Z2DA--542-22-133 | IV Cleanroom Replacement Coatesville VA Medical Center

Closed Solicitation Posted

This opportunity was awarded. See the award notice from .

Solicitation number
36C24424B0001
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

Notice details come from SAM.gov. Updated .

About this opportunity

This solicitation seeks proposals to replace an intravenous (IV) cleanroom at the Coatesville VA Medical Center in Coatesville, Pennsylvania. The Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 4 will award a firm-fixed price contract to design and construct a new modular cleanroom envelope, install a new air handling unit and louver, and complete associated demolition and facility modifications. Proposals are due January 3, 2024 and will be evaluated on technical qualifications, past performance, and price. The contract is to be awarded by the end of 2023 with project completion within 300 calendar days of the notice to proceed.

This procurement is set aside for Service-Disabled Veteran-Owned Small Businesses. Offerors must be verified through the SBA and registered in SAM.gov. The NAICS code for the design-build construction is 236220 and PSC code R499 covers the facility replacement. The construction value is estimated between $2-5 million. Work will be performed at the Coatesville VA Medical Center located in Coatesville, Pennsylvania and is to be completed within 300 calendar days of notice to proceed.

Notice text

2 versions

Update #2 · Latest ·

IV Cleanroom

RFI Amendment #1

010000 paragraph 1.11.D Is this requirement applicable to this project? If it is, are we using the definition in 013323 paragraph 1.2.B for shop drawings?

Yes to both questions.

010000 paragraph 1.12 Is a warranty management plan submittal necessary in addition to the warranties provided in a closeout binder at the end of the project?

Yes

010000 paragraphs 1.22 & 1.23 Will all general contractors be held to these very specific photographic documentation specifications?

Yes, as applicable to all work areas and disciplines utilized in the accomplishment of this project.

013323 paragraph 1.3.C Can a submittal register be provided for this project prior to the bid date?

Yes.

013323 paragraph 1.8.A Some specifications very clearly denote when O&M documents are required, and others do not. Do all specification sections require O&M documentation?

Yes, All specification sections will need a designated set of O&M documentation

013526 paragraph 1.7.C Can the SSHO for the general contractor also be the superintendent?Â

No.

013526 paragraph 1.8.B Is the SSHO considered a CP and required to have OSHA 30 training within the past 5 years?

Yes.

013526 paragraph 1.8.D Is there a 5-year expiration for OSHA 10 cards?

No. Per OSHA - OSHA 10 Cards do not expire. However, It is the GC's responsibility to ensure all other workers on site are competent and knowledgeable of OSHA 10 Requirements.

014500 paragraph 1.3.A.2 & 3 Are these submittals required

Yes.

015719 paragraph 1.4.A.1 Will all general contractors be held to the standards of this Environmental Protection Plan specification?

Yes.

017419 paragraph 1.5.B Will all general contractors be held to the standards of this Demolition Debris Management Plan?

Yes.

079200 paragraph 1.4.E Are cured samples applicable?

Yes.

092900 paragraph 1.4.D Are samples applicable?

Yes.

099100 paragraph 1.3.D Are sample panels applicable?

Yes.

260511 paragraph 1.12.F Does division 26 require a separately submitted closeout binder, or can it be part of the entire closeout binder for the whole project?

Yes, Maintenance and operations Manuals require their own binder.

Regarding all Commissioning, Specification Sections 01.91.00 and 07.08.00, please confirm, the VA will engage the Commissioning Agent under a separate contract.

The construction contractor shall provide an independent commissioning agent as part of the contract.

Are the wood doors rotary cut red oak?

All of the corridor doors in the older buildings are red oak. Offerors should bid the project with this material for wood doors.

For hollow metal door 211Y, what is the size of the terminated stops?

6"

What type of electric lock is being used on the wood doors?

Electric strike

What type of lite kit does the metal door get?

Half Lite

Any off-shift requirements for this Project?

Yes. While normal working hours is the intent, the contractor should anticipate the possibility of work required off-shift. These off-shift activities hours shall be indicated in the construction schedule by the contractor and approved by the COR. Any unscheduled or unforeseen construction activities requiring off-shift work shall be requested at least two weeks in advance of the off-shift requirement and approved by the COR prior to commencing work. Off-shift work may include, but not be limited to, safety or comfort of building occupants during any construction activity (examples: demolition work or deliveries of large equipment/materials or placing large equipment/material in-place or hauling or cutting/patching or welding or drilling) which is deemed unsafe or excessively loud or any other construction activity (inside or outside) that disrupts on-going building operations deemed inappropriate by the COR and/or the contractor. The contractor shall track all off-shift hours separately and report these to the COR on a weekly basis. Any unforeseen or unscheduled construction activity requiring off-shift work shall not be considered a delay to the contractor caused by the government and no claims will be consider for any additional off-shift work that may be required.Â

Is the cleanroom ceiling to be panelized vs. a drop grid?

If yes, then does it need to be walkable for access to terminal filters and lights?

In either case, what is the height of the structure to suspend the ceiling from? And from the drawings I believe the cleanroom ceiling is to be 9' AFF - please confirm

In the case of a walkable ceiling, is there access to the top of the ceiling after construction from elsewhere, or do we need to provide access panels in the ceiling, to enter through the cleanroom?

a. The cleanroom ceiling surrounding the Ante Room - ISO CLASS 8 211 and IV ROOM - ISO CLASS 7 210 rooms is specified to be a gasketed drop grid.

b. Height to existing structure above is 11'-8 1/8"". 8'-0"" APC-1 ceilings specified on A112 D6 of the Construction Documents.

c. Not a walkable ceiling surrounding the Ante Room - ISO CLASS 8 211 and IV ROOM - ISO CLASS 7 210 .

The spec calls out Portafab 300 Series, which is anodized aluminum studs and that is what the pics of the existing room shows. However, it is available in white powder-coated - which makes it all white - advise if we should include.

All white is acceptable.

Please provide the Acoustical Ceiling Specifications.

See Attached.

Please confirm the Building and Room to where the Bids shall be delivered.

Bids should be mailed/delivered to Coatesville VAMC, Building 70. The bid opening will take place at 10:00 AM in the conference room of Building 70.

Can you confirm that this is a hard bid? We ve been conditioned by bidding mostly in Altoona, Pittsburgh, Lebanon and NY that most bids are delivered via email, so I wanted to verify.

This will be a hard bid. Bids should be delivered in person or mailed to the address in block 8 on page 1 of the contract by 10:00 AM on 1/3/2024. Bids received after this time will not be accepted.

Are there any special directions concerning what you d like submitted with the bid documents? For instance, should the price be put in box 17 on the SF1442 or should it be placed in the Price/Cost Schedule on page 6 of the solicitation? Should all 74 pages of the solicitation be submitted, or should we stick to the SF1442, perhaps the Price/Cost Schedule and the necessary VAAR and FAR Regulations that require signature? Page 19 indicates that FAR 52.204-7 is incorporated so paragraph d of FAR 52.204-8 on page 22 applies, and we have completed the reps and certs on SAM. I do see that page 7 requires an EMR letter.

The bid should be submitted in Block 17 on page 2 of the SF1442. The following documents shall be provided in physical form at the time of bid opening:

a. SF1442 with information completed by offeror on page 2.

b. EMR Rating

c. VAAR 852.219-75 (Page 42-44) completed by offeror.

d. FAR 52.209-7 (Page 26-27) paragraph (b) filled out by offeror.

e. Amendment 0001 acknowledgment

Further documentation identified in the solicitation may be requested prior to award

Update #1 ·

?.## STATEMENT OF WORK

Statement of Work

Project # 542-22-133

IV Cleanroom Replacement in Building 2

This project constructs a new IV Cleanroom in Building 2, on the second floor to replace the existing cleanroom in accordance with the drawings and specifications. The new IV Cleanroom envelope will be of modular construction based upon Portafab company as name brand or equal to Series 300.  The construction includes the installation of a new air handling unit in the attic mechanical room together with the provision of a new louver in the attic. The project also includes all associated plumbing, fire protection, lighting and electrical, all controls, slab penetrations and modifications, and demolition. The work includes all labor, equipment, and materials. Contractor shall be required to provide the Brand Name or Equal products which must be compatible with existing systems throughout the Coatesville VAMC. All name brand products mentioned in this SOW are used as a basis of design, for fair bidding. They may be replaced with other products upon submittal approval.

The following list of brand name or equal to items would be used as a basis for design in the construction of this project:

Portafab Series 300

Krueger 5100HF

CS-1 and CS-1a Bell & Gossett Model e614

CS-2 Bell & Gossett Model e613.

Spirax Sarco Model CSG-HSEF020-PNP3C1-1F2F-AET-3SPLN17-AESL, Capacity 50 lbs./hr., 15 psig

Greenheck Model BSQ-90

Type A: CLEAR -VU LIGHTING #CLF14-S1-N-UV-HAL-LFSS304-CIA-IBBMS

Type B: COLUMBIA LIGHTING # MDS-4-40-VW-F-W-E-U

RING SS201 - CLEAN/PROCEDURE ROOM STATION

RING AA918 DESK/WALL MASTER STATION

RING RM5000EX INTERCOM CONTROL SYSTEM

All work shall be performed in compliance with all Federal, State and Local laws and regulations. In addition, work will meet equipment s manufacturer specifications and generally accepted industry standards of good workmanship. All waste materials shall be disposed of by the contractor in compliance with Federal State and Local laws and regulations. Contractors will supply receipts for quantities of material removed for recycling accountability.

No work may proceed by the contractor until all material cut sheets are submitted to the COR for approval.

A site visit will be conducted with perspective offerors to visually assess existing conditions.

Contractor will have 300 days after receipt of NTP to complete the project.

Files

File Type Posted
S06 - Ceiling Specs.pdf PDF
S04 - Site Visit Sign-In.pdf PDF
36C24424B0001 0001.docx DOCX document
VA IV Cleanroom Issue 4 Bid Drawings R2.pdf PDF
36C24424B0001_2.docx DOCX document
Specifications 100_ Design R4 100323.pdf PDF
AHU Design Submittal - Coatesville VAMC 9-26-2023.pdf PDF

Notice history

Notice Type Posted
Z2DA--542-22-133 | Construction IV Cleanroom Replacement Award Award Notice
This notice Latest solicitation Solicitation
Z2DA--542-22-133 | IV Cleanroom Replacement Coatesville VA Medical Center Original Pre-Solicitation

Similar open opportunities

Opportunity Agency Set-aside Responses due
Z1DA--542-21-113 | Correct FCA Deficiencies – Repair Water Penetration | NCO 4 Construction East (VA-26-00020634) (VA-26-00020634) Coatesville VA Medical Center 1400 Black Horse Hill Road 19320 · NAICS 238190 Veterans Integrated Service Network 4 Service-Disabled Veteran-Owned Small Business
Z1DA--528A8-22-814 Consolidate Primary Care Reception Albany VA Medical Center Albany, New York 12208, United States · NAICS 236220 Veterans Integrated Service Network 2 Service-Disabled Veteran-Owned Small Business
Z2DA--657-24-102JC - Prepare Site for New PET/CT, Room A011D 915 N Grand Blvd, St. Louis, MO 63106 · NAICS 236220 Veterans Integrated Service Network 15 Service-Disabled Veteran-Owned Small Business
540-22-700 | EHRM Infrastructure Upgrades Construction | Clarksburg WV Clarksburg, West Virginia 26301, United States · NAICS 236220 VA Technology Acquisition Center Service-Disabled Veteran-Owned Small Business
Z1DA--36C25227B0026 Remodel ENT Procedure Rooms CON 537-19-109 Jesse Brown VA Medical Center 820 South Damen Avenue Chicago, Illinois 60612, United States · NAICS 236220 Veterans Integrated Service Network 12 Service-Disabled Veteran-Owned Small Business

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors