Project 580-22-106 GE Drawings dated 10-5-22.pdf

PDF 2 MB Posted

Attached to
Y1DA--580-22-106 Site Prep Extremity-MRI Federal contract opportunity
Solicitation number
36C25623R0087
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 16

About this file

This document provides construction drawings and specifications for an MRI-extremity site preparation project at the Michael E. DeBakey VA Medical Center in Houston, Texas. The project includes demolition, mechanical, electrical, plumbing, structural, fire protection, radiation protection, medical gas, and interior finishes work to prepare for replacement of an existing MRI-extremity located on the first floor of Building 100. The construction work must be completed according to all applicable safety and healthcare standards without disrupting medical center operations. Interested service-disabled veteran-owned small businesses will have the opportunity to bid on providing all labor, materials, equipment, and supervision for the project through a firm-fixed-price construction contract with a period of performance of 300 calendar days. The solicitation is anticipated to be issued as a request for proposals on or around June 13, 2023, with a 90-day proposal acceptance period required and bid and performance bonds to be submitted. The project value is estimated between $1 million and $2 million under NAICS code 236220.

View the file

Other files for this federal contract opportunity

Other files attached to Y1DA--580-22-106 Site Prep Extremity-MRI, newest first.
File Type Posted
Site Visit Sign-In Sheet.jpg JPG image
36C25623R0087 0001.docx DOCX document
Project 580-22-106 Specifications.pdf PDF
SF24 Bid Bond.pdf PDF
36C25623R0087_1.docx DOCX document
Project 580-22-106 Drawings.pdf PDF
Project 580-22-106 PDC Drawings dated 4-26-22.pdf PDF
Project 580-22-106 Statement of Work.docx DOCX document
DOL Wage Determinations TX20230253 dated 5-5-23.pdf PDF
36C25623R0087.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

01 - C1 - Cover Sheet 02 - C2 - Disclaimer - Site Readiness 03 - A1 - General Notes 04 - A2 - Equipment Layout 05 - A3 - Section Views 06 - A4 - Acoustic - Proximity Limits 07 - A5 - RF shielding 08 - A6 - Equipment Details (1) 09 - A7 - Equipment Details (2) 10 - A8 - Equipment Details (3) 11 - A9 - Delivery 12 - A10 - Shield Requirements (1) 13 - A11 - Shield Requirements (2) 14 - A12 - Shield Requirements (3) 15 - S1 - Structural Notes

Houston, TX

V. Kelly Martin

SIGNA ARTIST

AKD

GE Healthcare

GE does not take responsibility for any damages resulting from changes on drawings made by others. Errors may occur by not referring to the complete set of final issue drawing. GE cannot accept responsibility for any damage due to the partial use of GE final issue drawings, however caused. All dimensions are in millimeters unless otherwise specified. Do not scale from printed pdf files. GE accepts no responsibility or liability for defective work due to scaling from these drawings. /29MRI-M234700-FIN-02-C.DWG

713-248-7733

Vinson.martin@ge.com

JDT 5670001

USA

1/4"=1'-0"

5227866

24/Feb/2023

Michael E Debakey VA Medical Center

A mandatory component of this drawing set is the GE Healthcare Pre Installation manual. Failure to reference the Pre Installation manual will result in incomplete documentation required for site design and preparation.

Pre Installation documents for GE Healthcare products can be accessed on the web at: www.gehealthcare.com/siteplanning -

Drawn by Verified by PIM Manual

SheetFormat

Concession

Scale DateFile Name

RevS.O. (GON)

A3

REV DATE MODIFICATIONS

A 05/Oct/2022 Final (DC-358849) B 17/Jan/2023 Revision (DC-371139 & DC-371314) C 24/Feb/2023 Revision (DC-376739)

FINAL STUDY

16 - S2 - Structural Layout 17 - S3 - Structural Details 18 - M1 - Mechanical Layout 19 - M2 - HVAC-Venting 20 - M3 - Chilled Water 21 - M4 - Cryogenics (1) 22 - M5 - Cryogenics (2) 23 - E1 - Electrical Notes 24 - E2 - Electrical Layout 25 - E3 - Electrical Elevations 26 - E4 - Electrical Details 27 - E5 - Electrical Details (2) 28 - E6 - Power Requirements 29 - E7 - Interconnections

· A mandatory component of this drawing set is the GE Healthcare Pre-installation manual. Failure to reference the Pre-installation manual will result in incomplete documentation required for site design and preparation.

· The items on the GE Healthcare Site Readiness Checklist DOC1809666 are REQUIRED to facilitate equipment delivery to the site. Equipment will not be delivered if these requirements are not satisfied.

CUSTOMER SITE READINESS REQUIREMENTS

REQUIRED MANUALS FOR SYSTEM PRE-INSTALLATION

Description Document Number*

Product specific Pre-installation Manual Refer to cover page Magnet Room Venting 5850263

RF Shielded Room Pre-installtion Requirements for MR systems 5850260 IEC Electromagnetic Compatibility 5850261

Acoustic Room Details 5850262 Magnet Venting Conformance Assessment Form 2705036 *documents can be accessed in multiple languages at https://customer-doc.cloud.gehealthcare.com/#/cdp/dashboard

· Any deviation from these drawings must be communicated in writing to and reviewed by your local GE Healthcare installation project manager prior to making changes.

· Make arrangements for any rigging, special handling, or facility modifications that must be made to deliver the equipment to the installation site. If desired, your local GE Healthcare installation project manager can supply a reference list of rigging contractors.

· New construction requires the following;

1. Secure area for equipment,

2. Power for drills and other test equipment,

3. Restrooms.

· Provide for refuse removal and disposal (e.g. crates, cartons, packing)

· It is required to minimize vibrations within the scan room. It is the customer's responsibility to contract a vibration consultant/engineer to implement site design modifications to meet the GE vibration specification. Refer to the system Pre-installation manual for vibration specifications.

DISCLAIMER

GENERAL SPECIFICATIONS

· GE is not responsible for the installation of developers and associated equipment, lighting, cassette trays and protective screens or derivatives not mentioned in the order.

· The final study contains recommendations for the location of GE equipment and associated devices, electrical wiring and room arrangements. When preparing the study, every effort has been made to consider every aspect of the actual equipment expected to be installed.

· The layout of the equipment offered by GE, the dimensions given for the premises, the details provided for the pre-installation work and electrical power supply are given according to the information noted during on-site study and the wishes expressed by the customer.

· The room dimensions used to create the equipment layout may originate from a previous layout and may not be accurate as they may not have been verified on site. GE cannot take any responsibility for errors due to lack of information.

· Dimensions apply to finished surfaces of the room.

· Actual configuration may differ from options presented in some typical views or tables.

· If this set of final drawings has been approved by the customer, any subsequent modification of the site must be subject to further investigation by GE about the feasibility of installing the equipment. Any reservations must be noted.

· The equipment layout indicates the placement and interconnection of the indicated equipment components.

There may be local requirements that could impact the placement of these components. It remains the customer's responsibility to ensure that the site and final equipment placement complies with all applicable local requirements.

· All work required to install GE equipment must be carried out in compliance with the building regulations and the safety standards of legal force in the country concerned.

· These drawings are not to be used for actual construction purposes. The company cannot take responsibility for any damage resulting therefrom.

CUSTOMER RESPONSIBILITIES

· It is the responsibility of the customer to prepare the site in accordance with the specifications stated in the final study. A detailed site readiness checklist is provided by GE. It is the responsibility of the customer to ensure all requirements are fulfilled and that the site conforms to all specifications defined in the checklist and final study. The GE Project Manager of Installation (PMI) will work in cooperation with the customer to follow up and ensure that actions in the checklist are complete, and if necessary, will aid in the rescheduling of the delivery and installation date.

· Prior to installation, a structural engineer of record must ensure that the floor and ceiling is designed in such a way that the loads of the installed system can be securely borne and transferred. The layout of additional structural elements, dimensioning and the selection of appropriate installation methods are the sole responsibility of the structural engineer. Execution of load bearing structures supporting equipment on the ceiling, floor or walls are the customer's responsibility.

THE UNDERSIGNED, HEREBY CERTIFIES THAT I HAVE READ AND APPROVED THE PLANS IN THIS DOCUMENT.

DATE NAME SIGNATURE

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center C2 - Disclaimer - Site Readiness 02-

MAGNETIC INTERFERENCE SPECIFICATIONS

TYPCIAL MOVING MAGNETIC MASS DISTANCE RADIALLY DISTANCE AXIALLY

Carts, Gurneys 100-400 lbs [45-182 kg] 3 Gauss line 3 Gauss line

Forklifts, small elevator, cars, minivans vans, pickup trucks, ambulances (objects greater than 400 lbs [182 kg]) 15.5 FT 4.72 M 21.0 FT 6.4 M

Buses and trucks (dump, tractor trailer, utility, fire trucks) 18.1 FT 5.52 M 24.5 FT 7.47 M

· The customer must establish protocols to prevent persons with cardiac pacemakers, neurostimulators, and biostimulation devices from entering magnetic fields of greater than 5 gauss (exclusion zone).

· Main power transformers must remain outside the 3 gauss field. EMI < 40mG RMS AC at the magnet location.

EMI <9.39mG DC.

· Potential exists under fault conditions that the 5 gauss line may expand radially to 11.48 ft. [3.5 m] and axially to 14.76 ft. [4.5 m] for 1 second or less. It should be noted that normal rampdowns or magnet rundown unit initiated quenches will not cause the magnetic field to expand.

· It is recommended every site consider the event of a quench and plan accordingly (such as placing 5 gauss warning signs at expanded locations).

· The ferrous metal objects listed below must not move into or inside of the moving metal sensitivity line during scans.

Please refer to pre-installation checklist in pre-installation manual listed on the cover sheet for items critical to image quality.

1. The layout should be arranged so that the 5g line is contained to the magnet room. If not possible, a barrier is recommended to prevent entry to the 5g field area.

2. The spaces around, above, and below the magnet must be reviewed for effects of the 5g, 3g, 1g, and .5g fields.

Refer to the proximity limit chart in the MR pre-installation manual referenced on the cover sheet.

3. For moving metal, the restriction lines typically extend outside of the MR space. Please confirm there are no moving metal concerns within these areas.

4. For vibration, analysis to be completed as required per pre-installation manual.

5. For EMI, review the site for the location of the main electrical feeders, AC devices, or distribution systems. An

EMI study is recommended if large AC systems are nearby.

6. Details of the floor below the magnet must be reviewed. The structural engineer must verify that the quantity of steel in the volume 10ft [3.1m] x 10ft [3.1m] x 13in [.3m] deep (below the magnet) does not exceed the allowable steel content as given in the MR pre-installation manual referenced on the cover sheet.

7. Remove, cover, or fill-in abandoned ducts or troughs from the Equipment and Magnet rooms. Access/computer room flooring in the Equipment room can either be removed or assessed and reinforced to support heavier cabinets.

Responsibility for the coordination, design, engineering, and site preparation resides with the customer and their project architects and contractors. GE does not, by providing reviews and furnishing comments and assistance, accept any responsibility beyond its obligations as defined in the MR system, sale/purchase agreement.

MRI SITE PLANNING REMINDERS

Broadband RF noise is a single transient or continuous series of transient disturbances caused by an electrical discharge. Low humidity environmental conditions will have higher probability of electrical discharge. The electrical discharge can occur due to electrical arcing (micro arcing) or merely static discharge. Some potential sources capable of producing electrical discharge include:

· Loose hardware/fasteners vibration or movement (electrical contunuity must always be maintained)

· Flooring material including raised access flooring (panels & support hardware) and carpeting

· Electrical fixtures (i.e. Lighting fixtures, track lighting, emergency lighting, battery chargers, outlets)

· Ducting for HVAC and cable routing

· RF shield seals (walls, doors, windows etc.)

For additional information regarding image quality, refer to the pre-installation manual listed on the cover sheet.

IMAGE QUALITY CONSIDERATIONS

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A1 - General Notes 03

N

S

EW

0.05mT (0.5G)

0.1mT (1G)

0.3mT (3G)

0.5mT (5G)

1mT (10G)

20mT (200G)

MAGNET ROOM

1C-396A

(considering RF shield walls)

EQUIPMENT

ROOM

1C-394

CONTROL ROOM

1C-396

LOCATED ELSEWHERE

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A2 - Equipment Layout 041/4"=1'-0"

6'-6"

6'-7"

14'-0"

1'-6"

'-0

'-7

'-4

2' -8

0'-11"

21'-11"

6' -8

3' -0

1' -3

'-2

7'-1"

0' -3

LOCATED ELSEWHERE

2' -4

1 2"

BY ITEM DESCRIPTION

MAX

HEAT

OUTPUT

(BTU/h)

WEIGHT

(lbs)

MAX

HEAT

OUTPUT

(W)

WEIGHT

(kg)

A 1 MAGNET (MAG) (IPM) 8184 9870 2400 4476

A 2 REAR PEDESTAL (PED) - 212 - 96

A 3 PATIENT TABLE (PT) - 463 - 210

A 4 MAGNET RUNDOWN UNIT (MRU) - 7 - 3.2

A 5 BLOWER BOX (MG6) 1534 46 450 21

A 6 PENETRATION CABINET (PEN) 10690 639 3135 290

A 7

SECONDARY PENETRATION WALL

(SPW) - - - -

A 8 POWER, GRADIENT, RF CABINET (PGR) 20927 3144 6137 1426

A 9 HEAT EXCHANGER CABINET (HEC) 3410 1349 1000 612

A 10 MAGNET MONITOR (MON) 818 10 240 4.5

A 11 CRYOCOOLER COMPRESSOR (CRY) 1705 276 500 125

A 12 OPERATOR CONSOLE (GOC) 4944 176 1450 80

A 13 MAGNETIC RESONANCE

ELASTOGRAPHY (MRE) 481 53 141 24.22

A 14 CCTV MONITOR - 8 - 3.5

A 15 CCTV CAMERA - 0 - 0.1

A 16 MUSIC SYSTEM - - - -

A 17 INJECTOR HEAD ON PEDESTAL - 95 - 43

A 18 INJECTOR CONTROL 675 18 198 8

A 19 INJECTOR POWER SUPPLY 600 7 176 3

B 20 MAIN DISCONNECT PANEL (MDP) 900 130 264 59

B 21 DIMPLEX CHILLER 204593 4300 59998 1950

B 22 WATER FILTER - - - -

B 23 REMOTE GRAPHIC DISPLAY (RGD) - - - -

B 24 MANUAL WATER BYPASS - - - -

B 25 FULL UPS 150 kW 23529 1206 6900 547

B 26 FULL UPS BATTERY CABINET - 3502 - 1588

B 27 MAINTENANCE BYPASS PANEL - 126 - 57

A 28 PNEUMATIC PATIENT ALERT - 0.5 - 0.2

D 29 METAL DETECTOR (HAND HELD) - - - -

C 30 MINIMUM OPENING FOR EQUIPMENT DELIVERY IS 40 IN. W X 82 IN. H, CONTINGENT

ON A 72 IN. CORRIDOR WIDTH

C 31 MINIMUM OPENING FOR DELIVERY IS 43 IN. W X 82 IN. H, CONTINGENT ON A 96 IN.

CORRIDOR WIDTH

C 32 COUNTER TOP FOR EQUIPMENT- PROVIDE GROMMETED OPENINGS AS REQUIRED TO

ROUTE CABLES

C 33 MOBILE CLOSET - REFER TO PREINSTALL FOR REQUIREMENTS

C 34 MAGNET ACCESS 2.5M X 2.5M [98.5IN X 98.5IN]

C 35 SHELF

C 36 DEFINE RF SHIELD'S INSET ACCORDING TO PROVISIONS MADE BY THE RF SHIELD VENDOR

C 37 SECONDARY MAGNETIC SHIELDING

LEGEND

A GE Supplied D Available from GE B GE Supplied/contractor installed E Equipment existing in room

C Customer/contractor supplied and installed * Item to be reinstalled from another site 200 Gauss 5 Gauss 100, 50, 30, 10 Gauss 3, 1, 0.5 Gauss

Exam room height

Finished floor to slab height

Finished ceiling height 10'-3"

8'-4"

RF SHIELD - 100 dB ATTENUATION

For Accessory Sales: (866) 281-7545 Options 1, 2, 1, 2 or mail to: gehcaccessorysales@ge.com

MAGNETIC SHIELDING

Moving metal sensitivity line for CARS, MINIVANS, PICKUP

TRUCKS, AND AMBULANCES.

NOTE: Ferrous objects must not move into or inside of the moving metal sensitivity line during scans.

Moving metal sensitivity line for BUSES AND TRUCKS (DUMP, TRACTOR TRAILER, UTILITY, FIRE TRUCKS)

0.05mT (0.5G)

0.1mT (1G) 0.3mT (3G)

0.5mT (5G) 1mT (10G)

20mT (200G)

0.05mT (0.5G)

0.1mT (1G)

0.3mT (3G)

0.5mT (5G) 1mT (10G) 20mT (200G)

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A3 - Section Views 051/4"=1'-0"

SIDE VIEW WITH MAGNETIC FIELD

Warning! 5 Gauss line outside the Magnet room limits.

MAGNET ROOM

CONTROL ROOM

± 0'-0"

FRONT VIEW WITH MAGNETIC FIELD

MAGNET ROOM

Warning! 5 Gauss line outside the Magnet room limits.

+ 8'-4"

+ 9'-5" + 10'-3"

± 0'-0"

3' -6

3' -6

Unistrut

'-2

0' -1

1 2"

+ TBD

+ 8'-4"

+ 9'-9" + 10'-3"

Warning! 5 Gauss line outside the Magnet room limits.

Warning! 5 Gauss line outside the Magnet room limits.

6'-7"

3'-2" 3'-9"

4'-5"

10'-0"

MAGNETIC PROXIMITY LIMITS

STEEL MASS LIMITS TO MAGNET ISOCENTER (3x3 m [10x10 ft] AREA UNDER MAGNET) Limits Of Steel Mass Distance Below Top Surface Of Floor kg/m² lbs/ft² mm in 0 0 0 - 76 0-3

9.8 2 76 - 127 3-5

14.7 3 127 - 254 5-10

39.2 8 254 - 330 10-13

98.0 20 330+ 13+

The actual field strength can be affected by Magnetic shielding, Earth's magnetic field, other magnetic fields and stationary or moving metal. This information must be used to evaluate potential site interaction of GE Healthcare equipment with other non-GE Healthcare equipment. Magnetic shielding can be installed to prevent interaction between the magnet and nearby sensitive devices. The GE Healthcare Project Manager of Installation (PMI) can work with the customer to coordinate the magnetic shielding site evaluation. The customer is responsible for installation of all magnetic shielding.

The customer must provide detail defining ferrous material below the magnet to the Project Manager so the GE Healthcare MR Siting and Shielding team can review for compliance.

Gauss (mT) Limit Equipment

0.5 gauss (0.05mT) Nuclear camera

1 gauss (0.1mT) Positron Emission Tomography scanner, Linear Accelerator, Cyclotrons, Accurate measuring scale, Image intensifiers, Bone Densitometers, Video display (tube), CT scanner, Ultrasound, Lithotriptor, Electron microscope, Digital X-Ray

3 gauss (0.3mT) Power transformers, Main electrical distribution transformers 5 gauss (0.5mT) Cardiac pacemakers, Neurostimulators, Biostimulation devices

10 gauss (1mT) Magnetic computer media, Line printers, Film processor, X-ray tubes, Emergency generators, Commercial laundry equipment, Food preparation area, Water cooling equipment, HVAC equipment, Major mechanical equipment room, Credit cards, watches, and clocks, Air conditioning equipment, Fuel storage tanks, Motors greater than 5 horsepower

50 gauss (5mT) Metal detector for screening, LCD panels, Telephones

No Limit Digital Detectors

ISOGAUSS PLOTS

The isogauss contour plots illustrated represent the performance of the system and the steel room shield described in the final shield design dated OCT. 03, 2022. The actual magnetic flux density at any point in the vicinity of the magnet when installed may vary from the contour plots due to factors such as the concentrating effects of nearby ferrous objects and ambient magnetic fields, including the earth's magnetic field. Therefore, the contours shown are only approximations of actual flux densities found at corresponding distances from the magnet's isocenter.

0 10 20 30 40 50 60

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

FREQUENCY [Hz]

VI

BR

AT

IO

N A

M

PL

IT

U

DE

[m

/s ²]

Low Frequency Magnet Floor Vibration (Vibration Amplitude at Each Foot)

Acoustic and vibroacoustic information is provided for site planning and architectural design activities. It is the customer's responsibility to hire a qualified acoustic engineer for solutions to further attenuate this transmitted noise and vibration, if required. The actual room noise level may vary based on room design, optional equipment, and usage:

ACOUSTICS SPECIFICATIONS

FREQUENCY

(Hz)

AMPLITUDE

(m/s²)

2 0.05

10 0.05

20 0.35

30 0.35

35 0.2

50 0.2

Control Room: 62 dBA Equipment Room: 80 dBA Magnet Room: 127 dBA* (maximum sound pressure level at magnet bore isocenter)

* Frequency: 20 Hz to 20kHz

SOUND PRESSURE SPECTRAL DISTRIBUTION

20 25

.5 40 50 63 80 10

20SP

L

(d

BA

n or m al ize d

Frequency (1/3 Octave)

1/3 Band Relative SPL

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A4 - Acoustic - Proximity Limits 06

PENETRATION PANEL WITH SPW

MAGNET

ROOM

FINISHED

WALL

EQUIPMENT

ROOM

SPW

PEN

BB

Minimum closet height:

2007 [79 in]

RF SHIELD

NOTES:

1) Maximum distance between PEN and SPW is 2743 [108 in]

2) SPW may be on either side of the PEN cabinet.

SCALE 1:30

468±5

[18.4 in±.2 in]

[2.4 in]

468±5

[18.4 in±.2 in]

EQUIPMENT ROOM FLOOR

RF common ground stud must be located near the PEN panel openings

OPENING

FOR SPW

OPENING

FOR PEN

[7.8 in]

±5

[6 1.

in in

0±1

[7 .5 in in

EQUIPMENT

ROOM WALL

M

IN

. 4

[1 8.

in

MIN. 1669

[65.7 in]

[7.8 in]

[3 in

NOTES:

1) RF vendor must supply 40 fasteners for hole size of Ø7.25 mm [0.285 in] (total quantity for both) and install each penetration panel adaptor frame to the wall through the outer ring of holes in the plate.

2) Minimum exposed RF shield material (around entire opening)

[6 6.

in

[23.2 in]

40 x Ø7.5 [0.3 in] (note 1)

PEN PANEL CLOSET (PLAN VIEW)

PEN PANEL OPENINGS (EQUIPMENT ROOM SIDE)

PENETRATION PANEL FRAME

(supplied by GE)

(note 2)

MINIMUM MAGNET CEILING HEIGHT ( TOP VIEW )

[7 in

[5 in

[51 in]

[51 in]

MAGNET

GEOMETRIC

ISOCENTER

FRONT OF MAGNET

Shaded area indicates floor to ceiling minimum height of 2500 mm [98.42 in]. Special service procedures are required if ceiling height is between 2500 mm [98.42 in] and 2667 mm [105 in].

PENETRATION PANEL CLOSET

An enclosure (i.e. closet) must be provided to restrict access to the PEN panels and for storage of excess interconnections.

· The PEN closet must have a mechanical locking mechanism to restrict access to the PEN panels

· The PEN closet must maintain the minimum service area outside the 200 Gauss in the magnet room.

· PEN closet must allow free air exchange of 400CFM (680 m³/hour) between the Magnet room and PEN closet for MR system blowers. Airflow may be achieved through door louvers or other openings in the PEN closet that meet all other PEN closet requirements

A closet service hatch must be provided if the room does not allow the PEN panel blower box removal path to remain completely outside the 200 Gauss line.

NOTE: If the room size is sufficiently large so the SPW blower box can be removed without entering the 200 Gauss line, a closet service hatch is not required.

The closet service hatch must meet the following requirements:

· Must be located within the PEN closet on the RF wall allowing access to the Equipment room

· May be located anywhere within the PEN closet (between 254 [10 in] and 1524 mm [60 in] with unobstructed pass-through)

· Must be minimum 508x508 mm [20x20 in]

· Must maintain RF shield integrity for all service access

· May use any design (quick disconnect RF panel, blanker panel, hinged door, etc.) as long as all other requirements are met

· The closet service hatch removal must take less than 15 minutes (replacement must also take less than 15 minutes)

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center RF shielding 071:10

GLOBAL OPERATORS CABINET (GOC)

[7.7 in]

[2 3.

in

[15.8 in]755

[29.7 in]

[1 1.

in

[12.8 in]

[27.6 in]

[1

.3 in

[3

.1 in

[4 .1 in

[1

.9 in

377.5 [14.9 in]

Center of Gravity

SCALE 1:10

TOP VIEW

FRONT VIEWSIDE VIEW

PENETRATION CABINET CLEARANCE

TOP VIEW

FRONT VIEW

SIDE VIEW

Blower boxMAGNET ROOM

EQUIPMENT ROOM

EQUIPMENT

ROOM

MAGNET ROOM

50 ± 25mm from RF shield room wall required for proper connection to penetration panel and cabinet airflow

Center of gravity

M12 x 1.75mm Seismic bracket mounting holes

Penetration panel

M12 x 1.75mm Seismic bracket mounting holes

Installation clearance

Airflow clearance

Installation clearance

Airflow clearance

Service clearance

Service clearance

Service clearance

Service clearance

Blower box

Service clearance

Cable drop area

[2 in]

[3 in

[4 in]

[23.6 in]

[15 in]

[15 in]

[3 in

[1 in

[17.3 in]

[15 in]

[4 in]

[7 5.

in

[3.9 in]

[15.8 in]

[6.5 in]

[10.6 in]

[10.6 in]

[14.1 in]

[35 in]

[36 in]

[6 in

[2 in

[1 in

[7 5.

in

[2 3.

in

[3 in

[36 in]

[36 in]

[1

.6 in

[23.6 in]

[15 in]

[1

.6 in

MAGNET ENCLOSURE

Center of gravity

Note:

Center of gravity is approximate and includes the GE Healthcare supplied VibroAcoustic Dampening Kit, but does not include cryogens, gradient assembly, side mounted electronics, or enclosures.

Enclosure dimensions are for reference only, NOT FOR SITE PLANNING USE.

1997.4 [78.6 in]

32.3 [1.3 in]

[4 2.

in

[4 2.

in

[.8 in

[1.3 in]

[9 2.

in

[97.2 in]

PATIENT TRANSPORT TABLE (PT)

SIDE VIEW

TOP VIEW

[89.6 in]

[2 6.

in

[4 0.

in

[59.1 in]

[4 in

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A6 - Equipment Details (1) 08

POWER, GRADIENT, RF CABINET (PGR)

Center of gravity

TOP VIEW A: Airflow clearance B: Dolly installation clearance C: Service clearance D: Cable strain relief E: Cable clearance F: Seismic bracket mounting holes

(all M12 x 1.75mm) B

A

A

C

B

B

C

B

B B

D

E

FRONT VIEW SIDE VIEW

[7 0.

in

[4 in

[58.27 in]

[31.73 in]

[8 2.

in

[7

4.

in

0.

[2

8.

in

561.8 [22.1 in]

499.3 [19.66 in]

[6 in

] 38

[1 in

[36 in]

[15 in]

[15 in]

[34.33 in]

[4 in]

[26.8 in]

[15.8 in]

[7.9 in]

[1

.6 in

[6.5 in]

[10.6 in]

[10.6 in]

[1

.6 in

FF

CRYOCOOLER COMPRESSOR ( CRY )

SCALE 1:20

A: Maintenance space B: Installation Clearance

Center of gravity

A

A

B

B

A

A B

A B

A, B

[3 6.

in

] 15

[6 in

[1

1.

in

[31.5 in]

[2 3.

in

[3 6.

in

[2

6.

in

[9 .8 in

[1 3.

in

[9.1 in]

[19.1 in]

[10.9 in]

[4.1 in]

[6 in]

[17.7 in]

[6 in]

Input power terminal

Input power terminal

TOP VIEW

FRONT VIEW SIDE VIEW

[15 in]

[34.6 in]

[15 in]

[4 in

[7 0.

in

[3 4.

in

[7 4.

in

[2

8.

in

[1 9.

in

[17.3 in]

[17.5 in]

[34.3 in]

[36 in]

[1 in

[24 in]

[6 in]

[24 in]

TOP VIEW TOP VIEW

OR

monitormonitor

Cable Tray:

bottom must be at least 483mm [19.02in] above cabinet

HEAT EXCHANGER CABINET ( HEC )

Additional service clearance required for Magnet Monitor.

It can be installed on either side of the HEC.

A: Airflow clearance B: Installation clearance C: Service clearance D:Seismic bracket mounting holes

(all M12 x 1.75mm)

Center of gravity

A A A

A

B B

C

C

C

C

C C C

Shipping without fluid Shipping with fluid

FRONT VIEW SIDE VIEW

D170 [6.7 in]

[6.5 in]

[10.6 in]

[10.6 in]

[10.6 in]

[10.6 in]

D

[1

.6 in

C

[6 in]

C

SECONDARY PENETRATION WALL (SPW)

FRONT VIEW SIDE VIEWTOP VIEW

SCALE 1:20

Magnet Room

Equipment Room

[23.6 in]

[3 in

[3 in

[1 8.

in

[1

3.

in

Space for hoses

[17.6 in]

[6 0.

in

[6.2 in]

[7.8 in]

[12 in]

Installation clearance, either side

[8 in]

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A7 - Equipment Details (2) 09

SCALE 1:20

MAGNET MONITOR AND MAGNET RUNDOWN UNIT (MRU)

[5.6 in]

[1.2 in]

[7 .7 in

[11.3 in]

[8.3 in]

[2

.1 in

]105 [4.1 in]

[3 in]

[15 in]

[1 0.

in

[1 1.

in

[.5 in

[.5 in

[11 in]

[.9 in]

[.9 in]

[6 in]

[6 in]

MAGNET MONITOR

MAGNET RUNDOWN UNIT (MRU)

[5 in]

Ø 26 mm [1 in] cable access

Mounting:

Ø7[Ø.3 in]

Ø5[Ø.2 in]

Ø10

[Ø .4 in

Mounting:

Front view: Side view:

Front view: Side view:

The bottom edge of the MRU must be mounted 1524 ±25 mm [60 ± 1 in] above the magnet room floor.

[4 .7 in

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A8 - Equipment Details (3) 10

This is only a partial list of items required for delivery of the magnet. For a complete checklist refer to the pre-installation manual referenced on cover sheet.

· 24/7 chilled water and 480v power for shield/cryo cooler

· 24/7 120v power for the magnet monitor

· Phone lines for magnet monitoring and emergency use

· Operational magnet room exhaust fan

· Cryogen venting (if roof hatch, completed within 24 hrs)

CRITICAL ITEMS FOR MAGNET DELIVERY

GRADIENT COIL REPLACEMENT

Front view of the BRM Gradient Side view of the BRM Gradient

The weight bearing structure of the site should support any additional weight of the main replacement parts occurring during maintenance of the magnet, throughout the whole lifecycle of the MR.

EQUIPMENT

DIMENSIONS

LxWxH WEIGHT NOTE mm in kg lbs

Replacement BRM gradient coil assembly on a shipping cradle/cart 991x2536x1499 39x99.84x59 1449 3194

Initial gradient coil assembly is shipped installed in the magnet. Shipping/installation cart is used to install re-placement coil assembly only.

[99.8 in]

[5 in

[39 in]

DELIVERY

STORAGE CONDITIONS

· System components except the magnet should be stored in a cleaned room:

· Temperature = -30 to 60°C [-22 to 140], relative humidity < 90% non condensing.

· Material should not be stored for more than 90 days.

· The magnet will be delivered after GE validation of the site.

INSTALLATION AND DELIVERY ACCEPTANCE

· A survey of the site established by the customer and GE will make the decision for the delivery time.

· This survey of the site (a form is made available by GE) is only to check if the apparent conditions of the site allow the equipment to be delivered.

· If the site is not ready, GE can delay the delivery time.

FRONT VIEW OF MAGNET SIDE VIEW OF MAGNET

[94 in]

STRAIGHT PATH

ROUTING

· The customer is solely liable for routing of components from dock to final site.

· GE must be able to move system components in or out with no need to uncrate or disassemble any of the components. The entire passageway must be cleared, adequately lighted and free from dust.

· The floor and it surfacing must be able to withstand the live load of components and handling equipment.

· Floor surfacing must be continuous.

· The customer must protect any fragile flooring surfaces.

SPECIFICATIONS FOR MAGNET ROUTING

· Floor must be able to withstand a moving load of 4823kg [10632 lb]

· Recommended opening height: 2.5m [98.5in], width: 2.5m [98.5in]. If recommended dimensions cannot be met refer to pre-installation manual for detailed specifications.

· Maximum slope: 30°

PATH WITH 90 DEGREE TURN

[98.5 in]

[98.5 in]

[83.5 in]

[90.7 in]

[8 8.

in

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A9 - Delivery 11

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A10 - Shield Requirements (1) 121:100

PERFORMANCE STATEMENT FOR MAGNETIC SHIELD DESIGN SHIELDING PLAN [REFERENCE ONLY-NOT TO SCALE]

ADDITIONAL NOTES

· Attractive force of a magnet on the ceiling and walls of the steel shield will be a maximum of 30% of the shield's weight. Support structure must account for these additional forces.

· Attractive forces on the floor or sub-floor may be sufficient to lift shield's weight. Support structure must prevent this with adhesive or bracing.

· M36 Silicon steel is used in this shield. see shielding requirements sheets for specifications.

All surfaces are 1/8" M36

GE HEALTHCARE ("GE") provides this magnetic shielding design solely for use with GE's Magnetic Resonance System ("System"). This magnetic shielding design applies solely to the magnetic resonance suite configured as shown in this drawing set.

The magnetic shield, shown in this set, when constructed in strict accordance with this magnetic shield design will:

1. Contain the MRI-generated magnetic induction of 5 gauss within 11'-0" plan north of the magnet's isocenter.

("Design Objectives"), subject to and conditioned upon the following:

1. The magnetic field strength measurements are conducted using techniques which account for the effects of the earth's magnetic field and;

2. The magnetic field strength measurements are conducted using techniques which account for magnetic field distortions caused by magnetic materials in the vicinity of the measurement and;

3. Customer demonstrates to GE that the shielding material and fabrications are in strict accordance with GE's design and specifications and;

4. Customer demonstrates to GE that the shielding is installed in strict accordance with GE's design, specifications and requirements and;

5. Customer demonstrates to GE that (i) the System has not been and is not being subjected to improper or extraordinary use, (ii) the System has not been and is not being subjected to improper maintenance, and (iii) Customer has complied with all applicable instructions or recommendations of GE related to the System and magnetic shield; and

6. Customer operates the System at or below its specified field strength.

GE will re-perform the magnetic shielding design services, to the extent that GE determines the design services fail to meet the Design Objectives specified above, provided GE is notified in writing by Customer of such failure within one (1) week after the date of availability of the System for first use, subject to all other terms and conditions of the agreement between GE and Customer for the purchase or lease of the System.

The foregoing sets forth Customer's exclusive remedies and GE's sole liability for claims based on the failure of the magnetic shielding design services to meet the Design Objectives. In no event shall GE be liable for special or consequential damages.

The magnetic field gradients resulting from the magnetic shield's effect on the System's magnet will allow the System's magnet to meet published homogeneity specifications upon completion of required installation procedures.

The final performance of the magnetic shield set forth in this drawing set depends on strict adherence to the construction and material requirements detailed herein. Any deviation from these requirements will degrade shielding effectiveness. For this reason, GE recommends that only established magnetic shielding vendors be used. Vendors should be required to provide documentary evidence to demonstrate that their shielding product complies with GE's construction requirements, material requirements, chemical composition specifications, and annealing specifications; or the material's DC magnetization curve at high induction is equivalent to GE's specification on SH2.. A guarantee that the finished shield will meet the Customer's fringe field containment requirements should be sought from the shielding vendor.

The isogauss contours shown on this drawing set indicate the magnetic fringe field containment which corresponds to the predicted interaction between the System and the magnetic shield design described. GE creates these isogauss contour plots only to predict if the shield design meets the Design Objectives. All other effects such as the superposition of the earth's magnetic field, residual magnetism, and localized field concentration effects, due to ferromagnetic structural elements, are not considered or indicated. The exact location of a particular isogauss contour may differ from that shown due to: (1.) the above stated reasons and (2.) the resolution limits of the mathematical modeling techniques used to derive these results. Note also that for actively shielded Systems, fault conditions may exist where the isogauss contours expand temporarily (refer to your System's Pre-Installation Manual for exact information regarding fault expansions).

The Customer is responsible for the effect of the fringe fields produced by the System's magnet, and is responsible for the coordination of the magnetic shield design into existing or planned facilities, such as foundations/footings and other building components.

The Design Objectives have been formulated by GE from information, both written and verbal, obtained from the Customer, his agents, and/or representatives as part of the MR Site Evaluation process. The final decision on the Design Objectives criteria and the amount of magnetic field containment needed at this site rests solely with the Customer or his designated representatives. Please contact MR Siting/Shielding immediately with any corrections and/or additions related to this magnetic shield design.

GE recommends that your final shield fabrication drawings be forwarded for review. Any deviations from the design specified in this drawing set must be brought to the attention of GE. Please allow up to two weeks for the review process.

Send drawings to your Project Manager as identified on the cover sheet.

MichaelEDVAMC_20234700_run02_HoustonTX_15IPM-vf.xlsm

6'-7" 6'-7"

5' -0

'-7

N

S

EW

TYPICAL SHIELD - GENERAL VIEW

TYPICAL SHIELD OPENINGS

A. WINDOW 4'-8" WIDE x 4'-2" HIGH (1422 mm WIDE x 1270 mm HIGH)

B. DOOR 4'-8" WIDE x 7'-6" HIGH (1422 mm WIDE x 2286 mm HIGH) FOR

C. SIGNA 3T, 1.5T PENETRATION PANEL 2'-2" WIDE x 6'-2" HIGH (660 mm WIDE x 1879 mm HIGH)

C.A. FOR MR 1.5T HDe, SIGNA EXPLORER/CREATOR PENETRATION PANEL 5' -1" WIDE x 6'-7" HIGH (1550 mm WIDE x 2000 mm HIGH)

C.B. FOR DV 3T AND 1.5T TYPICAL OPENING FOR 2 PENETRATION PANEL 6'-0"

WIDE x 10'-0" HIGH (1830 mm WIDE x 3050 mm HIGH)

C.C. FOR SIGNA PIONEER/VOYAGER OPENING FOR 2 PENETRATION PANEL 7'-2"

WIDE x 10'-0" HIGH (2150 mm WIDE x 3050 mm HIGH)

A

BC

TYPICAL SHIELD - GENERAL VIEW DETAILS

- TYPICAL LAPPING PLATE DETAIL -

- TYPICAL BEVELED CORNER DETAIL -

- ALTERNATE LAPPING PLATE DETAIL -

DETAIL D

DETAIL F

DETAIL C

MAX. GAP 1/8" (3 MM)

MAX. GAP 1/8" (3 MM)

LAPPING PLATE 6" (152 MM)

MAX. GAP 1/8" (3 MM)

(76 MM) x 3" (76 MM) x 1/3 PLATE THICKNESS ANGLE PLATE x LENGTH OF JOINT

GRAIN DIRECTIONSHORTER

DIMENSION

SLOTTED OR OVERSIZED HOLES (TYPICAL)

LAP PLATE MINIMUM 6" (152 MM)

MAXIMUM 1/8" (3 MM) GAP

SHIELD SUPPORT STEEL

3" (76 MM) x 3" (76 MM)

1/3 PLATE THICKNESS

DETAIL B

(SEE FABRICATION NOTE 8)

(IGNORE FOR M36)

- TYPICAL STAGGERED LAYER DETAIL -

- TYPICAL CORNER DETAIL -

- PLATE DETAIL -

DETAIL A

DETAIL E

WIDE x 1/3 PLATE THICKNESS x LENGTH OF JOINT

LAPPING PLATE MINIMUM 3"

LONGER DIMENSION

WIDE x 1/3 PLATE THICKNESS X LENGTH OF JOINT

LAPPING PLATE MINIMUM

(SEE FABRICATION NOTE 8)

ANGLE x LENGTH OF JOINT

SHIELD PLATES

TYPICALLY 5 LAYERS TO

EQUAL 1/8" [3MM]

SHIELD PLATES

TYPICALLY 5 LAYERS TO

EQUAL 1/8" [3MM]

SHIELD PLATES

TYPICALLY 5 LAYERS TO

EQUAL 1/8" [3MM]

SHIELD PLATES

TYPICALLY 5 LAYERS TO

EQUAL 1/8" [3MM]

SHIELD PLATES

TYPICALLY 5 LAYERS TO

EQUAL 1/8" [3MM]

SHIELD PLATES

TYPICALLY 5 LAYERS TO

EQUAL 1/8" [3MM]

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A11 - Shield Requirements (2) 131:100

ISOMETRIC DRAWING [REFERENCE ONLY-NOT TO SCALE]

All surfaces are 1/8" M36

AutoCAD SHX Text

= MAGNETIC FLUX DIRECTION

AutoCAD SHX Text

BOLD LINES INDICATE LOCATIONS FOR REQUIRED LAPPING PLATES

AutoCAD SHX Text

DETAIL F

AutoCAD SHX Text

(SEE FABRICATION NOTE 7)

AutoCAD SHX Text

= GRAIN DIRECTION (CAN BE IGNORED FOR M36)

AutoCAD SHX Text

DETAIL E

AutoCAD SHX Text

DETAILS B, C OR D

AutoCAD SHX Text

DETAIL F

AutoCAD SHX Text

BORE

AutoCAD SHX Text

DIRECTION

FABRICATION

1. The design of the shield support system is the responsibility of the customer's architect, structural engineer, and shield vendor.

2. The magnetic shield design must be adhered to exactly. Any deviation from dimensional design, fabrication technique or material specification must be assessed and documented by GE MR Siting.

3. Building and shield support structures should be outside of the steel plates to preclude detrimental effects on homogeneity. If it is absolutely necessary to have support structure on the inside of the shield, a homogeneity analysis must be performed and documented by GE MR siting.

4. For all 1.5T and 3T MR systems, use of non-magnetic materials for concrete floor reinforcement is recommended, unless a homogeneity analysis has been performed and documented by GE MR Siting. For older 1.5T systems with the Magnishield option, signa 0.5T systems, MR Max systems and Vectra systems, conventional reinforcing and structural members are acceptable.

5. A secondary plate must be magnetically isolated from the primary shield. The use of non-magnetic structural members such as aluminum or non-magnetic stainless steel is acceptable.

6. If the magnetic shield is installed inside an RF shield, both the shield and its support system must be isolated from ground by 1,000 ohms or greater and meet all the GE RF room requirements as specified in the appropriate GE pre-installation manual.

7. The minimum finished floor to finished ceiling clearance in the magnet room is shown on sheet A2. Clearance dimensions do not include space required for the magnetic shield, RF shield, shield structure, and finished ceiling materials.

8. Locations of the penetration panel, window, and door openings are established on the final layout. Changes must be documented by GE MR Siting and Planning.

9. Penetration panel mounting details are specified in the appropriate GE site planning direction.

10. Consult GE MR Siting for guidance on any issues relating to magnetic shielding.

11. There must be a minimum of 8 inch [200mm] between the magnetic shield and the area where the magnetic field is being reduced. For example if a hallway has a 5 gauss restriction goal, the magnetic shield should be positioned 8 inch [200mm] inside the hallway surface. Larger buffer spaces are preferred but 8 inch [200mm] is the recommended minimum.

12. If the magnetic shield is supported by the parent room structure, the parent room structure should be wood/concrete/aluminum whenever possible. If the parent wall structure is steel, then the magnetic shield needs to be decoupled from the steel studs by a minimum of 5/8 inch [16mm] gypboard, plywood, furring, or other non-magnetic media.

13. If the magnetic shield is supported inside the RF structure, care must be taken to avoid loose metal to metal contacts or dissimilar metal contacts. Refer to the RF section of the pre-installation manual for all requirements. Magnetic decoupling media may also be needed if the magnetic shield is attached directly to an RF shield that is primarily steel.

1. Magnetic shield ceilings, floors, walls, etc. Will consist of multiple layers to make of the design's required thickness.

2. Slotted or oversized bolt holes are recommended to assure adequate play for positioning plates tightly together. Refer to detail a.

3. If multiple layers of steel are used, some lapping plates can be eliminated by staggering plate separations. Refer to detail B. This is the recommended method for multiple layers.

4. If layers are not staggered or a single layer of the required thickness is used, lapping plate must be used to cover plate separation. Lapping plate is required at each corner, plate separations perpendicular to the magnet bore's axis. This include all vertical seams in the sidewalls, all cross seams in floor and ceiling, all horizontal and vertical seams in the front and rear surfaces and all corners where a front or rear surface meet any other surface. Refer to details C, D, E, F, and shield isometric.

5. Gaps in the shield which are perpendicular to the flux path are critical and should be kept to a minimum. Maximum separation between plates is 1/8 inch (3mm).

6. Hole diameter of 8" [200mm] or less, when separated from another hole by a minimum of 3 times the diameter of the hole is allowed as needed. Holes for cryogen venting, HVAC, etc. Must be assessed and documented by GE MR Siting.

7. Angled joints do not need to have beveled mating edges. Refer to detail E.

8. When lapping plate material is required, it should extend at least 3" [76mm] to each side of the separation, and should have a total minimum thickness of 1/3 the thickness of the shield thickness.

9. Any separation between plates which exceed 1/8" [3mm] cannot be corrected by placing a lapping plate over the separation or using full seam welds across the separation. Corrections for this situation must be assessed and documented by GE MR Siting.

10. Lapping plate can be attached by tack welding or bolting.

11. Shield plates are to be free of loose scale. Failure to remove scale can result in loose scale and steel particles becoming lodged in the magnet.

The preferred method for removal of scale is sandblasting. Removal of scale using wire brush is permitted. Shot blasting is prohibited.

12. Shield plates must be rigidly supported to prevent movement resulting from air pressure or other environmentally induced changes which can alter the magnet's homogeneity or systems performance.

13. Shield plate must be well anchored to the support structure by plug welding or bolting.

14. Full seam welding is not recommended for attaching shield plates to support structure or adjacent shield plates.

15. Minimum penetration full seam welds are allowed when the magnetic shield also serves as the RF shield.

16. Full seam welding is not allowed on silicon steel sheets.

17. Minimize the need for flame cutting of steel plate after annealing, as material's magnetic properties can be altered as a result of stress induced by this process.

18. Loose ferrous components become dangerous projectiles when the magnet is energized.

The actual performance of the magnetic shield depends upon strict adherence to the following requirements.

SHIELD NOTES

MAGNETIC SHIELD CONSTRUCTION REQUIREMENTS

GENERAL

1.0

FIGURE 1

0.1

IN

DU

CT

IO

N

KI

LO

GA

U

SS

ES

0.5 1.0

MAGNETIZING FORCE - OERSTEDS

5.0 10 50 100 500 0.6

IN

DU

CT

IO

N

TE

SL

A 0,8

1.2

1.4

1.6

1.8

ARMCO DI-MAX M-36 CR FP

D-C Magnetization curve at high induction Test: AS; 50/50; A596

Excessive stresses may be produced in electrical steel by improper arrangement of coils in storage, as indicated.

Impproper handling of strip, sheets or long laminations can introduce stress that impair magnetic properties of the steel.

Coil set introduced in handling strip or long laminations.Unintentional introduction of excessive stress in steel during storage of coils.

FIGURE 2

ARMCO DI-MAX M36 (or material with equivalent magnetic properties) CR FP BARE (non coated) non-oriented electrical steel (2.25% maximum silicon content, 0.004% maximum carbon content, 0.65% maximum aluminum content).

· Material must meet all requirements of ASTM A677 for fully-processed non-oriented electrical steel.

· Material must meet or exceed magnetic properties as illustrated in figure 1 when tested per ASTM A596 as sheared with laminations in test specimen half parallel, half transverse to the rolling direction.

· Material must not be handled with electromagnetic equipment.

· Each sheet of material must lie flat under it's own weight.

· Material supplier must provide manufacturers certification with shipment.

NOTES:

· Shield thickness is not to scale.

· Magnetic properties of electrical steels are especially sensitive to stress. substantial reductions in magnetic properties can be caused by only a few hundred pounds per square inch. therefore, prevention of stresses due to elastic or plastic deformation during handling is critical.

· Refer to figure 2 for some examples of situations to avoid.

· Each 1/8" (3 mm) shield thickness consists of 5 layers of 24 gauge (0.0250 inch [0.64mm]) M36 steel.

(SILICON-011310)

SILICON STEEL MATERIAL REQUIREMENTS

Radius may be too small

Corner may be too sharp

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center A12 - Shield Requirements (3) 141:100

VIBRATION TRANSMITTED THROUGH VIBROACOUSTIC MAT

.5 16 20 25

.5 40 50 63 80 10

0.025

Pe ak

A pl itu de

En ev el op e [m

/s ²]

1/3 Octave Frequency [Hz]

0.05

0.075

0.1

0.125

0.15

0.175

0.2

0.225

0.25

EXCITATION FREQUENCY [Hz]

AC

CE

LE

RA

TI

O

N g

's (1

0¯ ⁶)

Ab ov e am bi en t b as el in e

MAGNET STEADY-STATE VIBRATION SPECIFICATIONS

VIBRATION SPECIFICATIONS

Excessive vibration can affect MR image quality. Vibration testing must be performed early in the site planning process to ensure vibration is minimized. Both steady state vibration (exhaust fans, air conditioners, pumps, etc.)

and transient vibrations (traffic, pedestrians, door slamming, etc.) must be assessed.The Magnet cannot be directly isolated from vibration. Any vibration issue must be resolved at the source.

Transient vibration levels above the specified limits in the MR Site Vibration Test Guidelines must be analyzed. Any transient vibration that causes vibration to exceed the steady-state level must be mitigated.

10 20 30 40 50

5 15 25 35 45

· All units that are wall mounted or wall supported are to be provided with supports where necessary. Wall supports are to be supplied and installed by the customer or his contractors.

· Dimensions are to finished surfaces of room.

· Certain MR procedures require an extremely stable environment to achieve high resolution image quality.

Vibration is known to introduce field instabilities into the imaging system. The vibration effects on image quality can be minimized during the initial site planning of the mr suite by minimizing the vibration environment. See PROXIMITY LIMITS, PATIENT TABLE DOCK ANCHOR MOUNTING REQUIREMENTS AND VIBROACOUSTIC DAMPENING KIT details for additional information.

· Standard steel studs, nails, screws, conduit, piping, drains and other hardware are acceptable if properly secured. Any loose steel objects can be violently accelerated into the bore of the magnet. Careful thought should be given to the selection of light fixtures, cabinets, wall decorations, etc. To minimize this potential hazard. For safety, all removable items within the magnet room such as faucet handles, drain covers, switch box cover plates, light fixture components, mounting screws, etc. must be non-magnetic. If you have a specific question about material, bring it to the attention of your GE project manager of installations.

· Floor levelness refer to MAGNET ROOM FLOOR SPECIFICATIONS DETAIL, this floor levelness requirement is important for accurate patient table docking.

· Non-movable steel such as wall studs or hvac components will produce negligible effect on the active shield magnet.

· Customer's contractor must provide all penetrations in post tension floors.

· Customer's contractor must provide and install any non-standard anchoring. Documents for standard anchoring methods are included with GE equipment drawings for geographic areas that require such documentation.

· Customer's contractor must provide and install hardware for "through the floor" anchoring and/or any bracing under access floors. This contractor must also provide floor drilling that cannot be completed because of an obstruction encountered while drilling by the GE installer such as rebar etc.

· Customer's contractor to provide and install appropriate supports for the storage of excess cables.

· It is the customer's responsibility to perform any floor or wall penetrations that may be required. The customer is also responsible for ensuring that no subsurface utilities (e.g., electrical or any other form of wiring, conduits, piping, duct work or structural supports (i.e. post tension cables or rebar)) will interfere or come in contact with subsurface penetration operations (e.g. drilling and installation of anchors/screws) performed during the installation process. To ensure worker safety, GE installers will perform surface penetration operations only after the customer's validation and completion of the "GE surface penetration permit"

STRUCTURAL NOTES

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center S1 - Structural Notes 15

ITEM DESCRIPTION

(GE SUPPLIED / CONTRACTOR INSTALLED)

1 Vibroacoustic dampening kit (see floor structural detail)

2 Magnet curtain kit

(CONTRACTOR SUPPLIED & INSTALLED)

3 Patient table dock rebar free area 4 Structural wall backing for Main Disconnect Panel

5 Structural wall backing for Magnet Rundown Unit

6 Structural wall backing for Main Bypass Panel

7 Two I-beam runners are to support chiller feet at both ends. Refer to Dimplex manual.

8 Structural floor support for UPS Battery

9 Floor mounting point for full ups

DateRev /29MRI-M234700-FIN-02-C.DWG C 24/Feb/2023SIGNA ARTISTMichael E Debakey VA Medical Center S2 - Structural Layout 161/4"=1'-0"

MAGNET ROOM

EQUIPMENT ROOM

CONTROL ROOM

LOCATED ELSEWHERE

1 1

1 1

LOCATED ELSEWHERE

6'

-5

4' -1

0"

6'-6"

'-0

CABLE CONCEALMENT

SCALE 1:25

[3 7.

in

[1 4.

in

[5.3 in]

[47.1 in]

Vent pipe

Magnet isocenter

CABLE

TRAYS

FR

O

N T

O F

M

AG

N

ET

MAGNET ROOM FLOOR SPECIFICATIONS

Floor levelness must be 3 mm between high and low spots in the rectangular area shown.

The finished floor must support the weight of all components (e.g., patient table, gradient coil replacement cart) throughout operation and service life.

[9 2.

in

[245 in]

[103.3 in]

PATIENT TABLE DOCK ANCHOR MOUNTING REQUIREMENTS

· The RF Shield vendor must design and install the dock anchor bolt

· The dock anchor hole must be drilled after the Magnet is installed

· The dock anchor must…

This is the start of the file's text. The full file is on GovTribe.