THE EXPRESS WRITTEN

PERMISSION OF PDC FACILITIES, INC.

700 WALNUT RIDGE DRIVE

HARTLAND, WISCONSIN

53029

PH: (262) 367-7700

FAX:(262) 367-7744

CARING MR SUITE

DESIGN AND

CONSULTATION

SHEET NUMBER

of

E300

CONTROL ROOM

AVB BRP PAS LC-W S-OO

Michael E. DeBakey VA

Medical Center

2002 Holcombe Blvd

Houston, TX 77030

DL

4/26/22

15 15

Grounded 110 vac source by EC

Equipped with 20 amp circuit breaker

LAN DC IN

FIBER

19VDC

40W

MAX

2.1 A

H D

M I

O U

T

H D

M I

I N

P U

T

A U

D I O

I N

P U

T

A U

D I O

O U

T

G N

D

I N

I N

I N

To: Customer Supplied

Music Device

V

D C

O

U T

12 VDC OUT

P-cf

Included with AVB/PAD

AVB AVB

MDV (A)

AUX (A)

To: Patient Audio

System (PAS)

Front View Back View

F/O (D)

RCA Cable Shown

Cable varies according to Manufacturer of Patient Audio System (PAS)

12V

OUT

BRP

HDMI (A)

Faceplate for AVB

Located in Under Counter in Control Room by EC

Ref: Drawing E102

EO-q

115vac 15A

(power cord)

BACK OF TABLET

Grounded 115 vac source by EC equipped with 15 amp circuit breaker

Located Near Lighting Controller (LC-w)

E0-dCat6

LC-w

Typical Cat-6 Faceplate by EC

Located Near Lighting Controller (LC-w)

Ref: Drawing E102

115vac 15A

(power cord)

Typical Momentary Switch

Leviton 1081 or Similar

(Located in Control Room by EC)

Ref: Drawing E102

ON

ON

S-oo

19 VDC OUT

Included with AVB

115vac 15A

(power cord)

115vac 15A

(power cord)

Included with BRP

PAD

CAT-6 (C)

CAT-6 (B)

Included with LC-w

V

D C

O

U T

Control Room

Electrical contractor is responsible for

4625 T100 Rev-1
Sheets and Views
Model
4625 A100 Rev-1
Sheets and Views
Model
4625 A101
Sheets and Views
Model
4625 E001
Sheets and Views
Model
4625 E002 Rev-1
Sheets and Views
Model
4625 E100
Sheets and Views
Model
4625 E101
Sheets and Views
Model
4625 E102
Sheets and Views
Model
4625 E200
Sheets and Views
Model
4625 E201 Rev-1
Sheets and Views
Model
4625 E202
Sheets and Views
Model
4625 E203
Sheets and Views
Model
4625 E204
Sheets and Views
Model
4625 E205
Sheets and Views
Model
4625 E300
Sheets and Views
Model

File details come from the government source that posted it. Updated .