This is a Presolicitation Notice for preventative maintenance and service of Philips imaging equipment at the Evansville Health Care Center. The Department of Veterans Affairs Network Contracting Office is seeking a contractor to provide comprehensive repair, parts replacement, software updates, and scheduled preventive maintenance services for radiology and ultrasound equipment. The solicitation number is 36C25527Q0053, with a response deadline of December 2, 2026, at 8:00 AM Central Time. The contracting office is located in Leavenworth, Kansas (zip code 66048), with the place of performance at Evansville Health Care Center in Evansville, Indiana. The Product Service Code is J065 (Maintenance and Repair Services), and the NAICS code is 811210 (Electrical and Electronic Equipment Repair and Maintenance).

The contract requires the contractor to provide repair services during normal VA business hours (8:00 AM to 4:30 PM, Monday-Friday, excluding Federal Holidays) with a 30-minute telephone response time and 4-hour on-site arrival commitment. Equipment uptime guarantees are set at 97 percent or greater for radiology equipment and 95 percent or greater for ultrasound equipment. The contractor must supply all necessary replacement parts (including tubes, probes, detectors, grids, tables, and batteries) at no additional cost and deliver replaced parts on-site by 10:30 AM the following day. All operational and technical documentation, software updates, preventive maintenance procedures, and compliance with NFPA-99, OSHA, and CDRH standards must be provided by the contractor. The Point of Contact is Contract Specialist Tracie Raggs at tracie.raggs@va.gov.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Presolicitation Notice Presolicitation Notice

SUBJECT*
Imaging Service & Maintenance

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
66048
SOLICITATION NUMBER*
36C25527Q0053
RESPONSE DATE/TIME/ZONE
12-02-2026 08:00 CENTRAL TIME, CHICAGO, USA
ARCHIVE
10 DAYS AFTER THE RESPONSE DATE

SET-ASIDE

PRODUCT SERVICE CODE*
J065
NAICS CODE*
811210
PLACE OF PERFORMANCE
Evansville Health Care Center

6211 East Waterford Blvd.

Evansville IN

POSTAL CODE
47715
COUNTRY
USA

CONTACT INFORMATION

CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

Network Contracting Office (NCO) 15 3450 S 4th Street Trafficway

Leavenworth KS 66048

POINT OF CONTACT*

Contract Specialist Tracie Raggs tracie.raggs@va.gov

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS
tracie.raggs@va.gov

EMAIL DESCRIPTION

DESCRIPTION

This is for preventative maintenance and service of Philips Imaging equipment. See the solicitation that follows for full details and requirements.

I. SERVICES TO BE PROVIDED:

A. Repairs

1. Service Hours:

All repairs will be performed during normal VA business hours (8:00 AM – 4:30 PM, Monday-Friday), excluding Federal Holidays unless otherwise specified. Contractor may work outside normal business hours by arrangement with the COR if such services are provided without additional charge to the government. Any overtime charges must be approved by the COR or designee prior to the initiation of overtime work.

2. Response Time:

A technical support representative will respond by telephone to the VA within thirty (30) minutes of the original service call. The contractor will arrive on-site for repair within four (4) hours of the original service call. Radiology equipment shall have a guaranteed uptime of 97% or greater. Ultrasound Equipment shall have an uptime of 95% or greater, ensuring minimal downtime for critical diagnostic operations.

3. Documentation and Resources:

The VA shall not provide service manuals or service diagnostic software to the contractor. The contractor shall obtain, have on file, and make available to its FSE(s) all operational and technical documentation (e.g., operational and service manuals, schematics, parts lists) necessary to meet the performance requirements of this contract. The location and listing of the service data manuals, by name, and/or the manuals shall be provided to the CO upon request. Any charges for parts, manuals, tools, or necessary software for maintenance/repairs are included within this contract and its agreed-upon price unless otherwise stated in writing.

B. Parts

1. Replacement Parts:

The contractor shall provide all necessary replacement parts to ensure the equipment performs within the manufacturer's specifications. This includes, but is not limited to, coverage for tubes, ultrasound probes, detectors, grids, tables, and batteries. All replacement parts must meet or exceed the factory specifications to maintain compatibility with existing systems and support future performance and reliability upgrades. Any parts removed and replaced by the contractor will become the property of the contractor. Replaced parts must be delivered on-site no later than 10:30 AM the following day.

C. Software Updates/Upgrades

1. Software Maintenance:

The contractor will furnish, free of charge, all OEM software or hardware upgrades to the equipment purchased by the VA, which are intended to correct safety risks. Software updates offering enhancements to previously purchased software features shall also be covered under this service agreement if they do not require hardware modifications or additions.

D. Scheduled Maintenance

1. Preventive Maintenance (PM):

The contract will include comprehensive scheduled preventive maintenance inspections (PM) as required by manufacturer specifications.

2. Scheduling:

A mutually agreeable time for the inspections will be scheduled through the COR, Healthcare Technology Management (HTM) staff, or the Radiology supervisor/CT Lead.

3. Inspection and Maintenance Procedures:

The contractor will perform PM Services to ensure that equipment listed above functions in conformance with the latest published editions of NFPA-99, OSHA, and CDRH standards. The contractor shall provide and utilize procedures and checklists with worksheet originals indicating work performed and actual values obtained (as applicable) provided to the COR or HTM staff upon completion of the PM. Preventive maintenance procedures will be submitted to the COR for approval prior to initiation of this service and kept on file. PM services shall include, but are not limited to:

· Cleaning of equipment

· Reviewing operating system diagnostics to ensure system operation meets manufacturer's specifications

· Calibrating and lubricating the equipment

· Performing non-emergent remedial maintenance

· Testing and replacing faulty or worn parts and/or parts likely to become faulty, fail, or become worn

· Inspecting and replacing necessary cables and bushings

· Measuring, adjusting, and calibrating for optimal image quality

· Inspecting all mechanical components including, but not limited to, probes, cables, mounting hardware, chains, belts, bearings, tracks, interlocks, clutches, motors, monitors, keyboards for mechanical integrity, safety, performance

· Returning equipment to full and proper operating condition per manufacturer specifications

· Conducting an electrical safety inspection (as applicable) and documenting the results.

4. Compliance:

The contractor, their personnel, and their subcontractors shall be subject to Federal laws, regulations, standards, and VA Directives and Handbooks regarding information and information system security as delineated in this contract.