Scientific Automated -80°C Ultra-Low Temperature (ULT) Freezer System

Closed Solicitation Posted

Solicitation number
RFQ-NHLBI-03869
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

Place of performance
Bethesda, Maryland 20892, United States

Notice details come from SAM.gov. Updated .

About this opportunity

The National Institutes of Health National Heart, Lung, and Blood Institute seeks quotations for a scientific automated -80°C ultra-low temperature freezer system, with proposals due July 8, 2026. This is a commercial item procurement issued as a combined synopsis and solicitation under FAR Subpart 12.6, with no set-aside applied.

Place of performance is Bethesda, Maryland 20892. The freezer must maintain storage temperatures at -80°C ± 3°C with enhanced stability of ± 1.5°C and temperature homogeneity control. The facility has structural ceiling constraints: two beams at approximately 11 feet 7 inches height create a working clearance of 11 feet 6 inches, requiring vendor accommodation despite the stated 12-foot ceiling height in technical requirements. Fire code compliance requires a fire suppression system with sprinkler penetrations in the freezer ceiling or sidewalls per detailed engineering drawings. The system must be capable of picking samples at both -20°C and -80°C temperatures. The NHLBI Biospecimen Core currently manages approximately 35,000 biospecimens linked to 7,700 participant visits annually across 60 supported laboratories and oversees 300 manual freezers. Projected return on investment analysis forecasts approximately $1 million in annual savings by Year 2 and full cost recovery within five years, primarily through reductions in labor (88%), energy (67%), and floor space (66%).

Quotations must be all-inclusive using DDP Prepay terms with all freight and import fees incorporated in the quote submitted by the deadline; the government's system does not permit post-award cost adjustments. Vendors should provide a comprehensive itemized price list of tray options and estimate costs for filling approximately 80 percent of the freezer capacity. The NAICS code is 334516, Analytical Laboratory Instrument Manufacturing. The PSC is 6515, Medical and Surgical Instruments, Equipment, and Supplies.

Notice text

2 versions

Update #2 · Latest ·

See attached: Fire suppression system drawings, Ceiling clearances drawings and Answers to Questions Part 1 & Part 2.

Update #1 ·

Please see attached doc.

Files

File Type Posted
C201217 Penetrations Plan.pdf PDF
NHLBI Site Ceiling Clearances.pdf PDF
RFQ-NHBLI-03869 - ANSWERS - 1 2 6 - PART 1.docx DOCX document
RFQ-NHLBI-03869 - PART 2.docx DOCX document
RFQ-NHLBI-03869.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
AN11--501 - Dry Ice - Base 4 Department of Veterans Affairs New Mexico VA Health Care System Albuquerque, New Mexico 87108, United States · NAICS 325120 Veterans Integrated Service Network 22 No set-aside
Remove and Replace Walkin Cooler and Freezer NAICS 333415 — Total Small Business
36C25227Q0048 - coolers and freezer North Chicago, Illinois 60064, United States · NAICS 333415 Veterans Integrated Service Network 12 Service-Disabled Veteran-Owned Small Business Sole Source
NCSS - Bioruptor Pico Sonication Device NAICS 334516 Office of Logisitics and Acquisition Operations No set-aside
Preventative Maintenance and Service Support for Thermo Fisher Scientific HERA VIOS 160i CO2 and VIOS 250i CO2 Incubators Bethesda, Maryland 20892, United States · NAICS 811210 Office of Logisitics and Acquisition Operations No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors